    Definition of instability:

    Particles in the linac are driven by magnetic lenses and RF cavities; the correspondingforces are generally periodic, sometimes with adiabatic changing parameters. In this

    way the accelerator designer determines a beam with the same periodicity, in order to

    have a simple transport all over the structure.

    In other words it is required as a nominal (or unperturbed) situation:

    with number of particles in the elementary volume of phase space, since are the three

    spatial coordinated

    are the associated moments; tis time and Tis the period.

    This is the result achieved for single particle dynamics in the first lessons, where the

    periodic beta function defines the envelope equal to itself period after period. But if the

    beam intensity increases the electromagnetic forces determined by the beam cannot be

    neglected, the condition (1) has to be preserved in presence of external plus internal

    beam forces. In the next section we shall say something more about internal forces.

    Beam breakup:

    A very well known instability affecting beam quality in electron linacs is the beam

    breakup (BBU). When the bunch current is high it can happen that the transverse

    oscillation of the head of the bunch causes a resonant growth of the transverse

    oscillation of the tail, with consequent deformation of the bunch, emittance growth and

    possibly beam losses. The emittance growth is particularly detrimental in the case of a

    linear collider where the luminosity is affected.

    Electromagnetic fields generated by the beam :

    Each particle in the beam feels the effect of the electromagnetic field generated by the

    other particles in the field boundaries due to the accelerator elements. The

    mathematical problem corresponds to the solution of the Maxwell equations in the

    presence of the beam charges in motion, with accelerator boundary conditions. In

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    2/19

    practice the problem can often be simplified, in consideration of beam energy and

    acceleration frequency.

    Linear Accelerator

    View larger with caption

    The Electron Gun

    The electron gun, is where electron acceleration begins. The electrons start out

    attached to the molecules in a plate of barium aluminate or other thermionic materials

    such as thorium. This is the cathode of the electron gun. A cathode is a surface that

    has a negative electrical charge. In linac electron guns this charge/ electrons are

    emitted on heating the cathode. Barium aluminate is a "thermionic" material; this means

    that it's electrons tend to break free of their atoms when heated. These electrons "boil"

    near the surface of the cathode. The gate is like a switch. It consists of a copper screen,

    or "grid," and is an anode. An anode is a surface with a positive electrical charge. Every

    500 millionth of a second the gate is given a strong positive charge that causes

    electrons to fly toward it from the cathode in tremendous numbers. As these electrons

    reach the gate, they become attracted even more strongly by the main anode, and pass

    through the gate. Because the gate is pulsing at a rate of 500 million times per second

    (500 MHz), the electrons arrive at the anode in loose bunches, a 500 millionth of a

    second apart. The anode is a torus (a doughnut) shaped to create an electromagnetic

    http://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=cvr1.jpg&pg=linachttp://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=cvr1.jpg&pg=linachttp://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=cvr1.jpg&pg=linachttp://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=cvr1.jpg&pg=linachttp://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=cvr1.jpg&pg=linac
  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    3/19

    field that guides most of the electrons through the hole into the next part of the

    accelerator, called the buncher. B. The Buncher The purpose of the buncher is to

    accelerate the pulsing electrons as they come out of the electron gun and pack them

    into bunches. To do this the buncher receives powerful microwave radiation from the

    klystron. The microwaves accelerate the electrons in somewhat the same way that

    ocean waves accelerate surfers on surfboards. Look at the following graph:

    The yellow-orange disks are electrons in the buncher. The curve is the microwave

    radiation in the buncher. The electrons receive more energy from the wave--more

    acceleration--depending on how near they are to the crest of the wave, so the electrons

    riding higher on the wave catch up with the slower ones riding lower. The right-hand

    wave shows the same group of electrons a split second later. On the front of the wave,

    the two faster electrons have almost caught up with the slower electron. They won't

    pass it though, because they are now lower on the wave and therefore receive less

    acceleration. 8

    The higher electron on the back of the wave gets just enough acceleration to match the

    speed of the wave, and is in the same position as it was on the left-hand wave. This

    represents the last electron in the bunch. The lower electron on the back of the wave

    gets too little energy to keep up with the bunch and ends up even lower on the right-

    hand wave. Eventually it will fall back to the electron bunch forming one wave behind.

    C. The Linac The linac itself is just an extension of the buncher. It receives additional

    RF power to continue accelerating the electrons and compacting them into tighter

    bunches. Electrons enter the linac from the buncher at a velocity of 0.6c--that's 60% of

    the speed of light. By the time the electrons leave the linac, they are traveling very close

    to the speed of light.

