ionisation techniques

41
Ionization Techniques In mass Spectroscopy

Upload: malla-reddy-college-of-pharmacy

Post on 27-May-2015

74 views

Category:

Health & Medicine


2 download

DESCRIPTION

Ionisation techniques

TRANSCRIPT

Page 1: Ionisation techniques

Ionization Techniques In mass Spectroscopy

Page 2: Ionisation techniques

components of mass spectrometer includes

Ion SourceMass Analyzer

Ion collection System or Ion detector.

Data Handling System

Vacuum System

Inlet System

Page 3: Ionisation techniques

ION SOURCE

Liquids and solids are first converted in to gases from the gaseous sample, ions are produced in a Box like enclosure called Ion Source.

Function • Produces ion without mass discrimination of the sample.• Accelerates ions into the mass analyzer.

Classification of ion sources

Gas Phase Sources. • Electron Impact Ionization (EI).• Chemical Ionization (CI).• Field Ionizations (FI).

Page 4: Ionisation techniques

• Plasma desorption (PD).• Fast Atom Bombardment (FAB).• Thermospray Ionization (TS).• Secondary Ion Mass Spectrometry (SIMS).

Desorption Sources.

• Field Desorption (FD).• Electrospray Ionization (ESI).• Matrix assisted desorption/Ionisation (MALDI).

Page 5: Ionisation techniques

Gas Phase Ionization Methods

Electron Impact Ionization

Page 6: Ionisation techniques

It is the most widely used and highly developed method. It is also known as Electron bombardment or Electron Ionization.

Electron impact ionization source consists of a ionizing chamber which is maintained at a pressure of 0.005 torr and temperature of 200 ± 0.25 degrees.

Electron gun is located perpendicular to chamber.

Electrons are emitted from a glowing filament (tungsten or rhenium) by thermionic emission and accelerated by a potential of 70 V applied between the filament and anode.

These electrons are drawn in the ionization chamber through positively charged slits.

The number of electrons is controlled by filament temperature and energy .

The sample is brought to a temperature high enough to produce molecular vapors.

Page 7: Ionisation techniques

The gaseous Neutral molecules then pass through the molecular leaks and enter the ionization chamber.

The gaseous sample and the electrons collide at right angles in the chamber and ions are formed by exchange of energy during these collisions between electron beam and sample molecules.

Page 8: Ionisation techniques

M Analyte molecule e- Electrons

M.+ Molecular ions

The positive ions formed in the chamber are drawn out by a small potential difference (usually 5eV) between the large repeller plate (positively charged) and first accelerating plate (negatively charged).

Strong electrostatic field (400 – 4000 V) applied between the first and second accelerating plates accelerates the ions according to their masses (m1, m2, m3 etc) to their final velocities.

The ions emerge from the final accelerating slit as a collimated ribbon of ions. The energy and velocity of ions are given by :-

zV = ½ (m1v1) = ½ (m2v2) = ½ (m3v3) z = charge of the ionV = accelerating potential v = velocity of ion

Page 9: Ionisation techniques

Advantages

• Gives molecular mass and also the fragmentation pattern of the sample.• Extensive fragmentation and consequent large number of peaks gives structural information.• Gives reproducible mass spectra.• Can be used as GC/MS interface.

Disadvantages

• Sample must be thermally stable and volatile.• A small amount of sample is ionized (1 in 1000 molecules).• Unstable molecular ion fragments are formed so readily that are absent from mass spectrum.

Page 10: Ionisation techniques

Chemical Ionization

In chemical ionization the ionization of the analyte is achieved by interaction of it’s molecules with ions of a reagent gas in the chamber or source.

Chemical ionization is carried out in an instrument similar to electron impact ion source with some modifications such as:-

• Addition of a vacuum pump.

• Narrowing of exit slit to mass analyzer to maintain reagent gas pressure of about 1 torr in the ionization chamber.

• Providing a gas inlet.

Page 11: Ionisation techniques
Page 12: Ionisation techniques

It is a two part process.

Step-I Reagent gas is ionized by Electron Impact ionization in the source. The primary ions of reagent gas react with additional gas to produce stabilized reagent ions.

step-II Reagent ions interact with sample molecules to form molecular ions.

In this technique the sample is diluted with a large excess of reagent gas.

Gasses commonly used as reagent are low molecular weight compounds

such as Methane, tertiary Isobutane, Ammonia, Nitrous oxide, oxygen and

hydrogen etc.

Depending upon the type of ions formed CI is categorized as:-

• Positive Chemical Ionization.

• Negative Chemical Ionization.

