DKIST Optical Alignment Plan
Procedure Number: PROC-0018
Authors: P. Sekulic, C. Liang, K. Gonzales, R. Hubbard, S. Craig, SEIC
Date: May 2017
Approved by:
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE II
Revision Control
Version Date Revised by Items changed
A 8 May 2017 Original release
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE III
Table of Contents
1 Introduction .............................................................................................. 1
1.1 Plan Overview ..........................................................................................................1
1.2 Exclusions ................................................................................................................1
1.3 Safety Issues.............................................................................................................1
2 Supporting Information ............................................................................ 3
2.1 Reference Documents ..............................................................................................3
2.2 Acronyms and Abbreviations ..................................................................................4
3 Alignment Plan Definition ........................................................................ 6
3.1 General Approach ....................................................................................................6 3.1.1 Prime and Gregorian Foci (M1 / M2) ........................................................... 6
3.1.2 OSS Transfer Optics ...................................................................................... 6
3.1.3 Coudé and FIDO............................................................................................ 7 3.1.4 Telescope Level .............................................................................................. 7
3.2 Alignment Simulation ..............................................................................................7 3.2.1 Sensitivity Analysis......................................................................................... 7 3.2.2 Tolerancing Analysis ..................................................................................... 8
3.2.3 Alignment Simulation ..................................................................................... 8
3.3 Alignment Preparation ...........................................................................................10
4 On-Sky Measurements ........................................................................... 11
4.1 Nighttime Star Observing ......................................................................................11
4.2 Daytime Star Observing .........................................................................................11
4.3 Lunar Observing ....................................................................................................11
4.4 Pointing Model and Beam Steering .......................................................................11
5 Optical Alignment Equipment ................................................................ 13
5.1 Required Equipment ..............................................................................................13
5.2 Metrology Requirements .......................................................................................14 5.2.1 Coordinate Measurement............................................................................. 14 5.2.2 Tilt Measurement ......................................................................................... 15 5.2.3 Wavefront Measurement .............................................................................. 15 5.2.4 Software ....................................................................................................... 16
6 M1 Alignment ......................................................................................... 18
6.1 Method Overview and Flowchart ..........................................................................18
6.2 Objectives and Preconditions .................................................................................19
6.3 Wavefront Measurement ........................................................................................20 6.3.1 Measure the Wavefront at the Initial EA ..................................................... 20 6.3.2 Measure the Wavefront at Other EAs .......................................................... 21 6.3.3 Analyze the Wavefront ................................................................................. 21
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE IV
6.4 Modal Correction ...................................................................................................21
7 M2 Alignment ......................................................................................... 23
7.1 Method Overview and Flowchart ..........................................................................23
7.2 Objectives and Preconditions .................................................................................24 7.2.2 Perform the Shack-Hartmann Alignment..................................................... 25 7.2.3 Perform the M2 Alignment ........................................................................... 25 7.2.4 Optimize the Wavefront................................................................................ 26
8 OSS Transfer Optics Alignment ............................................................. 27
8.1 Method Overview and Flow Chart ........................................................................27
8.2 Definition of the Telescope’s Altitude Axis ..........................................................29
8.3 M5 and M6 Alignment...........................................................................................30
8.3.1 M6 Alignment ............................................................................................... 31 8.3.2 Tower Stability Test ..................................................................................... 31
8.3.3 M5 Alignment ............................................................................................... 32
8.4 M3 Alignment ........................................................................................................32
8.5 Alt-Az Alignment ..................................................................................................33
8.5.1 Defining the Azimuth Axis ............................................................................ 33 8.5.2 Alt-Az Alignment .......................................................................................... 34
8.6 M7 Alignment and Coudé Characterization ..........................................................35 8.6.1 M7 Alignment ............................................................................................... 35 8.6.2 Coudé Rotation Characterization ................................................................ 36
8.6.3 Coudé Optics Alignment Preparation .......................................................... 36
8.7 M4 Alignment ........................................................................................................37
9 Coudé Transfer Optics Alignment .......................................................... 38
9.1 Method Overview and Flowchart ..........................................................................38
9.2 Alignment of M8, M9, and M10 Mirrors ..............................................................39
9.2.1 M8 ................................................................................................................ 39 9.2.2 M9 ................................................................................................................ 40 9.2.3 CMM Arm Setup .......................................................................................... 40 9.2.4 M10 .............................................................................................................. 41
10 Telescope Residual Error Alignment ..................................................... 42
10.1 Method ...................................................................................................................42
10.2 Cumulative Focus Error .........................................................................................42
10.3 Final Measurement and Correction ........................................................................42
11 FIDO Alignment ...................................................................................... 44
11.1 Method Overview and Flowchart ..........................................................................44
11.2 Alignment Procedures ............................................................................................46
11.2.1 WFC-BS1 ................................................................................................... 46 11.2.2 M9a ............................................................................................................ 47
11.2.3 CL2a ........................................................................................................... 48
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE V
11.2.4 CL2 ............................................................................................................. 49 11.2.5 CL3a ........................................................................................................... 50 11.2.6 CL3 and CL4 .............................................................................................. 51
11.2.7 DL-FM1 ..................................................................................................... 52
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE VI
List of Figures
Figure 1: Schematic representation of the alignment tool used for simulation and during
telescope alignment ............................................................................................................. 9
Figure 2: HASO4-BB typical dynamic range of individual Zernike ............................................ 16
Figure 1: M1 mirror optical alignment flowchart ......................................................................... 19
Figure 2 : M2 mirror optical alignment flowchart ........................................................................ 24
Figure 3: OSS transfer optics alignment flowchart....................................................................... 28
Figure 4: Example of an AC target ............................................................................................... 29
Figure 5: Position of the theodolite, AC1 and AC2 mirrors ......................................................... 29
Figure 6: Theodolite turned 60º upward to align M6.................................................................... 31
Figure 7: Position of the theodolite to align M5 ........................................................................... 32
Figure 8: AC3 autocollimation mirror and theodolite in the Coudé room ................................... 34
Figure 9 : Position of theodolite for M7 alignment ...................................................................... 36
Figure 10: M9 alignment configuration ........................................................................................ 37
Figure 11 : Coudé transfer optics alignment flowchart ................................................................ 39
Figure 12: M9 alignment configuration ........................................................................................ 40
Figure 13 : M10 alignment configuration ..................................................................................... 41
Figure 14 : Coudé beamsplitters and mirrors alignment flowchart .............................................. 46
Figure 15 : WFC-BS1 beamsplitter alignment configuration ....................................................... 47
Figure 16 : M9a alignment configuration ..................................................................................... 48
Figure 17 : CL2a beamsplitter alignment configuration ............................................................... 49
Figure 18 : CL2 beamsplitter alignment configuration................................................................. 50
Figure 19 : CL3a beamsplitter alignment configuration ............................................................... 51
Figure 20 : CL3 and CL4 beamsplitter alignment configuration .................................................. 52
Figure 21 : DL-FM1 alignment configuration .............................................................................. 53
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 1 OF 53
1 INTRODUCTION
1.1 Plan Overview
This document presents a plan for aligning the following optical components of the
DKIST telescope:
The primary and secondary mirrors, M1 and M2
The transfer optics of the optical support structure (OSS), which include M3, M4,
M5, and M6
The Coudé transfer optics, which include M7, M8, M9, and M10
The facility instrument distribution optics (FIDO), which include beamsplitters
WFC-BS1, CL2, CL3, CL4, CL2a, and CL3a and mirrors M9a (which feeds the
Cryo-NIRSP instrument) and DL-FM1 (which feeds the DL-NIRSP instrument)
NOTE: The procedures included in this plan are intended to provide a general outline of steps to be completed for each phase of alignment. A list of detailed procedures is presented in Section 2.1.
1.2 Exclusions
Separate alignment plans and procedures will be developed for all DKIST
instruments, for the wavefront correction system (WFC), and for the polarimetry
analysis and calibration (PA&C) Gregorian optical station (GOS). The PA&C GOS
procedure will be developed after M1 and M2 have been aligned.
1.3 Safety Issues
Project safety guidelines are described in SPEC-0030, “Conditions for Working at
the Site / Safety & Health Specification for Contractors.”
Many activities performed during alignment will require working on elevated
structures, a moving telescope, and a moving enclosure. These activities pose fall,
falling object, and pinch hazards. Only adequately trained personnel shall operate the
telescope and enclosure. Those moving the telescope and enclosure must follow
proper safety procedures to maintain a balanced structure; they must also coordinate
with other personnel working within the enclosure. All safety precautions must be
followed to mitigate associated risks.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 2 OF 53
Likewise, many activities associated with equipment installation and calibration will
require working with heavy equipment and lifting devices such as cranes, personnel
lifts, and forklifts, which pose pinch hazards. Only adequately trained personnel shall
operate lifting equipment, and all personnel must apply proper rigging and
standardized visual communication practices. All safety precautions must be followed
to mitigate associated risks.
