ge healthcarem07646000a30.oct.17this drawing is based on ...mplusmedtech.com/16001/pdf/2443.pdf ·...
TRANSCRIPT
-
·
·
·
·
·
GE
Hea
lthca
re
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
AutoCAD SHX TextSITE READINESS
AutoCAD SHX TextEQUIPMENT LAYOUT
AutoCAD SHX TextSTRUCTURAL LAYOUT
AutoCAD SHX TextSTRUCTURAL DETAILS
AutoCAD SHX TextELECTRICAL LAYOUT
AutoCAD SHX TextELECTRICAL SPECIFICATIONS
AutoCAD SHX TextELECTRICAL DETAILS
AutoCAD SHX TextEQUIPMENT DETAILS
AutoCAD SHX TextC1
AutoCAD SHX TextA1
AutoCAD SHX TextS1
AutoCAD SHX TextS2
AutoCAD SHX TextE1
AutoCAD SHX TextE2
AutoCAD SHX TextE3
AutoCAD SHX TextD1
AutoCAD SHX Text(Equipment locations, heat loads, component weights, environmental specs)
AutoCAD SHX Text(Structural support/mounting locations for floor/wall/ceiling, wall support elevations)
AutoCAD SHX Text(Floor and Ceiling loading information)
AutoCAD SHX Text(Contractor supplied wiring, interconnect methods, junction point locations and descriptions)
AutoCAD SHX Text(Maximum wiring run lengths, interconnect diagram, system power specifications)
AutoCAD SHX TextDiscovery NM630
AutoCAD SHX Text5491539-1EN
AutoCAD SHX TextAny deviation from these drawings must be communicated in writing to
AutoCAD SHX Textand reviewed by your local GE Healthcare Installation Project Manager
AutoCAD SHX Textprior to making changes.
AutoCAD SHX TextMake arrangements for any rigging, special handling, or facility
AutoCAD SHX Textmodifications that must be made to deliver the equipment to the
AutoCAD SHX Textinstallation site. If desired, your local GE Healthcare Installation Project
AutoCAD SHX TextManager can supply a reference list of rigging contractors.
AutoCAD SHX TextNew construction requires the following; 1. Secure area for equipment,
AutoCAD SHX Textanalysis, 4. Restrooms.
AutoCAD SHX TextProvide for refuse removal and disposal (e.g. crates, cartons, packing)
AutoCAD SHX Text2. Power for drills and other test equipment, 3. Capability for image
AutoCAD SHX TextContact a radiation physicist or consultant to specify radiation
AutoCAD SHX Textcontainment requirements.
AutoCAD SHX TextNuclear Medicine
AutoCAD SHX TextSite Planning
AutoCAD SHX Textcssh0000-29.JAN.2013
AutoCAD SHX TextSHEET
AutoCAD SHX TextC1
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDATE:
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSITE READINESS
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextGE Healthcare
AutoCAD SHX TextCustomer Site Readiness
AutoCAD SHX TextRequirements
AutoCAD SHX TextDrawing Index
AutoCAD SHX TextA mandatory component of this drawing set is the GE Healthcare
AutoCAD SHX TextPre Installation manual. Failure to reference the Pre Installation
AutoCAD SHX Textmanual will result in incomplete documentation required for
AutoCAD SHX Textsite design and preparation.
AutoCAD SHX TextPre Installation documents for GE Healthcare products can be
AutoCAD SHX Textaccessed on the web at:
http://www.gehealthcare.com/company/docs/siteplanning.htmlAutoCAD SHX Textwww.gehealthcare.com/siteplanning
AutoCAD SHX TextThese sheets are a document set and should not be separated.
AutoCAD SHX TextElectrical information and references are contained on all sheets.
AutoCAD SHX TextGE Equipment Delivery
AutoCAD SHX TextRequirements
AutoCAD SHX TextThe items on the GE Healthcare Site Readiness Checklist are REQUIRED to facilitate equipment
AutoCAD SHX Textdelivery to the IS site. Equipment will not be delivered if these requirements are not satisfied.
AutoCAD SHX Text* REQUIRED REFERENCE *
AutoCAD SHX TextPre Installation Manual
AutoCAD SHX TextThese drawings indicate the placement and interconnection of the listed
AutoCAD SHX Textequipment components. These drawings are not construction or site
AutoCAD SHX Textpreparation drawings. Customer remains ultimately responsible for
AutoCAD SHX Textpreparing the site to accommodate the operation of such equipment in
AutoCAD SHX Textcompliance with GE Healthcare's written specifications and all applicable
AutoCAD SHX Textfederal, state, and/or local requirements.
AutoCAD SHX TextCopyright 2009 General Electric Company - Proprietary to GE
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
-
·
·
··
GE
Hea
lthca
re
AutoCAD SHX TextNUCLEAR EXAM ROOM
AutoCAD SHX TextHOT LAB
AutoCAD SHX TextDECAY CLOSET
AutoCAD SHX TextRESTROOM
AutoCAD SHX Text 1 XELERIS WORKSTATION 55 lbs 255 btu M1014AW . S
AutoCAD SHX Text1
AutoCAD SHX Text 1 6 KVA UPS 125 lbs 1959 btu B640M - UPS -
AutoCAD SHX Text2
AutoCAD SHX Text 1 PATIENT TABLE 793 lbs 682 btu B81054 - NMT -
AutoCAD SHX Text3
AutoCAD SHX Text 1 TABLE SWING PLATE FOR - -
AutoCAD SHX Text4
AutoCAD SHX Text COLLIMATOR EXCHANGE
AutoCAD SHX Text 1 LIMIT OF TABLE TRAVEL - -
AutoCAD SHX Text5
AutoCAD SHX Text6
AutoCAD SHX Text 1 OPERATORS CONSOLE ON CART. 44 lbs 255 btu B4305G - AC -
AutoCAD SHX Text7
AutoCAD SHX Text 1 DISCOVERY NM 630 GANTRY 4828 lbs 4501 btu B81053 - NMC -
AutoCAD SHX Text8
AutoCAD SHX Text B81055
AutoCAD SHX Text B0574
AutoCAD SHX Text 4 DISCOVERY NM 630 727 lbs B81041 - -
AutoCAD SHX Text9
AutoCAD SHX Text COLLIMATOR CART
AutoCAD SHX Text 1 EMO PUSHBUTTON B81058 - EMO -
AutoCAD SHX Text10
AutoCAD SHX Text 1 ESTOP PUSHBUTTON B81061 - ESTP -
AutoCAD SHX Text OPERATORS CHAIR
AutoCAD SHX Text90
AutoCAD SHX Text 1 MAIN DISCONNECT CONTROL, 13 lbs - A C
AutoCAD SHX Text50
AutoCAD SHX Text CAT. NO. E4502SV
AutoCAD SHX Text COUNTERTOP WITH DRAWERS FOR MISCELLANEOUS ITEMS.
AutoCAD SHX Text60
AutoCAD SHX Text MINIMUM DOOR OPENING FOR EQUIPMENT DELIVERY IS
AutoCAD SHX Text61
AutoCAD SHX Text 55.1" W x 86.8" H [1399mm x 2204mm], CONTINGENT
AutoCAD SHX Text ON A 96" [2438mm] CORRIDOR WIDTH NOTE:DISMOUNTED
AutoCAD SHX Text DETECTOR SHIPPING OPTION (GE CAT.NO.H2506TR),MINIMUM
AutoCAD SHX Text DOOR WIDTH IS 39.4" W [1000mm].
AutoCAD SHX Text 30-AMP, 208V DISCONNECT WITH
AutoCAD SHX Text62
AutoCAD SHX Text LOCKOUT. GEXPRO CAT. NO.
AutoCAD SHX Text GEXNM630A2LV40A65KS (OR EQUIVALENT)
AutoCAD SHX Text OPTIONAL WALL PROTECTION FROM COLLIMATOR CART. ALSO,
AutoCAD SHX Text63
AutoCAD SHX Text FINISHED FLOORING COULD BE SUBJECT TO DAMAGE DURING
AutoCAD SHX Text MOVEMENT AND BEING PARKED FOR A LONG PERIOD.
AutoCAD SHX Text SUFFICIENT FLOORING MUST BE USED TO PREVENT DAMAGE.
AutoCAD SHX Texto GANTRY MUST BE LOCATED IN AMBIENT STATIC MAGNETIC FIELDS OF LESS
AutoCAD SHX TextTHAN ONE GAUSS TO GUARANTEE SPECIFIED IMAGING PERFORMANCE. AMBIENT
AutoCAD SHX TextAC MAGNETIC FIELDS MUST BE BELOW 0.01 GAUSS PEAK.
AutoCAD SHX Texto COMPUTER EQUIPMENT MUST BE LOCATED IN AMBIENT STATIC MAGNETIC
AutoCAD SHX TextFIELDS OF LESS THAN TEN GAUSS TO GUARANTEE DATA INTEGRITY.
AutoCAD SHX Texto MULTIFORMAT CAMERA EQUIPMENT MUST BE LOCATED IN AMBIENT STATIC MAGNETIC
AutoCAD SHX TextFIELDS OF LESS THAN THREE GAUSS TO OBTAIN SPECIFIED GEOMETRIC LINEARITY.
