    E. coliHappens

    Fecal Coliform Group

    - group of total coliform bacteria

    found in intestinal tracts of warm-

    blooded animals.

    E. co li

    - a species of bacteria within the fecalcoliform group

    - dominant bacteria found in waste ofhumans and warm-blooded animals.

    - produce a positive total coliform

    response- possess an enzyme called

    (-glucuronidase) which releasesfluorogen that is detected using a365 nm UV lamp.

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    3/36

    E. colias an indicator organism

    Determines the sanitary quality of water

    Impossible to test for ALL pathogenic microorganisms,

    so test for easily detectable indicator organisms.

    Hundreds of E. colistrains Most are non pathogenic (some beneficial)

    Some pathogenic strains

    Although generally not pathogenic, their presence

    indicates apathwayfor human pathogens (ex. Viruses,

    bacteria) to enter the water source.

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    4/36

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    5/36

    Analytical Methods

    EPA approved methods for testing for E. coliunder theGround Water Ruleinclude:

    Membrane Filtration (MF)

    - EPA Method 1604

    - Standard Methods 9222 G

    Multiple Tube/ Multiple Well Fermentation (enumeration)

    - Standard Methods 9221 F

    - Standard Methods 9223- Enzyme Substrate Methods

    - Quanti-Tray and Quanti-Tray 2000 (SM 9223 B)

    Link: www.epa.gov/ogwdw/methods/pdfs/methods/methods_gwr.pdf

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    6/36

    Enzyme Substrate Methods

    Enzyme based methodologies detect both total coliforms

    and E. colisimultaneously.

    Easy, rapid, accurate, economical

    Widely accepted as standard for microbiological analysisof water.

    Enzymes for Quanti-Tray method: Colilert, Colisure,Colilert-18

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    7/36

    Enzyme Substrate Methods

    Colilert:

    Ability to detect either presence/absence or to

    enumerate organisms.

    Detects a single, viable organism per sample

    Enhancements for enzyme expression

    Buffers, salts, micro-nutrients

    Suppression of non-coliforms

    Ex. Antibiotics

    Suppresses up to 2 million heterotrophs per 100 ml

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    8/36

    Equipment Needed for

    Method 9223 B

    Autoclave

    Sterilize TSB media for 15

    minutes at 119 - 121

    Sterilize blank water

    < 500 ml = 30 mins

    > 500 ml = 45 mins

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    9/36

    Equipment

    Autoclave Supplies

    Autoclave tape

    Autoclave bags : run waste

    cycle 45 mins at proper temp.

    Autoclave biological indicator

    (monthly QC)

    Log book: record time in/out,

    temp., cycle time

    Autoclave tape

    Autoclave bags

    BT Sure Biological

    Indicator

    http://www.vwrsp.com/catjpg/y00/y0062.jpg
  • 8/11/2019 E. Coli Enumeration in Drinking Water

    10/36

    Equipment

    Refrigerator

    Storage of reagents at 0-5 CTSB media

    Bacterial cultures

    ex. Kwik Stiks

    Refrigerator 0-5C

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    11/36

    Equipment

    Oven

    Sterilize measuring glassware

    for 2 hrs. at 180 C

    pipettes

    graduated cylinders

    Supplies:

    aluminum foil

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    12/36

    Equipment

    Incubator

    Incubate Quanti-Trays forE. colianalysis at 35C +/-

    0.5 C for indicated amount of

    time

    Incubator 35 C +/- 0.5

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    13/36

    Equipment

    Testing Supplies

    Quanti-Tray sealer and rubber

    inserts

    Quanti-Tray (57 wells)

    range: 1-200 MPN/100 mls

    OR

    Quanti-Tray 2000 (97 wells)

    range: 1-2419 MPN/100 mls

    Quanti-Tray Sealer and

    rubber inserts

    Quanti-Tray / Quanti-Tray 2000

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    14/36

    Equipment

    Testing Supplies (cont.)

    Pre-sterilized clear samplebottles with dechlorination

    chemicals

    Squeeze bottles for blankwater

    - used for dilutions

    Enzymes

    - ex. Colilert

    Long wave UV lamp

    Sterile sample bottle

    Squeeze bottles

    Long wave UV lamp

    365-366 nm

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    15/36

    Equipment

    Measuring Items

    Sterile graduated cylinders

    Sterile pipettes

    Balance for weighing dry

    media

    - if preparing TSB

    Sterile graduated cylinders

    Sterile pipettes

    Balance

    http://www.vwrsp.com/catjpg/041/041487.jpg
  • 8/11/2019 E. Coli Enumeration in Drinking Water

    16/36

    Equipment

    Misc. Items

    Pipette washer

    Conductivity meter

    -monthly QC of blank water

    pH meter

    -for checking pH of TSB

    Pipette washer

    pH meter

    Conductivity

    meter

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    17/36

    Reagents and Standards

    Colilert Snap packs for sample size 100 ML

    Sample turnsYELLOWwhen totalcoliform bacteria are present andfluoresces BLUEto indicate the presenceof E. coli

    Distilled Water/RODI Water Do NOTuse phosphate buffered rinse

    water with this method

    pH Buffers For calibration of pH meter used for

    monthly QC of RODI water

    DPD reagent packets For determination of residual chlorine

    (monthly QC on RODI water)

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    18/36

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    19/36

    Sample Collection

    Samples to berepresentative of the waterbeing tested Use aseptic technique for

    collection

    Keep sample containerclosed until collection begins Avoid contact with inside of

    bottle and/or cap

    Collect directly into sterilecontainer containing de-chlorinate agent Do not rinse the bottle

    Leave air space to allow formixing

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    20/36

    Sample Collection

    If not analyzed

    immediately:Recommended but not required

    to hold samples at

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    21/36

    Procedure

    Turn on Quanti-Tray Sealer

    Warm up time approx. 10 mins.