Important Guidelines for Handling Optics
At some point in the installation and alignment procedures, each optic will be
unprotected while work is undertaken to position it. During this time, the optical
surface is at risk for damage by inadvertent contact with lifting equipment, tools,
hands, clothing, and even airborne contaminants. Care must be taken to mitigate all of
these risks by following all steps and using all required protective equipment and
clothing when working near or above unprotected optics.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 3 OF 53
2 SUPPORTING INFORMATION
2.1 Reference Documents
File Name Title
Safety Documents
SPEC-0030 Conditions for Working at the Site / Safety & Health Specification for Contractors
Integration, Test, and Commissioning Procedures
PROC-0082 M1 Mirror Installation
PROC-0098 PF Wavefront Measurements
PROC-0097 TEOA/PF SHWS LT Align
PROC-0103 M2 and HS on-TMA LT Alignment
PROC-0117 GF Wavefront Measurements
PROC-0243 Alignment of optical targets
PROC-0244 Characterization of TMA axis, Coudé rotation and tower stability
PROC-0136 M3 Assy Installation
PROC-0245 M3 alignment
PROC-0141 M4 Assy Installation PROC
PROC-0246 M4 alignment
PROC-0134 M5 surrogate installation
PROC-0247 M5 surrogate alignment
PROC-0248 M5 alignment
PROC-0129 M6 Assy Install
PROC-0249 M6 alignment
PROC-0250 M6 tilt adjustment (beam pointing)
PROC-0135 M7 Assy Installation
PROC-0251 M7 alignment
PROC-0157 M8 Assy Installation
PROC-0252 M8 alignment
PROC-0158 M9 Assy Installation
PROC-0253 M9 alignment
PROC-0155 M10 Surrogate Installation
PROC-0254 M10 surrogate alignment
PROC-0255 M10 alignment
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 4 OF 53
File Name Title
PROC-0156 SHWS Installed and Aligned
PROC-0256 Coudé Wavefront Measurement
PROC-0164 M9a Assy Install/Align
PROC-0163 WFC-BS1 Assy Install/Align
PROC-0165 CL2a Assy Install/Align
PROC-0166 CL2 Assy Install/Align
PROC-0167 CL3a Assy Install/Align
PROC-0168 CL3 & CL4 Assy Install/Align
PROC-0169 DL-FM1 Assy Install/Align
PROC-0170 FIDO Beam Pointing verifications
Technical Notes
TN-0240 IT&C Focus Compensation Analysis
TN-0248 DKIST alignment simulation
2.2 Acronyms and Abbreviations
AOS: Adaptive optics system
BS: Beam splitter
CAD: Computer aided design
CMM: Coordinate measuring machine
FIDO: Facility instrument distribution optics
FOV: Field of view
GIS: Global interlock system
GOS: Gregorian optical station
IT&C: Integration, Testing, and Commissioning
LUT: Lookup table
NIRSP: Near-infrared spectro-polarimeter
OSS: Optical support structure
PA&C: Polarimetry analysis and calibration
SA: SpatialAnalyzer®
SHWS: Shack-Hartmann Wavefront Sensor
SMR: Spherically mounted retroreflector
TEOA: Top end optical assembly
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 5 OF 53
TMA: Telescope mount assembly
VBI: Visible-light broadband imager
WFC: Wavefront correction
WFCS: Wavefront correction system
WFE: Wavefront error
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 6 OF 53
3 ALIGNMENT PLAN DEFINITION
3.1 General Approach
Each optical component will be accurately positioned using a laser tracker, a
coordinate measuring machine (CMM), and/or a theodolite, followed by a Shack-
Hartmann wavefront sensor (SHWS) to achieve fine optical alignment of the
wavefront. This process will use three focal locations that precede the Coudé
laboratory: the prime focus (from the M1 mirror) and the Gregorian focus (between
M2 and M3).
The telescope coordinate system is defined by the rotation axis (altitude and azimuth)
of the telescope mount assembly (TMA). The mirrors will be initially positioned by
measuring their fiducials using a laser tracker with SpatialAnalyzer® (SA) software,
which integrates optical and mechanical computer aided design (CAD) models. The
TMA structure deflection at different elevation angles (EA) will be characterized by
measuring the position of mirrors with the laser tracker. These data will be analyzed
and processed to populate the look-up tables (LUT).
3.1.1 Prime and Gregorian Foci (M1 / M2)
At the prime focus, the SHWS will be used to measure the wavefront produced by
target stars. The objective of this effort is to characterize the focal plane by
identifying the best focus position and the center of the field of view (FOV). The
SHWS will also be used to characterize M1’s surface figure as function of EA. In
addition, the SHWS will be used to verify M1’s modal correction.
The M2 mirror will be aligned by placing the SHWS at the Gregorian focal plane.
The objectives are to align M2 to the M1 mirror to minimize WFE at Gregorian focal
plane and to measure the position the Gregorian focus for different elevation angles.
The WFE data and Gregorian focus position data will be used to refine M1 and M2
LUTs.
3.1.2 OSS Transfer Optics
The OSS transfer optics comprise three flat mirrors—M3, M5, and M6—and one off-
axis mirror, M4. The flat mirrors will be aligned using the laser tracker and a
theodolite to improve tilt accuracy. The off-axis mirror, M4, will be aligned using
only the laser tracker; however, the residual misalignment will be measured and
compensated for in the Coudé wavefront measurements.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 7 OF 53
3.1.3 Coudé and FIDO
Similarly, the Coudé optics and FIDO will be aligned using a laser tracker, theodolite,
and CMM arm. The main challenge for this phase of alignment will be to connect the
telescope and Coudé coordinate systems with minimal errors. To achieve this goal, a
number of targets will be attached permanently to non-rotating walls on the Coudé;
the laser tracker measurements of these targets can then be transferred automatically
into the telescope coordinate system.
3.1.4 Telescope Level
The wavefront for the entire telescope will be aligned at the Coudé level using the
beam from the M10 mirror. Night-sky measurements will be used to determine the
telescope performances, and the M2/M4 mirrors will be used to compensate for
residual astigmatism, coma, and focus to achieve fine optimization. The tilt of the
M3/M6 mirrors will also be optimized to adjust the boresight. For each wavefront
measurement—at the primary, Gregorian, and telescope foci—dedicated alignment
software will be used to compute the mirror shift and tilt needed to compensate for
the measured residual aberrations and to position the focal plane at its nominal
location.
3.2 Alignment Simulation
Three analyses will be conducted to inform the alignment strategy:
An analysis of the alignment sensitivities for each optical element
The results of Monte-Carlo tolerancing
The results of a simulation to verify the alignment procedure
This approach is designed to account for the opto-mechanical design, the error
budget, and the assembly procedure.
3.2.1 Sensitivity Analysis
The sensitivity analysis will be conducted using Zemax© OpticStudio™ (Zemax).
This software will be used to develop a sensitivity matrix for all degrees of freedom
for each optical element and for different field points. Sensitivities will be computed
for several parameters:
Fringe Zernike coefficients (focus Z4, astigmatism Z5 and Z6, coma Z7 and Z8,
and spherical Z9) in nine predefined field points
Focus position
Boresight error
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 8 OF 53
The analysis will reveal the most and least sensitive elements, providing a basis for
designing a preliminary alignment strategy and identifying compensators. The
alignment sensitivities of each mirror will be computed for the prime, Gregorian, and
telescope focal planes.
By default, Zemax uses the vertex position as the center of the coordinate systems.
When tilting an off-axis mirror, it creates, at first order, a large focus term, with
additional Dx or Dy decenters at second order. To eliminate this effect, macros will be
used to transform the coordinate system for each off-axis mirror (M1, M2, M4, and
M8) before computing sensitivities. The goal of this effort is to use the local vertex
position (rather than the global vertex position) as the center of rotation of each
optics. For flat mirrors (M3, M5, M6, M7, M9, and M10), Zemax centers the
coordinate system on the optical surface by default.
TN-0240, “IT&C Focus Compensation Analysis,” gives the results of this analysis.
3.2.2 Tolerancing Analysis
Next, a Monte-Carlo tolerancing analysis will be performed. The input for this
analysis is the tolerance of initial mechanical positioning of each mirror. This value is
determined from the mirror’s error budget during the use of the laser tracker for the
initial positioning. The results of this analysis will establish DKIST’s expected initial
performance prior to wavefront alignment.
Zemax will be used to perform the tolerancing at different alignment steps. Each step
will use the results of the previous alignment step as input. Tolerancing for each will
be performed for 1,000 trials that can be analyzed to determine WFE (root mean
square (RMS) and peak-to-valley), Zernike Z4 to Z9, the focus position, and the
boresight error.
