Vacuum PumpsTechnical Manual
DentalEZ MAKES YOUR PRACTICE PERFECT® ®
Section IV Vacuum Pumps
Technical Manual
iv-Page 1
Vacuum Pumps
VACUUM
PUMPS
Specifications ..........................................................iv-2
Certifications ..........................................................iv-3
Site Requirements CV-101 FS / CV-102 FS...........................................iv-4 MC-201 FS / MC-202 FS ........................................iv-5
Water Consumpsion Chart ..........................................................................iv-6
Recirculator Instructions CV-101 FS / CV-102 FS....................................... iv-7-9 MC-201 FS / MC-202 FS ............................... iv-10-12
Maintenance CV-101 FS / CV-102 FS........................................ iv-13 MC-201 FS / MC-202 FS ..................................... iv-14
Air / Water or Nitrous Oxide Separator AM-200 ................................................................... iv-15
Frequently Asked Questions Wire single pump to remote lighted switch .. iv-16 Wire dual pumps to remote lighted switch ... iv-17 One pump weak ................................................... iv-18 Buck-Boost Transformer data ...................... iv-19-20 Wire new transformer in old relay box .......... iv-21 Vacuum relay will not engage & pump will not turn on ................................... iv-22 Pump shuts down, cools then restarts ............ iv-23 Pump makes squealing / whistling sound ......... iv-24 Water leaks from anti-siphon valve .................. iv-25 Dual pump vacuum weak .................................... iv-26
Troubleshooting Vacuum Pump .................................................. iv-27-28 Electrical System ........................................................29
Wiring Schematics CV-101 FS / CV-102 FS........................................ iv-30 MC-201 FS / MC-202 FS ..................................... iv-31 Remote Low-Voltage Switch Control .............. iv-32 Two-Switch Dual Vacuum System ..................... iv-33 One-Switch Dual Vacuum System ..................... iv-34
System Repair Procedure CV-101 FS / CV-102 FS.......................................35-39 MC-201 FS / MC-202 FS ....................................40-44
Parts Lists CV-101 FS / CV-102 FS
Installation Kit .................................................... iv-45 Water Supply Line ............................................ iv-46 Vacuum Line ....................................................... iv-47 Waste Line ......................................................... iv-48 1 HP Single Pump Assembly ........................... iv-49 2 HP Sngle Pump Assembly ............................ iv-50 Electrical Box ..................................................... iv-51 Water Control Assembly ............................... iv-52 Anti-Siphon Valve Assembly ........................... iv-53 Water Solenoid Valve Assembly ................... iv-53 Intake Manifold Assembly ............................... iv-54 Primary/Secondary Filter................................. iv-54
MC-201 FS / MC-202 FS
Water Supply Line ............................................ iv-55 Vacuum Line ....................................................... iv-55 Waste Line ......................................................... iv-56 Electrical Connection Assembly .................... iv-56 Frame Assembly ................................................ iv-57 1 HP Dual Pump Assembly ............................. iv-58 2 HP Dual Pump Assembly ............................. iv-59 Intake Manifold Assembly ................................ iv-60 Exhaust Manifold Assembly ............................. iv-61 Water Recirculator, Exhaust Manifold Assembly (Optional) .................................. iv-61 Electrical Box ..................................................... iv-62 Water Control Assembly ................................ iv-63 Vacuum Control Assembly ............................. iv-64 Vacuum Gauge & Adapter ............................... iv-64 In-Line Filter Assembly .................................... iv-64 Anti-Siphon Valve Assembly ............................ iv-65 Water Solenoid Valve Assembly ..................... iv-65
Technical Manual
Section IV Vacuum PumpsPage iv-2
SPECIFICATIONS
Vacuum Pumps
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
2
This manual contains installation, operation, care andrepair instructions and user service information forthe CustomAir® Dual Wet Vacuum Pump (MC-201 FS/MC-202 FS).
The Dual Wet Vacuum Pump is designed to providetrouble-free service when installed, operated and caredfor according to the procedures set forth in this manual.
Section I Introduction
MotorPower Rating: 1 HP 2 HPVoltage: 208/230 208/230Amps (each motor): 7.5/7.5 15/15Cycle: 60 Hz.Phase: SingleRunning Speed: 3450 RPMAmbient Temperature Range: 10-40°C/50-104°FMax. Oper. Noise Level: 70-75 dB
VacuumMercury Pull (Sealed Sys.): Approx. 20-25" Hg. Adjust.One Pump OperationUsable CFM: 15 30Use factor (Number ofhigh-volume hosesopen simultaneously): 1.5 3Two Pump OperationUsable CFM: 30 60Use factor (Number ofhigh-volume hosesopen simultaneously): 3 6
DimensionsHeight: 23"Width: 24"Depth: 17-1/2"
Shipping Weight: 175 lbs. 215 lbs.
Water RequirementsGallons Per Min. (per pump): 1/2 1
Specifications
MC-201 FS MC-202 FS– Type of protection against electric shock: Class 1
Equipment
– Degree of protection against the ingress of water:Ordinary
– Equipment not suitable for use in the presence ofa flammable anesthetic mixture with air or withoxygen or nitrous oxide.
– Mode of operation: Continuous
– Environmental conditions for transport and storageRecommended ranges:
Classification
Medical Equipment withRespect to Electric Shock,Fire and MechanicalHazardsOnly in Accordance withUL60601-1 andCAN/CSA C22.2 No. 601.1and IEC60601-1-2
- Temperature range within - 40º C to + 70º C- Relative humidity range within 10% to 100%- Atmospheric pressure range within 500 to 1050 hPa
Explanation of Symbols used onEquipment:
Approvals:UL/CUL Classified: File E212679
53HN
= Attention, Consult Accompanying Documents
= Protective Earth Terminal
= Caution. Electrical Shock Hazard. Refer Servicingto Qualified Personnel
= Hot Surface
= European Certification
Installation, Operation and Care Manual
Single Wet Vacuum Pump
2
This manual contains installation, operation, care andrepair instructions and user service information forthe CustomAir® Single Wet Vacuum Pump (CV-101 FS/CV-102 FS).
The Single Wet Vacuum Pump is designed to providetrouble-free service when installed, operated andcared for according to the procedures set forth in thismanual.
Section I Introduction
MotorPower Rating: 1 HP 2 HPVoltage: 115/208/230 208/230Amps (each motor): 15/7.5/7.5 15/15Cycle: 60 Hz.Phase: SingleRunning Speed: 3450 RPMAmbient Temperature Range: 10-40°C/50-104°FWire Size: 12 GA.
VacuumMercury Pull (Sealed Sys.): Approx. 20-25" Hg. Adjust.Usable CFM: 15 30Use factor (Number ofhigh-volume hosesopen simultaneously): 1.5 3
Water RequirementsGallons Per Minute: 1/2 1
DimensionsHeight: 16" 20"Width: 12" 14"Depth: 9" 11"
Shipping Weight: 68 lbs. 78 lbs.
Approvals:UL/CUL Classified: File: E212679
Specifications
CV-101 FS CV-102 FS
= Attention, Consult Accompanying Documents
= Protective Earth Terminal
= Caution. Electrical Shock Hazard. Refer Servicingto Qualified Personnel
= Hot Surface
= European Certification
– Type of protection against electric shock: Class 1Equipment
– Degree of protection against the ingress of water:Ordinary
– Equipment not suitable for use in the presence ofa flammable anesthetic mixture with air or withoxygen or nitrous oxide.
– Mode of operation: Continuous
– Environmental conditions for transport and storageRecommended ranges:
Classification
Medical Equipment withRespect to Electric Shock,Fire and Mechanical HazardsOnly in Accordance withUL60601-1 andCAN/CSA C22.2 No. 601.1and IEC60601-1-2
- Temperature range within - 40º C to + 70º C- Relative humidity range within 10% to 100%- Atmospheric pressure range within 500 to 1050 hPa
Explanation of Symbols used onEquipment:
53HN
Section IV Vacuum Pumps
Technical Manual
iv-Page 3
CERTIFICATIONS
Vacuum Pumps
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
2
This manual contains installation, operation, care andrepair instructions and user service information forthe CustomAir® Dual Wet Vacuum Pump (MC-201 FS/MC-202 FS).
The Dual Wet Vacuum Pump is designed to providetrouble-free service when installed, operated and caredfor according to the procedures set forth in this manual.
Section I Introduction
MotorPower Rating: 1 HP 2 HPVoltage: 208/230 208/230Amps (each motor): 7.5/7.5 15/15Cycle: 60 Hz.Phase: SingleRunning Speed: 3450 RPMAmbient Temperature Range: 10-40°C/50-104°FMax. Oper. Noise Level: 70-75 dB
VacuumMercury Pull (Sealed Sys.): Approx. 20-25" Hg. Adjust.One Pump OperationUsable CFM: 15 30Use factor (Number ofhigh-volume hosesopen simultaneously): 1.5 3Two Pump OperationUsable CFM: 30 60Use factor (Number ofhigh-volume hosesopen simultaneously): 3 6
DimensionsHeight: 23"Width: 24"Depth: 17-1/2"
Shipping Weight: 175 lbs. 215 lbs.
Water RequirementsGallons Per Min. (per pump): 1/2 1
Specifications
MC-201 FS MC-202 FS– Type of protection against electric shock: Class 1
Equipment
– Degree of protection against the ingress of water:Ordinary
– Equipment not suitable for use in the presence ofa flammable anesthetic mixture with air or withoxygen or nitrous oxide.
– Mode of operation: Continuous
– Environmental conditions for transport and storageRecommended ranges:
Classification
Medical Equipment withRespect to Electric Shock,Fire and MechanicalHazardsOnly in Accordance withUL60601-1 andCAN/CSA C22.2 No. 601.1and IEC60601-1-2
- Temperature range within - 40º C to + 70º C- Relative humidity range within 10% to 100%- Atmospheric pressure range within 500 to 1050 hPa
Explanation of Symbols used onEquipment:
Approvals:UL/CUL Classified: File E212679
53HN
= Attention, Consult Accompanying Documents
= Protective Earth Terminal
= Caution. Electrical Shock Hazard. Refer Servicingto Qualified Personnel
= Hot Surface
= European Certification
File # 1371
Technical Manual
Section IV Vacuum PumpsPage iv-4
CV101FS / CV102FS
SITE
REQUIREMENTS
3Single Wet Vacuum Pump
The following Pre-installation information will assistin making a quick, easy and quality installation. How-ever, if there are any questions, contact a CustomAirtechnical service representative at 1-866-DTE-INFO.
Section II Pre-installation
Site Requirements
ElectricalAll electrical supply lines and control wiring should besupplied and installed by a licensed electrician accord-ing to local building codes.
Before the Single Wet Vacuum Pump can be properlyinstalled, the following utilities must be available:
WaterWater line must be installed by a plumber accordingto local building codes. Requirements: 1/2" gatevalve reduced to 1/8" FIP.
WasteWaste line must be installed by a plumber accordingto local building and health codes. Requirements:reduce connection to 1" FIP or floor sink. Whenusing floor sink, order a floor sink adapter.
Vacuum LinesVacuum lines must be installed by a plumber accord-ing to local building and electrical codes. Require-ments: 1" IPS, PVC, Schedule 40.
stloV *SF101-VCledoM SF201-VCledoM
V511 ybtceridderiWrepsanaicirtcele
.sedoclacolhtiwtiucricetarapeS
,rekaerbpma03.esahpelgnis
ebtonnaCV511otdetcennoc
.enil
V032/802 ybtceridderiWrepsanaicirtcele
.sedoclacolhtiwtiucricetarapeS
,rekaerbpma02.esahpelgnis
ybtceridderiW.naicirtcele
htiwtiucricetarapeS,rekaerbpma03
.esahpelgnis
Electrical Requirements
*Model CV-101 FS is manufactured and shipped for115V operation. This model may be converted to 230Voperation by qualified installation technicians only follow-ing connection diagrams inside the pump control box.
Low Voltage Control Line (Optional)If remote low voltage is desired, a licensed electricianshould install 18-2 thermostat wire from the pumplocation to the operatory switches. (Use 18-3 thermostatwire for lighted switch.)
Vacuum Line1" IPS, PVC,Schedule 40
Water Line1/2" Gate Valve
Reduced to1/8" FIP
Low Voltage Line24V (Optional)
Electrical115V or 230V
Connection Box
Waste Line"P" Trap Reduced
to 2" FIP
Floor SinkUse Floor Sink
Adapter
Section IV Vacuum Pumps
Technical Manual
iv-Page 5
MC201FS / MC202FS
SITE
REQUIREMENTS
3Dual Wet Vacuum Pump
epyTnoitcetorP)rotomhcae(
SF102-CMV032
SF202-CMV032
rekaerBtiucriC spmA02 spmA03
esuFyaleDemiT spmA21 spmA02
esuFdradnatS spmA52 spmA04
The following Pre-installation information will assistin making a quick, easy and quality installation. How-ever, if there are any questions, contact a CustomAirtechnical service representative at 1-866-DTE-INFO.
Section II Pre-installation
Site Requirements
WaterWater line must be installed by a plumber accordingto local building codes. Terminate water line in 1/2"FIP gate valve.
Waste
Motor Protection Chart
Low Voltage Control Line (Optional)
If remote low voltage is desired, a licensed electricianshould install 18-3 thermostat wire for lighted switch or18-2 thermostat wire for non-lighted switch from the pumplocation to the operatory switch located up to 150 feetaway. If the operatory switch is over 150 feet away fromthe pump location, use 16-3 thermostat wire for lightedswitch or 16-2 thermostat wire for non-lighted switch.
Before the Dual Wet Vacuum Pump can be properlyinstalled, the following utilities must be supplied:
ElectricalAll electrical supply lines and control wiring should besupplied and installed by a licensed electrician accord-ing to local building codes. The Dual Wet VacuumPump requires two separate circuits (one for eachpump motor) of the correct voltage for the systemordered. In addition to the thermal overload protectionbuilt into the motors, each circuit must be providedwith a circuit breaker, time delay fuse or standardfuse. See the Motor Protection Chart below forrecommended breaker or fuse amperage.
