DYNAMIC CALIBRATION SYSTEMShas been assessed and accredited in accordance with the standard
ISO/IEC 17025:2017''General Requirements for the Competence of Testing &
Calibration Laboratories''for its facilities at
PLOT NO.63, FIRST CROSS STREET, AUDCO NAGAR, KATTUPAKKAM, CHENNAI., CHENNAI,THIRUVALLUR, TAMIL NADU, INDIA
in the field of
CALIBRATIONCertificate Number: CC-2390
Issue Date: 31/07/2019 Valid Until: 30/07/2021
This certificate remains valid for the Scope of Accreditation as specified in the annexure subject to continuedsatisfactory compliance to the above standard & the relevant requirements of NABL.
(To see the scope of accreditation of this laboratory, you may also visit NABL website www.nabl-india.org)
Signed for and on behalf of NABL
N. VenkateswaranChief Executive Officer
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
Permanent Facility
1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC Current (45Hz to 1KHz)
1 A to 10 A 0.18% to 0.48% Using Standard 6 ½DMM by Direct Method
2 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC Current (45Hz to 1KHz)
100 µA to 100 mA 0.247% to 0.181% Using Standard 6 ½DMM by Direct Method
3 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC Current (45Hz to 1KHz)
100 mA to 1 A 0.18% Using Standard 6 ½DMM by Direct Method
4 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC High Voltage(50Hz)
1 kV to 30 kV 5.7% Using HV Probe &DMM by Direct Method
5 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC Voltage (45Hz to 1KHz)
1 V to 100 V 0.44% to 0.204% Using Standard 6 ½DMM by Direct Method
6 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC Voltage (45Hz to 1KHz)
100 mV to 1 V 0.452% to 0.440% Using Standard 6 ½DMM by Direct Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 1 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
7 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC Voltage (45Hz to 1KHz)
100 V to 1000 V 0.204% to 0.17% Using Standard 6 ½DMM by Direct Method
8 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Current(50Hz to 1KHz)
1 A to 20 A 0.41% to 0.39% Using StandardMultifunction Calibratorby direct method
9 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Current(50Hz to 1KHz)
3.3 mA to 1 A 0.282% to 0.41% Using StandardMultifunction Calibratorby direct method
10 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Current(50Hz to 1KHz)
33 µA to 330 µA 0.89% to 0.054% Using StandardMultifunction Calibratorby direct method
11 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Current(50Hz to 1KHz)
330 µA to 3.3 mA 0.054% to 0.282% Using StandardMultifunction Calibratorby direct method
12 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Voltage (50Hz to1KHz)
30 V to 300 V 0.045% to 0.055% Using StandardMultifunction Calibratorby direct method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 2 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
13 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Voltage (50Hz to1KHz)
300 V to 1000 V 0.055% to 0.11% Using StandardMultifunction Calibratorby direct method
14 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Voltage(50Hz to1KHz)
1 V to 30 V 0.06% to 0.045% Using StandardMultifunction Calibratorby direct method
15 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Voltage(50Hz to1KHz)
10 mV to 30 mV 0.17% to 0.155% Using StandardMultifunction Calibratorby direct method
16 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Voltage(50Hz to1KHz)
100 mV to 1 V 0.20% to 0.06% Using StandardMultifunction Calibratorby direct method
17 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
AC Voltage(50Hz to1KHz)
30 mV to 100 mV 0.155% to 0.20% Using StandardMultifunction Calibratorby direct method
18 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
Capacitance (1kHz) 1 µF to 30 µF 0.35% to 0.74% Using StandardMultifunction Calibratorby direct method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 3 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
19 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
Capacitance(1KHz) 1 nF to 10 nF 1.744% to 0.41% Using StandardMultifunction Calibratorby direct method
20 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
Capacitance(1KHz) 10 nF to 300 nF 0.41% to 0.948% Using StandardMultifunction Calibratorby direct method
21 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
Capacitance(1KHz) 300 nF to 1uF 0.948% to 0.35% Using StandardMultifunction Calibratorby direct method:
22 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
