page 1/17 Evaluation of new K2E K 2 EDTA Tubes Background: Greiner Bio-One has developed a newly designed MiniCollect ® tube offering an integrated collection scoop. The advantage of the new tube is that capillaries and funnels are not needed to facilitate blood transfer from the puncture site into the MiniCollect ® tube. The MiniCollect ® K 2 EDTA capillary blood collection tube is also featured with a co-molded cap which can easily be removed during the collection and sampling process. The interior of the MiniCollect ® K 2 EDTA tube wall is coated with dipotassium EDTA (K 2 EDTA). MiniCollect ® K 2 EDTA Blood Collection Tubes are used to collect, transport, store and evaluate capillary blood specimens for hematology tests. Study Objective: A clinical evaluation was carried out to compare the performance of the new MiniCollect K 2 EDTA tube in comparison to the old design of MiniCollect ® K 2 EDTA tube including 50 healthy and 65 pathological donors. Study design: The following tube types were used in this study: Sample ID Description A MiniCollect ® K2E K 2 EDTA 0.5 ml, (item No.: 450480), old design B MiniCollect ® K2E K 2 EDTA 0.25-0.5 ml (item No.: 450532), new design The study has been approved by Ethics Commission. Informed consent has been given by all participants. Directly after blood collection with venous blood, the tubes were carefully inverted 8 times according to the instructions for use for MiniCollect ® blood collection tubes. The tubes were transported to a laboratory within 6 hours after blood collection. A complete blood count was performed using the DxH 800 from Beckman Coulter. Analysis was performed with the instrument’s accompanying reagents. It is necessary to mix the samples immediately before analysis. All samples were mixed by a gentle tap immediately before analysis. Determined parameters: Leucocytes Erythrocytes Hemoglobin Hematocrit Platelets Mean cellular volume Mean cellular hemoglobin Mean cellular hemoglobin concentration Mean platelet volume (MTV) Red cell distribution with (RDW) Lymphocytes Neutrophiles Eosinophiles Monocytes Basophiles

Upload: others

Post on 22-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 1/17

Evaluation of new K2E K2EDTA Tubes

Background:

Greiner Bio-One has developed a newly designed MiniCollect® tube offering an integrated collection scoop. The advantage of the new tube is that capillaries and funnels are not needed to facilitate blood transfer from the puncture site into the MiniCollect® tube.

The MiniCollect® K2EDTA capillary blood collection tube is also featured with a co-molded cap which can easily be removed during the collection and sampling process.

The interior of the MiniCollect® K2EDTA tube wall is coated with dipotassium EDTA (K2EDTA).

MiniCollect® K2EDTA Blood Collection Tubes are used to collect, transport, store and evaluate capillary blood specimens for hematology tests.

Study Objective:

A clinical evaluation was carried out to compare the performance of the new MiniCollect K2EDTA tube in comparison to the old design of MiniCollect® K2EDTA tube including 50 healthy and 65 pathological donors.

Study design:

The following tube types were used in this study:

Sample ID Description

A MiniCollect® K2E K2EDTA 0.5 ml, (item No.: 450480), old design

B MiniCollect® K2E K2EDTA 0.25-0.5 ml (item No.: 450532), new design

The study has been approved by Ethics Commission. Informed consent has been given by all participants.

Directly after blood collection with venous blood, the tubes were carefully inverted 8 times according to the instructions for use for MiniCollect® blood collection tubes. The tubes were transported to a laboratory within 6 hours after blood collection. A complete blood count was performed using the DxH 800 from Beckman Coulter. Analysis was performed with the instrument’s accompanying reagents.

It is necessary to mix the samples immediately before analysis. All samples were mixed by a gentle tap immediately before analysis.

Determined parameters:

Leucocytes

Erythrocytes

Hemoglobin

Hematocrit

Platelets

Mean cellular volume

Mean cellular hemoglobin

Mean cellular hemoglobin concentration

Mean platelet volume (MTV)

Red cell distribution with (RDW)

Lymphocytes

Neutrophiles

Eosinophiles

Monocytes

Basophiles

Page 2: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 2/17

Conclusion:

Performance of the new MiniCollect® K2E K2EDTA tube with the co-molded cap and integrated scoop has been demonstrated in comparison to the old MiniCollect® K2E K2EDTA tube on the basis of the analytes tested.

Results for healthy donors:

A few analytical deviations have been found most likely due to separation processes of the whole blood specimen during automated measurement process. Results for pathological donors: One value of pathological donors is missing for determination of granulocytes. Four subjects have shown analytically significant deviations regarding hemoglobin and thrombocytes but in a clinically acceptable range. The correlation between the samples was found to be excellent (r > 0.99). Neither systematic deviations nor statistically significant differences have been found. High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes, however without any significant differences. Statistically significant deviations have been found in the measurement of MCV, platelets, neutrophiles without clinical significance (mean difference was in a clinical acceptable range) and comparable distribution of the data. No systematic deviations have been observed. A high distribution of basophile granulocytes has been observed in both samples. Neither statistically significant nor systematic deviations have been found between both samples. In summary, despite the deviations and results that have been found, the MiniCollect® K2E K2EDTA with the new design is substantially equivalent to the MiniCollect® K2E K2EDTA with the old design.

References:

(1) Greiner Bio-One. MiniCollect® K2E K2EDTA Tubes. Instructions for Use. Kremsmünster, Austria. 2016.

(2) Greiner Bio-One. MiniCollect® Product Manual. Kremsmünster, Austria. 2016.

(3) Guideline published by the Chamber Association for Medical Practitioners of the State of Germany concerning the quality assurance of quantitative analyses of Medical Laboratories, Germany (2001). Rev.2003

(4) ISO 6710:1995(E), Single-use containers for venous blood specimen collection. International Standard. 1995

(5) EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition, CLSI 2011.

