Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of BuprenorphineCEDIA Assay versus GC-MS and ELISA using Urine Samples from Patientsin SubstitutionTreatment Michael B6ttcher Arztpraxis f. Medizinische Mikrobiologie, Labordiagnostikund Hygiene, Dessau, Germany Olof Beck* Department of Medicine, Division of Clinical Pharmacology, Karolinska University Hospital, Stockholm, Sweden [ Abstract As buprenorphine becomes more clinically used in heroin substitution treatment, there is an increasing need for methods suitable for high-volume screening. In this study, a new immunochemical test based on CEDIA technology was evaluated for the use in clinical urine drug testing. The method was compared with an existing ELISA method and a gas chromatography--mass spectrometry (GC-MS) method on urine specimensfrom patients in heroin substitution treatment. The precision of the CEDIA assay was < 9% both within- and between- day at levels at and above the cutoff limit of 5 pg/t. The concordance in qualitative results with an existing ELISA method was 96.8%. The CEDIA measuring range was extended by diluting urine samples100-fold with saline, and the results agreed well (slope of regression line was 1.09, r2 = 0.968) with GC-MS. The sensitivity of CEDIA in detecting authentic specimen containing buprenorphine at levels > 5 pg/L was 99.5%. Cross-reactivity causing false-positive response was discovered in patients receiving prescribed dihydrocodeine. The urine concentration of total buprenorphine in urine from patients prescribed daily doses between 0.2 and 24 mg ranged from 0.5 to 2900 pg/L. The concentration of the metabolite norbuprenorphine was usually higher, and the median ratio of buprenorphine to norbuprenorphine was 0.23 (95% were below 1). We conclude that the CEDIA assay is suitable for application in high-volume screening of buprenorphine for urine drug testing. Introduction Buprenorphine (Subutex | is used as an analgesic drug and for detoxification and substitution therapy of opioid de- * Author to whom correspondence should be addressed: Dr. Olof Beck, Department of Medicine, Division of Clinical Pharmacology, Karolinska University Hospital, SE-17176 Stockholm, Sweden. pendence (1,2). Since 1996, starting in France, the use of buprenorphine in substitution therapy has increased and spread to over 30 countries, many in Europe but also in Asia, Australia, and the Americas. Buprenorphine offersan attractive alternative to methadone, with benefits including increased safety for respiratory depression, suppressed heroin use, and possibility of longer dosing interval (2). However,in combina- tion with the increased medical use, buprenorphine also occurs on the black market as an illicit drug (3,4), and fatalities due to poly-drug use have been reported (5,6). Buprenorphine is metabolized by dealkylation to nor- buprenorphine primarily by the cytochrome P450 isozyme CYP 3A4 (7). Both buprenorphine and norbuprenorphine are conjugated with glucuronic acid and subsequently excreted in urine during the course of several days (8). In urine drug testing, parent buprenorphine has become the primary ana- lytical target compound, and less focus has been devoted to the metabolite norbuprenorphine. The proportion of the two compounds in urine has not been well-documented in pa- tient material. The activity of the CYP 3A4 isozyme is sub- jected to variability influenced by both genetic and environ- mental factors, which will determine the proportion of norbuprenorphine that is being formed and subsequently ex- creted in urine (9,10). As a consequence of increased legal and illicit use of buprenorphine, there is a need for an analytical method for tox- icological monitoring of patients. In addition, an analytical method is necessary for compliance monitoring during sub- stitution therapy, which has been integrated in many methadone programs. A number of methods for urinary de- termination of buprenorphine are available, including radioimmunoassay, enzyme-linked immunosorbent im- munoassay (ELISA), thin-layer chromatography, high-pres- sure liquid chromatography, gas chromatography (GC), gas chromatography-mass spectrometry (GC-MS), and liquid Reproduction (photocopying) of editorialcontent of thisjournal is prohibited withoutpublisher's permission. 769

Upload: others

Post on 15-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Journal of Analytical Toxicology, Vol. 29, November/December 2005

Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ELISA using Urine Samples from Patients in Substitution Treatment

Michael B6ttcher Arztpraxis f. Medizinische Mikrobiologie, Labordiagnostik und Hygiene, Dessau, Germany

Olof Beck* Department of Medicine, Division of Clinical Pharmacology, Karolinska University Hospital, Stockholm, Sweden

[ Abstract As buprenorphine becomes more clinically used in heroin substitution treatment, there is an increasing need for methods suitable for high-volume screening. In this study, a new immunochemical test based on CEDIA technology was evaluated for the use in clinical urine drug testing. The method was compared with an existing ELISA method and a gas chromatography--mass spectrometry (GC-MS) method on urine specimens from patients in heroin substitution treatment. The precision of the CEDIA assay was < 9% both within- and between- day at levels at and above the cutoff limit of 5 pg/t. The concordance in qualitative results with an existing ELISA method was 96.8%. The CEDIA measuring range was extended by diluting urine samples 100-fold with saline, and the results agreed well (slope of regression line was 1.09, r 2 = 0.968) with GC-MS. The sensitivity of CEDIA in detecting authentic specimen containing buprenorphine at levels > 5 pg/L was 99.5%. Cross-reactivity causing false-positive response was discovered in patients receiving prescribed dihydrocodeine. The urine concentration of total buprenorphine in urine from patients prescribed daily doses between 0.2 and 24 mg ranged from 0.5 to 2900 pg/L. The concentration of the metabolite norbuprenorphine was usually higher, and the median ratio of buprenorphine to norbuprenorphine was 0.23 (95% were below 1). We conclude that the CEDIA assay is suitable for application in high-volume screening of buprenorphine for urine drug testing.