Capacitance(1KHz) 350 pF to 1 nF 3.876% to 1.744% Using StandardMultifunction Calibratorby direct method
23 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Source)
Inductance(1KHz) 100 µH to 1 H 1.2% Using Zeal DecadeInductance Box byDirect Method
24 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Capacitance 1 nF to 0.01 F 5.18% to 4.86% Using Standard 6 ½DMM by Direct Method
25 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Current 1 A to 10 A 0.085% to 0.21% Using Standard 6 ½DMM by Direct Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 4 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
26 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Current 1 mA to 100 mA 0.066% to 0.064% Using Standard 6 ½DMM by Direct Method
27 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Current 100 µA to 1 mA 0.096% to 0.066% Using Standard 6 ½DMM by Direct Method
28 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Current 100 mA to 1 A 0.064% to 0.085% Using Standard 6 ½DMM by Direct Method
29 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 1 kOhm to 100 kOhm 0.027% to 0.023% Using Standard 6 ½DMM by Direct Method
30 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 1 MOhm to 10 MOhm 0.017% to 0.047% Using Standard 6 ½DMM by Direct Method
31 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 10 MOhm to 100MOhm
0.047% to 0.95% Using Standard 6 ½DMM by Direct Method
32 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 10 Ohm to 100 Ohm 0.046% to 0.016% Using Standard 6 ½DMM by Direct Method
33 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 100 kOhm to 1 MOhm 0.023% to 0.017% Using Standard 6 ½DMM by Direct Method
34 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 100 MOhm to 300MOhm
Using Standard 6 ½DMM by Direct Method
35 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Resistance 100 Ohm to 1 kOhm 0.016% to 0.027% Using Standard 6 ½DMM by Direct Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 5 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
36 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Voltage 1 V to 10 V 0.004% to 0.0039% Using Standard 6 ½DMM by Direct Method
37 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Voltage 10 V to 100 V 0.0039% to 0.0063% Using Standard 6 ½DMM by Direct Method
38 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Voltage 100 mV to 1 V 0.01% to 0.004% Using Standard 6 ½DMM by Direct Method
39 ELECTRO-TECHNICAL- DIRECTCURRENT (Measure)
DC Voltage 100 V to 1000 V 0.0063% to 0.0064% Using Standard 6 ½DMM by Direct Method
40 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Current 1 A to 20 A 0.034% to 0.14% Using StandardMultifunction Calibratorby direct method
41 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Current 1 mA to 100 mA 0.02% to 0.019% Using StandardMultifunction Calibratorby direct method
42 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Current 100 mA to 1 A 0.019% to 0.034% Using StandardMultifunction Calibratorby direct method
43 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Current 190 µA to 1 mA 0.03% to 0.02% Using StandardMultifunction Calibratorby direct method
44 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Resistance 1 MOhm to 10 MOhm 0.014% to 0.069% Using StandardMultifunction Calibratorby direct method
45 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Resistance 1 kOhm to 1 MOhm 0.234% to 0.014% Using StandardMultifunction Calibratorby direct method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 6 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
46 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Resistance 10 MOhm to 300MOhm
0.069% to 0.42% Using StandardMultifunction Calibratorby direct method
47 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Resistance 100 Ohm to 1 Kilo Ohm 0.005% to 0.234% Using StandardMultifunction Calibratorby direct method
48 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Resistance 2 Ohm to 10 Ohm 0.062% to 0.016% Using StandardMultifunction Calibratorby direct method
49 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Voltage 1 V to 10 V 0.0019% to 0.0014% Using StandardMultifunction Calibratorby direct method
50 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Voltage 10 mV to 100 mV 0.024% to 0.0035% Using StandardMultifunction Calibrator
51 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Voltage 10 V to 1000 V 0.0014% to 0.0025% Using StandardMultifunction Calibratorby direct method
52 ELECTRO-TECHNICAL- DIRECTCURRENT (Source)
