FUJIFILM MEDICAL SYSTEMS USA, INC. VISIONARY SUITE FINAL
12
FUJIFILM MEDICAL SYSTEMS USA, INC. TYPICALS VISIONARY EQUIPMENT PLANS STAMFORD, CT 06801 VISIONARY SUITE FINAL SITE PLAN MAY 28, 2017 2017-219-TYP 1 of 12 HQ FINAL DRAWING PLANS - APPROVAL CUSTOMER DATE CUSTOMER SIGNATURE YOUR SIGNATURE APPROVES THIS FINAL DRAWING FOR ACCEPTANCE TO FORWARD TO YOUR ARCHITECT - CONTRACTOR TO IMPLEMENT INTO THEIR FINAL DRAWING SET
YOUR SIGNATURE APPROVES THIS FINAL DRAWING FOR ACCEPTANCE TO FORWARD TO
YOUR ARCHITECT - CONTRACTOR TO IMPLEMENT INTO THEIR FINAL DRAWING SET
AutoCAD SHX Text
PROFESSIONAL DESIGN SERVICES
AutoCAD SHX Text
STAMFORD, CT 06902
AutoCAD SHX Text
419 WEST AVENUE
AutoCAD SHX Text
CUSTOMER
AutoCAD SHX Text
PRODUCT NAME
AutoCAD SHX Text
DRAWING SET
AutoCAD SHX Text
DATE
AutoCAD SHX Text
REVISION/DATE
AutoCAD SHX Text
PROJECT NUMBER
AutoCAD SHX Text
SHEET
AutoCAD SHX Text
INSTALLATION COORD:
AutoCAD SHX Text
PHONE CONTACT:
AutoCAD SHX Text
REGION:
AutoCAD SHX Text
1
AutoCAD SHX Text
COVER SHEET
AutoCAD SHX Text
2
AutoCAD SHX Text
GENERAL NOTES
AutoCAD SHX Text
3
AutoCAD SHX Text
FINAL EQUIPMENT PLAN
AutoCAD SHX Text
4
AutoCAD SHX Text
DETAIL DRAWING
AutoCAD SHX Text
5
AutoCAD SHX Text
DETAIL DRAWING
AutoCAD SHX Text
6
AutoCAD SHX Text
ELECTRICAL DIAGRAM, NOTES
AutoCAD SHX Text
7
AutoCAD SHX Text
FINAL ELECTRICAL PLAN
AutoCAD SHX Text
8
AutoCAD SHX Text
FINAL ELECTRICAL ISOMETRIC
AutoCAD SHX Text
9
AutoCAD SHX Text
WALLDUCT DETAILS
AutoCAD SHX Text
10
AutoCAD SHX Text
GENERAL STRUCTURAL PLANS
AutoCAD SHX Text
11 - 12
AutoCAD SHX Text
GENERAL STRUCTURAL NOTES
Customer to verify freight size and integrity of any and all elevator(s) used for routing Fujifilm Medical Systems USA equipment.
Fujifilm Medical Systems USA representative.Customer to coordinate routing of Fujifilm Medical Systems USA equipment prior to equipment arrival with local
plans without prior notification.
to design and incorporate our final site plan drawings to your architectural construction documents. Fujifilm Medical Systems USA
It is the customer's responsibility to secure a storage area for equipment on site prior and during installation.
Fujifilm Medical Systems USA reserves the right to make changes to equipment and detailed specification regarding thesepreliminary plans shall not be used for construction documents, please refer to the latter final site plan drawings.
This drawing set is the sole property of Fujifilm Medical Systems USA. It's use is authorized for the customer architect of record
These drawings are provided with the understanding by providing such documents, does not alter any of the terms and conditions
damages resulting (from any matter) from the use of the attached drawings.Also Fuji Medical Systems USA or their agents will not be held liable for any direct, indirect, special, incidental or consequentialFujifilm Medical Systems USA makes no warranty or representation, either express or implied with respect to the attached drawings.
architects, engineers, or other design construction professional in the preparation of official construction documents.requirements. The information contained in this drawing set is provided for the use of customers and their designatedThese drawings are reference drawings (only), illustration of general equipment layout and facility-related equipment
Minimum equipment entry door height 80 INCHES
Minimum equipment entry door width 42 INCHES (recommended)
Passageways and Corridors 60 INCHES
The customer shall supply network drops for IP application.All network outlets will installed and set to 100 base T full duplex and checked prior to equipment install.
Customer to provide an IT representative readily available during the equipment installation period.
All power connections shall be dedicated independent connections to the power distribution panel.
All power shall be guaranteed by the owner to meet the power specificationsnormally associated powering medical electronic equipment.
Please refer to your final site plan drawing for site specific power requirements.
Equipment ambient temperature range 59 - 93 Degrees Fahrenheit.
Relative humidity should be between 45 - 95 percent non-condensing.
The room the equipment is installed within should be kept dust free.
of said sale between Fujifilm Medical Systems USA and their customer.
The customer is responsible for the complete an accurate incorporation of all specifications and requirements from these
siting and placement of purchased Fujifilm Medical Systems USA equipment.These drawing are not to be used as construction documents. It is the sole responsibility of the customer to ensure the correct
reference drawings to your architect of record for use of site preparation. Sufficiency of final site plans and specifications, specifically including
Worldwide Healthcare Designs, Inc. (WHD) is Fujifilm Medical Systems USA sole designer for said project.