AutoCAD SHX Texto CONSOLE EQUIPMENT MUST BE LOCATED IN AMBIENT STATIC MAGNETIC FIELDS
AutoCAD SHX TextOF LESS THAN TEN GAUSS TO OBTAIN SPECIFIED GEOMETRIC LINEARITY.
AutoCAD SHX Text INCLUDING FOR EXAMPLE OVERNIGHT, WEEKENDS, AND HOLIDAYS.
AutoCAD SHX Texto ENVIRONMENTAL CONDITIONS LISTED ABOVE MUST BE MAINTAINED AT ALL TIMES
AutoCAD SHX Text ALLOWABLE TEMPERATURE CHANGE OF 5%%D F (3%%D C)/HOUR, MAXIMUM ROOM
AutoCAD SHX Text ABOVE RESTRICTIONS ARE NOT EXCEEDED.
AutoCAD SHX Texto AMBIENT OPERATING TEMPERATURE: 64%%D TO 79%%D F, (18%%D TO 26%%D C) MAXIMUM
AutoCAD SHX Texto HUMIDITY: 30 TO 60 PERCENT NON-CONDENSING, MAXIMUM ALLOWABLE CHANGE OF
AutoCAD SHX Texto ALTITUDE: NOT TO EXCEED 9843 FT. (3000M) ABOVE SEA LEVEL.
AutoCAD SHX Texto THE ENVIRONMENT FOR THE ELECTRONICS CABINET MUST BE CONTROLLED SO THE
AutoCAD SHX Texto DO NOT RESTRICT THE AIR INTAKE OR AIR EXHAUST OF THE SYSTEM COMPONENTS.
AutoCAD SHX Text TEMPERATURE GRADIENT 5%%DF, (3%%D C).
AutoCAD SHX Text 5 PERCENT/HOUR.
AutoCAD SHX TextLICENSING MUST BE ARRANGED BY THE CUSTOMER EARLY IN THE PLANNING PROCESS AND THEN
AutoCAD SHX TextNECESSARY FOR EQUIPMENT CALIBRATION. APPROPRIATE REGULATORY COMPLIANCE AND
AutoCAD SHX TextTHIS EQUIPMENT INVOLVES THE USE OF RADIOACTIVE ISOTOPES, INCLUDING THOSE SOURCES
AutoCAD SHX TextIMPORTANT CUSTOMER READINESS ALERT:
AutoCAD SHX TextDEMONSTRATED/AVAILABLE FOR EQUIPMENT INSTALLATION.
AutoCAD SHX TextASSIGNED BY THE HOSPITAL NET ADMIN IF CONNECTING TO THE HOSPITAL LAN
AutoCAD SHX TextHOSTNAME
AutoCAD SHX TextIP
AutoCAD SHX TextAE TITLE
AutoCAD SHX TextDICOM PORT
AutoCAD SHX TextACQUISITION HOST
AutoCAD SHX TextPROCESSING HOST
AutoCAD SHX TextHARDCOPY HOST
AutoCAD SHX TextLAN NET MASK
AutoCAD SHX TextGATEWAY TO OTHER NETWORKS
AutoCAD SHX TextOTHER
AutoCAD SHX TextHUB OR SWITCH
AutoCAD SHX TextPREPARE ADEQUATE NETWORK SOCKETS IN THE PROPER LOCATIONS TO SUPPORT ALL ACQUISITION, LOCAL AND REMOTE WORKSTATION. IT DEPARTMENT MUST ASSIGN DEDICATED IP ADDRESSES (NOT DHCP) NOTE THE ADDRESSES BELOW FOR THE ACQUISITION, LOCAL AND REMOTE WORKSTATIONS. PREPARE BROADBAND CONNECTIVITY LINE AND DEDICATED IP ADDRESSES FOR INSITE CONNECTIVITY. REFER TO TABLE ON A1 PAGE
AutoCAD SHX Text1
AutoCAD SHX Text2
AutoCAD SHX Text3
AutoCAD SHX Text4
AutoCAD SHX Text5
AutoCAD SHX Text7
AutoCAD SHX Text8
AutoCAD SHX Text9
AutoCAD SHX Text10
AutoCAD SHX Text6
AutoCAD SHX Text90
AutoCAD SHX Text50
AutoCAD SHX Text60
AutoCAD SHX Text61
AutoCAD SHX Text62
AutoCAD SHX Text63
AutoCAD SHX Text60
AutoCAD SHX TextEQUIPMENT LAYOUT
AutoCAD SHX TextEQUIPMENT LAYOUT
AutoCAD SHX TextGE EQUIPMENT LISTING
AutoCAD SHX TextARE TO BE INSTALLED BY THE CUSTOMER OR HIS CONTRACTOR.
AutoCAD SHX TextTHE FOLLOWING ITEMS, WHICH HAVE BEEN ORDERED FROM GE HEALTHCARE,
AutoCAD SHX TextPLAN
AutoCAD SHX TextSTRC
AutoCAD SHX Text(* = EXISTING/REINSTALL)
AutoCAD SHX TextITEM DESCRIPTION
AutoCAD SHX TextREFER TO SHEET "D"
AutoCAD SHX TextNO.
AutoCAD SHX TextITEM
AutoCAD SHX TextWEIGHT
AutoCAD SHX TextQUANTITY ORDERED
AutoCAD SHX TextEQUIPMENT CROSS
AutoCAD SHX TextREFERENCE CHART
AutoCAD SHX TextP = PREAPPROVAL
AutoCAD SHX TextC = CALCULATIONS/
AutoCAD SHX TextS = SPECIFICATIONS
AutoCAD SHX TextSEISMIC
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextMAGNETIC INTERFERENCE SPECIFICATIONS
AutoCAD SHX TextSITE ENVIRONMENT SPECIFICATIONS
AutoCAD SHX TextGENERAL SPECIFICATIONS
AutoCAD SHX TextANCILLARY ITEMS
AutoCAD SHX Text(* INDICATES EXISTING)
AutoCAD SHX TextITEM DESCRIPTION
AutoCAD SHX TextITEMS
AutoCAD SHX TextCUSTOMER/CONTRACTOR SUPPLIED AND INSTALLED
AutoCAD SHX TextNO.
AutoCAD SHX TextITEM
AutoCAD SHX TextBE INSTALLED BY OTHERS.
AutoCAD SHX TextNOTE: LOCAL CONDITIONS MAY DICTATE THAT ITEMS IDENTIFIED IN THIS CATEGORY
AutoCAD SHX TextTHE FOLLOWING ITEMS ARE AVAILABLE FROM GE HEALTHCARE
AutoCAD SHX TextTECHNOLOGIES. CONTACT YOUR LOCAL GE HEALTHCARE
AutoCAD SHX TextSERVICE REPRESENTATIVE FOR PRICING AND AVAILABILITY.
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextA1
AutoCAD SHX TextHEAT OUTPUT
AutoCAD SHX TextDETAIL
AutoCAD SHX TextNO.
AutoCAD SHX Text ONLY
AutoCAD SHX TextSTATUS
AutoCAD SHX TextELEC
AutoCAD SHX TextPLAN
AutoCAD SHX TextPER
AutoCAD SHX TextEQUIPMENT ON ORDER FROM GE HEALTHCARE, INSTALLED BY GE HEALTHCARE,
AutoCAD SHX TextSHEET
AutoCAD SHX TextNFSH-1002
AutoCAD SHX Text PENDING APPROVAL
AutoCAD SHX Texto THE REQUIRED CEILING HEIGHT INDICATED ON THESE PLANS IS TO ENSURE EQUIPMENT
AutoCAD SHX Text FUNCTION IS NOT INHIBITED. CONSULT WITH YOUR LOCAL GEHC IS SPECIALIST
AutoCAD SHX Text REGARDING ACCEPTABILITY OF OTHER CEILING HEIGHTS.
AutoCAD SHX Texto CHECK ALL DOOR OPENINGS AND HALLWAYS FROM DELIVERY LOCATION TO WHERE
AutoCAD SHX Text EQUIPMENT IS TO BE INSTALLED TO ENSURE THE ROUTE PHYSICALLY AND STRUCTURALLY
AutoCAD SHX Texto RADIATION PROTECTION REQUIREMENTS ARE NOT INDICATED ON THIS PLAN. WHERE
AutoCAD SHX Text NEEDED PER NATIONAL OR LOCAL CODE THEY SHALL BE SPECIFIED BY A QUALIFIED
AutoCAD SHX Texto THE DEVELOPMENT OF THE EQUIPMENT LAYOUT, ROOM DIMENSIONS, MECHANICAL AND
AutoCAD SHX Text ELECTRICAL SUGGESTIONS IS PREDICATED UPON THE BEST INFORMATION OBTAINABLE
AutoCAD SHX Text FROM THE SITE, COUPLED WITH THE CUSTOMER'S KNOWN DESIRES. ARCHITECTURAL OR
AutoCAD SHX Text ELECTRICAL CHANGES INCLUDING RELOCATION OF EQUIPMENT ILLUSTRATED ON THIS
AutoCAD SHX Text DRAWING IS ALLOWED ONLY WITH NOTIFICATION, IN WRITING, AND REVIEW BY GEHC
AutoCAD SHX Text SERVICE DEPARTMENT. EQUIPMENT OPERATION, SERVICEABILITY, AND RESTRICTING CABLE
AutoCAD SHX Text LENGTHS, ETC., MAKE THIS ESSENTIAL FOR A PROPER IS. GEHC RESERVES
AutoCAD SHX Text THE RIGHT TO MAKE ON THE JOB CHANGES BECAUSE OF CUSTOMER REQUIREMENTS
AutoCAD SHX Text AND/OR OBSTACLES IN CONSTRUCTION, ETC..
