Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M. De Lazzari Laboratori Nazionali di Legnaro HIAT 09, Venice, 12 June 2009

Upload: halle-jocelyn

Post on 19-Jan-2016

216 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Improved on line performance of the installed ALPI Nb sputtered QWRs

A.M. Porcellato, S. Stark,

F. Stivanello, L. Boscagli

F. Chiurlotto, D. Giora,

M. De Lazzari

Laboratori Nazionali di Legnaro

HIAT 09, Venice, 12 June 2009

Page 2: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

A.M.Porcellato, Lea Workshop, Padua April 7th 2005 2

ALPI

TANDEM

PIAVE

ECR

Page 3: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

ALPI resonators

CR10 CR9 CR8 CR7 CR6 CR5 CR4 CR3 B2

CR12 CR13 CR14 CR15CR16 CR17 CR18 CR19 CR20 B4

COLD BOX

B3

, 80 MHz, full Nb =0.13, 160 MHz, Nb/Cu

, 160 MHz,Nb/Cu , 160 MHz, Pb/Cu

Page 4: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

CR20 resonators are in ALPI since 1998T are drilled from a billet of OFHC Cu, 99.95% No brazed joints The average operational accelerating field is 6 MV/m

Installed in CR20 and CR19Very simple shape External beam ports jointed by In gaskets Rounded shorting plate Capacitive coupler

=0.13; f= 160 MHz

ALPI high QWRs

Page 5: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

CR19 cavitiesinstalled ALPI in 2001, similar in inner shape to CR20 cavities made in SeCu circumferential brazed joint brazed collar and supports

ALPI high QWRs

Page 6: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

CR19 and CR20

year

Ea@

7W [

MV

/m]

CR19 and CR20 Better performance in CR20Lower performance in CR19 (similar shape but different construction technology) 2 cavities to be repaired because of a stuck coupler and rf line damage,

High

QWRs

Page 7: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Brazed joints (especially the ones in the outer resonator surface) Flat shorting plate Beam ports shape Inductive coupler (hole in high current region)

ALPI =0.11 160 MHz

Limited the reached performance to 4.8MV/m @7W, a lower level than the ones of high resonators.

ALPI medium QWRs

Originally Pb plated; Nb sputtered in between 1998-2003

PbNb

Page 8: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

QCR7 and CR8

CR9 and CR10

CR12 and CR13

MV/m

CR16 and CR17

CR18

Q-curves of medium cavities installed in ALPI cryostats

Q

MV/m

Page 9: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Ea@

7W [M

V/m

] CR7 and CR8

CR9 and CR10

CR12 and CR13

CR14 and CR15

CR16 and CR17

CR18

year

Operational Ea in QWRs of medium cryostats

Ea@

7W [M

V/m

]rinsing

rinsing

rinsing

year

Page 10: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

,

Iimprovements maintaining old substrates More accurate “Lifting” of

the old Cu substrate (the opening of all the trapped enclaves was not well performed in the first produced resonators)

Better tuning plate contact to avoid Indium joint (by modifying the end plate and doubling its fixing screws)

HPW resonator rinsing

Cleaner assembling condition before and after sputtering

Longer high power rf conditioning

Page 11: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

We have To align the resonators with

their beam port open to air. To close the cryogenic circuits

after the resonator assembling It would surely help:

– Avoid cryostat venting to air (because of cryogenic circuits leaks)

– Perform high pressure rinsing after resonator alignment (possible if we have not the In joint)

– Longer rf and He conditioning (5 MV/m reachable @ 7W in old substrates)

Drawbacks /possible improvements

Page 12: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Nb/Cu Average Ea @7W: 4.83MV/m Nb/Cu Operational Ea : 4.70 MV/m

Pb/Cu Average Ea: 2.48 MV/m

1999 2001* 2000* 1999* 2003 2002* 2003 2000 2002 2001 2003 2001

Nb/Cu ALPI QWR Performance

1998

* Recent cryostat maintenance

Cryostat installation year

Page 13: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

old

New medium QWRs

We have 4 new cavities ready with– New beam port design – A rounded shorting plate– A capacitive coupler– Without holes in high current regions– Without brazing in the outer resonator body

Page 14: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

New medium

QWR performance

4 READY

4 cavities produced (MBD2 sputtered twice) Improvement due to cathode optimization Tested up to 3MV/m in laboratory Ea between 5.5 and 6.5 MV/m @7W expected on line Better results possible but we prefer to install them in CR15 as soon as possible

