directvision® instruction manual€“system2.pdf · directvision® instruction manual page iii...
TRANSCRIPT
DirectVision® Instruction Manual
Page ii
DirectVision® Instruction Manual
Page iii
Table of Contents
GENERAL GUIDANCE .......................................................................... 1
INTRODUCTION ...................................................................................... 1
INDICATIONS FOR USE ............................................................................. 1
PRODUCT OVERVIEW ......................................................................... 2
DIRECTVISION® SYSTEM ........................................................................... 2
DIRECTVISION® GUIDE ............................................................................ 2
DIRECTVISION® CATHETER ....................................................................... 2
WARNINGS & CAUTIONS ................................................................... 3
DOCUMENTATION CONVENTIONS .............................................................. 3
GENERAL WARNINGS .............................................................................. 3
GENERAL CAUTIONS ............................................................................... 5
CATHETERIZATION INSERTION PROCEDURE ....................................... 6
SPECIFIC WARNINGS ............................................................................... 6
SPECIFIC CAUTIONS ................................................................................ 7
SPECIFIC NOTES ..................................................................................... 8
USER & ASSISTANT RESPONSIBILITIES ......................................................... 9
QUICK START INSTRUCTIONS FOR DIRECTVISION® ....................................... 14
Setup DirectVision® System ......................................................... 14
Setup DirectVision® Catheter and Guide ..................................... 14
Catheterize Patient ..................................................................... 15
Post Catheterization Procedure .................................................. 15
DETAIL INSTRUCTIONS FOR DIRECTVISION® ................................................ 16
CATHETER REMOVAL PROCEDURE ................................................... 37
WARNINGS AND CAUTIONS .................................................................... 37
PROCEDURAL STEPS .............................................................................. 37
CLEANING, DISINFECTING & STERILIZING ......................................... 39
INTRODUCTION .................................................................................... 39
GENERAL WARNINGS AND CAUTIONS ....................................................... 39
DIRECTVISION® GUIDE ........................................................................... 40
Specific Warnings ........................................................................ 40
Specific Cautions ......................................................................... 41
DirectVision® Instruction Manual
Page iv
Specific Note ................................................................................ 41
Cleaning Procedure ..................................................................... 42
Disinfecting Procedure ................................................................ 43
Sterilization Recommendations ................................................... 43
Sterilizing Procedure .................................................................... 43
Additional Warnings .................................................................... 44
CAMERA HANDPIECE ............................................................................. 44
Specific Warning .......................................................................... 44
Specific Cautions.......................................................................... 45
Cleaning Procedure ..................................................................... 45
Disinfecting Procedure ................................................................ 46
Sterilizing Procedure .................................................................... 47
CAMERA CONTROL UNIT ........................................................................ 47
Specific Cautions.......................................................................... 48
Cleaning Procedure ..................................................................... 48
Disinfecting Procedure ................................................................ 48
Sterilizing Procedure .................................................................... 48
DISPLAY MONITOR ............................................................................... 49
Specific Cautions.......................................................................... 49
Cleaning Plastic Enclosure Procedure .......................................... 50
Cleaning Color LCD Display Procedure ........................................ 50
Disinfecting Procedure ................................................................ 50
Sterilizing Procedure .................................................................... 50
CART .................................................................................................. 51
Specific Caution ........................................................................... 51
Cleaning Procedure ..................................................................... 51
Disinfecting Procedure ................................................................ 52
Sterilizing Procedure .................................................................... 52
DEFINITIONS .................................................................................... 52
REFERENCES .................................................................................... 53
DirectVision® Instruction Manual
Page 1 of 53
General Guidance
Introduction Thank you for purchasing the PercuVision® DirectVision® System. This manual is intended to serve as a guide for assembling, preparing components for use, operating and handling, maintaining, cleaning, disinfecting and sterilizing the DirectVision® System and reusables. All of these guidances are explained in this manual. It is imperative that you read and understand the contents of this manual prior to use. Keep this instruction guide for reference.
Indications for Use The DirectVision® Catheter is indicated for use in the drainage and/or collection and/or measurement of urine. It is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as the nephrostomy tract.
The DirectVision® Guide is indicated for use in the visual examination of the urethra and bladder when placing the DirectVision® Catheter.
Warning: Federal (USA) Law restricts this device to use by or on the order of a physician.
!
DirectVision® Instruction Manual
Page 2 of 53
Product Overview
DirectVision® System
Compact camera with integrated LED light source
Custom designed optics
Control features for image optimization
Cart with LCD color monitor
DirectVision® Guide Endoscopic quality image
Fiber-optic bundle to transport light and images
1.5mm flexible design for easy insertion into the
DirectVision® Catheter
Quick, easy and secure connection to the camera and
catheter
DirectVision® Catheter 3-way / tri-lumen Foley design: vision, voiding and balloon
inflation
Irrigation enabled
100% silicone
1cm insertion markers
Multiple French sizes for various voiding capacities
10cc balloon
DirectVision® Instruction Manual
Page 3 of 53
Warnings & Cautions
Documentation Conventions Please read this manual and follow its instructions carefully. The words: Warning, Caution, and Note carry special meaning and should be reviewed carefully:
Warning: The personal safety of the patient or user may be involved. Disregarding this information could result in injury to the patient or user.
Caution: Special service procedures or precautions must be followed to avoid damage to the equipment.
Note: Special information to make maintenance easier and important information more clear.
An exclamation point within a triangle alerts the user of risk of danger; attention, consult company documents.
General Warnings It is important to read and understand this instruction manual. Failure to follow these instructions may cause damage to the product or harm to the patient and/or the user. The user should do the following:
1. Warning: Read this instruction manual thoroughly before
assembling and using the product.
2. Warning: Be a qualified user and properly trained in
DirectVision® catheterization placement procedure.
3. Warning: Federal (USA) Law restricts this device to use by
or on the order of a physician.
4. Warning: Keep the distal tip of DirectVision® Guide and DirectVision® Catheter in your field of vision at all times during the procedure (Create & maintain the sterile field).
