a-dec dental furniture - prosites, inc.c1-preview.prosites.com/106982/wy/docs/adec preference...

98
A-dec Dental Furniture PRE-I NSTALLATION GUIDE For Preference Collection, ® Preference ICC, ® and Preference Slimline

Upload: truongnguyet

Post on 04-Dec-2018

223 views

Category:

Documents


0 download

TRANSCRIPT

A-dec Dental Furniture

Pre-InstallatIon GuIde

For Preference Collection,® Preference ICC,® and Preference Slimline™

Copyright

© 2012 A-dec Inc. All rights reserved.

A-dec Inc. makes no warranty of any kind with regard to this mate-rial, including, but not limited to, the implied warranties of merchant-ability and fitness for a particular purpose. A-dec Inc. shall not be held liable for any errors contained herein or any consequential or other damages concerning the furnishing, performance or use of this material. The information in this document is subject to change without notice. If you find any problems in the documentation, please report them to us in writing. A-dec Inc. does not warrant that this document is error-free.

No part of this document may be copied, reproduced, altered, or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or by any information storage and retrieval system, without prior written permission from A-dec Inc.

Trademarks and Additional Intellectual Property Rights

A-dec, the A-dec logo, A-dec 500, A-dec 300, Cascade, Cascade Master Series, Century Plus, Continental, Decade, ICX, ICV, Performer, Preference, Preference Collection, Preference ICC, and Radius are registered in the United States and other countries. A-dec 200, Preference Slimline, and reliablecreativesolutions are also trademarks of A-dec Inc. None of the trademarks or trade names in this document may be reproduced, copied, or manipulated in any manner without the express, written approval of the trademark owner.

Certain touchpad symbols are proprietary to A-dec Inc. Any use of these symbols, in whole or in part, without the express written consent of A-dec Inc., is strictly prohibited.

Regulatory Information

The Regulatory Information and Specifications document is delivered with A-dec dental device equipment as mandated by agency requirements. If you need this information, please go to the Document Library at www.a-dec.com.

Product Service

For service information, contact your local authorized A-dec dealer. To find your local dealer, go to www.a-dec.com.

86.0142.00 Rev T ii

A-dec Dental Furniture Pre-Install

INTRODUCTION ......................Things to Note ...................................

Preference Collection® Return PAccessory Console and PlumbiDocument Information.............Other Sources of Information..Pre-Installation Notes and Cau

Electrical Specifications ....................Electrical Rough-In ..................Electrical Configuration ............

Floor Mounting..................................Concrete Floors ..........................Non-Concrete Floors .................

PREFERENCE COLLECTION ........Estimated Shipping Weights ...........Wiring Diagrams...............................

5543 Central Console: 99", 106"5531 Accessory Console Floor-M5731 Upper Storage Unit Wall-5580 Treatment Console Floor-

42" Center Module (base on69" Center Module w/1 sin96" Center Module w/2 sin

Floor-Mount ....................................... 14Floor-Mount ....................................... 14Insert Wall-Mount ............................ 14............................................................... 15............................................................... 16............................................................... 16............................................................... 16............................................................... 17Console ................................................ 17Storage Unit......................................... 19............................................................... 21olumn Insert with Pass-Through.......... 21olumn Insert with Non-Pass-Through .. 22

............................................................... 23

............................................................... 23

............................................................... 25

............................................................... 27Console................................................ 27Console................................................ 28Console................................................ 29Console................................................ 30Console................................................ 31Console ................................................ 32 Template............................................. 33

C O N T E N T S

ation Guide

......................................... 1

............................................................. 2olicy................................................... 2

ng Templates .................................... 2.............................................................. 2.............................................................. 3tions .................................................... 3............................................................. 4.............................................................. 4............................................................. 5............................................................. 6............................................................. 6............................................................. 7

......................................... 9

........................................................... 10

........................................................... 11, and 113" w/sink............................ 11

ount................................................ 12Mount ............................................... 12Mountly)..................................................... 13k module (base only)..................... 13k modules (base only) ................... 13

5580, 5562, and 56625580 Treatment Console 5562 Treatment Column 5662 Treatment Column

Utility Specifications.....................Loop Vent Diagram ......................

Universal 5543 ........................5531 Accessory Console .......

Mounting Specifications ..............5631 Wall-Mount Accessory 5730 Dispenser Unit or 5731

Framing Dimensions ....................5662 34" and 42" Treatment C5662 34" and 42" Treatment C

Framing Specifications .................5731 Upper Storage Insert ....X-Ray Storage Insert ..............

Templates and Utilities ................5531 and 5631 30" Accessory 5531 and 5631 44" Accessory 5531 and 5631 58" Accessory 5531 and 5631 72" Accessory 5531 and 5631 86" Accessory 5631 Wall-Mount Accessory 5531 Floor-Mounted Utilities

iii 86.0142.00 Rev T

A-dec

555555

555556

Full-Si

PREFEEstimaUtility

El

SpottinUtility

SWSLSSSWMMMMMMMMMMM

................................................................64

................................................................64

................................................................65

................................................................66

................................................................67

................................................................68

................................................................69

................................................................70

................................................................71

................................................................72

...........................................73

................................................................74

................................................................75ount.......................................................75ations....................................................76ount.......................................................76................................................................76................................................................77................................................................79................................................................80................................................................81................................................................81................................................................82................................................................83................................................................84................................................................84................................................................85l-Mount .................................................86................................................................87................................................................88................................................................88................................................................90................................................................92

Dental Furniture Pre-Installation Guide

31 and 5631 Wall-Mounted Utilities............................................... 3443 Round/Square-End Central Console........................................ 3562 Treatment Column 34" and 42" .................................................. 36

Wall-Mounted Utilities (optional) ............................................. 3780 Treatment Console 42", 69", and 96".......................................... 3880 Treatment Console Plumbing (Wall-Mounted Utilities)........ 3962 Treatment Insert 34" and 42" ...................................................... 40ze Template Reference ..................................................................... 41

RENCE ICC ............................................................43ted Shipping Weights ...................................................................... 44 Specifications .................................................................................... 45ectrical ................................................................................................. 46

Ventilation Fan in Clean Module, Sterile Storage Area ........... 46Ancillary Devices ........................................................................... 46General Considerations ................................................................ 46Optional Dryer ............................................................................... 46Canadian Wiring Kit ..................................................................... 46Task Lights...................................................................................... 47

g Back Panel Locations ................................................................... 48 Templates.......................................................................................... 49D2 Module ....................................................................................... 49

DW2 / SLDW3 Module................................................................... 50D3 Module ......................................................................................... 51S0 Module ........................................................................................ 52

SRU Module....................................................................................... 53SLW Module ...................................................................................... 54SLD Module ....................................................................................... 55SRW Module ...................................................................................... 56SRD Module....................................................................................... 57SSD Module ....................................................................................... 58PS4 Module ........................................................................................ 59DM5 Module...................................................................................... 60SMD Module...................................................................................... 61DK0 Module....................................................................................... 62WMD Module.................................................................................... 63

Back Panel Templates...................CSD3 Module .........................CSS4 Module ..........................CLS5 Module..........................CTP4 Module..........................C2S2B Module........................CSS1 Module ..........................CST1-12 Module.....................CST1-25 Module.....................CRNR Module........................

PREFERENCE SLIMLINE ..........Estimated Shipping Weights.......Mounting Specifications ..............

5636 Utility Console, Wall-MFraming and Reinforcing Specific

5636 Utility Console, Wall-MFraming............................Reinforcing ......................

Utility Specifications.....................5536 18" Utility Console ........5536 34" Utility Console ........