— IMPORTANT —All vacuum systems must be installed accord-
ing to local building and electrical codes.Vacuum Line
(IPS, PVC,Schedule 40)
Water Line(1/2" Gate Valve
Reduced to3/8" FIP)
Low Voltage Line(24V) (Optional)
Electrical(230V)
ConnectionBoxes (2)
Waste Line(Drain)
Waste line must be installed by a plumber accordingto local building and health codes. For requirements,see Typical Installation Options, Page 4 and WasteLine, Page 6.
Vacuum LinesVacuum lines be installed by a plumber according tolocal building codes. Vacuum lines and risers to beIPS, PVC, Schedule 40, unless local codes requireanother material such as copper. Slope all vacuumlines toward the vacuum producer 1 inch for every 20foot run. Make all vacuum line connections usinglong radius sweep fittings.
— CAUTION —A free-flowing discharge system is required forproper operation of the vacuum system. Thedual pump may leak at the anti-siphon valve ifthe discharge system is restricted. Ensure the
vacuum system is installed and cleaned in accor-dance to the instructions in this manual.
Technical Manual
Section IV Vacuum PumpsPage iv-6
WATER
CONSUMPTION
Wet Vacuum Pumps
Model # Gal. per hour w/o recirculator Gal. per hour w/recirculatorCV-101 30 5 - 6CV--102 60 9 - 10
MC-201 w/one pump running 30 5 - 6MC-201 w/both pumps running 60 9 - 10MC-202 w/one pump running 60 9 - 10
MC-202 w/both pumps running 120 18 - 20Note: Figures are for continuous operation.
Vacuum Pump Water Consumption Chart
Section IV Vacuum Pumps
Technical Manual
iv-Page 7
CV-101FS / CV102FS
RECIRCULATOR
INSTRUCTIONS
Page 1 of 3
INSTALLATION INSTRUCTIONS
���DentalEZ Group 2500 Highway 31 South Bay Minette, AL 36507 1-866-DTE-INFO
October, 2001
64710022CSingle Pump Water Recirculator
�������������������������������������������������������
�����������������������������������������������������������������������������������
Mounting
�� ��������������������������������������������������������������������������������������
�� �������������������������������������������������������������������������������������������
�� ���������������������������������������������������������������������������������������������������
�� ���������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
�� ����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
10
RemoveHose
1
2
13
To Drain
3
4
17 7
9
6
8
16
18
3
Technical Manual
Section IV Vacuum PumpsPage iv-8
CV-101FS / CV102FS
RECIRCULATOR
INSTRUCTIONS
Page 2 of 3
Pump Exhaust
�� ���������������������������������������������������������������������������������������������
�� �����������������������������������������������������������������������������������������������������������������������
�� ��������������������������������������������������������������������������������������������������������������������
Water Supply
�������������������������������������������������������������������������������������������������
�� �������������������������������������������������������������������
�� ����������������������������������������������������������������������
�� ���������������������������������������������������������������������
�� ��������������������������������������������������
�� �������������������������������������������������������������������������������������������������������������������
� ������������������������������������������� ������������������������������������������
�� ������������������������������������������������������������������������������������������������������� ��������������������������
�� �����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
�� ������������������������������������������������������������������������������������
�� ������������������������������������������������������������������������������������������������������
��� ����������������������������������������������������������������������������������������������������
��� �������������������������������������������������������������������������������������������������������������������������
Operation Check
�� ���������������������������������������������������������������������������������������
�� ��������������������������
Single Pump Water Recirculator
(On next page)
Section IV Vacuum Pumps
Technical Manual
iv-Page 9
CV-101FS / CV102FS
RECIRCULATOR
INSTRUCTIONS
Page 3 of 3
64710022C
View A
View B
View C
10
11
18
11
19
12
101410
13
Housing ElbowThreaded Hole
5
15
10
Technical Manual
Section IV Vacuum PumpsPage iv-10
MC-201 FS / MC-202 FS
RECIRCULATOR
INSTRUCTIONS
Page 1 of 3
INSTALLATION INSTRUCTIONS
���DentalEZ Group 2500 Highway 31 South Bay Minette, AL 36507 1-866-DTE-INFO
October, 2001
64710012CDual Pump Water Recirculator
�������������������������������������������������������
�����������������������������������������������������������������������������������
Mounting
�� ������������������������������������������������������������������������������������
�� ���������������������������������������������������������������������������
�� �����������������������������������������������������������������������������������������������������������������
�� ������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
�� ����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
10
RemoveHose
RemoveHose
1
2
15
13
To Drain
3
4 5 17 7
9
6 8
16
18
�������������������������������������������������������������������������������������������������������������������������������������������������������
Section IV Vacuum Pumps
Technical Manual
iv-Page 11
RECIRCULATOR
INSTRUCTIONS
MC-201 FS / MC-202 FS
Page 2 of 3
Pump Exhaust
�� �����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
�� ��������������������������������������������������������������������������������������������������������������������������������������������������������������������������
Water Supply
�������������������������������������������������������������������������������������������������
�� �������������������������������������������������������������������
�� ����������������������������������������������������������������������
�� ���������������������������������������������������������������������
�� �������������������������������������������������������������������������������������������������������������������
� ������������������������������������������� ������������������������������������������
�� ������������������������������������������������������������������������������������������������������� ��������������������������
�� �����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
�� ������������������������������������������������������������������������������������
�� ���������������������������������������������������������������
�� ������������������������������������������������������������������������������������
��� ���������������������������������������������������������������������������������������������������������������������������������������
��� ��������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������
Operation Check
�� ���������������������������������������������������������������������������������������
�� ��������������������������
Dual Pump Water Recirculator
(On next page)
Technical Manual
Section IV Vacuum PumpsPage iv-12
RECIRCULATOR
INSTRUCTIONS
MC-201 FS / MC-202 FS
Page 3 of 3
64710012C
View A
View B
View C
10
11
10
11
19
12
101410
13
Housing ElbowThreaded Hole
Section IV Vacuum Pumps
Technical Manual
iv-Page 13
MAINTENANCE
CV-101 FS / CV-102 FS
19Single Wet Vacuum Pump
Section VI Operation & Care
— WARNING —Before starting cleaning procedures, make
certain to put on eye protection, a mask andpuncture-resistant nitrile gloves.
Weekly Maintenance
Clean In-Line Filter
Turn OFF the pump motors.
Carefully unscrew the lower bowl from the filtertop and lift out the screen.
Using water, flush the bowl assembly and anycontaminated sediment.
Submerge the screen/bowl assembly into a high-level chemical disinfectant solution. Follow thedisinfectant manufacturer's recommendation for timeinterval required to achieve disinfection.
Remove the screen/bowl assembly from thedisinfectant solution and rinse using tap water.
Properly dispose of the contaminated sedimentand disinfectant solution waste.
Replace the screen and make sure the gasket is inplace before replacing the bowl.
Check all connecting vacuum, waste and waterlines for tightness.
Inspect Operatory FiltersCheck and clean all operatory and secondaryfilters weekly. Follow the equipment manufacturer'srecommendations.
Inspect Vacuum SystemCheck the system weekly for water leaks and looseor broken connections.
Monthly Maintenance
Check Vacuum Level
The vacuum level of the CustomAir Single WetVacuum Pump is preset at the factory.
Recommended Operating Vacuum Levels:10" Hg. General Dentistry19" Hg. Surgery
The pressure gauge will indicate the vacuum levelprovided by the pump. If it becomes necessary toadjust this setting, the following steps can betaken:
NOTE: The evacuator should be left ON whensetting the vacuum level. Also, make sure all hosesin the operatory are closed.
Turn the lock nut counterclockwise to loosen.
For greater vacuum level, turn the screwclockwise; for less vacuum level, turn screwcounterclockwise.
Set to desired vacuum level and tighten locknut.
Clean SurfaceUsing a damp cloth, remove any surface dust anddirt.
— CAUTION —A free-flowing discharge system is required forproper operation of the vacuum system. The
pump systems may leak at the anti-siphon valveif the discharge system is restricted. Ensure thevacuum system is installed and cleaned in accor-
dance to the instructions in this manual.
Technical Manual
Section IV Vacuum PumpsPage iv-14
MC-201 FS / MC-202 FS
MAINTENANCE
21Dual Wet Vacuum Pump
Section VI Operation & Care
— WARNING —Before starting cleaning procedures, make
certain to put on eye protection, a mask andpuncture-resistant nitrile gloves.
Weekly Maintenance
Clean In-Line Filter
Turn OFF the pump motors.
Carefully unscrew the lower bowl from the filtertop and lift out the screen.
Using water, flush the bowl assembly and anycontaminated sediment.
Submerge the screen/bowl assembly into a high-level chemical disinfectant solution. Follow thedisinfectant manufacturer's recommendation for timeinterval required to achieve disinfection.
Remove the screen/bowl assembly from thedisinfectant solution and rinse using tap water.
Properly dispose of the contaminated sedimentand disinfectant solution waste.
Replace the screen and make sure the gasket is inplace before replacing the bowl.
Check all connecting vacuum, waste and waterlines for tightness.
Inspect Operatory FiltersCheck and clean all operatory and secondaryfilters weekly. Follow the equipment manufacturer'srecommendations.
Inspect Vacuum SystemCheck the system weekly for water leaks andloose or broken connections.
Monthly Maintenance
Check Vacuum Level
The vacuum level of the CustomAir Dual WetVacuum Pump is preset at the factory.
Recommended Operating Vacuum Levels:10" Hg. General Dentistry19" Hg. Surgery
The pressure gauge will indicate the vacuum levelprovided by the pump. If it becomes necessary toadjust this setting, the following steps can betaken:
NOTE: Turn On one pump — vacuum leveladjustments are made with only one pump operative.
Turn the lock nut counterclockwise to loosen.
For greater vacuum level, turn the screwclockwise; for less vacuum level, turn screwcounterclockwise.
Set to desired vacuum level and tighten locknut.
NOTE: The evacuator should be left ON whensetting the vacuum level. Also, make sure all hosesin the operatory are closed.
Clean SurfaceUsing a damp cloth, remove any surface dust anddirt.
Section IV Vacuum Pumps
Technical Manual
iv-Page 15
AM-200
AIR/
WATER
SEPARATOR
Model: AM-200Air/Water or Nitrous Oxide Separator
(PN: 64656200)
Compatible with all single ordual water injected vacuumpumps. Parts provided for unitto be adapted to 3/4", 1" or1-1/2" exhausts.
2500 Highway 31 SouthBay Minette, Alabama 36507
1-866-DTE-INFO
Air discharge vent to outsidesupplied by plumber
3/4" or1"
Exhaust
1-1/2"Exhaust
CleanOutPlugWater to
Drain
PN: 64710078C
A standard 2", no-hub coupling onthe air discharge vent eliminatesany adaptor and allows directhookup to a standard vent pipe.
The water trap is inside, whichmakes plugging virtually impossible.
The entire separator is made ofcorrosion-proof PVC plastic.
The separator should bemounted near the vacuum pumpwith the center of the exhaustsinlets on the chamber no morethan 2 feet above the bottom ofthe vacuum pump.
NOTE: This product should beinstalled by a competent technicianor plumber. The installation shouldmeet all local health and plumbingcodes. This device requires a separateatmospheric vent for maximumsafety and effectiveness.
Technical Manual
Section IV Vacuum PumpsPage iv-16
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
How is a single pump wired to a remote lighted switch?
IMPORTANT NOTE:If not using a DentalEZMaster Control Panel Kit,proceed as follows:
1. Position the masterswitch in the desiredlocation.
2. Run the appropriate wiresto their respectiveeuipment location. (Seewiring chart below.)
3. Connect wires accordingto the wiring diagram onthe right.
WIRING CHART
— WARNING —Before proceeding with any electrical installation, comply with and
maintain all applicable local electrical code(s) and regulations.
— NOTICE —All wiring between the main control box and equipment should be Class B, low-
voltage. In most cases, a conduit is not required when using this type of wiring.
CV-101 3 18 AWG 16 AWGCV-102 3 18 AWG 16 AWG
Model No. No. of W ires W ire Under 150 ft. W ire Over 150 ft.
Wall-Mounted Switches Shown from
Back of Switch Plate Vacuum
Junction Box
Line Voltage(To Source)
Load(To Vacuum Pump)
Vacuum PumpRelay Box
Brown
Yellow
Black
L1
L2 T2
T1
LEGENDBrown = LightYellow = ReturnBlack = Power
FAQ #1
Section IV Vacuum Pumps
Technical Manual
iv-Page 17
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
L1
L2 T2
T1
Wall-Mounted Switch
Shown from Back of
Switch Plate
Junction Box
Vacuum PumpRelay Box
Brown
YellowBlack
ON/OFF Switches Mounted on Dual Frame
Vacuum
Brown
YellowBlack
Line Voltage(To Source)
Load(To Vacuum
Pump)
L1
L2 T2
T1
Line Voltage(To Source)
Load(To Vacuum
Pump)
Ret
urn
Light Common
Pow
er
Light Common
Vacuum PumpRelay Box
MC-201 6 18 AW G 16 AW GMC-202 6 18 AW G 16 AW G
How are dual pumps wired to a remote lighted switch?
IMPORTANTNOTE: If notusing a DentalEZMaster ControlPanel Kit, proceed asfollows:
WIRING CHART
— WARNING —Before proceeding with any electrical installation, comply with and
maintain all applicable local electrical code(s) and regulations.
— NOTICE —All wiring between the main control box and equipment should be Class B, low-
voltage. In most cases, a conduit is not required when using this type of wiring.
Model No. No. of Wires Wire Under 150 ft. Wire Over 150 ft.