    1. Gridded Electron Gun:-

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    4/19

    Controls dose rate rapidly and accurately. Permits precise beam control for dynamic

    treatments, since gun can be gated. Removable for cost - effective replacement.

    2. Energy Switch

    Patented switch provides energies within the full therapeutic range at consistently high

    stable dose rates, even with low energy x ray beams. Ensures optimum performance

    and spectral purity at both energies.

    3. Wave Guide

    High efficiency, side coupled standing wave accelerator guide with demountable

    electron gun and energy switch

    4. Achromatic 3 field bending magnet

    Unique design with fixed +/- 3 % energy slits ensures exact replication of the input beam

    for every treatment. The 270 degree bending system, coupled with varians 3

    dimensional servo system, provides for a 2mm circular focal spot size for optimal portal

    imaging.

    5. Real time beam control steering system

    Radial and transverse steering coils and a real time feed back system ensures that

    beam symmetry is within +/- 2%at all gantry angles.

    6. Focal spot size

    Even at maximum dose rate and gantry angle the circular focal pot remains less

    than 2mm, held constant by a focus solenoid. Assures optimum image quality for portal

    imaging

    7. 10 port carousel

    New electron scattering foils provide homogenous electron beams at therapeutic

    depths. Extra ports allow for future development of specialized beams.

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    5/19

    8. Ion chamber

    Dual sealed ion chambers with 8 sectors for rigorous beam control provide two

    independent channels, impervious to changes in temperature and pressure. Beam

    Dosimetry is monitored to be within +/- 2% for long term consistency and stability.

    9. Asymmetric Jaws

    Four independent collimators provide flexible beam definition of symmetric and

    asymmetric fields. 10. Multi-Leaf Collimator Dynamic full field resolution 120 leaf MLC

    with dual redundant safely read out for most accurate conformal beam shaping for IMRT

    treatments.

    WORKING AND DESCRIPTION MEDICAL LINEAR ACCELERATOR : - With the

    advances in linear accelerator models and working condition, the disadvantages of

    earlier cobalt machine based gamma radiation are overcome. The main advantage of

    linac over Co 60 is skin sparing, means the radiation affects the skin surface. Most of x-

    ray energy goes to tumour; secondly there is minimal scatter of x-ray energy outside the

    beam. Sharply defined x-ray beam minimizes the side effects of treatment, so only a

    small amount of radiation travel to other parts of the body. Linac allows sharpness of the

    beam edge, allows a very precise treatment and adjacent tissues spared. And this linac

    may be programmed to treat with electron rather than x rays for special situations.

    Most linear Accelerators make use of isocentric mount, modern linacs use 100 cm

    isocentre, i.e the source of x-ray is 100 cm from the axis of rotation, so the patient is

    positioned with the center of tumour on the axis of rotation. Treatment delivered using

    several angular positioned, without realigning the patient simply by rotating the unit. For

    low energy medical linacs, having energy range of 4 6 MeV, the accelerating tube can

    be made short enough to allow the arrangement. The electrons travel down the tube

    and strike the target, produce x ray beam symmetrical about the line of source to center

    of the machine. If electron beam required the target is moved to the side to allow the

    electron beam emerge thro a thin window

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    6/19

    For high energy machines, the accelerating tube must be made longer. RF power from

    klystron is brought to the wave guide through the axis of machine using a rotating

    vacuum seal. The electron beam is bent thro an angle of 60 degree passes thro

    quadruplet focusing magnets and then bent 90 degree in the head to hit a target or

    emerge through a window for electron beam. In high energy medical linacs there are

    some important components required to control and shape the beam from a Linac used

    in both electron and photon modes. High energy electron from wave guide is focused at

    the target which is thick enough to stop the electron. When we use electron mode of

    operation in the linac, the tungsten target on the sliding mount moved towards right side

    and allowing the electron to pass through an open window. Just below the target one

    more sliding mount is located, which contains a variety of electron scattering foil for

    different electron energy and a port for photon mode. These beams passes through a

    conical hole in the primary collimator made of heavy metal. Below this primary

    collimator a secondary slide containing the flattening filter of tungsten and a one section

    for a electron beam. Ion chamber is positioned near the secondary slide to monitor the

    dose and a quadrant is connected to a servo mechanism to control the beam optics. A

    light localizer is placed below the ion chamber which is retraced from the beam after

    patient position under the beam.