Left-Side Delivery...........Right-Side Delivery ........

5537 18" Utility Console ........5537 34" Utility Console ........

Left-Side Delivery...........Right-Side Delivery ........

5636 18" Utility Console WalOptional Utilities ............

5636 34" Utility Console ........Left-Side Delivery...........Right-Side Delivery ........

Full-Size Template Reference......

1

86.0142.00 Rev T 1

A-dec Dental Furniture Pre-Install

I N T R O D U C T

This guide provides detailed techniplumbing specifications, and diagr

Information presented in this docu

• Specify required plumbing and u• Specify framing and construction• Wire individual units• Mount and secure cabinetry

ation Guide

I O N

cal information on electrical, utility, ams, as well as mini templates.

ment will assist you to:

tilities requirements

2 86.0142.00 Rev T

A-dec

T H

As youspecifi

• Theconsengi

• Shipthisor c

• Locashal

• All uregu

• For wal

• Thebe p

• Thenot

Return Policynot be returned for credit.

Plumbing Templatesmbols:

r standard and optional stubs.

op or extent of console.

bers, electricians, carpenters, ho are involved in designing and s that accommodate Preference

. structural or architectural engineer, sonry walls. Ensure that all requirements.

ed to be, and should not be, used as a al, state, federal code, and American onsibility of the user to integrate the esented herein with the regulations ing and inspection authorities.

Dental Furniture Pre-Installation Guide

I N G S T O N O T E

use this document to guide you through the technical cations, note the following:

minimum requirements for attachment and for wall truction have been determined by an independent structural neering company. Documented detail is available from A-dec.ping weights, hanging weights and mounting specifications in

document are for standard Preference products. Modifications ustomizations may affect these specifications.l regulations provide that licensed plumbers and electricians

l install utilities.tilities must conform to NFPA, local fire codes and local lations, as applicable.

the accessory consoles, utilities must be plumbed through the l or subbase of the cabinet.re are several areas in the console subbase where the utilities can lumbed. drawings provided in this guide are for reference only. They are drawn to scale.

Preference Collection® Preference Collection products can

Accessory Console and These templates included target sy

Indicates datum points fo

Indicates end of countert

Document Information

Intended AudienceThis guide is for contractors, plumdesigners, architects and others wremodeling dental treatment roomCollection products.

Consult a licensed professional, i.efor mounting to metal studs or mainstallations conform to local code

DisclaimerThis written material is not intendsubstitute for compliance with locwith Disabilities Act. It is the respA-dec equipment specifications prand codes overseen by local plann

CAUTION Consult a licensed professional, i.e. structural or architectural engineer, for mounting to metal studs or masonry walls.

86.0142 3

INTRODUCTIONOther Sources of Information

OtheThe AA-decinformPrefer

A-deContaand ef

Full-SProvidPreferspottijob sitprovidsubba

InstaContacabine

A-deProvididentiselecti

and Cautions

truction should be completed before stallation (including finished floors,

ighting, paint, etc.).

ulation requires licensed plumbers nstall utilities. All plumbing and rm to prevailing local codes.

ner and method for accessing all is the responsibility of the design

e project (dental dealer, architectural ctors). Utilities must be accessible ools.

.00 Rev T

r Sources of Information-dec Dental Furniture Pre-Installation Guide is just one offering in ’s documentation set that provides technical product ation, which covers proper installation and maintenance of

ence Collection. Other associated documents include:

c Design Toolbox ins drawing and visualization blocks that you can use to quickly ficiently layout a treatment room space.

ize Templates e accurate footprints for the Preference Collection and

ence SlimlineTM product lines. They are designed to assist you in ng and positioning A-dec products whether on the construction e or in a completed treatment room. The full-size templates e plumbing and utility locations, mounting clip positions, metal

se structure and critical dimensions.

llation Guides in procedures designed to help you install Preference Collection try.

c Equipment Catalog es comprehensive product detail that walks you through

fying the cabinet models, choosing models and options and ng accent and decor color

Pre-Installation Notes

NOTE Job site consdental equipment inceiling, plumbing, l

CAUTION Local regand electricians to iutilities must confo

CAUTION The manutilities within the wteam working on thservices, and contrawithout the use of t

4 86.0142.00 Rev T

A-dec

E L E

Elec

A contand flefinishe

Use thfrom telectri

The lowires,

ple

Dental Furniture Pre-Installation Guide

C T R I C A L S P E C I F I C A T I O N S

trical Rough-In

ractor should supply the x-ray junction box, wiring, conduit x conduit. The conduit must protrude 3" (76.2 mm) from the d floor.

e flex conduit, as required, to connect the 3" (76.2 mm) piece he floor to the junction box. Wire as per local and national cal codes.

w voltage conduit is used for routing data lines and touchpad not for line voltage.

Figure 1. Electrical Rough-In Exam.

86.0142 5

INTRODUCTIONcifications Electrical Configuration

ElecTable ence Collection task lights:

o’clock console device: Doctor’s, Assistant’s, and

Ancil

A-dec

A-dec

A-dec

Monit

Intra

Curin

Amal

Ultra

X-ray

.00 Rev T

Electrical Spe

trical Configuration1. Operating Current for Ancillary Devices

Task Lighting

Power switch locations for Prefer

• Stack switch located on the 12• Power switch located on each

12 o’clock task light

lary Device Operating Current (Amps)@ 110-120 VAC

Preference Collection task light 0.3

300-watt power supply 2.8

17-watt power supply 0.2

or, typical 1.1

oral camera, typical 2.0

g light 2.0

gamator, typical 2.3

sonic scaler, typical 0.9

viewer, typical 0.4

NOTE Connect incoming power to each cabinet to a 20-amp circuit breaker. Table 1 shows typical current draw for various devices that might be installed in a cabinet. Always consult device name plate for actual power requirements. Mobile equipment may have substantial power requirements and should be considered when determining total treatment room branch circuits.

6 86.0142.00 Rev T

A-dec

F L O

ConTable 2

Figure

Ancho Part Number

Prefe -13 x 5-1/2" 038.034.00

Prefe -13 x 5-1/2" 038.034.00

5543 2-13 x 5-1/2" 038.034.00

A-dec -13 x 4-1/4" 038.004.00

Dental Furniture Pre-Installation Guide

O R M O U N T I N G

crete Floors. Concrete Floor Mounting Chart

2. Floor-Mounting - Concrete

r Dimension ADiameter Hole

Dimension BMinimum Depth Fastener

rence Collection Subbase Clips 1/2" (12.7 mm) 3-1/2" (88.9 mm) Stud, anchor, 1/2

rence ICC Leg Anchor 1/2" (12.7 mm) 3-1/2" (88.9 mm) Stud, anchor, 1/2

X-Ray Column Anchor 1/2" (12.7 mm) 3-1/2" (88.9 mm) Stud, anchor, 1/

500® 541/545 Mount 1/2" (12.7 mm) 2-3/4" (69.9 mm) Stud, anchor, 1/2

NOTE Dimension C per architectural drawings. Anchors are only approved for use in slabs greater than 5" (127 mm) thick. If concrete does not meet minimum requirements, consult with a licenced engineer for alternate anchors.

86.0142 7

INTRODUCTIONloor Mounting Non-Concrete Floors

NonTable

Figure.