!
DirectVision® Instruction Manual
Page 4 of 53
5. Warning: Read and understand the care, cleaning, disinfecting, and sterilizing instructions in this manual. Failure to follow these instructions may cause damage to the product or harm to the patient and/or user.
6. Warning: To minimize the risk of patient infection, the DirectVision® Guide must be cleaned and sterilized prior to the first use and every subsequent use. See the Cleaning, Disinfecting & Sterilizing Section in this manual for details.
7. Warning: Do not use this product until you have sterilized the necessary components.
8. Warning: Never look directly at the light emitted from the light source of the camera handpiece or the DirectVision® Guide. Eye damage may result.
9. Warning: Never leave the DirectVision® System powered and unattended. Heat from the light source could ignite drapes or other materials.
10. Warning: Do not store or use the DirectVision® System in an environment that contains flammable gases which can result in an explosion.
11. Warning: The DirectVision® System must only be repaired by personnel authorized by PercuVision®. Unauthorized repairs could compromise the safety and integrity of the product and voids the warranty.
Failure to follow the instructions in this manual could void the warranty.
PercuVision® reserves the right to improve our product. As a result, product may not conform in detail to published design or specifications. Improvements are subject to change without notice. Please contact PercuVision® customer care department for information on improvements and new products.
DirectVision® Instruction Manual
Page 5 of 53
General Cautions 1. Caution: Carefully unpack the shipment and check for
physical damage. If damage is observed or suspected, please refer to the Warranty, Service and Claims Section in DirectVision® Technical Reference Manual (PL000072) for resolution.
2. Caution: Do not use the DirectVision® System until you have verified that it is properly assembled, setup and functioning correctly.
3. Caution: Use only in conjunction with other Type BF or CF rated medical electrical equipment approved to IEC 60601-1 or equivalent standard.
4. Caution: The DirectVision® System has been designed and tested to comply with IEC 60601-1(ed. 3.1)including requirements for electromagnetic compatibility with other devices. Make sure that all devices operated in the vicinity comply with the relevant EMC requirements. Radio systems, mobile telephones, etc. are possible sources of interference as they may emit higher levels of electromagnetic energy.
5. Caution: Never attempt repairs or maintenance to the DirectVision® System before reviewing the Warranty, Service and Claims Section of DirectVision®Technical Reference Manual (PL000072).
6. Caution: To prevent damage to the DirectVision® Guide and camera handpiece, always handle with care. Do not drop, crush, cut, or bend any portion of the visual guide beyond a 2” radius. Although the visual guide is flexible, bending beyond the specified radius could cause breakage in the fiber optic bundle, which reduces its useful life. Kinking the visual guide could break the entire fiber optic bundle which can render it non-operational.
7. Caution: Do not expose the DirectVision® System to excessive heat which can cause damage.
DirectVision® Instruction Manual
Page 6 of 53
8. Caution: Never place heavy objects on the DirectVision® System.
9. Caution: The DirectVision® System and accessories must be disposed of according to local regulations, ordinances and hospital practices.
Catheterization Insertion Procedure
Specific Warnings 1. Warning: A user should stop catheterization upon observing
pelvic trauma. The system is not recommended if in the opinion of a qualified physician, use would endanger the patient, e.g. due to the patient’s general condition.
2. Warning: This information is not a substitute for proper training. Use of this product without proper training could result in patient or user injury.
3. Warning: Always remove visual guide before balloon
inflation.
4. Warning: Before any catheterization procedures and
sterilization processes, inspect the DirectVision® System for
damage. Do not use damaged DirectVision® products in a
catheterization procedure.
5. Warning: Always inspect the visual guide before a
procedure. Never use a visual guide that has burs, sharp
edges, cracked or disconnected jacket and/or strain
relief(s), or is kinked.
6. Warning: Never insert the DirectVision® Guide into a
patient without first placing it into a PercuVision® catheter.
7. Warning: Only use a DirectVision® Guide with a
DirectVision® Catheter. Failure to do so can result in patient
injury.
!
!
!
!
!
!
!
DirectVision® Instruction Manual
Page 7 of 53
.
8. Warning: After each use, remove any observable body
fluids and/or mucus or incrustations from the DirectVision®
Guide with 70% isopropyl alcohol, place it in the sterilization
container and send it to the sterilization department for
immediate cleaning and sterilization.
9. Warning: To minimize the risk of patient infection, the
DirectVision® Guide must be sterilized by your facility prior
to the first use and after every subsequent use. See the
Cleaning, Disinfecting & Sterilizing Section of this manual
for sterilization details. Always use a new sterile
DirectVision® Catheter with each catheterization.
10. Warning: Be sure to remove the DirectVision® Guide from
the patient prior to discharge of a defibrillator. Failure to
follow this warning could result in patient injury or damage
to the DirectVision® Guide.
11. Warning: Do not use ointments or petroleum based
lubricants. They will damage the silicone catheter and may
cause balloon rupture.
Specific Cautions 1. Caution: To prevent damage to the DirectVision® System,
always handle the visual element and camera with care. Do not drop, crush, cut, or bend any portion of the visual element beyond a 2” radius. Although the visual guide is flexible, bending beyond the specified radius could cause breakage in the fiber optic bundle, which reduces its useful life. Kinking the visual guide could break the entire fiber optic bundle which can render it non-operational.
!
!
!
!
DirectVision® Instruction Manual
Page 8 of 53
2. Caution: Be sure to either irrigate the visual guide lumen of
the DirectVision® Catheter with sterile irrigation water (not
saline solution) or lubricate the visual guide directly with
surgical lubricant prior to inserting the DirectVision® Guide.
This will lubricate the catheter’s visual guide lumen and
facilitate placement of the visual guide. Failure to do so
can cause damage to the visual guide and/or catheter.
Specific Notes 1. Note: Once the powerbutton for the Monitor is turned ON,
it remains ON even when the power cord is not plugged into a wall receptacle.