Once the tolerancing is complete for each focus (prime, Gregorian, and telescope),
the best-, mean-, and worst-case results will be used as input to simulate wavefront
alignment and identify the corresponding system. This system will then be used as
input for the next tolerancing step.
The results of this analysis will be detailed in TN-0248, “DKIST Alignment
Simulation.”
3.2.3 Alignment Simulation
The goal of this final step is to identify the best achievable optical performances. It
entails simulating the optical alignment using a tool that computes the required
motions (decenters and tilts) to compensate for the measured residual WFE, focus
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 9 OF 53
position, and boresight error. The simulation will provide the data needed to define
the alignment specifications and to confirm the efficacy of the alignment strategy and
compensators.
The alignment tool will be used during both the mirror alignment simulation and the
on-site IT&C optical alignment. Developed specifically for DKIST, this tool
automatically identifies the required mirror compensations (Dx, Dy, Dz, Tx, Ty, and Tz)
and telescope pointing corrections to minimize the measured WFE (Zernike
decomposition), focus position, and boresight error. As inputs, it uses the sensitivities
computed for mirror alignment and telescope pointing, along with Zernike, focus
position, and boresight error values that are either measured or generated by the
Monte-Carlo simulation. Three matrices must be created to use this tool:
S, the sensitivity matrix, which contains the computed sensitivities (from Zemax)
of the parameters that will be measured (Zernike in nine field points, focus
position, and boresight error)
M, the measurement matrix, which contains the measured WFE (Zernike
decomposition) at the center of the field and at different field points, as well as the
focus position and boresight error
C, the compensation matrix, which contains the mirror motions and pointing
corrections
Figure 1: Schematic
representation of the alignment
tool used for simulation and
during telescope alignment
The objective is to find the
compensation matrix (C) using
the measured data (M) and
sensitivities (S). When these
motions are applied to the
mirrors, new measurements are
performed to generate a new matrix M that converges toward zero: S×C = –M and
then C = –S-1×M. Since matrices S and M are known, matrix C can be computed with
the pseudo-inverse function.
Individual tools have been created for each phase of alignment—at the prime focus
(M1), Gregorian focus (M1/M2), and telescope focal plane (M1 to M10).
The results of these individual alignments are detailed in TN-0248, “DKIST
Alignment Simulation.”
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 10 OF 53
3.3 Alignment Preparation
Preparation is critical to performing the alignment procedures effectively and
efficiently. The powerful SA software will be used for the following tasks:
Merging the optical Zemax design and mechanical CAD model.
Defining and creating the as-built alignment fiducials on the optical mounts to
establish their locations with respect to the optical surface.
Creating the coordinate systems that will be measured with the laser tracker or
CMM arm.
Extracting the values that will be used as success criteria during alignment.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 11 OF 53
4 ON-SKY MEASUREMENTS
The on-sky procedures described in this section refer to nonsolar observing—the
equipment referenced here has not been designed for solar illumination.
Solar alignment and thermal testing are covered in other procedures.
4.1 Nighttime Star Observing
Almost all alignment activities will be performed at night on bright stars. This offers
several advantages. First, abundant data is available to provide full-sky coverage for
calibrating the LUTs for both WFE and beam steering. Second, a fully functional
thermal system is not required for nighttime measurements; consequently, they can
begin earlier in the project schedule. Another important consideration is the location
of the pole star. Not only is Polaris visible but it is also very close to the optimal EA
of 15° for best seeing. This will minimize the measurement error from other
sources—for example, the telescope bearing run out—as well as from gravitational
and trajectory errors. It may be possible to perform the alignment with the drives
powered off, which would remove telescope jitter. Finally, alignment activities can be
conducted in parallel with other IT&C activities, assuming the facility is in a state
that will allow observing.
4.2 Daytime Star Observing
Even when the sun cannot be observed, there are sufficient bright stars visible during
the day to allow some comparison with the nighttime results and to provide the data
required to populate the LUTs for thermal effects on the structure. The GIS will be
temporarily modified to prohibit anyone from entering the solar disc and solar
avoidance zones. This will prevent accidental solar illumination of the primary
mirror.
4.3 Lunar Observing
Although details are not yet available, there is a plan to observe the moon. Lunar
observation offers two main advantages. First, it allows tracking to be assessed on a
non-sidereal target. Second, because the full field of view is illuminated, aperture
“masks” can be used on the mirrors as a backup alignment check.
4.4 Pointing Model and Beam Steering
At each stage in the alignment process and throughout the IT&C phase, telescope
pointing runs will be conducted. The goal of these runs is to update the pointing
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 12 OF 53
model and, when light reaches the Coudé, the LUTs for the beam-steering optics (M3,
M6) to maintain alignment.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 13 OF 53
5 OPTICAL ALIGNMENT EQUIPMENT
5.1 Required Equipment
The following equipment is needed to perform the IT&C optical alignment:
Qty Item
Laser Trackers and CMM Arms
1 API OmniTrac2
1 FARO Vantage E
1 API Axxis6-200™ (3.2m range)
1 FARO Edge (2.7m range)
Theodolite and Related Equipment
1 Leica T3000A plus X-Y stages
Theodolite supports for the Nasmyth optical table (+X), M5 assembly interface tower, M3 assembly interface, and Coudé room (floor and VBI/WFC optical tables)
1 Video eyepiece module for the theodolite
1 Optical table with four legs (at Nasmyth) to support the theodolite and targets
Software
1 SpatialAnalyzer Software
Wavefront Sensor and Related Equipment
1 Shack-Hartmann wavefront sensor — HASO4-BB (Imagine Optic)
1 Imaging system with support for the three measurement configurations — prime focus (f /2), Gregorian focus (f /13), and telescope focus in the Coudé
Mirrors and Targets
Spherically mounted retroreflectors (SMRs) for laser trackers, magnetic bases, and other associated toolings
1 Alignment mirror for M5 assembly re-alignment
1 Pentaprism for M7 alignment
3 Alignment autocollimation mirror targets (named AC1, AC2, and AC3) with tip-tilt / centering mounts
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 14 OF 53
5.2 Metrology Requirements
5.2.1 Coordinate Measurement
Laser Trackers
The laser tracker is an essential tool for making optical alignment measurements in
large volumes. This instrument measures the 3D coordinates of large objects by
tracking a laser beam to SMR targets that attached to the object. Based on these
coordinates, laser trackers can then measure the geometry and position of objects. The
DKIST optics contracts specify a number of line-of-sight requirements that will
permit the SMR targets to be seen from the planned mounting locations of the laser
tracker.
Laser trackers will be used in the interface metrology and TMA assembly; they will
also be used to position the mirror into the telescope structure, measure structure
deformation due to TMA rotation, align instrument tables in the Coudé laboratory,
and align instruments. Two laser trackers will be available during IT&C:
API Omnitrac 2 (OT2). The typical volumetric accuracy of this laser tracker is
±40 μm in the 5 m range, ±65 μm in the 10 m range, and ±115 μm in the 20 m
range. This laser tracker features a long measurement range—up to 50 m.
FARO Vantage E. The typical volumetric accuracy of this laser tracker is ±45
μm in the 5 m range, ±70 μm in the 10 m range, and ±120 μm in the 20 m range.
The Vantage E can measure up to 25 m. It features an integrated, high-resolution,
two-camera system that can help locate a specific target quickly and efficiently.
This system operates in all lighting conditions, from complete darkness to bright
sunlight.
CMM Arms
The CMM is a portable articulated arm that has six axes equipped with rotary
encoders. A probe attached to the arm of the instrument touches the object to measure
the coordinates of the contact point. The CMM arm would significantly reduce
integration time, particularly for bench-mounted optics, instruments, and the WFC. It
will be used to align local optics, Coudé instruments, and WFC optics in the Coudé
laboratory. The following data are for the full range of the arm; however, accuracy
and repeatability improve for smaller volumes.
API Axxis6-200. The accuracy of the Axxis6-200 arm is attributable to a
mechanical design that includes high-precision bearings; titanium joints; a
lightweight, aeronautical, aluminum structure; and carbon fiber tubes. Its
volumetric accuracy is ±64 μm over a full range of the 3.2 m. Repeatability over
the full range is ±45 μm.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 15 OF 53
FARO Edge 2.7m. The Edge can be used remotely with WiFi connection and has
a built-in touchscreen and an on-board operating system. Its volumetric accuracy
is ±41 μm over a full 2.7 m range. Repeatability over the full range is ±29 μm.
5.2.2 Tilt Measurement
For measuring the angle between two mirrors accurately, the Leica T3000A
theodolite offers a very good solution. The T3000A can be used as an alignment
telescope and as an autocollimation system; it measures tilt with a theoretical
accuracy of ±0.5 arcsec (less than ±2 arcsec in a working environment). An
illuminated green cross is collimated and projected at infinity; when reflected on a
mirror, the theodolite measures the tilt change (horizontal and vertical). It can also
measure the horizontal and vertical angles between two flat mirrors. This theodolite
will be upgraded by adding a CCD camera in front of the eyepiece, which allows
remote tilt measurement.