LEGENDBrown = LightYellow = ReturnBlack = Power
1. Position themaster switchin the desiredlocation.
2. Run theappropriatewires to theirrespectiveeuipmentlocation. (Seewiring chartbelow.)
3. Connectwiresaccording tothe wiringdiagram onthe right.
FAQ #9
Technical Manual
Section IV Vacuum PumpsPage iv-18
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
Cle
an o
r re
plac
e ch
eck
valv
e on
the
rig
ht s
ide
of t
he in
take
man
ifold
.PN
: 645
6819
1, 3
/4"
PN: 6
4568
193,
1"
Why
is o
ne p
ump
wea
k?D
oes
the
dual
pum
p ha
vea
wat
er r
ecir
cula
tor?
Doe
s th
e pu
mp
have
suffi
cien
t w
ater
sup
ply?
Che
ck li
nes
from
rec
ircu
lato
rto
inta
ke li
nes.
Are
the
y cl
ogge
d?
Che
ck w
ater
flow
from
res
tric
tor
toth
e im
pelle
r ho
usin
g.Is
flow
suf
ficie
nt?
Hav
e el
ectr
icia
n pr
ovid
epr
oper
pow
er s
uppl
y.
Cor
rect
wat
er s
uppl
y:1
HP
= 1
/2 g
al. p
er m
in.
2 H
P =
1 g
al. p
er m
in.
Cle
an o
r re
plac
eflo
w r
estr
icto
rs.
PN: 6
4568
192,
1 H
PPN
: 646
2201
0, 2
HP
No
Yes
Cle
an li
nes
and
mak
e su
rere
circ
ulat
or c
lean
er is
used
onc
e a
mon
th.
With
the
left
pum
p ru
nnin
g,re
mov
e th
e in
take
hos
e fr
om t
heri
ght
pum
p. Is
vac
uum
pre
sent
?
Yes
No
Is t
he v
olta
ge t
o th
eva
cuum
pum
ps c
orre
ct?
Che
ck im
pelle
r ho
usin
gfo
r w
ear
or b
uild
up.
Yes
No
Yes
No
Yes
No
No
No
Yes
Yes
With
the
rig
ht p
ump
runn
ing,
rem
ove
the
inta
ke h
ose
from
the
left
pum
p.Is
vac
uum
pre
sent
?
Cle
an o
r re
plac
e ch
eck
valv
e on
the
left
sid
e of
the
inta
ke m
anifo
ld.
PN: 6
4568
191,
3/4
"PN
: 645
6819
3, 1
"
FAQ
#11
Section IV Vacuum Pumps
Technical Manual
iv-Page 19
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
FAQ #12Page 1 of 2
Technical Manual
Section IV Vacuum PumpsPage iv-20
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
FAQ #12Page 2 of 2
Section IV Vacuum Pumps
Technical Manual
iv-Page 21
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
How is a new transformer wired in an old relay box?(V Series)
GR
EEN
&Y
ELLO
W
LIG
HT
BLU
E
TER
MIN
AL
BLO
CK
(646
1502
3)
RED
BRO
WN REL
AY
(645
6813
3)
TR
AN
SFO
RM
ER(6
4637
018)
RED
SPA
RE
FUSE
HO
LDER
(645
6100
5)
RPL
. 115
V F
USE
(64
5681
31)
RPL
. 230
V F
USE
(64
5681
30)
FUSE
HO
LDER
(645
6816
0)
BLA
CK
(646
2508
3)BR
OW
N(6
4625
085)
PUR
PLE
(646
2507
3)
YEL
LOW
(646
2508
4)
BRO
WN
&W
HIT
E(6
5625
074)
BLA
CK
(646
2507
7)
BLU
E (6
4625
063)
YEL
LOW
(646
2508
2)
TER
MIN
AL
BLO
CK
JUM
PER
(64
6150
21)
BRO
WN
(646
2508
9)
CIR
CU
ITBR
EAK
ER
RPL.
15
AM
P B
REA
KER
(380
2-12
0)RP
L. 8
AM
P B
REA
KER
(380
1-91
1)
65
43
02
1
RED
WHITEBLUEBROWNBLACK
65
43
21
Yellow
Transformer
Fuse Holder
Alt. VoltageFuse Holder
Brown/White
Purple
Black
TerminalBlock
Black
Black
Blue
Yellow
Black/White
Brown
Red
White
Green
Relay
Blue
Black
Yellow
Brown
Yellow
New Transformer Old Relay Box
FAQ #13
Technical Manual
Section IV Vacuum PumpsPage iv-22
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
Why
wo
n't
the
vacu
um r
elay
eng
age
and
the
pum
p do
esn'
t tu
rn o
n?St
artin
g Le
tter
of S
eria
l #A
, C, D
, S
or V
AC
Vol
tage
acr
oss
2ye
llow
wir
es o
n re
lay
read
s <
20V
or
>20
V
AC
Vol
tage
acr
oss
2ye
llow
wir
es o
n re
lay
read
s >
17V
or
<17
V
Rep
lace
rel
ayPN
: 645
6813
3
Rep
lace
tra
nsfo
rmer
PN: 6
4568
195
Hav
e el
ectr
icia
nin
stal
l a b
uck/
boos
ttr
ansf
orm
er(2
if a
dua
l sys
tem
)
AC
Vol
tage
acr
oss
light
blu
e an
d br
own
wir
es o
n le
ft s
ide
of r
elay
is <
200V
on
230V
pum
p or
<10
0V o
n 11
5V p
ump
AC
Vol
tage
acr
oss
blac
k an
d pu
rple
wir
es o
n le
ft s
ide
of r
elay
is<
208V
on
230V
pum
p or
<11
0V o
n 11
5V p
ump
C, D
, S o
r VA
>17V
<17V
>20V
<20V
Yes
Yes
No
No
FAQ
#23
Section IV Vacuum Pumps
Technical Manual
iv-Page 23
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
Che
ck li
ne fr
omth
e re
circ
ulat
or.
Is it
clo
gged
?
Why
do
es t
he v
acuu
m p
ump
shut
dow
n, c
oo
l the
n re
star
t?D
oes
pum
p ha
vepr
oper
vol
tage
? Hav
e el
ectr
icia
npr
ovid
e pr
oper
vol
tage
.D
oes
pum
p ha
vesu
ffici
ent
wat
er s
uppl
y?
Che
ck s
haft
rota
tion.
Isit
smoo
th?
Rep
lace
pum
p. P
robl
emis
in w
indi
ngs
of m
otor
.
Che
ck w
ater
from
flow
rest
rict
or. I
s th
ere
suffi
cien
t w
ater
?1
HP
= 1
/2 g
al. p
er m
in.
2 H
P =
1 g
al. p
er m
in.
Do
low
erre
build
.*
No
Yes
Doe
s pu
mp
have
a re
circ
ulat
or?
Hav
e pl
umbe
r su
pply
prop
er w
ater
req
uire
men
ts.
Yes
No
Rep
lace
mot
or.*
Che
ck im
pelle
r ho
usin
gfo
r w
ear
or b
uild
up.
Is t
here
wea
r or
bui
ld u
p?
Yes
No
Yes N
oYe
s
No
No
No
Yes
Yes
* Fo
r pa
rt #
info
rmat
ion
cont
act T
ech
Serv
ice a
t 1-8
66-D
TE-IN
FO.
Cle
ar c
log
and
mak
esu
re r
ecir
cula
tor
clea
ner
is u
sed
once
a m
onth
.
Cle
an o
r re
plac
eflo
w r
estr
icto
rs.
PN: 6
4568
192,
1 H
PPN
: 646
2201
0, 2
HP
FAQ
#29
Technical Manual
Section IV Vacuum PumpsPage iv-24
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
Rep
lace
pum
pm
otor
or
pum
pca
rtri
dge.
*
Why
do
es p
ump
mak
e sq
ueal
ing/
whi
stlin
g so
und?
Che
ck V
olta
geIs
it c
orre
ct?
Cor
rect
ele
ctri
cal
prob
lem
s.R
emov
e in
take
hos
e fr
om p
ump.
Doe
s no
ise
go a
way
?
Cor
rect
wat
erflo
w p
robl
ems.
Rec
onne
ct in
take
hos
e.T
hen
disc
onne
ct t
he h
ose
to t
he v
acuu
m r
elie
f val
ve.
Doe
s no
ise
go a
way
?
No
Yes
Insp
ect
chec
kva
lve.
Is it
wor
king
pro
perl
y?
Man
ually
rot
ate
the
pum
p sh
aft.
Doe
s th
e sh
aft
feel
rou
gh?
Yes
No
Rep
lace
che
ck v
alve
.3/
4" -
PN
: 645
6819
11"
- P
N: 6
4568
193
Cal
l Tec
hnic
alSe
rvic
e.
Yes
No
Yes
No
No
No
Yes
Yes
Che
ck w
ater
flow
to
pum
p. Is
it c
orre
ct?
Che
ck o
r re
plac
eva
cuum
rel
ief v
alve
.PN
: 646
2200
1
* Co
ntac
t Tec
h Se
rvice
at 1
-866
-DTE
-INFO
.
FAQ
#32
Section IV Vacuum Pumps
Technical Manual
iv-Page 25
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONSW
hy d
oes
wat
er le
ak fr
om
the
ant
i-si
pho
n va
lve?
Sing
le o
rdu
al p
ump?
Is p
ump
runn
ing?
Are
bot
h an
ti-si
phon
valv
es le
akin
g?
Rep
lace
or
rebu
ildan
ti-si
phon
val
ve.
Reb
uild
kit
- PN
: 645
6801
6N
ew V
alve
- P
N: 6
4622
011
Dua
lS
ingl
e
Turn
pum
p of
f. D
isco
nnec
t dr
ain
line
and
let
exha
ust
into
buc
ket.
Turn
pum
p on
. Doe
s le
akin
g st
op?
No
Che
ck fl
ow t
hrou
gh fl
owre
stri
ctor
s. Is
it c
orre
ct?
1 H
P -
1/2
GPM
2 H
P -
1 G
PM1
HP
w/r
ecir.
- .1
3 G
PM2
HP
w/r
ecir.
- .2
5 G
PM
Are
pum
ps r
unni
ng?
Yes
Yes
No
No
No
Yes
Rep
lace
wat
er s
olen
oid
valv
e.24
V- P
N: 6
4568
156
115V
- P
N: 6
4622
084
220V
- P
N: 6
4622
085
Yes
Yes
Hav
e pl
umbe
r cl
ear
obst
ruct
ion
in d
rain
line
.R
epla
ce fl
ow r
estr
itor.
1 H
P -
PN: 6
4568
192
2 H
P -
PN: 6
4622
010
1 H
P w
/rec
ir. -
PN
: 646
2201
82
HP
w/r
ecir.
- P
N: 6
4622
019
No
No
Yes
No
Yes
Is p
ump
runn
ing?
Turn
pum
p/s
off.
Dis
conn
ect
drai
n lin
e an
d le
t ex
haus
tin
to b
ucke
t. Tu
rn p
ump/
s on
. Doe
s le
akin
g st
op?
Rep
lace
wat
er s
olen
oid
valv
e.24
V- P
N: 6
4568
156
115V
- P
N: 6
4622
084
220V
- P
N: 6
4622
085
FAQ
#33
Technical Manual
Section IV Vacuum PumpsPage iv-26
Vacuum Pumps
FREQUENTLY
ASKED
QUESTIONS
Is t
he in
-line
filte
r cl
ean?
Why
is v
acuu
m fr
om
dua
l pum
p w
eak?
Are
wat
er li
nes
from
rec
ircu
lato
r to
inta
ke li
nes
clog
ged?
Are
bot
hpu
mps
run
ning
? Is t
here
a s
uffic
ient
am
ount
of
fres
h w
ater
goi
ng in
to t
he p
umps
?1
HP
= 1
/2 g
al. p
er m
in. p
er p
ump
2 H
P =
1 g
al. p
er m
in. p
er p
ump
No
Cle
an o
r re
plac
eth
e ch
eck
valv
e.PN
: 645
6819
1 -
3/4"
che
ck v
alve
PN: 6
4568
193
-1"
che
ck v
alve
Rem
ove
the
inta
ke h
ose
from
the
pum
p th
at is
not
runn
ing.
Is t
here
vac
uum
pres
ent
on t
hat
hose
?
Yes
No
Yes
No
Yes
No
No
No
Con
tact
Tec
hnic
alSe
rvic
e.1-
866-
DT
E-IN
FO
No
Yes
Yes
Cle
an o
r re
plac
e th
ein
-line
filte
r sc
reen
.
Loca
te a
nd r
epai
r le
aks
in t
he o
pera
tory
line
.
Cle
an b
lock
ages
/cl
ear
rest
rict
ions
.
Doe
s th
e du
alpu
mp
have
aw
ater
rec
ircu
lato
r?
Yes
Yes
Cle
an c
logg
edlin
es a
nd c
lean
out
wat
erre
cicu
lato
r. D
oes
vacu
um in
crea
se?
Any
res
tric
tions
in t
he e
xaus
t lin
e?
FAQ
#43
Rem
ind
offic
est
aff t
he im
port
ance
of u
sing
rec
ircu
lato
rcl
eane
r(P
N: 6
4568
129)
once
a m
onth
.
No
Unh
ook
the
inta
kelin
e fr
om t
he o
pera
tory
line
and
seal
off
the
vacu
um r
elie
f muf
fler.
Doe
s va
cuum
incr
ease
?
Is t
he a
mou
nt o
f fre
sh w
ater
goin
g in
to t
he p
umps
cor
rect
?1
HP
= .1
3 ga
l. pe
r m
in. p
er p
ump
2 H
P =
.25
gal.
per
min
. per
pum
p
Cle
an o
r re
plac
eth
e flo
w r
estr
icto
rs.
PN: 6
4568
192
-1
HP
=1/
2 ga
l. pe
r m
in.
PN: 6
4622
010
-2
HP
= 1
gal
. per
min
.
Cle
an o
r re
plac
eth
e flo
w r
estr
icto
rs.
PN: 6
4622
018
- 1
HP
PN: 6
4622
019
- 2
HP
NO
TE
: If t
he r
ecirc
ulat
orw
as fi
eld
inst
alle
d,th
ere
may
be
2 flo
wre
stric
tors
on
each
Pum
p.