All network drops will be installed and marked on the connection prior to installation of the equipment.
It is the customer's responsibility to be in compliance with all local, state and federal safety, electrical, and building
This shall include any and all expenses for room and structural alterations. Site preparation shall be in accordanceIt is the customer's sole responsibility and their expenses for room preparation of said room for this equipment sale.
Any time delay from the removal of the latter may result in an extended installation period.
The customer shall obtain all permits and required licenses needed for Federal, State and Local regulations,
The customer should provide Fujifilm Medical Systems USA a construction time-line chart and also
Power - 3 phase, 480 VACBreaker - 80 amp for 80kW generator
All doors, windows and radiation protection installed and finished.
All electrical conduits, raceway, junction boxes and outlets installed and operational.
flush with ceiling grid and operational.Please have walls painted, baseboards installed, all floors tiled, ceiling tiles installed and light fixtures installed
All plumbing must be installed, finished and operational. Any plumbing connections in the immediate area of
All contractors supplied cables must be pulled and terminated at designated locations.
X-ray Procedure and Control Room - a dust free environment.
and free of lateral or longitudinal interference.All support structures used for install and equipment support must be installed, level (within one-eight inch end to end),
Incoming power to X-Ray room breaker panel connected and operational.
the new installation that is not being used must be capped and covered.
with this set of final site plans and specifications provided by Fujifilm Medical Systems USA.
codes relevant to Fujifilm Medical Systems USA final site plans, equipment and installation.
Radiation protection is not shown on these set of plans, However the customer at their own expense shall obtain a
Also the customer shall bear any expense in complying with any related rules, regulations and ordinances.in connection with the construction. installation and operation of their purchased equipment.
The customer shall advise a Fujifilm Medical Systems USA of any adverse conditions at or near the sitethat could affect the installation scheduling and service support team travel.
but not limited to the accuracy of dimensions (based on information provided), shall be the sole responsibility of the customer.
It is the customer's sole responsibility to properly remove and dispose of any hazardous materials - at their expense.
It is the customer's responsibility to properly remove and make the site conditions safe for installation and operation.
follow-up with scheduled construction meeting notice for project area(room).This is to insure coordination between contractor(s) with delivery and installation of purchased equipment.
All dimensions are measured from finished surfaces, unless otherwise specified on drawing.
We recommend that during equipment delivery and installation, for the customer to protect finished floors, walls andany areas, including other equipment within area of install with protective covering.
Installation will not commence until room is clean and dust free. All contractors equipment must be removed prior to install.Fujifilm Medical Systems USA will not be responsible for any lost, damaged tools and materials from customer's contractors.
licensed radiation physicist to specify need of radiation protection.
Installation will not commence until room is clean and dust free. All contractors equipment must be removed prior to install.Fujifilm Medical Systems USA reserves the right to refuse delivery and installation due to the room is not properly preparedper accordance with Fujifilm Medical Systems USA final site plans.
The customer or contractor shall supply and install all materials and other features called-out from Fujifilm Medical Systems USA final site plans.
FUJIFILM MEDICAL SYSTEMS USA. Inc. TYPICALSVISIONARY EQUIPMENT PLANS
STAMFORD, CT 06801
VISIONARYSUITE
WHD-A.A.
MAY 28, 2017
NOT TO SCALE
2017-219-TYP
2 of 12
AutoCAD SHX Text
%%UDRAWING SUBJECT TO CHANGE
AutoCAD SHX Text
%%uEQUIPMENT ROUTING
AutoCAD SHX Text
%%UDISCLAIMER
AutoCAD SHX Text
Network Notes
AutoCAD SHX Text
%%UEnvironmental Notes
AutoCAD SHX Text
%%UElectrical Requirements
AutoCAD SHX Text
%%UGENERAL NOTES - site prep requirements
AutoCAD SHX Text
EQUIPMENT
AutoCAD SHX Text
PROFESSIONAL DESIGN SERVICES
AutoCAD SHX Text
STAMFORD, CT 06902
AutoCAD SHX Text
REFERENCE
AutoCAD SHX Text
NOT FOR
AutoCAD SHX Text
CONSTRUCTION
AutoCAD SHX Text
DRAWINGS ONLY
AutoCAD SHX Text
419 WEST AVENUE
AutoCAD SHX Text
PROJECT
AutoCAD SHX Text
REVISION
AutoCAD SHX Text
SCALE
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DRAWN BY
AutoCAD SHX Text
PROJECT NUMBER
AutoCAD SHX Text
SHEET
AutoCAD SHX Text
GENERAL NOTES
CUSTOMER-CONTRACTOR TO PROVIDE VACUUM AND OXYGEN OUTLETS PER THEIR
CUSTOMER-CONTRACTOR TO PROVIDE DESK, CHAIR, AND CABINETRY PER THEIR NEEDS.
CUSTOMER 'S PHYSICIST TO DETERMINE AREA SHIELDING REQUIREMENTS.
BY THE CUSTOMER. CONTROL ROOM MAXIMUM WALL HEIGHT NEEDS TO BE 7 '-6 ' '.LEAD LINED WINDOWS-WALLS, CABINETRY, COUNTER AND SINKS ARE TO BE PROVIDED
PLEASE VERIFY ROOM DIMENSIONS TO THE ACTUAL SITE CONDITIONS.