2. Note: The DirectVision® catheterization procedure involves two people (i.e. User and assistant).
3. Note: The DirectVision® catheterization procedure requires double gloving or a glove change.
4. Note: PercuVision® does not recommend inflating the balloon prior to catheter insertion.
5. Note: For the balloon volume 10cc = 10ml.
DirectVision® Instruction Manual
Page 9 of 53
User & Assistant Responsibilities
User’s Responsibilities Assistant’s
Responsibilities
Setup DirectVision® System
Lock cart wheels
Plug power plug of cart into wall
receptacle
Connect camera handpiece to
CCU
Turn on CCU & monitor
Place an IV pole next to the
DirectVision® System
Hang irrigation bag on the IV
pole
Setup DirectVision® Catheter and Guide
Receive wrapped DirectVision®
Guide
Unwrap DirectVision® Guide
Prep patient and create a sterile
field
Spike the irrigation bag
Present sterile catheter package
DirectVision® Instruction Manual
Page 10 of 53
Remove DirectVision® Catheter from package
Turn on irrigation
Lubricate the visual guide lumen with sterile irrigant OR Lubricate the visual guide directly with surgical lubricant
Turn off irrigation
User’s Responsibilities Assistant’s
Responsibilities
Connect irrigation tubing to center port of catheter
Lubricate the catheter (with sterile gel) except for the very tip (visual guide opening)
Remove lid of sterilization box which contains sterile
DirectVision® Guide
Retrieve DirectVision® Guide from sterilization box
Present visual guide connector to camera handpiece connector
Present camera handpiece connector to visual guide connector
Connect camera handpiece to visual guide
Lock visual guide to camera handpiece
Insert DirectVision® Guide into catheter
Lock visual guide to catheter
DirectVision® Instruction Manual
Page 11 of 53
Confirm system operation (light at tip of visual guide and image on monitor)
Perform white balance Press white balance button
Perform focus adjustment Adjust focus ring on camera handpiece
Orient the catheter curve tip and the visual guide alignment “dot” to the 12:00 o’clock position (curved tip up)
User’s Responsibilities Assistant’s
Responsibilities
Catheterize Patient
Start irrigation
Insert the catheter Hold connected camera handpiece-visual guide
Stop irrigation
Unlock visual guide from catheter
Remove visual guide from catheter
Present connected visual guide-camera handpiece to assistant
Receive connected visual guide-camera handpiece
Cap visual guide lumen
DirectVision® Instruction Manual
Page 12 of 53
Advanced catheter hub to meatus
Inflate balloon
Disconnect irrigation tubing
Connect urine collection bag & verify urine flow
DirectVision® Instruction Manual
Page 13 of 53
User’s Responsibilities Assistant’s
Responsibilities
Post Catheterization Procedure
Unlock & separate visual guide from camera handpiece
Clean camera handpiece and cable
Place camera handpiece in the camera holder.
Wipe off used visual guide and place in sterilization box
Close lid of sterilization box
Send visual guide to sterilization department
Dispose of waste
Turn off power
Unplug power cord
Clean DirectVision® System
Unlock cart wheels & transport cart to holding area
DirectVision® Instruction Manual
Page 14 of 53
Quick Start Instructions for DirectVision®
Setup DirectVision® System
1. Lock cart wheels and plug in power cord
2. Connect camera handpiece to CCU and turn on CCU & monitor
3. Locate and place an IV pole next to the DirectVision System
4. Hang catheter irrigation solution bag on the IV pole
Setup DirectVision® Catheter and Guide
5. Open visual guide sterile box wrap
6. Prep patient and create a sterile field
7. Connect irrigation tubing to IV bag
8. Remove DirectVision® Catheter from package
9. Instill 1-2ml of sterile irrigant into the visual guide lumen (via irrigation tubing) OR lubricate visual guide directly with surgical lubricant
10. Stop irrigation and connect irrigation tubing to center port of catheter
11. Lubricate the catheter (with sterile gel) except for the very tip (visual guide opening)
12. Connect DirectVision® Guide to camera handpiece and lock into place
13. Insert the DirectVision® Guide into visual guide port of DirectVision® Catheter and lock into place. Focus
14. Inspect the visual guide image for brightness and clarity prior to each use. Discontinue use of any visual guide which has an unacceptable image after troubleshooting
15. Orient the catheter curve tip and the visual guide alignment “dot” to the 12:00 o’clock position (curved tip up)
DirectVision® Instruction Manual
Page 15 of 53
Catheterize Patient
16. Start irrigation and insert the catheter into meatus
17. Advance the catheter while viewing the video image on the monitor
Note: If a DirectVision® Guide fails to provide an adequate visual representation of the urethra to the clinician, the catheterization procedure should be discontinued
18. Steer the catheter by rotating the catheter and manipulating the penis (to keep full image of urethra whenever possible)
19. Advance the catheter until the tip enters the bladder (the video image will go dark or yellow)
20. Stop irrigation. Unlock and remove the DirectVision® Guide and cap the lumen
21. Advance the hub of the catheter to meatus before inflating balloon with no more than 10ml of sterile water
Note: Inflating the balloon should not cause pain to the patient
22. Disconnect irrigation tubing and connect the urine collection bag
Post Catheterization Procedure
23. Unlock and detach the DirectVision® Guide from the camera handpiece
24. Place visual guide in its tray for immediate cleaning and re-sterilization by sterile processing
25. Power down the equipment, remove camera handpiece from CCU, unplug power and unlock wheel casters. Move system to holding location (restock as needed)
26. Properly dispose of waste materials
DirectVision® Instruction Manual
Page 16 of 53
Detail Instructions for DirectVision® 1. Setup the DirectVision® System
Position DirectVision® System in room to perform catheterization procedure near a 110/120 VAC wall socket.
Lock the cart wheels.
Warning: Failure to lock the wheels could result in cart movement during the catheterization procedure. Injury to the patient could result.
Unwind the power cord, and plug it into an approved medical grade 110/120 VAC wall receptacle, keeping the cord away from traffic areas to prevent tripping.
