“End station A setup” data analysis
Josef Uher
![Page 2: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/2.jpg)
Outline
Introduction to setup and analysis
Quartz bar start counter
MA and MCP PMT in the prototype
![Page 3: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/3.jpg)
What was analyzed
quartz bar start counter4 pad MCP - PMT
ADC
ConstantFractionDiscriminator
prototype3x MCP PMT (Burle)
2x MA PMT (Hamamatsu)
light source -pulse laser “Pilas”
LeadingEdgeDiscriminator
TDC
TDC
![Page 4: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/4.jpg)
Tasks for data analysis software
develop a timing strategies for start counters
develop a calibration procedure for the prototype (determine time offsets for each of all modules, determine single electron timing resolutions)
help to tune the setup and find errors
![Page 5: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/5.jpg)
What can affect timing resolution?
fluctuation of avalanche recoil of photoelectrons from MCP
surface charge sharing cross talk electronics noise smear of pulse source (laser) ...
![Page 6: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/6.jpg)
150
170
190
210
230
250
270
290
310
330
350
10 100 1000Frequency [Hz]
Sigm
a [p
s]
The best resolutionMean resolution
Average improvement of the timing resolution
~22%
Timing resolution vs. Pilas trigger frequency
(measured with MA PMT Hamamatsu)
![Page 7: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/7.jpg)
Quartz bar start counter
![Page 8: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/8.jpg)
Quartz bar start counter
quartz bar start counter4 pad MCP - PMT
ADC
LeadingEdgeDiscriminator
TDC
Configuration with ADC gives possibility to correct measured time on pulse height off-line.
=> better control than in case of CFD
![Page 9: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/9.jpg)
ADC correction – how it works?
t1 t2 Time
Am
plitu
de
TDC threshold
TD
C s
pectr
aA
DC
vs.
TD
C
Before correction After correction
1=250 ps 1=100 ps
![Page 10: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/10.jpg)
Time resolution vs. number of electronsQuartz bar start counter
10
100
1000
0.00 50.00 100.00 150.00 200.00 250.00 300.00
ADC signal [photoelectrons]
Nar
row
er S
igm
a [
ps]
Narrower sigma - new data (100x amplification)
Narrower sigma - old data (10x amplification)
Quartz bar start counter timing resolution
Improvement for single photonsfrom ~350 ps to ~138 ps.
![Page 11: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/11.jpg)
Prototype
![Page 12: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/12.jpg)
Burle MCP-PMT, raw TDC spectra
![Page 13: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/13.jpg)
Hamamatsu MA-PMT, raw TDC spectra
![Page 14: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/14.jpg)
MCP and MA PMT timing resolution
SLOT 3 Best resolution Average
resolution
[ps] 194.25 250.5
MA PMT Hamamatsu
SLOT 4 Best resolution Average
resolution
[ps] 156.5 258.75
MCP PMT Burle
(raw data only, single photon mode, no corrections)
±58
±43
![Page 15: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/15.jpg)
Pad timing offset mapsM
A P
MT H
am
am
ats
uWithout TDC link offset
MC
P P
MT B
url
eTDC 1 TDC 2 TDC 3 TDC 4Start
Peak p
osit
ion
Peak p
osit
ion
Peak p
osit
ion
Peak p
osit
ion
Delay ~1ns
![Page 16: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/16.jpg)
Cross talk
![Page 17: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/17.jpg)
Cross talk treatment
Possible sources of “false pulses”: charge sharing recoiled electrons internal PMT cross talk (wiring) cross talk in amplifiers cross talk in CFD photons bouncing in prototype box
![Page 18: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/18.jpg)
Cross talk can appear in any pad of MCP or MA PMT
Cut in certain time after the first pulse
Cross talk treatmentM
CP
or
MA
PM
T
![Page 19: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/19.jpg)
Cross talk treatment
Time [cnt]
Even
ts
[-]
Example of cross talk treatment with time window 15 counts.
MC
P P
MT B
url
e
![Page 20: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/20.jpg)
Cross talk treatment
Time [cnt]
Even
ts
[-]
Example of cross talk treatment with time window 15 counts.
MC
P P
MT B
url
e
![Page 21: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/21.jpg)
Cross talk treatment
The time window changed from 15 to 27 counts
Time [cnt]
Even
ts
[-]
Example of cross talk treatment with time window 27 counts.
MC
P P
MT B
url
e
![Page 22: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/22.jpg)
Time window 15 cnt
Time window 27 cnt
Cross talk only
Cross talk treatment
Time [cnt]
Even
ts
[-]
Cross talk spectra with time window 15 and 27 counts.
MC
P P
MT B
url
e
![Page 23: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/23.jpg)
Time window 15 cnt
Time window 27 cnt
Cross talk treatment
Time [cnt]
Even
ts
[-]
Corrected TDC spectra with time window 15 and 27 counts.
MC
P P
MT B
url
e
![Page 24: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/24.jpg)
Timing resolution as a function of cross talk time window
180
190
200
210
220
230
240
250
260
270
0 10 20 30 40
Cross talk window [cnt]
Sig
ma
[ps]
TDC peak height as a function of cross talk time window
2000
2500
3000
3500
4000
4500
5000
5500
6000
6500
7000
0 10 20 30 40
Cross talk window [cnt]
Pea
k h
eig
ht
[-]
Cross talk treatment
Raw TDC spectrum Time window 1 Time window 31
Time [cnt]
Even
ts
[-]
narrow = 190 ps
![Page 25: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/25.jpg)
MCP PMT Burle after cross talk correction
SLOT 4 Best resolution Average
resolution
[ps] 135.12 225.00 improvement 14% 13%
Multiple hits Cross talk “frequency”
Raw data max.: ~7Corrected data: ~5
![Page 26: “End station A setup” data analysis Josef Uher. Outline Introduction to setup and analysis Quartz bar start counter MA and MCP PMT in the prototype](https://reader035.vdocuments.us/reader035/viewer/2022062423/56649f0d5503460f94c21e14/html5/thumbnails/26.jpg)
Conclusion
Influence of Pilas laser on timing resolution discovered. Higher Pilas trigger frequency needed – another Pilas laser will be tested.
Repeated measurement of quartz bar start counter timing resolution – still needs more work
All slots in the prototype will be tested in near future
Effective cross talk treatment was implemented in data analysis code