How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? ZEISS Quick Help: CIRRUS HD-OCT This guide reviews remedies for three standard error messages that may occur during the Start-Up Check process: Scanning module error, preventative maintenance error and Archive volume error. Scanning module error Message: "Errors were detected in the CIRRUS HD-OCT scanning module. You cannot acquire or review scans until the problem is corrected. Please contact Carl Zeiss Meditec Customer Service for assistance.” (Figure 1) Error detail: Click Details to open the Details Dialog box. Scroll to the bottom of the window to view the error detail (Figure 2). Figure 1 Figure 2 Error Message Error Detail

Upload: others

Post on 24-Aug-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: How do I troubleshoot Start-Up Check errors on a CIRRUS HD ...€¦ · How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT Page 2

How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? ZEISS Quick Help: CIRRUS™ HD-OCT

This guide reviews remedies for three standard error messages that may occur during the Start-Up Check process: Scanning module error, preventative maintenance error and Archive volume error.

Scanning module error

Message: "Errors were detected in the CIRRUS HD-OCT scanning module. You cannot acquire or review scans until the problem is corrected. Please contact Carl Zeiss Meditec Customer Service for assistance.” (Figure 1)

Error detail: Click Details to open the Details Dialog box. Scroll to the bottom of the window to view the error detail (Figure 2).

Figure 1

Figure 2

Error Message

Error Detail

Page 2: How do I troubleshoot Start-Up Check errors on a CIRRUS HD ...€¦ · How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT Page 2

How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT

Page 2 of 5

Troubleshoot: Disconnected controller

1. Click the Windows Logo in the left corner of the menu bar, select Control Panel and then select Network and Sharing Center. From the left-hand navigation column of the Network and Sharing Center window, click Change adapter settings (Figure 3) to reveal all existing Networking Connections — External, Internal and Wireless (Figure 4).

2. Locate the Internal Network icon (Figure 5a) and confirm that it indicates Enabled. If it indicates Disabled as shown below, right-click on the icon and select Enable (Figure 5b).

Figure 3

Figure 4

Figure 5b

Click here to see the list of Network adapters.

Figure 5a

Page 3: How do I troubleshoot Start-Up Check errors on a CIRRUS HD ...€¦ · How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT Page 2

3. Return to the Internal Network icon, right-click on the icon and select Properties from the drop-down menu (Figure 6a). From the list of connection items under the Networking tab, select Internet Protocol Version 4 (TCP/IPv4), then click Properties (Figure 6b) and verify the following Internet Protocol Properties: (Figure 7)

• "Use the following IP address" radio button should be selected – IP address is configured to 192.168.52.100 – Subnet mask is configured to 255.255.255.0

• "Use the following DNS server address" radio button should be selected – Preferred DNS server field is empty – Alternate DNS server field is empty.

Note: If the error message persists after you verify that the Internet Protocol Properties are correct, contact Carl Zeiss Meditec Technical Support at 1-800-341-6968.

How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT

Figure 6b

Page 3 of 5

Figure 7

Figure 6a

Page 4: How do I troubleshoot Start-Up Check errors on a CIRRUS HD ...€¦ · How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT Page 2

How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT

Preventive maintenance errorMessage: The instrument is prompting a notice for preventative maintenance (Figure 1).

Error Message

Troubleshoot: Preventive maintenance

This message does not indicate a problem with the CIRRUS HD-OCT instrument but rather serves as a reminder to schedule a preventative maintenance appointment for the instrument. ZEISS recommends regular preventative maintenance service. To schedule preventative maintenance service, contact Carl Zeiss Meditec Technical Support at 1-800-341-6968.

Note: Only ZEISS certified, factory-trained Carl Zeiss Meditec personnel may perform a preventative maintenance.

Figure 1

Page 4 of 5

Page 5: How do I troubleshoot Start-Up Check errors on a CIRRUS HD ...€¦ · How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT Page 2

How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT

Archive volume errorMessage: “Archive volume is not available” or “Archive volume has not been created” (Figure 1).

Note: This error is associated with the yellow Status message located near the bottom left corner of the application screen (Figure 2).

Troubleshoot: Archive volume

This message indicates that the instrument does not have a current archive, either one is not available or one has not been created.

Archive volume is not available: This message indicates that the instrument has an archive that is not set as current, or its archive location is not available. If the archive location on the network is not available, consult your Network Administrator.

Archive volume has not yet been created: This message indicates that an archive has not been created for the instrument. For assistance with the creation of an archive volume, refer to ZEISS Quick Help document How do I create a new (non-DICOM) archive on my CIRRUS HD-OCT?

Carl Zeiss Meditec AG Goeschwitzer Strasse 51–5207745 JenaGermanywww.zeiss.com/med

Carl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, CA 94568USAwww.zeiss.com/med

Refer to the CIRRUS HD-OCT User Manual Instructions for Use for safe and effective operation of the instrument.

Page 5 of 5

SER.

9124

Pr

inte

d in

the

Unite

d St

ates

. CZ

-VIII

/201

7 US

A Ed

ition

.Th

e co

nten

ts o

f the

refe

renc

e gu

ide

may

diff

er fr

om th

e cu

rrent

stat

us o

f app

rova

l of t

he p

rodu

ct o

r ser

vice

offer

ing

in y

our c

ount

ry. P

leas

e co

ntac

t our

regi

onal

repr

esen

tativ

e fo

r mor

e in

form

atio

n. S

ubje

ct to

cha

nge

in d

esig

n an

d sc

ope

of d

elive

ry

and

as a

resu

lt of

ong

oing

tech

nica

l dev

elop

men

t. CI

RRUS

is a

eith

er tr

adem

ark

or re

gist

ered

trad

emar

k of

Car

l Zei

ss M

edite

c AG

or o

ther

com

pani

es o

f the

ZEI

SS G

roup

in G

erm

any

and/

or o

ther

cou

ntrie

s. ©

Carl

Zeiss

Med

itec,

Inc.

, 201

7. A

ll rig

hts r

eser

ved.

Figure 2

Figure 1