!
DirectVision® Instruction Manual
Page 17 of 53
Connect camera handpiece to the CCU. o Retrieve the camera handpiece from the camera
holder.
o Remove the Plug Cap (metal cover) from Camera Handpiece Plug end.
DirectVision® Instruction Manual
Page 18 of 53
Note: The camera cord may be controlled by threading it through one loop of an IV pole.
o Connect the Camera Handpiece Plug to the camera connector receptacle on the CCU’s front panel. Red alignment dot on camera connector plug should align to the red dot on the CCU receptacle.
Caution: Be careful how you align the Camera Handpiece Plug into the CCU Connector Receptacle. Failure to align pins with the corresponding pin hole could cause damage to the pins.
Turn on the CCU by pressing the power ON button. CCU LED should illuminate.
DirectVision® Instruction Manual
Page 19 of 53
Note: The display monitor’s power switch should be ON. If this is not the case, the display monitor must be turned ON. Turn on the Monitor power by pressing the Power Button on the back of the monitor. The monitor LED Indicator should illuminate GREEN.
Hang an irrigation bag on an IV pole placed next to the DirectVision® System.
2. Prepare DirectVision® Guide
Receive wrapped visual guide.
Verify visual guide was sterilized (color of external sterilization indicator)
Unwrap DirectVision® Guide box using sterile technique.
3. Prep the patient for procedure (create and maintain a sterile field).
Note: PercuVision® recommends double gloving, put on a mask, apply antiseptic such as Betadine to genitalia, instill
Wrapped sterilization tray
Closed sterilization tray
Visual Guide in an open sterilization tray
Preparing the DirectVision® Guide for Use
DirectVision® Instruction Manual
Page 20 of 53
lidocaine gel into urethra, apply a penile clamp, open a fenestrated sterile drape and properly position it on the patient. These steps should be performed in accordance to standard hospital protocol. Details for this step are not presented in this document.
4. Connect irrigation tubing to IV bag.
Assistant connects irrigation tubing to the irrigation bag (i.e. Spike the irrigation bag).
Assistant removes the DirectVision® Catheter by peeling back the outer, unsterile packaging and presents the sterile catheter to user.
DirectVision® Instruction Manual
Page 21 of 53
Remove the catheter from the packaging.
Note: The uncoated catheter does not have inner packaging.
Instill 1-2ml of sterile irrigation water from irrigation
tubing into the visual guide lumen of the DirectVision®
Catheter. Water will lubricate the catheter’s visual
guide lumen and facilitate the passage of visual guide.
The visual guide may also be lubricanted directly using
surgical lubricant.
DirectVision® Instruction Manual
Page 22 of 53
Stop irrigation. Connect irrigation tube to the irrigation/drainage lumen of the DirectVision® catheter.
5. Lubricate the catheter shaft with sterile gel.
Note: Lubricate about 8 inches (20 cm) of the catheter
shaft with a lubricant approved by your healthcare facility
for 100% silicone catheters. Do not lubricate the first 0.25
inches of the tip in order to avoid obscuring the visual guide
tip.
6. Connect DirectVision® Guide to the camera handpiece.
Remove lid of sterilization box which contains sterile DirectVision® Guide
Retrieve Directvision® Guide from Sterilization box
Assistant removes the camera handpiece from the camera handpiece holder on the cart's upper column
DirectVision® Instruction Manual
Page 23 of 53
Assistant removes Protection Cap from camera handpiece and places it on the top shelf of cart.
Align the DirectVision® Guide with the camera handpiece. Open the lock. Connect the visual guide with camera handpiece making sure that the lock remains open.
DirectVision® Instruction Manual
Page 24 of 53
Assistant closes the lock on camera handpiece once the visual guide and camera handpiece are engaged.
The camera handpiece and visual guide are now securely coupled and ready for use.
7. Confirm System Operation.
Turn the illumination knob to maximum brightness.
Confirm that the tip of the visual guide is illuminated and an image appears on the monitor. If the tip of the visual guide is not illuminated or an image does not appear on the monitor, go to the Troubleshooting Section of the DirectVision® Technical Reference Manual (PL000072) to resolve the problem.
DirectVision® Instruction Manual
Page 25 of 53
8. Insert DirectVision® Guide into the visual guide lumen of DirectVision® Catheter.
Warning: The DirectVision® Guide must only be inserted
into the visual guide lumen of the DirectVision® Catheter.
Grasp the catheter and if the visual guide lumen cap is affixed, remove it to be able to insert the DirectVision® Guide.
Insert the DirectVision® Guide into visual guide lumen of the DirectVision® Catheter. Line-up the visual guide with the guide port.
Caution: When inserting the DirectVision® Guide into
catheter’s visual guide lumen, advance slowly and carefully,
rotating the silicone visual guide port. If the visual guide
encounters resistance, withdraw the visual guide about an
inch then resume insertion. Straighten catheter tip as
needed either by resting the catheter on the sterile field or
by manually straightening the tip.
Caution: The visual guide can be damaged by kinking the
polyimide tubing when inserting the visual guide into the
catheter’s visual guide lumen. While holding the catheter
manifold with one hand, always apply the insertion force
with your other hand less than 1 inch from the catheter’s
visual guide lumen.
Silicone visual guide port
DirectVision® Instruction Manual
Page 26 of 53
Note: Use the image on the monitor display to thread the
visual guide through the retaining ring located about 1.75
inches from the end of the catheter tip.
Continue inserting the DirectVision® Guide into the visual guide lumen until the end of visual guide touches the luer-lock on the end of the manifold. The tip of the DirectVision® Guide should be positioned at the tip of the catheter.
Lock the DirectVision® Guide in place by holding the
catheter luer-lock stable and twisting the metal fitting
of the visual guide into catheter's plastic luer-lock.
.
DirectVision® Instruction Manual
Page 27 of 53
Inspect the visual guide image for brightness and clarity
prior to each use. Discontinue use of any visual guide
which has an unacceptable image.