Introduction

Buprenorphine (Subutex | is used as an analgesic drug and for detoxification and substitution therapy of opioid de-

* Author to whom correspondence should be addressed: Dr. Olof Beck, Department of Medicine, Division of Clinical Pharmacology, Karolinska University Hospital, SE-17176 Stockholm, Sweden.

pendence (1,2). Since 1996, starting in France, the use of buprenorphine in substitution therapy has increased and spread to over 30 countries, many in Europe but also in Asia, Australia, and the Americas. Buprenorphine offers an attractive alternative to methadone, with benefits including increased safety for respiratory depression, suppressed heroin use, and possibility of longer dosing interval (2). However, in combina- tion with the increased medical use, buprenorphine also occurs on the black market as an illicit drug (3,4), and fatalities due to poly-drug use have been reported (5,6).

Buprenorphine is metabolized by dealkylation to nor- buprenorphine primarily by the cytochrome P450 isozyme CYP 3A4 (7). Both buprenorphine and norbuprenorphine are conjugated with glucuronic acid and subsequently excreted in urine during the course of several days (8). In urine drug testing, parent buprenorphine has become the primary ana- lytical target compound, and less focus has been devoted to the metabolite norbuprenorphine. The proportion of the two compounds in urine has not been well-documented in pa- tient material. The activity of the CYP 3A4 isozyme is sub- jected to variability influenced by both genetic and environ- mental factors, which will determine the proportion of norbuprenorphine that is being formed and subsequently ex- creted in urine (9,10).

As a consequence of increased legal and illicit use of buprenorphine, there is a need for an analytical method for tox- icological monitoring of patients. In addition, an analytical method is necessary for compliance monitoring during sub- stitution therapy, which has been integrated in many methadone programs. A number of methods for urinary de- termination of buprenorphine are available, including radioimmunoassay, enzyme-linked immunosorbent im- munoassay (ELISA), thin-layer chromatography, high-pres- sure liquid chromatography, gas chromatography (GC), gas chromatography-mass spectrometry (GC-MS), and liquid

Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission. 769

Page 2: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

chromatography-mass spectrometry (LC-MS) (11-20). How- ever, automated high-volume screening tests for buprenor- phine have not been available. The aim of this study was to evaluate whether a new buprenorphine CEDIA assay (cloned enzyme-donor immunoassay) could fulfill the need for a rapid and cost-effective urinary screening assay.

Materials and Methods

Chemicals Buprenorphine, morphine, morphine-3-glucuronide,

codeine, norbuprenorphine, dihydrocodeine, hydrocodone, buprenorphine-d4, and norbuprenorphine-d3 were obtained as methanol solutions from Promochem GmbH (Wesel, Ger- many). Codeine-6-glucuronide, dihydrocodeine-6-glucuronide, dihydromorphine, dihydromorphine-3-glucuronide, dihydro- morphine-6-glucuronide, dihydronormorphine, and dihy- dronorcodeine were obtained as solid material from Lipomed GmbH (Bad S~ickingen, Germany) and dissolved in pure water or methanol. Working solutions were prepared by dilutions with blank urine or saline (0.9%). The experiment for ob- taining cross-reactivity data used typically 4-7 prepared solu- tions for each compound using linear regression analysis. For compounds with low cross-reactivity, one data point was used.

Helix pomatia I~-glucuronidase/arylsulfatase (supplied in liquid form) was obtained from Roche GmbH (Mannheim, Ger- many); acetic anhydride was from Fluka (Buchs, Switzerland); methanol, ethyl acetate (suprasolve grade), sodium acetate, ammonia (32%), and acetic acid were from Merck GmbH (Darmstadt, Germany); acetonitrile was from Riedel-de Haen (Seelze, Germany); and 4-dimethylaminopyridine (crystalline grade) was from Sigma Chemical (Taufkirchen, Germany). All chemicals were of analytical grade unless otherwise stated.

Urine samples Urine samples were collected from outpatients in heroin

substitution treatment with methadone, buprenorphine, or dihydrocodeine assumed to be in steady-state. In total, 1552 samples were obtained from approximately 600 patients, con- sisting of 70% males and having an age range of 18-54 years. The samples were collected under supervision and sent to the laboratory for routine urine drug testing. Three different cri- teria were used for inclusion in various parts of the study: pre- scription of Subutex with known daily dose, prescription of other substitution medication (methadone or dihydrocodeine), or a specific request of urine buprenorphine testing. The ethics committee at the Karolinska Institute approved the study.

CFDIA assay Reagents for the CEDIA buprenorphine test were supplied by

Microgenics (Fremont, CA). The test was performed on Hitachi 911 and 912 (Roche Diagnostics, Indianapolis, IN) in- struments with a test protocol supplied by Microgenics and used a sample volume of 10 tJL. Calibrators (levels: 0, 5.8, 21.3, 51.5, and 78.2 pg/L) and controls (levels: 3.9, 6.4, 40.2, and 65.1 IJg/L) supplied by Microgenics were used. During the study,

770

Journal of Analytical Toxicology, Vol. 29, November/December 2005

calibration was done on a daily basis together with four dif- ferent quality control samples every 25th sample and following calibration.