2.
6.
5.
4.
3.
1.
( IF APPLICABLE )
REQUIREMENTS.
NEW WALLS
EXISTING WALLS
EXISTING WALLS TO BE REMOVED
CUSTOMER TO SUPPLY ALL CONSTRUCTION
13 '-0 ' ' STANDARD RAILS SHOWN
PLAN NORTH
N
EQUIPMENT PLAN
*
SUPPLIED-INSTALLED BY MEDICAL MANUFACTURERSUPPLIED-INSTALLED BY CUSTOMER-CONTRACTOR
INCLUDE (1) 30 AMP BREAKER (208 VAC) FOR "DS4"- HUBBELL CONNECTOR (L630R)
AutoCAD SHX Text
DS2
AutoCAD SHX Text
DS1
AutoCAD SHX Text
SUPPLY HUBBELL CONNECTOR (L620R), LOCATE AT 16" A.F.F.
AutoCAD SHX Text
LOCATE PER LOCAL CODES/OTHERS.
AutoCAD SHX Text
JB1
AutoCAD SHX Text
FD1
AutoCAD SHX Text
G-4000 WIREMOLD DUCT, 2" H. x 4" W., SURFACE MOUNT FLUSHED WITH FINISHED FLOOR.
AutoCAD SHX Text
DUCT MUST HAVE REMOVABLE COVERS AND MIN. ONE (1) DIVIDER THROUGH-OUT.
AutoCAD SHX Text
- LOCATE PER LOCAL CODE/OTHERS. (INCLUDE NIPPLE, FLEX-HOSE, ETC..)
AutoCAD SHX Text
MOUNT BOTTOM OF WIREMOLD FLUSH WITH FINISHED FLOOR. MUST HAVE CABLE ACCESS TO WALLDUCT "WD1".
AutoCAD SHX Text
RISER HAVE REMOVABLE COVERS AND MIN. (1) DIVIDER THROUGH-OUT.
AutoCAD SHX Text
RISER DUCT, 10" x 3 1/2", FROM FINISHED FLOOR TO ABOVE FINISHED CEILING. SURFACE MOUNT
AutoCAD SHX Text
WITH FINISHED WALL, RISER MUST HAVE ACCESS OPENING TO WALLDUCT "WD1".
AutoCAD SHX Text
CONTRACTOR MUST SUPPLY CONDUIT FITTINGS FOR CONDUIT CONNECTIONS. (IF APPLICABLE)
AutoCAD SHX Text
LOCATE PER LOCAL CODE/OTHERS.
AutoCAD SHX Text
J-BOX, 10" x 10" x 4", BOTTOM OF BOX FLUSH MOUNTED 3" A.F.F. AND FLUSH WITH FINISHED WALL.
AutoCAD SHX Text
J-BOX TO HAVE CUT-OUT AND ACCESS ON FACE OF "JB1". CABLE ACCESS THROUGH WALLDUCT "WD1".
AutoCAD SHX Text
LOCATE PER LOCAL CODE/OTHERS.
AutoCAD SHX Text
CONTRACTOR MUST SUPPLY CONDUIT FITTINGS FOR CONDUIT CONNECTIONS. (IF APPLICABLE)
AutoCAD SHX Text
JB2
AutoCAD SHX Text
JB3
AutoCAD SHX Text
(2) J-BOX, 6" x 6" x 6", MOUNT BOX FLUSH-IN FLOOR. TOP OF BOX FLUSH MOUNT WITH FINISHED FLOOR.
AutoCAD SHX Text
J-BOX TO HAVE CUT-OUT AND ACCESS ON FACE OF "JB3". MUST HAVE CABLE ACCESS TO FLOOR DUCT
AutoCAD SHX Text
LOCATE PER LOCAL CODE/OTHERS.
AutoCAD SHX Text
"FD1". CONTRACTOR MUST SUPPLY CONDUIT FITTINGS FOR CONDUIT CONNECTIONS. (IF APPLICABLE)
AutoCAD SHX Text
J-BOX, 8" x 8" x 4", BOTTOM OF J-BOX FLUSH MTD. 24" A.F.F. AND FLUSH WITH FINISHED WALL.
AutoCAD SHX Text
J-BOX TO HAVE CUT-OUT AND ACCESS ON FACE OF "JB2".
AutoCAD SHX Text
LOCATE PER LOCAL CODE/OTHERS.
AutoCAD SHX Text
CONTRACTOR MUST SUPPLY CONDUIT FITTINGS FOR CONDUIT CONNECTIONS. (IF APPLICABLE)
AutoCAD SHX Text
WD1
AutoCAD SHX Text
WALLDUCT, 10" H. x 3 1/2" W., SURFACE MOUNT WITH FINISHED FLOOR, DUCT MUST HAVE REMOVABLE
AutoCAD SHX Text
COVERS AND MIIN. (1) DIVIDER THROUGH-OUT. DUCT MUST HAVE ACCESS OPENING TO RISERS "R1" AND
AutoCAD SHX Text
CONTRACTOR MUST SUPPLY CONDUIT FITTINGS FOR CONDUIT CONNECTIONS. (IF APPLICABLE)
AutoCAD SHX Text
WALLDUCT MUST HAVE ACCESS THORUGH CONTROL ROOM WALL (IF APPLICABLE).