Cle
an o
r re
plac
eth
e flo
w r
estr
icto
rs.
PN: 6
4622
018
- 1
HP
PN: 6
4622
019
- 2
HP
NO
TE
: If t
he r
ecirc
ulat
orw
as fi
eld
inst
alle
d,th
ere
will
be 2
flow
rest
ricto
rs o
n th
e un
it.
Any
blo
ckag
es in
the
drai
n lin
e?
Yes
No
Yes
No
Yes
No
Yes
Section IV Vacuum Pumps
Technical Manual
iv-Page 27
Wet Vacuum Pumps
TROUBLESHOOTING
Installation, Operation and Care Manual
Single Wet Vacuum Pump
20
motpmyS )s(esuaCelbissoP noituloS
lliwpmuPnurton
nekorbroesoolrofseriwllakcehCwolkcehC.esufkcehC.snoitcennoc
sdaelegatlov )seriwkcalbdnawolley(.stlov42rof
fieesotgnihctawffodnanorewopnruT.ylreporpsetareporabrekaerbyaler
ssarbgninesoolybevlavdionelostseThtiwtuoswolfretawfI.edisthgirnotun
wolfretaW.fforewopnrut,norewop.potsdluohs :noituaC tonoD etarepo
esuaceb,emitdednetxenarofpmupehtdluocretawtuohtiwpmupehtgninnur
.egamadlanretniesuac
ehtecalper,tneserptonsiegatlovfI.esufroremrofsnart eesoslA(
).trahCsmelborPlacirtcelE
dna,ylreporpgnitarepotonyalerfI,doogerastraprehtollagnimussa
.yalerehtecalper lacirtcelEeesoslA().trahCsmelborP
rewophtiwpotstonseodwolfretawfI.evlavdionelosehtecalper,ffo
spotsrotoMtonlliwro
trats
-woldnaffo-tucniam,rekaerbtiucriCehtnierasehctiwsyrotarepoegatlov
FFO .noitisop
.nideggulptonsitinU
.seriwnekorbroesooL
.roticapacdabagnitacidnismuhrotoM
.rotomysionrothgiT
.gnitaehrevosirotoM
nisehctiwsecalP NO .noitisop
.tinunigulP
.seriwecalperroriaper,nethgiT
.roticapacecalpeR
sipmuperusekaM.sgniraebkcehC,rO.rotomehtotdemmihsylreporp
.pmupehtnisirbedynaevomer
sievlavfeilermuucavehterusekaMsahrotomehtdnaylreporpdetsujda
.ylppusretawdnanoitalitnevtneiciffus.egatlovenilwolrofkcehC
• Model Name
• Model Number
• Serial Number
• Installation Date
The following troubleshooting charts should be usedwhen attempting to isolate CustomAir Single WetVacuum Pump operational problems.
If the problem is not addressed in the trouble shoot-ing chart or cannot be isolated by performing thesuggested procedures, contact your local DentalEZfull-service dealership.
Service Instruction
Vacuum Pump
Be prepared to supply the following product informa-tion:
Section VII User Service Information
(On Page 29)
(On Page 29)
Technical Manual
Section IV Vacuum PumpsPage iv-28
Wet Vacuum Pumps
TROUBLESHOOTING
21Single Wet Vacuum Pump
motpmyS )s(esuaCelbissoP noituloS
tonpmuPgninnur
foesuacebtcepsus
lacirtcelemelborp
esufnwolbsahylppusrewopniaM.rekaerbtiucricro
ropmupgnihcaertonrewoP.egatlovtcerrocni
.xoblortnocehtniesufnwolB
ehtedisnisnoitcennocesooL.xoblortnoc
.rekaerbtiucricteserroesufnwolbecalpeR
laudehtfoedisehtnoxobnoitcennockcehCpmupehtgnihcaersirewopyfirevottenibac
.egatlovtcerrocehtsiylppusrewopniamdna
detaremasehthtiwesufehtecalpeR.yrotcafehtmorfenoehtsayticapac
yllausivdnarevocxoblortnocehtevomeR.snoitcennocesoolroftcepsni ehtotrefeR(
rofrevocxoblortnocehtedisnicitamehcstahtrofemehcsgniriwdnastnenopmoceht
).xobralucitrap
motpmyS )s(esuaCelbissoP noituloS
onrowoLmuucav
retlifytriD
tondeggolcylbmessalortnocretaW.pmuphguorhtogotretawgniwolla
enilmuucavnekorbroesooL.snoitcennoc
gniwolladeggolcsevlavkcehcgniwSrehtoehthguorhtnoitcusotpmup
.pmup
.tuonrowsipmuP
.ylppusretawtneiciffusnI
fisretlifyradnocesdnayrotareponaelC.yrassecen
pmupnruT FFO tneverpotyletaidemmiretawgolcnunehT.egamadlanretni
.ylbmessalortnoc
enilmuucavecalperroriaper,nethgiT.snoitcennoc
fi,dnaevlavehtfopotehtffoekaTeht,esiwrehtO.sirbedevomer,elbissop
.decalperebdluohsevlav
.noitcesriaperpmupotrefeR
ehtotstcennoctierehwylppuskcehClortnocretawehtkcehc,nehT.tenibacevlavdioneloslortnocretawehT.evlav
kcehC.nodenrutsipmupehtnehwsneporetawehtgnitcennocsidybnoitcnufsiht
ybdnagnisuohmuucavehtsretnetahtenil,nehT.evlavehtrednureniatnocagnidloh
aebdluohserehT.pmupehtnonrut.retawfomaertsydaets
TONOD:NOITUAC PMUPEHTEVAELsierehtfI!RETAWTUOHTIWGNINNURlortnocretawehtecalper,wolfretawon
.evlavdionelos
Section VII User Service Information
Vacuum Pump (Continued)
Electrical System
Section IV Vacuum Pumps
Technical Manual
iv-Page 29
21Single Wet Vacuum Pump
motpmyS )s(esuaCelbissoP noituloS
tonpmuPgninnur
foesuacebtcepsus
lacirtcelemelborp
esufnwolbsahylppusrewopniaM.rekaerbtiucricro
ropmupgnihcaertonrewoP.egatlovtcerrocni
.xoblortnocehtniesufnwolB
ehtedisnisnoitcennocesooL.xoblortnoc
.rekaerbtiucricteserroesufnwolbecalpeR
laudehtfoedisehtnoxobnoitcennockcehCpmupehtgnihcaersirewopyfirevottenibac
.egatlovtcerrocehtsiylppusrewopniamdna
detaremasehthtiwesufehtecalpeR.yrotcafehtmorfenoehtsayticapac
yllausivdnarevocxoblortnocehtevomeR.snoitcennocesoolroftcepsni ehtotrefeR(
rofrevocxoblortnocehtedisnicitamehcstahtrofemehcsgniriwdnastnenopmoceht
).xobralucitrap
motpmyS )s(esuaCelbissoP noituloS
onrowoLmuucav
retlifytriD
tondeggolcylbmessalortnocretaW.pmuphguorhtogotretawgniwolla
enilmuucavnekorbroesooL.snoitcennoc
gniwolladeggolcsevlavkcehcgniwSrehtoehthguorhtnoitcusotpmup
.pmup
.tuonrowsipmuP
.ylppusretawtneiciffusnI
fisretlifyradnocesdnayrotareponaelC.yrassecen
pmupnruT FFO tneverpotyletaidemmiretawgolcnunehT.egamadlanretni
.ylbmessalortnoc
enilmuucavecalperroriaper,nethgiT.snoitcennoc
fi,dnaevlavehtfopotehtffoekaTeht,esiwrehtO.sirbedevomer,elbissop
.decalperebdluohsevlav
.noitcesriaperpmupotrefeR
ehtotstcennoctierehwylppuskcehClortnocretawehtkcehc,nehT.tenibacevlavdioneloslortnocretawehT.evlav
kcehC.nodenrutsipmupehtnehwsneporetawehtgnitcennocsidybnoitcnufsiht
ybdnagnisuohmuucavehtsretnetahtenil,nehT.evlavehtrednureniatnocagnidloh
aebdluohserehT.pmupehtnonrut.retawfomaertsydaets
TONOD:NOITUAC PMUPEHTEVAELsierehtfI!RETAWTUOHTIWGNINNURlortnocretawehtecalper,wolfretawon
.evlavdionelos
Section VII User Service Information
Vacuum Pump (Continued)
Electrical System
Wet Vacuum Pumps
TROUBLESHOOTING
Installation, Operation and Care Manual
Single Wet Vacuum Pump
22
115 Volt PumpFuse replacement: Use BUSS MDL or MDQ 3/10 A,Littlefuse 313 3/10 A Slo-blo.
230 Volt PumpFuse replacement: Use BUSS MDL or MDQ 15/100 A,Littlefuse 313 15/100 A Slo-blo.
motpmyS )s(esuaCelbissoP noituloS
seodrotoMybtratston
hctiws
ehtniliocroremrofsnartevitcefeD.yaler
ehtnihsup,ecivedevitcudnoc-nonagnisUlliwrotomehtfienimretedotyalerehtnosbat
,tnempiuqetsetlacirtcelegnisunehT.tratsyradnocesehtfotuognimocegatlovehtyfirev
,V12nahtsselsitifI.remrofsnartehtfoedis,V12nahteromsitifI.remrofsnartehtecalper
.yalerehtecalper
Electrical System (Continued)
Section VII User Service Information
Relay Box
Fuse Replacement
— WARNING —For continued protection against fire
hazard: Correct fuse must be used.
GREEN&YELLOW
LIGHT BLUE
TERMINALBLOCK
(64615023)
RED
BROWN
RELAY(64568133)
TRANSFORMER(64637018)
RED
SPAREFUSE
HOLDER(64561005)
RPL. 115V FUSE (64568131)RPL. 230V FUSE (64568130)
FUSEHOLDER
(64568160)
BLACK(64625083)
BROWN(64625085)
PURPLE(64625073)
YELLOW(64625084)
BROWN&WHITE(65625074)
BLACK(64625077)
BLUE (64625063)
YELLOW(64625082)
TERMINAL BLOCKJUMPER (64615021)
BROWN(64625089)
CIRCUITBREAKER
RPL. 15 AMP BREAKER (3802-120)RPL. 8 AMP BREAKER (3801-911)
6 5 4 3 02 1
RED
PN: 64529265A
WH
ITE
BLUE
BRO
WN
BLAC
K
Technical Manual
Section IV Vacuum PumpsPage iv-30
CV-101FS / CV102FS
SCHEMATICS
WIRING
23Single Wet Vacuum Pump
VacuumFilter
Relief Valve
Intake ManifoldGauge
Pump
Water
Wiring Schematic
Air/Water Flow Circulation Diagram
The Dual Wet Vacuumelectrical control system isa low-voltage (24V) circuit.This system also providesautomatic control of thewater supply system. Thewiring schematic shows forelectrical rough-in. (Refer tothe Installation section ofthis manual.) (See Pages 3and 4 for further informa-tion on rough-in and finalhook-up locations.)
The circulation diagramshows the relationship ofall the major assemblies tothe total system. All con-necting lines are markedwith a symbol to indicatetheir function: vacuum line,water line or waste line.The assemblies are indi-vidually broken down andfurther explained in othersections of this manual.
Section VII User Service Information
.oN .ytQ .oNtraP noitpircseD1 2 53186546 retaW,retliF2 2 65186546 )ylnO(dioneloS,evlaV3 2 11022646 nohpiS-itnA,evlaVA4 2 29186546 PH1--rotalugeRwolF,evlaVB4 2 01022646 PH2--rotalugeRwolF,evlaV
.oN .ytQ .oNtraP noitpircseD5 1 04054546 muucaV,retliF6 1 95186546 feileRmuucaV,relffuM7 1 10022646 feileRmuucaV,evlaV
1
7
23
4
65
Relay
Coil
Switch 24V
Transformer
Primary
Fuse 1Relay 1
OL
L1
G
Relay 2L2G
Motor1
Section IV Vacuum Pumps
Technical Manual
iv-Page 31
MC-201 FS / MC-202 FS
SCHEMATICS
25Dual Wet Vacuum Pump
Water
Vacuum
Waste
Relief Valve
Intake ManifoldGauge
Exhaust Manifold
Pump1
Pump2
Wiring Schematic
Air/Water Flow Circulation Diagram
The Dual Wet Vacuumelectrical control system isa low-voltage (24V) circuit,designed to provide twototally independent controlsystems for each pumpunit. This system alsoprovides automatic controlof the water supply systemfor each pump. The wiringschematic shows forelectrical rough-in. (Refer tothe Installation section ofthis manual.) (See Pages 3and 4 for further informa-tion on rough-in and finalhook-up locations.)
The circulation diagramshows the relationship ofall the major assemblies tothe total system. All con-necting lines are markedwith a symbol to indicatetheir function: vacuum line,water line or waste line.The assemblies are indi-vidually broken down andfurther explained in othersections of this manual.
Section VII User Service Information
.oN .ytQ .oNtraP noitpircseD
1 2 21022646 ffotuhSretaW,evlaV
2 2 19186546 kcehCgniwS,evlaV3 2 A11022646 nohpiS-itnA,evlaVA4 2 29186546 PH1--rotalugeRwolF,evlaV
B4 2 01022646 PH2--rotalugeRwolF,evlaV
.oN .ytQ .oNtraP noitpircseD
5 2 65186546 )ylnO(dioneloS,evlaV
6 2 53186546 retaW,retliF7 1 95186546 feileRmuucaV,relffuM8 1 10022646 feileRmuucaV,evlaV
17
5
81
2
3
2
43
4
65
6
Relay 1
Relay 2
Solenoid 1
Switch 1
Solenoid 2
Switch 2
24V
24V
Transformer 1
Transformer 2
Primary
Primary
Fuse 1
Fuse 2
Relay 1A
OL
L1
G
Relay 1BL2G
L3
Motor1
Relay 2A
OL
Motor2 G
Relay 2BL4
WIRING
Technical Manual
Section IV Vacuum PumpsPage iv-32
Wet Vacuum Pumps
SCHEMATICS
WIRING
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
8
Section III Installation
IMPORTANT NOTE: If not using a DentalEZ Master Control Panel Kit,proceed as follows:
1. Position the master switch in the desired location.
2. Run the appropriate wires to their respective euipment location. (See wiring chartbelow.)
3. Connect wires according to the wiring diagrams on the next two pages.
Remote Low-Voltage Switch Control
WIRING CHART
— WARNING —Before proceeding with any electrical installation, comply with and
maintain all applicable local electrical code(s) and regulations.