Reagent for opiate CEDIA screening (cutoff 300 ng/mL) was obtained from Microgenics and was applied on a Hitachi 911 instrument according to kit instructions.

ELISA assay The ELISA assay of buprenorphine was performed using

reagents and calibrators from Diagnostix (Mississauga, ON, Canada). The procedure was performed manually according to kit instructions, and the final result was obtained by using a Labsystems Multiskan (Helsinki, Finland) plate reader (450 nm). A cutoff limit of 5 pg/L was used. Calibrating points used were 0, 5, 10, and 20 pg/L.

GC-MS procedure A 3-rnL aliquot of the urine sample was fortified with 0.05

mL internal standard solution (methanol containing 50 ng buprenorphine-d4 and 100 ng norbuprenorphine-d3), 0.3 mL of 1.7M acetate buffer, pH 4.75 (pH was controlled to be 4.5-5.0), and 30 IJL (3000 Fishman units) of Helixpornatia enzyme. The mixture was vortex mixed and incubated overnight in a water bath at 37~ (or at 2 h at 56~ This procedure was docu- mented to be optimal for the hydrolysis of both buprenor- phine and norbuprenorphine (21).

The analytes were then extracted using a solid-phase ex- traction (SPE) procedure. The mixed phase (nonpolar/SCX) SPE cartridge (SPEC DAU 3 mL, Varian, Darmstadt, Germany) was activated by treatment with 0.2 mL methanol, followed by application of the sample. The procedure was performed using a vacuum manifold (approximate flow 1 mL/min). Washing was performed with 0.5 mL 0.1M acetic acid and subsequently 0.5 mL methanol. The cartridge was vacuum dried with aspi- ration for 5 rain. Subsequently, the elution of analytes was performed with 0.8 mL freshly prepared methanol/ammonia (9:1, v/v). Thereafter, 25 pL of 4-dimethylaminopyridine (0.5 mg/mL in acetonitrile) was added, and the extract was evapo- rated to dryness under nitrogen (22). The dry residue was redissolved in 100 pL acetic anhydride, vortex mixed, screw- capped, and incubated for 5 rain in a 500 W microwave oven. The reaction mixture was evaporated at 65~ in a heating block under nitrogen, the residue redissolved in 701JL of ethyl acetate, and 1 IJL was injected into the GC-MS.

The final extract was analyzed using a QP5050 Shimadzu GC-MS using automated splitless injection (280~ (au- tosampler AOC20s). The column was a 15-m (0.25-ram i.d., 0.25-pro film thickness) DB-5MS (J&W Scientific, Promochem GmbH) with helium (grade 5.0) as carrier gas. Initial temper- ature was 130~ for I rain, then increased 30~ to a final temperature of 300~ The instrument operated in the selected ion monitoring mode measuring m/z 420 (quantitation ion), 452, and 453 (qualifier ions) for buprenorphine; m/z 424 (quan- titation ion), 456, and 457 (qualifier ions) for buprenorphine- d4; m/z 440 (quantitation ion) and 422 (qualifier ion) for nor- buprenorphine; and m/z 443 (quantitation ion) and 425 (qualifier ion) for norbuprenorphine-d3.

Standards were prepared in blank pooled urine covering the

Page 3: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Journal of Analytical Toxicology, Vol. 29, November/December 2005

concentration range from 2 to 2000 pg/L. The limits of detection (LOD) and quantification (LOQ) were

calculated according to DIN 32645 (23) from the lower end of the calibration curves using B.E.N. software version 2.03 (Arvecon GmbH, Heidelberg, Germany). Linearity for buprenorphine and norbuprenorphine was shown in the range from 2 to 2000 pg/L with this software. This measuring range was extended by 1:5 dilution of urine with saline. For buprenor- phine, the LOD was 0.5 pg/L, the LOQ 1.5 pg/L, and the cor- relation coefficient r 2 = 0.9998. For norbuprenorphine, the LOD was 2 pg/L, the LOQ 6 pg/L, and the correlation coeffi- cient r 2 = 0.9985. Quality control of the GC-MS method was performed using certified urine control material (Medichem, Stuttgart, Germany) in every series with target values at 18.1 pg/L for buprenorphine and 84.8 pg/L for norbuprenorphine. The observed mean value for buprenorphine was 19.4 pg/L [coefficient of variation (CV) 9.0%, n = 13] and for nor- buprenorphine 84.9 pg/L (CV 9.2%, n = 12).

The method has been part of the GTFCh (German Society for Toxicological and Forensic Chemistry) proficiency test programme since 2001 and gives comparable results to other laboratories using GC-MS and LC-MS methods. Representa- tive chromatograms are shown in Figure 1.

Results

Validation of the CEDIA assay The within- and between-day variability in quantification at

5 different levels was studied using the 4 control samples and

the cutoff calibrator supplied by Microgenics. At the levels ex- ceeding the 5 pg/L cutoff limit, a variability lower than 11% was observed (Table I).