Warning: Inspect the tip of the catheter once the visual guide is inserted into the catheter and locked into place. Make sure it is not protruding 1mm or more beyond the catheter tip.
Confirm that the tip of the visual guide is illuminated and an image appears on the monitor. If the tip of the visual guide is not illuminated or an image does not appear on the monitor, go to the Troubleshooting Section of the DirectVision® Technical Reference Manual (PL000072) to resolve the problem.
Aim the illuminated tip of the visual guide to within approximately 5mm of a white object. If the white light requires color correction, perform White Balance Adjustment summarized in the Camera Control Unit Calibration Procedures Section of DirectVision® Technical Reference Manual (PL000072).
!
DirectVision® Instruction Manual
Page 28 of 53
Confirm image quality and camera orientation. If the focus is not sharp, perform Focus summarized in the Camera Control Unit Calibration Procedures Section of DirectVision® Technical Reference Manual.
9. Warning: Do not use the DirectVision® System if the video image is not displayed on the monitor or if the video image is not sharp.
10. Orient the catheter tip to the 12:00 o’clock position.
Note: The alignment “dot” on the catheter manifold (irrigation/drainage port) is on the same plane as the angled catheter tip. When the catheter tip is positioned to the 12:00 o’clock position, the catheter tip is curved up and the “dot” on the catheter manifold is also in the 12:00 o’clock position.
DirectVision® Instruction Manual
Page 29 of 53
Catheter Manifold
Tip of catheter is curved up
Catheter dotaligned to 12 O’clock
DirectVision® Instruction Manual
Page 30 of 53
11. Catheterize the patient.
Turn on irrigation water and verify flow.
Carefully insert the catheter tip into the patient’s meatus. Observe the image of the urethra on the display monitor.
Note: When initially inserting the catheter, the “dot” on the visual guide luer fitting and the “dot” on catheter irrigation/drainage port should both be aligned to 12:00 o’clock position. This alignment orientation ensures the tip of the catheter is pointing up and orientates the relative position of the image displayed on the monitor to the 12:00 o’clock position.
Note: If a DirectVision® Guide fails to provide an adequate visual representation of the urethra to the clinician, the catheterization procedure use should be discontinued.
Steer the catheter by manipulating the penis or rotating the catheter. Advance the catheter carefully while viewing the video image on the monitor. As you advance, try to maintain a “bull’s-eye” view of the urethra.
DirectVision® Instruction Manual
Page 31 of 53
View in sequence 5 basic urethral landmarks:
o Upward angulation of Bulbous urethra o External sphincter o 6:00 o’clock verumontanum o Upward angulation of prostate o Bladder opening
Continue advancing the catheter until the tip enters the bladder (the video image will go dark or yellow).
Stop irrigation.
12. Remove visual guide from catheter.
Unlock the DirectVision® Guide by rotating the luer-lock on the catheter clockwise until it becomes loose. Detach from the catheter.
Warning: Once the luer-lock is unlocked and while the catheter is in the patient, do not advance or reinsert the DirectVision® Guide in the visual guide lumen.
While holding on to the visual guide’s connector, gently pull the visual guide out of the catheter, about one inch, while holding the manifold with other hand to prevent catheter removal.
Change hand position from grasping the visual guide connector to grasping the one inch of exposed visual guide shaft.
!
DirectVision® Instruction Manual
Page 32 of 53
Continue removing the visual guide by grasping the exposed visual guide shaft and gently pulling until the entire shaft is removed from the catheter.
Affix the visual guide lumen cap to the open visual guide lumen.
Pass the visual guide attached to the camera to the assistant.
13. Inflate the balloon.
Continue advancing the catheter until the catheter hub reaches the meatus.
Verify that the catheter hub is touching patient’s meatus.
Insert 10ml syringe with sterile water into balloon port.
DirectVision® Instruction Manual
Page 33 of 53
Inflate balloon with 10ml sterile water.
Warning: Only use sterile water to inflate balloon. Do not exceed 10ml of liquid volume for balloon inflation.
Remove the syringe from the balloon port. Dispose of
the empty syringe.
Gently pull catheter until inflation balloon is snug against bladder neck.
14. Disconnect irrigation tubing.
Disconnect irrigation tubing from the catheter.
Connect urine collection bag and verify urine flow.
15. Place camera handpiece into camera holder.
Assistant unlocks the camera handpiece and gently pulls the camera handpiece apart from the visual guide.
DirectVision® Instruction Manual
Page 34 of 53
Assistant retrieves the Protection Cap from top shelf of cart, places it on the front of the camera handpiece and locks into place.
DirectVision® Instruction Manual
Page 35 of 53
Assistant attaches Plug Cap to Camera Handpiece Plug.
Clean the camera handpiece as described in the Cleaning, Disinfecting & Sterilizing Procedures Section of this manual.
Place camera handpiece into camera holder.
16. Send visual guide to sterilization department.
Clean the visual guide as described in the Cleaning, Disinfecting & Sterilizing Procedures Section of this manual.
Place the DirectVision® Guide into an approved sterilization tray. See Table 1 in the Cleaning, Disinfecting & Sterilizing Procedures Section for list of recommended sterilization trays.
DirectVision® Instruction Manual
Page 36 of 53
Send the DirectVision® Guide in its sterilization tray to
sterilization services for immediate cleaning and
sterilization.
17. Return the DirectVision® System to its holding area.
Remove and properly dispose of waste.
Press OFF Button on CCU.
Note: The display monitor is not turned off.
Unplug cart’s power plug from 110/120 VAC wall receptacle. Wind power cord around hooks on the neck of the cart.
Clean CCU, monitor and cart as described in the Cleaning, Disinfecting & Sterilizing Section of this manual.
Unlock cart wheels and transport cart to its holding area.
DirectVision® Instruction Manual
Page 37 of 53
Catheter Removal Procedure
Warnings and Cautions Warning: Should balloon rupture occur, care should be taken to assure that all balloon fragments are removed from the patient.
