design and fabrication of a fiber optic …web.mit.edu/vkarthik/www/old-site/btp-report.pdfs....
TRANSCRIPT
DESIGN AND FABRICATION OF A FIBER OPTIC PROBE FOR FLOURESCENCE MEASUREMENT
B.TECH PROJECT-2006
V.KARTHIK (2002055)
&
S. SANGAMESHWAR RAO (2002335)
ENGINEERING PHYSICS
SUPERVISORS
PROF. B.P.PAL
&
DR.M.R.SHENOY
DEPARTMENT OF PHYSICS INDIAN INSTITUTE OF TECHNOLOGY DELHI
Chapter 1 1.1 Introduction
In numerous industrial situations, online remote monitoring becomes imperative because
of the hazardous nature of the environment. Chemical plants dealing with toxic materials,
high radiation environment in a nuclear plant, high hydraulic pressure region of the oil
fields, etc are some of the various examples where unmanned monitoring devices will be
of great help. Optical fiber probes, often called optrodes, have caught attention recently
for such applications. Coupled with laser light, one can realize very efficient fiber
instruments for probing purposes. There exist many other methods where optrodes can be
used- fluorescent based temperature sensors [1], in clinical applications such as
fluorescence spectroscopy of tissues [2-6], endoscopy [2], fluorescence based oxygen
sensing [7] and potassium monitoring in blood [8],
Fiber optic probes can be used for remote collection of fluorescence, scattering or Raman
signals from a sample. A great deal work has been done in designing the fiber optic
probes for such applications especially for clinical applications. They can be used for
optical spectroscopy of tissues, as flexible catheters with outer diameter less than 0.5 mm
[2]. While the diameter of the probe has to be very small for such clinical applications, it
is not a stringent requirement in the case of probes for other applications. With the basic
features from these clinical optrodes and incorporating further changes with no limitation
in size, very efficient optrodes can be developed.
Good simulation models for propagation of light in tissues also exist in the literature [9-
11]. Further, some of the basic features in these models can be used for developing
simulation models for light propagation in turbid medium. These simulation models can
be used in designing optimum configuration of probe.
The kind of optrodes that we have tried to fabricate can be used to collect fluorescence
signal from an irradiated turbid sample placed in a cell. Using a Microscope Objective,
laser light is coupled into a fiber. The outlet of this fiber illuminates the sample (in our
case the turbid medium) at the other end. A bundle of fibers, called the optrode, collects
the scattered and the fluorescence light. The collected signal can be analyzed using
spectroscopic methods. This data is very useful in obtaining important information about
the sample studied.
1.2 An overview of the project:
The aim of the project was to design and fabricate an optimum probe for maximum
collection of fluorescence from an irradiated liquid. To design an optimum configuration,
good simulation models are required. So we started our project with developing Monte-
Carlo simulation code for light propagation in turbid medium. Very soon we found out
that to compare results of simulation with experiment, we need to input the values of the
optical parameters of the turbid medium in the simulation. It was found that the existing
methods were difficult to implement and so we had developed a new method to estimate
these parameters. This was the most time consuming part of the project.
Our next task was to study different configurations for maximum collection efficiency.
Some of the results from simulation were used in the initial design aspect. The optrode
was further optimized by minimizing noise from the input signal. Finally, we have
fabricated a fiber probe as per the optimum design.
1.3 Organization of the thesis:
Chapter 2 includes the Monte-Carlo techniques to study the propagation of photons in
turbid medium under different conditions like absorption, scattering and fluorescence.
Chapter 3 gives a brief overview of different experimental arrangements that were used
in estimating the optical parameters. An introduction to different methods in estimating
the optical parameters and a novel method developed by us are explained is this chapter.
The design issues considered in developing the optrode and some of the designs are
explained in Chapter 5. Finally, the experimental results for fluorescence, conclusion and
scope for future study are discussed in Chapter 6.
1.4 Important concepts and definitions
Optrode: The optical fiber device, which is used to probe or sense a signal based on light
detection is called as optrode.
Fig: A simple optrode used for spectroscopy. The central fiber is used for illuminating
the sample and the outer fibers are used for collecting the fluorescent light.
Turbid medium: It is a liquid medium containing lot of suspended particles which act as
scattering centers. When light is made to pass though such medium, there will be random
scattering of light due to these suspended particles.
Some of the naturally found turbid mediums are fog, clouds and milk
Light on interaction with medium-Different mechanisms
1. Absorption – The atoms of the medium absorb the incident photons having
certain energy. In such a case, the atoms absorb and re-emit the absorbed photon.
As known the emission process is an isotropic process and hence the emitted
photons move in all directions with equal probability.
Absorption coefficient (µa): It is defined as the probability of photon absorption
per unit infinitesimal path length
2. Scattering: – In this case the incident photons behave like a particle and interact
with the atom of the medium causing scattering. The process of scattering
depends on the type of particle with which the atoms interact. Depending on the
medium the photon scattering can either be isotropic or anisotropic. According to
the Scattering model used by us, the anisotropy term is expressed by a g-factor
which is explained in next chapter.
Scattering Coefficient (µs): It is defined as the probability of photon scattering per
unit infinitesimal path length.
Total interaction coefficient (µt): The sum of the absorption coefficient (µa) and
the scattering coefficient (µs) is termed as the total interaction coefficient µt.
ast µµµ +=
Albedo: It is defined as the fraction of light getting scattered and is expressed as
t
sbµµ
=
3. Fluorescence: Fluorescence is the phenomenon in which absorption of light of a
given wavelength by a fluorescent molecule is followed by the emission of light
at longer wavelengths. The distribution of wavelength-dependent intensity that
causes fluorescence is known as the fluorescence excitation spectrum, and the
distribution of wavelength-dependent intensity of emitted energy is known as the
fluorescence emission spectrum.
1.5 Fluorescence Spectroscopy
1.5.1 Introduction
Spectroscopy is the study of the interaction of the electromagnetic radiation with
matter. Optical spectroscopy deals with interactions of electromagnetic radiation with
matter that occur at the UV, VIS, near-infrared (NIR) and infrared (IR) wavelengths.