AutoCAD SHX Texto ALL WORK TO BE IN COMPLIANCE WITH NATIONAL AND LOCAL BUILDING SAFETY CODES.
AutoCAD SHX Texto DIMENSIONS ARE TO FINISHED SURFACES OF ROOM
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text WILL ACCOMODATE THE EQUIPMENT AS SHIPPED.
AutoCAD SHX Text RADIOLOGICAL PHYSICIST.
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX Text(PER HOUR)
AutoCAD SHX TextThis equipment layout indicates the placement and interconnection of the indicated equipment components. There may be federal, state, and/or local requirements that could impact the placement
AutoCAD SHX Textof these components. It remains the Customer's responsibility for ensuring the site and final equipment placement complies with all applicable federal, state, and/or local requirements.
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
AutoCAD SHX TextSCALE: 1/4" = 1'-0"
AutoCAD SHX TextGON 4566821 DATED 13.Oct.17
AutoCAD SHX TextEXISTING CEILING HEIGHT = 8'-0"
AutoCAD SHX TextGE Project Manager:
AutoCAD SHX TextTelephone:
AutoCAD SHX TextTHE GE HPI TECHNICAL SUPPORT GROUP IS AN ADDITIONAL RESOURCE THAT CAN PROVIDE ANSWERS FOR GENERAL GE PRODUCT SITING QUESTIONS AND CAN BE REACHED AT (877)-305-9677 OR MAILTO:[email protected] [email protected]
AutoCAD SHX TextBRIAN FERRELL
AutoCAD SHX Text337-371-7202
AutoCAD SHX Text(LOCATED ELSEWHERE)
AutoCAD SHX Text 1 48" TABLE - -
AutoCAD SHX Text11
AutoCAD SHX Text11
-
GE
Hea
lthca
re
AutoCAD SHX TextNUCLEAR EXAM ROOM
AutoCAD SHX TextHOT LAB
AutoCAD SHX TextDECAY CLOSET
AutoCAD SHX TextRESTROOM
AutoCAD SHX Text FLOOR REQUIREMENTS:
AutoCAD SHX Text1
AutoCAD SHX Text FLOOR LEVELNESS IN THE EXAM ROOM MUST BE LEVEL WITHIN
AutoCAD SHX Text 1/8 IN. [3 MM] OVER 120 IN. [3048 MM]. FLOOR FLATNESS
AutoCAD SHX Text IN THE EXAM ROOM MUST HAVE NO DEVIATIONS GREATER THAN
AutoCAD SHX Text 3/16" [0.5 CM] OVER 60 IN. [150 CM]. REFER TO FULL SIZE
AutoCAD SHX Text FLOOR TEMPLATE FOR ALL DRILLING/FLOOR PENETRATION
AutoCAD SHX Text REQUIREMENTS IN ORDER TO ENABLE MOUNTING OF SYSTEM
AutoCAD SHX Text FLOOR ANCHORS. CONCRETE MUST HAVE A MINIMUM CURE
AutoCAD SHX Text STRENGTH OF fc = 4350 psi (30 MPa) at 28 DAYS (CURING
AutoCAD SHX Text TIME) FOR 25/30 CONCRETE, AND MUST BE AT LEAST 140mm
AutoCAD SHX Text (5.5") THICK. IT IS THE CUSTOMER'S RESPONSIBILITY TO
AutoCAD SHX Text HAVE APPROPRIATE TESTS PERFORMED TO DETERMINE AND
AutoCAD SHX Text MEASURE CONCRETE STRENGTH. IF THE DISCOVERY NM 630
AutoCAD SHX Text SYSTEM IS INSTALLED ON A FLOOR TYPE THINNER THAN A
AutoCAD SHX Text 140mm (5.5") CONCRETE FLOOR, THE CUSTOMER SHALL, AT ITS
AutoCAD SHX Text EXPENSE, PROVIDE ACCEPTABLE ANCHORING AND MOUNTING
AutoCAD SHX Text METHODS THAT MEET ALL STRUCTURAL SPECIFICATIONS
AutoCAD SHX Text PROVIDED IN THE PRE-INSTALLATION MANUAL.
AutoCAD SHX Text VIBRATION REQUIREMENTS:
AutoCAD SHX Text THE MAXIMUM STEADY STATE VIBRATION TRANSMITTED THROUGH
AutoCAD SHX Text THE FLOOR SHOULD NOT EXCEED 0.001 PER METER SQUARED RMS
AutoCAD SHX Text MAXIMUM SINGLE FREQUENCY ABOVE AMBIENT BASELINE FROM
AutoCAD SHX Text 0.5 TO 80 Hz (MEASURED IN ANY 1 HOUR DURING A NORMAL
AutoCAD SHX Text OPERATING PERIOD). THE BEHAVIORAL CHARACTERISTICS MUST
AutoCAD SHX Text BE SUCH THAT ANY MEASURABLE TRANSIENT DISTURBANCE MUST
AutoCAD SHX Text ALSO BE MINIMIZED TO LESS THAN 0.01 PER METER SQUARED
AutoCAD SHX Text PEAK-TO-PEAK.
AutoCAD SHX Text1
AutoCAD SHX TextTYPICAL WALL SUPPORT ELEVATIONS
AutoCAD SHX TextSTRUCTURAL SUPPORT METHODS
AutoCAD SHX Text(* INDICATES EXISTING)
AutoCAD SHX TextITEM DESCRIPTION
AutoCAD SHX TextITEMS
AutoCAD SHX TextCUSTOMER/CONTRACTOR SUPPLIED AND INSTALLED
AutoCAD SHX TextNO.
AutoCAD SHX TextITEM
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextSTRUCTURAL LAYOUT
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextSTRUCTURAL LAYOUT
AutoCAD SHX TextS1
AutoCAD SHX TextSHEET
AutoCAD SHX TextNFSH-1002
AutoCAD SHX TextSTRUCTURAL NOTES
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX Texto ALL UNITS THAT ARE WALL MOUNTED OR WALL SUPPORTED ARE TO BE PROVIDED WITH
AutoCAD SHX Text SUPPORTS WHERE NECESSARY. WALL SUPPORTS ARE TO BE SUPPLIED AND INSTALLED BY
AutoCAD SHX Text THE CUSTOMER OR HIS CONTRACTORS. SEE PLAN AND DETAIL SHEETS FOR SUGGESTED
AutoCAD SHX Text LOCATIONS AND MOUNTING HOLE LOCATIONS.
AutoCAD SHX Texto FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST BE LEVEL TO
AutoCAD SHX Text SPECIFICATIONS. (IF NOT SPECIFIED ELSEWHERE ON THIS SHEET THE FLOOR LEVELNESS
AutoCAD SHX Texto DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
AutoCAD SHX Texto FOR SEISMIC REGIONS ENSURE SUPPORTS SPAN THREE MEMBERS.
AutoCAD SHX Text SHOULD BE 1/8 IN. [3 MM] IN 10 FT. [3.05 M].
AutoCAD SHX Text DOCUMENTS FOR STANDARD ANCHORING METHODS ARE INCLUDED WITH GE EQUIPMENT
AutoCAD SHX Text DRAWINGS FOR GEOGRAPHIC AREAS THAT REQUIRE SUCH DOCUMENTATION.
AutoCAD SHX Texto CUSTOMERS CONTRACTOR MUST PROVIDE AND INSTALL ANY NON-STANDARD ANCHORING.
AutoCAD SHX Texto CUSTOMERS CONTRACTOR MUST PROVIDE ALL PENETRATIONS IN POST TENSION FLOORS.
AutoCAD SHX Text OBSTRUCTION ENCOUNTERED WHILE DRILLING BY THE GE INSTALLER SUCH AS REBAR ETC.
