Energy resolution of several scintillating crystalsusing differents readout systems
Measurements performed during Jul-Sept 2007 at the GSI
Martın Gascon
Particle Physics DepartamentUniversity of Santiago de Compostela
5th. October, 2007
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 1
![Page 2: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/2.jpg)
Contents
Tests performed
1 Energy resolution calibration using 1 cm3 CsI(Tl)
2 Energy resolution of CsI(Na) and BGO using APDs
3 Comparison between APDs and PIN Diodes
4 Energy resolution of LaBr3 using PMTs
5 Energy resolution of LaCl3 using PMTs
6 Non-Uniformities measurements
7 Best values obtained for energy resolution
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 2
![Page 3: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/3.jpg)
1. Energy resolution calibrationusing 1 cm3 CsI(Tl)
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 3
![Page 4: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/4.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDExperimental setup
The detector is electrically isolated by a grounded Faraday box
The detector and Preamplifier are placed in a metalic box with a humidity control system
The system is temperature monitorized
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 4
![Page 5: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/5.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDBias Voltage Optimization
Bias Voltage Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.300 937.3 ± 0.3 4.87 ± 0.08 937.1 ± 0.4 4.90 ± 0.08320 1254 ± 0.5 5.12 ± 0.09 1258 ± 0.5 5.16 ± 0.10340 1698 ± 0.7 5.37 ± 0.10 1705 ± 0.6 4.83 ± 0.09360 2357 ± 0.9 5.23 ± 0.09 2366 ± 0.9 4.79 ± 0.09380 3255 ± 1.3 5.14 ± 0.09 3266 ± 1.2 4.75 ± 0.08400 4517 ± 1.8 4.97 ± 0.08 4527 ± 1.6 4.57 ± 0.08420 6211 ± 2.3 4.62 ± 0.08 6222 ± 2.2 4.54 ± 0.08430 7275 ± 2.7 4.74 ± 0.09 7285 ± 2.7 4.71 ± 0.08
Table: Photopeak position vs Bias Voltage
Spectra for each Bias Voltage are taken twice,in order to check stability and to reduceuncertainties
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 5
![Page 6: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/6.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDBias Voltage Optimization
Not saturated between 300 V and 420 V
There is no minimum in the energy resolution
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 6
![Page 7: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/7.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDBias Voltage Optimization
Bias Voltage Amp. gain Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.360 135 6307 ± 2.3 4.51 ± 0.08 6304 ± 2.3 4.81 ± 0.09370 113 6330 ± 2.3 4.71 ± 0.08 6332 ± 2.3 4.59 ± 0.08380 96 6330 ± 2.2 4.61 ± 0.08 6326 ± 2.4 4.62 ± 0.08390 81 6335 ± 2.4 4.85 ± 0.09 6340 ± 2.3 4.74 ± 0.08400 69 6296 ± 2.4 4.92 ± 0.09 6302 ± 2.3 4.74 ± 0.09410 59 6333 ± 2.3 4.71 ± 0.09 6333 ± 2.3 4.83 ± 0.09
Table: Photopeak position vs Bias Voltage
Varying Bias Voltage and Amplifier gain to keepconstant the photopeak channel
There is minimum of energy resolution around380 V
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 7
![Page 8: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/8.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDAmplifier Gain
Linear along the whole dynamic range
the higher the Amplifier gain, the better the energy resolution
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 8
![Page 9: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/9.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDAcquisition Time
There is a statistical deficit below 40 seconds
60 seconds give us a good energy resolution
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 9
![Page 10: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/10.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDShaping Time
3 µs and 6 µs are a good trade-off
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 10
![Page 11: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/11.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDFinal energy resolution
Temp. oC Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.24.3 ± 0.1 6156 ± 2.3 4.96 ± 0.09 6154 ± 2.4 4.87 ± 0.0924.3 ± 0.1 6159 ± 2.4 5.08 ± 0.09 6156 ± 2.4 5.04 ± 0.0924.3 ± 0.1 6159 ± 2.3 4.75 ± 0.09 6152 ± 2.3 4.65 ± 0.0924.3 ± 0.1 6157 ± 2.5 4.97 ± 0.09 6155 ± 2.4 4.93 ± 0.0924.3 ± 0.1 6151 ± 2.3 4.81 ± 0.09 6155 ± 2.4 4.86 ± 0.09
Table: Ten spectra acquired for st = 3µs, t = 50s, HV = 380V and G = 105
Temp. oC Photopeak ch. Energy Resol. Photopeak ch. Energy Resol.24.3 ± 0.1 6220 ± 2.2 4.59 ± 0.08 6221 ± 2.3 4.68 ± 0.0824.3 ± 0.1 6219 ± 2.3 4.74 ± 0.09 6219 ± 2.2 4.68 ± 0.0824.3 ± 0.1 6219 ± 2.2 4.60 ± 0.08 6222 ± 2.2 4.59 ± 0.0824.3 ± 0.1 6220 ± 2.3 4.65 ± 0.08 6221 ± 2.2 4.63 ± 0.0824.3 ± 0.1 6216 ± 2.2 4.55 ± 0.08 6219 ± 2.3 4.73 ± 0.09
Table: Ten spectra acquired for st = 6µs, t = 50s, HV = 380V and G = 96