Page 13: Ionisation techniques

1. Positive Chemical Ionization

In this technique positive ions of the sample are produced. In positive chemical ionization gasses such as Methane, Ammonia, Isobutane etc are used

For example Ammonia is used as reagent gas. First ammonia radical cations are generated by electron impact and this react with neutral ammonia to form ammonium cation (reactive species of ammonia CI).

NH3 NH3.+ + 2 e-

NH3.+ NH4

+ + NH2

NH4+ reacts with the sample molecules by proton transfer or

Adduct formation to produce sample ions

M + NH4+ [M + H]+ + NH3 Proton

transfer

M + NH4+ [M + NH4]+ Adduct

formation

e-

Page 14: Ionisation techniques

When Methane is used as Reagent gas. Methane is ionized by electron impact:

CH4 + e- CH4+ + 2e-

Primary ions react with additional reagent gas molecules to produce stabilized reagent ions:

CH4+ + CH4 CH5+ + CH3

CH3+ + CH4 C2H5+ + H2

The reagent ions then react with the sample molecules to ionize the sample molecules:

CH5+ + MH CH4 + MH2+ (Proton transfer)

CH3+ + MH CH4 + M+ (hydride abstraction)

CH4+ + MH CH4 + MH+ (Charge transfer)

Page 15: Ionisation techniques

Negative chemical ionization is counterpart of Positive chemical ionization.

Negative ions of the sample are formed and oxygen and Hydrogen are used

as reagent gasses.

This method is used for ionization of highly electronegative samples.

The negative ions are formed by following reactions :-

Resonance electron capture M + e- M-

Dissociative electron capture

RCl + e- R + Cl-

H2O + e- H + OH-

The ion molecule reaction occurring between negative ion formed in the

chamber source and the sample molecule include:-Charge transfer. Hydride transfer.Anion- Molecule adduct formation.

2. Negative Chemical Ionization

Page 16: Ionisation techniques

Advantages

• Used for high molecular weight compounds.• Used for samples which undergo rapid fragmentation in EI.

Limitations

• Not suitable for thermally unstable and non-volatile samples.• Relative less sensitive then EI ionization.• Samples must be diluted with large excess of reagent gas to prevent primary interaction between the electrons and sample molecules

Page 17: Ionisation techniques

Field Ionization

FI is used to produce ions from volatile compounds that do not give molecular ions by EI.

It produces molecular ions with little or no fragmentation.

Page 18: Ionisation techniques

Application of very strong electric field induces emission of electrons.

Sample molecules in vapour phase is brought between two closely spaced electrodes in the presence of high electric field (107 - 108 V/cm) it experiences electrostatic force.

If the metal surface (anode) has proper geometry (a sharp tip, cluster of tips or a thin wire ) and is under vacuum (10-6 torr) this force is sufficient to remove electrons from the sample molecule without imparting much excess energy.

The electric field is produced by applying high voltage ( 20 KV ) to these specially formed emitters ( made up of thin tungsten wire ).

Page 19: Ionisation techniques

In order to achieve high potential gradients necessary to effect ionization, the anode is activated by growing carbon microneedles or whiskers.

These whiskers are 10 micro meters in length and greater than 1µm in diameters. These whiskers are capable of removing valence electrons from the organic molecules by quantum mechanical tunneling mechanism.

As concentration of sample molecules is high at the anode ion-molecule reactions often occur which results in formation of protonated species (M+H)+. Thus both M+ and (M+H) + is observed in FI spectrum.

Page 20: Ionisation techniques

These cations are accelerated out of the source and their mass is analyzed by analyzer .

Advantages• As fragmentation is less, abundance of molecular ions (M+) is enhanced, hence this method is useful for relative molecular mass and empirical formula determination.

Disadvantages • Not suitable for thermally unstable and non volatile samples.• Sensitivity is les than EI ion source.• No structural information is produced as very little fragmentation occurs

Page 21: Ionisation techniques

In field desorption method a multitipped emitter (made up of tungsten wire with carbon or silicon whiskers grown on its surface).

The electrode is mounted on a probe that can be removed from the sample compartment and coated with the solution of the sample.

The sample solution is deposited on the tip of the emitter whiskers either by dipping the emitter into analyte solution or by using a microsyringe.

The probe is then reinserted into the sample compartment which is similar to CI or EI unit.

Then the sample is ionized by applying a high voltage to the emitter.

In some cases it is necessary to heat the emitter by passing a current through the wire to evaporate the sample.

DESORPTION SOURCES

Field desorption

Page 22: Ionisation techniques

Ionization takes place by quantum mechanical tunneling mechanism which involves transfer of ions from the sample molecule to the anode (emitter).