AutoCAD SHX Text MUST ALSO PROVIDE FLOOR DRILLING THAT CANNOT BE COMPLETED BECAUSE OF AN
AutoCAD SHX Texto CUSTOMERS CONTRACTOR MUST PROVIDE AND INSTALL HARDWARE FOR "THROUGH THE
AutoCAD SHX Text FLOOR" ANCHORING AND/OR ANY BRACING UNDER ACCESS FLOORS. THIS CONTRACTOR
AutoCAD SHX Texto IT IS THE CUSTOMER'S RESPONSIBILITY TO PERFORM ANY FLOOR OR WALL
AutoCAD SHX TextPENETRATIONS THAT MAY BE REQUIRED. THE CUSTOMER IS ALSO RESPONSIBLE FOR
AutoCAD SHX TextENSURING THAT NO SUBSURFACE UTILITIES (E.G., ELECTRICAL OR ANY OTHER FORM OF
AutoCAD SHX TextWIRING, CONDUITS, PIPING, DUCT WORK OR STRUCTURAL SUPPORTS (I.E. POST TENSION
AutoCAD SHX TextCABLES OR REBAR)) WILL INTERFERE OR COME IN CONTACT WITH SUBSURFACE
AutoCAD SHX TextPENETRATION OPERATIONS (E.G. DRILLING AND INSTALLATION OF ANCHORS/SCREWS)
AutoCAD SHX TextPERFORMED DURING THE INSTALLATION PROCESS. TO ENSURE WORKER SAFETY, GE
AutoCAD SHX TextINSTALLERS WILL PERFORM SURFACE PENETRATION OPERATIONS ONLY AFTER THE
AutoCAD SHX TextCUSTOMER'S VALIDATION AND COMPLETION OF THE "GE SURFACE PENETRATION PERMIT"
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
AutoCAD SHX TextEXISTING CEILING HEIGHT = 8'-0"
AutoCAD SHX TextSCALE: 1/4" = 1'-0"
AutoCAD SHX TextGE Project Manager:
AutoCAD SHX TextTelephone:
AutoCAD SHX TextTHE GE HPI TECHNICAL SUPPORT GROUP IS AN ADDITIONAL RESOURCE THAT CAN PROVIDE ANSWERS FOR GENERAL GE PRODUCT SITING QUESTIONS AND CAN BE REACHED AT (877)-305-9677 OR MAILTO:[email protected] [email protected]
AutoCAD SHX TextBRIAN FERRELL
AutoCAD SHX Text337-371-7202
-
GE
Hea
lthca
re
AutoCAD SHX Text DISCOVERY NM 630/BRIVO NM 615 GANTRY AND TABLE ANCHOR/LEVELING
AutoCAD SHX TextREV. DATE: 15.JAN.13
AutoCAD SHX TextOM630A
AutoCAD SHX TextTABLE CENTER OF GRAVITY 390KG [859.80 lbs]
AutoCAD SHX TextNM GANTRY CENTER OF GRAVITY 2190KG [4828.12 lbs]
AutoCAD SHX TextNM GANTRY FRONT PADS 845KG [1862.91 lbs] LOAD PER PAD
AutoCAD SHX TextNM GANTRY REAR PADS 250KG [551.15 lbs] LOAD PER PAD
AutoCAD SHX TextNM GANTRY MAIN ANCHORING
AutoCAD SHX TextMEASUREMENT LEGEND FEET AND [MILLIMETERS]
AutoCAD SHX Text DISCOVERY NM 630/BRIVO NM 615 GANTRY AND TABLE ANCHOR/LEVELING
AutoCAD SHX TextREV. DATE: 01.DEC.16
AutoCAD SHX TextOM630B
AutoCAD SHX TextDRILL
AutoCAD SHX TextHOLE
AutoCAD SHX TextHOLE
AutoCAD SHX TextDEPTH
AutoCAD SHX TextDRILLING
AutoCAD SHX TextMETHOD
AutoCAD SHX TextANCHOR
AutoCAD SHX TextTYPE
AutoCAD SHX TextTORQUE,
AutoCAD SHX TextNm
AutoCAD SHX TextSECTION
AutoCAD SHX TextFIG 2-12
AutoCAD SHX Text35
AutoCAD SHX TextFIG 2-10
AutoCAD SHX TextHOLE PURPOSE
AutoCAD SHX TextMAIN ANCHOR
AutoCAD SHX TextALTERNATIVE ANCHOR
AutoCAD SHX TextNo.
AutoCAD SHX TextX
AutoCAD SHX TextY
AutoCAD SHX Text1
AutoCAD SHX Text1
AutoCAD SHX Text-0.00
AutoCAD SHX Text405.00
AutoCAD SHX Text2
AutoCAD SHX Text2
AutoCAD SHX Text-0.00
AutoCAD SHX Text-405.00
AutoCAD SHX Text3
AutoCAD SHX Text3
AutoCAD SHX Text-810.00
AutoCAD SHX Text405.00
AutoCAD SHX Text4
AutoCAD SHX Text4
AutoCAD SHX Text-810.00
AutoCAD SHX Text-405.00
AutoCAD SHX Text5
AutoCAD SHX Text5
AutoCAD SHX Text-80.00
AutoCAD SHX Text365.00
AutoCAD SHX Text6
AutoCAD SHX Text6
AutoCAD SHX Text-80.00
AutoCAD SHX Text-365.00
AutoCAD SHX Text7
AutoCAD SHX Text7
AutoCAD SHX Text-810.00
AutoCAD SHX Text530.00
AutoCAD SHX Text8
AutoCAD SHX Text8
AutoCAD SHX Text-810.00
AutoCAD SHX Text-530.00
AutoCAD SHX Text90.00
AutoCAD SHX TextANCHORED PART
AutoCAD SHX TextNM GANTRY
AutoCAD SHX TextPOCKET FOR COLLIMATOR CART PIN
AutoCAD SHX Text 3" [76]DEPTH 1.3" [35]
AutoCAD SHX TextNM FOV
AutoCAD SHX TextCOORDINATES TABLE ORIGIN
AutoCAD SHX TextNM GANTRY
AutoCAD SHX TextHILTI HSL-3-G M10/40
AutoCAD SHX TextNM GANTRY
AutoCAD SHX TextUSE METAL DRILLING TEMPLATE
AutoCAD SHX TextLEVELING AND FLATTENING FLOOR AREA DETAIL
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextSUPPORT DETAIL
AutoCAD SHX TextB3000X
AutoCAD SHX TextREV. DATE: 04.May.2012
AutoCAD SHX TextLEVEL BUBBLE
AutoCAD SHX TextFLOOR
AutoCAD SHX TextSURFACE FLATNESS
AutoCAD SHX TextMEASUREMENT DIRECTIONS
AutoCAD SHX Text SELF LEVELING LASER
AutoCAD SHX TextLIGHT
AutoCAD SHX TextREFERENCE
AutoCAD SHX TextPOINT
AutoCAD SHX TextRULER
AutoCAD SHX TextCONCRETE FLOOR
AutoCAD SHX TextSLOPE MEASUREMENTS
AutoCAD SHX TextCUSTOMER/CONTRACTOR IS RESPONSIBLE FOR MEASURING FLOOR LEVELNESS/FLATNESS AND PROVIDING THE RESULTS TO A GEHC REPRESENTATIVE (FE OR PM) BEFORE FINISHED FLOORING IS INSTALLED. SEE S1 PAGE AND SPECIFICATIONS FROM PREINSTALLATION MANUAL FOR MORE INFORMATION. FLOOR REQUIREMENTS ARE CRITICAL TO INSTALLATION/OPERATION OF SYSTEM!
AutoCAD SHX TextSLOPE OUT OF
AutoCAD SHX TextACCEPTABLE LIMIT
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextSTRUCTURAL DETAILS
AutoCAD SHX TextS2
AutoCAD SHX TextSHEET
AutoCAD SHX TextNFSH-1002
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
-
GE
Hea
lthca
re
AutoCAD SHX TextONE 4" CND. BELOW FLOOR
AutoCAD SHX TextONE 2" CND. BELOW FLOOR
AutoCAD SHX TextONE 3/4" CND. BELOW FLOOR
AutoCAD SHX TextA
AutoCAD SHX TextB
AutoCAD SHX TextA
AutoCAD SHX TextA2
AutoCAD SHX TextFINISHED CEILING
AutoCAD SHX TextFINISHED FLOOR
AutoCAD SHX TextA
AutoCAD SHX TextUPS
AutoCAD SHX TextB
AutoCAD SHX TextFINISHED CEILING
AutoCAD SHX TextFINISHED FLOOR
AutoCAD SHX TextEMO
AutoCAD SHX TextNUCLEAR EXAM ROOM
AutoCAD SHX TextHOT LAB
AutoCAD SHX TextDECAY CLOSET
AutoCAD SHX TextRESTROOM
AutoCAD SHX TextA MAIN DISCONNECT 1 PANEL - INCLUDED IN ORDER
AutoCAD SHX TextA2 MAIN DISCONNECT 1 30-AMP 208V DISCONNECT WITH
AutoCAD SHX Text AVAILABLE FROM LOCKOUT. GEXPRO CAT. NO.
AutoCAD SHX Text GEXPRO, CALL: GEXNM630A2LV40A65KS
AutoCAD SHX Text 800-200-9760 (OR EQUIVALENT)
AutoCAD SHX Text EXTENSION 3825 OR
AutoCAD SHX Text LOCAL GE
AutoCAD SHX Text PROJECT MANAGER
AutoCAD SHX TextAC ACQUISITION MOBILE 1 EXTERNALLY CONNECTED
AutoCAD SHX Text CART
AutoCAD SHX TextEMO EMO PUSHBUTTON 1 COVERPLATE ELEC-8
AutoCAD SHX Text 1 4 X 4 X 4 IN BOX
AutoCAD SHX Text 1 3/4 IN. DIA CHASE NIPPLE
AutoCAD SHX TextESTP ESTOP 1 SAME ROUTING AS 'AC'
AutoCAD SHX Text PUSHBUTTON
AutoCAD SHX TextNMC NUCLEAR MEDICINE 1 SUITABLE BUSHING & LOCKNUT ELEC-9
AutoCAD SHX Text CAMERA
AutoCAD SHX TextNMT PATIENT TABLE 1 OPENING IN CONDUIT MUST BE CUT
AutoCAD SHX Text FLUSH WITH FINISHED FLOOR
AutoCAD SHX TextUPS UPS CABINET 1 COVERPLATE ELEC-8
AutoCAD SHX Text 1 2 1/2 IN. DIA. CHASE NIPPLE
AutoCAD SHX Text 1 6 X 6 X 4 IN. BOX
AutoCAD SHX Text250
AutoCAD SHX Text100
AutoCAD SHX Text150
AutoCAD SHX Text200
AutoCAD SHX Text50
AutoCAD SHX TextLENGTH
AutoCAD SHX TextIN FEET
AutoCAD SHX TextRUN
AutoCAD SHX Text173 to 250 VAC 50 to 60hz
AutoCAD SHX TextREV. DATE: 03.May.16
AutoCAD SHX Text3
AutoCAD SHX Text208 VAC
AutoCAD SHX Text6
AutoCAD SHX Text10
AutoCAD SHX Text6
AutoCAD SHX Text8
AutoCAD SHX Text240 VAC
AutoCAD SHX TextFEEDER TABLE - Discovery NM 630
AutoCAD SHX TextPOWER SUPPLY VOLTAGE
AutoCAD SHX Text4
AutoCAD SHX Text8
AutoCAD SHX Text8
AutoCAD SHX Text10
AutoCAD SHX Text4
AutoCAD SHX Texto RECOMMENDED FEEDER SIZES FROM POWER SOURCE TO MAIN DISCONNECT.