Anch Part Number

Prefe ead screw, 3/8 x 2-1/2" 003.067.00

Prefe ead screw, 1/2 x 3-1/2" 003.048.00

5543 ead screw, 1/2 x 3-1/2" 003.048.00

Casca ead screw, 1/2 x 3-1/2" 003.048.00

A-dec ead screw, 1/2 x 3-1/2" 003.048.00

.00 Rev T

F

-Concrete Floors3. Non-Concrete Floor Mounting Chart

3. Floor-Mounting - Non-Concrete

or Dimension DDiameter Hole

Dimension EMinimum Embedment Depth Fastener

rence Collection Subbase Clips 9/32" (7.1 mm) 1-1/2" (38.1 mm) Lag screw, hex-h

rence ICC Leg Anchor 3/8" (9.5 mm) 2-1/2" (63.5 mm) Lag screw, hex-h

X-Ray Column Anchor 3/8" (9.5 mm) 3-1/4" (82.6 mm) Lag screw, hex-h

de® 3171/3175 Mount 3/8" (9.5 mm) 3-1/4" (82.6 mm) Lag screw, hex-h

541/545 Mount 3/8" (9.5 mm) 3-1/4" (82.6 mm) Lag screw, hex-h

NOTE Dimension F, per architectural drawings, must exceed minimum embedment depth for lag screw application. If the flooring material does not meet minimum requirements, consult your licensed contractor about adding reinforcements.

8 86.0142.00 Rev T

A-dec

Dental Furniture Pre-Installation Guide

2

86.0142.00 Rev T 9

A-dec Dental Furniture Pre-Install

P R E F E R E N C E

Preference Collection is a completethat can be custom configured to m

This section of the Dental Furniture

• Estimated Shipping Weights• Wiring Diagrams• Utility Specifications• Loop Vent Diagram• Mounting Specifications• Framing Dimensions• Framing Specifications• Full-Size Template Reference

ation Guide

C O L L E C T I O N

line of modular dental furniture eet various needs.

Pre-Installation Guide includes:

10 86.0142.00 Rev T

A-dec

E S T

PrefTable 4

Mode

5562.

5562.

5662.

5662.

5580.

5580.

5580.

5543.

5543.

5543.

5531.

5531.

5531.

5531.

5531.

5731.

5730.

5732.

5732.

5731.

Kilograms

47

53

77

75 / 80

103 / 110

132 / 141

165 / 176

203 / 216

Dental Furniture Pre-Installation Guide

I M A T E D S H I P P I N G W E I G H T S

erence Collection . Estimated Shipping Weights (laminate, solid surface / Quartz)

l Number Pounds Kilograms

34 500 227

42 621 282

34 424 192

42 519 235

42 677 / 692 307 / 314

69 770 / 794 349 / 360

96 930 / 964 422 / 437

SP 920 / 940 417 / 427

SB 1080 / 1090 490 / 495

TB 1800 / 1810 816 / 823

30 211 / 221 96 / 100

44 296 / 310 134 / 141

58 380 / 400 172 / 182

72 475 / 499 215 / 227

86 580 / 609 263 / 277

22 60 27

29 78 35

43 223 101

49 252 114

22 85 39

Model Number Pounds

5731.29 107

5731.34 116

5731.42 169

5631.30 165 / 175

5631.44 228 / 242

5631.58 291 / 311

5631.72 364 / 388

5631.86 447 / 476

86.0142 11

PREFERENCE COLLECTIONal Console 99", 106", and 113" w/sink

W I

55499",

Note: Locate incoming power on this side only if wire count is exceeded in the other box.

.00 Rev T

Wiring Diagrams 5543 Centr

R I N G D I A G R A M S

3 Central Console 106", and 113" w/sink

Figure 4. Wiring Diagram - 5543 Central Console

Note: To be connected by an electrician. See page 5.

Note: When the unit is installed, the electrician connects the jumper wires to the incoming wires. See page 5.

Note: To be connected by an electrician. See page 5.

A-dec

12 86.0142.00 Rev T

553

5531Floor

30", 44

5731Wall-

22", 29

pper Storage Unit

5731

Dental Furniture Pre-Installation Guide

1 and 5731

Accessory Console -Mount

", 58", 72", and 86" w/sink

Upper Storage Unit Mount

", 34", and 42" w/task light

Figure 5. Wiring Diagram - 5531 Accessory Console and 5731 U

Note: Provided if unit has floor plumbing.

5531

86.0142 13

PREFERENCE COLLECTIONWiring Diagrams 5580

558

Treat42" C(base

Treat69" Cw/ 1

Treat96" Cw/ 2

dules

.00 Rev T

0

ment Console Floor-Mount enter Module only)

ment Console Floor-Mount enter Module sink module (base only)

ment Console Floor-Mount enter Modulesink modules (base only)

Figure 6. Wiring Diagram - 5580 Treatment Console Center Mo

A-dec

14 86.0142.00 Rev T

558

5580Floor42", 69

5562Floor34" an

5662Wall-34" an

and 5662 Treatment Columns

Dental Furniture Pre-Installation Guide

0, 5562, and 5662

Treatment Console -Mount", and 96"

Treatment Column -Mountd 42"

Treatment Column InsertMountd 42"

Figure 7. Wiring Diagram - 5580 Treatment Console and 5562

Note: To be connected by electrician; see Electrical Rough-In on page 4.

Note: When the unit is installed, the electrician connects the jumper wires to the incoming wires; see Electrical Rough-In on page 4.

86.0142 15

PREFERENCE COLLECTIONpecifications 5580, 5562, and 5662

U T

Table

Utilit

Air .5 scfm (71 sl/min) minimum, during min] peak intermittent flow). Flush lines o dental equipment.

Wate 5 gpm (5.7 l/min) minimum, not to es prior to making connections to dental

Centr Hg (34 ± 7 KPa) and 9 scfm (255 sl/min) ms 4.5 ± 1 inches of Hg (16 ± 3.5 KPa) imum.

Elect ments listed throughout this section.

Gravi 1/4" or 1-1/2"

Nitro ufacturer.

Oxyg ufacturer.

.00 Rev T

Utility S

I L I T Y S P E C I F I C A T I O N S

5. Utility Specifications - Preference Collection

y Finish Requirement

Protrude 1" (25.4 mm) for 3/8" compression 80-125 psi (550 to 860 KPa), 2normal use (7.5 scfm [210 sl/prior to making connections t

r Protrude 1" (25.4 mm) for 3/8" compression 60 ± 20 psi (410 ±140 KPa), 1.exceed 40°C (104°F). Flush linequipment.

al vacuum Protrude 1" (25.4 mm) for 5/8" (16 mm) O.D. tubing and other specified by the central vacuum supplier

Wet Systems: 10 ± 2 inches ofminimum. Dry/Semi-dry Systeand 12 scfm (340 sl/min) min

rical Protrude 1" (25.4 mm) Refer to the electrical require

ty drain Protrude 1" (25.4 mm) Basket strainer connects to 1-flanged tailpiece.

us Oxide Specified by the product manufacturer Specified by the product man

en Specified by the product manufacturer Specified by the product man

16 86.0142.00 Rev T

A-dec

L O

Univ

Round

Figure

sole

Dental Furniture Pre-Installation Guide

O P V E N T D I A G R A M

ersal 5543

and square centrals

8. Loop Vent Diagram - Preference Collection 5543 and 5531

5531 Accessory Con

35" sink module

86.0142 17

PREFERENCE COLLECTION 5631 Wall-Mount Accessory Console

M O

563

Meta

1-1/4"

Top: Afull leconne

Bottom

Wood

Top: #Tableprovidto rein

Bottom

aching to Wall

29-1/8"(739.7 mm)

.00 Rev T

Mounting Specifications

U N T I N G S P E C I F I C A T I O N S

1 Wall-Mount Accessory Console

l Studs

x 3-1/2"x 22 ga minimum

dd minimum 2-1/2" x 22 ga stiffener between studs and behind ngth of wall-mounting cleat. Reference Table 6 for required top ction points.