This results in formation of positive ions which are radical ions (M+) and cations attached species such as (M+Na)+. (M+Na)+ are produced during desorption by attachment of trace alkali metal ions present in analyte.

Advantages

• Works well for small organic molecules, low molecular weight polymers and petrochemical fractions.

• Sensitive to alkali metal contamination.

• Sample must be soluble in a solvent.

• Not suitable for thermally unstable and non volatile samples.

• Structural information is not obtained as very little fragmentation occurs.

Disadvantages

Page 23: Ionisation techniques

Electrospray ionization

The method generates ions from solution of a sample by creating fine spray of charged droplets.

A solution of sample is pumped through a fine, charged stainless steel capillary needle at a rate of few microlitres/minute.

The needle is maintained at a high electric field (several kilovolts) with respect to cylindrical electrode.

Page 24: Ionisation techniques

The liquid pushes itself out of the capillary as a mist or aerosol of fine charged droplets. Set of aerosol droplets is produced by a process involving formation of Taylor cone and a jet from the tip of this cone.

Page 25: Ionisation techniques

These charged droplets are then passed through desolvating capillary where the solvent is evaporated in the vacuum and attachment of charge to the analyte molecules takes place.

Desolvating capillary uses warm nitrogen as nebulising gas. The desolvating capillary is maintained under high pressure.

As the droplets evaporate the analyte molecules comes closer together.

These molecules become unstable as the similarly charged molecules comes closer together and the droplets explode once again.

This is referred as Coulombic fission. The process repeats itself until the analyte is free from solvent and is lone ion.

The ion then moves to the mass analyzer.

Page 26: Ionisation techniques

In this method quassimolecular ions are produced by addition of a proton (hydrogen ion) to give (M+H)+ or other cations such as sodium ion (M+Na)+ or removal of hydrogen ion (M-H).

Multiply charged ions are often observed and these ions are even electron species indicating that electrons have neither been added nor removed.

Advantages

• Used for analysis of high molecular weight biomolecules such as polypeptides, proteins, oligonucleotides and synthetic polymers.

• Can be used along with LC and capillary electrophoresis.

• softest ionization technique.

Disadvantage• Multiply charged ions are confusing and needs careful interpretation.

• Sensitive to contaminants such as alkali metals or basic compounds.

• Not suitable for low polarity compounds.

Page 27: Ionisation techniques

Matrix Assisted Laser Desorption (MALDI)In this method ionization is carried out by bombarding a laser beam on the sample dissolved in a matrix solution.

Page 28: Ionisation techniques

Matrix is used in MALDI to

• Absorb the laser energy.• Prevent analyte agglomeration.• Protect analyte from being destroyed by direct laser beam.

Matrix

Matrix consists of a crystallized molecules of which the most commonly used are :-

• 3,5 – dimethoxy – 4 – hydroxy cinnamic acid (sinapinic acid).• α – cyano – 4 – cinnamic acid (α – cyano or α – matrix).• 2,5 – dihydroxy benzoic acid (DHB)

Page 29: Ionisation techniques

Solution of the matrix is made in a mixture of highly purified water and another organic compound (acetonitrile or ethanol).

Triofluoro acetic acid (TFA) is also added. If sinapinic acid is used as a matrix the solution is prepared by adding:

20 mg/ml of sinapinic acid. Water: acetonitrile: TFA (50:50:0.1).

Matrix solution is then mixed with the analyte to be investigated.

The organic compound acetonitrile dissolves hydrophobic proteins present in the sample while water dissolves hydrophilic proteins.

The solution is then spotted in a air tight chamber on the tip of the sample probe.

With a vacuum pump the air is removed and vacuum is created which leads to evaporation of the solvent leaving behind a layer of recrystalized matrix containing analyte molecules.

Page 30: Ionisation techniques

Now the laser beam (EMR) is shooted to the sample, the range of uv radiation used is 360-390nm.due to the absorbing substance is present in matrix ,it absorbs radiation or energy and thus it transfers some of its energy to sample molecule where by the molecular ions are formed and then accelerate to analysers.

Page 31: Ionisation techniques

Plasma desorption

Plasma desorption produces molecular ions from the samples coated on a thin foil when a highly energetic fission fragments from the Californium-252 “blast through” from the opposite side of the foil.

The fission of Californium-252 nucleus is highly exothermic and the energy released is carried away by a wide range of fission fragments which are heavy atomic ion pairs.

Ion pair fission fragments depart in opposite directions.

Each fission of this radio active nucleus gives rise to two fragments traveling in opposite directions (because necessity of momentum conversation).

Page 32: Ionisation techniques

A typical pair of fission fragments is 142Ba18+ and 106TC22+, with kinetic energies roughly 79 and 104MeV respectively.