AutoCAD SHX Text GROUNDING POINT AND ALWAYS TRAVEL IN THE SAME CONDUIT WITH THE FEEDERS
AutoCAD SHX Text GROUND WILL RUN FROM THE EQUIPMENT BACK TO THE FACILITY POWER SOURCE/MAIN
AutoCAD SHX Texto THE GROUNDING CONDUCTOR WILL BE THE SAME SIZE AS THE POWER FEEDER. THIS
AutoCAD SHX Texto CALCULATIONS BASED UPON NOMINAL VOLTAGE, WIRE SIZE IN AWG.
AutoCAD SHX Text AND NEUTRAL.
AutoCAD SHX TextSIZE OF FEEDERS AND GROUND WIRES (AWG)
AutoCAD SHX TextUPS > A 2-BLACK, 1-GREEN (REFER TO FEEDER TABLE FOR SIZE)
AutoCAD SHX TextA2 > UPS 2-BLACK, 1-WHITE, 1-GREEN (SIZE AS REQUIRED)
AutoCAD SHX Text1 PHASE > A2 2-BLACK, 1-WHITE, 1-GREEN (SIZE AS REQUIRED)
AutoCAD SHX TextELECTRICAL OUTLET LEGEND
AutoCAD SHX TextCUSTOMER/CONTRACTOR SUPPLIED AND INSTALLED
AutoCAD SHX TextITEMS. HEIGHT ABOVE FLOOR DETERMINED BY
AutoCAD SHX TextLOCAL CODES UNLESS OTHERWISE SPECIFIED.
AutoCAD SHX TextDUPLEX HOSPITAL GRADE, DEDICATED OUTLET 120-V, SINGLE PHASE OUTLET 20 AMP
AutoCAD SHX TextDEDICATED TELEPHONE LINE(S) (SEE ELECTRICAL DETAIL ELEC-1 OR ELEC-67)
AutoCAD SHX TextNETWORK OUTLET (SEE ELECTRICAL DETAILS ELEC-83 AND ELEC-84 OR ELEC-87)
AutoCAD SHX TextA
AutoCAD SHX TextA2
AutoCAD SHX TextAC
AutoCAD SHX TextEMO
AutoCAD SHX TextESTP
AutoCAD SHX TextNMC
AutoCAD SHX TextNMT
AutoCAD SHX TextUPS
AutoCAD SHX TextREV DATE:15.APR.14
AutoCAD SHX TextADDITIONAL CONDUIT RUNS FOR
AutoCAD SHX TextDISCOVERY NM630/BRIVO NM615
AutoCAD SHX Text(BY CONTRACTOR)
AutoCAD SHX TextCONDUITS REQUIRED FOR BASE SYSTEM
AutoCAD SHX Text(CONDUITS ARE LOCATED ABOVE CEILING)
AutoCAD SHX TextA2
AutoCAD SHX TextTO
AutoCAD SHX TextCONDUIT AS REQ'D
AutoCAD SHX TextUPS
AutoCAD SHX TextA2
AutoCAD SHX TextTO
AutoCAD SHX TextCONDUIT AS REQ'D
AutoCAD SHX Text208-V 1 30 AMP POWER
AutoCAD SHX TextUPS
AutoCAD SHX TextTO
AutoCAD SHX TextA
AutoCAD SHX TextCND AS REQ'D REFER
AutoCAD SHX TextSIZE PER FEEDER TABLE
AutoCAD SHX TextNOTE: SEE E2 PAGE FOR MAXIMUM RUN LENGTHS
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextELECTRICAL LAYOUT
AutoCAD SHX TextINSTALLED BY THE CUSTOMER'S ELECTRICAL CONTRACTOR
AutoCAD SHX TextTHE FOLLOWING MATERIALS ARE TO BE SUPPLIED AND
AutoCAD SHX TextJUNCTION POINT DESCRIPTIONS
AutoCAD SHX TextELECTRICAL PLAN
AutoCAD SHX TextHARDWARE
AutoCAD SHX TextDESCRIPTION
AutoCAD SHX TextPOINT
AutoCAD SHX TextQTY.
AutoCAD SHX TextCONTRACTOR SUPPLIED AND INSTALLED WIRING
AutoCAD SHX TextQUANTITY, WIRE SIZE/COLOR
AutoCAD SHX TextWIRE RUN, FROM - TO
AutoCAD SHX TextDETAIL NO., SHT. E3
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextE1
AutoCAD SHX TextSHEET
AutoCAD SHX TextNFSH-1002
AutoCAD SHX TextELECTRICAL CONTRACTOR SHALL RING OUT, TAG AND TERMINATE ALL WIRES AT BOTH ENDS.
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX Text BE SUPPLIED AND INSTALLED BY CUSTOMERS ELECTRICAL CONTRACTOR.
AutoCAD SHX Text FLOOR AS POSSIBLE TO REDUCE RUN LENGTH.
AutoCAD SHX Texto ALL JUNCTION BOXES, CONDUIT, DUCT, DUCT DIVIDERS, SWITCHES, CIRCUIT BREAKERS, CABLE TRAY, ETC., ARE TO
AutoCAD SHX Texto CONDUIT AND DUCT RUNS SHALL HAVE SWEEP RADIUS BENDS
AutoCAD SHX Texto CONDUITS AND DUCT ABOVE CEILING OR BELOW FINISHED FLOOR MUST BE INSTALLED AS NEAR TO CEILING OR
AutoCAD SHX TextJUNCTION POINT NOTES
AutoCAD SHX Texto ALL DUCTWORK MUST MEET THE FOLLOWING REQUIREMENTS:
AutoCAD SHX Text 1. DUCTWORK SHALL BE METAL WITH DIVIDERS AND HAVE REMOVABLE, ACCESSIBLE COVERS.
AutoCAD SHX Texto ALL OPENINGS IN ACCESS FLOORING ARE TO BE CUT OUT AND FINISHED OFF WITH GROMMET MATERIAL BY THE
AutoCAD SHX Text 2. DUCTWORK SHALL BE CERTIFIED/RATED FOR ELECTRICAL POWER PURPOSES.
AutoCAD SHX Text 3. DUCTWORK SHALL BE ELECTRICALLY AND MECHANICALLY BONDED TOGETHER IN AN APPROVED MANNER.
AutoCAD SHX Text 4. PVC AS A SUBSTITUTE MUST BE USED IN ACCORDANCE WITH ALL LOCAL AND NATIONAL CODES.
AutoCAD SHX Text CUSTOMERS CONTRACTOR.
AutoCAD SHX Texto GENERAL CONTRACTOR TO INSERT PULL CORDS FOR ALL CABLE RUN CONDUITS BETWEEN THE
AutoCAD SHX TextEQUIPMENT ROOM AND THE OPERATORS CONTROL ROOM.
AutoCAD SHX Texto CEILING MOUNTED JUNCTION BOXES ILLUSTRATED ON THIS PLAN MUST BE INSTALLED FLUSH WITH FINISHED CEILING.
AutoCAD SHX Texto 10 FOOT PIGTAILS AT ALL JUNCTION POINTS.
AutoCAD SHX Texto ALL WIRING MUST BE THHN OR TFFN STRANDED COPPER THERMOPLASTIC 600 VOLT OR EQUIVALENT INSULATION.
AutoCAD SHX Text ALUMINUM OR SOLID WIRES ARE NOT ALLOWED.
AutoCAD SHX Texto GROUNDING IS CRITICAL TO EQUIPMENT FUNCTION AND PATIENT SAFETY. SITE MUST CONFORM TO WIRING
AutoCAD SHX Text SPECIFICATIONS SHOWN ON THIS PLAN.
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
AutoCAD SHX TextEXISTING CEILING HEIGHT = 8'-0"
AutoCAD SHX TextSCALE: 1/4" = 1'-0"
AutoCAD SHX TextGE Project Manager:
AutoCAD SHX TextTelephone:
AutoCAD SHX TextTHE GE HPI TECHNICAL SUPPORT GROUP IS AN ADDITIONAL RESOURCE THAT CAN PROVIDE ANSWERS FOR GENERAL GE PRODUCT SITING QUESTIONS AND CAN BE REACHED AT (877)-305-9677 OR MAILTO:[email protected] [email protected]
AutoCAD SHX TextBRIAN FERRELL
AutoCAD SHX Text337-371-7202
-
GE
Hea
lthca
re
AutoCAD SHX TextTRANSIENT LEVEL EXCEEDS THIS VALUE.)