— NOTICE —All wiring between the main control box and equipment should be Class B, low-
voltage. In most cases, a conduit is not required when using this type of wiring.
CV-101 3 18 AWG 16 AWGCV-102 3 18 AWG 16 AWGDV-301 3 18 AWG 16 AWGDV-302 3 18 AWG 16 AWGMC-201 6 18 AWG 16 AWGMC-202 6 18 AWG 16 AWG
Air Compressor 3 18 AWG 16 AWGMWCV Solenoid Valve 2 18 AWG 16 AWGMWCV Transformer 2 18 AWG 16 AWG
Model No. No. of Wires Wire Under 150 ft. Wire Over 150 ft.
Section IV Vacuum Pumps
Technical Manual
iv-Page 33
MC-201 FS / MC-202 FS
SCHEMATICS
WIRING
9Dual Wet Vacuum Pump
Section III Installation
Two Switch, Dual Vacuum System
IMPORTANT NOTE: Refer to wiring chart on Page 8 for length and gauge of wire.
Wall-Mounted Switches
Shown from Back of
Switch Plate(CustomAir
MP400)
Vacuum 1
Junction Box
Vacuum PumpRelay Box
Brown
YellowBlack
ON/OFF Switches Mounted on Dual Frame
Vacuum 2
Brown
YellowBlack
Line Voltage(To Source)
Load(To Vacuum
Pump)
L1
L2 T2
T1
Vacuum PumpRelay Box
Line Voltage(To Source)
Load(To Vacuum
Pump)
L1
L2 T2
T1
Ret
urn
Ligh
t C
omm
on
Pow
er
Ret
urn
Ligh
t C
omm
on
Pow
er
32
Technical Manual
Section IV Vacuum PumpsPage iv-34
SCHEMATICS
WIRING
MC-201 FS / MC-202 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
10
One Switch, Dual Vacuum System
IMPORTANT NOTE: Refer to wiring chart on Page 8 for length and gauge of wire.
L1
L2 T2
T1
Wall-Mounted Switch
Shown from Back of
Switch Plate
Junction Box
Vacuum PumpRelay Box
Brown
YellowBlack
ON/OFF Switches Mounted on Dual Frame
Vacuum
Brown
YellowBlack
Line Voltage(To Source)
Load(To Vacuum
Pump)
L1
L2 T2
T1
Line Voltage(To Source)
Load(To Vacuum
Pump)
Ret
urn
Light Common
Pow
er
Light Common
Vacuum PumpRelay Box
32
Section IV Vacuum Pumps
Technical Manual
iv-Page 35
CV-101FS / CV102FS
SYSTEM
REPAIR
PROCEDURE
Installation, Operation and Care Manual
Single Wet Vacuum Pump
24
Section VII User Service Information
Tools Required:
• Wire Stripper/Crimper
• 1/4", 5/16", 11/32" and 3/8"Nut Driver
• 7/16", 1/2", 9/16" and 15/16"Open-end Wrench
• 1-3/4" Socket and Torque Wrench
• 15/16" Socket with Ratchet
• Hammer
• Flat-blade Screwdriver
• Wire Cutter
• Needle-nose Pliers
• Channel-lock Pliers
• Paint Scraper
• Pump Motor Holding Fixture:DTE# 64546001
• Pop Riveter
• Bench Vise
• Bearing Seal Lubricant
• Red Loctite/Pipe Sealant:Item# 57141
Dust Cover
If the motor does not start after checking through thetroubleshooting charts, the motor may be defective.If the pump must be removed for factory repair orreplacement, perform the following vacuum pumpsystem repair procedure:
Test Procedure — DiagnosticDetermine if the required repair procedure iselectrical or mechanical.
1. Unplug the electrical supply cord.
2. Using the manual valve, turn OFF the watersupply.
3. Remove the dust cover located at the topcenter of the motor.
Vacuum Pump System Repair Procedure
4. Using a flat-blade screwdriver, engage the slotin the drive shaft and rotate the shaft to checkfor free movement:• If the shaft moves smoothly, proceed to
Electrical Box Removal.• If the shaft is difficult to turn or is jammed,
proceed to Pump Removal/Disassembly.
Electrical Box Removal1. Remove the electrical box cover plate.
— WARNING —Do not attempt these repairs in the dental office.
Fuse
Electrical BoxCover Plate
Drive Shaft Slot
Technical Manual
Section IV Vacuum PumpsPage iv-36
CV-101FS / CV102FS
SYSTEM
REPAIR
PROCEDURE
25Single Wet Vacuum Pump
Section VII User Service Information
5. Remove the water line from the pump housing.
6. Remove the bolt from the bottom of theelectrical box where the bracket is attached.
7. Remove the two screws and nuts from theterminal board located at the top of theelectrical box (old style).
8. Remove the chase nipple from inside the boxwhere it screws to the motor.
Pump Removal/Disassembly1. Remove the motor cover plate (top of motor).
2. Shut OFF the valve supplying water to thepump.
3. Disconnect the vacuum and exhaust lines at thepump and install line closure plugs.
4. Disconnect the two motor wire leads from theelectrical terminal.
Wires
Water Line
Bolt
ChaseNipple
Vacuum Intake
Water Line
2. Check the wire leads to the motor for breaksor bad connections.
3. Check the fuse and check for loose or brokenwires in the electrical box.
4. Disconnect the wires from the top of themotor and pull through the box.
Pump MotorHolding Fixture
5. Disconnect the waste and input lines from thepump and cap off the lines using the closureplugs provided.
6. Disconnect the vacuum intake manifold.
7. Disconnect the water line at the pump.
8. Disconnect the low voltage wires from theremote switch circuit.
9. Mount the motor in the pump motor holdingfixture by aligning the slot in the motor's shaftto the holding blade. Then connect the holddown brackets and tighten so the motor willnot rotate in the fixture.
Two Screws
Section IV Vacuum Pumps
Technical Manual
iv-Page 37
CV-101FS / CV102FS
SYSTEM
REPAIR
PROCEDURE
Installation, Operation and Care Manual
Single Wet Vacuum Pump
26
2. Press the seal into the housing with theceramic side facing outward.
Ceramic Side
MotorShaft
Seal
3. Carefully install the housing onto the motorshaft.
4. Install the black spring seal with the nylonside in contact with the ceramic washer.
NOTE: Lubrication of shaft will easeinstallation.
5. Apply loctite to the threads of the motor shaft.
6. Screw the impeller onto the shaft with the flatside facing the motor.
7. Torque the impeller to 27 in. lbs.
8. Align the housing so that the electrical boxmounting hole is opposite of the small elbowin back of the pump housing.
Section VII User Service Information
10. Take off the base by removing the bolts thatsecure it to the housing. Then inspect the basefor excessive scoring or side wear.
Pump to Motor Assembly1. Apply lubricating oil to the motor shaft, seal
and housing.
Bolts
Impeller
Housing
12. Remove the housing bolts and take off thehousing by pulling it off the shaft.
13. Remove the water seal spring assembly theninspect the housing for pitting or scoring.
11. Remove the impeller by unscrewingcounterclockwise. Then inspect the impellerfor any pitting or scoring.
Technical Manual
Section IV Vacuum PumpsPage iv-38
CV-101FS / CV102FS
SYSTEM
REPAIR
PROCEDURE
27Single Wet Vacuum Pump
4. Install the bolts and required shims.
5. Tighten and check for free spin of theimpeller by removing the motor from thefixture and rotating the impeller. (For properfit, shims may need to be added or deleted.)
Section VII User Service Information
Shims(Left Side)
Shims(Right Side)
ShimsandBolt
Housing/Impeller Spacing1. To properly space the impeller to the housing,
slide two or three .031" shims between thehousing and the motor at one corner.
2. Slide other shims of various sizes diagonallyacross from the corner in Step 1, until theimpeller is snug with the housing.
3. Remove and add up all the sizes of the shimsused. EXAMPLE: A total of .100" is used,divide that by 2 to obtain .050". Use thisamount minus an additional .005" from 1 HPpumps and .010" from 2 HP for each corner.
NOTE: For further explanation, seeShimming Procedure.
SHIMMING PROCEDURE
A. Insert shims between the motor and thehousing at two opposite mounting legs untilsnug.
B. Remove and add up the total thickness ofthe shims from both sides.
EXAMPLE: Left Side + Right Side Shims Shims
.031 .031+.015 +.031+.015 +.005
.061 .067
.061 + .067 = .128
C. Divide the total by 2 and subtract:.005 for 1 HP motor.010 for 2 HP motor
EXAMPLE: .128 : 2 = .064 .064 .005 = .059
D. Use the best combination of shims for therequired shim dimension.
EXAMPLE: RequiredShim
Dimension .059
E. Install a bolt and equal shim combinationsat all four locations.
Combine Shims.031
+.015+.010+.002.058
Section IV Vacuum Pumps
Technical Manual
iv-Page 39
CV-101FS / CV102FS
SYSTEM
REPAIR
PROCEDURE
Installation, Operation and Care Manual
Single Wet Vacuum Pump
28
Base
Gasket
Section VII User Service Information
6. Install the base with the gasket by aligning theintake and discharge ports. Then tighten thebolts (do not over tighten).
7. Set the pump upright and insert a screwdriverinto the slot on top of the motor and checkfor free spin and smooth operation of thepump.
Vacuum Pump Operation Testing1. To check for proper operation of the vacuum
pump, connect to an electrical source, waterand waste line.
2. Start the pump, then block the suction side ofthe pump and the vacuum relief valve. Thevacuum level should read between 20-25 in. hg.Shims may need to be added or deleted forproper vacuum reading. (Refer to the Housing/Impeller Spacing Section.)
3. Set the vacuum level by unblocking thevacuum relief valve and adjusting the valve to10-12 in. hg.
4. Operate the pump for approximately onehour continuously.
5. Check for any water leaks, electrical problemsand consistent vacuum level.
(On previous page)
Technical Manual
Section IV Vacuum PumpsPage iv-40
MC-201 FS / MC-202 FS
SYSTEM
REPAIR
PROCEDURE
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
26
Section VII User Service Information
Vacuum Pump System Repair Procedure
Tools Required:
• Wire Stripper/Crimper
• 1/4", 5/16", 11/32" and 3/8"Nut Driver
• 7/16", 1/2", 9/16" and 15/16"Open-end Wrench
• 1-3/4" Socket and Torque Wrench
• 15/16" Socket with Ratchet
• Hammer
• Flat-blade Screwdriver
• Wire Cutter
• Needle-nose Pliers
• Channel-lock Pliers
• Paint Scraper
• Pump Motor Holding Fixture:DTE# 64546001
• Pop Riveter
• Bench Vise
• Bearing Seal Lubricant
• Red Loctite/Pipe Sealant:Item# 57141
Dust Cover
DriveShaftSlot
If the motor does not start after checking through thetroubleshooting charts, the motor may be defective.If the pump must be removed for factory repair orreplacement, perform the following steps:
1. Unplug the electrical supply cord.
2. Using the manual valve, turn OFF the watersupply.
3. Disconnect the water line to the pump beingremoved.
4. Disconnect the waste and input lines from thepump being removed and cap off the linesusing the closure plugs provided.
5. Disconnect the low voltage wires from theremove switch circuit.
6. Remove the wing nuts.
Pump Removal
Test Procedure — DiagnosticDetermine if the required repair procedure iselectrical or mechanical.
1. Disconnect the pump electrical cords fromthe receptacles.
2. At the needle valves, turn OFF the mainwater supply to the cabinet.
3. Remove the dust cover located at the topcenter of the motor.
4. Using a flat-blade screwdriver, engage the slotin the drive shaft and rotate the shaft to checkfor free movement:• If the shaft moves smoothly, proceed to
Electrical Box Removal.• If the shaft is difficult to turn or is jammed,
proceed to Pump Disassembly.
Section IV Vacuum Pumps
Technical Manual
iv-Page 41
MC-201 FS / MC-202 FS
SYSTEM
REPAIR
PROCEDURE
27Dual Wet Vacuum Pump
3. Check the fuse and check for loose or brokenwires in the electrical box.
4. Disconnect the wires from behind the face-plate of the cabinet and pull through the box.
5. Remove the two water lines from the watermanifold.
Section VII User Service Information
7. Remove the two screws and nuts from theterminal board located at the top of theelectrical box.
8. Remove the chase nipple from inside the boxwhere it screws to the motor.
Pump Disassembly1. Remove the motor cover plate (top of motor).
2. Remove the front plate from the cabinet.
3. Remove the nut from the vacuum relief valve.
4. Disconnect the three electrical box wires fromthe switch located behind the faceplate of thecabinet.
Motor Pump Assembly Removal1. Disconnect the pump's vacuum and exhaust
lines and install line closure plugs.
2. Disconnect the pump's water line.
Wires
Water Line
Water Line
2 Screws andNuts
Bolt
ChaseNipple
Vacuum Intake
HoseClamps
Water LineWing Nuts
Electrical Box RemovalNOTE: Most electrical repairs can be performedwithout removing the pump from the cabinet.