(6) EP09-A2-IR: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline — Second Edition (Interim Revision). CLSI 2011.

(7) H01-A6: Tubes and Additives for Venous and Capillary Blood Specimen Collection; Approved Standard – Sixth Edition CLSI 2011

(8) H04-A6: Procedures and Devices for the Collection of Diagnostic Capillary Blood Specimens – Approved Standard – Sixth Edition CLSO 2011

(9) RILIBÄK: Guideline of the German Medical Association for Quality Assurance

Page 3: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 3/17

Results in detail:

Leucocytes (WBC) Normal range: 4-10 109/l Healthy donors

Donor

T/µ

l

1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 47 490

2

4

6

8

10

12 A0h

B0h

Pathological donors

Donor

T/µ

l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64

0

5

10

15

20

25

30

35 A0h B0h

Page 4: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 4/17

Erythrocytes (RBC)

Normal range: 4.4 - 5.9 T/l Healthy donors

Donor

M/µ

l

1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 47 490

1

2

3

4

5

6

7 A0h B0h

Pathological donors

Donor

M/µ

l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

1

2

3

4

5

6 A0h B0h

Page 5: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 5/17

Hemoglobin

Normal range: 13-18 g/dl Healthy donors

Donor

g/d

l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

2

4

6

8

10

12

14

16

18 HGB [g/dl] A0h HGB [g/dl] B0h

Pathological donors

Donor

g/d

l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

2

4

6

8

10

12

14

16

18

HGB [g/dl] A0h HGB [g/dl] B0h

Page 6: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 6/17

Hematocrit

Normal range: (m) 40 - 54 % (f) 37 - 47 % Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

10

20

30

40

50

60

HKT [%] A0h HKT [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

10

20

30

40

50

60

HKT [%] A0h HKT [%] B0h

Page 7: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 7/17

Thrombocytes (platelets)

Normal range: 109/l Healthy donors

Donor

G/l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

50

100

150

200

250

300

350

400

PLT [G/l] A0h PLT [G/l] B0h

Pathological donors

Donor

G/l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

50

100

150

200

250

300

350

400 PLT [G/l] A0h PLT [G/l] B0h

Page 8: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 8/17

Mean cellular volume (MCV)

Normal range: 50 - 150 fL Healthy donors

Donor

fl

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

20

40

60

80

100

120 MCV [fl] A0h MCV [fl] B0h

Pathological donors

Donor

fl

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

20

40

60

80

100

120

MCV [fl] A0h MCV [fl] B0h

Page 9: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 9/17

Mean Corpuscular Hemoglobin (MCH)

Normal range: 27 - 35 pg Healthy donors

Donor

pg

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

5

10

15

20

25

30

35 MCH [pg] A0h

MCH [pg] B0h

Pathological donors

Donor

pg

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64

0

5

10

15

20

25

30

35

40

MCH [pg] A0h MCH [pg] B0h

Page 10: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 10/17

Mean Corpuscular Hemoglobin Concentration (MCHC)

Normal range: 29-36 g/dl Healthy donors

Donor

g/d

l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

5

10

15

20

25

30

35

40

MCHC [g/dl] A0h MCHC [g/dl] B0h

Pathological donors

Donor

g/d

l

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

5

10

15

20

25

30

35

40

MCHC [g/dl] A0h MCHC [g/dl] B0h

Page 11: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 11/17

Mean platelet volume (MTV)

Normal range: 7-12 fl Healthy donors

Donor

fl

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

2

4

6

8

10

12

14

MTV [fl] A0h MTV [fl] B0h

Pathological donors

Donor

fl

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

2

4

6

8

10

12

14

MTV [fl] A0h MTV [fl] B0h

Page 12: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 12/17

Red cell distribution with (RDW)

Normal range: 12.9-18.7 % Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

2

4

6

8

10

12

14

16

18

EVB [%] A0h EVB [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

2

4

6

8

10

12

14

16

18

20

22

EVB [%] A0h

EVB [%] B0h

Page 13: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 13/17

Lymphocytes Normal range: 20-40 % Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

10

20

30

40

50

60

LYMPH [%] A0h LYMPH [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

10

20

30

40

50

LYMPH [%] A0h LYMPH [%] B0h

Page 14: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 14/17

Monocytes

Normal range: 2-10 % Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49

0

2

4

6

8

10

12

14

16

18

20

22

MONO [%] A0h MONO [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64

0

2

4

6

8

10

12

14

16

18

20

22

24

26

28

MONO [%] A0h MONO [%] B0h

Page 15: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 15/17

Eosinophile Granulocytes

Normal range: 1-4 % Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

1

2

3

4

5

6

7

8

EO [%] A0h EO [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

1

2

3

4

5

6

7

8

EO [%] A0h EO [%] B0h

Page 16: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

page 16/17

Basophile Granulocytes

Normal range: 0-1 % Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490,0

0,5

1,0

1,5

2,0

2,5

3,0

3,5

4,0

4,5 BA [%] A0h BA [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640,0

0,2

0,4

0,6

0,8

1,0

1,2

1,4

1,6

1,8

2,0

BA [%] A0h BA [%] B0h

Page 17: Evaluation of new K2E K EDTA Tubes - GBO€¦ · Sample ID Description ... High single deviations have been shown for lymphocytes, monocytes, eosinophile and basophile granulocytes,

ENT24 Rev.0 page 17/17

Neutrophiles

Normal range: 55-75% Healthy donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 490

10

20

30

40

50

60

70

80 NE [%] A0h NE [%] B0h

Pathological donors

Donor

%

1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 640

10

20

30

40

50

60

70

80 NE [%] A0h

NE [%] B0h