DC Voltage 100 mV to 1 V 0.0035% to 0.0019% Using StandardMultifunction Calibratorby direct method
53 ELECTRO-TECHNICAL- DIRECTCURRENT(Source,Measure)
DC Resistance 10 Ohm to 100 Ohm 0.016% to 0.005% Using StandardMultifunction Calibratorby direct method
54 ELECTRO-TECHNICAL-ELECTRICALEQUIPMENT(Measure)
Timer 60 s to 9000 s 0.81s to 65s Using Digital Timer byComparison Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 7 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
55 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
B-Type 600 °C to 1800 °C 5.32°C Using PortableCalibrator by DirectMethod
56 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
E-Type (-) 50 °C to 800 °C 2.42°C Using PortableCalibrator by DirectMethod
57 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
J-Type (-) 200 °C to 1100 °C 3.3°C Using PortableCalibrator by DirectMethod
58 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
K-Type (-) 200 °C to 1350 °C 6.32°C Using PortableCalibrator by DirectMethod
59 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
N-Type (-)200 °C to 1200 °C 4.03°C Using PortableCalibrator by DirectMethod
60 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
PT-100 (-) 200 °C to 800 °C 1.40°C Using PortableCalibrator by DirectMethod
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 8 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
61 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
R-Type 250 °C to 1700 °C 6.47°C Using PortableCalibrator by DirectMethod
62 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
S-Type 250 °C to 1700 °C 5.68°C Using PortableCalibrator by DirectMethod
63 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
T-Type (-) 100 °C to 400 °C 0.69°C Using PortableCalibrator by DirectMethod
64 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
E-type -200 °C to 1000 °C 0.86°C Using StandardMultifunction Calibratorby direct method
65 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
J-type -200 °C to 1200 °C 0.86°C Using StandardMultifunction Calibratorby direct method
66 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
K-type -200 °C to 1350 °C 1.10°C Using StandardMultifunction Calibratorby direct method
67 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
N-type -200 °C to 1300 °C 1.20°C Using StandardMultifunction Calibratorby direct method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 9 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
68 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
PT-100 -200 °C to 800 °C 0.97°C Using Standard Multi-function Calibrator bydirect method
69 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
R-type 250 °C to 1750 °C 1.75°C Using Standard Multi-function Calibrator bydirect method
70 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
S-type 300 °C to 1750 °C 1.75°C Using StandardMultifunction Calibratorby direct method
71 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
T-type -250 °C to 400 °C 0.93°C Using Standard Multi-function Calibrator bydirect method
72 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)
Frequency 300 kHz to 1 MHz 0.04% to 0.10% Using Standard 6 ½DMM by Direct Method
73 ELECTRO-TECHNICAL- TIME &FREQUENCY(Measure)
Frequency 5 Hz to 300 kHz 1.16% to 0.04% Using Standard 6 ½DMM by Direct Method
74 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)
Frequency 120 Hz to 500 MHz 0.0006% to 0.05% Using StandardMultifunction Calibratorby direct method
75 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)
Oscilloscope(AmplitudeDC Signal)
1 mV to 130 V 4.84% to 0.06% Using StandardMultifunction Calibratorby direct method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 10 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
76 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)
Oscilloscope(AmplitudeSquare Wave)
3 mVp-p to 130 Vp-p 2.6% to 0.12% Using Standard Multifunction Calibrator bydirect method
77 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)
Oscilloscope(Bandwidth)
50 kHz to 500 MHz 8.09% Using Standard Multi-function Calibrator bydirect method
78 ELECTRO-TECHNICAL- TIME &FREQUENCY (Source)
Oscilloscope(TimeMarker)
2 ns to 5 s 0.58% Using Standard Multi-function Calibrator bydirect method
79 MECHANICAL-ACCELERATION ANDSPEED
Speed - (Non ContactType)Tachometers &RPM indicators
100 rpm to 95000 rpm 0.98% of Rdg Using DigitalTachometer & RPMSource By ComparisonMethod as per SANASTR 45-02.
80 MECHANICAL-ACOUSTICS
SoundSound LevelMeter
94 & 114 dB 1.30dB Using Sound LevelCalibrator as per IS15575 / OIML-R-58 &BL-WI-M15.