    Vigorously shake water sample bottle 25 times within 7 seconds.

    Interval between shaking and measuring the test portion should

    not exceed 3 minutes.

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    22/36

    Procedure

    Aseptically remove lid and adjust sample volume to the

    calibrated 100 ml line on sample container.

    Test requires the use of 100 ml of sample

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    23/36

    Procedure

    Aseptically add 1 packet of Colilert reagent to the 100

    ml test bottle**If sample flashes blue: excessive chlorine and invalid for analysis

    Re-cap the bottle and shake until reagent is mostly

    dissolved.

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    24/36

    Procedure

    Use one hand to hold open the Quanti-Tray or Quanti-

    Tray/2000

    Well side is facing the palm of the hand.

    Squeeze upper part of tray so it bends toward the palm.

    Gently pull foil tab to open the tray.

    Avoid touching inside of tray or foil tab.

    Pour 100 ml sample into the tray.

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    25/36

    Procedure

    Tap small wells 2-3 times to release air bubbles.

    Label back of tray with sample ID and dilution used.

    Place tray with sample into rubber insert so that wells

    sit within the cutouts

    Place rubber insert on the input shelf of sealer. Slide rubber insert with tray

    into the sealer

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    26/36

    Procedure

    Once sealed, incubate the tray/trays for 24 hrs at

    35 +/- 0.5 C

    After 24 hrs, if fluorescence is questionable, incubate

    for an additional 4 hrs.

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    27/36

    Counting and Calculations

    Quanti-Tray (51 wells) and Quanti-Tray/2000 (97 wells)Counting Ranges:

    Quanti-Tray: max. of 200 MPN/ 100 mls sample

    Quanti-Tray 2000: max. of 2,419 MPN/ 100 mls sample

    Count both small and large yellow wells.

    Use color comparator to confirm positive result.

    Document these as total coliform positive.

    Use the UV lamp to check for fluorescence.

    If no wells fluoresce, negative for E. coli

    If wells do fluoresce, positive for E. coli

    Count small and large fluorescing wells

    Refer to table for MPN

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    28/36

    Counting and Calculations

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    29/36

    Counting and Calculations

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    30/36

    Quality Control Summary

    Daily Monthly Quarterly Annually

    Negative Control Autoclave Biological

    Indicator Check

    Autoclave Timer Working

    Thermometers

    Verification

    Positive Control Balance Check

    Class S1 weights

    Autoclave

    Thermometer

    MRT1/3 years by

    OEPA auditor

    Incubator Temps

    2/Day4Hrs apart

    Reagent Water

    pH, Cl & Cond.

    Reagent Water

    Metals

    (Cd,Cr,Cu,Pb,Ni,Zn)

    Refrigerator Temp

    1/Day

    Balance Service

    Outside Vendor

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    31/36

    Quality Control

    Monthly QC

    Autoclave Biological Indicator Checks

    Verifies autoclave is sterilizing properly

    Reagent Water Analysis ( Blank Water)

    Residual ChlorineNot Detected

    Conductivity - < 2 mhos/cm

    pH5.50 to 7.50 s.u.

    Balance Calibration Check

    (used for weighing medias)

    BT Sure Biological

    Indicator

    BT Sure

    Incubator

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    32/36

    Quarterly QC

    Autoclave Timer Calibration

    Autoclave External Thermometer

    Calibration

    Annual QC

    Thermometer Calibrations

    **Exception:

    MRT = every 3 years at EPA

    recertification for drinking

    water analysis/ and when

    new MRT received.

    Reagent Water Contamination

    Analysis

    Cd, Cr, Cu, Pb, Ni, Zn

    Balance Service Check

    Outside Contractor

    Quality Control

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    33/36

    Quality Control Summary

    Media Bottles

    TSB

    Positive & Negative

    Verification

    Each new lot

    PRIOR TO USE

    Sterility Verification

    Each new lot

    PRIOR TO USE

    Colilert

    Each new lot

    PRIOR TO USE

    3 Control Bacteria

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    34/36

    QC per each new lot pr ior to use

    Sample bottle sterility checks: each new lot

    Use TSB media (Tryptic Soy Broth) Test 1% of each box received for growth

    Quality Control

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    35/36

    Quality Control

    Test each new batch (lot number) of mediaprepared/ received

    TSB media check

    1 positive control (using E. coli) ,1 negative control (no inoculation)

  • 8/11/2019 E. Coli Enumeration in Drinking Water

    36/36

    Quality Control

    Colilert check

    ( Each new lot received)

    Inoculation with 3 control bacteria:

    One control bacteria mustbe E.coli

    total coliform (+), E. coli (+)

    One control bacteria can be

    Pseudomonas aeruginosa( or other

    non-coliform)

    total coliform (-), E. coli (-)

    One control bacteria can be Klebsiella

    pneumoniae( or other coliform)

    total coliform (+), E.coli (-)