3 µs: Mean value: 4.89± 0.09 - Best value: 4.65± 0.09
6 µs: Mean value: 4.64± 0.09 - Best value: 4.55± 0.08
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 11
![Page 12: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/12.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDEnergy resolution vs Energy
1170 keV is part of Comptom edge of 1332 keV of 60Co
The energy resolution is function of the square root of the incident energy photon
We used a 60Co, 137Cs and 152Eu sources
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 12
![Page 13: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/13.jpg)
Energy resolution calibration using 1 cubic cm CsI(Tl) + APDEnergy resolution
The reported energy resolution obtained is 4.55 % is similar to the obtained in”Optimization of energy resolution obtained with CsI(Tl) crystals for the R3BCalorimeter” M. Gascon et al., Submited to IEEE Transactions on NuclearScience.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 13
![Page 14: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/14.jpg)
Albert Einstein
A theory issomething nobodybelieves, except theperson who made it.An experiment issomething everybodybelieves, except theperson who made it.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 14
![Page 15: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/15.jpg)
2. Energy resolution ofCsI(Na) and BGO using APDs
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 15
![Page 16: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/16.jpg)
Energy resolution of CsI(Na) and BGO using APDsExperimental setup
Frustum shape crystals (170 mm length)
All are encapsulated into aluminium shielding
9x9 and 15x15 windows (1mm thick) (Face A and B)
Delivered by Scionix Holland.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 16
![Page 17: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/17.jpg)
Energy resolution of CsI(Na) and BGO using APDsExperimental setup
The detector is electrically isolated by a grounded Faraday box
The detector and Preamplifier are placed in a metalic box with a humidity control system
The system is temperature monitorized
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 17
![Page 18: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/18.jpg)
Energy resolution of CsI(Na) and BGO using APDsCsI(Na) - Shaping Time - Face A
between 1 and 3 µs the energy resolution is approximately constant
3 µs requires the lowest amplifier gain
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 18
![Page 19: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/19.jpg)
Energy resolution of CsI(Na) and BGO using APDsCsI(Na) - Acquisition Time - Face A
80 seconds is a good trade-off
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 19
![Page 20: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/20.jpg)
Energy resolution of CsI(Na) and BGO using APDsCsI(Na) - Amplifier Gain - Face A
Linear along the whole amplifier dynamic range
500 is the situable one
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 20
![Page 21: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/21.jpg)
Energy resolution of CsI(Na) and BGO using APDsCsI(Na) - Amplifier Gain - Face A and B
Face B provides more light collection
The higher the light collection, the better the energy resolution
300 is selected for Face B
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 21
![Page 22: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/22.jpg)
Energy resolution of CsI(Na) and BGO using APDsCsI(Na) - Acquisition Time - Face A and B
The higher the light collection, the better the energy resolution
80 seconds is a good trade-off
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 22
![Page 23: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/23.jpg)
Energy resolution of CsI(Na) and BGO using APDsBGO - Shaping Time - Face B
between 1 and 2 µs the energy resolution is approximately constant
1 µs requires lower amplifier gain
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 23
![Page 24: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/24.jpg)
Energy resolution of CsI(Na) and BGO using APDsBGO - Amplifier Gain - Face B
Linear along the whole dynamic range
500 is selected, above this value noise is amplified.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 24
![Page 25: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/25.jpg)
Energy resolution of CsI(Na) and BGO using APDsBGO - Acquisition Time - Face B
60 seconds is a good trade-off
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 25
![Page 26: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/26.jpg)
African Proverb
Not to knowis bad. Not towish to know isworse.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 26
![Page 27: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/27.jpg)
3. Comparison betweenAPDs and PIN Diodes
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 27