81 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Angle Plate / BoxAngle(Squareness,Flatness, Parallelism)
400 x 300 x 500 mm 6.3µm Using Co-ordinatingMeasuring Machine
82 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Bench Centre (Co-Axiality & Parallelism)
Up to 3000 mm 5.0µm Using Mandrel & DialGauge
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 11 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
83 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
BevelProtractor(LC:0°5')
0 ° to 360 ° 17minutes per arc Using Profile Projector
84 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Bore Gauge 0 to 1.2 mm 2.4µm Using Dial CalibrationTester & ElectronicComparator
85 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
CalipersDial/Vernier/Digital/GearTooth(LC:0.01/0.02mm)
0 to 2000 mm 15.6µm Using Caliper Checker& Length Bar with endJaws
86 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Coating Thicknessgauge (LC:0.1µm)
0 to 10 mm 4.0µm Using Electroniccomparotor and Foils
87 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Coating thicknessspecimen/Foil
upto 10 mm 0.91 µm Using Electroniccomparotor
88 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
CombinationSet(Protractor,centrehead,square head)
0 to 180° 30minutes per arc Using Profile Projector
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 12 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
89 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
ComparatorStand(Flatness,Parallelism,Perpendicular)
0 to 500 mm 9.8µm Using CMM
90 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Cylindrical SettingMasters
2 to 200 mm 1.6µm Using LengthMeasuring Machineand cylinder master
91 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
DepthGaugeVernier/Digital(LC:0.01 /0.02 mm)
0 to 1000 mm 10.2µm Using Slip gauge &Length Bars
92 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
DepthMicrometer(LC:0.01mm)
0 to 1000 mm 5.8µm Using Slip Gauge /Length bar
93 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Dial SnapGauges(LC:0.01 mm)
0 to 200 mm 5.98µm Using slip gauges longslip gauges
94 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Dial Thickness Gauge /Pistol Caliper(LC:0.01mm)
0 to 200 mm 6.8µm Using Slip Gauges &Long Slip Gauges
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 13 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
95 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Engineers Try Square& GraniteSquare(Squareness,Flatness,Parallelism)
0 to 600 mm 5.0µm Using Co-ordinatingMeasuring Machine
96 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
External MicrometerL.C. 0.001 mm
> 300 to 1500 mm 7.7µm Using Slip Gauge,Long Slip Gauge &Length Bars
97 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
External MicrometerL.C. 0.001 mm
0 to 300 mm 1.6µm Using 0 Grade Gaugeblocks & Length Bars
98 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Feeler Gauge 0.025 to 2 mm 0.6µm Using LengthMeasuring Machine
99 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Flush Pin / DepthGauge
0.2 to 200 mm 3.2µm Using Slip Gauge &Electronic Comparator
100 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Gap/Width Gauge 0.2 to 200 mm 3.9 µm Using Slip Gauge &Electronic Comparator
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 14 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
101 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Groove DialGauge(Internal /External)
0 to 200 mm 11.9µm Using Slip Gauges &Long Slip Gauges
102 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Height Gauge(Dial/Vernier/Digital)L.C.: 0.01 mm
0 to 1500 mm 13.8µm Using Caliper Checker,Slip Gauge & LengthBars
103 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
InclinometerDigital /Analog / Dial
0 to 90° & 0 to 180° 3.0sec Using Precision SpritLevel, Slip gauge &Sine Bar
104 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Internal Micrometer /StickMicrometer(LC:0.01mm)
5 to 1100 mm 14.26µm Using Slip Gauge /Length Bars &Electronic Comparator
105 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Lever Type DialGaugeL.C.: 1 µm
0 to 2.5 mm 2.0µm Using Dial CalibrationTester
106 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Measuring Pins/Thread MeasuringWires
0.17 to 20 mm 0.8µm Using Slip Gauge &Electronic Comparator
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 15 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
107 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Micrometer Setting rod 25 to 1500 mm 7.7µm Using Length Bars &Electronic Comparator
108 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Plain Plug Gauge 1 to 400 mm 3.6µm Using Slip Gauge &Electronic Comparator
109 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Plain Ring Gauge > 100 to 400 mm 4.0µm Using LengthMeasuring Machineand master ring
110 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Plain Ring Gauge 2 to 100 mm 1.50µm Using LengthMeasuring Machine
111 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Plunger Dial Gauge/Digital DialGaugesL.C.: 0.1µm
0 to 100 mm 3.5µm Using Dial CalibrationTester & Slip Gaugeblocks
112 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Radius Gauge up to 50 mm 6.60µm Using Profile Projector
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 16 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
113 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Sine Bar/Sine Centre Up to to 600 mm 9.9 um and 10 Sec Using Co-ordinatingMeasuring Machine
114 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Slip GaugeAccessories
Up to 300 mm 3.9µm Using Electronic HeightGauge
115 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Snap gauge 2 to 500 mm 3.6µm Using Slip Gauge &Long Slip Gauge
116 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Spline Plug Gauge 5 to 200 mm 3.2µm Using LengthMeasuring Machine