AutoCAD SHX TextOF THE PATIENT AND PERSONNEL.
AutoCAD SHX TextPOWER SUPPLY REGULATION MUST BE 4 PERCENT OR BETTER.
AutoCAD SHX TextTHE MAXIMUM ALLOWABLE TRANSIENT AMPLITUDE IS 2.5 TIMES
AutoCAD SHX TextTHE RMS LINE VOLTAGE. (FILTERS MAY BE REQUIRED IF
AutoCAD SHX TextIT IS RECOMMENDED THAT THE POWER SUPPLY BE MONITORED
AutoCAD SHX TextTO ASCERTAIN THE AVERAGE LINE VOLTAGE, SURGES, SAGS,
AutoCAD SHX TextIMPULSES AND FREQUENCY OF THE SUPPLY VOLTAGE. THE
AutoCAD SHX TextANALYSIS OF A SIMULATED LOAD, USING A POWER SYSTEMS
AutoCAD SHX TextANALYZER CAPABLE OF THE ABOVE SPECIFICATIONS, SHOULD
AutoCAD SHX TextBE CARRIED OUT OVER A CONTINUOUS SEVEN DAY PERIOD PRIOR
AutoCAD SHX TextTO INSTALLATION. THE RESULTS OF THIS ANALYSIS SHOULD
AutoCAD SHX TextBE REVIEWED WITH THE LOCAL SERVICE REPRESENTATIVE TO
AutoCAD SHX TextDETERMINE WHETHER A VOLTAGE/FREQUENCY STABILIZER, POWER
AutoCAD SHX TextLINE PROTECTOR OR FILTERS ARE REQUIRED TO BE INSTALLED
AutoCAD SHX TextBY THE PURCHASER, AS PART OF THE PREINSTALLATION WORK,
AutoCAD SHX TextTO COMPLY WITH THE ABOVE ELECTRICAL REQUIREMENTS.
AutoCAD SHX TextIS THE LIGHTING IN THE ROOM TO ALLOW SAFE EVACUATION
AutoCAD SHX TextTYPE UPS IS RECOMMENDED. EMERGENCY POWER RECOMMENDED
AutoCAD SHX TextPOWER. IF CONTINUOUS OPERATION IS REQUIRED AN ON-LINE
AutoCAD SHX TextFROM POWERLINE DISTURBANCES BY SWITCHING TO EMERGENCY
AutoCAD SHX TextSERIOUS DISRUPTION OF EQUIPMENT OPERATION CAN RESULT
AutoCAD SHX TextEMERGENCY POWER IS NOT RECOMMENDED FOR THE SYSTEM.
AutoCAD SHX TextEMERGENCY
AutoCAD SHX TextPOWER
AutoCAD SHX TextREGULATION
AutoCAD SHX TextPOWER
AutoCAD SHX TextSUPPLY TEST
AutoCAD SHX TextMAXIMUM CURRENT
AutoCAD SHX TextDISCOVERY NM 630/BRIVO NM 615
AutoCAD SHX Text(AMPS)
AutoCAD SHX TextMAXIMUM DAILY VOLTAGE VARIATION MUST FALL WITHIN ONE OF
AutoCAD SHX Text173 to 250 VAC, NOMINAL LINE VOLTAGE OF 208-V 50 Hz or
AutoCAD SHX TextTHAN ONE SECOND TO WITHSTAND SWITCH-ON SURGE.
AutoCAD SHX TextCIRCUIT BREAKERS SHOULD HAVE A TIME DELAY OF GREATER
AutoCAD SHX TextOF 5 CYCLES AND FREQUENCY OF 10 TIMES PER HOUR.
AutoCAD SHX Text5 PERCENT OF RATED LINE VOLTAGE AT A MAXIMUM DURATION
AutoCAD SHX TextMAXIMUM ALLOWABLE TRANSIENT VOLTAGE EXCURSIONS ARE
AutoCAD SHX TextSURGES, LOAD SWITCHING, STATIC ELECTRICITY ETC. CAN CAUSE
AutoCAD SHX TextBE HELD TO A MINIMUM. TRANSIENTS CAUSED BY LIGHTNING,
AutoCAD SHX TextVOLTAGE TRANSIENT OR IMPULSE ON THE INCOMING POWER MUST
AutoCAD SHX TextSCAN ABORTS OR, IN EXTREME INSTANCES, COMPONENT FAILURE
AutoCAD SHX TextVOLTAGE
AutoCAD SHX TextRANGE %%u+%%u 10%
AutoCAD SHX TextINPUT
AutoCAD SHX TextIN THE COMPUTER SUBSYSTEM.
AutoCAD SHX TextTRANSIENT
AutoCAD SHX TextCURRENT
AutoCAD SHX TextDEMAND
AutoCAD SHX TextVOLTAGES/
AutoCAD SHX Text*
AutoCAD SHX Text187-230
AutoCAD SHX Text208
AutoCAD SHX TextTHE RANGES IN TABLE A.
AutoCAD SHX TextALLOWABLE
AutoCAD SHX TextTABLE A
AutoCAD SHX TextVOLTAGE
AutoCAD SHX TextNORMAL
AutoCAD SHX TextNOMINAL
AutoCAD SHX Text25
AutoCAD SHX TextPROTECTION
AutoCAD SHX Text30-A
AutoCAD SHX Text(REV. DATE 28.FEB.14)
AutoCAD SHX Text*
AutoCAD SHX TextOVERCURRENT
AutoCAD SHX TextMINIMUM STANDARD
AutoCAD SHX TextTHE POWER REQUIREMENTS MAY VARY.
AutoCAD SHX TextOPTIONAL FULL SYSTEM UPS IS APPLIED WITH THIS SYSTEM
AutoCAD SHX TextTHESE SPECIFICATIONS APPLY TO THE BASE SYSTEM. IF AN
AutoCAD SHX TextNOTE:
AutoCAD SHX TextREQUIRED POWER SUPPLY: WYE DISTRIBUTION
AutoCAD SHX TextUPS CIRCUIT BREAKER AND CONDUCTORS MUST BE SIZED FOR
AutoCAD SHX TextGEHC
AutoCAD SHX TextUPS
AutoCAD SHX TextREQUIREMENTS
AutoCAD SHX TextRECOMMENDED
AutoCAD SHX TextTHE MAXIMUM INPUT CURRENT OF THE UPS. MAXIMUM INPUT
AutoCAD SHX TextCURRENT OF THE UPS. MAXIMUM INPUT CURRENT IS FULL LOAD
AutoCAD SHX TextCURRENT PLUS THE MAXIMUM BATTERY CHARGING CURRENT.
AutoCAD SHX TextCONSULT UPS NAMEPLATE AND INSTRUCTIONS TO DETERMINE
AutoCAD SHX TextREQUIREMENTS.
AutoCAD SHX Text60Hz, +/- 3Hz, 6 KVA.
AutoCAD SHX TextALL INSTALLATIONS. RANGE OF LINE VOLTAGES:
AutoCAD SHX TextPRIMARY DEDICATED SINGLE PHASE SOURCE IS REQUIRED FOR
AutoCAD SHX TextECG
AutoCAD SHX Text10' [3.05M]
AutoCAD SHX Text46' [14.02M]
AutoCAD SHX TextPATIENT
AutoCAD SHX TextA
AutoCAD SHX Text39.37' [12.00M]
AutoCAD SHX TextAC
AutoCAD SHX Text37.37' [11.39M]
AutoCAD SHX TextNMC
AutoCAD SHX TextNMT
AutoCAD SHX TextEMO
AutoCAD SHX Text61.61' [18.77M]
AutoCAD SHX Text1 PHASE POWER
AutoCAD SHX TextA2
AutoCAD SHX TextUPS
AutoCAD SHX TextESTOP
AutoCAD SHX Text61.61' [18.77M]
AutoCAD SHX TextREV DATE: 06.Jan.16
AutoCAD SHX TextSTANDARD: 29.5' [9.0M]
AutoCAD SHX TextLONG:47.6' [14.5M]
AutoCAD SHX Text(FRU: 5407925)
AutoCAD SHX Text(FRU: 5426855)
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextELECTRICAL SPECIFICATIONS
AutoCAD SHX TextPOWER SPECIFICATIONS
AutoCAD SHX TextINTERCONNECT DIAGRAM
AutoCAD SHX TextELECTRICAL NOTES
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextE2
AutoCAD SHX TextSHEET
AutoCAD SHX TextALL WIRES SPECIFIED SHALL BE COPPER STRANDED, FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
AutoCAD SHX TextLONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR STUBBED CONDUIT ENDS.
AutoCAD SHX TextALL CONDUCTORS, POWER, SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR DUCT SYSTEM. ELECTRICAL
AutoCAD SHX TextCONTRACTOR SHALL RING OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS MUST BE CONTINUOUS COPPER
AutoCAD SHX TextSTRANDED AND FREE FROM SPLICES.
AutoCAD SHX TextWIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER SIZES MAY BE REQUIRED BY LOCAL CODES.