AutoCAD SHX Text
LOCATE PER LOCAL CODE/OTHERS.
AutoCAD SHX Text
EXISTING
AutoCAD SHX Text
EXISTING
AutoCAD SHX Text
J-BOX "JB3".
AutoCAD SHX Text
JB3
AutoCAD SHX Text
FROM "JB1" TO "JB2"
AutoCAD SHX Text
(1) 2 1/2" CONDUIT
AutoCAD SHX Text
EQUIPMENT
AutoCAD SHX Text
PROFESSIONAL DESIGN SERVICES
AutoCAD SHX Text
STAMFORD, CT 06902
AutoCAD SHX Text
REFERENCE
AutoCAD SHX Text
NOT FOR
AutoCAD SHX Text
CONSTRUCTION
AutoCAD SHX Text
DRAWINGS ONLY
AutoCAD SHX Text
419 WEST AVENUE
AutoCAD SHX Text
PROJECT
AutoCAD SHX Text
REVISION
AutoCAD SHX Text
SCALE
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DRAWN BY
AutoCAD SHX Text
PROJECT NUMBER
AutoCAD SHX Text
SHEET
AutoCAD SHX Text
ELECTRICAL ISOMETRIC PLAN
POWER WIRES FROM FUJI MEDICALSIGNAL CABLES. (SEE RACEWAY LAYOUT, DIAGRAM).
ALL TROUGH SHALL BE SQUARE-D, WALKER DUCT (OR EQUAL) FLOOR AND WALL TRENCH DUCT.SIZES SHALL BE AS SPECIFIED ON DRAWINGS, OR PROVIDE EQUIVALENT NOMINAL SIZES AVAILABLEFROM MANUFACTURER (SEE TROUGH LAYOUT, DIAGRAM). FLOOR TROUGH TO BE GASKETED TO
ALL TROUGH SHALL BE FLUSH WITH FINISHED SURFACES UNLESS OTHERWISE SPECIFIED AND PROVIDED WITH REMOVABLE COVERS THE LENGTH. COVERS SHALL BE CLEAR OF ANYOBSTUCTION FOR INSERTION OF PRE-PLUGGED CABLES. IF IT IS NECESSARY THAT THE TROUGHBE SURFACE MOUNTED, WHEN NOT INDICATED ON THESE FUJIFILM MEDICAL DRAWINGS, ALL EQUIPMENT
PLACEMENT AND TRAVEL CLEARANCE MUST BE VERIFIED TO PREVENT EQUIPMENT AND TROUGHINTERFERENCE AND POSSIBLE EQUIPMENT DAMAGE. CONTACT FUJIFILM MEDICAL PROFESSIONAL DESIGNSERVICES FOR PROPER VERIFICATION OF EQUIPMENT PLACEMENT AND TROUGH CLEARANCES.
REMOVABLE TYPE BARRIER STRIP AND CROSSOVERS MAY BE REQUIRED FOR SEPARATION OF A.C.POWER WIRES OR VIDEO COAX CABLES FROM FUJIFILM MEDICAL SIGNAL CABLES. (SEE TROUGH LAYOUT,
PAINT ALL EXPOSED AREAS OF TROUGH. CONSULT WITH CUSTOMER ON MATCHING OR ACCENT COLORS.
WHEN ORDERING TROUGH DUCT AND REQUIRED FITTINGS FROM SQUARE-D OR WALKER PAKERSBURG,USE THE APPROPRIATE CATALOG NUMBERS FOR EACH ITEM TO EXPEDITE DELIVERY.
"SWEEP ELBOWS" FOR CORNER BENDS SHALL BE PROVIDED. THE FOLLOWING SQUARE-D CATALOGNUMBERS (OR EQUAL): "RWT-SFEE" FOR FLUSH MOUNTED TROUGH. "RWT-SSEES" FOR SURFACEMOUNTED TROUGH. NOTE: WHEN ORDERING, SPECIFY 10" OR 18" TROUGH USAGE.
ALL RACEWAY SHALL BE SURFACE MOUNTED ON FINISHED SURFACES UNLESS OTHERWISE SPECIFIEDAND SHALL BE PROVIDED WITH REMOVABLE COVERS, THE FULL LENGTH. COVERS SHALL BE CLEAROF ANY OBSTRUCTION FOR INSERTION OF PRE-PLUGGED CABLES.
PAINT ALL EXPOSED AREAS OF RACEWAY. CONSULT WITH CUSTOMER ON MATCHING OR ACCENT
ALL RACEWAY SHALL BE "WIREMOLD" BRAND (OR EQUAL) FOR FLOOR, WALLS, OR CEILING, UNLESSOTHERWISE SPECIFIED. (SEE RACEWAY LAYOUT, DIAGRAM FOR RECOMMENDED SIZES).
REMOVABLE TYPE BARRIER STRIP AND CROSSOVERS MAY BE REQUIRED FOR SEPARATION OF A.C.
DIAGRAM)
COLORS.
PREVENT ENTRY OF LIQUIDS.