The CEDIA calibrators were compared to laboratory made urine standards used for the GC-MS assay in the concentration range 3-50 pg/L, and the recovery was 92% (n = 5, CV 4.0%). The result from CEDIA analysis of the certified urine control material used for the GC-MS method with assigned value of 18.1 pg/L buprenorphine was 17.5 pg/L (mean, n = 8, between- day CV 3.0%).

The CEDIA response curve in the measuring interval was documented using calibrators from Microgenics covering the range from 0 to 78.2 pg/L (Figure 2).

The agreement of the CEDIA and ELISA assays was studied in 221 samples (Table II). There was a 96.8% agreement in qualitative results between the methods. In three samples (2.7% of all positives), CEDIA produced a false-positive

Table I. Variability in the CEDIA Quantification of Buprenorphine in Urine

Assigned Within-Day Between-Day Value Observed mean CV (pg/L) N value (pg/L) (%) N

Observed mean CV value (pg/L) (%)

3.9 3.1 17.0 10 3.9 11.0 32 5.8 6.2 6.3 10 5.6 8.1 10 6.4 6.4 4.1 10 6.8 8.0 31

40.2 40.9 2.7 10 40.7 3.0 31 65.1 64.7 11.0 10 67.7 7.0 29

A B

Bup

. J ~ t # z 42o

nor-Bup

910 915 915 10.0 Time (rain) Time (min)

nor-Bup

9'.o 91s 91s Time (min)

, v z j ~

ld.o "lime (rain)

Bup,,.d 4

.Vz4~

nor-Bup-d3

__ . ~ . m~' 42e

9~o gls 91s lo',o Time (min) Time (min)

BuP~4

9.0 9.6 Time (rain)

nor-Bup-d3

J

s:s to'.o Time (rain)

Figure 1. Chromatograms from the GC-MS analysis of two authentic patient samples: patient prescribed 0.4 mg Subutex daily and had a concentration of buprenorphine of 9.5 pg/L and of norbuprenorphine of 56 pg/L (creatinine was 286 mg/dL) (A) and patient prescribed 4 mg Subutex daily and had a concen- tration of buprenorphine of 1.9 tJg/L and of norbuprenorphine of 21 pg/L (creatinine was 46 mg/dL) (B).

771

Page 4: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Journal of Analytical Toxicology, Vol. 29, November/December 2005

4500

m 3000

@ "o

1500 . . . . . . . . . . . . . . . . 0 10 20 30 40 50 60 70 80

Concentration (pg/L)

Figure 2. This CEDIA response curve in the range of 0-78 pg/L in urine ma- trix was obtained using kit calibrators as applied on a Hitachi 912 instru- ment. The response was non-linear and continued to increase at levels ex- ceeding 100 pg/L. In the instrument application the calibration was performed using non-linear regression.

Table II. The Agreement of CEDIA and ELISA Assays in Detecting Buprenorphine in Authentic Patient Urine Samples at the Cutoff Limit of 5 pg/L

ELISA

Negative Positive

CEDIA Negative 105 4* Positive 3 t 109

* GC-MS: no buprenorphine was detected in two samples and two contained 5 and 26 pg/L, respectively.

* GC-MS: no buprenorphine detected.

response as the GC-MS confirmation was clearly negative. In one sample, the CEDIA result was false negative as the sample contained 26 ng/mL of buprenorphine according to GC-MS.

The agreement between the CEDIA and GC-MS methods in quantifying buprenorphine was studied in the range from 0 to 6000 pg/L using 300 samples collected from 72 patients with known Subutex doses. The CEDIA measuring range was extended by dilution of samples 10- or 100-fold with saline. The slope of the regression line was 1.09 with an intercept of-13 on the CEDIA axis (Figure 3A). The median ratio between CEDIA and GC-MS was 0.96 (n = 300). In the sample with the highest concentration of buprenorphine, a considerable and repro- ducible mismatch between methods was observed (GC-MS 2936 pg/L and CEDIA 5990 pg/L). In the undiluted samples, it was again observed that the quantitative results of the two methods were correlated, but low values from CEDIA were observed in the upper part of the measuring range.

By pooling data for all samples with a buprenorphine con- centration > 5 pg/L according to GC-MS or ELISA (n = 400), the sensitivity of CEDIA for detecting positive sampies at the cutoff level of 5 pg/L was calculated to be 99.5%.

A total of 1011 samples from about 400 different patients who were not receiving prescribed buprenorphine but received other heroin substitution treatments was used to estimate specificity of the CEDIA assay and to estimate frequency of abuse in this population. Thirty samples (3.0%) were found to be positive by CEDIA at the cutoff level of 5 pg/L. One of these had detectable buprenorphine by GC-MS but at a low level (0.6 pg/L together with 80 pg/L norbuprenorphine). Fourteen of these 30 samples were from patients prescribed dihydrocodeine and contained high amounts of this substance and metabolites (according to GC-MS), indicating a cross-reactivity. An additional 20 samples selected from different patients prescribed dihydrocodeine in

A 3000 r . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . r---~

, / 1~176176 .5/ . [

/ , . u a y= 1.o9,,--13 02l/- . . . . , , ; o , , ~

. J l

t , .

