Image Reconstructionfrom Projections
J. Anthony Parker, MD PhD
Beth Israel Deaconess Medical Center
Boston, Massachusetts
Caveat Lector
![Page 4: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/4.jpg)
ProjectionSingle Slice
Axial
![Page 5: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/5.jpg)
Single Axial Slice: 3600
collimator
Ignoring attenuation, SPECT data are projections
![Page 6: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/6.jpg)
Attenuation: 180o = 360o
keV
150
100
80
60
50
x
Tc-99m
htl(140 keV) ≈ 4 cm
![Page 7: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/7.jpg)
Cardiac Perfusion Data Collection Special Case - 180o
AxialCoronal / Sagittal
Multiple simultaneous axial slices
![Page 8: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/8.jpg)
Dual-Head General-Purpose Gamma Camera: 900 “Cardiac” Position
2 heads: 900 rotation = 1800 data
1
2
![Page 9: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/9.jpg)
Inconsistent projections“motion corrected”
![Page 10: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/10.jpg)
Original data
![Page 11: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/11.jpg)
0
0
0
00
0
0
0
Single Axial Slice: 3600
![Page 12: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/12.jpg)
Sinogram: ProjectionsSingle Axial Slice
0 60
060
pro
ject
ion
an
gle
x
x
x
![Page 13: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/13.jpg)
Uniformity & Motion on Sinogram
1 h
ea
d2
4 m
in
2 h
ea
ds
12
min
12
min
![Page 14: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/14.jpg)
Reconstruction by Backprojection
Backprojection tails
![Page 15: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/15.jpg)
Backprojection2 projections2 objects
![Page 16: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/16.jpg)
projection tailsmerge resulting
in blurring
![Page 17: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/17.jpg)
Projection -> Backprojection of a Point
(1/r)
backprojectionlines add atthe point
tails spread point out
![Page 18: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/18.jpg)
Projection -> Backprojection
![Page 19: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/19.jpg)
Projection->Backprojection Smooths
Smooths or “blurs” the image
(Low pass filter)
((Convolution with 1/r))
Nuclear Medicine physics
Square law detector adds pixels
-> always blurs
Different from MRI (phase)
![Page 20: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/20.jpg)
(Projection-Slice Theorem)“k-space (k,)”
detail
lowfrequency
spatial frequency domainspatial domain2D Fouriertransform
![Page 21: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/21.jpg)
Spatial Frequency Basis Functionsf(u,v) ≠ 0, single u,0f(u,v) ≠ 0, single 0,v
f(u,v) ≠ 0, single u = v
![Page 22: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/22.jpg)
Projection -> Backprojection: k-space
1/k
(Density ofslices is 1/k)
(Fourier Transform of 1/r <-> 1/k)
one projectionmultiple projections
![Page 23: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/23.jpg)
Image Reconstruction: Ramp Filter
Projection -> Backprojection
blurs with 1/r in object space
k-space 1/k ( 1/r<-> 1/k)
Ramp filter
sharpen with k
(windowed at Nyquist frequency)k
k
![Page 24: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/24.jpg)
Low Pass Times Ramp Filter
Low pass,Butterworth– noise
Ramp –reconstruct
![Page 25: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/25.jpg)
What’s Good about FPB
Ramp filter exactly reconstructs projection
Efficient
(Linear shift invariant)
(FFT is order of n log(n)
n = number of pixels)
“Easily” understood
![Page 26: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/26.jpg)
New Cardiac Cameras
Solid state - CZT: $$$, energy resolution
scatter rejection, dual isotope
Pixelated detector: count rate &
potential high resolution
poorer uniformity
Non-uniform sampling: sensitivity
potential for artifacts
Special purpose design
closer to patient: system resolution
upright: ameliorates diaphragmatic attenuation
![Page 27: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/27.jpg)
![Page 28: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/28.jpg)
Collimator Resolution*
Single photon imaging (i.e. not PET)
Collimators: image formation
Sensitivity / resolution trade-off
Resolution recovery hype
“Low resolution, high sensitivity ->
image processing = high resolution”
Reality - ameliorates low resolution
Steve Moore: “Resolution: data = target object”
Can do quick, low resolution image
* not resolution from reduced distance due to design
![Page 29: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/29.jpg)
Dual Head: Non-Uniform Sampling
![Page 30: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/30.jpg)
Activity Measurement: Attenuation
keV
150
100
80
60
50htl(140 keV) ≈ 4 cm
![Page 31: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/31.jpg)
Attenuation Correction: Simultaneous Emission (90%) and Transmission (10%)