: #9 x 2-1/4" S.D. screw in each stud

Studs

10 x 3" wood screws into studs or wood blocking. Reference 6 for required top connection points. If available studs do not e adequate mounting locations, wood blocking must be added force the framing.

: #10 x 3" wood screws in each stud

Figure 9. 5631 Mounting Cleat Att.

WARNING If the ends of the wall-mounting cleat are more than 10" (254 mm) from a direct stud attachment, you must install a toggle wall fastener to attach the cleat to the gypsum wallboard.

Install the additional toggle wall fastener approximately 10" (254 mm) from the direct stud attachment point. If the proper number of stud attachments cannot be made, you must change the module position or reinforce the wall framing. Refer to page 2 for required consultation.

18 86.0142.00 Rev T

A-dec

Table 6

Figure

Size

30" 762 m

44" 1118

58" 1473

72" 1829

86" 2184

Dental Furniture Pre-Installation Guide

. Hanging Weight Chart (laminate, solid surface / Quartz)

10. 5631 Mounting Specifications

Cabinet Weight Live Load Total Top Connection Points Required

Wall Studs Required

m135 / 145 lb.61 / 66 kg

200 / 190 lb.91 / 86 kg

335 lb. 152 kg

2 2

mm198 / 212 lb.90 / 96 kg

330 / 316 lb. 150 / 143 kg

528 lb. 239 kg

3 2

mm261 / 281 lb.118 / 127 kg

400 / 380 lb. 181 / 172 kg

661 lb. 300 kg

6 3

mm324 / 348 lb. 147 / 159 kg

350 / 326 lb.159 / 148 kg

674 lb.306 kg

7 3

mm387 / 416 lb.176 / 189 kg

500 / 471 lb.227 / 214 kg

887 lb.402 kg

8 4

86.0142 19

PREFERENCE COLLECTION0 Dispenser Unit or 5731 Storage Unit

573

Meta

1-1/4"

Top: 2

T.W. a

Bottom

Wood

Std or

Top: #

Bottom

.00 Rev T

Mounting Specifications 573

0 Dispenser Unit or 5731 Storage Unit

l Studs

x 22 ga min

2" (558.8 mm), 29" (736.6 mm), 34" (863.6 mm), and 42" (1066.8 mm)

nchors in stud flange

: #9 x 2-1/4" S.D. screw in each stud

Studs

Better/D.F. or S.P.

10 x 3" wood screws

: #10 x 3" wood screws in each stud

WARNING For .22 or .29 modules (22-5/8" [574.7 mm] or 29-5/8" [752.5 mm] wide), you must mount the module anchor strips directly to wall studs in two locations (see Figure 12).

For .32 or .42 modules (34" [86.3 mm] or 42-1/8" [1070 mm] wide), you must mount the anchor strip directly to wall studs in a minimum of three locations (see Figure 12). If the ends of the anchor clip are more than 10" (254 mm) from a direct stud attachment, you must install an additional toggle bolt anchor to attach the anchor strip to the gypsum wallboard. Install the additional toggle bolt anchor approximately 10" (254 mm) from the direct stud attachment point.

If you cannot make the proper number of stud attachments, you must change the module position or reinforce the wall framing. Refer to page 2 for required consultation.

20 86.0142.00 Rev T

A-dec

Figure

Figure

Dental Furniture Pre-Installation Guide

11. 5730 and 5731 Mounting Cleat Attaching to Wall

12. 5730 and 5731 Mounting Specifications

68-1/2"(1740 mm)

86.0142 21

PREFERENCE COLLECTIONent Column Insert with Pass-Through

F R

566Inse

Knowrough

Table

Thereplacem

662 Treatment Column Pass-Through

Catal

ShelfPaneMSTC

Flip-dAcceMSTC

Flip-dShelfMSTC

.00 Rev T

Framing Dimensions 5662 34" and 42" Treatm

A M I N G D I M E N S I O N S

2 34" and 42" Treatment Column rt with Pass-Through

ledge of the product configuration is necessary to determine -in dimensions.

7. 5662 34" and 42" (with Pass-Through) Configuration

has been 1/2" (12.7 mm) overall added to the dimensions for ent of cabinet.

Figure 13. Framing Dimensions - 5

og Options Pass-Through Storage

and Accessory lC.15

72-1/2" Top of Upper 50-7/8" Bottom of Upper

own Door and ssory Panel C.16

72-1/2" Top of Upper50-7/8" Bottom of Upper

own Door and C.17

72-1/2" Top of Upper 50-7/8" Bottom of Upper

Note: Blocking required here (typical)

22 86.0142.00 Rev T

A-dec

566Inse

Knowlrough-

Table 8

There placem

662 Treatment Column with

Catal

Shelf Panel

Flip-dAccesMSTC

Flip-dMSTC

Dental Furniture Pre-Installation Guide

2 34" and 42" Treatment Column rt with Non-Pass-Through

edge of the product configuration is necessary to determine in dimensions.

. 5662 34" and 42" (with Non-Pass-Through) Configuration

has been 1/2" (12.7 mm) overall added to the dimensions for ent of cabinet.

Figure 14. Framing Dimensions - 5Non-Pass-Through

og Options Non-Pass-Through Storage Cap Only

and Accessory MSTCC.15

72-1/2" (1841.5 mm) Top of Upper

56-1/4" (1428.8 mm) Top of Cap

own Door and sory Panel C.16

72-1/2" (1841.5 mm) Top of Upper

56-1/4" (1428.8 mm) Top of Cap

own Door and Shelf C.17

72-1/2" (1841.5 mm) Top of Upper

56-1/4" (1428.8 mm) Top of Cap

Note: Blocking required here (typical)

86.0142 23

PREFERENCE COLLECTIONfications 5731 Upper Storage Insert

F R

573

Minimbeen dDocumpage 2

Meta

Minim1-1/

Typic#9 x

Wood

MinimStan

Typic#10

.00 Rev T

Framing Speci

A M I N G S P E C I F I C A T I O N S

1 Upper Storage Insert

um requirements for attachment and for wall construction have etermined by an independent structural engineering company. ented detail is available from A-dec customer service. Refer to

for required consultation.

l Stud Wall

um stud:4" x 3-1/2" x 22 ga @ 24" (609.6 mm) o.c.

al Attachment: 2-1/4" self-drilling screws (4 required)

Stud Wall

um stud:dard or better 2" x 4" @ 24" (609.6 mm) o.c.

al Attachment: x 3" self-drilling screws (4 required)

NOTE All dimensions are for rough-in opening with 1/2" (12.7 mm) extra.

24 86.0142.00 Rev T

A-dec

Figure

Dental Furniture Pre-Installation Guide

15. Framing Specifications - 5731 Upper Storage Insert

86.0142 25

PREFERENCE COLLECTION Specifications X-Ray Storage Insert

X-R

Meta

The x-Consuassistaminim

Wood

The x-Consufor assSouthself-dr

Refer is dep

.00 Rev T

Framing

ay Storage Insert

l Stud Wall

ray unit manufacturer requires additional jamb framing. lt the x-ray manufacturer and your local building contractor for nce. The x-ray module requires 1-1/4" x 3-1/2" x 22 gage studs, um. Attach with eight #9 x 2-1/4" self-drilling screws.