There are three aspects to a spectroscopic measurement: irradiation of a sample with
electromagnetic radiation; measurement of the absorption, spontaneous emission
(fluorescence, phosphorescence) and/or scattering (Rayleigh elastic scattering, Raman
inelastic scattering) from the sample; and analysis and interpretation of these
measurements [12].
1.5.2 Fluorescence [12]:
Figure 1 displays an energy level diagram with ground (S0) and excited (S1) electronic
states as well as vibrational energy levels within each electronic state of a molecule.
When a molecule is illuminated at an excitation wavelength that lies within the
absorption spectrum of that molecule, it will absorb the energy and be activated from its
ground state (S0) to an excited state (S1). The molecule can then relax back from the
excited state to the ground state by generating energy either nonradiatively or radiatively,
depending on the local environment.
In a nonradiative transition, relaxation occurs by thermal generation (dashed arrows). In a
radiative transition, relaxation occurs via fluorescence at specific emission wavelengths
(solid arrow). Fluorescence generation occurs in three steps: thermal equilibrium is
achieved rapidly as the electron makes a nonradiative transition to the lowest vibrational
level of the first excited state; the electron then makes a radiative transition to a
vibrational level of the ground state; and finally the electron makes a nonradiative
transition to the lowest vibrational level of the ground state.
1.5.3 Importance of Fluorescence Spectroscopy [13]
Fluorescence detection has three major advantages over other light-based investigation
methods: high sensitivity, high speed, and safety. The point of safety refers to the fact
that samples are not affected or destroyed in the process, and no hazardous byproducts
are generated.
Sensitivity is an important issue because the fluorescence signal is proportional to the
concentration of the substance being investigated. Optical fluorescence spectroscopy is
often applied in the analytical laboratory to measure molecular concentrations of
fluorophores in clear or turbid samples[14]. Whereas absorbance measurements can
reliably determine concentrations only as low as several tenths of a micromolar,
fluorescence techniques can accurately measure concentrations one million times smaller
-- pico- and even femtomolar. Quantities less than an attomole (<10-18 mole) may be
detected. Using fluorescence, one can monitor very rapid changes in concentration.
Changes in fluorescence intensity on the order of picoseconds can be detected if
necessary.
Because it is a non-invasive technique, fluorescence does not interfere with a sample.
Especially in clinical applications, this non-invasive nature is of great help. The
excitation light levels required to generate a fluorescence signal are low, reducing the
effects of photo-bleaching, and living tissue can be investigated with no adverse effects
on its natural physiological behavior.
1.5.4 Instrumentation for fluorescence spectroscopy
Laser: Lasers, which have a very narrow spectral output, are generally used as
monochromatic excitation light sources for fluorescence spectroscopy. Lamps also can be
used as quasimonochromatic excitation light sources for fluorescence spectroscopy when
coupled with a monochromator or narrow-bandpass filter.
Laser Monochromator Detector
Lamps provide the option of wavelength tenability over the UV/VIS spectral range, but
the disadvantage is that the power of these lamps is very less compared to the lasers.
Probe–Illumination and collection geometry: With respect to illumination and
collection of light from tissue, two different approaches may be considered: the contact
approach, where fiber-optic probes are placed directly in contact with the turbid medium;
and the noncontact approach, where a series of lenses are used to project the light onto
the turbid medium and collect it though lenses and fiber optic robe arrangement. A
detailed design analysis of the probe geometry is discussed in chapter 4.
Monochromator: A monochromator is used to spectrally disperse light. A
monochromator presents one wavelength or bandpass at a time of the input light from its
exit slit. Monochromators in fluorescence spectroscopy can be used as filters in
conjunction with arc lamps to produce a series of monochromatic outputs for sample
illumination (if only a few wavelengths are needed, narrow-bandpass filters may be more
appropriate) or can be used to disperse the emitted fluorescence light into its respective
wavelengths.
The key components of a monochromator/spectrograph are: an entrance slit; a
collimating lens; a grating (dispersing unit) for wavelength selection; a focusing lens; and
an exit slit
Detectors
Generally the fluorescence spectroscopy is performed in UV/VIS range. So Si photodiode
detectors are used. The fluorescence signal is very weak of the order of nano watts. So
generally a photomultipler tube (PMT) is used in conjunction with Si photodiode
detector. Sometimes CCDs are also used for detection.
Chapter 2
Monte-Carlo model of light transport in turbid medi um
2.1 Monte Carlo Simulation
Monte Carlo Simulation is a powerful method that is used to solve various physical
problems by constructing a stochastic model in which the expected value of a certain
random variable or a combination of several variables is equivalent to the value of a
physical quantity to be determined. This expected value is then determined by the
average of multiple independent samples representing the random variable introduced
above. The random variable can be generated using different kind of distribution which
can be realized using programming techniques. Monte Carlo simulations thus offer a
flexible, yet rigorous approach to photon transport in a turbid solution: which can score
multiple physical quantities simultaneously. The method describes local rules, of photon
propagation that are expressed, in the simplest case, as probability distributions that
describe the step size of photon movement between sites of photon-particle interaction in
the medium, and the angles of deflection in a photon’s trajectory when a scattering event
occurs. However, the method is statistical in nature and as such, relies on calculating the
propagation of a large number of photons (e.g.50 000) by the computer.
2.2 Modeling the Laser Source
The initial task in hand was to simulate the output of a laser source. As evident the laser
light has a Gaussian profile. This was simulated by making the photons to be emitted out
of the fiber core with a radial distribution given by the a Gaussian probability function
shown below-
P(r) = (1/ σ√ 2π) Exp (-r2/2σ2)
However, the fibers that we were using produced multiple modes. Hence it was estimated
that the laser light profile given by a Super-Gaussian distribution will prove to give a
better solution. The super Gaussian distribution function is given by the following
equation.
P(r) = C Exp (-r4/2)
where C is the normalized constant. The above equation is for a variance of 1 unit.
However a random number with the required variance can be achieved by simply
multiplying the random number generated using the equation above, by the required
variance. Generating a random number with a Super-Gaussian distribution involved a bit
of tedious work as it requires developing a time efficient algorithm. Unlike other standard
distributions like the Gaussian, which is a built-in-function in the Matlab directories,
there is no provision for generating Super-Gaussian random numbers.