5.2.3 Wavefront Measurement
A WFS will be required during IT&C to align the telescope mirrors and help refine
the LUTs. The instrument will be used at multiple focal locations to:
Characterize the primary focal plane
Align M2 at Gregorian focus
Optimize alignment at Coudé focal plane
Measure the wavefront on the night-sky at Gregorian and Coudé
To properly size and condition the input beam into the WFS, a collimating optic will
be required for the primary (F/2), Gregorian (F/13), and Coudé foci.
The HASO4-Broadband (Imagine Optic™), an SHWS, has been selected for
wavefront measurement. It features 50x68 microlenses and a 5.2x7.0 mm2 sensor.
The absolute RMS wavefront accuracy is λ/100 over the full bandwidth (400–1100
nm), with a minimum of 6 nm RMS. The repeatability is λ/200. The HASO4-
Broadband has a relatively large Zernike dynamic range, which is required for
DKIST alignment (see Figure 2). Its tilt dynamic range of ±3° allows a large initial
wavefront measurement before starting alignment.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 16 OF 53
Figure 2: HASO4-BB typical dynamic range of individual Zernike
The SHWS and associated optics will be mounted on a hexapod that will be attached
to following locations:
The top end optical assembly (TEAO), for the prime focus
The M3 interface, for the Gregorian focus
An optical table in the Coudé laboratory
Three SMRs will be calibrated to establish the location of the SH detector plane
within the telescope coordinate system when they are measured using the laser
trackers. The total position accuracy depends on the accuracy of the following:
The relation between the focus term (Z4) measurement and the Z position of the
SHWFS is determined by the SHWFS accuracy and the optical design. Since the
SH accuracy is 6 nm RMS, the position accuracy is <1 micron at prime focus and
0.05mm at Gregorian focus.
The environmental stability (thermal, vibration) value, as determined during
IT&C.
Calibration (< 0.05 mm, as determined using the CMM arm).
The laser tracker measurements (~0.1mm).
Assuming these conditions are met, the final position accuracy should be
approximately ±0.2 mm.
5.2.4 Software
SA software will be used for several purposes during IT&C. First, it will be used to
acquire measurements and process data from the laser trackers, CMM arms, and
theodolite. Because SA can integrate measurements from different instruments in the
same coordinate system, it facilitates the simultaneous use of the laser tracker and
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 17 OF 53
CMM arm. In addition, SA will keep acquiring measurements in the same coordinate
system even when an instrument is moved to a different position. Another important
role of this software will be to merge the mechanical CAD and optical Zemax
models. This provides several benefits for defining the metrology procedure, adding
the as-built mirrors, and extracting the data that will be used during alignment.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 18 OF 53
6 M1 ALIGNMENT
6.1 Method Overview and Flowchart
During M1 polishing, four SMRs will be attached to the mirror to register the location
of its final optical surface. Initially, M1 will be installed in its cell and aligned onto
the TMA at its nominal position using the laser tracker and SMRs.
To facilitate the M1 alignment process, the SHWS will be mounted onto the TEOA
frame near the primary focus, and the wavefront from the pole star will be measured
at night. Similar wavefront measurements will also be performed at different EAs, for
several reasons: to determine the primary focus position for these angles, to verify
that they are within the expected range, and to verify that M1’s surface figure at
different EAs falls within the expected range after correction using the LUT. During
this step, the SHWS will also be used to verify the ability of the M1 cell to inject
specified Zernike, or mirror modes, onto the mirror’s surface.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 19 OF 53
Position of the SH wrt primary focus at different Alt angles
Multiple measurement of a target star along the axial direction
Measurement at the best focus in the center and at different field positions
Analysis and determination of the center of the FOV
Wavefront measurement at optimum Alt angle
WF sensor
Measurement of M1 and SH position with the LT at different Alt angles
Programmation of the hexapod position for different Alt angles
Change of Alt angle
Primary focus position for different Alt angles (axially and laterally)
M1 Mirror surface figure at different Alt angles, after M1 LUT correction
Alignment of M1 mirror at the nominal position
M1 mirror installed into M1 cell
M1 cell installed onto TMA
Data collection and analysis
Injection of known M1 mode into M1 mirror using LUT
Retake of the wavefront measurement
Change of M1 mode
M1 response to modes injectionAnalysis of wavefront measurements
Laser tracker
Figure 1: M1 mirror optical alignment flowchart
6.2 Objectives and Preconditions
Objectives
To determine the axial position of the primary focus
To determine the lateral position of the primary focus (the center of the FOV)
To verify the LUTs—primarily the elevation LUT—for the M1 cell
To verify the M1 modal correction
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 20 OF 53
Preconditions
M1 is installed in its cell, the cell is installed on the TMA, and the nominal
position of the M1 mirror has been established using the laser tracker.
The TMA can point and track correctly.
The TEOA frame is mounted on the TMA and the SHWS is mounted on the
TEOA frame near the primary focus.
6.3 Wavefront Measurement
The position of the SHWS at different EAs will be programmed into the hexapod
based on the expected primary focus position, as determined by the M1 laser tracker
measurement and the TMA deflection analysis. Those positions will be measured
directly using the laser tracker and the SMRs that are attached to the SHWS
assembly.
6.3.1 Measure the Wavefront at the Initial EA
For this initial set of wavefront measurements, start with the telescope pointing at the
pole star at an EA of approximately 20° above the horizon.
1. Using the SHWS viewer, center the SHWS onto the pole star and take multiple
wavefront measurements along the axial direction (through focus measurement).
2. Analyze the measurement data and determine the position of best focus, where the
Z4 term is zero or at a minimum. Compare this with the expected position.
3. Using the best axial focus position, take a set of wavefront measurements on a
grid of lateral positions by first moving the SHWS to a predetermined position
and then pointing the telescope to center the star.
NOTE: The grid and step size will be based on the expected accuracy of the primary focus after it is aligned using the laser tracker.
4. Analyze the measured wavefront and compare it to the predicted one based on the
Zemax model at multiple lateral positions to determine the center of the FOV1.
5. Analyze the measured wavefront and compare it to the predicted shape of the
mirror surface.
6. Confirm and fine-tune the M1 cell LUT.
1 Off-axis aberration and LUT error can be distinguished by looking at the wavefront over multiple fields
and comparing it to the designed wavefront.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 21 OF 53
6.3.2 Measure the Wavefront at Other EAs
1. Start at the first elevation angle (EA1) and lock onto a target star using telescope
pointing.
2. Use the SHWS viewer to center the SHWS onto the target star. Take multiple
wavefront measurements along the axial direction (through focus measurement).
3. Analyze the measurement data and determine the position of best focus. Compare
this with the expected position.
4. Using the best axial focus position, take a set of wavefront measurements on a
grid of lateral positions by first moving the SHWS to a predetermined position
and then using telescope pointing to center the star.
NOTE: The grid and step size will be based on the expected accuracy of the primary focus after it is aligned using the laser tracker.
5. Analyze the measurement data and determine the center of the FOV2.
6. Analyze the measured wavefront and compare it to the predicted mirror surface
shape; also, confirm and fine-tune the M1 cell LUT.
7. Move on to EA2 and repeat steps 1–6, and then repeat these steps for a number of
EAs.
8. After you have measured the wavefront at all of the EAs, confirm the primary
focus position for each (both axially and laterally). Also, confirm that…
The primary focus positions are within the expected range.
The mirror surface figure at each EA is within the expected range after
correction using the M1 LUT.
NOTE: Multiple iterations may be needed to fine-tune the surface figure LUT for M1.
6.3.3 Analyze the Wavefront
The detail analysis of the wavefront will be presented in the document PROC-0098.
6.4 Modal Correction
Once you complete the steps outlined in Section 6.3, use the SHWS to verify the M1
cell’s ability to inject specified Zernike or mirror modes onto the mirror.
2 Off-axis aberration and LUT error can be distinguished by looking at the wavefront over multiple fields
and comparing it to the designed wavefront.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 22 OF 53
1. Establish a baseline by taking wavefront measurements using the pole star as the
target, approximately 20° above horizontal pointing.
NOTE: Baselines should be repeated periodically to monitor possible changes.
2. Inject a known mode into the M1 mirror (an input wavefront) using the LUT via
the M1 control system (CS).
3. Retake the wavefront measurements.
4. Analyze the measured wavefront and compare it to the input wavefront.
5. Repeat the measurement for as many modes as desired.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 23 OF 53
7 M2 ALIGNMENT
7.1 Method Overview and Flowchart
Initially, the M2 mirror will be mounted into the TEOA and aligned, using a laser
tracker and SMRs, relative to the measured primary focus position. As described for
M1, three SMRs will be attached to the M2 mirror during polishing to register the
location of the optical surface. The SHWS will be mounted near the Gregorian focal
plane (near the M3 mirror mount).