Warning: Do not resterilize the DirectVision® Catheter.
Note: A catheter that is removed from a patient may be a potential biohazard. Handle and dispose of the catheter in accordance with accepted medical practice and applicable laws and regulations.
Procedural Steps 1. Setup for procedure.
Put on clean gloves in accordance to hospital protocol.
Retrieve 10cc syringe.
2. Deflate balloon.
Gently insert a syringe in the catheter valve. Never use more force than is required to make the syringe “stick” in the valve.
Allow the pressure within the balloon to force the plunger back and fill the syringe with water.
If you notice slow or no balloon deflation, gently reseat the syringe. Use only gentle aspiration to encourage deflation as needed. Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.
If the balloon does not deflate, resolve the problem by following hospital procedures and/or calling a physician.
Remove all 10ml sterile water from inflated balloon.
Disconnect syringe from balloon port.
3. Withdraw the catheter slowly from patient. If the catheter does not gently withdraw from the patient, follow hospital guidelines and/or notify a physician.
DirectVision® Instruction Manual
Page 38 of 53
4. Dispose of the catheter and urine bag following facility guidelines.
DirectVision® Instruction Manual
Page 39 of 53
Cleaning, Disinfecting & Sterilizing
Introduction Medical equipment devices are essential for the care of patients. Some are typically designed for reuse, but can transmit pathogens if any steps in reprocessing, cleaning, disinfecting, or sterilizing fail. The vast majority of pathogens are present in organic matter, e.g. visible soil. The first step in reprocessing is cleaning. It is the most important. Failure to remove soil creates the potential to transmit infection and can compromise the efficacy of subsequent disinfection or sterilization. Decontamination is the process in which microorganisms are removed or destroyed in order to render an object safe. Decontamination includes:
• Cleaning
• Disinfecting
• Sterilizing
General Warnings and Cautions 1. Warning: Read these instructions thoroughly before
cleaning, disinfecting and/or sterilizing the product. Failure to follow these instructions may cause damage to the product or harm to the patient and/or user.
2. Caution: Ultrasonic cleaning should not be used to clean any part of the DirectVision® System.
3. Caution: The cleaning, disinfecting and sterilizing methods presented in this document are validated. Any deviation is not validated and is not recommended.
DirectVision® Instruction Manual
Page 40 of 53
DirectVision® Guide
Table 1: Recommended Cleaning and Sterilizing Conditions for the DirectVision® Guide
Specific Warnings
1. Warning: Always inspect the DirectVision® Guide before performing a cleaning and/or sterilization procedure. Reject a visual guide that has burs, sharp edges, cracked or disconnected jacket and/or strain relief(s), or is kinked.
Catheter Hub Lock
Camera Connector
Internal Stop
Camera Channel
LightChannel
Sterilization System or
Method
Cycle(s)/Paramenters Tray or Pouch
ASP Sterrad 100s Full Cycle
APTIMAX®
Instrument Tray
(13835) or equivalent
ASP Sterrad 200 Full Cycle
APTIMAX®
Instrument Tray
(13835) or equivalent
ASP Sterrad 100 NX Duo Cycle
APTIMAX®
Instrument Tray
(13835) or equivalent
Steris Amsco V Pro-1 Lumen CycleSteris Tray (VP0003)
or equivalent
Lumen Cycle
Non-Lumen Cycle
Lumen Cycle
Non-Lumen Cycle
Flexible Cycle
Steris System 1E Standard Operating CycleStaris C1140E, C1200E,
or C1220E
Staris Tray (VP0003)
or equivalent
Sterilization EnvironmentDevice Name Cleaning
Solutions
Disinfecting
Solution
DirectVision
® GuideEnzol™ ---
Steris Amsco V Pro-1PlusSteris Tray (VP0003)
or equivalent
Steris Amsco V-Pro maX
DirectVision® Instruction Manual
Page 41 of 53
2. Warning: Only clean or sterilize the DirectVision® Guide as specified in this manual. Failure to follow the instructions provided in this manual could result in infection or damage to the DirectVision® Guide.
Specific Cautions
1. Caution: To prevent damage to the DirectVision® Guide always handle with care. Do not drop, crush, cut, or bend any portion of the visual guide beyond a 2” radius.
2. Caution: Machine washing is not recommended. It increases likelihood of damage to the DirectVision® Guide.
3. Caution: When cleaning the DirectVision® Guide, use a soft sponge and/or soft bristle brush to clean surfaces and crevices.
4. Caution: Inspect the DirectVision® Guide for damage following each use.
5. Caution: The DirectVision® Guide should never be exposed to temperatures greater than 55o C (131o F). It can cause damage.
6. Caution: Never place heavy objects on the DirectVision® System.
7. Caution: The DirectVision® Guide must be disposed of according to local regulations, ordinances and hospital practices.
Specific Note
1. Note: The cleaning and sterilization procedure can affect performance of the DirectVision® Guide. If a DirectVision® Guide ceases to provide adequate illumination or a clear image of the urethra, discontinue use of that DirectVision® Guide.
DirectVision® Instruction Manual
Page 42 of 53
Cleaning Procedure
1. Prior to cleaning, disassemble the DirectVision® Guide from camera handpiece.
2. Prepare an enzymatic detergent, such as Enzol™ (1 oz detergent/gallon water), using warm tap water and following the manufacturers instructions.
3. Note: The enzyme solution should be changed when it becomes grossly contaminated (bloody and/or turbid).
4. Immerse the DirectVision® Guide in an enzymatic solution for at least one minute.
5. Soak a clean soft cloth in the detergent and wipe entire DirectVision® Guide to remove any visible evidence of soil or debris.
6. Using a soft bristled brush (such as M-16 BRUSH-DBL END),
scrub the DirectVision® Guide for a minimum of 30 seconds,
paying particular attention to the visual guide coupler’s two
lenses.
7. Use only a soft cloth to clean the tip of the DirectVision® Guide lens.
8. Remove the DirectVision® Guide from the detergent and wipe with a clean soft cloth that is dipped in ambient sterile water to remove any evidence of detergent.