Generation of Super-Gaussian Random Numbers
The basic steps involved in the generation of Super-Gaussian random number are-
1. To choose randomly (Uniform distribution) a value of r that lies inside the core of
a fiber i.e. 0 to core radius.
2. Calculate the value of the super Gaussian distribution at r (normalization constant
C=1) and store it in a variable f.
3. Next generate a new uniformly distributed random number between 0 and 1 and
store its value in y. If the value of the variable y is less than the f then r is chosen
else it is discarded.
4. On performing the above steps large number of time we get the values of r that
follow a super Gaussian distribution.
2.3 Simulation model for scattering and absorption process
Light transport can be modeled by two different methods- weighted approach or now
weighted approach [1]. In weighted approach, after each propagation step, a part of the
packet is absorbed and a part is scattered. But in non–weighted approach, either the
whole photon packet is absorbed or scattered i.e., each photon packet is being considered
as a single photon. The weighted approach gives a good accuracy over non-weighted
approach. So in all our modeling we considered weighted approach.
The scattering process is modeled using the Monte Carlo Techniques based on variable
step size method. The steps involved are-
Figure 1: Flowchart for the variable stepsize Monte Carlo technique.
2.3.1 Photon Initialization
The simulation is performed for a fixed number of photon packets that are made to emit
randomly from the central fiber of the optrode. The modeling is done by considering one
Launch New Photon Packet with intial weight wi
Move distance s
Absorb
−= −
t
aii µ
µωω 11
Scatter (Using Henyey-Greenstein
function)
Photon Collected by
detector
Pi = Pi-1 + wi
wi < wcritical
Yes
No
Yes
No
End
Last Photon
photon packet at a time. The weight of the photon packet is initialized to one. A point is
randomly chosen on the fiber (r,Ψ) which is obtained by generating a Super-Gaussian
random number r and a uniformly distributed random number phi. Next the photon
direction is randomly chosen within the numerical aperture of the fiber. The direction
cosines of the photon are stored using variables (ux, uy, uz).
2.3.2 Generating propagating distance
The step size of the photon packet is calculated based on a sampling of the probability
distribution for the photon’s free path s (0 < s < ∞), which is the step size. According to
the definition of interaction coefficient µt, the probability of photon-particle interaction
per unit path length in the interval (s’, s’ + ds’) is:
or
where P{} gives the probability for the condition inside the {} to hold The above
equation can be integrated over s’ in the range (0,s1) and can lead to an exponential
distribution, where P{s ≥ 0} = 1 is used:
The probability density function of free path s is:
On using the above equation the step size with the above exponential distribution can be
calculated to be
where ξ is a uniformly distributed random number between 0 and 1. Note that the average
of the random numbers generated above gives the mean free path of the photon-particle
interaction in the turbid medium.
The photon free path or the step size is estimated using the above exponential distribution.
A particular value of ut is thus chosen for performing the simulation. This distance is then
converted into the coordinate system by multiplying s with the direction cosines. After
moving a distance s, the position of the photon is updated.
2.3.3 Absorption
After each propagating step, the weight of the packet is split into two parts- the absorbed
part and the scattered part. The absorption of the photon packet is taken into account by
decreasing the weight of the photon packet according to the following equation:
2.3.4 Scattering Model
In this case the incident photons behave like a particle and interact with the atom of the
medium causing scattering. The process of scattering depends on the type of particle with
which the atoms interact. Depending on the medium, the photon scattering can either be
isotropic or anisotropic. According to the scattering model used by us, the anisotropy
term is expressed by a g-factor. The model assumes a diffused scattering with the cosine
of the scattering angle given by a probability distribution which is a function of the g
factor. The probability distribution for the cosine of the deflection angle, cosθ, is
described by the scattering function-
where the anisotropy factor g can have a value between -1 and 1. A value of 0 denotes
isotropic scattering, -1 as back scattering and 1 as forward scattering. Using the above
equation the value of cosθ as a function of a random number ξ (uniformly distributed
random number between 0 and 1) can be calculated to be-
−= −
t
aii µ
µωω 11
Steps during scattering
The remaining part of the photon packet left over after absorption undergoes scattering as
follows:
1. The scattering model as discussed above is used in identifying the scattering angle
of the photon (θ), which can lie in the range 0 to 180oC. The angle θ depends on
the g factor which can be chosen between -1 to 1. Again a particular value for g is
chosen for the simulation.
2. Along with the angle θ, a Ψ angle is also calculated. Ψ essentially helps in
realizing a 3D motion of the photons. It is clear that Ψ follows a uniform random
distribution and can have values between 0 and 360o.
3. Next the direction cosine of the photon is updated using the values of the
scattering angle θ and Ψ as obtained above.
4. This process (from step 3 to step 7) is repeated again and again till one of the
termination conditions is encountered. Different termination conditions are used
for handling two cases which is discussed in the next sections.
5. Once the termination condition is achieved a new photon is then incident into the
turbid medium and the above steps are repeated.
2.3.5 Termination Conditions:
Two cases are considered in evaluating the scattering parameter g and ut.. Both these
cases are realized using different termination conditions that are mentioned below-
Case 1: Termination Condition for Diffused Reflection
a. The photon after multiple scattering goes back into the collecting fiber.
This is done by monitoring the z value. If the z value becomes negative
then the position of the photon is calculated and compared with the probe
dimension.
b. If the photon moves forward beyond a critical distance, then the photon is
terminated. In this model we have assumed that once the photons cross a
critical distance Zcritical, chances that they scatter back and be collected is
minimal. Typical values choosen by us is about 10 cm which is quite
justifiable.
c. Also a radial limit is put on the motion of the photons i.e. if the photon
crosses a certain critical distance Rcritical then we terminate the loop and
discard the photons.