117 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Spline Ring Gauge 10 to 200 mm 4.8µm Using LengthMeasuring Machine
118 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Sprit level/Block level upto 600 mm 6.9 µm using precision spritlevel, slip gauge & sinebar
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 17 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
119 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Steel Rule/Plastic RuleMeasuring Tape / PiTape
0 to 100 mtrs 88vLµm Using Digital Tape &Scale MeasuringDevice
120 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
StraightEdge(Straightness,Parallelism)
0 to 1000 mm 10.8µm Using Co-ordinatingMeasuring Machine
121 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Surface Plate Up to 3000 x 3000 mm 1.7+Vl+w/150µm Using Precision SpritLevel, Slip Gauge &Sine Bar
122 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Taper Plain PlugGauge(Diameter,Taper angle)
Up to to 500 mm 4.2 µm,21" Using LengthMeasuring Machine
123 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Taper Plain RingGauge(diameter, Taperangle)
Up to to 300 mm 4.8µm,21" Using LengthMeasuring Machine
124 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Taper Scale 0 to 200 mm 25.8µm Using Profile Projector
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 18 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
125 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Taper Thread PlugGauge(Diameter,Taper angle)
5 to 200 mm 1.60µm,21" Using LengthMeasuring Machine
126 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Taper Thread RingGauge(minor &effective diameter &taper angle)
5 to 200 mm 2.10 µm, 21" Using LengthMeasuring Machine
127 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Test MandrelPlain/Taper(Cylindricity, Concentricity &Straightness)
0 to 600 mm 4.0µm Using Dial Gauges
128 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Test Sieves 0.02 to 80 mm 6.6µm Using Profile Projector
129 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Thread Pitch Gauge /Angle Gauge
0.2 to 7.0 mm 36 Sec Using Profile Projector
130 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Thread PitchGauge(Angle Gauge)
0.2 to 7.0 mm 6.6µm Using profile Projector
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 19 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
131 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Thread Plug Gauge(effective and majordiameter)
> 100 to 300 mm 2.90µm Using LengthMeasuring Machine
132 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Thread PlugGauge(effective &major diameter)
2 to 100 mm 2.4µm Using LengthMeasuring Machine
133 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Thread RingGauge(effective andmajor diameter)
3 to 300 mm 2.6µm Using LengthMeasuring Machine
134 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Three Point Micrometer 5 to 200 mm 3.9µm Using Setting RingGauges
135 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Ultrasonic ThicknessGauge L.C. 0.01 mm
0 to 200 mm 11.3µm Using Slip GaugeBlocks
136 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Vee Block Symmetricity Up to to 200 mm 6.0µm Using Mandrel &Electronic HeightGauge
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 20 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
137 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Vee Block(Parallelism) Up to to 200 mm 5.3µm Using Mandrel &Electronic HeightGauge
138 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
VeeBlock(Perpendicularity)
Up to to 200 mm 5.4µm Using Mandrel &Electronic HeightGauge
139 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Wield Fillet / AutoDepth Gauge
0 to 200 mm 6.6µm Using Profile Projector
140 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Caliper Checker 0 mm to 600 mm 7.16µm Using Electronic HeightGauge
141 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Dial Calibration Tester(LC:0.0001 mm)
0 mm to 50 mm 1.96µm Using Gauge Block &Electronic Comparator.Procedure as per IS9483.
142 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Electronic HeightGaugeL.C: 0.1µm
0 mm to 1000 mm 7.0µm Using Long Slip GaugeBlocks & Length Bars.Procedure as per IS2921.
143 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Electronic Probe WithDRO / LVDTL.C.:0.0001mm
0 mm to 100 mm 0.70µm Using K Grade SlipGauges
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 21 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
144 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Floating Carriage DiaMeasuring MachineLC:0.1µm(Alignment ofcenter to base fcdm,Micrometer head,overall accuracy fcdm)
0 mm to 100 mm 4.16µm,0.77µm,1.61µm
Using Gauge Blocks,Mandrel & ElectronicHeight Gauge
145 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Length Bar 125 mm to 500 mm 4.66µm Using LengthMeasuring Machineand procedure basedon ISO 3650 & IS7014.
146 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Long Slip Gauges 125 mm to 500 mm 4.66µm Using LengthMeasuring Machineand procedure basedon ISO 3650 & IS7014.:
147 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
MicrometerHead(LC:0.0001 mm)
0 mm to 50 mm 1.96µm Using Gauge Block &Electronic Comparator.Procedure as per IS9483.:
148 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Profile Projector /VMM(Angular)
0 ° to 360 ° 3.2 "sec of arc Using Angular GlassScale
149 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Profile Projector /VMM(Linear)
0 mm to 300 mm 1.5µm Using Linear GlassScale
150 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Profile Projector /VMM(Magnification)
10 x to 100 x 0.7% Using Glass Scales
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 22 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
151 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Slip Gauge 0 mm to 25 mm 0.15µm Using Gauge BlockCalibrator & K GradeSlip Gauges.Procedure based on IS2984.