Page 15: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Since 2001 most of the ALPI equivalent voltage (V) is provided by Nb/Cu cavities

Operation at Ea determined by the available cryogenic power

No frequency tracking or fast tuning required No degradation with time; average fields is still improving

Low have to be locked at Ea <3 .5 MV/m in ALPI Higher operational Ea expected by increasing the rf

driving power; new rf cryostat lines and a L.N cooled coupler required

qL

WEa

max

LEV a

L

Cavities at work

W=energy gainq=State of charge

Page 16: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

16

0

2

4

6

8

10

Cavity

Ea [

MV

/m]

Bulk Nb resonator perforamance

Nb, 80 MHz, =0.056, 12+(4) resonators Excellent performance: 6.7 MV/m @ 7W Installed in between 1995-2000Mechanical dampers effectively reduce their sensitivity to acoustic vibrations Sensitivity to He bath pressure drifts (1Hz/mbar) made the cavity locking tricky when the resonators were cooled by the very noisy ALPI cryogenic system Operating up to 3.5 MV/m in ALPI;4.2 MV/m in PIAVE; 2/3 out of workUpgraded tuners, amplifiers, bottom plates… New L.N2 cooled couplers coming soonExpected 5 MV/m in lock condition

Low section

Page 17: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

HIAT 2005 17

CR5 L He bath pressure

from 7Jun 03 , 00:49 to 11 Jun 03 , 23:34

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

1 24 47 70 93 116hr

Bar

CR5 Max

CR5 Min

Recording of L He bath pressure change rate 21 Jul 05

0

0.05

0.1

0.15

0.2

0.25

0.3

0 2 4 6 8 10 12 14 16 18 20 22 24

Time [hours]

bar

B2 CR5 CR6 CR7 CR8

Pressure excursions in

ALPI cryostats

Page 18: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Surface finishing and chemical treatments

. Frequency adjustment by electro-

polishing Electro-polishing (20m, 2

hours, phosphoric acid +butanol, computer controlled)

Rinsing (water, ultrasonic water, HPR)

Chemical polishing (10m, 4 min, SUBU5)

Passivation (sulphamic acid) Rinsing (water, ultrasonic

water, HPR) Drying (ethanol, nitrogen)

Page 19: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Biased Nb sputtering

• Good vacuum•Backing

• No discharges• High substrate

temperature

Cathode: Nb tube

Cu base

Sputtering chamber

Sputtering performed in about 12 steps of 15’Production cycle asks for 9 days

Page 20: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Nb sputtering advantages

Mechanical stability (mechanical vibrations are not an issue) Frequency not affected by changes He bath (p <0.01Hz/mbar, no

frequency tracking) Reduced over-coupling (smaller amplifier, coupler do not need cooling, rf

lines have reduced size and limited rf dissipation) High thermal stability (less prone to hot spots, conditioning easier) Stiffness (in case of loss of isolation vacuum leak…) Absence of Q-disease (less demand on cryogenic system cooling velocity and

reliability) Insensitivity to small magnetic fields (no magnetic shielding) High Q of the N.C. cavity (easier coupling in N.C state, better multipactor

conditioning) Lower X-ray production (lower power available for field emission) Absence of In vacuum joints (vacuum leaks less probable) Price (both material and construction)

Page 21: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

Conclusion

The Nb sputtering technology showed to be very effective in producing reliable resonators, which have high performance, are very steadily phase locked and are easy to put into operation. Even better results were obtained using suitable substrates. The high number of produced and operational resonators and the reliability of the sputtering.

Thanks for your attention!

Page 22: Improved on line performance of the installed ALPI Nb sputtered QWRs A.M. Porcellato, S. Stark, F. Stivanello, L. Boscagli F. Chiurlotto, D. Giora, M

A.M.Porcellato, Lea Workshop, Padua April 7th 2005 22

Cryostat maintenance

•Change of the leaking cryogenic valve (external actuator)

•Viton sealing in the cryostat beam line valves, from resonator integral conditioning (changed + shielded through stainless steel rings)

•Gaskets in the He circuits (changed + fixed by silver plated screws)

•Leaks on the cryostat upper thermal shield

•RF lines (Mechanical adjustments + Cu re-plating)