Case 2: Simulation Technique for Modeling Transmitted Light
a. The photon after multiple scattering goes back into the collecting fiber and
hence not detected. This is done by monitoring the z value. If the z value
becomes negative then photon is discarded and the loop terminated
b. If the photon moves forward beyond a distance, i.e. the point at which we
are detecting the photon (Detector). In such a case the photon is taken into
account and a new photon is then made to incident onto the medium.
c. A radial limit is put on the motion of the photons i.e. if the photon crosses
a certain critical distance Rcritical then we terminate the loop and discard the
photons.
2.4 Verification of the Program Code
In order to check the simulation, the incident
photons are made to fall on the medium and the
distribution of the photons scattered back on to the
fiber is plotted on a graph. In the first trial we
received a non-uniform pattern as can be seen in
Figure 3.1. This made us debug the program code,
until we received a uniform distribution of the back
scattered light. Figure 3.2 shows the distribution
obtained after the corrections, indicating that the
program code is working well and can be used for
simulating different working conditions. The empty
circle at the center is the illuminating fiber.
2.5 Simulation model for fluorescence
In the earlier model, photon on interaction with medium is either scattered or absorbed. In
the model for fluorescence, once the photons are absorbed by fluorescent medium, a new
photon at different wavelength is emitted.
The ‘g’, ‘µs’ and ‘µa’ for an incident photon and an emitted fluorescent photon are
different and their values have to be used accordingly. The important steps in the
simulation model are as follows:
Figure 3.1
Figure 3.2
1) For scoring photon absorption, a two-dimensional homogeneous grid system is set
up in the r and z directions. The grid separations are Ar and Az in the r and z
directions, respectively.
2) Photons from the illuminating fiber propagate through the medium experiencing
both scattering and absorption as per the model in 2.3. At each stage of absorption,
the weight of the photon packet that is absorbed is stored in the local grid element
(Ar i, Azi).
3) After all the photons are exhausted (say 1,00,000 photons) , the simulation for
fluorescence is executed. From each local grid element, photons are emitted
isotropically, which corresponds to the fluorescence signal. The number of
photons that are emitted from each grid element are equal to the closest integer to
the weight of the local grid element,
4) Once the fluorescence photon is emitted, the propagation steps are same as one in
section 2.4, except that the photon now travels with a different ‘g’, ‘µs’ and ‘µa’.
5) The terminations conditions are applied as per the probe design.
Chapter 3
Estimation of the optical properties of a turbid medium
Importance of finding the optical parameters
We started our project, performing experiments on the measurement of the diffused
reflection. The collected power was measured for different input powers. However,
simulation of the above experiment requires the values of the three parameters g , µs and
µa of the liquid. Using the above experiment, it is not feasible to determine all the above
interaction parameters. Thus, a new experiment had to be set up to identify these
parameters.
Before proceeding, it is important to understand some of the fundamental aspects of the
light-matter interaction. As was discussed in the last chapter, two mechanisms viz.
scattering and absorption occur when light interacts with a medium. The scattering
mechanism is characterized by the parameters g and µs. The g parameter depends on the
way the photon interacts with the atoms of the medium and hence is independent of the
concentration of the turbid medium. However, it depends on the size of the scattering
particles in the medium and is also a strong function of the wavelength. The absorption
mechanism on the other hand is characterized by the parameter µa. Both µa and µs depend
on the concentration of the medium i.e. if the medium is dense then the photons interact
with the particles at a much shorter distance (photon free path) and thus intensity drops
drastically. The sum of µa and µs gives us the total interaction coefficient denoted by µt.
Different experiments (as discussed below) were tried in an attempt to identify these
parameters for simulating the light propagation in a turbid medium.
Experiments for estimation of optical parameters
Commonly used methods
A lot of literature survey was done to identify such experiments. We came across two
most widely used methods. The most common method was the integrating sphere
technique.
Integrating sphere technique: The spectrophotometer measures direct transmission of
light through a sample as a function of wavelength. When equipped with an integrating
sphere attachment, the spectrophotometer can also measure the diffuse reflection and
total transmission of a sample. Three different measurements (collimated transmission,
diifuse transmission and diffuse reflection) are therefore available as a function of
wavelength.
Once these measurements are done, then simulations are used to determine the optical
parameters. The most common simulation code is the inverse-adding doubling method [].
experiments.
The drawback of the method is that most
commonly available spectrophotomters does
not have integrating sphere attachment for
liquid samples; the attachment is generally
provided only for thin films.
1) Diffused reflectance
When the hole is (not) closed, detector gives (diffused transmittance.. (2)) total transmittance
Total - diffused gives
collimated transmittance.(3)
The other drawback is that, even if such an arrangement was set-up for liquid samples,
then there will be significant losses of light from the side walls of the liquid container (as
shown in fig.), which will lead to errors in the estimated optical parameters. In our
method, these losses are eliminated.
The other method that can be used for estimation of the optical parameters is by time
resolved spectroscopy analysis. However, it is expensive and a complex set-up.
Experimental methods developed by us
As the existing methods were difficult to implement, we were left to explore a new and
simple way of identifying the parameters. As a first attempt, we neglected the
contribution due to the absorption by the liquid i.e. µa = 0; µt = µs. This simplified our
problem to a two variables problem viz g and µs.
I. Method for pure scattering case
To determine the scattering
parameters a new experiment was
designed. This involved the
measurement of the Diffused
Transmittance for different
apertures. Fig.1 shows the
experimental set-up. Here a laser
beam was coupled to the fiber from
one end. The other end of the fiber
was mounted on an XYZ translator stage with its tip inserted into the turbid medium. The
turbid medium was placed in a cylindrical glass cell with a flat bottom; the cell was
constructed by fixing a glass cover- slip at one end of a small glass cylinder. A variable
aperture followed by a detector was placed just below the cell. The transmitted power for
different apertures was measured at different value of ‘z’. Optical alignment and
Detector
MO
Power meter Turbid
solution z
He-Ne Laser (543 nm)
Fig. 1: Set-up for diffused transmission experiment
Variable Aperture
zx
y
positioning of the fiber tip with the aperture and the detector was achieved using X-Y
translation.
Fig 2 shows the
normalized output for
two different
concentrations of
turbidity. Using Fig.2
the intensity drop was
measured at a distance
Z=0.3mm for each
aperture values.
Aperture 4 and aperture
5 give similar readings.