152 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Slip Gauge 25 mm to 50 mm 0.18µm Using Gauge BlockCalibrator & K GradeSlip Gauges.Procedure as per IS2984.
153 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Slip Gauge 50 mm to 75 mm 0.21µm Using Gauge BlockCalibrator & K GradeSlip Gauges.Procedure as per IS2984.
154 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Slip Gauge 75 mm to 100 mm 0.25µm Using Gauge BlockCalibrator & K GradeSlip Gauges.Procedure based on IS2984.
155 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Specimens 0.05 µm to 12.5 µm 6.0% Using SurfaceRoughness Tester:
156 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Surface RoughnessTester
0.05 µm to 12.5 µm 6.0% Using SurfaceRoughness specimen
157 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Verification of Co-ordinating MeasuringMachine
0 mm to 2000 mm 5.18µm Using Long SlipGauges & Length bars.Procedure as per ISO10360.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 23 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
158 MECHANICAL-DUROMETER
Durometers Shore A 0 Shore A to 100 ShoreA
1.0Shore A Using Shore HardnessCalibrator As perASTM D2240-05
159 MECHANICAL-DUROMETER
DurometersShore D 0 Shore D to 100 ShoreD
1.0Shore D Using Shore HardnessCalibrator As perASTM D2240-05
160 MECHANICAL-FORCE PROVINGINSTRUMENTS
Load cells (with/without indicator),Proving Rings,Dynamometer(Compression andTension)Class 0.5 andcoarser
0.5 N to 100 N 0.04% Using Dead WeightForce CalibrationMachine with StainlessSteel / Dead Weightsand Loading hangers.As per IS 4169:2014/ISO 376:2011
161 MECHANICAL-FORCE PROVINGINSTRUMENTS
Load cells (with/without indicator),Proving Rings,Dynamometer(Compression andTension)Class 00 andcoarser
100 N to 5000 N 0.04% Using Dead WeightForce CalibrationMachine with StainlessSteel / Dead Weightsand Loading hangers.As per IS 4169:2014/ISO 376:2011
162 MECHANICAL-MOBILE FORCEMEASURING SYSTEM
Push Pull Gauge /Force Gauge
0 N to 2000 N 0.04% of Rdg Using Slotted StandardWeights As Per ASTME4-16
163 MECHANICAL-PRESSUREINDICATING DEVICES
Pressure(Gauge,Switches, Sensors,Transducers)
0 to 1000 bar 0.46% of Rdg Using Digital PressureIndicators & PressureComparator pump ByComparison Method Asper DKD-R-6-1
164 MECHANICAL-PRESSUREINDICATING DEVICES
Vaccum (Gauge,Switches, Sensors,Transducers)
-1 bar to 0 bar 1.2% of rdg Using Digital Vaccum &Pressure ComparatorPump DKD-R-6-1
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 24 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
165 MECHANICAL-TORQUEGENERATINGDEVICES
Torque Multiplier 0 Nm to 5000 Nm 0.90% of Rdg Using Digital TorqueWrench Sensor WithIndicator ByComparison Method
166 MECHANICAL-TORQUEGENERATINGDEVICES
Torque Wrench /Torque ScrewDrivers(Type-I Class-A,B,C,D,E Type-IIClass-A,B,C,D,E,F,G)
0 Nm to 3000 Nm 1.04% of rdg Using Digital TorqueWrench Sensor WithIndicator As per ISO6789:2003
167 MECHANICAL-TORQUEMEASURINGDEVICES
Torque Calibrator,Torque TransducerWith / WithoutIndicator, TorqueMeter, Torque Tester(b)Class 0.2 & Coarser
5 Nm to 2000 Nm 0.04% Using Dead WeightTorque CalibrationSystem consisting ofLever Arm andStainless steel /Aluminum Deadweights. As per BS:7882.: 2008
168 MECHANICAL-TORQUEMEASURINGDEVICES
Torque Calibrator,Torque TransducerWith / WithoutIndicator, TorqueMeter, Torque Tester(a)Class 0.5 & Coarser
0.05 Nm to 5 Nm 0.03% Using Dead WeightTorque CalibrationSystem consisting ofPulley and AluminumDead weights. As perBS: 7882:2008
169 THERMAL- SPECIFICHEAT & HUMIDITY
Humidity Sensors,Thermo Hygrometers,Data Logger/ Recorder
20 % RH to 95 % RH 1.9% RH Using DigitalHygrometer, HumidityGenerator byComparison Method
170 THERMAL-TEMPERATURE
Glass thermometers 30 °c to 200 °c 1.8°c Using 3 wire RTD withprecision digitalThermometer LiquidBath (Oil), byComparison Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 25 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
171 THERMAL-TEMPERATURE
IR Thermometer (Non-Contact Type)RadiationThermometer.