Gd-153 rods T1/2 240 d e.c. 100% 97 keV 29% 103 keV 21%
2 heads: 900 rotation = 1800 data
![Page 32: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/32.jpg)
Semi-erect: Ameliorates Attenuation
![Page 33: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/33.jpg)
Leaning Forward, < 500 Pounds
![Page 34: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/34.jpg)
Digirad: Patient RotatesX-ray Attenuation Correction
![Page 35: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/35.jpg)
CT: Polychromatic Beam -> Dose
keV
150
100
80
60
50
![Page 36: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/36.jpg)
X-ray Tube Spectra
bremsstrahlung
characteristic X-rays
e- interaction:- ionization- deflection
X-ray tube: electrons on Tungsten or Molybdenum
![Page 37: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/37.jpg)
Digirad X-ray Source: X-rays on Lead
74W
82Pb
X-rays interaction- ionization- no 10 bremsstrahlung
![Page 38: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/38.jpg)
Digirad X-ray Spectrum
![Page 39: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/39.jpg)
New Cardiac Cameras
D-SPECT CardiArc Digirad GE
Detector CZT* NaI(Tl) CsI(Tl) CZT*
Electronics SS* PMT PD*? SS*
Pixelated Y N Y Y
Collimation holes slits*? holes pinholes
Non-uniform Y* Y* ~N Y*
Limited angle Y Y N ~N
Closer to pt Y Y Y ~N
AC N CT? CT* CT
Position ~semi semi erect supine
![Page 40: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/40.jpg)
Soft Tissue Attenuation: Supine
breast
lung
![Page 41: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/41.jpg)
Soft Tissue Attenuation: Prone
breast
![Page 42: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/42.jpg)
Soft Tissue Attenuation: Digirad Erect
breast
post
![Page 43: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/43.jpg)
Sequential Tidal-Breathing Emission and Average-Transmission Alignment
![Page 44: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/44.jpg)
Sensitivity / Resolution Trade-Off
Non-uniform sampling -> sensitivity
Special purpose design -> resolution
Advantages
Throughput at same noise
Patient motion - Hx: 1 head -> 2 head
Cost
Non-uniform sampling -> artifacts
History: 7-pinhole - failed
180o sampling - success
Sequential emission transmission
![Page 45: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/45.jpg)
What’s Wrong with FilteredBackprojection, FBP, for SPECT
Can’t model:
Attenuation
Scatter
Depth dependant resolution
New imaging geometries
(Linear shift invariant model)
![Page 46: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/46.jpg)
Solution
Iterative reconstruction
Uses:
Simultaneous linear equations
Matrix algebra
Can model image physics
(Linear model)
![Page 47: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/47.jpg)
Projections as Simultaneous Equations(Linear Model)
But, exact solution for a largenumber of equations isn’t practical
![Page 48: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/48.jpg)
Iterative Backprojection Reconstruction
Af
n
p
fn-1^ pn-1
^ en-1^
fn^
+
- x
+
f0^
r
H
H
A
object data
projection backprojection
estimate
model
error
estimate
estimateddata
estimate +backprojected
error
![Page 49: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/49.jpg)
Reconstruction, H, can be Approximate
Af
n
p
fn-1^ pn-1
^ en-1^
fn^
+
- x
+
f0^
r
H
H
A
![Page 50: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/50.jpg)
Accuracy of Model, A, is Key
Af
n
p
fn-1^ pn-1
^ en-1^
fn^
+
- x
+
f0^
r
H
H
A
^
![Page 51: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/51.jpg)
Model, A, is Well-known PhysicsProblem: Model of the Body
^
Tc-99m half-tissue layer: 4 cm
![Page 52: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/52.jpg)
Attenuation Map Gd-153 Transmission
Map adds noise to reconstructionand can introduce artifacts
![Page 53: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/53.jpg)
Iterative ReconstructionNoise is “Blobby”
![Page 54: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/54.jpg)
What’s Good About Iterative Reconstruction
Able to model:
Data collection, including new geometries
Attenuation
Scatter
Depth dependant resolution
Fairly efficient given current computers
(Iterative solution, e.g. EM, reasonable)
(OSEM is even better)
((OSEM has about 1/nsubsets of EM iterations))
![Page 55: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/55.jpg)
What’s Wrong with Iterative Reconstruction
(Complicated by ill conditioned model)
((Estimating projections not object))
Noise character bad for oncology
To model attenuation & scatter
- need to measure attenuation
- adds noise
![Page 56: Image Reconstruction from Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu](https://reader034.vdocuments.us/reader034/viewer/2022051819/551b535a550346dd1a8b6054/html5/thumbnails/56.jpg)
Conclusions
Filtered backprojection, FBP
Efficient
(Models noise)
“Easy” to understand
Iterative reconstruction, OSEM
Moderately efficient
Models noise, attenuation, scatter,
depth dependant resolution,
and new cameras