During this phase of alignment, the LUT for the M1 mirror will be refined to optimize
the wavefront at Gregorian focus. As described for M1, the wavefront measurement
at Gregorian focus will be acquired at night on Polaris; the SHWS will be used to
determine the position of best focus, where the Zernike Z4 term (defocus) is at a
minimum. This measured wavefront will be compared to the wavefront by predicted
by the Zemax model at multiple lateral positions to determine the M1/M2 alignments
and the position (lateral and axial) of the Gregorian focus. The same measurement
will be repeated at different EAs, and an M2 LUT will be established to provide a
stable Gregorian focus position for these angles. Zernike measurements obtained in
the field and at different EAs will provide input data for the optical alignment tool;
they will also define the optimal position of M2 (6° of freedom) at each EA that
minimizes the WFE.
The M2 LUT may require further adjustments to balance between improving the
positioning of the Gregorian focus and improving the WFE. Three LUTs may be
created to effect this process:
An LUT that optimizes the WFE at Gregorian focus, but changes the focus with
EA.
AN LUT that keeps the Gregorian focus at the same position for all EAs, but
gives a degraded WFE.
An LUT that compromises between WFE and Gregorian focus positon.
Some of the residual WFEs can be corrected by updating the M1 LUT. After updating
the M1 LUT, additional iterations of wavefront measurement will be performed to
confirm the effectiveness of the updates.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 24 OF 53
Laser tracker
Multiple measurement of a target star along the axial direction
Measurement at the best focus in the center and at different field positions
Analysis and determination of the center of the FOV
M2 alignment to optimize wavefront error
WF sensor
Mounting of the SH wavefront sensor onto M3 mirror mount
Change of Alt angle
Gregorian focus position for different Alt angles (axially and laterally)
Alignment of M2 mirror at the nominal position
M1 mirror aligned at primary focus
M2 mirror mounted onto TEOA
Alignment of the SH at the nominal Gregorian focus with the LT
Improvement of Gregorian focus position
Updating of M1 LUT
Telescope performances at Gregorian focal plane
Final wavefront measurement at Gregorian focus
Figure 2 : M2 mirror optical alignment flowchart
7.2 Objectives and Preconditions
Objectives
Establish a LUT for the M2 hexapod as a function of EA
Align the M2 mirror to M1 mirror
Position the Gregorian focus at its nominal position.
Refine the LUT for the M1 mirror to optimize the wavefront at Gregorian focus
Preconditions
The M1 mirror is aligned according to procedures defined in Section 6.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 25 OF 53
The M2 mirror is mounted onto the TEOA and has been aligned with the laser
tracker. The initial mirror position should be near the hexapod’s center of travel.
7.2.2 Perform the Shack-Hartmann Alignment
1. Mount the SHWS onto the M3 mirror mount interface.
2. Align the SHWS to the expected Gregorian focus using the laser tracker.
7.2.3 Perform the M2 Alignment
The initial LUT for the M2 hexapod motion will be computed based on the Zemax
model and the movement of the primary focus position measured during M1
alignment (see Section 6).
1. As described for M1, start at the initial EA using the pole star as the target. Use
the SHWS viewer to center the instrument onto the target star. Take multiple
wavefront measurements along the axial direction (through focus measurement).
2. Analyze the measurement data and determine the position of best focus, where the
Zernike Z4 term (defocus) is zero or at a minimum. Compare this with the
expected position.
3. Using the best axial focus position, take a set of wavefront measurements on a
grid (possibly 5x5) of lateral positions by first moving the SHWS to a
predetermined position, and then using telescope pointing to center the star.
4. Analyze the measured wavefronts and compare to the predicted wavefront based
on the Zemax model at multiple lateral positions to determine the center of the
FOV3.
5. Move on to EA2 and repeat the wavefront measurements. Repeat them for a
number of EAs.
6. After wavefront measurements for all telescope EAs are complete, confirm that
the Gregorian focus positions (axial and lateral) have been determined and that
they are within the expected range. In addition, confirm the wavefront
performance of combined M1 and M2 at those positions.
7. Examine the WFEs for these EAs and compute their respective Gregorian focus
positions relative to other telescope reference points.
NOTE: You may need to further adjust the M2 LUT to balance between improving the positioning of the Gregorian focus and improving the WFE. For example, if you observe a larger-than-expected amount of astigmatism and/or coma, you can correct
3 Off-axis aberration and LUT error can be distinguished by looking at the wavefront over multiple fields
and comparing it to the designed wavefront.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 26 OF 53
using the M2 position (and let the Gregorian focus position drift) or use M1 to correct the wavefront and fix the focus position.
7.2.4 Optimize the Wavefront
Once the M2 position LUT is established and provides a stable (or known) Gregorian
focus for different EAs, review the measured wavefront data to identify residual
errors. Some residual WFEs can be corrected by updating the M1 LUT. After
updating the LUT, be sure to perform additional wavefront measurement iterations to
confirm effectiveness.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 27 OF 53
8 OSS TRANSFER OPTICS ALIGNMENT
8.1 Method Overview and Flow Chart
The OSS transfer optics include four mirrors—M3, M4, M5, and M6—that must be
aligned with respect to the telescope’s altitude and azimuth. Alignment will be
conducted in several stages that include defining the altitude and azimuth axes using a
theodolite and a set of mirrored targets; axis orientation will be determined based on
tilt of these targets while axis position is based on their centers. In addition to the four
OSS mirrors, the alignment procedure for M7 is included here because it can be used
to define the azimuth axis.
The altitude axis will be defined based on the Nasmyth platform and the azimuth axis
will be defined based on the Coudé laboratory. The theodolite will be positioned in
both of these locations, which will also facilitate characterization of the mechanical
stability of the telescope’s altitude-azimuth rotation.
Telescope pointing runs will be conducted at each stage of this process to update the
LUTs for the beam-steering optics (M3, M6) and to maintain the position of the
beam.
Mirrors M3, M4, M5, and M6 will be positioned into their mounts using a laser
tracker. The flat mirrors (M3, M5, and M6) will be accurately tilt-aligned with respect
to the altitude axis using the theodolite and mirrored targets. A (flat) mirrored target,
AC1, will be positioned at the M4 location during these alignment procedures;
consequently, the M4 mirror will be integrated last, after the procedures have been
completed.
Various alignment tests will be conducted using the theodolite located in the Coudé
laboratory. The rotation of the altitude-azimuth axis will be characterized when the
Coudé is static and the telescope is rotating. The theodolite will be used to confirm
the stability of folded (90°) beam produced by when M5 and M6 mirrors are aligned
on the altitude axis. If the azimuth and altitude axes are perfectly perpendicular
(within the ±2 arcsec accuracy of the theodolite), the beam’s tilt should remain
constant during alt-az rotation; if not, the theodolite will measure this deviation and
the mirrors will be adjusted to minimize it. When the Coudé is rotating, the tilt will be
optically measured with respect to the TMA and compensated as needed.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 28 OF 53
Figure 3: OSS transfer optics alignment flowchart
Theodolite, AC1, AC2
Alignment of M6 angle with respect to Alt axis
Measurement of tower tilt stability when TMA Alt is rotated
Mounting and alignment of M5 assembly with laser tracker
Alignment of M5 angle with respect to Alt axis
Alignment of AC1 and AC2 on the Alt axis
AC1, AC2, laser tracker and theodolite install
Mounting and alignment of M6 assembly with laser tracker
Mounting of AC3 on the tower
Alignment of AC3 on the Az axis
M6 assembly
Laser tracker
Definition of the Alt axis
Tower stability with TMA Alt rotation
M5 assembly
AC3
Definition of the Az axis
Characterization of Alt-Az rotation and perpendicularity with respect to M5-M6
Adjustment of M6 tilt
M7 assembly mounting and alignment in Coudé coordinate system wrt Az axis
Coudé rotation characterizationTilt and decenter of the Coudé rotation
axis wrt Az axis
Alignment of theodolite on Alt axis and alignment of AC1 on M4 local tilt
Mounting and alignment of M4 assembly with laser tracker
M4 assembly
M7 assembly
Mounting and alignment of M3 assembly with laser tracker
Alignment of M3 angle with theodolite and GOS target
M3 assembly
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 29 OF 53
8.2 Definition of the Telescope’s Altitude Axis
Before any of the mirror assemblies (M3, M4, M5, and M6) are mounted, the
telescope’s altitude axis will be defined using mirror targets AC1 and AC2 and a
theodolite. These flat mirrors will be reflective and feature a target that defines their
centers. Their tilts will be used to define the tilt of the altitude axis, and their centers
will be used to define the axis position. The outside diameters of the ACs are
concentric to the reticle within 0.03 mm and cylindrical to within 0.008mm. They will
be mounted on a tool that allows X-Y tip-tilt and X-Y centering.