Stud Wall

ray unit manufacturer requires additional jamb framing. lt the x-ray manufacturer and your local building contractor istance. The x-ray module requires #2 Douglas Fir, Larch or

ern Pine studs, minimum. Attach with eight #9 x 2-1/4" illing screws.

to page 2 for required consultation. Routing of electrical for x-ray endent on the x-ray manufacturer requirements.

NOTE The 41-1/2" (1054.1 mm) reference dimension is the framing requirement when placing the x-ray module above the 5531 or 5631 at standard height. This allows the x-ray module to be opened without interfering with the sink faucet. If the wall construction limits this height dimension, a special, shorter faucet handle may be ordered as a special part from Delta (call 1.800.345.3358). The chrome handle part number is RP17443. The white handle part number is RP17443WH. Using this faucet allows the reference dimension to be lowered to1-7/8" (47.6 mm) to 39-3/8" (1000.1 mm).

26 86.0142.00 Rev T

A-dec

Figure

Dental Furniture Pre-Installation Guide

16. Framing Specifications - X-Ray Storage Insert

NOTE Dimensions for rough-in opening include 1/2" (12.7 mm) extra, for ease of fit. The header and structural framing is configured per local code.

86.0142 27

PREFERENCE COLLECTION 5531 and 5631 30" Accessory Console

T E

55330" Figure

.00 Rev T

Templates and Utilities

M P L A T E S A N D U T I L I T I E S

1 and 5631 Accessory Console17. Templates and Utilities - 5531 and 5631 30" Accessory Console

28 86.0142.00 Rev T

A-dec

55344" Figure

Dental Furniture Pre-Installation Guide

1 and 5631 Accessory Console18. Templates and Utilities - 5531 and 5631 44" Accessory Console

86.0142 29

PREFERENCE COLLECTION 5531 and 5631 58" Accessory Console

55358" Figure

.00 Rev T

Templates and Utilities

1 and 5631 Accessory Console19. Templates and Utilities - 5531 and 5631 58" Accessory Console

30 86.0142.00 Rev T

A-dec

55372" Figure

Dental Furniture Pre-Installation Guide

1 and 5631Accessory Console20. Templates and Utilities - 5531 and 5631 72" Accessory Console

86.0142 31

PREFERENCE COLLECTION 5531 and 5631 86" Accessory Console

55386" Figure

.00 Rev T

Templates and Utilities

1 and 5631Accessory Console21. Templates and Utilities - 5531 and 5631 86" Accessory Console

32 86.0142.00 Rev T

A-dec

563WallFigure

Dental Furniture Pre-Installation Guide

1 -Mount Accessory Console22. Templates and Utilities - 5631 Wall-Mount Accessory Console

29 1/8"[739.7 mm]

32 7/8"[833 mm]

86.0142 33

PREFERENCE COLLECTION531 Floor-Mounted Utilities Template

553FlooFigure

1

.00 Rev T

Templates and Utilities 5

1r-Mounted Utilities Template23. Templates and Utilities - 5531 Floor-Mounted Utilities

34 86.0142.00 Rev T

A-dec

553WallFigure

Dental Furniture Pre-Installation Guide

1 and 5631-Mounted Utilities24. Templates and Utilities - 5531 and 5631 Wall-Mounted Utilities

86.0142 35

PREFERENCE COLLECTION43 Round/Square-End Central Console

554RouFigure

.00 Rev T

Templates and Utilities 55

3nd/Square-End Central Console25. Templates and Utilities - 5543 Round/Square-End Central Console

36 86.0142.00 Rev T

A-dec

55634" Figure

t configuration supports the use of livery systems, available 2005.

Dental Furniture Pre-Installation Guide

2 Treatment Column and 42"26. Templates and Utilities - 5562 Treatment Column 34" and 42"

This produc541/545 deSeptember

86.0142 37

PREFERENCE COLLECTION 5562 Treatment Column 34" and 42"

Wall-(opti

Figure

.00 Rev T

Templates and Utilities

Mounted Utilities onal)

27. Templates and Utilities - 5562 Wall-Mounted Utilities

38 86.0142.00 Rev T

A-dec

558TreaFigure

Dental Furniture Pre-Installation Guide

0tment Console 42", 69", and 96"

28. Templates and Utilities - 5580 Treatment Console 42", 69", and 96"

86.0142 39

PREFERENCE COLLECTIONole Plumbing (Wall-Mounted Utilities)

558TreaFigure

.00 Rev T

Templates and Utilities 5580 Treatment Cons

0tment Console Plumbing (Wall-Mounted Utilities)

29. Templates and Utilities - 5580 Treatment Console Plumbing (Wall-Mounted Utilities)

40 86.0142.00 Rev T

A-dec

56634" Figure

is product configuration supports the of 541/545 delivery systems.

Dental Furniture Pre-Installation Guide

2 Treatment Insert and 42"30. Templates and Utilities - 5662 Treatment Insert 34" and 42"

Thuse

86.0142 41

PREFERENCE COLLECTION 5662 Treatment Insert 34" and 42"

F U

Listedcorresfull-siA-dec

Table

* The u

Temp Part Number

5531 86.0173.00

5531 86.0174.00

5531 86.0175.00

5531 86.0176.00

5531 86.0177.00

5531 N/A*

5543— 86.0145.00

5562— 86.0147.00

5580— 86.0146.00

5662— 86.0180.00

.00 Rev T

Full-Size Template Reference

L L - S I Z E T E M P L A T E R E F E R E N C E

below are the titles for each full-size template and the ponding page numbers showing a miniature version. To order a ze template, have the template part number ready and contact customer service at 1.800.547.1883 or 1.503.538.7478.

9. Full-Size Templates - Preference Collection

tilities appear on all 5531 and 5631 templates.

late Title Page Number

and 5631—30" Accessory Console page 27

and 5631—44" Accessory Console page 28

and 5631—58" Accessory Console page 29

and 5631—72" Accessory Console page 30

and 5631—86" Accessory Console page 31

and 5631 Utilities* page 32 — 34

Round/Square-End Central Console with Utilities page 35

Treatment Column and Wall-Mounted Utilities 34" and 42" with Utilities page 36 — 37

Treatment Console 42", 69", and 96" with Utilities page 38 — 39

Treatment Insert 34" and 42" with Utilities page 40

42 86.0142.00 Rev T

A-dec

Dental Furniture Pre-Installation Guide

3

86.0142.00 Rev T 43

A-dec Dental Furniture Pre-Install

P R E F E R E N C E

The A-dec Preference ICC is organiconfigurations, which means it cansystem specifically designed for a p

This section of the Dental Furniture

• Estimated Shipping Weights• Utility Specifications• Spotting Back Panel Locations• Utility Templates• Back Panel Templates

ation Guide

I C C

zed into nineteen simple module deliver the flexibility to create a ractice.