    Types of External Beam Radiation Therapy Conventional external Beam radiation

    therapy

    The science of radiation oncology and medical physics has developed standard

    approaches to dose delivery. In many cancer cases the treatment approach may be

    very similar and allows for conventional treatment. For example, many tumors can be

    treated with a single field from the front and a single field from the back or with two

    fields from the opposite sides. These are examples of parallel opposed fields. The

    combination of fields helps to uniformly deliver dose across the tumor. Sometimes 3 or

    4 fields will be used. Occasionally, the gantry of the linear accelerator will rotate during

    treatment using what is called arc therapy. 3-D Conformal Radiation Therapy

    Through the advancement of imaging technology enhanced images of the body allow

    for programming of treatment beams to conform better to the shape of a tumor. Hence

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    7/19

    treatment is more effective and side effects are reduced. By treating with large numbers

    of beams each shaped with a multileaf collimator (MLC) or cerrobend block, radiation

    dose is delivered uniformly and conformally to the tumorIntensity Modulated

    Radiation Therapy (IMRT)

    IMRT is one of the latest advancements in radiation therapy. This new approach to

    treatment allows for dose sculpting and even distribution of delivery to avoid critical

    structures while delivering precise uniform treatment. In this technique, the multileaf

    collimator (MLC) moves and modulates the radiation as the linac treats the patient.

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    8/19

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    9/19

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    10/19

    With the improvement in computer field, diagnostic methods and other breakthroughs in

    the modern electronics systems, made possible a variety of treatment technique and

    methods to treat the patient with more accurate and less side effects or nil side effects,

    sparing the critical organs in and around the tumour region. The electronic Portal

    Imaging device (EPID) made possible to verify the treatment portal without delivering a

    high dose and make correction where ever it is mandatory. Computer assisted

    treatment delivery and record verify systems reduces the errors and assures the

    accurate treatment delivery with reduced time frame, and with the less man power.

    Newer treatment modalities like IGRT, DART and other high resolution treatment

    modalities made possible the treatment of organs involved with motion due to breath

    and circulations. Robotic arm based radiosurgery made the skull based radiosurgery

    programme a grand success with high accuracy.

    Technology-driven delivery methods

    Technological developments in computers and accelerator designs, particularly inverse

    treatment planning and dynamic multi-leaf collimation systems (dMLCs), have given the

    radiotherapy community the ability to deliver conformal and intensity-modulated

    radiation therapy (IMRT) treatments. With the implementation of image-guided radiation

    therapy (IGRT), there is a potential to deliver inhomogeneous dose distributions that

    increase the dose inside the target by 20-30% over the minimal peripheral dose, while

    decreasing possible normal tissue complications by collapsing the planning target

    volume (PTV) onto the clinical target volume (CTV), as is often done in Stereotactic

    radiosurgery.

    Image-guided intensity-modulated and adaptive helical TomoTherapy

    Image-guided IMRT is redefining the practice of radiation oncology. Traditional methods

    of implementing beam intensity modulation have included individually designed

    compensators, static multi-leaf collimators (MLC), dynamic MLC and sequential (serial)

    tomotherapy. Helical Tomotherapy provides added functionality to enhance the

    application of IMRT. It facilitates adaptive radiotherapy and conformal avoidance. These

    advances improve normal tissue sparing and permit dose reconstruction and

    verification, thereby allowing significant biologically effective dose escalation and

    reduced radiation toxicity. Recent radiobiological findings can be translated into altered

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    11/19

    fractionation schemes that aim to improve the local control and long-term survival. The

    intrinsic capability of helical Tomotherapy for 14

    megavoltage CT (MVCT) imaging for IMRT image-guidance is an added feature aiding

    in further

    What is this equipment used for?

    How does the equipment work?

    Who operates this equipment?

    How is safety ensured?

    What is this equipment used for?

    A linear accelerator (LINAC) is the device most commonly used forexternal beam

    radiation treatmentsfor patients with cancer. The linear accelerator is used to treat all

    parts/organs of the body. It delivers high-energyx-raysto the region of the patient's

    tumor. These x-ray treatments can be designed in such a way that they destroy the

    cancer cells while sparing the surrounding normal tissue. The LINAC is used to treat all

    body sites, using conventional techniques,Intensity-Modulated Radiation Therapy

    (IMRT),Image Guided Radiation Therapy (IGRT),Stereotactic Radiosurgery (SRS) and

    Stereotactic Body Radio Therapy (SBRT).