50 °c to 1200 °c 3.99°c Using Standard IRThermometer & Blackbody Source byComparison Method
172 THERMAL-TEMPERATURE
RTDs, Thermocouples(with and withoutindicators), DataLoggers, Recorders,Transmitters, DigitalThermometers
-196 °C to -196 °C 1.26°C Using 4 wire RTD withtemperature calibrator,Liquid nitrogen Bath, byComparison Method
173 THERMAL-TEMPERATURE
RTDs, Thermocouples(with and withoutindicators), DataLoggers, Recorders,Transmitters, DigitalThermometers
-80 °C to 250 °C 2.4°C Using 4 wire RTD withhigh precision digitalThermometer lowtemperature bath byComparison Method
174 THERMAL-TEMPERATURE
Thermocouples (withand without indicators),Data Loggers,Recorders,Transmitters, DigitalThermometers
250 °C to 1200 °C 3.65°C Using Standard S-typeThermocouple WithPortable Calibrator &Dry Block ByComparison Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 26 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
Site Facility
1 ELECTRO-TECHNICAL-ALTERNATINGCURRENT (< 1 GHZ)(Measure)
AC High Voltage(50Hz)
1 kV to 30 kV 5.7% Using HV Probe &DMM by Direct Method
2 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
B-Type 600 °C to 1800 °C 5.32°C Using PortableCalibrator by DirectMethod
3 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
E-Type (-) 50 °C to 800 °C 2.42°C Using PortableCalibrator by DirectMethod
4 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
J-Type (-) 200 °C to 1100 °C 3.3°C Using PortableCalibrator by DirectMethod
5 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
K-Type (-) 200 °C to 1350 °C 6.32°C Using PortableCalibrator by DirectMethod
6 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
N-Type (-)200 °C to 1200 °C 4.03°C Using PortableCalibrator by DirectMethod
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 27 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
7 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
PT-100 (-) 200 °C to 800 °C 1.40°C Using PortableCalibrator by DirectMethod
8 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
R-Type 250 °C to 1700 °C 6.47°C Using PortableCalibrator by DirectMethod
9 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
S-Type 250 °C to 1700 °C 5.68°C Using PortableCalibrator by DirectMethod
10 ELECTRO-TECHNICAL-TEMPERATURESIMULATION(Measure)
T-Type (-) 100 °C to 400 °C 0.69°C Using PortableCalibrator by DirectMethod
11 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
E-type -200 °C to 1000 °C 0.86°C Using StandardMultifunction Calibratorby direct method
12 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
J-type -200 °C to 1200 °C 0.86°C Using StandardMultifunction Calibratorby direct method
13 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
K-type -200 °C to 1350 °C 1.10°C Using StandardMultifunction Calibratorby direct method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 28 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
14 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
N-type -200 °C to 1300 °C 1.20°C Using StandardMultifunction Calibratorby direct method
15 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
PT-100 -200 °C to 800 °C 0.97°C Using Standard Multi-function Calibrator bydirect method
16 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
R-type 250 °C to 1750 °C 1.75°C Using Standard Multi-function Calibrator bydirect method
17 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
S-type 300 °C to 1750 °C 1.75°C Using StandardMultifunction Calibratorby direct method
18 ELECTRO-TECHNICAL-TEMPERATURESIMULATION (Source)
T-type -250 °C to 400 °C 0.93°C Using Standard Multi-function Calibrator bydirect method
19 MECHANICAL-ACCELERATION ANDSPEED
Speed - (Noncontact)Centrifuge,RPM Source &Indicators
100 rpm to 20000 rpm 0.60% of rdg Using DigitalTachometer ByComparison Method asper SANAS TR 45-02
20 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Bench Centre (Co-Axiality & Parallelism)
Up to 3000 mm 5.0µm Using Mandrel & DialGauge
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 29 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
21 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Sprit level/Block level upto 600 mm 6.9 µm using precision spritlevel, slip gauge & sinebar
22 MECHANICAL-DIMENSION (BASICMEASURINGINSTRUMENT,GAUGE ETC.)
Surface Plate Up to 3000 x 3000 mm 1.7+Vl+w/150µm Using Precision SpritLevel, Slip Gauge &Sine Bar
23 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Electronic HeightGaugeL.C: 0.1µm
0 mm to 1000 mm 7.0µm Using Long Slip GaugeBlocks & Length Bars.Procedure as per IS2921.