When such a high energy fission fragments passes through the sample foil, extremely rapid localized heating occurs, producing a temperature in the range of 10000K.

Consequently, the molecules in this plasma zone are desorbed, with the production of both positive and negative ions.

These ions are then accelerated out of the source in to the analyzer system.

Page 33: Ionisation techniques

Fast Atom Bombardment

It is an ionization technique in which the analyte and non-volatile liquid matrix mixture is bombarded by a high energy beam of inert gas such as Argon or Xenon..

Page 34: Ionisation techniques

This technique is used for ionization of polar high molecular weight compounds such as polypeptides. Commonly used matrices include :-

GlycerolMonothioglycerolCarbowax2,4 – dipentyl phenol3 – nitrobenzyl alcohol (3 – NBA)

These solvents easily dissolve organic compounds and do not evaporate in vacuum.

The bombarding beam consists of Xenon or Argon atoms of high translational energy.

This beam is produced by first ionizing the Xenon (or Argon atoms with electrons to give Xenon radical cations.

Xe + e- = Xe.+ +2e-

The radical cations are then accelerated to 6 – 10 KeV to give radical cations of high translational energy (Xe)++, which are then passed through a chamber containing Xenon atoms at a pressure of 10-5 torr.

Page 35: Ionisation techniques

During this passage high energy cation obtain electrons from Xenon atoms to become high energy atoms (Xe). The lower energy ions are removed by electrostatic deflector.

(Xe)++ Xe.+ + Xe

(Xe).+ + Xe (Xe) + Xe.+

The analyte is dissolved in the liquid matrix such as glycerol and applied as a thin layer on the sample probe shaft.

The mixture is bombarded with the high energy beam of Xenon atoms. Xenon ionizes the glycerol molecules to give glycerol ions. These ions react with the surrounding glycerol molecules to produce (G+H)+ as reactant ions.

The sample molecules then undergo proton transfer or hydride transfer or ion-pair interaction with reactant ions to give quassimolecular or psuedomolecular ions such as (M+H)+, (M-H)- or (M+G+H)+.

These ions are then extracted from slit lens system designed to collect ions and directed to mass analyzer.

Page 36: Ionisation techniques

Advantages

• Used for ionization of polar high molecular weight samples.• Provides rapid heating of samples and reduces sample fragmentation.• Rapid ionization.

Disadvantages

• Difficult to distinguish between low molecular weight compounds.• Compounds must be soluble in liquid matrix.• Not good for multiply charged compounds

Page 37: Ionisation techniques

Secondary ion mass spectrometry

Secondary ion mass spectrometry is nearly identical to FAB except the primary ionizing beam is an ion beam rather than a neutral atom beam. The Cesium or Argon ions are most commonly used.

The source consists of a cylindrical grid and a vertically placed ion gun or filament.

Argon or Cesium gas is ionized by heating the filament to produce monoenergetic noble gas ions. The ion gun can produce an ion beam of diameter ranging from 0.1mm to 1mm.

The ions are accelerated to a potential of 300 to 3000 eV.

This ion beam is then bombarded on to the surface of the sample.

This results in the formation of secondary sample ions by charge transfer interaction between the sample molecules and the primary gas ions.

Page 38: Ionisation techniques

The ions formed in the cylindrical grid are then extracted from one end and focused on the target or mass analyzer by an electrostatic lens system.

Advantages• Higher sensitivity.• Selection of Beam diameter permits for rapid transition from a small.• surface analysis with a small beam to a large surface area.

Thermal ionization or Surface ionization

Thermal surface ionization source is useful for inorganic solid materials.

Samples are coated on a tungsten ribbon filament and then the filament is heated until the sample is evaporates.

As the sample evaporates it undergoes ionization.

Page 39: Ionisation techniques

The probability of ionization is predictable and is a function of work function of :-

• Ionization potential of the sample E1• Work function of the filament material Φ• Filament temperature T

This can be summarized as followsn+/n0 = exp[z(Φ – E1)/KT]

z = electronic chargeK = Boltzmann’s constantn+ = Number of ions formedn0 = Number of neutral species

Page 40: Ionisation techniques

References1. Principles of Instrumental analysis. Fifth Edition by Douglas. A.

Skoog, F. James Holler and Timothy A. Nieman. Page No. 499 – 511.

2. Instrumental Methods Of Analysis. Seventh Edition by Willard Meritt. Page No. 468 – 74.

3. http://www.chem.ox.ac.uk/spectroscopy/mass-spec/Lecture/oxmain_lectureCI.html

4. http://www.astbury.leeds.ac.uk (A.E. Ashcroft's MS web pages and tutorial)