Also due to error in
experiment aperture 1 is
neglected. The radius
for the three apertures
(2, 3 & 4) is – 325µm,
655µm & 1035µm
respectively. Also a rough
estimate of 8% loss in the
collected power due to
the glass slide was
observed while
performing the
experiment. Hence, in an
attempt to match the
experimental results with
the simulation, an 8%
increase is made to the
Table 1: Collected no of photons for 30000 incident photons
25% turbidity
0
0.2
0.4
0.6
0.8
1
1.2
0 100 200 300 400 500 600 700 800
Distance(Z in micormeter)
Po
wer
e(N
orm
aliz
ed t
o 1
)
aperture2
aperture3
aperture4
aperture5
50% turbidity
0
0.2
0.4
0.6
0.8
1
1.2
0 50 100 150 200 250 300 350 400 450 500
Distance (Z in micrometer)
Po
wer
( N
orm
aliz
ed t
o 1
)
aperture1
aperture2
aperture3
aperture4
aperture5
Fig 2: Normalized Collected Power at different Z for two different turbidity
collected powers, expressed in no of photons collected for 30000 incident photons. To
find out g and µt we had adopted the following method-
The experimental ratio of output to input power at a particular distance ‘Z ‘ was noted for
different apertures and compared it with the ratio obtained by simulation (output number
of photons to the input number of photons at the same ‘Z’ for the corresponding
apertures). Table 1 shows the observed results for Z = 0.3mm for three apertures namely
2, 3 and 4.
Next for each aperture, the
possible values of ‘g’ and ‘µt’
were noted from the simulation
results and plotted in Fig 3. The
curves for different apertures
intersect at a particular value of ‘g’
and ‘µt’, because of the fact that,
for a given turbidity ‘g’ and ‘µt’
remains same. From Fig. 3, we
notice that the curves intersect at a
common point around g = 0.9 & M.F.P (mean free path = 1/ µt) = 77nm. However,
simulations at such low values of M.F.P were not completely performed as it requires
large computational time. Hence we need to confirm the results by performing
simulations for lower M.F.P‘s as well.
With these calculations we are able to estimate our parameters ‘g’ and ‘µs’ qualitatively.
But the arrangement for the above setup esp. alignment of the aperture with the detector
was tedious and time consuming.
Now with an idea of the above method we tried to move to a three variable problem.
Another experiment on Collimated transmission was setup and few minor changes were
-1.0 -0.8 -0.6 -0.4 -0.2 0.0 0.2 0.4 0.6 0.8 1.0
02468
101214161820222426283032343638
Pho
ton
Mea
n F
ree
Pat
h(M
.F.P
)- u
m
g
Aperture 2 Aperture 3 Aperture 4
Fig 3: Plot of M.F.P Vs g for different Apertures
made to the existing Diffused transmission experiment for estimating the parameters ‘g’,
‘µa’ and ‘µs’.
II. Collimated Transmission experiment
Collimated transmission experiment
was setup to identify the total
scattering coefficient ut. A green He-Ne
(543 nm) laser beam was made to pass
through a certain length of the turbid
liquid placed in the wedge (Fig.4) and
the collimated light was measured
using a photo detector placed far away from the sample to ensure that no scattered power
was collected. The measured power was normalized with a reference liquid, usually water
or the solvent with which the turbid solution was prepared. The experiment was
performed for different sample lengths by moving the cell across the laser beam and the
average value of µt was calculated.
III. Method by considering both scattering and absorption mechanism
The three parameters ‘g’, ‘µa’ and
‘µs’ could also be written in a new
form as ‘µt’ , ‘g’ and ‘b’. A term
‘albedo’ denoted by b is introduced
which is defined as the fraction of
light getting scattered and is
expressed as.
t
sbµµ
=
The value of µt is measured using the
collimated transmission experiment. An advantage of writing the original parameters to
Power meter
Detector
He-Ne Laser (543 nm) MO
Turbid solution z
x
y
z
Fig. 5: Set-up for diffused transmission experiment
Power meter
He-Ne Laser (543 nm)
Turbid solution in a fabricated cell
Scattered Light
Fig. 4: Set-up for collimated transmission experiment
Milk (g=0.91, b = 0.865)
0.8
0.85
0.9
0.95
1
1.05
0.7 0.75 0.8 0.85 0.9 0.95 1
g
alb
edo
(b)
z=6mm
z=9mm
z=12mm
0.91
new form is that both the parameters ‘b’ and ‘g’ can have values between 0 and 1 only.
This makes the simulation analysis much easier. These two parameters were obtained
using the diffused transmission setup (Fig 5). In this setup, the aperture was removed and
the detector of a known active area was placed directly below the cell. The transmitted
power for different values of z was measured.
The ‘diffused
transmission’
power collected
by the detector
was measured
for three
different values
of z and was
normalized with
respect to the
corresponding
power, when the
turbid liquid was replaced by
the reference sample (DI water or the solvent with which the turbid medium was
prepared). A Monte Carlo simulation was developed with the above measured value of µt
and values of g and b between 0 and 1 were used as an input. Using the Monte Carlo
simulation (discussed in chapter 2), we determined the corresponding power collected
within the aperture of radius R. For each value of z, the experimentally measured
normalized power was matched with the simulation result for several possible
combinations of b and g. Figure 6 shows the three curves, corresponding to the three
different values of z, representing the possible combinations (g,b). For a given turbid
medium, since the optical properties are constant and are independent of the choice of z,
the common point of intersection give us the values of b and g for that liquid.
Fig. 6: Plot of (g,b) for different values of z (6mm, 9mm ,&12mm)
Results:
Experiments on milk
Experiments were performed on milk (1 ml milk in 40ml water) and the results were
parametric optimized with the simulation. The possible pairs of (g ,b) for each z are
identified and plotted in figure 3. The curves intersect at a common point g=0.93, b=0.87.
The results for g were in agreement with I.V. Yaroslavsky et al [4], which gives a good
confirmation of our method.
Experiments on gelusil:
Experiments were performed with two different concentrations of gelusil. The results
are shown in table1.