9. Dry the DirectVision® Guide with a clean soft cloth. Use filtered pressurized air (no more than 40psi) to dry crevices and hard to reach areas.
10. Visually examine each DirectVision® Guide to ensure cleanliness.
DirectVision® Instruction Manual
Page 43 of 53
Disinfecting Procedure
Not required.
Sterilization Recommendations
During reprocessing ensure the following:
1. Comply with labeling and instructions for use for the STERRAD®, Steris Amsco® V-PRO®, or Steris Systems 1E system and parameters (listed in Table 1).
2. Use only the recommended cycles listed in Table 1 for
respective sterilization systems listed in Table 1.
3. Use only recommended trays listed in Table 1 for respective
sterilization systems listed in Table 1.
Sterilizing Procedure
1. The DirectVision® Guide is sterilized with vapor hydrogen peroxide sterilization. See Table 1 for recommended sterilization systems and conditions.
2. Warning: To ensure proper sterilization of the DirectVision® Guide, follow the sterilization procedure provided by the manufacturer for the respective system listed in Table 1. Failure to follow the sterilization procedure defined by the manufacturer could result in infection or damage to the DirectVision® Guide.
3. Warning: Use of sterilization systems or conditions not listed in Table 1 may result in shortened useful life for the DirectVision® Guide.
4. Warning: Always sterilize the DirectVision® Guide in a tray/container listed in Table 1. Failure to use the correct container could result in patient infection or damage to the DirectVision® Guide.
DirectVision® Instruction Manual
Page 44 of 53
Additional Warnings
1. Warning: To minimize the risk of patient infection, the
DirectVision® Guide must be sterilized prior to the first use
and after every subsequent use.
2. Warning: Clean the DirectVision® Guide immediately after
use to prevent incrustation. Failing to do so could
compromise future sterility and possibly cause damage to
the visual guide.
3. Warning: During cleaning, immerse the DirectVision®
Guide in an enzymatic solution for at least 1 minute while
cleaning or patient infection may result.
Camera Handpiece
Table 2: Recommended cleaning and disinfecting conditions for the camera handpiece
Specific Warning
1. Warning: To minimize the risk of bioburden contamination,
take caution to not contaminate the camera handpiece
LED Output
Camera Input
Camera Handpiece
Focus
Sterilization System Cycle(s) Tray
Camera
HandpieceEnzol® CIDEX® OPA --- --- ---
Component
Name
Cleaning
Solution
Disinfecting
Solution
Sterilization Environment
DirectVision® Instruction Manual
Page 45 of 53
after the sterile drape (if applicable) has been removed or
before processing.
Specific Cautions
1. Caution: The Plug Cap, which may be wired to the Camera
Handpiece Plug , should be locked into place during cleaning
and disinfecting procedures. Failing to cover the Camera
Handpiece Plug while soaking will damage the wiring.
2. Caution: Machine washing is not recommended. It increases likelihood of damage to the camera.
3. Caution: The camera handpiece shall not be cleaned in an
ultrasonic bath.
4. Caution: Ensure that detergent and disinfectant are
compatible with each other. Avoid constantly changing the
selected detergent-disinfectants.
Cleaning Procedure
Cleaning the camera handpiece is necessary for hygienic reasons. It can be done manually or by soaking.
Use only mild cleaning substances for dissolving the impurities. These cleaning agents must have been approved by PercuVision® for such use. While using the different substances, pay attention to the details provided by PercuVision® with respect to concentration and soaking time.
Plug Cap
Camera Handpiece Plug
DirectVision® Instruction Manual
Page 46 of 53
1. Prepare the cleaning bath solution using the Enzol™
enzymatic solution.
2. Lock Plug Cap on Camera Handpiece Plug to protect cable
from damage.
3. Remove the Protection Cap from the camera handpiece if it
is locked to the camera handpiece. This will expose the
protective optical surfaces of the illumination and camera
ports for cleaning.
4. For mechanical cleaning, use soft cloth, cotton swabs and
special brushes. For channels and cavities, it is especially
advisable to use suitable brushes.
5. Clean dirt off the optical surface with a cotton swab soaked
in 70% isopropyl alcohol or neutral cleanser.
6. After cleaning, rinse all parts with distilled water. Ensure
that all residues of cleaning agents have been washed away.
7. Dry the camera with a tissue or soft absorbent cloth.
Disinfecting Procedure
The necessity for disinfection of the camera handpiece is to be determined by the user based upon facility policy for equipment.
1. Prepare the disinfecting bath solution using CIDEX® OPA.
Note: In order to ensure the entire camera is properly disinfected, all surfaces must be exposed to CIDEX®.
Protection Cap
Camera Handpiece
Lock OPEN
DirectVision® Instruction Manual
Page 47 of 53
2. Completely immerse the camera handpiece in the disinfecting solution.
3. Soak in bath according to bath manufacturer’s instructions.
4. After soaking, thoroughly rinse camera handpiece with sterile water.
5. Dry camera handpiece appropriately, with compressed air, for its next use. Ensure lens is free of residue.
6. Lock the Protection Cap to the Camera Handpiece if the Protective Cap is available.
Sterilizing Procedure
1. Caution: The camera handpiece is never sterilized.
Camera Control Unit
Table 3: Recommended cleaning condition for the Camera Control Unit (CCU)
Protection Cap
Camera Handpiece
Lock OPEN
Sterilization System Cycle(s) Tray
CCUIsopropyl
Alcohol (IPA)--- --- --- ---
Component
Name
Cleaning
Solution
Disinfecting
Solution
Sterilization Environment
DirectVision® Instruction Manual
Page 48 of 53
Specific Cautions
1. Caution: Do not immerse the CCU into liquids.
2. Caution: Machine washing is not recommended. It increases likelihood of damage to the CCU.
2. Caution: Do not use scouring agents or solvents since they could damage the paint or labels on the CCU housing.
3. Caution: Do not use wet sponges or clothes since this could lead to liquid reaching the electrical parts which could cause damage.