NOT TO SCALE
THIS DETAIL IS A RECOMMENDATION ONLY. CONTRACTOR TO DETERMINE
THIS DETAIL APPLIES TO JUNCTION BOXES INDICATED ON FUJIFILM MEDICAL'S FINAL
OPEINGS OF THESE BOXES SHALL FEATURE EXPOSED REMOVEABLE COVER
ACTUAL METHODS BASED ON SPECIFIC SITE CONDITIONS.
DRAWINGS WITH LETTER DESIGNATIONS JB1, JB2, ETC.
PLATES AS DESIGNATED.
WITH 2 DIVIDERS FOR
GROMMETTED OPENING FOR
VERTICAL RACEWAY RISER FORTUBERCRANE CABLES.
JUNCTION BOX AND CONDUITWITHIN WALL
TROUGH TUB TO MATCH JUNTION BOXOPENING THROUGH REAR OF RACEWAY/
SURFACE RACEWAY / TROUGH TUB
OPENING
GROMMET SHARP EDGES
COVER PLATE CUTAWAY VIEW
COVER PLATE
WALL LINE
CABLE SEPARATIONS.
CABLE ACCESS.
FUJIFILM MEDICAL SYSTEMS USA. Inc. TYPICALSVISIONARY EQUIPMENT PLANS
STAMFORD, CT 06801
VISIONARYSUITE
WHD-A.A.
MAY 28, 2017
1/4"=1'-0
2017-219-TYP
9 of 12
AutoCAD SHX Text
ITEMS LISTED BELOW SHALL BE SUPPLIED AND INSTALLED BY CONTRACTOR.
AutoCAD SHX Text
ITEMS LISTED BELOW SHALL BE SUPPLIED AND INSTALLED BY CONTRACTOR
AutoCAD SHX Text
%%URACEWAY NOTES
AutoCAD SHX Text
%%UTROUGH NOTES
AutoCAD SHX Text
%%UFLUSH J.BOX & CONDUIT ADAPTION TO SURFACE MOUNTED RACEWAY
AutoCAD SHX Text
%%UGENERAL NOTES
AutoCAD SHX Text
EQUIPMENT
AutoCAD SHX Text
PROFESSIONAL DESIGN SERVICES
AutoCAD SHX Text
STAMFORD, CT 06902
AutoCAD SHX Text
REFERENCE
AutoCAD SHX Text
NOT FOR
AutoCAD SHX Text
CONSTRUCTION
AutoCAD SHX Text
DRAWINGS ONLY
AutoCAD SHX Text
419 WEST AVENUE
AutoCAD SHX Text
PROJECT
AutoCAD SHX Text
REVISION
AutoCAD SHX Text
SCALE
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DRAWN BY
AutoCAD SHX Text
PROJECT NUMBER
AutoCAD SHX Text
SHEET
AutoCAD SHX Text
WALLDUCT DETAILS
STRUCTURAL PLAN
OR SUPPLY AN ACCESS THROUGH CEILING TO SERVICE EQUIPMENT.PLEASE SUPPLY A SUSPENDED FINISHED CEILING. PREFERED FOR SERVICING TUBE SUSPENSION.
BOLTS, ANCHORS, PULL-TESTING, WATER PROOFING MATERIAL, ETC.
CHEST STAND SYSTEM BASE, SURFACE MOUNTED WITH THE FINISHED FLOOR. INSTALLED BY FUJIFILM 'SINSTALLATION TEAM. STRUCTURAL CONTRACTOR TO PROVIDE ANY ADDITIONAL MATERIALS,
STRUCTURAL CONTRACTOR MUST COORDINATE WITH THE CUSTOMER'S ARCH. OF RECORD
STRUCTURAL LEGEND
TO ENSURE NO PROTRUSIONS AT ALL EXIST FROM BELOW THE FINISHED CLG. THE ENTIRELENGTH OF TRAVEL FOR THE SIEMEN 'S BRIDGE. ANY AND ALL EQUIPMENT MUST BE OF THERECESSED TYPE. ANY DEVIATION MUST BE BROUGHT TO THE ATTENTION OF THE LOCALFUJIFILM SERVICE REP.
FUJIFILM 'S CEILING SYSTEMS RAILS (SEE DETAIL SHEET).
CODE DESCRIPTION
BOLTS, ANCHORS, PULL-TESTING, WATER PROOFING MATERIAL, ETC.INSTALLATION TEAM. STRUCTURAL CONTRACTOR TO PROVIDE ANY ADDITIONAL MATERIALS, TABLE BASE, SURFACE MOUNTED WITH THE FINISHED FLOOR. INSTALLED BY FUJIFILM 'S
TYPICAL UNISTRUT P-1001K OR P-5501K, OR EQUAL, FLUSH MOUNTED WITH FINISHEDCEILING. PAINT TO MATCH FINISHED CEILING. BOLT SPACING OF EQUIPMENT SHALL DETERMINETYPE OF UNISTRUT TO BE USED. VERIFY WITH LOCAL FUJIFILM MEDICAL REPRESENTATIVE.
(CUSTOMER TO DETERMINE IF EXISTING STRUCTURE CAN SUPPORT NEW SUPPORT RAILS AND(SUPPLIED AND INSTALLED BY THE CUSTOMER 'S CONTRACTOR).