We can see that as the
concentration of the turbid
medium increases ( i.e., as µt
increases), µs and µa also
increases and ‘g’ remains nearly
same. This validates the method
and the argument that ‘g’ does
not depend on the concentration
of the sample and only µs, µa
depend on the concentration of
the sample.
µt (cm-1) g b µs (cm-1) µa (cm-1) 2.32 0.91 0.885 2.05 0.267 14.07 0.93 0.96 13.5 0.57
0.80 0.85 0.90 0.95 1.00
0.78
0.80
0.82
0.84
0.86
0.88
0.90
0.92
0.94
0.96
0.98
1.00
1.02
g = 0.91b = 0.885
Gelusil
Z = 6 mm Z = 9 mm Z = 12 mm
Alb
edo
(b)
Anisotropy Factor (g)
µt = 2.32 cm-1
Experiments on Ag nanoparticles
For both milk and gelusil, the ‘g’ factor was more than 0.9. In order to check the theoretical argument that ‘g’ factor depends on the dimensions of the scattering particles, we have performed experiments on silver nanoparticles. These nanoparticles have dimension much less than the wavelength of the laser light, so theoretically the scattered light should be more isotropic compared to the case when the dimensions of the scattering particles are of the same size as that of wavelength. The silver nanoparticles that we used in our experiment were borrowed from thin film lab, and their was around ~150 nm as estimated from SEM.
From the above graph, we can see that the ‘g’ factor for silver nanoparticles is around 0.65 and this validates the method and the argument that ‘g’ factor depends on the dimensions of the scattering particles.
Silver nanoparticles
0.4
0.5
0.6
0.7
0.8
0.9
1
1.1
0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1
g
Alb
edo
(b
)
z =3mm
z =6mm
z =9mm
Chapter 4
Probe Design Analysis
There are several issues involved in designing an optimum probe for collecting
fluorescence signal. The foremost issue is the probe size i.e. how big should the probe be
and under what circumstances could it be used? For e.g if we need to use designing a
probe for endoscopy, then we need to consider the probe size as the deciding factor for
fabrication rather than on improving the signal at the cost of a larger probe. In endoscopy,
we have a bundle of fibers, with one or few fibers as the illumination centers and the rest
serving as the collection centers. The collection efficiency is improved by the use of a
ball lens, but however the collection of the fluorescence signal is limited by the small
numerical aperture of the fiber. Thus only a small fraction of the fluorescence signal gets
coupled to the collecting fibers.
But if there are no size constraints then we can design a probe whereby the collection
efficiency can be improved by using a lens system that can couple more fluorescence
signal to the fiber. In this project, we have tried to perform experiments on fluorescence
and determine the possible optimized probe geometries. There are two ways in which the
probe could be designed - (1) In proximity mode or (2) remote mode (optrode not in
contact with the liquid).
Proximity Mode
In continuation to our work before the mid-semester evaluation, we have tried to study
the light propagation through fluorescence liquid (Rhodamine 6G and Basonyl Rot 542)
and tried to calculate the expected signals for different configurations of the probe. In the
proximity mode, the optrode was kept very close to the liquid. Rhodamine 6G has an
absorption maximum at 529.5nm and fluorescence excitation peak at 556nm. Basonyl
Rot has an absorption maximum at 542nm and an excitation peak at 600nm. For the He-
Ne laser source at 543nm, Basonyl Rot 542 gives a stronger signal as compared to
Rhodamine 6G and thus was preferred for experimental purposes.
The different configurations tried are listed below:
1. Simple geometry – In this case the probe was simply immersed in a liquid and
simulation analysis was done to calculate the measured signal (Fig 1). Experiment
was also performed for this configuration and the following plot was obtained.
2. Concave Reflector- In this case a concave reflector was placed inside the turbid
medium as shown in Fig 4.2. A Monte Carlo model simulating such an
experimental setup was developed with two input parameters L, distance between
the immersed optrode tip and the center of the concave mirror and R, radius of
curvature of the mirror. Fig 4.3 shows the variation of the fluorescent signal for
different values of R for a given L. The experiment was performed for three
Figure 4.2: Experimental Setup with a simple probe geometry
Power meter
Detector Laser
MO
Fiber chuck
Turbid solution
Figure 4.1: Experimental Setup with a simple probe geometry
Power meter
Detector Laser
MO
Fiber chuck
Turbid solution
different values of L. The x- axis is shown in terms of the ratio R/L rather that R.
We notice that the signal is strongest when the ration R/L lies between 1.4-1.7.
Thus it is appropriate to take the R/L ratio to be 1.5 as the optimized condition for
efficient collection of the fluorescent signal. i.e. L = 2R/3.
R,L Optimization Chart
0
0.1
0.2
0.3
0.4
0.5
0.6
1 1.2 1.4 1.6 1.8 2 2.2 2.4
Ratio of R/L
Flu
ore
scen
t sig
nal
( in
% o
f in
cid
ent
ligh
t)
L=3mmL=5mmL=4mm
Next the length L was varied to measure the optimum L so that maximum
fluorescence signal was obtained. Fig 4.4 shows that plot indicating that for the
given liquid with the interaction parameters (‘g’=0.9, ‘µs’=2, ‘µa’=3 ), maximum
signal was collected when L = 2.5mm.
3. Concave Reflector with a Condenser Lens: A lens was added to the previous
setup and collected signal was calculated for L=2R/3. Simulation was developed
for this experiment. The output of the simulation yields the results in Fig 4.5. An
Figure 4.4: Fluorescent Signal Vs L for R/L=3/2
Figure 4.3: Fluorescent Signal Vs R/L ratio
0
0.1
0.2
0.3
0.4
0.5
0.6
0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000
Distance(L)
Flu
ore
scen
ce s
ign
al (i
n %
of t
he
inci
den
t lig
ht)
increase in the signal was obtained for 2 different values of the focal length i.e.
3cm and 6cm of the condenser lens (Bi-convex lens).
Based on the above experiments we got a good idea of some aspects that are discussed
below-
• The fluorescence signal collected by the optrode is restricted to only few
millimeter of the liquid length i.e. majority of the signal collected by the optrode
is within 2-3 mm of the liquid length below the optrode tip.