24 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Profile Projector /VMM(Angular)
0 ° to 360 ° 3.2 "sec of arc Using Angular GlassScale
25 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Profile Projector /VMM(Linear)
0 mm to 300 mm 1.5µm Using Linear GlassScale
26 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Profile Projector /VMM(Magnification)
10 x to 100 x 0.7% Using Glass Scales
27 MECHANICAL-DIMENSION(PRECISIONINSTRUMENTS)
Verification of Co-ordinating MeasuringMachine
0 mm to 2000 mm 5.18µm Using Long SlipGauges & Length bars.Procedure as per ISO10360.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 30 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
28 MECHANICAL-HARDNESS TESTINGMACHINES
Verification of BrinellHardness TestingMachine HBW 10/3000
10/3000 HBW to10/3000 HBW
2.0% HBW Using HBW HardnessBlocks As per IS6506.2.2012
29 MECHANICAL-HARDNESS TESTINGMACHINES
Verification of BrinellHardness TestingMachine HBW 5/750
5/750 HBW to 5/750HBW
1.4% HBW Using HBW HardnessBlocks As per IS6506.2.2012
30 MECHANICAL-HARDNESS TESTINGMACHINES
Verification of RockwellHardness TestingMachine
0 HRC to 100 HRC 1.04HRC Using HRC HardnessBlocks As per IS1586.2.2012 By DirectMethod
31 MECHANICAL-HARDNESS TESTINGMACHINES
Verification of RockwellHardness TestingMachine
0 HRWB to 100 HRWB 1.3HRWB Using HRWB HardnessBlocks As per IS1586.2.2012 By DirectMethod
32 MECHANICAL-HARDNESS TESTINGMACHINES
Verification of VickersHardness TestingMachine HV 30
30 HV to 30 HV 2.2% HV Using HV HardnessBlocks As per IS6507.2.2012
33 MECHANICAL-PRESSUREINDICATING DEVICES
Pressure(Gauge,Switches, Sensors,Transducers)
0 to 1000 bar 0.46% of Rdg Using Digital PressureIndicators & PressureComparator pump ByComparison Method Asper DKD-R-6-1
34 MECHANICAL-PRESSUREINDICATING DEVICES
Vaccum (Gauge,Switches, Sensors,Transducers)
-1 bar to 0 bar 1.2% of rdg Using Digital Vaccum &Pressure ComparatorPump DKD-R-6-1
35 MECHANICAL- UTM,TENSION CREEPAND TORSIONTESTING MACHINE
Universal TestingMachine(CompressionMode)
1 kN to 1000 kN 0.41% of Rdg Using Load Cells WithIndicators as per IS1828 (Part I) :2015
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 31 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019
S.No Discipline / Group Quantity Measured/Instrument
Range / Frequency * CalibrationMeasurementCapability(±)
Remarks
36 THERMAL- SPECIFICHEAT & HUMIDITY
HumidityEnvironmentalChambers, HumidityChambers, HumidityGenerators
30 %RH to 93 %RH 1.9%RH Using DigitalHygrometer, HumidityData Logger byComparison Method
37 THERMAL-TEMPERATURE
Indicator ofEnvironmentalChamber, DeepFreezers, Furnace,Oven, Vacuum Oven,BOD Incubator,Autoclave, CentrifugeChamber
200 °C to 1200 °C 4.8°C Using Standard N-TypeThermocouple withTemperature Calibrator& Indicator byComparison Method(Single point)
38 THERMAL-TEMPERATURE
Indicator ofEnvironmentalChamber, DeepFreezers, Furnace,Oven, Vacuum Oven,BOD Incubator,Autoclave, CentrifugeChamber
-80 °C to 200 °C 2.5°C Using Standard RTD(Pt 100) withTemperature Calibrator& Indicator byComparison Method(Single Point)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
Laboratory Name DYNAMIC CALIBRATION SYSTEMS, PLOT NO.63, FIRST CROSS STREET, AUDCONAGAR, KATTUPAKKAM, CHENNAI., CHENNAI, THIRUVALLUR, TAMIL NADU ,INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2390 Page No. : 32 / 32
Validity 31/07/2019 to 30/07/2021 Last Amended on 14/08/2019