Pre-Installation Guide includes:

44 86.0142.00 Rev T

A-dec

E S T

Pref

Table 1

5590

SWD2

SLDW

SSD3

SWS0

MSRU

MSLW

MSLD/

MSSD

MPS4

MDM5

MSMD

MDK0

MWMD

Kilograms

166

166

149

167 / 172

166

166

102

169

181 / 195

45

Dental Furniture Pre-Installation Guide

I M A T E D S H I P P I N G W E I G H T S

erence ICC

0. Estimated Shipping Weights (laminate, solid surface / Quartz)

Pounds Kilograms

349 158

2 / SLDW3 302 137

329 149

308 / 318 140 / 145

285 / 295 129 / 134

/MSRW 247 / 257 112 / 117

MSRD 285 / 295 129 / 134

195 / 205 89 / 93

249 / 259 113 / 118

203 / 213 92 / 97

141 / 151 64 / 69

108-325 / 112-346 49-134 / 51-157

141 / 151 64 / 69

5590 Pounds

CSD3 365

CSS4 366

CLS5 329

CTP4 369 / 379

C2S2B 366

CSS1 366

CST1.12 225

CST1.25 372

CRNR 398 / 428

Ship Kit 100

86.0142 45

PREFERENCE ICCtility Specifications Preference ICC

U T

Table

Utilit

Air 2.5 scfm (71 sl/min) minimum, during /min] peak intermittent flow). Flush tions to dental equipment.

Wate .5 gpm (5.7 l/min) minimum, not to lines prior to making connections to

Centr f Hg (34 ± 7 KPa) and 9 scfm (255 sl/min) tems 4.5 ± 1 inches of Hg (16 ± 3.5 KPa) nimum.

Elect rements listed throughout this section.

Gravi 1-1/4" or 1-1/2" flanged tailpiece.

.00 Rev T

U

I L I T Y S P E C I F I C A T I O N S

11. Utility Specifications - Preference ICC

y Finish Requirement

Protrude 1" (25.4 mm) for 3/8" compression 80-125 psi (550 to 860 KPa),normal use (7.5 scfm [210 sllines prior to making connec

r Protrude 1" (25.4 mm) for 3/8" compression 60 ± 20 psi (410 ±140 KPa), 1exceed 40°C (104°F). Flush dental equipment.

al vacuum Protrude 1" (25.4 mm) for 5/8" (15.9 mm) O.D. tubing and other specified by the central vacuum supplier

Wet Systems: 10 ± 2 inches ominimum. Dry/Semi-dry Sysand 12 scfm (340 sl/min) mi

rical Protrude flex conduit to 24" (609.6 mm) Refer to the electrical requi

ty drain Protrude 1" (25.4 mm) Basket strainer connects to

46 86.0142.00 Rev T

A-dec

Elec

Venti

One (1storag100-12

Ancil

Providutility manufrecept

Gene

1/2" (1contrabe mad20 A b

er is 14 A at 100-120VAC, 50-60 Hz cuit breaker.

nce wiring according to the rence Collection Canadian

ultation.

Dental Furniture Pre-Installation Guide

trical

lation Fan in Clean Module, Sterile Storage Area

) 120VAC outlet is required for each clean module, sterile e area. The operating current for the ventilation fan is 0.3 A at 0VAC, 50-60 Hz for each ventilation fan.

lary Devices

e electrical supply for ancillary devices in the space noted on the templates and back panel templates. Refer to device acturer for supply requirements. There are no electrical duplex acles supplied with a Preference ICC cabinet.

ral Considerations

2.7 mm) flexible conduit, (120VAC, 3 wire) supplied by ctor to protrude 24" (609.6 mm) at rough-in. Connection to e in junction box. Preference ICC cabinets are approved for

reakers.

Optional Dryer

The operating current for the dryand requires a dedicated 20 A cir

Canadian Wiring Kit

Install additional ground impedainstructions provided in the PrefeWiring Kit, p/n 86.0461.00.

Refer to page 2 for required cons

NOTE A-dec supplies junction box when a sink module is chosen. Contractor supplies junction box otherwise.

86.0142 47

PREFERENCE ICCUtility Specifications Electrical

Task

The opwith t

PowerSingle120 VA

Regarone duconfigadditiinstru

Figure

In caspowercannoconfig

Once indivitechni

eiving/Clean Modules Without

n modules with the Auto door option, but s required for each module. Only applies middle modules.

.00 Rev T

Lights

erating current for task lights is 3 A at 100-120VAC, 50-60 Hz he maximum number of task light options.

Requirements—ICC Without Middle Module Task Light, double or triple receiving/clean module applications require

C flush-mounted duplex box for task light power.

dless of the number of modules (single, double or triple), only plex is required to power all the lighting. Standard

urations place power in the left-hand most modules, with onal modules powered using the standard installation ctions shipped with the product.

31. Receiving/Clean Modules Without Middle Modules

es where the left-hand end modules cannot house the duplex , you can relocate power to the right-hand end module. You t place the duplex in the center module of a triple module uration.

you meet the upper task light requirements, refer to the dual module templates provided in the custom Preference ICC cal information packet for anciliary power requirements.

Figure 32. Duplex Positioning—RecMiddle Modules

1) Air supply stub for receiving/cleawithout middle modules. One stub ito receiving/clean modules without

1

48 86.0142.00 Rev T

A-dec

S P O

To spo

1. EsStewi

2. Finthfoumoonca

3. Ondimedi

4. Vesucuth

ations

Dental Furniture Pre-Installation Guide

T T I N G B A C K P A N E L L O C A T I O N S

t back panel cut-out locations:

tablish the proper spacing between the left side of the 5590 rilization Center and the nearest left-hand reference point thin the room or building where you will place the unit (#1).d the location of the specific module being installed (#3) within

e configured cabinet by adding up the widths of the modules nd between the left end of the cabinet and the left end of the dule (#2). (The individual module widths can be found listed

the cabinet illustrations.) Remember to add the width of the end p on the left side of the cabinet.ce a total spotting location has been determined (sum of

stances #1 & #2), mark that point along the edge where the wall ets the floor. This location becomes the base point for all the

mensions found on this module template.rify the exact module configured within the cabinet, and make re you have the correct template for that module. Apply the t-out dimensional information found on the selected template to e proper location on the wall.

Figure 33. Spotting Back Panel Loc

86.0142 49

PREFERENCE ICCUtility Templates SWD2 Module

U T

SWD

Deale

Air, eladjace

Locat

SWD2 Module

1

Floor Floor

.00 Rev T

I L I T Y T E M P L A T E S

2 Module

r-Specified Requirements

ectrical, and water are non-isolated and accessible from the nt sink module.

ion 1: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 34. Back Panel Template -

Floor

50 86.0142.00 Rev T

A-dec

SLD

Air, el

SLDW2 / SLDW3 Module

Floor Floor

Dental Furniture Pre-Installation Guide

W2 / SLDW3 Module

ectrical, and water are accessed from adjacent sink module.

Figure 35. Back Panel Template -

Floor

86.0142 51

PREFERENCE ICCUtility Templates SSD3 Module

SSD

Air is

Locat

SSD3 Module

Floor Floor

4”mm]

10 1/4”[260 mm]

13 1/2”[343 mm]

.00 Rev T

3 Module

accessed from adjacent sink module.

ion 1: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 36. Back Panel Template -

Floor

5 3/[146

52 86.0142.00 Rev T

A-dec

SWS

Air is a

Locati

SWS0 Module

Floor Floor

Dental Furniture Pre-Installation Guide

0 Module

ccessed from adjacent sink module.

on 1: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 37. Back Panel Template -

Floor

86.0142 53

PREFERENCE ICCUtility Templates MSRU Module

MSR

Deale

Modulisted

Locat

Locat

Locat

U Module

Floor Floor

3

2

1

.00 Rev T

U Module

r-Specified Requirements

le either uses utilities from adjacent sink module OR uses utilities as shown:

ion 1: Hot/cold waterion 2: Drainion 3: Refer to manufacturer’s equipment specifications for

utility requirements.

