3D Echo BasicsConcept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of
cardiac structures. Real time 3D echo provides anatomically realistic visualization
of structures. Decrease the time required for complete image acquisition. 3D Echo can be viewed from various projections by rotation of
images. A limitation- the information acquired as 3D dataset, must be
displayed as 2D image.
![Page 3: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/3.jpg)
![Page 4: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/4.jpg)
![Page 5: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/5.jpg)
Linear scanning approach
• Earliest approach to dynamic 3D echo.• Based on 3D concepts used in CT and MR imaging.• Acquiring parallel and equidistantly placed 2D images
using transducer mounted on a sliding carriage.• Movement of transducer by a computer controlled
stepper motor.
Fan- like scanning approach • Transducer moved in a fan like arc at prescribed
angles.( manually or by stepper motor)
![Page 6: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/6.jpg)
Hardware used for fanlike acquisition
![Page 7: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/7.jpg)
Rotational scanning approach
• Transducer is rotated in a semicircle around the central axis of imaging plane.
![Page 8: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/8.jpg)
Disadvantages of Initial models– Time consuming– High Cost– Large Machines– Slow processor
RT3DE in 2002– Dense Matrix Array Transducer – 3000 elements– Fast processor
![Page 9: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/9.jpg)
Real Time 3D – Pyramidal Volume
![Page 10: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/10.jpg)
3D volumetric probe- Pyramidal Volume
![Page 11: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/11.jpg)
Technical Issues in Real Time 3D
Full volume images are typically obtained at frame rate of 20-40 Hz.( vary as a
function of depth & size of the volume.)
Either sacrifice frame rate for image quality (spatial resolution) or spatial resolution
for frame rate
Min Frame rate required 20 Hz – 20 frames per sec
Each frame contains beams
1 beam requires .2ms for scanning
If 100 beams / frame – 20 ms time – frame rate – 50Hz
250 beams / frame – 50 ms time – 20 Hz.
![Page 12: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/12.jpg)
Innovations to overcome the limitations
1. Increase the distance between beams.
Blurring of structures that are not adequately sampled.
2. Parallel receive beam processing – Increase frame rate
• 4 or more simultaneous receive beams for each transmitted beam
3. Limit the total number of beams by scanning a smaller volume - 3D
zoom mode
4. Use the patient’s ECG signal to acquire smaller subvolumes for each
R-R interval, and then stitch these subvolumes to produce the
larger volume -full volume dataset
5. Broaden the transmit beam
![Page 13: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/13.jpg)
Transducers
Phased array transducer
Sparse Matrix Array Transducer
Microbeam Former
Dense Matrix Array Transducer
![Page 14: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/14.jpg)
Sparse Matrix Array Transducer:
• Only a small percentage of 2500 (50х50) elements are electrically connected or acoustically active.
• Usually strategically placed 256 elements will be active.
• Loss of signal to noise ratio.• THI not supported- incapable of creating sufficient MI
necessary to create tissue harmonic image.
![Page 15: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/15.jpg)
Micro-beam forming
• Electrically group, small arrays of elements (patches).
• Typical patch contain approx 25 elements(5х5).
• 128 micro-beamformer with 128 wires connecting the transducer to mainframe beamformer- provide a fully sampled array of 3200 elements.
![Page 16: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/16.jpg)
![Page 17: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/17.jpg)
Matrix Array TEE Transducer distal tip
![Page 18: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/18.jpg)
Modes
![Page 19: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/19.jpg)
Real time Mode
• Narrow angle 600:300
• Higher Resolution(high line density
& narrow scan volume)
• No stitching Artifact
• Visualisation of valves,small masses,
vegetations.
• Real time Interventional guidance
![Page 20: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/20.jpg)
Zoom Mode
• Focused – 300: 300
• Enface view of MV
• Masses-
Thrombi,vegetations
• Eliminate need for cropping.
• Low spatial resolution due
to low line density.
![Page 21: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/21.jpg)
Full volume mode
• Covers wider region 900:900 upto
1040 .
• 4 or 7 live 3D subvolumes stitched
together.
• Near real time
• Higher temporal resolution
• Spatial relationship of cardiac
structures
• Chamber Quantification
![Page 22: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/22.jpg)
Colour Flow Mode
• Limited angle 60:60
• Limited Temporal resolution
• Shape & extention of jets
![Page 23: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/23.jpg)
![Page 24: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/24.jpg)
Basic 3D analysis
• 3D orientation
– manual free rotation of the dataset to provide the best perspective
of the structures of the heart
• Cropping
– manually moving a cutting plane from outside the 3D volume
towards its center- provide a view from the cutting plane
– cropping can be performed either before (during) or after data
acquisition
• Slicing
– extraction of image planes from the 3D volume in different modes
![Page 25: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/25.jpg)
Full Volume
![Page 26: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/26.jpg)
![Page 27: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/27.jpg)
Artifacts
![Page 28: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/28.jpg)
Stitch artifact
• Lines of disagreement between two neighboring subvolumes.
![Page 29: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/29.jpg)
Dropout artifacts
•Losses of 3D surface information due to poor echo signal intensity.•Structures- too thin to reflect.•Appropriate gain settings, colour doppler can identify false defects.