Figure 4: Example of an AC target
1. Position the theodolite on an optical table on the Nasmyth platform at the +X side
of the mount. It should be fixed on the altitude axis, facing M4.
2. Place the AC1 mirror on the –X trunnion of the mount. It should be on the
rotating part of the TMA, at the optical surface position of M4 (located at
X = 3949.660 mm from the azimuth axis).
Figure 5: Position of the theodolite, AC1 and AC2 mirrors
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 30 OF 53
NOTE: The AC1 mount should enable tip-tilt and centering adjustments.
3. Position AC2 on the +X trunnion.
4. Rotate the TMA in the altitude direction over its full range (7° to 90°) in 10°
increments and use the theodolite to measure the AC1 tilt and center at each EA.
The AC1 mirror should rotate with the telescope while the theodolite stays fixed
on the Nasmyth floor.
5. Align AC1 iteratively on the altitude axis by adjusting its tip-tilt and centering so
it has no tilt or center variation in the theodolite during rotation.
6. Align the theodolite tilt on the altitude axis by adjusting it so the autocollimated
reticle is centered.
7. Follow the same procedure to center the theodolite on the altitude axis: In visual
mode, center the AC1 target into the theodolite by shifting the theodolite in the
two directions (without changing the tilt).
8. Place AC2 on the +X trunnion of the TMA and align it on the altitude axis using
the same method you used for AC1.
NOTE: You must be able to remove AC2 and replace it in the same location accurately.
The measurements for AC1 and AC2 should be identical.
This test will identify the difference between the rotation axes of the two
TMA trunnions and provide a second reference to reposition the theodolite on
the altitude axis.
8.3 M5 and M6 Alignment
The M5 and M6 mirrors will be aligned using a laser tracker and theodolite. This
procedure requires replacing the theodolite’s eyepiece with a camera to visualize and
acquire measurements during telescope rotation. Tilt will be measured with the
theodolite in autocollimation mode, with the focus on infinity, and centering will be
measured in visual mode, with the focus on the target surface. Because the focus can
only be adjusted manually, the tilt and centering must be aligned separately in two
steps (tilt first, then centering).
M6 will be aligned first, since the theodolite must be positioned at the M5 assembly,
followed by stability testing, and then alignment of M5.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 31 OF 53
8.3.1 M6 Alignment
The M6 assembly will be integrated on the TMA tower and positioned in the
telescope coordinate system using a laser tracker. The tilt will be aligned using the
theodolite attached to the tower.
1. Position the theodolite on the tower on a dedicated support at the M5 assembly
interface.
Set the height of the theodolite on the altitude axis.
Position the theodolite to allow autocollimation on the M6 optical surface.
2. Allow the theodolite to autocollimate onto the AC1 mirror.
3. Set the theodolite at 60º upwards via a 90° pentaprism.
4. Tilt-align M6 on theodolite so it is 30º from the altitude axis. The horizontal tilt of
M6 should be the same as the altitude axis.
5. Check the M6 position using the laser tracker and realign if necessary by
monitoring the tilt with the theodolite.
Figure 6: Theodolite turned 60º upward to align M6
8.3.2 Tower Stability Test
This test entails rotating the TMA and measuring the tilt and decenter change on AC1
to characterize accurately the stability of the tower. In this test configuration, the
theodolite will be attached to the tower and aligned on AC1.
1. Autocollimate the theodolite onto the AC1 mirror.
2. Rotate the telescope mount around the altitude axis.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 32 OF 53
3. Use the theodolite to acquire the tilt and position of AC1 during rotation at
different EAs.
If the tower is stable during rotation, the measurement should also be stable.
If not, the theodolite measurement will show the residual tilt and decentering
with respect to the altitude axis.
8.3.3 M5 Alignment
M5 will be aligned using the same method as M6, but with the theodolite placed at a
different position.
1. Dismount the theodolite and its support from the tower.
2. Position the theodolite on the altitude axis close to the M6 mount hole, facing
AC1.
3. Integrate the M5 assembly onto the tower and align it with the laser tracker in the
telescope coordinate system.
4. Allow the theodolite to autocollimate onto AC1 and set it horizontally at 180º and
downward at 15º.
5. Align the M5 tilt onto the theodolite; the horizontal tilt should be the same as the
altitude axis.
6. Check the M5 position with the laser tracker and realign if necessary by
monitoring the tilt with the theodolite.
Figure 7: Position of the theodolite to align M5
8.4 M3 Alignment
Once M5 is aligned, the M3 assembly interface can be freed from the TMA structure.
M3 alignment can be initiated after the AC1 target has been aligned. The M3
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 33 OF 53
assembly will be mounted and initially aligned in the telescope coordinate system
using the laser tracker and attached SMRs. The tilt will be aligned with respect to
AC1 using the theodolite in autocollimation mode with the focus on infinity. In order
to center the theodolite on the altitude axis close to the M3 mirror, a support tool will
need to be attached to the TMA structure between AC1 and M3.
The following steps describe the alignment sequence:
1. Align the theodolite tilt on AC1; you do not need to center the theodolite on the
elevation axis.
2. Rotate the theodolite horizontally and vertically to point toward the M3 optical
surface.
3. Align the tilt of M3 with respect to the theodolite.
8.5 Alt-Az Alignment
8.5.1 Defining the Azimuth Axis
Another mirrored target, AC3, will be used to define the azimuth axis. This target will
be aligned perpendicular to the azimuth axis and, like the AC1/AC2 targets, it must
enable tip-tilt and centering adjustments. The theodolite will be positioned in the
Coudé to align AC3 when the telescope rotates in azimuth. The following steps
describe the main stages of the procedure:
1. Position the AC3 mirror on the azimuth axis.
2. Position the theodolite on the Coudé floor and align its tilt using Coudé
references.
3. Most likely, the M7 mirror will be available.
If so, follow the alignment procedure described in Section 8.6 so it can be
used as a fold mirror during the alt-az alignment process (Section 0).
If not, use a small fold mirror (with a diameter of at least 10mm) mounted at
the M7 position and aligned 45º vertically using the theodolite. The horizontal
tilt can be defined using the Coudé references.
4. Rotate the TMA in the azimuth direction over ±180° in 10° increments. The
Coudé floor should not rotate.
5. Using the theodolite at different azimuth angles, measure the tilt and center of
AC3.
6. Align AC3 iteratively by adjusting its tip-tilt and centering to remove and tilt and
center variation during azimuth axis rotation.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 34 OF 53
Figure 8: AC3 autocollimation mirror and theodolite in the Coudé room
8.5.2 Alt-Az Alignment
1. Align the theodolite on the azimuth axis by adjusting its tilt so that the AC3
autocollimated reticle is centered.
2. Align the theodolite position by centering it on the azimuth axis, in visual mode:
Center the AC3 target in the theodolite eyepiece by shifting the theodolite.
Do not change the theodolite’s tilt.
3. Remove AC3.
The theodolite stays fixed so the beam emanating from it passes through M7,
M6, and M5.
This beam should autocollimate on AC1 and return into the theodolite.
4. Rotate the TMA in the alt-az directions. The Coudé does not rotate at this stage.
5. Using the theodolite, confirm that the beam orientation does not change with
TMA rotation.
AC1 has been aligned on the altitude axis and AC3 on the azimuth axis. M5
and M6 have been aligned on the altitude axis to produce a 90° folded beam.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 35 OF 53
If the azimuth and altitude axes are perfectly perpendicular (within the
theodolite accuracy ±2 arcsec), the alt-az measurement should remain constant
when the telescope rotates.
If the axes are not perfectly perpendicular, the theodolite will measure this
deviation during alt-az rotation.
6. Adjust the tilt of M6 to correct and reduce this deviation.
8.6 M7 Alignment and Coudé Characterization
8.6.1 M7 Alignment
M7 will be aligned with respect to the azimuth axis using a laser tracker and the three
SMRs that are attached to the mirror substrate. All of the measurements discussed
below reference the Coudé coordinate system.
1. If a fold mirror was used in Section 8.5.1, remove it; otherwise, proceed to step 2.
2. Install the M7 assembly in the Coudé laboratory.
3. Position the laser tracker between M7 and M8.
4. Measure the SMRs and align the mirror to its nominal position.
5. Replace the laser tracker with the theodolite.
6. Align the theodolite tilt using the Coudé references.
7. Using either a pentaprism or dedicated theodolite support, point at 45º directly
toward the M7 optical surface.
If you use a 45º pentaprism, place it between the theodolite and the mirror.
A dedicated theodolite support requires a tool but is more accurate.
8. Align the tilt of M7 mirror with the theodolite.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 36 OF 53
Figure 9 : Position of theodolite for M7 alignment
8.6.2 Coudé Rotation Characterization
The Coudé rotation axis should be characterized with the respect to the azimuth axis
as follows:
1. Rotate the Coudé floor over its full ±180° range in 10° increments.
2. Use the theodolite to measure the autocollimation tilt and center of AC1 through
M5, M6, and M7.