AutoCAD SHX TextIT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED, AS REQUIRED IN ACCORDANCE WITH NATIONAL AND LOCAL
AutoCAD SHX TextELECTRICAL CODES.
AutoCAD SHX TextLOCAL OR NATIONAL CODES.
AutoCAD SHX TextCONVENIENCE OUTLETS ARE NOT ILLUSTRATED. THEIR NUMBER AND LOCATION ARE TO BE SPECIFIED BY OTHERS.
AutoCAD SHX TextLOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO THE SYSTEM CONTROL, THE POWER DISTRITBUTION UNIT AND
AutoCAD SHX TextONE ON EACH WALL OF THE PROCEDURE ROOM. USE HOSPITAL APPROVED OUTLET OR EQUIVALENT.
AutoCAD SHX TextGENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED. CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT FROM
AutoCAD SHX TextFOR GREATER THAN STANDARD CABLE LENGTHS (REFER TO THE INTERCONNECTION DIAGRAM FOR MAXIMUM USABLE
AutoCAD SHX TextLENGTHS POINT TO POINT).
AutoCAD SHX TextCONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND LOCAL
AutoCAD SHX TextA SPECIAL GROUNDING SYSTEM IS REQUIRED IN ALL PROCEDURE ROOMS BY SOME NATIONAL AND LOCAL CODES. IT IS
AutoCAD SHX TextRECOMMENDED IN AREAS WHERE PATIENTS MIGHT BE EXAMINED OR TREATED UNDER PRESENT, FUTURE, OR EMERGENCY
AutoCAD SHX TextPERSONNEL TO DETERMINE THE AREAS REQUIRING THIS TYPE OF GROUNDING SYSTEM.
AutoCAD SHX TextROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE NEED
AutoCAD SHX TextELECTRICAL CODES.
AutoCAD SHX TextCONDITIONS. CONSULT THE GOVERNING ELECTRICAL CODE AND CONFER WITH APPROPRIATE CUSTOMER ADMINISTRATIVE
AutoCAD SHX TextCONDUIT SIZES SHALL BE VERIFIED BY THE ARCHITECT, ELECTRICAL ENGINEER OR CONTRACTOR, IN ACCORDANCE WITH
AutoCAD SHX TextNOTE 9:
AutoCAD SHX TextNOTE 8:
AutoCAD SHX TextNOTE 6:
AutoCAD SHX TextNOTE 7:
AutoCAD SHX TextNOTE 5:
AutoCAD SHX TextNOTE 4:
AutoCAD SHX TextNOTE 2:
AutoCAD SHX TextNOTE 3:
AutoCAD SHX TextNOTE 1:
AutoCAD SHX TextNFSH-1002
AutoCAD SHX TextTHE MAXIMUM POINT TO POINT DISTANCES ILLUSTRATED ON THIS DRAWING MUST NOT BE EXCEEDED.
AutoCAD SHX TextNOTE 10:
AutoCAD SHX TextOVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO CEILING MOUNTING COMPONENTS AND WIRING IF HIGH WATTAGE BULBS
AutoCAD SHX TextARE USED. RECOMMEND LOW WATTAGE BULBS NO HIGHER THAN 75 WATTS AND USE DIMMER CONTROLS (EXCEPT MR).
AutoCAD SHX TextDO NOT MOUNT LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING MOUNTED ACCESSORIES WILL BE PARKED.
AutoCAD SHX TextPHYSICAL CONNECTION OF PRIMARY POWER TO GE EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL CONTRACTOR
AutoCAD SHX TextNOTE 11:
AutoCAD SHX TextWITH THE SUPERVISION OF A GE REPRESENTATIVE. THE GE REPRESENTATIVE WOULD BE REQUIRED TO IDENTIFY THE
AutoCAD SHX TextPHYSICAL CONNECTION LOCATION, AND INSURE PROPER HANDLING OF GE EQUIPMENT.
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX TextALUMINUM OR SOLID WIRES ARE NOT ALLOWED.
AutoCAD SHX TextGEHC CONDUCTS POWER AUDITS TO VERIFY QUALITY OF POWER BEING DELIVERED TO THE SYSTEM. THE CUSTOMER'S
AutoCAD SHX TextNOTE 12:
AutoCAD SHX TextELECTRICAL CONTRACTOR IS REQUIRED TO BE AVAILABLE TO SUPPORT THIS ACTIVITY.
AutoCAD SHX TextDIAGRAM KEY
AutoCAD SHX Text59' [18M] MAXIMUM RUN LENGTH BETWEEN JUNCTION POINTS.
AutoCAD SHX TextGE FURNISHED CABLE RUNS. ROUTE IN EMPTY
AutoCAD SHX TextCUSTOMER/CONTRACTOR SUPPLIED WIRING. ROUTE IN
AutoCAD SHX Text Feet [Meters]
AutoCAD SHX TextCONDUIT OR RACEWAY.
AutoCAD SHX TextADEQUATE CONDUIT OR RACEWAY.
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
-
GE
Hea
lthca
re
AutoCAD SHX TextE3
AutoCAD SHX TextALL ITEMS ILLUSTRATED ARE TO BE FURNISHED AND INSTALLED BY CUSTOMER OR THEIR CONTRACTOR.
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextELECTRICAL DETAIL
AutoCAD SHX TextINSITE CONNECTION (TYPICAL)
AutoCAD SHX TextELEC-1
AutoCAD SHX TextFINISHED FLOOR
AutoCAD SHX TextFINISHED CEILING
AutoCAD SHX TextTO BE DETERMINED
AutoCAD SHX Text1" CONDUIT FROM J.B.
AutoCAD SHX TextTO ABOVE FINSHED CEILING.
AutoCAD SHX TextSINGLE GANG J.B.
AutoCAD SHX TextCOVERPLATE WITH TWO
AutoCAD SHX TextTELEPHONE RECEPTACLES
AutoCAD SHX TextREV. DATE: 04/24/02
AutoCAD SHX TextONE TELEPHONE RECEPTACLE AND
AutoCAD SHX TextOR
AutoCAD SHX TextONE NETWORK RECEPTACLE
AutoCAD SHX TextONE OF THE FOLLOWING TWO SELECTIONS MUST BE INSTALLED AT THE LOCATION SHOWN ON THE
AutoCAD SHX TextELECTRICAL PLAN (SHEET E1) FOR GE INSITE CONNECTION BASED UPON SYSTEM CONFIGURATION.
AutoCAD SHX TextA) ONE INTERNET ACCESSIBLE VIRTUAL PRIVATE NETWORK (VPN) CONNECTION WITH A STATIC IP
AutoCAD SHX Text ADDRESS, AND ONE TELEPHONE LINE - DEDICATED-DIRECT-DIALING, VOICE GRADE.
AutoCAD SHX TextOR
AutoCAD SHX TextB) TWO TELEPHONE LINES - ONE DEDICATED DIRECT-DISTANCE-DIALING, VOICE GRADE AND
AutoCAD SHX Text ONE A DEDICATED DATA LINE.
AutoCAD SHX TextBOX WITH COVERPLATE AND NETWORK JACK
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextELECTRICAL DETAIL
AutoCAD SHX TextELEC-83
AutoCAD SHX TextCOVERPLATE
AutoCAD SHX TextBOX
AutoCAD SHX TextREV. DATE: 10/06/98
AutoCAD SHX TextNETWORK JACK
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextELECTRICAL DETAIL
AutoCAD SHX TextNETWORK CONNECTION (TYPICAL)
AutoCAD SHX TextELEC-84
AutoCAD SHX TextFINISHED FLOOR
AutoCAD SHX TextFINISHED CEILING
AutoCAD SHX TextTO BE DETERMINED
AutoCAD SHX TextLOCAL AREA NETWORK
AutoCAD SHX Text1/2" CONDUIT FROM J.B.
AutoCAD SHX TextTO ABOVE FINSHED CEILING.
AutoCAD SHX TextSINGLE GANG J.B.
AutoCAD SHX TextCOVERPLATE WITH NETWORK
AutoCAD SHX TextRECEPTACLE
AutoCAD SHX TextREV. DATE: 17.Jun.16
AutoCAD SHX TextBOX WITH COVERPLATE (TYPICAL)
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextELECTRICAL DETAIL
AutoCAD SHX TextELEC-8
AutoCAD SHX TextOUTLET BOX
AutoCAD SHX TextHARDWARE
AutoCAD SHX TextREV. DATE: 09/30/94
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextELECTRICAL DETAIL
AutoCAD SHX TextCONDUITS THRU-FLOOR (TYPICAL)
AutoCAD SHX TextFINISHED FLOOR
AutoCAD SHX TextELEC-9
AutoCAD SHX TextHARDWARE
AutoCAD SHX TextREV. DATE: 08/08/94
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextELECTRICAL DETAILS
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextSHEET
AutoCAD SHX TextNFSH-1002
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
-
··
GE
Hea
lthca
re
AutoCAD SHX TextD1
AutoCAD SHX TextM1014AW
AutoCAD SHX TextXELERIS WORKSTATION
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextCOMPUTER
AutoCAD SHX TextKEYBOARD
AutoCAD SHX Text21" COLOR MONITOR
AutoCAD SHX TextREV. DATE: 01/24/08
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX Text18.8"
AutoCAD SHX Text[470mm]
AutoCAD SHX Text6.7"
AutoCAD SHX Text[170mm]
AutoCAD SHX Text1.57"
AutoCAD SHX Text[40mm]
AutoCAD SHX Text19.8"
AutoCAD SHX Text[503mm]
AutoCAD SHX Text18.1"
AutoCAD SHX Text[460mm]
AutoCAD SHX Text20.1"
AutoCAD SHX Text[510mm]
AutoCAD SHX TextTOWER
AutoCAD SHX TextFLAT PANEL MONITOR
AutoCAD SHX Text8.30"
AutoCAD SHX Text[210mm]
AutoCAD SHX Text17.07"
AutoCAD SHX Text[449mm]
AutoCAD SHX Text20.3"
AutoCAD SHX Text[515mm]
AutoCAD SHX TextT.B.D.