Figure 38. Utility Templates - MSR

Floor

54 86.0142.00 Rev T

A-dec

MSL

Deale

Locati

Locati

Locati

Locati

Locati

Locati

W Module

3

4

5

1

2

6

Floor Floor

Dental Furniture Pre-Installation Guide

W Module

r-Specified Requirements

on 1: 24" (609.6 mm) flexible conduit and isolated 120V, 20 A service, with no GFCI required to support Preference ICC

on 2: Hot/cold wateron 3: Air

on 4: Drain

on 5: Vacuum for counter mounted HVEon 6: Refer to manufacturer’s equipment specifications for

utility requirements.

Figure 39. Utility Templates - MSL

NOTE If you have an HVE, thermal washer, ultrasonic, or any other equipment requiring utilities, please see location 5 and 6. All utilities below are optional.

Floor

86.0142 55

PREFERENCE ICCUtility Templates MSLD Module

MSL

Deale

Locat

Locat

Locat

Locat

Locat

Locat

D Module

Floor Floor

1

2

3

4

5

6

.00 Rev T

D Module

r-Specified Requirements

ion 1: 24" (609.6 mm) flexible conduit and isolated 120V, 20 A service, with no GFCI required to support Preference ICC

ion 2: Hot/cold waterion 3: Air

ion 4: Drain

ion 5: Vacuum for counter mounted HVEion 6: Refer to manufacturer’s equipment specifications for

utility requirements.

Figure 40. Utility Templates - MSL

NOTE If you have an HVE, thermal washer, ultrasonic, or any other equipment requiring utilities, please see location 5 and 6. All utilities below are optional.

Floor

56 86.0142.00 Rev T

A-dec

MSR

Deale

Locati

Locati

Locati

Locati

Locati

Locati

W Module

3

4

5

1

2

6

Floor Floor

Dental Furniture Pre-Installation Guide

W Module

r-Specified Requirements

on 1: 24" (609.6 mm) flexible conduit and isolated 120V, 20 A service, with no GFCI required to support Preference ICC

on 2: Hot/cold wateron 3: Air

on 4: Drain

on 5: Vacuum for counter mounted HVEon 6: Refer to manufacturer’s equipment specifications for

utility requirements.

Figure 41. Utility Templates - MSR

NOTE If you have an HVE, thermal washer, ultrasonic, or any other equipment requiring utilities, please see location 5 and 6. All utilities below are optional.

Floor

86.0142 57

PREFERENCE ICCUtility Templates MSRD Module

MSR

Deale

Locat

Locat

Locat

Locat

Locat

Locat

D Module

3

4

5

1

2

6

Floor Floor

.00 Rev T

D Module

r-Specified Requirements

ion 1: 24" (609.6 mm) flexible conduit and isolated 120V, 20 A service, with no GFCI required to support Preference ICC

ion 2: Hot/cold waterion 3: Air

ion 4: Drain

ion 5: Vacuum for counter mounted HVEion 6: Refer to manufacturer’s equipment specifications for

utility requirements.

Figure 42. Utility Templates - MSR

NOTE If you have an HVE, thermal washer, ultrasonic, or any other equipment requiring utilities, please see location 5 and 6. All utilities below are optional.

Floor

58 86.0142.00 Rev T

A-dec

MSS

Deale

Locati

Locati

Locati

Locati

Locati

Locati

D Module

3 6

4

5

2

1

Floor Floor

Dental Furniture Pre-Installation Guide

D Module

r-Specified Requirements

on 1: 24" (609.6 mm) flexible conduit and isolated 120V, 20 A service, with no GFCI required to support Preference ICC

on 2: Hot/cold wateron 3: Air

on 4: Drain

on 5: Vacuum for counter mounted HVEon 6: Refer to manufacturer’s equipment specifications for

utility requirements.

Figure 43. Utility Templates - MSS

NOTE If you have an HVE, thermal washer, ultrasonic, or any other equipment requiring utilities, please see location 5 and 6. All utilities below are optional.

86.0142 59

PREFERENCE ICCUtility Templates MPS4 Module

MPS

No de

MPS4 Module

Floor Floor

Available space for the placement of utilities to support ancillary devices within the unit

.00 Rev T

4 Module

dicated utilities required.

Figure 44. Back Panel Template -

Floor

60 86.0142.00 Rev T

A-dec

MDM

No de

MDM5 Module

Floor Floor

Dental Furniture Pre-Installation Guide

5 Module

dicated utilities required.

Figure 45. Back Panel Templates -

86.0142 61

PREFERENCE ICCUtility Templates MSMD Module

MSM

No de

MSMD Module

Floor Floor

.00 Rev T

D Module

dicated utilities required.

Figure 46. Back Panel Template -

62 86.0142.00 Rev T

A-dec

MDK

No de

MDK0 Module

Floor Floor

Dental Furniture Pre-Installation Guide

0 Module

dicated utilities required.

Figure 47. Back Panel Template -

Floor

86.0142 63

PREFERENCE ICCUtility Templates MWMD Module

MW

No de

MWMD Module

.00 Rev T

MD Module

dicated utilities required.

Figure 48. Back Panel Template -

64 86.0142.00 Rev T

A-dec

B A

CSD

Deale

Locati

Locati

CSD3 Module

1 2

Floor Floor

Dental Furniture Pre-Installation Guide

C K P A N E L T E M P L A T E S

3 Module

r-Specified Requirements

on 1: Non-isolated 120V Quad, 20 A, with no GFCI required for ventilation fan and 120V ancillary

on 2: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 49. Back Panel Templates -

Floor

86.0142 65

PREFERENCE ICCBack Panel Templates CSS4 Module

CSS

Deale

Locat

Locat

CSS4 Module

1 2

Floor Floor

.00 Rev T

4 Module

r-Specified Requirements

ion 1: Non-isolated 120V quad, 20 A, with no GFCI required for ventilation fan and 120V ancillaries

ion 2: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 50. Back Panel Template -

Floor

66 86.0142.00 Rev T

A-dec

CLS5

Deale

Locati

Locati

CLS5 Module

1 2

Floor Floor

Dental Furniture Pre-Installation Guide

Module

r-Specified Requirements

on 1: Non-isolated 120V Quad, 20 A, with no GFCI required for ventilation fan and 120V ancillaries

on 2: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 51. Back Panel Templates -

Floor

86.0142 67

PREFERENCE ICCBack Panel Templates CTP4 Module

CTP

Deale

Locat

Locat

CTP4 Module

1 2

Floor Floor

Available space for the placement of utilities to support ancillary devices within the unit

.00 Rev T

4 Module

r-Specified Requirements

ion 1: Non-isolated 120V Quad, 20 A, with no GFCI required for ventilation fan and 120V ancillaries

ion 2: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 52. Back Panel Template -

Floor

68 86.0142.00 Rev T

A-dec

C2S

Deale

Locati

Locati

C2S2B Module

Floor Floor

Dental Furniture Pre-Installation Guide

2B Module

r-Specified Requirements

on 1: Non-isolated 120V Quad, 20 A, with no GFCI required for ventilation fan and 120V ancillaries

on 2: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 53. Back Panel Template -

Floor

86.0142 69

PREFERENCE ICCBack Panel Templates CSS1 Module

CSS

Deale

Locat

Locat

CSS1 Module

1 2

roolFroolF

.00 Rev T

1 Module

r-Specified Requirements

ion 1: Non-isolated 120V Quad, 20 A, with no GFCI required for ventilation fan and 120V ancillaries

ion 2: Refer to manufacturer’s equipment specifications for utility requirements.