FUJIFILM MEDICAL SYSTEMS USA. Inc. TYPICALSVISIONARY EQUIPMENT PLANS
STAMFORD, CT 06801
VISIONARYSUITE
WHD-A.A.
MAY 28, 2017
1/4"=1'-0
2017-219-TYP
10 of 12
AutoCAD SHX Text
004
AutoCAD SHX Text
002
AutoCAD SHX Text
005
AutoCAD SHX Text
005
AutoCAD SHX Text
006
AutoCAD SHX Text
004
AutoCAD SHX Text
002
AutoCAD SHX Text
003
AutoCAD SHX Text
001
AutoCAD SHX Text
004
AutoCAD SHX Text
L
AutoCAD SHX Text
C
AutoCAD SHX Text
BACK
AutoCAD SHX Text
005
AutoCAD SHX Text
BACK
AutoCAD SHX Text
L
AutoCAD SHX Text
C
AutoCAD SHX Text
L
AutoCAD SHX Text
C
AutoCAD SHX Text
006
AutoCAD SHX Text
%%U9'-4" A.F.F.
AutoCAD SHX Text
%%URECOMMENDED
AutoCAD SHX Text
%%UFIN. CLG. HT.
AutoCAD SHX Text
%%U9'-6" A.F.F.
AutoCAD SHX Text
%%UEXISTING
AutoCAD SHX Text
%%UFIN. CLG. HT.
AutoCAD SHX Text
UNDER CEILING TILE
AutoCAD SHX Text
PLEASE DO NOT INSTALL
AutoCAD SHX Text
SPRINKLER HEADS,
AutoCAD SHX Text
ALONG BRIDGE TRAVEL.
AutoCAD SHX Text
003
AutoCAD SHX Text
001
AutoCAD SHX Text
001
AutoCAD SHX Text
001
AutoCAD SHX Text
001
AutoCAD SHX Text
001
AutoCAD SHX Text
001
AutoCAD SHX Text
003
AutoCAD SHX Text
003
AutoCAD SHX Text
003
AutoCAD SHX Text
EQUIPMENT
AutoCAD SHX Text
PROFESSIONAL DESIGN SERVICES
AutoCAD SHX Text
STAMFORD, CT 06902
AutoCAD SHX Text
REFERENCE
AutoCAD SHX Text
NOT FOR
AutoCAD SHX Text
CONSTRUCTION
AutoCAD SHX Text
DRAWINGS ONLY
AutoCAD SHX Text
419 WEST AVENUE
AutoCAD SHX Text
PROJECT
AutoCAD SHX Text
REVISION
AutoCAD SHX Text
SCALE
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DRAWN BY
AutoCAD SHX Text
PROJECT NUMBER
AutoCAD SHX Text
SHEET
AutoCAD SHX Text
GENERAL STRUCTURAL PLAN
CUSTOMERS RESPONSIBILITY TO HAVE A RADIATION SURVEY PERFORMED IF NEEDED.RADIATION SHIELDING IS NOT SHOWN ON THE SET OF PLANS, HOWEVER IT IS THE
THE CUSTOMER OR HIS CONTRACTORS SHALL SUPPLY AND INSTALL MATERIALS AND OTHER THE CUSTOMER IS RESPONSIBLE FOR ALL UTILITY TIES CONNECTIONS AND SERVICES.
THE CUSTOMER IS RESPONSIBLE FOR ALL ROOM PREPARATION COSTS, FEES, PERMITS AND
BEAR SOLE RESPONSIBILITY FOR COMPLIANCE WITH APPLICABLE CODES ( NEC, OSHA, LOCAL, THE OPERATION OF THE EQUIPMENT. THE CUSTOMERS ARCHITECT OR ITS AGENT SHALL ATTACHED RECOMMENDED LAYOUT, CONFIGURES THE LAYOUT PLAN TO ALLOW OPTIMUM OF EVERY EFFORT HAS BEEN MADE TO ASSURE THAT THE EQUIPMENT DEPICTED ON THE
SYSTEMS U.S.A., INC WILL NOT BE RESPONSIBLE FOR DESIGNS AND CONSTRUCTION THAT DO THE ELECTRICAL AND STRUCTURAL REQUIREMENTS FOR THAT EQUIPMENT. FUJIFILM MEDICAL THESE DRAWING SUGGEST PLACEMENT OF THE PURCHASED EQUIPMENT, AS WELL DESCRIBE
ARCHITECT OF RECORD OR AGENT SHALL PREPARE CONSTRUCTION DOCUMENTS.THE ATTACHED DRAWINGS ARE NOT FOR CONSTRUCTION PURPOSES. THE CUSTOMERS
CONSTRUTION.
BRACE UNISTRUT OR POWER STRUT ATTACHED TO VERTICAL COLUMNS AS SHOWN ON DETAIL.