AutoCAD SHX TextT.B.D.
AutoCAD SHX TextT.B.D.
AutoCAD SHX Text37.9 LBS/18 KG
AutoCAD SHX Text1465 BTU/500 W
AutoCAD SHX Text3.4 LBS/1.54 KG
AutoCAD SHX Text68.3 LBS/31 KG
AutoCAD SHX Text369 BTU/135 W
AutoCAD SHX TextPLAN VIEW
AutoCAD SHX TextFRONT VIEW
AutoCAD SHX TextSIDE VIEW
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX Text6KVA UPS
AutoCAD SHX TextB640M
AutoCAD SHX TextINDICATES CENTER OF GRAVITY
AutoCAD SHX TextNOTE:
AutoCAD SHX TextINDICATES AIR FLOW
AutoCAD SHX TextREV. DATE: 16.OCT.2012
AutoCAD SHX TextB81054
AutoCAD SHX TextDISCOVERY NM 630/BRIVO NM 615 TABLE
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextREV. DATE: 21.FEB.12
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextOPERATORS CONSOLE
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextREV. DATE: 15.FEB.12
AutoCAD SHX TextB4305G
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextTOP VIEW
AutoCAD SHX TextSIDE VIEWS
AutoCAD SHX TextB81053
AutoCAD SHX TextDISCOVERY NM 630 GANTRY AND TABLE
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextREV. DATE: 22.SEP.11
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextGAMMA CAMERA GANTRY DELIVERY
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextREV. DATE: 22.SEP.11
AutoCAD SHX TextB81055
AutoCAD SHX TextTHE CUSTOMER MUST : - PROVIDE AN AREA, ADJACENT TO THE SUITE, FOR DELIVERY AND UNLOADING OF THE GEMS EQUIPMENT. - ENSURE THAT THE DIMENSIONS OF ALL DOORS, CORRIDORS, CEILING HEIGHTS, ARE SUFFICIENT TO ACCOMMODATE THE MOVEMENT OF GEMS EQUIPMENT FROM THE DELIVERY AREA TO THE SPECIFIC ROOMS OF THE SITE. - ENSURE THAT THE ACCESS ROUTE WILL ACCOMMODATE THE WEIGHTS OF THE EQUIPMENT AND ANY TRANSPORTATION, LIFTING AND RIGGING EQUIPMENT, - IF THE PARKING AND DOCK FACILITIES ARE ON PROPERTY WICH DOES NOT BELONG TO THE CUSTOMER, ENSURE THAT ALL NECESSARY STEPS HAVE BEEN TAKEN TO ENSURE THEIR TEMPORARY USE BY GEMS.
AutoCAD SHX TextTHE GANTRY IS SHIPPED ON A DOLLY EQUIPPED WITH ELEVATING CASTERS. THE ABSOLUTE MINIMUM DIMENSIONS FOR EQUIPMENT DELIVERY ARE : W = 2260MM, D = 1360MM, H = 2000MM.
AutoCAD SHX TextGAMMA CAMERA DELIVERY NEED 1400 MM MINIMUM OPENING
AutoCAD SHX TextCAUTION
AutoCAD SHX TextSYSTEM WORKING AREA TO BE MARKED WITH YELLOW CAUTION TAPE.
AutoCAD SHX TextSYSTEM WORKING AREA TO BE MARKED WITH YELLOW CAUTION TAPE.
AutoCAD SHX Text CAUTION
AutoCAD SHX TextTHE SYSTEM WORKING AREA IS A "CAUTION AREA" INSIDE WHICH ONLY AUTHORIZED PERSONNEL ARE PERMITTED ACCESS. NO UNAUTHORIZED PERSONS ARE ALLOWED INSIDE THIS AREA. THE FLOOR CLEARANCE AREA SHOULD BE CLEARLY MARKED OFF AROUND THE CAMERA TO PREVENT OBSTACLES (FOR EXAMPLE WHEEL CHAIRS) FROM GETTING TOO CLOSE AND COLLIDING WITH THE SYSTEM DURING ITS AUTOMATIC OPERATION. NO ITEMS OF ANY KIND MAY BE PRESENT WITHIN THIS AREA DURING THE AUTOMATIC OPERATION OF THE SYSTEM. A MORE DECORATIVE FLOORING DESIGN MAY BE USED IN LIEU OF YELLOW CAUTION TAPE. HOWEVER, FINAL FLOORING MUST MEET FLOOR SPECIFICATIONS AS DESIGNATED IN THE PREINSTALLATION MANUAL.
AutoCAD SHX TextB0574-
AutoCAD SHX TextDISCOVERY NM630/BRIVO NM615 SYSTEM WORKING AREA
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextREV. DATE: 10.JUN.13
AutoCAD SHX TextNOTE: ALL DIMENSIONS IN INCHES BRACKETED DIMENSIONS [] EQUAL MILLIMETERS
AutoCAD SHX TextB81041
AutoCAD SHX TextDISCOVERY NM 630 AND NM/CT 670 COLLIMATOR CART
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX Text%%USIDE VIEW
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextREV. DATE: 22.SEP.11
AutoCAD SHX Text%%UFRONT VIEW
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextEMO PUSHBUTTON
AutoCAD SHX TextFRONT VIEW
AutoCAD SHX TextTOP VIEW
AutoCAD SHX TextSIDE VIEW
AutoCAD SHX TextREV. DATE: 22.SEP.11
AutoCAD SHX TextB81058
AutoCAD SHX TextDETAIL NOT TO SCALE
AutoCAD SHX TextEQUIPMENT DETAIL
AutoCAD SHX TextE-STOP PUSHBUTTON
AutoCAD SHX TextFRONT VIEW
AutoCAD SHX TextTOP VIEW
AutoCAD SHX TextSIDE VIEW
AutoCAD SHX TextREV. DATE: 18.MAR.13
AutoCAD SHX TextB81061
AutoCAD SHX TextREVISION HISTORY:
AutoCAD SHX TextPROJECT TITLE:
AutoCAD SHX TextSHEET TITLE:
AutoCAD SHX TextTHIS PLAN IS SUBMITTED TO SUGGEST LOCATION OF GE HEALTHCARE EQUIPMENT
AutoCAD SHX TextAND ASSOCIATED APPARATUS, ELECTRICAL WIRING DETAILS AND ROOM ARRANGEMENTS.
AutoCAD SHX TextIN PREPARING THIS PLAN, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS
AutoCAD SHX TextTO ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. IT IS NOT TO BE USED FOR
AutoCAD SHX TextACTUAL CONSTRUCTION PURPOSES, HOWEVER, AND THE COMPANY CANNOT ACCEPT
AutoCAD SHX TextRESPONSIBILITY FOR ANY DAMAGES RESULTING THEREFROM.
AutoCAD SHX TextMODALITY TYPE:
AutoCAD SHX TextEQUIPMENT DETAILS
AutoCAD SHX TextPROJECT REVISION
AutoCAD SHX TextDATE:
AutoCAD SHX TextCHECKED BY:
AutoCAD SHX TextDRAWN BY:
AutoCAD SHX TextSHEET
AutoCAD SHX TextNFSH-1002
AutoCAD SHX TextTHIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX TextHealthcare Project Implementation - Design Center
AutoCAD SHX TextMilwaukee,
AutoCAD SHX TextWisconsin
AutoCAD SHX TextDISCOVERY NM630
AutoCAD SHX TextCHILDRENS HOSPITAL
AutoCAD SHX TextOF NEW ORLEANS
AutoCAD SHX TextNEW ORLEANS, LOUISIANA
AutoCAD SHX TextM076460
AutoCAD SHX Text00A
AutoCAD SHX Text30.Oct.17
AutoCAD SHX TextKMS
AutoCAD SHX TextRET
AutoCAD SHX TextThis drawing is based on Sketch No.: CHNO-NM-D630-1st Floor.dxf
AutoCAD SHX TextRQ - DC-33356
AutoCAD SHX TextPIM R3
AutoCAD SHX TextGON NO:
AutoCAD SHX TextGON DT:
AutoCAD SHX Text4566821
AutoCAD SHX Text13.Oct.17
Sheets and ViewsNUC-M076460-00A-Cover_SheetNUC-M076460-00A-Equip_LayoutNUC-M076460-00A-Struct_LayoutNUC-M076460-00A-Struct_DetailsNUC-M076460-00A-Elec_LayoutNUC-M076460-00A-Elec_SpecsNUC-M076460-00A-Elec_DetailsNUC-M076460-00A-Equip_Details