150

100

B : ~gQ*

�9 % '

�9 A A

e �9 a �9 �9

�9 : j J ' . " : . : . �9 50~ "*S' :" "

�9 �9 �9 �9 . oe ~ �9

: ",0~

0 500 1000 1500 2000 2500 3000 50 100 150 Buprenorphine GC-MS (lag/L) Buprenorphine GC-MS (lag/L)

Figure 3. The agreement of CEDIA and GC-MS in quantifying buprenorphine in authentic patient urine samples. The CEDIA measuring range was extended by diluting urine with saline; all data are presented with the exception of one outlier, which was not included (n = 299, see text) (A). Data are from samples in the low concentration range (n = 157); the slope of the linear regression curves for different ranges was below 1 because of low values from CEDIA in the high end of its measuring range; the highest correlation coefficient (0.970) was obtained for data in the range 0-50 pg/L (slope 0.85); and the circle data point rep- resents values obtained in undiluted samples and triangle data points values after 10-fold dilution (B).

772

Page 5: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Journal of Analytical Toxicology, Vol. 29, November/December 2005

daily doses of 14--80 mg were therefore investigated, and all were false positive in the CEDIA assay with responses ranging from 5 to 45 lag/L. In 15 of the other false-positive samples (1.5%), no cause for the response could be revealed in this in- vestigation. However, 14 of these samples were out of range in the CEDIA opiate immunoassay, which could indicate a cross-re- activity from codeine (see Table III). Most of these samples had a low (5.0-9.1 lag/L) CEDIA response for buprenorphine in com- bination with a high creatinine concentraion.

The cross-reactivity of various analogue substances was studied (see Materials and Methods) in urine matrix and the data is presented in Table III. In immunoassays, one has to con- sider that different cross-reacting substances, when present si- multaneously and at different concentrations, do not always add to the assay response in an arithmetical or predictable way (24). We therefore conducted an experiment with a saline sample spiked with reference materials to the following con- centrations: 7.14 mg/L dihydromorphine- 3-glucuronide, 7.14 mg/L dihydro- morphine-6-glucuronide, 14.3 mg/L dihydronorcodeine, 14.3 mg/L dihy- drocodeine-6-glucuronide, 14.3 mg/L codeine, 14.3 mg/L dihydromorphine, and 14.3 mg/L dihydronormorphine. The CEDIA assay gave a result of 5.2 ]~g/L, which corresponds to the expected result if the total cross-reactivity would be cal- culated from the individual substance and its cross-reactivity at the given concen- trations.

As can be seen in Figure 4A, urine sam- ples from patients on Subutex therapy may contain levels of buprenorphine below the 5 lag/L cutoff limit. In this ma- terial, 4.7% of the samples had such low levels, giving a calculated sensitivity for detecting buprenorphine intake of 95.3%.

By applying a 2 lag/L cutoff limit, an increase in sensitivity to 98.3% could have been obtained. However, when doing that, the rate of false-positive results for CEDIA increased from 3% to 15%, which was observed in urine samples from patients re- ceiving substitution treatments other than Subutex and may be the result of more complex sample matrices that contain ad- ditional drugs in high concentrations.

Pharmacokinetic observations The concentration of buprenorphine and norbuprenorphine

in the 300 samples from patients prescribed Subutex ranged from 0.5 to 2936 laglL and 4.0 to 4462 ]JglL, respectively (Figure 4). A majority (51%) of the samples had a buprenor- phine concentration below 100 IJg/L. After correcting the urine concentrations with creatinine, a dose-dependent urinary ex- cretion of both buprenorphine and norbuprenorphine could be observed (Figure 5).

Table III. Cross-Reactivity of Related Substances in the CEDIA Assay

Tested Linear Concentration Range Regression Cross-Reactivity

Substance (mg/L) (r z) (%)

Morphine 0.78-50 0.990 0.010 Morphine-3-glucuronide 50 - 0.002 Norcodeine 50 - 0.005 Codeine 6.25-50 0.997 0.016 Codeine-6-glucuronide 6.25-I00 0.991 0.010 Dihydrocodeine 12.5-I00 0.995 0.014 Dihydrocodeine-6-glucuronide 6.25-I00 0.995 0.010 Dihydromorphine 12.5-I00 0.983 0.007 Dihydromorphine-3-glucuronide 12.5-I 00 0.992 0.005 Dihydromorphine-6-glucuronide 12.5-I00 0.974 0.005 Dihydronormorphine 12.5-I 00 0.989 0.004 Hydrocodone 6.25-I00 0.999 0.013 Dihydronorcodeine 12.5-I00 0.998 0.004 Norbuprenorphine 0.31-5.0 0.984 0.08

~- 15o

| |

--~.100

L,,

A B

100 1000 2000 3000 Buprenorphine (pg/L)

L 40

== 30

'~ 20

= lo

100 1000 2000 3000 4000 Norbuprenorphine (pg/L)

Figure 4. The distribution of urinary buprenorphine and norbuprenorphine concentrations obtained from 72 patients receiving from 0.2 to 24 mg buprenorphine per day.

773

Page 6: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Buprenorphine concentrations were generally lower than norbuprenorphine with a median and mean value of the ratio of 0.23 and 0.34, respectively, and was found not to be related to the dose (Figure 6 and Table IV). Fifteen samples from 8 pa- tients had a ratio > 0.99. One patient with a constant dose of 4 mg/day produced 6 of 12 samples with a ratio between 1.21 and 3.16. The ratio between buprenorphine and norbuprenorphine was found to be variable within some individuals, whereas in others, the ratio appeared to be more stable (Figure 7).