4. Caution: Only clean front bezel of the CCU.
5. Caution: Do not clean the back of CCU which has video and power connections. There is danger of electric shock.
Cleaning Procedure
1. The CCU front panel can be cleaned by wiping with a moist cloth.
2. Prepare a pH neutral cleanser or isopropyl alcohol
3. All exterior surfaces of the CCU can be wiped by a soft cloth which has been moistened with water and soft soap suds or isopropyl alcohol. Difficult to remove stains can be removed using a soft cleaner of ammoniac basis.
Disinfecting Procedure
1. Caution: The CCU is never disinfected.
Sterilizing Procedure
1. Caution: The CCU is never sterilized.
DirectVision® Instruction Manual
Page 49 of 53
Display Monitor
Table 4: Recommended cleaning conditions for the MonitorEnclosure and the LCD Display Panel components of the Display Monitor
Specific Cautions
1. Caution: Do not use thinner, benzene, alcohol or such on the monitorenclosure. These solvents can damage the enclosure.
2. Caution: Do not use organic solvents such as acetone and toluene to clean the color LCD panel. These solvents can damage the LCD panel.
3. Caution: Do not apply insecticides and other volatile items to the monitor enclosure. Do not leave rubber and vinyl products or such in contact with the enclosure for long hours. This can damage the enclosure and cause paint to peel.
4. Caution: Treat the color LCD panel with care. Do not rub the
Sterilization System Cycle(s) Tray
Plastic
(Enclosure)pH=7 Solution --- --- --- ---
LCD Display
Panel
Isopropyl
Alcohol (IPA)--- --- --- ---
Component
Name
Cleaning
Solution
Disinfecting
Solution
Sterilization Environment
DirectVision® Instruction Manual
Page 50 of 53
LCD panel surface with a rough item or hit it on the surface. Do not strongly press the LCD panel surface. This can lead to unevenness in the screen and also to failure of the product.
Cleaning Monitor Enclosure Procedure
1. Remove the power plug from the monitor and AC wall receptacle for security.
2. Lightly wipe off dirt on the monitor enclosure with a hard-
wrung soft cloth soaked in a neutral cleaning solution.
3. Allow wet surfaces to completely dry before reconnecting power plugs.
Cleaning Color LCD Display Procedure
Recommended cleaners for the color LCD panel are: isopropyl alcohol (without abrasive), non-ammonic glass cleaner and watered-down neutral cleaning solution.
1. Remove the power plug from the monitor and AC wall
receptacle for security.
2. When the screen has dust on the color LCD panel surface,
wipe it off with soft moist cloth.
3. Lightly wipe-off dirt on the monitor enclosure with a hard-
wrung soft cloth soaked in a neutral cleaning solution.
4. Allow wet surfaces to completely dry before reconnecting power plugs.
Disinfecting Procedure
1. Caution: The monitor is never disinfected.
Sterilizing Procedure
1. Caution: The monitor is never sterilized.
DirectVision® Instruction Manual
Page 51 of 53
Cart
Table 5: Recommended cleaning condition for the Cart
Specific Caution
1. Caution: Do NOT use the following chemicals to clean your cart: acetone, mineral spirits, abrasive cleaners, paint thinner or any other harsh or toxic chemicals.
Cleaning Procedure
1. Verify that your cart is unplugged from the wall outlet before cleaning.
2. Use a soft, clean cloth to clean the cart.
3. Use a cleaner of 80% denatured alcohol and 20% water.
4. Allow the cart to dry completely before plugging the power plug into a wall outlet.
Sterilization System Cycle(s) Tray
CartIsopropyl
Alcohol (IPA)--- --- --- ---
Component
Name
Cleaning
Solution
Disinfecting
Solution
Sterilization Environment
DirectVision® Instruction Manual
Page 52 of 53
Disinfecting Procedure
1. Caution: The Cart is never disinfected.
Sterilizing Procedure
1. Caution: The Cart is never sterilized.
Definitions cc: Cubic Centimeters (volume, i.e. cm3; equivalence 1cc = 1ml )
CCU: Camera Control Unit
Cleaning: Cleaning is a process, usually involving detergent or enzymatic presoak, that removes foreign material (e.g. dirt, mucus, blood stains or microorganisms) from an object. Cleaning is the most essential step in reprocessing medical instruments and equipment.
Disinfecting: Any process, chemical or physical, that destroys pathogens such that an item is safe to handle for its intended use.
Disinfectant: A disinfectant is a chemical agent that destroys most pathogens but may not kill bacterial spores. Chemical disinfection should only be used if heat treatment is impractical or if it may cause damage to the equipment. There is a broad spectrum of chemical disinfectants that have different antimicrobial activities. Most of them do not necessarily kill all microorganisms or spores that are present on an inanimate object but instead reduce the number of microorganisms to a level that is not harmful to health. Disinfectants are used on inanimate objects only and not on living tissue. Chemicals used to kill microorganisms on skin or living tissue are known as antiseptics.
IPA: Isopropyl Alcohol
ml: milliliters (volume; equivalence 1ml = 1cc )
mm: millimeter
Sterilant: An agent that destroys all viable forms of microbial life to achieve sterilization.
DirectVision® Instruction Manual
Page 53 of 53
Sterilization: Sterilization methods remove or destroy all forms of microbial life, including bacterial spores, by either physical or chemical processes. It is recommended that any instrument or equipment that comes in contact with the blood stream or with subdermal tissues is cleaned and sterilized in between each use. Sterilization is accomplished principally by steam under pressure, by dry heat, or by chemical sterilants. The choice of the method for sterilization depends on a number of factors including the type of material to be sterilized is made of, the number and type of microorganisms involved, the classification of the item, and availability of sterilization methods.
References DirectVision®Technical Reference Manual, PL000072.
PercuVision® LLC
2030 Dividend Drive
Columbus, Ohio 43228
(877) 913-6333
www.percuvision.com
PL000070 Rev 01