1. Remove the electrical box cover plate.
2. Check the wire leads to the motor for breaksor bad connections.
Fuse
Electrical BoxCover Plate
6. Remove the bolt from the bottom of the boxwhere the bracket is attached.
3. Shut OFF the valve at the water controlmanifold.
4. Remove the wing nuts from the pump's base.(four nuts - 2 HP pump; three nuts - 1 HP pump)
Technical Manual
Section IV Vacuum PumpsPage iv-42
MC-201 FS / MC-202 FS
SYSTEM
REPAIR
PROCEDURE
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
28
Section VII User Service Information
5. Disconnect the two motor wire leads from theelectrical terminal.
8. Remove the impeller by unscrewingcounterclockwise. Then inspect the impellerfor any pitting or scoring.
9. Remove the housing bolts and take off thehousing by pulling it off the shaft.
10. Remove the water seal spring assembly.
11. Inspect the housing for pitting or scoring.
Bolts
Impeller
Housing
Pump MotorHolding Fixture
7. Take off the base by Removing the bolts thatsecure it to the housing. Then inspect the basefor excessive scoring or side wear.
6. Mount the motor in the pump motor holdingfixture (DTE# 64546001) by aligning the slot inthe motor's shaft to the holding blade. Thenconnect the hold down brackets and tightenso the motor will not rotate in the fixture.
Section IV Vacuum Pumps
Technical Manual
iv-Page 43
SYSTEM
REPAIR
PROCEDURE
MC-201 FS / MC-202 FS
29Dual Wet Vacuum Pump
Pump to Motor Assembly1. Apply lubricating oil to the motor shaft, seal
and housing.
Housing/Impeller Spacing1. To properly space the impeller to the housing,
slide two or three .031" shims between thehousing and the motor at one corner.
Section VII User Service Information
Ceramic Side
MotorShaft
Seal
Shims(Left Side)
Shims(Right Side)
ShimsandBolt
2. Press the seal into the housing with theceramic side facing outward.
3. Carefully install the housing onto the motorshaft.
4. Install the black spring seal with the nylonside in contact with the ceramic washer.
NOTE: Lubrication of the motor shaft willease installation.
5. Apply loctite to the threads of the motorshaft.
6. Screw the impeller onto the shaft with the flatside facing the motor.
7. Torque the impeller to 27 in. lbs.
8. Align the housing so that the small elbow isopposite the electrical box mounting hole inthe motor.
2. Slide other shims of various sizes diagonallyacross from the corner in Step 1, until theimpeller is snug with the housing.
3. Remove and add up all the sizes of the shimsused. EXAMPLE: A total of .100" is used,divide that by 2 to obtain .050". Use thisamount minus an additional .005" from 1 HPpumps and .010" from 2 HP for each corner.
NOTE: For further explanation, seeShimming Procedure.(On next page)
Technical Manual
Section IV Vacuum PumpsPage iv-44
SYSTEM
REPAIR
PROCEDURE
MC-201 FS / MC-202 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
30
4. Install the bolts and required shims.
5. Tighten and check for free spin of theimpeller by removing the motor from thefixture and rotating the impeller. (For properfit, shims may need to be added or deleted.)
Base
Gasket
SHIMMING PROCEDURE
A. Insert shims between the motor and thehousing at two opposite mounting legs untilsnug.
B. Remove and add up the total thickness ofthe shims from both sides.
EXAMPLE: Left Side + Right Side Shims Shims
.031 .031+.015 +.031+.015 +.005
.061 .067
.061 + .067 = .128
C. Divide the total by 2 and subtract:.005 for 1 HP motor.010 for 2 HP motor
EXAMPLE: .128 : 2 = .064 .064 .005 = .059
D. Use the best combination of shims for therequired shim dimension.
EXAMPLE: RequiredShim
Dimension .059
E. Install a bolt and equal shim combinationsat all four locations.
Combine Shims.031
+.015+.010+.002.058
Section VII User Service Information
6. Install the base with the gasket by aligning theintake and discharge ports. Then tighten thebolts (do not over tighten).
7. Set the pump upright and insert a screwdriverinto the slot on top of the motor.
8. Check for free spin and smooth operation ofthe pump.
Vacuum Pump Operation Testing1. To check for proper operation of the vacuum
pump, connect to an electrical source, waterand waste line.
2. Start the pump, then block the suction side ofthe pump and the vacuum relief valve. Thevacuum level should read between 20 -25 in hg.Shims may need to be added or deleted forproper vacuum reading. (Refer to the Housing/Impeller Spacing Section.)
3. Operate the pump for approximately onehour continuously.
4. Check for any water leaks, electrical problemsand consistent vacuum level.
(On previous page)
Section IV Vacuum Pumps
Technical Manual
iv-Page 45
PARTS
LISTS
CV-101FS / CV102FS
5Single Wet Vacuum Pump
Installation Kit
Section III Installation
.oN .ytQ .oNtraP noitpircseD
1 1 50040546 THM4/3xPIM4/3nolyN,rotpadA
2 2 97040546 BRAB4/3xPIM4/3,nolyN,rotpadA
3 1 30140546 knahS1xTPN1,CVP,rotpadA
4 2 40140546 knahS1xTPN4/3CVP,rotpadA
5 1 21061546 PIF8/1xPIM2/1,gnihsuB
6 6 30072546 TSS"1,pmalC
7 2 15033546 PIF4/3xesoHleviwS,gnilpuoC
8 1 28014546 PIF1xpilS1,CVP,woblE
9 1 51114546 knahS1xTPN1,CVP,woblE
01 2 71114546 knahS1xTPN4/3CVP,woblE
11 .tf8 01026546 "1,raelC,esoH
21 1 41077546 "1,tsuahxE,relffuM
31 2 40097546 noisserpmoC4/1,tuN
41 1 93097546 4/1,straPgnibuTnoisserpmoC
51 .tf4 60081646 4/1,nolyNlarutaN,ebuT
61 1 21022646 eldeeN,4/1x8/1,elgnA,evlaV
71 4 20042646 rebbuRkcalB,esoH,rehsaW
Each vacuum pump is supplied with an installation kit (PN: 64568036).
132 4
56
7
8
9 1011
12
13
14
15
1617
Technical Manual
Section IV Vacuum PumpsPage iv-46
PARTS
LISTS
CV-101FS / CV102FS
7Single Wet Vacuum Pump
Section III Installation
The water going to the unit acts as a pump sealantand cooling agent. When the vacuum pump is inoperation, the water supply must be on at all times.There are two methods of installation:
1/8" MIP Angle Valve Installation
The plumber supplies the water line and installs a 1/2"brass gate valve on the water supply line. Connect the1/4" nylon tubing and slide the nut and brass ferruleover the tubing. Push the tubing in the valve as far aspossible. Make sure the tapered end of the ferrule isfacing the end of the tubing and tighten the nut.
1
2
3
4
5
6
.oN .ytQ .oNtraP noitpircseD
1 1 * eniLretaW
2 1 * "2/1,ssarB,etaG,evlaV
3 1 21061546 xPIM"2/1,ssarB,gnihsuBPIF"8/1
4 .tf4 60081646 "4/1,nolyN,ebuT
5 1 93097546 gnibuTsnoisserpmoC"4/1,straP
6 1 21022646 ,"4/1x"8/1,elgnA,evlaVeldeen
* Plumber Supplies
Water Supply Line
Saddle Valve Installation
A saddle valve may be tapped into an existing coldwater line. Follow the instructions on the saddle valvepackage.
1
2
3
4
5
67 8
.oN .ytQ .oNtraP noitpircseD
1 1 * eniLretaW
7-2 1 * ,"4/1,evlaVelddaStiKetelpmoC
8 .tf4 60081646 "4/1,nolyN,ebuT
Angle Valve Installation
Saddle Valve Installation
* Plumber Supplies
Section IV Vacuum Pumps
Technical Manual
iv-Page 47
PARTS
LISTS
CV-101FS / CV102FS
Installation, Operation and Care Manual
Single Wet Vacuum Pump
8
Section III Installation
Vacuum Line
Vacuum lines must be installed by a local plumberaccording to local building codes. All vacuum linesand risers are recommended to be IPS, PVC, SCH.40. Type “M” copper should be used if local codedoes not allow the use of PVC.
Care should be taken to slope the lines 1" for every20' of run toward the vacuum pump(s). This allowswaste and liquids to flow with gravity, contributing tothe efficiency of the vacuum system.
Make all connections using long radius sweep fittings.To promote unrestricted flow of air and waste liquids
through the vacuum lines, directional flow connec-tions should be used.
Using 45º elbows for turns or avoiding obstructions isbest; however, do not make a trap in the line, doing sowill decrease the efficiency of the system.
All elbows and tees should be sized for the main lineand sized down with bushing reducers to accommo-date smaller lines.
Avoid sagging lines, which cause the formation oftraps in the system and prevent good air and wasteliquid flow.
Connect the evacuation system to the vacuum lineusing the hose and fittings supplied in the installa-tion kit.
1
2
3
4
5
.oN .ytQ .oNtraP noitpircseD
1 * ,CVP,SPI,eniLmuucaV04.HCS
2 1 51114546 "1x"1,woblE
3 1 30140546 "1x"1,CVP,retpadA
4 1 30072546 "1,leetSsselniatS,pmalC
5 .tf4 01026546 "1,ekatnI,esoH
— IMPORTANT —All vacuum systems must be installed accord-
ing to local building and electrical codes.
* Supplied by Plumber
Technical Manual
Section IV Vacuum PumpsPage iv-48
PARTS
LISTS
CV-101FS / CV102FS
9Single Wet Vacuum Pump
Section III Installation
Waste Line
— IMPORTANT NOTE —No part of the waste line should be morethan three (3) feet above the level of thewaste connection on the vacuum pump.
The waste line carries water from the pumps andliquid waste from the operatory to the building'ssewer system.The waste line should follow the most direct path tothe sewer connection with a minimum of bends andelevations, and must be installed according to localbuilding and plumbing codes.The exhaust connection should be made by either oftwo methods, floor sink connection or direct connec-tion to “P” trap, depending on local code and build-ing facilities.
1
2
3
4
Floor Sink Connection
Direct Connection to "P" Trap
12
Use floor sink adapter SA-100. Install as illustrated.
Use “P” trap-air gap, if required by local code. Installas illustrated.
.oN .ytQ .oNtraP noitpircseD
*1 1 ** "1,kniSroolF,CVP,retpadA
*2 1 20140546 "1x.SNI"1,CVP,retpadA
3 1 30072546 "1,leetSsselniatS,pmalC
4 .tf3 01026546 "1,tsuahxE,esoH
5 1 rebmulPseilppuS kniSroolF
.oN .ytQ .oNtraP noitpircseD
1 1 01026546 "1,tsuahxE,esoH
2 1 30072546 "1,leetSsselniatS,pmalC
3 1 * tuN/x"1,ssarB,woblE
4 1 * PIF"1xPIM"1,CVP,gnihsuB
5 1 * "2,paGriA,retpadA
6 1 * ,"2/1-1x"2,lleB,recudeR.vlaG
7 1 * .vlaG,esolCx"2/1-1,elppiN
8 1 * "2/1-1,”P”,parT* Floor Sink Adapter Parts* Supplied by Plumber
1
* Supplied by Plumber
3
4
5
6
7
8
Section IV Vacuum Pumps
Technical Manual
iv-Page 49
PARTS
LISTS
CV-101FS / CV102FS
Installation, Operation and Care Manual
Single Wet Vacuum Pump
12
1 HP Single Pump Assembly
.oN .ytQ .oNtraP noitpircseD
1 10168546 PH1,caViH
1 1 R05110546 V032,V511,tinUyaleR
2 1 69186546 nollag2/1,dlofinaMretaW
3 1 95057546 SF,PH1,rotoM
4 1 64186546 "2/1x"4/3,ssarB,rotpadA
5* sA.qeR
1004064620040646300406464004064650040646
kcihT500.leetS,mihSkcihT130.leetS,mihSkcihT510.leetS,mihSkcihT200.leetS,mihSkcihT010.leetS,mihS
6* 4 820-5161 "1x61--"8/3,wercSpaC
7* 1 99186546 PH1,ssarB,gnisuoH
8* 1 72186546 rebiF,teksaGgnisuoH
9* 1 32186546 yratoR,laeSgnisuoH
01* 1 89186546 PH1,ssarB,rellepmI
11* 1 86186546 PH1,ssarB,esaB
21* 4 61042646 "61/5,ratSlanretnI,rehsaW
31* 4 44011646 "1x81-"61/5,paC,wercS
31-5* 1 52186546 PH1,tiKdliubeRrewoL
41 1 46186546 eguaG/w,ekatnI,dlofinaMevlaVdna
51 2 40097546 ,ebuT"4/1,tuNnoisserpmoC
61 2 20001646 noisserpmoC,"4/1,eveelS
71 1 55186546 "4/1x"4/1,woblE
81 1 30086546 PH1,tiKteeFgnitnuoM
1 45041546 otxoBlacirtcelE,tekcarBrezilibatS,rotoM
NOTE: See the Operation/Maintenance section ofthis manual for detailed information on hookup andcare of the Single Wet Vacuum Pump.