    How does the equipment work?

    The linear accelerator uses microwave technology (similar to that used for radar) to

    accelerate electrons in a part of the accelerator called the "wave guide," then allows

    these electrons to collide with a heavy metal target. As a result of the collisions, high-

    energy x-rays are produced from the target. These high energy x-rays are shaped as

    they exit the machine to conform to the shape of the patient's tumor and the customized

    beam is directed to the patient's tumor. The beam may be shaped either by blocks that

    are placed in the head of the machine or by a multileaf collimator that is incorporated

    into the head of the machine. The patient lies on a moveable treatment couch and

    http://www.radiologyinfo.org/en/info.cfm?pg=linac#part_onehttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_onehttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_twohttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_twohttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_threehttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_threehttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_fourhttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_fourhttp://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=317http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=317http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=317http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=317http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=3http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=3http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=3http://www.radiologyinfo.org/en/info.cfm?pg=imrthttp://www.radiologyinfo.org/en/info.cfm?pg=imrthttp://www.radiologyinfo.org/en/info.cfm?pg=imrthttp://www.radiologyinfo.org/en/info.cfm?pg=imrthttp://www.radiologyinfo.org/en/info.cfm?pg=igrthttp://www.radiologyinfo.org/en/info.cfm?pg=igrthttp://www.radiologyinfo.org/en/info.cfm?pg=igrthttp://www.radiologyinfo.org/en/info.cfm?pg=stereotactichttp://www.radiologyinfo.org/en/info.cfm?pg=stereotactichttp://www.radiologyinfo.org/en/info.cfm?pg=stereotactichttp://www.radiologyinfo.org/en/info.cfm?pg=stereotactichttp://www.radiologyinfo.org/en/info.cfm?pg=stereotactichttp://www.radiologyinfo.org/en/info.cfm?pg=stereotactichttp://www.radiologyinfo.org/en/info.cfm?pg=igrthttp://www.radiologyinfo.org/en/info.cfm?pg=imrthttp://www.radiologyinfo.org/en/info.cfm?pg=imrthttp://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=3http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=317http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=317http://www.radiologyinfo.org/en/info.cfm?pg=linac#part_fourhttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_threehttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_twohttp://www.radiologyinfo.org/en/info.cfm?pg=linac#part_one
  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    12/19

    lasers are used to make sure the patient is in the proper position. The treatment couch

    can move in many beam comes out of a part of the accelerator called a gantry, which

    can be rotated around the patient. Radiation can be delivered to the tumor from any

    angle by rotating the gantry and moving the treatment couch.directions including up,

    down, right, left, in and out. The

    Who operates this equipment?

    View larger with caption

    The patient'sradiation oncologistprescribes the appropriate treatment volume and

    dosage. Themedical radiation physicistand thedosimetristdetermine how to deliver

    the prescribed dose and calculate the amount of time it will take the accelerator to

    deliver that dose.Radiation therapistsoperate the linear accelerator and give patients

    their daily radiation treatme

    How is safety ensured?

    Patient safety is very important and is assured in several ways.

    Before treatment is delivered to the patient, the treatment plan is developed and

    approved by the radiation oncologist in collaboration with the radiation dosimetrist and

    physicist. The plan is double-checked before treatment is given and quality-control

    procedures ensure that the treatment delivered is the same as was planned.

    Quality control of the linear accelerator is also very important. There are several

    systems built into the accelerator so that it will not deliver a higher dose than the

    radiation oncologist has prescribed. Each morning before any patients are treated, the

    http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=126http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=126http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=126http://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=astro24.jpg&pg=linachttp://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=216http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=216http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=216http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=297http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=297http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=297http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=298http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=298http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=298http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=299http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=299http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=299http://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=astro24.jpg&pg=linachttp://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=299http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=298http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=297http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=216http://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=astro24.jpg&pg=linachttp://www.radiologyinfo.org/en/photocat/gallery3.cfm?pid=1&image=astro24.jpg&pg=linachttp://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=126
  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    13/19

    radiation therapist performs checks on the machine using a piece of equipment called a

    "tracker" to make sure that the radiation intensity is uniform across the beam and that it

    is working properly. In addition, the radiation physicist conducts more detailed weekly

    and monthly checks of the linear accelerator.

    Modern linear accelerators also have internal checking systems to provide further safety

    so that the machine will not turn on until all the treatment requirements prescribed by

    your physician are perfect. When all the checks match and are perfect, the machine will

    turn on to provide your treatment.