If the Coudé rotation and azimuth axes are colinear, the theodolite
measurement should be stable.
If there is a residual tilt or decenter between the two axes, this test will
measure the residual tilt and decenter of the Coudé axis.
8.6.3 Coudé Optics Alignment Preparation
A flat alignment mirror, AM, will be set up with a tilt related to the M7 mirror. This
mirror will be placed onto the M9a optical table and used to align the tilt of the M9
mirror.
1. While autocollimated on AC1, rotate the theodolite horizontally by 90º and place
AM on the M9a optical table in the theodolite’s FOV.
2. Align the tip-tilt of AM to autocollimate on the theodolite.
M7 M8
Theodolite
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 37 OF 53
Figure 10: M9 alignment configuration
8.7 M4 Alignment
Once M3, M5, M6, and M7 are aligned and the TMA and Coudé rotations have been
characterized, AC1 can be dismounted and replaced with the M4 assembly. M4 will
be aligned using only a laser tracker; a theodolite cannot be used because this mirror
is not flat.
The initial alignment step is basically mechanical because M4 has low alignment
sensitivities; the mirror can be aligned using its fiducials.
If M4 has a centered target, centering can be verified and corrected using the
theodolite, which is already aligned on the altitude axis.
Th
AM
M7
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 38 OF 53
9 COUDÉ TRANSFER OPTICS ALIGNMENT
9.1 Method Overview and Flowchart
Mirrors M8, M9, and M10 will be aligned in the Coudé coordinate system.
M8 will be aligned using only a laser tracker, M9 will be aligned using laser tracker
and theodolite, and M10 will be aligned with the CMM arm and theodolite. Note that
M8 and M9 both feature SMRs that can be used to position these mirrors accurately
in the Coudé laboratory.
Because the tolerance required for beams feeding the different instruments is tight
(typically ±10 arcsec), the most sensitive alignment parameter is the tilt of the flat
mirrors, M9 and M10. Consequently, the theodolite will be used to accurately align
the tilt of M9 with respect to M7, and the M9 mirror angle will then be used as
reference to tilt-align all other Coudé flat mirrors (and beam-splitters). Aligning all of
the flat optics with respect to the same reference eliminates the accumulation of
measurement error.
Once M9 is aligned in the Coudé coordinate system, the theodolite can be used to
align M10 with respect to M9’s optical surface. M10, a deformable mirror, will be
mounted on the optical table that supports the wavefront correction system (WFCS)
and the visible-light broadband imager (VBI). M10 can then be aligned using the
CMM arm instead of the laser tracker.
Measurements will initially be referenced in the Coudé coordinate system; however,
all Coudé measurements will be converted to the telescope’s coordinate system using
SA software. This conversion will require a set of transfer coordinates. To develop
these coordinates, a laser tracker will be used to measure SMRs (in the Coudé
system) that are located on a nonrotating Coudé wall at zero degree of the encoder.
Three SMRs will also be attached to the WFCS/VBI optical table (where the CMM
arm will also be set up) and measured using both the laser tracker and CMM arm.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 39 OF 53
Laser tracker
Setup of the CMM arm and transfer of the coordinate system
Mounting and alignment of DM assembly in Coudé coordinate system
Alignment of M10 mirror angle wrt to M9 mirror with the theodolite
Mounting, alignment and calibration of the focusing optic and wavefront sensor
Mounting and alignment of M9 assembly in Coudé coordinate system
Mounting and alignment of M8 assembly in Coudé coordinate system
M8 assembly
Alignment of M9 mirror angle wrt to M7 mirror with the theodolite
Wavefront error measurement on the sky of selected stars in center, nominal and extended field at different Alt-Az angles
M2 optimization
Telescope performances at different Alt-AzFinal telescope characterization
M9 assembly
TheodoliteAM mirror
CMM arm
DM assembly
Focusing optic
Wavefront sensor
Figure 11 : Coudé transfer optics alignment flowchart
9.2 Alignment of M8, M9, and M10 Mirrors
9.2.1 M8
The M8 assembly will be mounted and then aligned using the laser tracker in the
Coudé coordinate system. Data extracted from the CAD model and as-built will be
entered into SA to calculate the real position of the vertex and optical axis with
respect to a local coordinate system that will be developed using the mirror SMRs.
During the alignment process, the mirror’s three SMR positions will be measured in
the Coudé coordinate system.
SA will be used to achieve two objectives during this process: to convert the
measured position of the vertex and optical axis to the telescope coordinate system
and to calculate the needed shift and tilt to position the M8 optical surface at its
nominal position.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 40 OF 53
9.2.2 M9
The M9 assembly will be mounted and aligned using the laser tracker in the Coudé
coordinate system. The M9 mirror angle will be aligned with respect to M7 using the
theodolite (see Section 8.6.1).
1. Place the theodolite at position “Th” and allow M9 to autocollimate.
2. Set up the theodolite.
3. Perform an autocollimation on AM.
4. Rotate the theodolite toward M9 by 89.3425º.
5. Align M9’s tilt to autocollimate on the theodolite.
6. Check the position of M9 with the laser tracker, and correct it if needed by
keeping the theodolite autocollimated on the M9 optical surface.
Figure 12: M9 alignment configuration
9.2.3 CMM Arm Setup
The Coudé optics and CMM arm must be mounted on the same optical table. This
step will ensure that the SMRs are accessible to the CMM arm on the WFCS/VBI
bench.
1. Set up the CMM arm on the bench.
2. Place on the three SMRs on the optical table and measure their position with the
laser tracker in the Coudé coordinate system.
3. Measure the three SMRs with CMM arm.
4. In SA, cophase the CMM arm and laser tracker so they are measuring in the
Coudé coordinate system.
Th
AM
M7
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 41 OF 53
9.2.4 M10
The M10 assembly will be mounted and aligned with the CMM arm in the Coudé
coordinate system. The tilt of M10 will be set with respect to M9.
NOTE: The VBI and WFC optics should not yet be mounted on the benches.
1. Mount and set up the M10 assembly on its bench.
2. Align the position of M10 with the CMM arm.
3. Place the theodolite at the intersection between the normal of the two mirrors, M9
and M10.
This intersection is represented on the following drawing by Th.
M9 will serve as reference to align M10’s tilt.
5. Setup the theodolite on a stand.
6. Autocollimate the theodolite on M9.
7. Rotate the theodolite toward M10 by 155.0000º.
8. Align M10 in autocollimation on the theodolite.
Figure 13 : M10 alignment configuration
M7
M9
M8
M10
Th
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 42 OF 53
10 TELESCOPE RESIDUAL ERROR
ALIGNMENT
10.1 Method
The telescope’s final WFE will be measured by adding calibrated, focusing optics
that will re-image the collimated beam emanating from M10. The SHWS will be used
to measure the WFE on the sky in the center of the field and in nominal and extended
fields. The WFE will be acquired for the entire range of alt-az angles, and the M2/M4
mirrors will be used to compensate for any residual aberrations—typically
astigmatism, coma, and focus—that are related to the manufacturing tolerances of the
Coudé mirrors. Misalignment should not induce significant aberrations because the
alignment sensitivities of M7, M8, M9, and M10 are negligible.
10.2 Cumulative Focus Error
The procedures described in Sections 6–9 of this plan will result in a fully aligned
telescope; however, residual power errors within each optic will cause a non-ideal
configuration. To minimize the effect on M2 and the PA&C Gregorian pinhole,
adjustments will be made only to M4, in two iterations, based on the following:
An estimate of the initial shift offset from the as-built test results for each optic
The final measurement at the M10 location, where M4 will be offset to
compensate for the residual error
Initial estimates for the offset, described in TN-0240, “IT&C Focus Compensation
Analysis,” are for less than 5mm of movement in M4.
10.3 Final Measurement and Correction
This last task will indicate how the telescope will perform after optical alignment and
before setup of the WFCS. General steps are outlined below.
1. Mount and align the imaging optic system needed to measure the telescope
wavefront.
2. Mount the WFS.
3. Calibrate the WFS and imaging optic system.
4. Use the WFS to measure the wavefronts of selected stars—on axis, in the center
of the field, and in the nominal and extended field, at different alt-az positions of
the telescope. These measurements will determine the telescope performances.
5. Optimize the telescope performances initially using M4 as compensator. You can
also use the M2 mirror to compensate for residual astigmatism, coma, and focus.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 43 OF 53
6. Measure the telescope’s focal length:
Select different known stars in the FOV.
Measure the relative position of the WFS to center these stars in the field.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 44 OF 53
11 FIDO ALIGNMENT
11.1 Method Overview and Flowchart
The feed optics consists of the following:
The beamsplitter for the WFC (WFC-BS1).