![Page 28: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/28.jpg)
Comparison between APDs and PIN DiodesExperimental setup
Filter + Preamp made by GSI
No chance of varying Bias Voltage
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 28
![Page 29: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/29.jpg)
Comparison between APDs and PIN DiodesAmplifier Gain
Bad energy resolution
PINs Don’t need High Voltage
Faster stabilization
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 29
![Page 30: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/30.jpg)
Comparison between APDs and PIN DiodesAcquisition Time
similar behavior to Hamamatsu
70 seconds is appropriated
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 30
![Page 31: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/31.jpg)
Comparison between APDs and PIN DiodesShaping Time
1 µs and 6 µs are minima
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 31
![Page 32: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/32.jpg)
Comparison between APDs and PIN DiodesExperimental setup
Canberra 2001A Preamplifier
PIN Diode S3590-08 from Hamamatsu
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 32
![Page 33: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/33.jpg)
Comparison between APDs and PIN DiodesBias Voltage
PINs require much lower Bias Voltage
Faster stabilization (stable between 40 V and 80 V)
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 33
![Page 34: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/34.jpg)
Comparison between APDs and PIN DiodesBias Voltage
60 V is selected
Energy resolution is 2 % worse for PIN diode
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 34
![Page 35: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/35.jpg)
Comparison between APDs and PIN DiodesAmplifier Gain
Linear along the whole dynamic range
750 is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 35
![Page 36: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/36.jpg)
Comparison between APDs and PIN DiodesAcquisition Time
Similar behavior to Hamamatsu
60 seconds is adecuated
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 36
![Page 37: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/37.jpg)
Comparison between APDs and PIN DiodesShaping Time
3 µs and 6 µs are good trade off for both
Pin Diode are too noisy, so that they can not resolve energies below 400 keV
Pin Diode has worse energy resolution, but similar trend for energies above 500keV
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 37
![Page 38: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/38.jpg)
4. Energy resolution of LaBr3
using PMTs
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 38
![Page 39: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/39.jpg)
Energy resolution of LaBr3 using PMTsR329 - Experimental setup
Brillance 380 (Saint Gobain)
R329 Hamamatsu PMT
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 39
![Page 40: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/40.jpg)
Energy resolution of LaBr3 using PMTsR329 - Bias Voltage
Last point is out of the curve
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 40
![Page 41: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/41.jpg)
Energy resolution of LaBr3 using PMTsR329 - Bias Voltage
Varying Bias Voltage and Amplifier gain to keep constant the photopeak channel
1050 V is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 41
![Page 42: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/42.jpg)
Energy resolution of LaBr3 using PMTsR329 - Amplifier Gain
Not linear for gain higher than 40
36 is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 42
![Page 43: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/43.jpg)
Energy resolution of LaBr3 using PMTsR329 - Acquisition Time
120 seconds is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 43
![Page 44: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/44.jpg)
Energy resolution of LaBr3 using PMTsR329 - Shaping Time
1 µs and 2 µs are good trade off
after 2 µs saturates
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 44
![Page 45: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/45.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Experimental setup
Brillance 380 (Saint Gobain)
XP1918 Photonis
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 45
![Page 46: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/46.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Bias Voltage
After 1100 V is saturated
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 46
![Page 47: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/47.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Amplifier Gain
Not linear from 40
36 is adecuated
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 47
![Page 48: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/48.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Acquisition Time