• Large amount of the fluorescence signal backscattered went into the illuminating
fiber which meant only small amount of power was collected by the collecting
fibers.
Figure 4.5: Fluorescent Signal Vs Focal Length of the Condenser (Biconvex) Lens for L=2500. R/L=3/2
Power meter
Detector Laser
MO
Fiber chuck
Turbid solution
Condenser Lens
Figure 4.4: Experimental Setup with a probe having reflector and lens system geometry
Optimized Bi-convex Lens
0.25
0.27
0.29
0.31
0.33
0.35
0.37
0 1 2 3 4 5 6 7 8 9
Focal Length (cm)
Flu
ore
scn
et s
ign
al (
in %
of
Inci
den
t L
igh
t)
• Since the reflector also needs to be placed within 2-3 mm, this requires us to use a
condenser with a high N.A so that illuminating signal focuses to the central region
between the condenser and the mirror.
Thus based on the above observations it was decided to use the configuration shown in
Fig 4.6. for the contact mode. Here we have two illuminating fibers at the side. The light
from these fibers are focused to a spot at the center in the liquid, using a high N.A
condenser (or Microscope objective). The collecting fibers are placed at the center of the
system. Such a configuration helps us to overcome the limitation seen in the earlier
experiments i.e. the fluorescence collected at the center of the optrode is also carried with
the other collecting fibers.
But the major factors limiting these configurations are that (a) the fiber N.A is small and
so only a very small fraction of the fluorescence signal is coupled to the fibers and (b)
due to finite focal length of the liquid, the fluorescent signal is reabsorbed as it
propagates through the liquid towards the collecting fiber. At this stage we came up with
an innovative idea of doing an N.A down-conversion which is explained in the remote
mode.
Remote Mode
In this case the optrode is placed at a distance from the liquid medium. The fluorescence
from the liquid is then collected using a lens system, aiding N.A down-conversion. Some
ideas were taken from the experimental setup on Laser Doppler Anemometry. The liquid
sample is placed in a capillary tube and the illuminating light is focused onto the tube
(see Fig 4.7). A line streak is formed as the light passes through the capillary. The lens
system is such that this line streak is collimated using a higher N.A. lens, which then is
focused onto the collecting fibers using another lens with a N.A same as that of the fiber.
An experiment was setup with the above concept. Such a N.A down-conversion system
(see Fig 4.8) gives a better signal and was found to yield higher signal strength than the
proximity case.
Since there is no restriction on the size of the probe, we have decided to use this approach
in designing the final probe. We also plan to use a reflector so that a larger fraction of the
fluorescence signal is coupled to the optrode.
Optrode Fiber N.A = 0.22
(Hemi-)Spherical Mirror
Fluorescent Sample (Line Streak)
Bi-convex lens (N.A = 0.7)
Bi-convex lens (N.A = 0.22)
Figure 4.8: N.A. Down-conversion using a lens system
Fabrication of optrode: The final design of the probe that was fabricated is as shown in the figure:
The important parts of the design are
1) Main tube
2) Illuminating arrangement
3) Capillary tube that holds the liquid sample
4) Collecting lens arrangement
5) Collecting optrode
Main tube:
The tube is made of PVC and it holds all the arrangements in it viz, illuminating and
collecting lens arrangements, capillary tube and collecting optrode.
Collecting optrode (5)
Main tube
Lens Mirror Small capillary tube
High NA Lens
Low NA Lens
Illuminating arrangement (2)
Collecting lens arrangement (4) Capillary
tube (3)
Illuminating arrangement:
The illumination arrangement consists of an illuminating fiber, a mirror and a lens. The
laser was coupled to one end of the illuminating fiber and the other end was passed
through a capillary tube (with a very small bore), which provides support for the fiber.
The light from this fiber end falls on mirror placed at 450 to the axis of the main tube.
The reflected light is then made to pass through a lens (focal length of 12.5 mm), which
focuses light on capillary tube.
Capillary tube:
A hole is made in main tube and the capillary tube runs straight though
this hole. The turbid medium whose fluorescence spectroscopy has to
be performed can be passed into this capillary tube from the openings
outside the main tube.
Collecting lens arrangement:
The function of this arrangement is to collect the fluorescent light from the liquid sample
and focus it onto the collecting optrode. The important design requirement of this
arrangement is to maximize the collection of the fluorescence light. This is achieved by a
system of two lenses as explained below:
1) The first lens is of very high numerical aperture (NA -.66). This distance between
this lens and the capillary tube equal to the focal length of the lens i.e., 2 cm. So,
this lens collimates all the fluorescent light that is collected by it.
2) The second lens is of numerical aperture 0.25 that is nearly equal to that of the
fiber (NA-0.22). This lens collects the collimated light from the first lens and
focuses at its focal length (10cm) on other side.
Fig: Cross sectional view of main tube
Collecting optrode:
The collecting optrode is a bundle of fibers that carries the fluorescent light to the
detector. The geometry of the fiber bundle has to be such that it collects the maximum
fluorescence light.
Because the incident light on the capillary tube is line streak, we find that the collected
fluorescence light is also a line of streak. So the best possible arrangement of the
collecting optrode is linear array of fibers. However, as we have already fabricated a
optrode with hexagonal arrangement of fibers, in all our initial experiments we have used
this circular optrode.
The important steps in the fabrication of the collecting optrode are as follows:
• The dimensions of the fibers used are as follows:
Core (silica) – 600µm
Cladding (silica) - 660µm
Jacket (silica) - 710µm
Numerical Aperture -0.22
• The optrode consists of three layers of fibers (1-6-12)
• The 19 fibers were cut from the spool, each one with a length of around 1m. All
the fibers were passed through short individual pieces of sleeves, so that these
fibers don’t break by getting stressed against the glass bore, into which the fibers
were inserted.
• Now to get the arrangement of 1-6-12 fibers, the fibers were simply placed
together and then a thread was tied from the position where the sleeves were
located and continued towards the end, so that the fibers automatically get into the
required hexagonal arrangement. The fibers were permanently held in this
arrangement by using an adhesive.