1
67
18
2
3
5
4
9
10
14
15
8
Section IV Illustrated Parts Breakdown
16
17
11
12
13
Technical Manual
Section IV Vacuum PumpsPage iv-50
PARTS
LISTS
CV-101FS / CV102FS
13Single Wet Vacuum Pump
2 HP Single Pump Assembly
Section IV Illustrated Parts Breakdown
.oN .ytQ .oNtraP noitpircseD
1 20168546 PH2,caViH
1 1 R05110546 V032,tinUyaleR
2 1 79186546 nollag1,dlofinaMretaW
3 1 06057546 SF,PH2,rotoM
4 1 74186546 "2/1x"4/3,ssarB,rotpadA
5* sA.qeR
1004064620040646300406464004064650040646
kcihT500.leetS,mihSkcihT130.leetS,mihSkcihT510.leetS,mihSkcihT200.leetS,mihSkcihT010.leetS,mihS
6* 4 820-5161 "1x61-"8/3,wercSpaC
7* 1 00286546 PH2,ssarB,gnisuoH
8* 1 82186546 rebiF,teksaGgnisuoH
9* 1 42186546 yratoR,laeSgnisuoH
01* 1 98186546 PH2,ssarB,rellepmI
11* 1 86186546 PH2,ssarB,esaB
21* 5 61042646 "61/5,ratSlanretnI,rehsaW
31* 5 44011646 "1x81-"61/5,paC,wercS
31-5* 1 62186546 PH2,tiKtliubeRrewoL
41 1 46186546 eguaG/w,ekatnI,dlofinaMevlaVdna
51 2 40097546 ,ebuT"4/1,tuNnoisserpmoC
61 2 20001646 noisserpmoC,"4/1,eveelS
71 1 55186546 "4/1x"4/1,woblE
81 1 40086546 PH2,tiKteeFgnitnuoM
1 45041546 otxoBlacirtcelE,tekcarBrezilibatS,rotoM
1
6
7
18
2
3
5
4
9
10
14
8
1617
11
12
13
15
Section IV Vacuum Pumps
Technical Manual
iv-Page 51
PARTS
LISTS
CV-101FS / CV102FS
Installation, Operation and Care Manual
Single Wet Vacuum Pump
14
Electrical Box
Section IV Illustrated Parts Breakdown
*NOTE: Electrical Box Assembly Kits: Part Nos.64501151, 115/230V (1 HP) and PN 64501157,230V (2 HP) are complete with water manifold attached.
1
2
345
15
16
14
18
17
.oN .ytQ .oNtraP noitpircseD
1 05110546 ,ylbmessAxoBlacirtcelEV032roV511
1 1 11031546 sissahCxoBlortnoC2 1 B11031546 revoCxoBlortnoC3 1 72092546 V42,woL,laceD4 1 47192546 gninraWkcohS/eriF,laceD5 1 28192546 margaiD.celEVD,laceD6 1 58192546 kcolBlanimreTVD,laceD7 1 40016546 redloHpilCesuF8 1 33186546 yticapaCPMA03,yaleR9 2 450-3261 "2/1x23-01,wercS01 1 06011646 dnuorG,neerG23-01,wercS11 4 58011646 spillihP"4/1x8,wercS
21 1 91111646 esuF,"61/5x23-6,wercSredloH
31 2 02111646 draoBlanimreT,23-6,wercS41 1 32051646 tcennoC.kQ7,drBlanimreT51 1 620-6361 ratsl'tnI01#,rehsaWkcoL61 1 59186546 V032/511,.snarTyramirP71 1 52092546 stloVwoLnoituaC,laceD81 1 30292546 gninraWegasUeriW,gaT91 2 30062646 yarGB-17,laedI,tuNeriW
1 86081646 kcalB,tiudnoC1 85042646 "8/7,rehsaW2 10087546 "2/1,esahC,elppiN1 37052646 "6,elpruP,.aG81,eriW
* 1 15110546 ,tiKylbmessAxoBlacirtcelE)PH1(V032/511
* 1 75110546 ,tiKylbmessAxoBlacirtcelE)PH2(V032
02 11
021-2083119-1083
roPMA51,rekaerBtiucriCPMA8,rekaerBtiucriC
.oN .ytQ .oNtraP noitpircseD
1 80023546 V032,droC
1
1
68192546
78192546
tnemecalpeResuF,laceDroV511,A01/3
tnemecalpeResuF,laceDV032,A001/51
1
1
52192546
62192546
zH06,A5.7,V032,laceDro)lauDPH1(
zH06,A51,V032,laceD)PH2(
1 794-9762 dnuorG,laceD
1 52013546 kcalB,ni-panS,rotcennoC
1 10097546 "2/1,gnikcoL,tuN
1
1
13186546
03186546
,olB-olS,pmA01/3,esuFro)4fo.gkp(V511
,olB-olS,pmA001/51,esuF)4fo.gkp(V032
2
6 1
7
10
11
8
9
12
13
19
20
Technical Manual
Section IV Vacuum PumpsPage iv-52
PARTS
LISTS
CV-101FS / CV102FS
15Single Wet Vacuum Pump
Water Control Assembly
Section V Repair Parts
The water control system of the CustomAir SingleWet Vacuum Pump provides the pump with the waterflow required for proper suction and supplementalcooling. Automatic and independent water control isprovided for the pump. Filters protect the systemfrom damage because of solid materials in the watersupply. Flow valves control the amount of watersupplied to the pump.
16
7
10
2 3
4 5
8
9
1312
11
14
.oN .ytQ .oNtraP noitpircseD
11
6918654679186546
PH1,.yssAlortnoCretaWPH2,.yssAlortnoCretaW
1 1 40014546 xebuT"4/1,ssarB,woblEPIM"8/1
2 1 53186546 "8/1,retliFretaW
3 1 20087546 PIM"8/1,ssarB,elppiN
4 1 65186546 zh06,v42,evlaVdioneloS
5 1 27040546 xPIM"8/1,ssarB,retpadAPIF"8/1
6 1 60114546 x.pmoC"4/1,ssarB,woblEPIM"8/1
7 1 45186546 thgiartS"6,ebutretaW
8 1 31013546 "8/1xebuT"4/1,rotcennoCPIM
9 1 50061546 ,ssarB,recudeR,gnihsuBPIF"8/1xPIM"8/3
01 1 11022646 "8/3,evlaVnohpys-itnA
11 1 20060646 noisserpmoCTSS,gnirpS
21 129186546
01022646
2/1,evlaVlortnoCwolFroPH1,MPG
,MPG1,evlaVlortnoCwolFPH2
31 1 48014546 xebuT"4/1,ssarB,woblEPIM"8/3
41 RA 80006646 tnalaeS175#,etitcoL*
* All Fittings to be sealed with liquid sealant unless fitting ispretreated from manufacturer with sealant.
Section IV Vacuum Pumps
Technical Manual
iv-Page 53
PARTS
LISTS
CV-101FS / CV102FS
Installation, Operation and Care Manual
Single Wet Vacuum Pump
16
Anti-Siphon Valve Assembly
Water Solenoid Valve Assembly
.oN .ytQ .oNtraP noitpircseD
1 11022646evlaVnohpiS-itnA
ylbmessA
3,2,1 1 61086546 tiKriapeRnohpiS-itnA
1
2
3
.oN .ytQ .oNtraP noitpircseD
1 38022646evlaVdioneloSretaW
V42,ylbmessA
1 1 88022646 ydoBevlaV
2 1 60020546 V42,ylbmessAnaC&lioC
Section V Repair Parts
A
64568156
Technical Manual
Section IV Vacuum PumpsPage iv-54
PARTS
LISTS
CV-101FS / CV102FS
17Single Wet Vacuum Pump
Section V Repair Parts
Intake Manifold Assembly
Primary / Secondary Filter
ASHCROFT
5
1015
20
25
30 0in.Hgvac
1
6
7
2
3
4
5
8
.oN .ytQ .oNtraP noitpircseD
1 46186546 caViH,yssAdlofinaMekatnI
1 1 88186546 "2,eguaGmuucaV
2 1 10027546 ssarB,dlofinaMekatnI
3 2 95011646 ,wercSpaCdaeHtekcoS"2x02-"4/1
4 1 90048546 ssarB,"8/3,gulP
5 2 30087546 ssarB,esolCx"4/3,elppiN
6 1 10022646 ssarB,evlaVfeileRmuucaV
7 1 73186546 rebiF,teksaGdlofinaM
8 1 95186546 relffuMfeileRmuucaV
.oN .ytQ .oNtraP noitpircseD
1 1 04054546 lwobraelchtiw,"4/3,yramirP,retliF
2 1 91186546 teksaG&neercS
3 1 02186546 teksaG&neercS,lwoB
1
2
3
Section IV Vacuum Pumps
Technical Manual
iv-Page 55
PARTS
LISTS
MC-201 FS / MC-202 FS
5Dual Wet Vacuum Pump
Section III Installation
Water Supply Line
Vacuum Line
The water going to the unit acts as a pump sealantand cooling agent. When the vacuum pump is inoperation, the water supply must be on at all times.The plumber supplies and installs a 1/2" gate valveon the water supply line. Connect the 3/8" nylontubing and fittings, which are supplied with the unit,as illustrated.
Vacuum lines must be installed by a local plumberaccording to local building codes. All vacuum linesand risers are recommended to be IPS, PVC, SCH.40. Type “M” copper should be used if local codedoes not allow the use of PVC.
Care should be taken to slope the lines 1" for every20' of run toward the vacuum pump(s). This allowswaste and liquids to flow with gravity, contributing tothe efficiency of the vacuum system.
Make all connections using long radius sweep fittings.To promote unrestricted flow of air and waste liquidsthrough the vacuum lines, directional flow connec-tions should be used.
Using 45º elbows for turns or avoiding obstructions isbest; however, do not make a trap in the line, doing sowill decrease the efficiency of the system.
All elbows and tees should be sized for the main lineand sized down with bushing reducers to accommo-date smaller lines.
Avoid sagging lines, which cause the formation oftraps in the system and prevent good air and wasteliquid flow.
Connect the evacuation system to the vacuum lineusing the hose and fittings supplied in the installationkit. The vacuum line can be installed from the right orleft side of the cabinet.
1
23
4
56
7
1
2 3 4
5
.oN .ytQ .oNtraP noitpircseD
1 1 * eniLretaW
2 1 * "2/1,ssarB,etaG,evlaV
3 1 14061546xPIM"2/1,ssarB,gnihsuB
PIF"8/3
4 1 76040546xPIM"8/3,ssarB,retpadA
ebuT"8/3
5 1 70001646,noisserpmoC,eveelS
"8/3,ssarB
6 1 72097546 ebuTnolyN,"8/3,ssarB,tuN
7 .tf4 64081646 "8/3,nolyN,ebuT
* Plumber Supplies
* Plumber Supplies
.oN .ytQ .oNtraP noitpircseD
1 * ,CVP,SPI,eniLmuucaV04.HCS
2 * 04.HCS,CVP,SPI,woblE
3 1 35186546 xPIM"1CVP,retpadAknahS"2/1-1
4 1 40072546 esoH"2/1-1,pmalC
5 .tf6 91026546 "2/1-1,esoH
Technical Manual
Section IV Vacuum PumpsPage iv-56
PARTS
LISTS
MC-201 FS / MC-202 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
6
Waste Line
Section III Installation
— IMPORTANT NOTE —No part of the waste line should be morethan three (3) feet above the level of thewaste connection on the vacuum pump.
The waste line carries water and liquid waste from theoperatory to the building's sewer system.The waste line should follow the most direct path tothe sewer connection with a minimum of bends andelevations, and must be installed according to localbuilding and plumbing codes.The waste line can be installed from the right or leftside of the cabinet.The exhaust connection should be made by either oftwo methods, floor sink connection or direct connec-tion to “P” trap, depending on local code and build-ing facilities.
1
23
4
FLOOR SINK CONNECTION
DIRECT CONNECTION TO “P” TRAP
1
2
Electrical Connection Assembly
Use floor sink adapter SA-200, Part No. 6-4504-013.Install as illustrated.
Use “P” trap-air gap, if required by local code. Installas illustrated.
.oN .ytQ .oNtraP noitpircseD
1 1 * "1,kniSroolF,CVP,retpadAtresnIdebraB
2 1 35186546 "2/1-1xPIM"1,CVP,retpadAknahS
3 1 30072546 "2/1-1,leetSsselniatS,pmalC
4 .tf6 91026546 "2/1-1,tsuahxE,esoH
.oN .ytQ .oNtraP noitpircseD
1 .tf6 91026546 "2/1-1,tsuahxE,esoH
2 1 30072546 "2/1-1,leetSsselniatS,pmalC
3 1 * "2/1-1woblE
4 1 * "2/1-1,elppiN
5 1 * "2,paGriA,retpadA
6 1 * "2/1-1x"2,lleB,recudeR
7 1 * esolCx"2/1-1,elppiN
8 1 * "2/1-1,”P”,parT
Complete the electrical hook-up, as the electricalcodes require, through the .375" hole in the boxcover.
.oN .ytQ .oNtraP noitpircseD
1 1 12031546 wercSetalPrevoC
2 1 48448526 xoB-J,etalPrevoC
3 1 06011646 wercSdnuorG
1
2
* Plumber Supplies
Section IV Vacuum Pumps
Technical Manual
iv-Page 57
PARTS
LISTS
MC-201 FS / MC-202 FS
11Dual Wet Vacuum Pump
Frame Assembly
Section IV Illustrated Parts Breakdown
The various parts of the CustomAir Dual WetVacuum Pump are attached to the frame assembly.The table identifies each assembly and the drawingindicates its location.
.oN .ytQ .oNtraP noitpircseD
1 1 etelpmoc,erusolcnE
2 1 15410546 ylbmessAdlofinaMekatnI
3 11
rolaceD102CMlaceD202CM
4 1 43186546 eguaGmuucaV
5 2 23186546 hctiwSrekcoR
6 1 36186546 ylbmessAdlofinaMretaW
7 4 16186546 tooFCVPlauD
8 6 72011646 "4/1-1,wercSlauD
9 1 26186546 ylbmessAdlofinaMetsaW
01 1 61065546 etihw,"2/1,temmorG
11 1 62031546 tiKnoitallatsnIlacirtcelE
21 1 19310546 ylbmessAlortnoCmuucaV
31 1 94186546 )4fo.gkp(tnuoMrotoM
41 2 71072546 TME"2,regnaH,pmalC
1 2
3
4
5
6
7
8 910
11
12
13
14
8
Technical Manual
Section IV Vacuum PumpsPage iv-58
PARTS
LISTS
MC-201 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
12
1 HP Dual Pump Assembly
Section IV Illustrated Parts Breakdown
The CustomAir Dual Wet Vacuum Pump is poweredby two water injected, impeller type vacuum pumps.This type of vacuum producer runs quietly andgenerates a high mercury pull, which is adjustablefrom a normal operating range of 10 to 12 inches ofmercury to a high range of 18 to 20 inches of mer-cury for surgical applications. There are two vacuumpump sizes: the 1 HP (model MC-201 FS) and the2 HP (model MC-202 FS, illustrated on Page 13). Thedual pump back-up system and the choice of vacuumpower ranges, make this a highly reliable and versatileevacuation system.