CORPORATION. ALL STRUT MEMBERS SHALL BE SUPPLIED AND INSTALLED BY
LOCATION OF VERTICAL COLUMNS SHALL VARY DEPENDING ON BUILDING
STRUT CHANNEL SUPPORT SHOULD BE A PART OF SPECIALTIES SPECIFICATIONS,
SPANNING MEMBER 'S UNISTRUT OR POWER STRUT FLUSH WITH FINISHED
VERTICAL COLUMNS: UNISTRUT OR POWER STRUT ATTACHED TO SPANNINGMEMBERS AS SHOWN ON GENERAL CEILING DETAILS SHEET. NUMBER AND
RECOMMENDED UNIVERSAL STRUT CHANNEL SUPPORT FRAMEALLOWS FOR TRANSVERSE AND LONGITUDINAL ADJUSTMENT AT TIME OF EQUIPMENTINSTALLATION. DETERMINATION OF THE ACTUAL DESIGN SHALL BE COORDINATEDAND APPROVED BETWEEN ENGINEER OF RECORD AND UNISTRUT OR POWER STRUT
CEILING. SPACED 2 '-2 ' ' O.C.
OMITTED
SEC. 10. NOT IRON SPECIFICATIONS.
CONTRACTOR.
O.C. TYPICAL
3
78 ' '
FEATURES CALLED OUT IN THE PLANS.
INSPECTIONS.
NATIONAL, RADIATION SAFETY ).
NOT FOLLOW THOSE GUIDELINES.
1
4
LARGE A STANDARD 30-INCH COUNTER TOP MAY NOT BE SUFFICIENT.COMPUTERS INCLUDING KEYBOARD DRAWS. SINCE SOME OF THE CRT MONITORS ARE VERY THE RESPONSIBILITY OF THE CUSTOMER TO PROVIDE ERGONOMIC DEVICES TO HOLD THESE IN AREAS THAT HAVE IIP'S PRINT SERVERS, QA 771WS OR SIMILAR DEVICES LOCATED, IT IS
MUST BE CAPPED OR COVERED PRIOR TO INSTALLATION OF THE FUJIFILM MEDICAL SYSTEMS EXISTING SERVICES THAT WILL NOT BE USED ( I.E. PULMBING, J BOXES, FLOOR DRAINS )
CABLES AND TERMINATIONS. THIS WORK SHALL BE PRIOR TO INSTALLATION START.THE CUSTOMER IS RESPONSIBLE FOR THE SUPPLY AND INSTALLATION OF ALL INTERCONNECT
WORK WHICH SHALL BE COMPLETED AFTER THE INSTALLATION OF THE PURCHASED SYSTEM.THE CUSTOMER OR HIS CONTRACTOR IS RESPONSIBLE FOR ANY FINAL PAINT OR TOUCH-UP
EQUIPMENT.
BRACE
REQUIRED FITTING
REQUIRED FITTING
VERTICAL COLUMN
FUJIFILM MEDICAL LONGITUDINALCEILING TRACK
SPANNING MEMBER
REQUIRED FITTING
FINISHED CEILING LINE
REFERENCE TO 2 '-2 ' ' IS SHOWNFOR 2 ' x 4 ' CEILING TILES ONLY.SPACE BETWEEN SUPPORT RAILSCAN VARY BETWEEN 34.5 ' ' TO 53.125 ' '(BASED ON INSTALL MANUAL)
DRAWINGS TO MATCH FUJIFILM MEDICALLONGITUDINAL CEILING TRACK SPACINGS.RAILS SUPPLIED AND INSTALLED
SPECIFIED ON SITE PLANNINGO.C. OR AS OTHERWISE
BY FUJIFILM MEDICAL INSTALLATIONS.
NOT TO SCALE
DETAILSSEE GENERAL
FUJIFILM MEDICAL SYSTEMS USA. Inc. TYPICALSVISIONARY EQUIPMENT PLANS
STAMFORD, CT 06801
VISIONARYSUITE
WHD-A.A.
MAY 28, 2017
1/4"=1'-0
2017-219-TYP
11 of 12
AutoCAD SHX Text
4
AutoCAD SHX Text
%%UMETHOD
AutoCAD SHX Text
3
AutoCAD SHX Text
2
AutoCAD SHX Text
2'-2"
AutoCAD SHX Text
1
AutoCAD SHX Text
%%USTRUCTURAL NOTES
AutoCAD SHX Text
%%UUNIVERSAL STRUT CHANNEL ASSEMBLY METHOD
AutoCAD SHX Text
EQUIPMENT
AutoCAD SHX Text
PROFESSIONAL DESIGN SERVICES
AutoCAD SHX Text
STAMFORD, CT 06902
AutoCAD SHX Text
REFERENCE
AutoCAD SHX Text
NOT FOR
AutoCAD SHX Text
CONSTRUCTION
AutoCAD SHX Text
DRAWINGS ONLY
AutoCAD SHX Text
419 WEST AVENUE
AutoCAD SHX Text
PROJECT
AutoCAD SHX Text
REVISION
AutoCAD SHX Text
SCALE
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DRAWN BY
AutoCAD SHX Text
PROJECT NUMBER
AutoCAD SHX Text
SHEET
AutoCAD SHX Text
GENERAL STRUCTURAL NOTES
.25 ' ' ( 6.4MM ) MINIMUM THICK WASHER OR ELONGATED WASHER PLATES MUST BE PLACED AT
STRUT CHANNEL LONGITUDINAL TRACK SUPPORT SHALL BE FASTENED TO STRUT CHANNEL SPANNINGMEMBER BY USE OF APPROPRIATE HARDWARE MANUFACTURED BY UNISTRUT OR POWER STRUT.