• The fibers are then passed through a short glass bore and the fiber arrangement is
held in place with respect to the glass bore by putting wax (Beeswax + rosin is
melted and introduced into the gap, which on cooling solidifies)
Adhesive
Arranging the fibers in hexagonal pattern
Outer glass bore Teflon
tape
Sleeve
Cross Section
• All these fibers are now passed through a second glass
bore of bigger size, so that the position of the sleeves
matches with the edge of the bore. A Teflon tape is
used at the end where the sleeves are present to ensure
that the fiber and the glass bore are held together. The
gap between the opening end of the first bore and the
second bore is again filled with wax.
• Rough grinding and polishing: Unlike the normal fibers,
these fibers cannot be spliced easily to get a circular cut,
because they are very brittle. So it is necessary to grind
and polish to get smooth ends. It is also necessary that
the surfaces of the polished fiber ends should be flat,
without formation of any wedges. So a disc with hole
was fabricated from the workshop. This disc is now
fixed to the end of the optrode. The grinding Figure 6.2: Arrangement for grinding and polishing
Glass slab
Figure 6.1: Fabrication of optrode
was then done with emery (aluminum oxide) of different grades - MA-1, MA-2,
and MA-3. As we increase the grade of the emery, the surface roughness is
reduced. Then polishing was done with a lapping agent and a velvet cloth as
polisher. The surface of the fiber end is checked under microscope and polishing
is continued until the pit size is considerably reduced.
• After one end of the optrode is polished, similar steps of arranging the fibers in a
glass bore and polishing were performed with the other end.
Chapter 6
Conclusion and Future Plan
6.1 Conclusion
• A Monte-Carlo simulation incorporating all three mechanism-scattering,
absorption and fluorescence is developed.
• A simple and novel method to determine the optical properties ‘b’ and ‘g’ of a
turbid solution is developed. The method gives a straightforward estimation of the
optical properties for any turbid medium containing both scattering and absorbing
particles. One of the advantages of this method is that the model can be used to
calculate the effective anisotropy coefficient g of a turbid medium containing
more than one kind of scattering particles. This effective value of g can then be
used to study the light propagation in the given medium.
• Different designs for the optrode have been studied.
6.2 Future Work
• For determination of optical parameters, further experiments are required to
improve the accuracy. Some of the changes to made in simulation are to
incorporate reflections losses at glass slide. Experiments with latex spheres
(whose size is already known from the suppliers) for determination of optical
properties will help in further improvement of the procedure.
• Many designs can be fabricated and studied based on our proposed designs
mentioned in chapter 4. By theoretically analysis (ray approach) it can be said that
using reflectors and lenses will improve the collection efficiency. But a lot of
simulation analysis is needed to find out the optimum dimensions of these
reflectors and lenses. So all the future Monte Carlo simulations should have the
provision for incorporating these components.
• One of the limitations during the project was to find out a monochromator with
good resolution and a filter to separate the input laser signal from fluorescence
signal. Once a good filter and monochromator are found, several new designs can
be tried (using ball lens) which might be very compact and similar to the optrodes
used for endoscopy applications.
References
1) Characteristics of a high-temperature fibre-optic sensor probe Z.Y. Zhang,
K.T.V. Grattan *, A.W. Palmer, B.T. Meggitt, Sensors and Actuators A 64
(1998) 231-236
2) Fiber Optic Probes in Optical Spectroscopy, Clinical Applications
3) Multiple-fiber probe design for fluorescence spectroscopy in tissue, T. Joshua
Pfefer, Kevin T. Schomacker, Marwood N. Ediger, and Norman S. Nishioka,
APPLIED OPTICS _ Vol. 41, No. 22 _ 1 August 2002, 4712-4721
4) Light Propagation in Tissue During Fluorescence Spectroscopy With Single-
Fiber Probes, T. Joshua Pfefer, Kevin T. Schomacker, Marwood N. Ediger, and
Norman S. Nishioka, IEEE JOURNAL ON SELECTED TOPICS IN QUANTUM
ELECTRONICS, VOL. 7, NO. 6, NOVEMBER/DECEMBER 2001, 1004-1012
5) Influence of the emission–reception geometry in laser-induced fluorescence
spectra from turbid media, Sigrid Avrillier, Eric Tinet, Dominique Ettori, Jean-
Michel Tualle, and Bernard Ge´ le´ bart, 1 May 1998 y Vol. 37, No. 13 y
APPLIED OPTICS, 2781-2787
6) Fiber-optic bundle design for quantitative fluorescence measurement from
tissue Brian W. Pogue and Gregory Burke, 1 November 1998 y Vol. 37, No. 31
y APPLIED OPTICS, 7429-7436
7) Optical fiber probes for fluorescence based oxygen sensing P.A.S. Jorge a,b,∗,
P. Caldas a,b,c, C.C. Rosa a,b, A.G. Oliva d, J.L. Santos, Sensors and Actuators
B 103 (2004) 290–299
8) An optimized optrode for continuous potassium monitoring in whole blood E.
Malavolti a, A. Cagnini a, G. Caputo b, L. Della Ciana b, M. Mascini Analytica
Chimica Acta 401 (1999) 129–136
9) L. Wang, S. L. Jacques, L. Zhenq, “MCML - Monte Carlo modeling of light transport
in multi-layered tissues”, Computer Methods and Programs in Biomedicine, vol. 47,
pp.131-146, 1995.
10) Accelerated Monte Carlo models to simulate fluorescence spectra from layered
tissues, Swartling et al., 714-727
11) Propagation of Fluorescent Light, A.J. Welch, PhD,* Craig Gardner, PhD,
Rebecca Richards-Kortum, PhD, Eric Chan, MS, Glen Criswell, BS, Josh
Pfefer, MS, and Steve Warren, PhD, Lasers in Surgery and Medicine 21:166–
178 (1997)
12) http://www.pti-nj.com/fluorescence_2.html
13) Noninvasive measurement of fluorophore concentration in turbid media with a
simple fluorescence_reflectance ratio technique, Robert Weersink, Michael S.
Patterson, Kevin Diamond, Shawna Silver, and Neil Padgett, 1 December 2001
_ Vol. 40, No. 34 _ APPLIED OPTICS, 6389-6395