.oN .ytQ .oNtraP noitpircseD
1 R10868546 PH1,ylbmessApmuPlauD
1 1 35110546 V032,tinUyaleR
2 1 69186546 nollag2/1,dlofinaMretaW
3 1 80023546 V032,.yssAdroCcirtcelE
4 1 95057546 SF,PH1,rotoM
5 1 55186546 "4/1x"4/1,ssarB,woblE
6 2 34186546 xesoh"1,ssarB,rotpadAPIM"2/1
7 2 93186546 teertS"2/1,ssarB,woblE
8* sA.qeR
1004064620040646300406464004064650040646
kcihT500.leetS,mihSkcihT130.leetS,mihSkcihT510.leetS,mihSkcihT200.leetS,mihSkcihT010.leetS,mihS
9* 1 99186546 PH1,ssarB,gnisuoH
01* 1 32186546 teksag(laeSretaWyratoR)dedulcni
11* 4 820-5161 "1x61--"8/3,wercSpaC
21* 1 89186546 ,decnalaB,ssarB,rellepmIPH1
31* 1 72186546 rebiF,teksaGgnisuoH
41* 1 76186546 PH1,ssarB,etalPesaB
51* 4 61042646 "61/5,rehsaWratSlanretnI
61* 4 44011646 "1x81--"61/5,wercSpaC
61-8* 1 52186546 PH1,tiKdliubeRrewoL
NOTE: See the Operation/Maintenance section ofthis manual for detailed information on hookup andcare of the Dual Wet Vacuum Pump.
1
6
7
11
12
2
3
4
5
910
13
14
15 16
8
58
Section IV Vacuum Pumps
Technical Manual
iv-Page 59
PARTS
LISTS
MC-202 FS
13Dual Wet Vacuum Pump
2 HP Dual Pump Assembly
Section IV Illustrated Parts Breakdown
.oN .ytQ .oNtraP noitpircseD
1 R20868546 PH2,ylbmessApmuPlauD
1 1 85110546 V032,tinUyaleR
2 1 79186546 nollaG1,dlofinaMretaW
3 1 80023546 V032,.yssAdroCcirtcelE
4 1 06057546 SF,PH2,rotoM
5 1 55186546 "4/1x"4/1,ssarB,woblE
6 2 78040546 xesoh"1,ssarB,rotpadAPIM"4/3
7 2 44186546 teertS"4/3,ssarB,woblE
*8 sA.qeR
1004064620040646300406464004064650040646
kcihT500.leetS,mihSkcihT130.leetS,mihSkcihT510.leetS,mihSkcihT200.leetS,mihSkcihT010.leetS,mihS
*9 1 00286546 PH2,ssarB,gnisuoH
* 01 1 42186546 teksag(laeSretaWyratoRdedulcni
* 11 4 820-5161 "1x61--"8/3,wercSpaC
* 21 1 98186546 ,decnalaB,ssarB,rellepmIPH2
* 31 1 82186546 rebiF,teksaGgnisuoH
* 41 1 86186546 PH2,ssarBetalPesaB
* 51 5 61042646 "61/5,rehsaWratSlanretnI
* 61 5 44011646 "1x81--"61/5,wercSpaC
61-8* 1 62186546 PH2,tiKdliubeRrewoL
1
6
7
10 11
2
3
4
5 8
9
1213
15 16
14
Technical Manual
Section IV Vacuum PumpsPage iv-60
PARTS
LISTS
MC-201 FS / MC-202 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
14
Intake Manifold Assembly, Dual
Section IV Illustrated Parts Breakdown
The intake manifold swing-check valves provideautomatic isolation for the back-up pump when it isnot operating. This prevents vacuum loss through theexhaust of the standby pump. When both pumps areoperating, the swing-check valves open to allowmaximum air flow. The manifold is constructed ofPVC and the valves are brass. Both materials are notaffected by chemical disinfectants.
.oN .ytQ .oNtraP noitpircseD
1 15410546 ylbmessAdlofinaMekatnI
1 3 71114546 knahS"1xTPN"4/3,woblE
2 2 1918654639186546
"4/3,evlaVkcehCgniwS"1,evlaVkcehCgniwS
3 3 60061546 xpilS"1recudeR,gnihsuBPIF"4/3
4 1 60073546 pilS"1CVP,ssorC
5 1 40014546 "8/1xebuT"4/1,ssarB,woblEPIM
6 3 24186546 esolC"4/3,ssarB,elppiN
7 1 89087546 gnoL"2x"1,CVP
8 1 80061646 pilS"1,eeT
9 1 79087546 gnoL"4x"1,CVP
01 2 92040546 pilS"1xPIF"1,CVP,rotpadA
11 1 69087546 gnoL"21x"1,CVP
1
67
2
11
4
5
8
To Left Pump
Intake
To Right Pump
To Gauge Assembly
To Vacuum Relief
10
3
9
▼
▲
▼
▲
▲
Section IV Vacuum Pumps
Technical Manual
iv-Page 61
PARTS
LISTS
MC-201 FS / MC-202 FS
15Dual Wet Vacuum Pump
Exhaust Manifold Assembly, Dual
Section IV Illustrated Parts Breakdown
The exhaust manifold delivers liquid waste materialsfrom both pumps to the drain line. The manifold isconstructed of PVC to prevent chemical corrosionand buildup of organic and mineral wastes. Theexhaust muffler connected to the drain line outsidethe cabinet assures quiet operation.
4
56
7
8
1
2
3
.oN .ytQ .oNtraP noitpircseD
4 1 22110546 tsuahxErotalucriceRretaW)lanoitpO(ylbmessAdlofinaM
5 1 13061546 "2/1-1,recudeRgnihsuBPIF"1xPILS
6 2 43114546 TPN"1,ssarB,woblE
7 2 12114546 PIM"8/1xyloP"4/1,woblE
8 2 12140546 esoH"1,rotpadA
.oN .ytQ .oNtraP noitpircseD
1 1 26186546 .yssAdlofinaMtsuahxE.dtS
2 2 43114546 "1xPIM"1,ssarB,woblEknahS
3 1 42048546 PIM"2/1-1,CVP,gulP
Water Recirculator Exhaust Manifold Assembly (Optional)
Technical Manual
Section IV Vacuum PumpsPage iv-62
PARTS
LISTS
MC-201 FS / MC-202 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
16
Electrical Box
Section IV Illustrated Parts Breakdown
*NOTE: Electrical Box Assembly Kits: PN 64501153,230V (1 HP) and PN 64501158, 230V (2 HP) arecomplete with water manifold attached.
.oN .ytQ .oNtraP noitpircseD
1 05110546 ,ylbmessAxoBlacirtcelEV032roV511
1 1 11031546 sissahCxoBlortnoC2 1 B11031546 revoCxoBlortnoC3 1 72092546 V42,woL,laceD4 1 47192546 gninraWkcohS/eriF,laceD5 1 28192546 margaiD.celEVD,laceD6 1 58192546 kcolBlanimreTVD,laceD7 1 40016546 redloHpilCesuF8 1 33186546 yticapaCPMA03,yaleR9 2 450-3261 "2/1x23-01,wercS01 1 06011646 dnuorG,neerG23-01,wercS11 4 58011646 spillihP"4/1x8,wercS
21 1 91111646 esuF,"61/5x23-6,wercSredloH
31 2 02111646 draoBlanimreT,23-6,wercS
41 1 32051646 kiuQ7,draoBlanimreTtcennoC
51 1 620-6361 ratsl'tnI01#,rehsaWkcoL
61 1 59186546 ,remrofsnarTyramirPV032/511
71 1 52092546 stloVwoLnoituaC,laceD81 1 30292546 gninraWegasUeriW,gaT91 2 30062646 yarGB-17,laedI,tuNeriW
1 86081646 kcalB,tiudnoC1 85042646 "8/7,rehsaW2 10087546 "2/1,esahC,elppiN
* 1 35110546 ,tiKylbmessAxoBlacirtcelE)PH1(V032/511
* 1 85110546 ,tiKylbmessAxoBlacirtcelE)PH2(V032
02 1 021-2083 pmA51rekaerBtiucriC02 1 119-1083 pmA8,rekaerBtiucriC
.oN .ytQ .oNtraP noitpircseD
1 80023546 V032,droC
1
1
68192546
78192546
tnemecalpeResuF,laceDroV511,A01/3
tnemecalpeResuF,laceDV032,A001/51
1 794-9762 dnuorG,laceD
1
1
52192546
62192546
zH06,A5.7,V032,laceDro)lauDPH1(
zH06,A51,V032,laceD)PH2(
1 52013546 kcalB,ni-panS,rotcennoC
1 10097546 "2/1,gnikcoL,tuN
1
1
13186546
03186546
,olB-olS,pmA01/3,esuFro)4fo.gkp(V511
,olB-olS,pmA001/51,esuF)4fo.gkp(V032
2
45
2
1
3
15
16
14
18
17
6 1
7
10
11
8
9
12
13
19
20
Section IV Vacuum Pumps
Technical Manual
iv-Page 63
PARTS
LISTS
MC-201 FS / MC-202 FS
17Dual Wet Vacuum Pump
Water Control Assembly
Section V Repair Parts
The water control system of the CustomAir Dual WetVacuum Pump provides the pumps with the waterflow required for proper suction and sopplementalcooling. Automatic and independent water control isprovided for each pump. Filters protect the systemfrom damage due to solid materials in the watersupply. Flow valves control the amount of watersupplied to each pump.
1
6
7
10
2 3
4 5
8
913
12
11
14
15
.oN .ytQ .oNtraP noitpircseD
11
6918654679186546
PH1,.yssAlortnoCretaWPH2,.yssAlortnoCretaW
1 1 40014546 xebuT"4/1,ssarB,woblEPIM"8/1
2 1 53186546 "8/1,retliFretaW
3 1 20087546 PIM"8/1,ssarB,elppiN
4 1 65186546 evlaVdioneloS
5 1 27040546 xPIM"8/1,ssarB,retpadAPIF"8/1
6 1 60114546 x.pmoC"4/1,ssarB,woblEPIM"8/1
7 1 45186546 thgiartS"6,ebutretaW
8 1 31013546 "8/1xebuT"4/1,rotcennoCPIM
9 1 50061546 ,ssarB,recudeR,gnihsuBPIF"8/1xPIM"8/3
01 1 11022646 "8/3,evlaVnohpys-itnA
11 129186546
01022646
2/1,evlaVlortnoCwolFroPH1,MPG
,MPG1,evlaVlortnoCwolFPH2
21 1 40114546 xebuT"4/1,ssarB,woblEPIM"8/3
31 2 11322646 TPNF"8/3,evlaVkcehC
41 1 19040546 rotpadA
51 1 60081646 "4/1,ebuT
Technical Manual
Section IV Vacuum PumpsPage iv-64
PARTS
LISTS
MC-201 FS / MC-202 FS
Installation, Operation and Care Manual
Dual Wet Vacuum Pump
18
Vacuum Control Assembly
Section V Repair Parts
The vacuum control assembly is attached to thecentral section of the front control panel. The valvein this assembly provides automatic regulation of thevacuum flow in the system. The vacuum level can beadjusted for different system requirements. The vacuumgauge is connected to a fitting on this assembly.
.oN .ytQ .oNtraP noitpircseD1 19310546 ylbmessAlortnoCmuucaV
1 1 71114546 "1xPIM"4/3,CVP,woblEknahS
2 1 10022646 evlaVfeileRmuucaV3 2 61014546 "4/3teertS,ssarB,woblE
4 1 80061546 xpilS"4/1-1,CVP,gnihsuBPIF"4/3
5 1 54186546 feileRmuucaV,tnemelEretliFevlaV
6 1 58011646 spillihP"4/1x8,wercS7 1 90012546 "4/1-1,CVP,paC
Vacuum Gauge and Adapter
In-Line Filter Assembly, 1"
.oN .ytQ .oNtraP noitpircseD1 1 43186546 ylbmessAeguaGmuucaV
*2 1 80013546 yloP"4/1xPIF"4/1,retpadA
NOTE: *Not included in assembly.
.oN .ytQ .oNtraP noitpircseD1 04910546 "1,ylbmessAretliFenIL-nI
1 1 71090546 "1,ydoBretliFeniL-nI
2 1 12186546 D"4/1-2,hseM02,"1,neercSteksaG&,L"4/1-4x
3 1 22186546D"4/1-2,hseM02,"1,neercS"1,lwoB&teksaG,L"4/1-4x
raelC,retliFeniL-nI4 1 14041546 redlo,retliFeniL-nI"1,tekcarB
52
2
35186546
30140546
"2/1-1xPIM"1CVP,retpadAroknahS
"1xPIM"1CVP,retpadAknahS
6 1 24041546 llaWretliFeniL-nI,tekcarBtnuoM
7 3 70011646 tolSnaP"2/1x01,wercS
1
6
7
2
3
45
12
1
5
62
3
4
7
Section IV Vacuum Pumps
Technical Manual
iv-Page 65
PARTS
LISTS
MC-201 FS / MC-202 FS
19Dual Wet Vacuum Pump
Anti-Siphon Valve Assembly
Water Solenoid Valve Assembly
Section V Repair Parts
.oN .ytQ .oNtraP noitpircseD
1 11022646evlaVnohpiS-itnA
ylbmessA
3,2,1 1 61086546 tiKriapeRnohpiS-itnA
1
2
3
.oN .ytQ .oNtraP noitpircseD
1 65186546evlaVdioneloSretaW
ylbmessA
A
2500 Highway 31 SouthBay Minette, Alabama 36507
1-866-DTE-INFOFax: (251) 937-0461www.dentalez.com
© DentalEZ Alabama, Inc.Printed in USAJanuary, 2010