CHANNEL LOCATED BETWEEN FINISHED CEILING LINE AND TOP OF FUJIFILM MEDICAL LONGITUDINAL CLG. TRACK.EXPOSED STRUT CHANNEL LONGITUDINAL TRACK SUPPORT SHALL BE 1.625 ' ' DEEP SLOTTED
NOT TO SCALE
SLOTTED HOLES FOR THIS CONNECTION.
FUJIFILM MEDICAL STRONGLY RECOMMENDS THAT THE CUSTOMER HIRE A STRUCTURAL ENGINEER(ENGINEER OF RECORD) WHO IS REGISTERED AND CERTIFIED IN THE STATE IN WHICH THE CUSTOMER'S SITE IS LOCATED. THE ENGINEER OF RECORD SHALL BE RESPONSIBLE FOR PROVIDING AND APPROVING A DESIGN FOR ADAPTING THE SUPPORT STRUCTURE TO THE NEW OR EXISTING BUILDING STRUCTURE. THE CEILING SUPPORTSTRUCTURE INCORPORATING THE DESIGN OF THE ENGINEER OF RECORD SHALL BE SUP-PLIED AND INSTALLED BY CONTRACTOR, AS PART OF THE IRON WORK OR SPECIALTIES
A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY UNITS. ITS TOP MEMBER IS FASTENED TO THE FLOOR SLAB ( OR BUILDING STEEL ) ABOVE THE X-RAY ROOM. THE BOTTOM OF ITS LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO WHICH THE STATIONARY CEILING TRACKS ARE FASTENED WITH MOUNTING HARDWARE SUPPLIED AND INSTALLED BY FUJIFILM MEDICAL.
THE FOLLOWING GENERAL DETAILS INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND LOGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO ACCOMMODATE THE VARIOUS ARCHITECTRAL CIRCUMSTANCES
THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUALEQUIPMENT LOADS FOR EACH ROOM ARE SPECIFIED ON THE FUJIFILM MEDICAL SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
FUJIFILM MEDICAL SUGGESTS THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE WHENEVER POSSIBLE. FUJIFILM MEDICAL WILL ALSO ACCEPT THE APPLICATION OF HILTI EXPANSIONANCHORS ( OR EQUAL ) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND MATERIALS
SHALL BE DETERMINED BY ENGINEER OF RECORD. ALL SUPPORT MEMBERS, ANCHOR BOLTS,AND HARDWARE SHALL BE SUPPLIED AND INSTALLED BY CONTRACTOR. THE LENGTH OF BOLTSSHALL BE DETERMINED TO MEET SPECIFIC SITE REQUIREMENTS BY ENGINEER OF RECORD.
FUJIFILM MEDICAL WILL ACCEPT THE USE OF UNISTRUT OR POWER STRUT SUPPORT MEMBERS FOR CONSTRUCTION, ONLY IF SUPPORT STRUCTURE IS DESIGNED AND CERTIFIED BY UNISTRUT OR POWER STRUT CORPORATION, IN CONJUNCTION WITH THEENGINEER OF RECORD. THESE STRUT CHANNEL MEMBERS MUST NOT BE WELDED TOGETHER.THE APPROPRIATE AVAILABLE HARDWARE SHALL BE APPLIED.
CENTERLINES OF THE TWO STRUCTURAL MEMBERS HOLDING THE SUSPENDED WEIGHTS ARESHOWN. UNISTRUT OR POWER STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS. THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN1 16 ' ' ( 1.6MM ) AND BRACED TO PREVENT MOVEMENT IN ANY DIRECTION. THEY MUST BESECURED TO THE SLAB ABOVE AND/OR TO THE BUILDING STRUCTURE TO THE HOLD THE DEAD
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING
WEIGHTS INDICATED.
DRAWINGS IS UNDERTAKEN.
CONTRACT.
ENCOUNTERED AT EACH INSTALLATION.
- CUSTOMER IS RESPONSIBLE FOR MEETING SITE CONDITIONS FOR STATIC LOADS,
( WIDTH, DEPTH, STRENGTH, ETC...)- SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS.IS CERTIFIED BY A LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.CUSTOMER CAN ALSO UTILIZE AN EXISTING SUPPORT STRUCTURE, ONLY IF THE STRUCTURE
FINISHED CEILING LINE
SPANNING MEMBER SUPPLIEDAND INSTALLED BY CONTRACTOR
FUJIFILM MEDICAL LONGITUDINALCEILING TRACKMOUNTING SURFACE OF
FUJIFILM MEDICAL LONGITUDINALCEILING TRACK
BOTTOM OF SPANNING MEMBERFLUSH WITH FINISHED CEILING LINE
EACH MEMBER 2 '-2 ' 'O.C.
LEVELNESS, ETC...)
FUJIFILM MEDICAL SYSTEMS USA. Inc. TYPICALSVISIONARY EQUIPMENT PLANS
STAMFORD, CT 06801
VISIONARYSUITE
WHD-A.A.
MAY 28, 2017
1/4"=1'-0
2017-219-TYP
12 of 12
AutoCAD SHX Text
%%UUNIVERSAL STRUT CHANNEL ASSEMBLY METHOD
AutoCAD SHX Text
%%UNOTE
AutoCAD SHX Text
%%UIMPORTANT NOTE
AutoCAD SHX Text
%%UGENERAL STRUCTURAL CEILING SUPPORT REQUIREMENTS