80 seconds is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 48
![Page 49: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/49.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Shaping Time
1 or 2 µs are selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 49
![Page 50: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/50.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Energy resolution
4.11 % is the lowest energy resolution without saturation.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 50
![Page 51: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/51.jpg)
Energy resolution of LaBr3 using PMTsXP1918 - Energy resolution
The presented by Saint Gobain resolutions could be saturated
The relation between Photopeak channel/Compton is higher
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 51
![Page 52: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/52.jpg)
5. Energy resolution of LaCl3using PMTs
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 52
![Page 53: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/53.jpg)
Energy resolution of LaCl3 using PMTsExperimental setup
R329 PMT from Hamamatsu
Brillance 350 - 10 cm LaCl3 (1 and 2 windows)
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 53
![Page 54: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/54.jpg)
Energy resolution of LaCl3 using PMTsR329 - Bias Voltage
After 1200 V is saturated
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 54
![Page 55: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/55.jpg)
Energy resolution of LaCl3 using PMTsR329 - Bias Voltage
Varying the gain and Bias Voltage
1050 V is adecuated
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 55
![Page 56: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/56.jpg)
Energy resolution of LaCl3 using PMTsR329 - Amplifier Gain
linear
40 is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 56
![Page 57: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/57.jpg)
Energy resolution of LaCl3 using PMTsR329 - Acquisition Time
120 seconds is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 57
![Page 58: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/58.jpg)
Energy resolution of LaCl3 using PMTsR329 - Shaping Time
1 or 2 µs are selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 58
![Page 59: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/59.jpg)
Energy resolution of LaCl3 using PMTsR329 - Amplifier Gain
Covering the window with aluminized mylar
linear
60 is selected
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 59
![Page 60: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/60.jpg)
Energy resolution of LaCl3 using PMTsR329 -Energy resolution
5.78 % is the best energy resolution achieved.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 60
![Page 61: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/61.jpg)
6. Non-Uniformities measurements
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 61
![Page 62: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/62.jpg)
Non-Uniformities measurementsExperimental setup
To collimate the source, we use Lead blocks of 5 cm length
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 62
![Page 63: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/63.jpg)
Non-Uniformities measurementsNon-Uniformity definitions
Non Uniformity in the Ligth Output
is one of the factors worsening the energy resolution
at least, 2 methods are used to define it: G and δ
Definition of G
G =LOmax − LOmin
LOmed(1)
Definition of δ
LOLOmed
= 1 + δ(x − xmed
xmed) (2)
Notes
δ-value is independent of the number of measuring points, but G does not
infinite number of measuring points means G = 2 · δExample:
For a 10 cm lenght crystal, if we measure 9 points every 1 cm, we get:G = 1.6 · δ
(1) D.M. Beylin, et al., Nucl. Inst. and Meth., vol. A541, pp 501-515, 2005.(2) G. Ren, X et al.,Nucl. Inst. and Meth., vol. A564, pp 364-369, 2006.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 63
![Page 64: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/64.jpg)
Non-Uniformities measurementsNon-Uniformity definitions
Non Uniformity in the Ligth Output
is one of the factors worsening the energy resolution
at least, 2 methods are used to define it: G and δ
Definition of G
G =LOmax − LOmin
LOmed(1)
Definition of δ
LOLOmed
= 1 + δ(x − xmed
xmed) (2)
Notes
δ-value is independent of the number of measuring points, but G does not
infinite number of measuring points means G = 2 · δExample:
For a 10 cm lenght crystal, if we measure 9 points every 1 cm, we get:G = 1.6 · δ
(1) D.M. Beylin, et al., Nucl. Inst. and Meth., vol. A541, pp 501-515, 2005.(2) G. Ren, X et al.,Nucl. Inst. and Meth., vol. A564, pp 364-369, 2006.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 63
![Page 65: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/65.jpg)