Of the 300 urine samples from Subutex-treated patients, 21% were found to be positive for opiates. These samples were from 44% of the patients, indicating relapse into heroin use during treatment. No difference existed between opiate positive and negative subgroups regarding dose (median dose 8.0 for both), buprenorphine/norbuprenorphine ratio (Table IV), or buprenorphine and norbuprenorphine excretion.

r _=

.J O1

_== ,,r

O e -

P O.

m

500 n = 2 6

400

300.

200

t00

A 35 152 87 26 35 1~2 ~'7 1500

_L 1201 pg/g 3@ 0 8~--L~'Q .E

U

1000 .J O)

.E

o Z 0 .2 - 2 .0 - 4 .0 - 12 - 0 .2 - 2 .0 - 4 .0 - t 2 - 1.2 3.4 11 24 1.2 3.4 11 24

Dose (mg/d)

Figure 5. The urinary excretion of buprenorphine and norbuprenorphine in patients categorized according to dose level (300 samples from 72 patients). The boxplots show the upper and lower quartile and the line (or inserted number) describes the data range.

.o

r

t .

Z

30 t 2.0

1,0

i

0.5

i 0

, . , , , �9 , �9 , . ,

4 8 12 16 20 24 Buprenorphine dose (mg/d)

Figure 6. The urinary buprenorphine/norbuprenorphine ratio presented as a function of the dose level (300 samples from 72 patients). Fifteen sam- ples from 8 patients had ratios above 1 (5%).

Journal of Analytical Toxicology, Vol. 29, November/December 2005

Discussion

This work demonstrates that the newly developed CEDIA assay for buprenorphine in urine is well suited for application in clinical testing and thereby provides the first high-volume screening test for buprenorphine. It was demonstrated that the quantitative results are in good agreement with GC-MS and that the measuring range can be extended by simply diluting the urine sample with saline. A 1:10 dilution was performed in 44% of the samples and 1:100 dilution in 3.7%. This possibility might be useful from the aspect of finding a urine sample that has been spiked with buprenorphine and are likely to contain much higher concentrations. Proof of tampering by lack of norbuprenorphine or an abnormally high buprenorphine/nor- buprenorphine ratio can be obtained from the confirmation analysis in such cases. The observation of low values for buprenorphine at the high end of the CEDIA measuring range, as compared to GC-MS, was surprising and might suggest that a more limited measuring range should be used. Tenfold dilution could be performed at responses above 50 to improve that quantitative correlation to GC-MS. A modified test with an extended measuring range (0-750 lag/L) would cover 96.3% of samples.

From a qualitative aspect, the CEDIA assay compares both with the existing ELISA test and GC-MS at the cutoff level at 5 lag/L. Unwanted cross-reactivity in samples from patients receiving dihydrocodeine was discovered but always with a re- sponse g 45 lag/L. Several metabolites of dihydrocodeine were studied, but all displayed a rather low cross-reactivity. Ac- cording to present knowledge of dihydrocodeine metabolism (25,26), the main metabolites were included in the cross- reactivity experiment. It is most likely that several dihy- drocodeine metabolites contributed to the observed cross-re- activity in the CEDIA assay. Because the slope of the CEDIA/GC-MS regression line was close to 1 and the cross- reactivity of norbuprenorphine (Table III) was low, it can be concluded that the CEDIA response corresponds to the sum of buprenorphine and conjugated buprenorphine only. Nor- buprenorphine does not contribute to the response in the CEDIA assay. This feature of having buprenorphine as the main analyte is similar to the ELISA assay (19).

Cirimele and co-workers (19) validated the ELISA assay used for comparison, against an LC-MS procedure. False-positive re- suits were observed at the 2 lJg/L but not at the 5 lag/L cutoff limit in that study, which is in some contrast to our findings. However, this might simply be related to the differences in material studied. The rate of false-positive results was low for

Table IV. Ratio of Buprenorphine to Norbuprenorphine in Opiate Negative and Positive Subgroups

All Samples Opiate Negative Opiate Positive

Mean 0.34 0.34 0.33 Median 0.23 0.22 0.24 5% Conf. limit 0.08 0.07 0.09 95% Conf. limit 0.99 1.07 1.03 N 300 237 63

774

Page 7: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Journal of Analytical Toxicology, Vol, 29, November/December 2005

both assays. One advantage with the CEDIA assay is the better precision that can be achieved at the cutoff limit. However, it has been observed in the authors' laboratories that the preci- sion of the CEDIA buprenorphine test during routine use is not as good as most other CEDIA urine tests for drugs of abuse.

Both GC-MS and LC-MS methods have been published for buprenorphine and norbuprenorphine measurement (16-18,20). The procedure described in this paper has the novel feature of using acetylation with 4-dimethylaminopyridine as the catalyst. This enabled a high yield of acetyl derivatives that had favorable GC and MS characteristics.