Five additional beamsplitters (CL2, CL2a, CL3, CL3a, and CL4)
Two mirrors (M9a and DL-FM1)
The procedure for aligning these optics will be similar to the one used for the Coudé
mirrors: Their position will be determined using the laser tracker or CMM arm and
their tilt will be aligned using the theodolite with respect to M9. The initial alignment
is trivial. The main challenge stems from the need for interchangeability between
CL2, CL2a, CL3, CL3a, and CL4 during telescope operation. In addition, the lack of
space between optical elements imposes constraints, which are accounted for in this
alignment plan.
All alignments will reference the Coudé coordinate system. The beamsplitters and
mirrors will be aligned using a laser tracker and CMM arm. The beamsplitter mounts
will be used as metrology surfaces. For flat surfaces (M9a, DL-FM1, and all
beamsplitters), tilt—the most sensitive alignment parameter— will be aligned using a
theodolite. Once M9 is aligned in the Coudé coordinate system, it will serve as the
reference for all of the beamsplitter tilts, an approach that is designed to eliminate
theodolite measurement errors.
The M9a mirror will be aligned using the laser tracker and theodolite. The other
optics—WFC-BS1, CL2, CL3, CL4, CL2a, CL3a, and DL-FM1—will be aligned
using the CMM arm and theodolite. The theodolite will be mounted on a stand for
M9a, WFC-BS1, CL3, and CL4 and on the optical table for CL2, CL2a, CL3a and
DL-FM1. The alignment sequence described below considers the position and space
needed for the theodolite. Note that M9a can be aligned at any time in the sequence.
NOTE: Before the beamsplitter alignment procedure can be initiated, all of the benches must be integrated at their position in the Coudé room and only M10 must be installed on its bench.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 45 OF 53
CMM arm
Alignment of M9a angle wrt to M9 mirror with the theodolite
Mounting and alignment of CL2a assembly with the CMM arm
Alignment of CL2a angle wrt to M9 mirror with the theodolite
Mounting and alignment of CL2 assembly with the CMM arm
TheodoliteAlignment of WFC-BS1 angle wrt to M9
mirror with the theodolite
Mounting and alignment of WFC-BS1 assembly with the CMM arm
WFC-BS1 assembly
Mounting and alignment of M9a assembly with the laser tracker
Alignment of CL2 angle wrt to M9 mirror with the theodolite
Laser tracker
M9a assembly
CL2a assembly
CL2 assembly
Mounting and alignment of CL3a assembly with the CMM arm
Alignment of CL3a angle wrt to M9 mirror with the theodolite
CL3a assembly
Setup the theodolite at the second position and transfer coordinate system
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 46 OF 53
Alignment of DL-FM1 angle wrt to CL2 with the theodolite
Mounting and alignment of CL3 and CL4 assemblies with the CMM arm
Alignment of CL3 and CL4 angles wrt to M9 mirror with the theodolite
CL3 and CL4 assemblies
Mounting and alignment of DL-FM1 assembly with the CMM arm
DL-FM1 assembly
Figure 14 : Coudé beamsplitters and mirrors alignment flowchart
11.2 Alignment Procedures
11.2.1 WFC-BS1
WFC-BS1 should be aligned after M10 because the theodolite will already be in close
proximity.
1. Mount and set up the WFC-BS1 assembly on the bench.
2. Align WFC-BS1 using the CMM arm.
3. Place the theodolite on a stand at the intersection between the normal of M9 and
WFC-BS1. M9 will serve as reference for aligning the tilt of WFC-BS1.
4. Set up the theodolite.
5. Autocollimate the theodolite on M9.
6. Rotate the theodolite toward WFC-BS1 by 154.5000º.
7. Align WFC-BS1 to be in autocollimation on the theodolite.
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 47 OF 53
Figure 15 : WFC-BS1 beamsplitter alignment configuration
11.2.2 M9a
M9a will be aligned using the laser tracker. Its tilt will be set with respect to M9 using
the theodolite:
1. Mount and set up M9a assembly.
2. Align M9a using the laser tracker.
3. Place the theodolite on a stand at the intersection between the normal of M9 and
M9a (coordinates TBD). M9 will serve as reference for aligning the tilt of M9a.
4. Set up the theodolite.
5. Autocollimate the theodolite on M9.
6. Rotate the theodolite toward M9a by 161.0000º.
7. Align M9a to be in autocollimation on the theodolite.
M9
WFC-BS1
Th
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 48 OF 53
Figure 16 : M9a alignment configuration
11.2.3 CL2a
CL2a will be aligned using the CMM arm. Its tilt will be set with respect to M9 using
the theodolite.
2. Mount and set up the CL2a assembly on its bench.
3. Align CL2a using the CMM arm.
4. Place the theodolite on the optical table at the intersection between the normal of
M9 and CL2a. M9 will serve as reference to align the tilt of CL2a.
5. Set up the theodolite.
6. Autocollimate the theodolite on M9.
7. Rotate the theodolite toward CL2a by 125.2434º.
8. Align CL2a to be in autocollimation on the theodolite.
M9
M9a
Th
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 49 OF 53
Figure 17 : CL2a beamsplitter alignment configuration
11.2.4 CL2
CL2 will be aligned using the CMM arm. Its tilt will be set with respect to M9 using
the theodolite:
1. Mount and set up the CL2 assembly on its bench.
2. Align CL2 using the CMM arm.
3. Place the theodolite on the optical table at the intersection between the normal of
M9 and CL2. M9 will serve as reference to align the tilt of CL2.
4. Set up the theodolite.
5. Autocollimate the theodolite on M9.
6. Rotate the theodolite toward CL2 by 125.7434º.
7. Align CL2 in autocollimation on the theodolite.
M9
CL2a
Th
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 50 OF 53
Figure 18 : CL2 beamsplitter alignment configuration
11.2.5 CL3a
CL3a will be aligned using the CMM arm. The tilt of CL3a will be set with respect to
M9 using the theodolite:
1. Mount and set up the CL3a assembly on its bench.
2. Align CL3a with the CMM arm.
3. Place the theodolite on the optical table at the intersection between the normal of
M9 and CL3a. M9 will serve as reference to align the tilt of CL3a.
4. Set up the theodolite.
5. Autocollimate the theodolite on M9.
6. Rotate the theodolite toward CL3a by 154.9991º.
7. Align CL3a in autocollimation on the theodolite.
M9 CL2
Th
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 51 OF 53
Figure 19 : CL3a beamsplitter alignment configuration
11.2.6 CL3 and CL4
CL3 and CL4 will be aligned using the CMM arm. Their tilt will be set with respect
to M9 using the theodolite. CL3 and CL4 can be tilt-aligned simultaneously because a
single theodolite position allows autocollimation on M9, CL3, and CL4.
1. Move the CMM arm to a new position on the bench that allows the measurement
of CL3, CL4, and the DL-FM1 mirror.
2. Measure the three SMRs using the CMM arm. These measurements will be
conducted in the Coudé coordinate system and used to create a new coordinate
system that is identical but shifted.
3. In SA, cophase the new coordinate system so the CMM arm is measuring in the
Coudé coordinate system.
4. Mount and set up the CL3 assembly on the bench.
5. Align CL3 using the CMM arm.
6. Mount and set up the CL4 assembly.
7. Align CL4 using the CMM arm.
8. Place the theodolite on a stand at the intersection between the normal of M9, CL3
and CL4 (coordinates TBD). M9 will serve as reference for aligning the tilt of
CL3 and CL4.
9. Set up the theodolite.
10. Autocollimate the theodolite on M9.
11. Rotate the theodolite toward CL3 by 154.4991º.
M9 CL3a
Th
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 52 OF 53
12. Align CL3 in autocollimation on the theodolite.
13. Repeat the autocollimation on M9.
14. Rotate the theodolite toward CL4 by 54.7566º.
15. Align CL4 in autocollimation on the theodolite.
Figure 20 : CL3 and CL4 beamsplitter alignment configuration
11.2.7 DL-FM1
DL-FM1 will be aligned using the CMM arm. Its tilt cannot be aligned directly with
respect to M9; instead, since its normal line intersects that of the CL2 beamsplitter,
DL-FM1 will be aligned with respect to CL2.
1. Mount the DL-FM1 assembly.
2. Align DL-FM1 using the CMM arm.
3. Place the theodolite on the optical table at the intersection between the normal of
DL-FM1 and CL2. CL2 will serve as reference for aligning the tilt of DL-FM1.
4. Set up the theodolite.
5. Autocollimate the theodolite on CL2.
6. Rotate the theodolite toward DL-FM1 by 120.7440º.
7. Align DL-FM1 in autocollimation on the theodolite.
M9 CL3
Th
CL4
DKIST | PROC-0018
DKIST Optical Alignment Plan
PAGE 53 OF 53
Figure 21 : DL-FM1 alignment configuration
DL-FM1
CL2
Th