    During treatment the radiation therapist continuously watches the patient through a

    closed-circuit television monitor. There is also a microphone in the treatment room so

    that the patient can speak to the therapist if needed. Port films (x-rays taken with the

    treatment beam) or other imaging tools are checked regularly to make sure that the

    beam position doesn't vary from the original plan.

    Safety of the staff operating the linear accelerator is also important. The linear

    accelerator sits in a room with lead and concrete walls so that the high-energy x-rays

    are shielded. The radiation therapist must turn on the accelerator from outside the

    treatment room. Because the accelerator only gives off radiation when it is actuallyturned on, the risk of accidental exposure is extremely low. The treatment room is

    shielded to such an extent that even pregnant women may safely operate linear

    accelerators.

    THEACCELERATING STRUCTURE AND THE BEAM

    Linear accelerators consist of two basic elements: the accelerating structure and

    the particle beam. The accelerating structure depends on the type of linac. The main

    types of linacs are

    :

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    14/19

    Electric field polarizes the drift tubes so that periodically the two ends are charged toopposite sign.

    When charging currents flow to the right along the tubes, an equal current flows to theleft along the inside wall of the outer cylinder.The tubes are supported at their centers by radial rods along which no current flows.

    DC linacs:

    like Van de Graafs, in which the structure consists of some kind

    of column of electrodes. These electrodes sustain a DC electric field which

    Accelerates a continuous stream of particles. DC linacs are limited to a few tens

    of MeV.

    Induction linacs:

    in which the accelerating electric fields are obtained, according

    to Faradays law, from changing magnetic fluxes. These changing magnetic

    fluxes are generated by large pulsed currents driven through linear arrays of

    magnetic toroids. The beam path, along which the electric field develops, can

    be considered as the single turn secondary of a transformer. Induction linacs

    are generally used in medium-energy high-current pulsed applications.

    RF linacs:

    the type on which we will concentrate here, can be categorized in

    a number of ways: low frequency (UHF), microwave frequency (L, S, C, or Xband),

    laser frequency; CW or pulsed; traveling-wave or standing-wave; room

    temperature or superconducting. In all these cases, the structure is a conducting

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    15/19

    Array of gaps, cavities or gratings along which rf waves with an electric field

    Parallel to the beam can be supported and built up through some resonant process.

    RF linacs are used for a wide spectrum of applications from injectors into

    Circular accelerators, to entire high-energy accelerators such as SLAC, medical

    Accelerators and many others.

    While the accelerating structure can be considered the heart of each individual

    Machine, it cannot work without its associated systems such~-as the power source, the

    Vacuum, cooling, support and alignment, and instrumentation and control systems. It

    is important to note that a technological breakthrough in any one of these associated

    Systems can have profound effects on the main design of the accelerator.

    The second basic element of the linac is, of course, the beam awn d the particles which

    Compose_ it. The vast majority of linacs today are elect& (for positron) machines; they -

    Number about 1400 accelerators of which a large percentage are commercial radiation

    Therapy machines. The other linacs accelerate protons (H*) and in a few cases ions.

    There are about 50 proton or ion linacs in existence in the world.

    Very little will be said in these lectures about the sources of these particles, guns,

    Ion sources, duoplasmatrons, polarized beams, strippers, positron radiators, etc. This

    Is an encyclopedic subject by itself with many specialties and sub-specialties which

    Can not profitably be summarized here. The reader should not conclude that because

    The subject has been left out, it is not of crucial importance to the design and operation

    Of a given linac. Not only does the source have an effect on how well the accelerator

    Can perform its function, but in some cases it determines how a new concept can or

    Cannot be approached. An example of this is the electron source for the linear collider

    For which a conventional electron gun cannot create a beam with a sufficiently small

    Remittance and must be followed by a damping ring to cool it down.

    The fundamental problems in beam dynamics are:

    1. Longitudinal bunching and stability,

    2. Focusing and transverse stability, and

    3. Steering and transport to a target or interaction area.

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    16/19

    Finally we note that instead of operating a SW-cavity in the -mode one can operate in

    the 0-modewith good efficiency.

    Again, forward and backward space harmonics add up in phase in every cell and both

    contribute to the acceleration. Since the electromagnetic fields in adjacent cells are in

    phase the separating walls can be left out without affecting the field distribution.