The present CEDIA assay was primarily intended for use with a 5 pg/L cutoff limit. In this study, a majority (96%) of the samples from patients being prescribed buprenorphine was above this limit. The biological noise in samples from other substitution treatments (mainly methadone) was significant because of unspecific reactivity or from other cross-reacting substances, and application of 2 pg/L as cutoff resulted in an unacceptably high rate of false-positive results. However, it is our experience that when monitoring Subutex-treated pa- tients, a cutoff limit of 2 IJg/L can be applied. In the material studied, this would have decreased the false-negative results by 75%, from 12 to 3 out of 300. In the study by Cirimele and co- workers (19), it was concluded that a 2 tJg/L cutoff limit was preferable over a 5 l~g/L cutoff limit because of minimal false- negative results. Because buprenorphine is available on the black market and is a subject of non-medical use, there is a need for including buprenorphine in general toxicology screening. The optimal cutoff limit for this application is un- known and cannot be estimated from present data. However, the observation that patients receiving prescribed buprenor- phine may test negative may indicate that a lower cutoff level than 5 pg/L should be used in compliance testing. A 1 IJg/L cutoff limit was previously proposed (13).

One essential clinical effect of buprenorphine is to replace and reduce heroin intake. It was therefore, unexpected to find a rather high opiate-positive rate (21%) in the samples from Subutex-treated patients. However, according to the meta-

A

o 3:

0 v v v �9 v

�9 v v

O v �9 v � 9 v e "

�9 �9 �9 � 9 v

v � 9 v �9 v v v �9 �9 O v

v Q

~ v �9

�9 v �9 ~ v u � 9 C ~ , v v ,~ �9 �9 , n , �9 v � 9 �9 ~

e-l=. v " �9 �9 v ~,, v v v" I I � 9 �9 .$ v t �9 u o �9 9 g 6 0

v v e,~ �9 � 9 w e �9 � 9 �9 - �9 �9 9 �9 �9 g

" .. ! o, o~ . ." . | . . . t ' . . . �9 .. . . , 0 " : . o , : : . ; , ~. S, o . . . . , , l i l t s . "1 "'. -! "o', " , !~,'~ .'"'o t I'..i't�9 i ' , , ,~l ..i:

10 20 30 4 0 50 60 70

P a t i e n t n u m b e r

analysis of buprenorphine made by Mattick and co-workers (2), the reduction of heroin use is dose-dependent and is ob- served at high doses (8 rag/day). However, because the mean dose in this study was 8 rag/day, both in the whole group and in the subgroup with opiate positive samples, this was some- what surprising.

The median ratio of buprenorphine and norbuprenorphine was found to be 0.23. However, there was a significant inter- and intra-individual variability. Because the elimination kinetics are slower for norbuprenorphine (27), the ratio is likely to be de- pendent on the time between sampling and dose intake. From this, one could expect high ratios to be related to low concen- tration. This was not observed in this material. Another factor is variability in activity of the enzyme responsible for the for- mation of norbuprenorphine. The CYP 3A class of enzymes is subjected to influence from both genetic and environmental fac- tors, and a significant variability of norbuprenorphine formation is therefore to be expected (9). The observed buprenorphine/ norbuprenorphine ratios and the dose-dependent urinary ex- cretion are in agreement with a recent report (28).

In conclusion, this study has shown that the CEDIA buprenorphine assay is suitable for high-volume screening of urine. The optimal cutoff limit for buprenorphine cannot be determined at present and will require future studies. At pre- sent we would recommend the use of a 5 pg/L cutoff limit. In connection with monitoring of patients on a low dose or in the case of dilute urine sample, a 2 IJg/L cutoff may be considered. In monitoring patients for buprenorphine, caution has to be taken for weak positive samples in combination with a positive opiate test as it may be due to cross-reactivity from dihy- drocodeine and possibly codeine.

References

A

20 v

GI

15 U)

o

o

Figure 7. Urinary buprenorphine/norbuprenorphine ratios and individual buprenorphine doses for all 72 patients are presented. Samples were collected over a 5-month period. Changes in dosing were based on clinical considerations.

1. E.J. Cone and K.L. Preston. Toxicological aspects of heroin sub- stitution treatrment. Ther. Drug Monit. 24:193-198 (2002).

2. R.P. Mattick, J. Kimber, C. Breen, and M. Davoli. Buprenorphine maintenance versus placebo or methadone mainte- nance for opioid dependence. Cochrane Database Syst. Rev. 4" CD002207 (2003).

3. Y. Obadia, V. Perrin, I. Feroni, D. Vlahov, and J.-P. Moatti. Injecting misuse of buprenorphine a~ong French drug users. Addiction 96:267-272 (2001).

4. World Health Organization Expert Com- mittee on Drug Dependence. World Health Organ. Tech. Rep. Ser. 915:1-26 (2003).

5. A. Tracqui, P. Kintz, and B. Ludes. Buprenorphine-related deaths among drug addicts in France: a report on 20 fa- talities. J. Anal. Toxicol. 22:430-434 (1998).

6. P. Kintz. Deaths involving buprenorphine: a compendium of French cases. Forensic Sci. Int. 121:65-69 (2001).

7. K. Kobayashi, T. Yamamoto, K. Chiba, M. Tani, N. Shimada, T. Ishizaki, and Y. Kuroiwa. Human buprenorphine

775

Page 8: Evaluation of Buprenorphine CEDIA Assay versus GC-MS and ......Journal of Analytical Toxicology, Vol. 29, November/December 2005 Evaluation of Buprenorphine CEDIA Assay versus GC-MS

Journal of Analytical Toxicology, Vol. 29, November/December 2005

N-dealkylation is catalysed by cytochrome P450 3A4. Drug Metab. Dispos. 26:818-821 (1998).