![Page 30: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/30.jpg)
Blurring and blooming artifacts
• Blurring refers to unsharp or hazy representation of thin structures- appear thicker.
• Eg: Mitral leaflets, mitral valve apparatus.• Blooming: thickened or excessive representation of
high echo density structures like mechanical prosthesis, pacemaker leads.
• Strongly related to the line density.
![Page 31: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/31.jpg)
Blurring and blooming artifacts
![Page 32: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/32.jpg)
Gain artifacts
![Page 33: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/33.jpg)
Protocols
![Page 34: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/34.jpg)
Protocols
![Page 35: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/35.jpg)
Applications
![Page 36: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/36.jpg)
Left ventricular function
• Unlike 2D echo, there is no geometrical assumptions.
• Left ventricular volume can be calculated by,
• 1. Method of discs
• 2. Directly sum the volumetric picture elements enclosed by
endocardial borders of 3D structure.
![Page 37: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/37.jpg)
Global LV function - EDV measurement
![Page 38: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/38.jpg)
Global LV function - ESV measurement
![Page 39: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/39.jpg)
LV Mass
![Page 40: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/40.jpg)
• Currently, 3D TTE and TEE assessment of LV volumes and ejection fraction is recommended over the use of 2D echocardiography, as it has been clearly demonstrated to provide more accurate and reproducible measurements.
European Heart Journal – Cardiovascular Imaging (2012) 13, 1–46 .(EAE/ASE RECOMMENDATIONS)
![Page 41: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/41.jpg)
Parametric imaging
• Assessing segmental contraction of LV.
• 800 endocardial data points - to develop a polar map of
endocardial surface of LV.
• Endocardial motion is displayed as shades of
– Blue - positive excursion values - inward motion
– Red - negative excursion values - outward motion
– Black - no motion
• Akinetic or dyskinetic myocardium - black or red color
• Normal or hypokinetic segment - shades of light to dark blue
![Page 42: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/42.jpg)
Parametric imaging
Normal Apical MI
![Page 43: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/43.jpg)
LA Volume
![Page 44: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/44.jpg)
RV volume & function
• RV has a complex geometrical shape.• RV inflow, outflow and apex do not align in a single 2D plane.• More heavily trabeculated. Normal reference values
Indexed RV EDV 49±10 ml/m²
Indexed RV ESV 16±6 ml/m²
RV EF 67±8%
![Page 45: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/45.jpg)
![Page 46: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/46.jpg)
Valvular Disease
![Page 47: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/47.jpg)
MVP - Mitral Regurgitation
• Accurate anatomy – Scallop
• Surgical View
• Vena contracta area – direct measurement
• TTE is good enough
![Page 48: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/48.jpg)
Important parameters derived by RT3DE before MV repair
1. Anterior-posterior diameter (DAP)
2. Anterolateral-posteromedial diameter (DAlPm)
3. 3D curvilinear length of posterior and anterior leaflet
4. Exposed area of leaflets (A3DE)
5. Minimal area of leaflets within the saddle-shaped annulus (A3Dmin)
6. Volume of leaflet prolapse (VProl)
7. Maximal prolapse height (HProl)
8. Length of antero-lateral and postero-medial chordae tendinae
![Page 49: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/49.jpg)
![Page 50: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/50.jpg)
![Page 51: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/51.jpg)
![Page 52: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/52.jpg)
MVP Multiple jets
![Page 53: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/53.jpg)
![Page 54: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/54.jpg)
Mitral Stenosis
• MVA at smallest orifice
• Better assessment for BMV
• Planimetry from Apical window
• More accurate than Gorlin
• New GOLD STANDARD
![Page 55: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/55.jpg)
Localizing smallest orifice
![Page 56: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/56.jpg)
![Page 57: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/57.jpg)
![Page 58: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/58.jpg)
![Page 59: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/59.jpg)
Aortic Regurgitation
TEE is better
Vena contracta area
Better delineation of Etiology
![Page 60: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/60.jpg)
Aortic Regurgitation
![Page 61: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/61.jpg)
![Page 62: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/62.jpg)
Aortic Stenosis
• AVA measurement
• Assess the noncircular LVOT area
![Page 63: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/63.jpg)
BCAV
DEGENERATIVE
RHEUMATIC
NORMAL
![Page 64: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/64.jpg)
![Page 65: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/65.jpg)
Other Applications
![Page 66: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/66.jpg)
Dysynchrony
![Page 67: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/67.jpg)
VSD
![Page 68: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/68.jpg)
ASD
![Page 69: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/69.jpg)
Thrombus
![Page 70: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/70.jpg)
Myxoma
![Page 71: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/71.jpg)
Intervention
![Page 72: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/72.jpg)
![Page 73: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/73.jpg)
![Page 74: Dr Prasanth S. 3D Echo Basics Concept - Early 1980s. Conventional 2D echo requires cognitive 3D reconstruction of cardiac structures. Real time 3D](https://reader037.vdocuments.us/reader037/viewer/2022102900/551c31cc5503469e4f8b6393/html5/thumbnails/74.jpg)
References
• Feigenbaum’s Echocardiography
• Clinical Echocardiography;Otto
• 3D Echo Thomas Buck, Andreas Franke,Mark J. Monaghan
• ASE 2012 Guidelines
• ESC – 3D Echobox
• Cardiology clinics 2007 vol 25