    An additional analysis of the mode indicates that there are cell-to-cell phase errors due

    to Q-losses, and moreover it is rather sensitive to dimensional or frequency errors. on

    the other hand, is far less sensitive to these effects. For these reasons, for room

    temperature lilacs one often operates in the /2 or 2/3 mode. However, we note there is

    technique to mitigate against the sensitivity disadvantages of the mode. Namely, we

    add coupling cavities to either side of the main accelerating cavities in a such a way that

    the chain uses a /2 mode while the main-to-main cell phase shift is

    Machine Specifications and Purchase Agreement

    The recommendations of the equipment selection committee are followed by the

    development of comprehensive machine specifications and a binding purchase

    agreement. If the equipment is purchased through a bid process, then the machine

    specifications are developed before the decision is made to select a manufacturer.

    Otherwise, the final specifications may be developed in close collaboration with the

    manufacturers representative. All manufacturers have developed product specifications

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    17/19

    for the functional performance of their equipment in response to requirements from

    potential users and in commercial competition with other manufacturers.

    These are available in the form of product data and specification sheets, which can

    serve as a good starting point for the purchase agreement. Any special requirement can

    then be added as an addendum. This saves lot of time and effort from being expended

    in repetitive work. An example of an addendum to the purchase agreement is presented

    in the Appendix, which illustrates how special requests are included in the agreement.

    It is imperative that the facility physicist develop a comprehensive acceptance testing

    document with a detailed test procedure to verify each term of the agreement and

    machine specifications. This document should be shared with the manufacturers

    representative before the installation begins so that all ambiguities are clarified in

    advance. It is not prudent to depend on the manufacturer-supplied acceptance test

    procedure exclusively. However, it should be reviewed thoroughly before acceptance

    testing. IEC publication 976, entitled Medical Electron

    Accelerators: Functional Performance Characteristics is an excellent resource to set

    up test procedures.

    It is essential that the physicist review the facility layout with the planning and

    installation department of the accelerator manufacturer. They can provide very useful

    information on work flow, equipment layout and special requirements. A joint meeting ofthe equipment planning coordinator, architect, contractor, and physicist in the earlier

    stages of construction is very helpful and productive. This meeting can resolve all

    potential problems regarding electrical power supply, conduit layout, air conditioning,

    and chilled water requirements for the machine. The shielding design and its final

    approval are solely the responsibility of the physicist, even if

    generic vault design and shielding barrier thicknesses are available from other sources.

    IV. Accelerator Installation

    The physicist and facility engineer (if available) should work closely with the installation

    engineer. A close collaboration during the installation can reduce the acceptance testing

    time considerably. It is important that the facility personnel do not interfere in the work of

    the installation engineer but observe the progress in the background. As soon as the

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    18/19

    accelerator is capable of producing a radiation beam, a series of tests should be

    conducted to assure the safety

    of all concerned. These include

    Testing of door interlocks

    Testing of proper operation of the emergency off switches

    A preliminary calibration of the machine output in all modes

    Modern-Day Linac Acceptance Testing and Commissioning - J.R. Palta, Ph.D.

    A radiation survey in both controlled and uncontrolled areas around the treatment

    vault at the highest available dose rate and under worst irradiation conditions (without

    phantom)

    A full radiation survey including the photon and neutron leakage measurements will still

    have to be completed to comply with regulatory requirements after a full calibration. The

    preliminary survey is done to assure the safety of individuals during the acceptance

    testing and commissioning.

    V. Acceptance Testing

    The installation is followed by acceptance testing by the physicist to ensure that the

    machine meets the product specifications and the purchase agreement. These tests are

    conducted

    according to the acceptance testing procedure agreed on between the manufacturers

    representative and the facility physicist. Each facility should have the necessary

    equipment for acceptance testing. This includes a 3-D water phantom scanner with

    computer interface, ion

    chambers, and electrometer X ray films; film laser scanner; and precision level. It is

    important to know that each machine comes with the functional performance test values

    performed in test cells in the factory. These are helpful for comparison during

    acceptance testing. IEC Report 977

    provides suggested values of functional performance that all manufacturers voluntarily

    comply with. A summary of the suggested values of functional performance is given in

    the Table. Some of these values are required to be more stringent for special

    application of the linac. For example, it is not unusual to require a radiation isocenter

    tolerance within 1 mm diameter of the

  • 7/30/2019 Intro Duction of Linear Accelrator by Asif Rasheed

    19/19

    linac scheduled to be used for high precision radiation therapy and radiosurgery.