8. E.J. Cone, C.W. Gorodetzky, D. Yousefnejad, W.F. Buchwald, and R.E. Johnson. The metabolism and excretion of buprenor- phine in humans. Drug Metab. Dispos. 12" 577-581 (1984).

9. J.K. Lamba, Y.S. Lin, E.G. Schuetz, and K.E. Thummel. Genetic contribution to variable human CYP3A-mediated metabolism. Adv. Drug Deliv. Rev. 54:1271-1294 (2002).

10. L. Wojnowski. Genetics of the variable expression of CYP3A in humans. Ther. Drug Monit. 26" 192-199 (2004)

11. C.W. Hand, K.E. Ryan, S.K. Dutt, R.A. Moore, J. O'Connor, D. Talbot, and H.J. McQuay. Radioimmunoassay of buprenor- phine in urine: studies in patients and in a drug clinic. J. Anal. Toxicol. 13" 100-104 (1989).

12. D. Martinez, M.C. Jurado, and M. Repetto. Analysis of buprenor- phine in plasma and urine by gas chromatography. J Chromatogr. 528" 459-463 (1990).

13. L. Debrabandere, M. Van Boven, and P. Daenens. Development of a radioimmunoassay for the determination of buprenorphine in biological samples. Analyst 118:137-143 (1993).

14. E. Schleyer, R. Lohmann, C. Rolf, A. Gralow, C.C. Kaufmann, M. Unterhalt, and W. Hiddemann. Column-switching solid-phase trace-enrichment high-performance liquid chromatographic method for measurement of buprenorphine and norbuprenor- phine in human plasma and urine by electrochemical detection. J. Chromatogr. 614" 275-283 (1993).

15. G. Alemany, A. Gamundi, C. Rossello, and M.C. Nicolau. A simple thin-layer chromatographic method for the analysis of buprenorphine in urine. Biomed. Chromatogr. 10:146-147 (1996).

16. A. Tracqui, P. Kintz, and P. Mangin. HPLC/MS determination of buprenorphine and norbuprenorphine in biological fluids and hair samples. J. Forensic Sci. 42" 111-114 (1997).

17. A.M. Lisi, R. Kazlauskas, and G.J. Trout. Gas chromatographic- mass spectrometric quantitation of urinary buprenorphine and norbuprenorphine after derivatization by direct extractive alky- lation. J. Chromatogr. B 692:67-77 (1997).

18. F. Vincent, J. Bessard, J. Vacheron, M. Mallaret, and G. Bessard. Determination of buprenorphine and norbuprenorphine in urine and hair by gas chromatography-mass spectrometry. J. Anal. Tox- icol. 23:270-279 (1999).

19. V. Cirimele, P. Kintz, S. Lohner, and B. Ludes. Enzyme im- munoassay validation for the detection of buprenorphine in urine. J. Anal. Toxicol. 27:103-105 (2003).

20. R. Kronstrand, T.G. Selden, and M. Josefsson. Analysis of buprenorphine, norbuprenorphine, and their glucuronides in urine by liquid chromatography-mass spectrometry. J. Anal. Tox- icol. 27:464-470 (2003).

21. S. Feng, M.A. EISohly, and D.T. Duckworth. Hydrolysis of con- jugated metabolites of buprenorphine. I. The quantitative enzy- matic hydrolysis of buprenorphine-3-13-D-glucuronide in human urine. J. Anal. Toxicol. 25:589-593 (2001).

22. H. Trauer, H.-J. Wehran, R.K. MiJller, and R. Vock. Katalysierte an- alytische Acetylierung--effizient und schnell. Toxichem. Krimtech. 65:97 (1998).

23. M. Kolb, A. Bahr, S. Hippich, and W. Schulz. Calculation of de- tection limit, identification limit and determination limit ac- cording to DIN 32645 with the aid of a computer program. Acta Hydrochim. Hydrobiol. 21: 308-311 (1993).

24. S. Iwersen-Bergmann and A. Schmoldt. Direct semiquantitative screening of drugs of abuse in serum and whole blood by means of CEDIA DAU urine immunoassays. J. Anal. Toxicol. 23: 247-256 (1999).

25. R.E. Aderjan and G. Skopp. Formation and clearance of active and inactive metabolites of opiates in humans. Ther. Drug Monit. 20" 561-569 (1998).

26. M. Balikova, V. Maresova, and V. Habrdova. Evaluation of uri- nary dihydrocodeine excretion in human by gas chromatog- raphy-mass spectrometry. J. Chromatogr. B 752:179-186 (2001).

27. J.J. Kuhlman, Jr., S. Lalani, J. Magluilo, Jr., B. Levine, W.D. Darwin, R.E. Johnson, and E.J. Cone. Human pharmacokinetics of intra- venous, sublingual, and buccal buprenorphine. J. Anal. Toxicol. 20" 369-378 (1996).

28. S. George, C. George, and M. Chauhan. The development and application of a rapid gas chromatography-mass spectrometry method to monitor buprenorphine withdrawal protocols. Forensic Sci. Int. 143:121-125 (2004).

Manuscript received November 8, 2004; revision received February 3, 2005.