Non-Uniformities measurements10 cm CsI(Tl) + APD + PMT
Spectra were taken using Cs-137 and Co-60 sources
Fitting the photopeak channel from the 1332 MeV gamma of a Co-60 source
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 64
![Page 66: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/66.jpg)
Non-Uniformities measurements10 cm CsI(Tl) + APD + PMT
G = 16.7 %; δ = 9.28% for 10 cm CsI(Tl) + HAMA
G = 9.6 %; δ = 5.33% for 10 cm CsI(Tl) + XP1918
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 65
![Page 67: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/67.jpg)
Non-Uniformities measurements10 cm CsI(Tl) + APD + PMT
Increasing the PMT gain and plotting as a function of distance to APD
Using a 1332 keV of Cobaltum 60 source (5 min/point)
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 66
![Page 68: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/68.jpg)
Non-Uniformities measurements10 cm CsI(Tl) + APD + PMT
Doing the sum of the photopeak channel for both
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 67
![Page 69: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/69.jpg)
Non-Uniformities measurementsCsI(Na) - Non Uniformity in light collection - Face B
Collimated source of 60Co by 5 cm of Lead
Face A produced few light to perform this test
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 68
![Page 70: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/70.jpg)
7. Best values obtainedfor energy resolution
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 69
![Page 71: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/71.jpg)
Best values obtained for energy resolutionEnergy resolution
Crystal lenght Readout Premplifiera Shaping Bias Volt. Energy Resolutionb
CsI(Tl) 1 cm APDc Canb. 2001 6 µs e-m 4.50%
CsI(Tl) 1 cm Pin GSI-Preamp 6 µs - 18.7%
CsI(Tl) 1 cm Pin Canb. 2001 6 µs e-m 6.45%
CsI(Tl) 10 cm APD Canb. 2001 6 µs e-m 5.36%
LaBr3 3 cm R329 - 2 µs Wenzel 4.54%
LaBr3 3 cm XP1918 - 2 µs Wenzel 4.11%
CsI(Na) 17 cm APD Canb. 2001 3 µs e-m 17.1%
BGO 17 cm APD Canb. 2001 1 µs e-m 23.7%
LaCl3 10 cm R329 - 1 µs Wenzel 5.70%
LaCl3 (2w) 10 cm R329 - 1 µs Wenzel 8.35%
aThe Amplifier used was Ortec 572bEnergy Resolution for the 662 keV photon of a Cs-137 sourcecAPD is S8664-1010 from Hamamatsu
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 70
![Page 72: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/72.jpg)
Conclusions
The calibration showed that the energy resolution value of 4.5 % is repeated with different electronicchains
The enery resolution of CsI(Na) and BGO are bad because its bad quality
The used window for the frustum like crystals produced an energy resolution worsening
There is difference of aproximately 4 % in the energy resolution between Face A and B in frustum likecrystals
The Filter and Preamplifier made by GSI for Pin Diodes don’t offer an energy resolution comparable toCanberra Preamplifier
Pin Diodes don’t need High Voltage, are stable between 40 and 80 V, have an energy resolution of 6.5% for 662 keV photons
Pin Diodes has worse energy resolution for low shaping times and for incident energies below 400 keVthan APDs.
LaBr3 produces so many light in a so short time that saturates common PMT working at tipical voltagevalues
Working at linear gain, the best energy resolution obtained for LaBr3 was 4.11 % for XP1918 PMT fromPhotonis
The best energy resolution for LaCl3 was 5.78 % for the R329 PMT from Hamamatsu
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 71
![Page 73: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/73.jpg)
Bibliography
Technical Proposal for the Design, Construction, Commissioning and Operation of R3B.http://www-land.gsi.de/r3b/
H. Alvarez−Pol et al. A first proposal for the geometry of the Total Absorption Calorimeterdesign at R3B. Internal Note: R3B CAL 01/05. http://www.usc.es/genp/
C.W.E. van Eijk, P. Dorenbos, E.V.D. van Loef, K. Kramer, H. Gudel ”Energy resolution ofsome new inorganic-scintillator in gamma-ray detectors” Radiation Measurements, vol. 33, pp521-525, 2001.
Hamamatsu Photonics: Photomultiplier Tubes and Related Products
D.M. Beylin, A.I. Korchagin, A.S. Kuzmin, L.M. Kurdadze, S.B. Oreshkin, S.E. Petrov, B.A.Shwartz, ”Study of the radiation hardness of CsI(Tl) scintillation crystals,”Nucl. Inst. and Meth., vol. A541, pp 501-515, 2005.
G. Ren, X. Chen, S. Lu, Z.Li, X. Xue, D. Shen, ”Non-uniformity of light output in large-sizedCsI(Tl) crystals grown by non-vacuum Bridgman method”,Nucl. Inst. and Meth., vol. A564, pp 364-369, 2006.
Martın Gascon Energy Resolution of several scintillating crystals 5th. October, 2007 72
![Page 74: Energy resolution of several scintillating crystals ...fpsalmon.usc.es/r3b/talks/gsi_JulSep07.pdf · Energy resolution calibration using 1 cubic cm CsI(Tl) + APD Energy resolution](https://reader034.vdocuments.us/reader034/viewer/2022050505/5f969b898590352eae54f84d/html5/thumbnails/74.jpg)