Figure 54. Back Panel Template -

roolF

70 86.0142.00 Rev T

A-dec

CST

No de

CST1-12 Module

Flooror

Dental Furniture Pre-Installation Guide

1-12 Module

dicated utilities required.

Figure 55. Back Panel Template -

Flo

86.0142 71

PREFERENCE ICCck Panel Templates CST1-25 Module

CST

No de

CST1-25 Module

Floorr Floor

.00 Rev T

Ba

1-25 Module

dicated utilities required.

Figure 56. Back Panel Template -

Floo

72 86.0142.00 Rev T

A-dec

CRN

No de

CRNR Module

FloorFloor Floor

Dental Furniture Pre-Installation Guide

R Module

dicated utilities required.

Figure 57. Back Panel Template -

4

86.0142.00 Rev T 73

A-dec Dental Furniture Pre-Install

P R E F E R E N C E

The A-dec Preference Slimline utilispace issues in treatment rooms whor constraining.

This section of the Dental Furniture

• Estimated Shipping Weights• Mounting Specifications• Framing and Reinforcing Specifi• Utility Specifications• Full-Size Template Reference

ation Guide

S L I M L I N E

ty consoles are designed to solve ere traditional cabinetry is too large

Pre-Installation Guide includes:

cations

74 86.0142.00 Rev T

A-dec

E S T

PrefTable 1

Mode

5536.

5536.

5537.

5537.

5636.

5636.

5731.

Dental Furniture Pre-Installation Guide

I M A T E D S H I P P I N G W E I G H T S

erence Slimline 2. Estimated Shipping Weights (laminate, solid surface / Quartz)

l Number Pounds Kilograms

18 70 / 75 32 / 34

34 117 / 125 53 / 57

18 75 / 80 34 / 36

34 126 / 134 57 / 61

18 56 / 61 25 / 28

34 90 / 98 41 / 45

34 x 9" 67 30

86.0142 75

PREFERENCE SLIMLINEns 5636 Utility Console, Wall-Mount

M O

563

Meta

1-1/4"

Top: Afull lerequirand/o

Bottom

Wood

Two m

Top: Rwood

Bottom

Refer

taching to Wall

.00 Rev T

Mounting Specificatio

U N T I N G S P E C I F I C A T I O N S

6 Utility Console, Wall-Mount

l Stud Walls

x 3-1/2" x 22 ga minimum.

dd minimum 2-1/2" x 22 ga stiffener between studs and behind ngth of mounting cleat. Reference Table 13 on page 77 for ed quantity of T.W. anchors to be installed through studs r stiffener.

: #9 x 2-1/4" S.D. screw in each stud.

Stud Walls

inimum std or better/D.F. or S.P.

eference Table 13 on page 77 for required quantity of #10 X 3" screws to be installed through studs and/or stiffener.

: #10 x 3" wood screws into two studs minimum.

to page 2 for required consultation.

Figure 58. 5636 Mounting Cleat At

WARNING Consult a licensed professional, i.e. structural or architectural engineer, for mounting to metal stud or masonry walls.Ensure that all installations conform to local code requirements. Mounting structures must be designed to support a minimum shear load of 100 lb. (445N) and a minimum moment of 112 ft-lbs. (152 N-m).

27 1/4” (692 mm)

(distance from floor

to bottom of cleat)

76 86.0142.00 Rev T

A-dec

F R AS P E

563

Fram

Minimbeen dDocumrequir

MetalMinim1-1/4"

Typica#9 x 2-

WoodMinim2x4 @

Typica#10 x 3

nt

Dental Furniture Pre-Installation Guide

M I N G A N D R E I N F O R C I N G C I F I C A T I O N S

6 Utility Console, Wall-Mount

ing

um requirements for attachment and for wall construction have etermined by an independent structural engineering company. ented detail is available from A-dec. Refer to page 2 for

ed consultation.

Stud Wallum stud: (31.8 mm) x 3-1/2" (89 mm) x 22 ga @ 24" o.c.

l Attachment:1/4" (57.1 mm) self-drilling screws.

Stud Wallum stud standard or better:24" (609.6 mm) o.c.

l Attachment:" (76.2 mm) self-drilling screws.

Figure 59. 5636 Wall Reinforceme

Reinforcement or bracing added by contractor

Existing wall stud

86.0142 77

PREFERENCE SLIMLINEns 5636 Utility Console, Wall-Mount

Rein

You m our local building contractor, state and local ction. Refer to page 2 for required consu

Table

MetaMeta ions and local and state codes.

WoodWoo ions and local and state codes.

Utili

18'' (

34'' (

.00 Rev T

Framing and Reinforcing Specificatio

forcing

ust make attachments to stud walls and reinforcements when mounting the utility console. Consult ycodes when mounting a utility console to walls with something other than wood or metal stud construltation.

13. Minimum Number of Direct Attachments

l Stud Wallsl stud constructed walls must be reinforced or braced per building contractors recommendat

Stud Wallsd stud constructed walls must be reinforced or braced per building contractors recommendat

ty Console Length Screws to Mount Cleat

457.2 mm) 2

863.6 mm) 4

CAUTION Be sure the wall studs, reinforcement/bracing supports a minimum of:

- 150 lb. (68 kg) for a 18" (457.2 mm) utility console

- 200 lb. (91 kg) for a 34" (863.6 mm) utility console

78 86.0142.00 Rev T

A-dec

Figure

Figure

Dental Furniture Pre-Installation Guide

60. Reinforcing - 5636 18" Utility Console Wall-Mount

61. Reinforcing - 5636 34" Utility Console Wall-Mount

86.0142 79

PREFERENCE SLIMLINEns 5636 Utility Console, Wall-Mount

U T

Table

Utilit

Air 5 scfm (71 sl/min) minimum, during in] peak intermittent flow. Flush lines dental equipment.

Wate gpm (5.7 l/min) minimum, not to es prior to making connections to dental

Centr Hg (34 ± 7 KPa) and 9 scfm (255 sl/min) s 4.5 ± 1 inches of Hg (16 ± 3.5 KPa) and

.

Elect ments listed throughout this section.

.00 Rev T

Utility Specificatio

I L I T Y S P E C I F I C A T I O N S

14. Utility Specifications - Preference Slimline

y Finish Requirement

Protrude 1" (25.4 mm) for 3/8" compression 80-125 psi (550 to 860 KPa), 2.normal use (7.5 scfm [210 sl/mprior to making connections to

r Protrude 1" (25.4 mm) for 3/8" compression 60 ± 20 psi (410 ±140 KPa), 1.5exceed 40°C (104°F). Flush linequipment.

al vacuum Protrude 1" (25.4 mm) for 5/8" (16 mm) O.D. tubing and other specified by the central vacuum supplier

Wet Systems: 10 ± 2 inches of minimum. Dry/Semi-dry System12 scfm (340 sl/min) minimum

rical Protrude 1" (25.4 mm) Refer to the electrical require

80 86.0142.00 Rev T

A-dec

553

This p

Figure

Dental Furniture Pre-Installation Guide

6 18" Utility Console

roduct configuration supports the use of 541/545 delivery systems.

62. Utility Specifications - 5536 18" Utility Console

86.0142 81

PREFERENCE SLIMLINEtions 5536 34" Utility Console

553

Left-

This p

Figure

.00 Rev T

Utility Specifica

6 34" Utility Console

Side Delivery

roduct configuration supports the use of 541/545 delivery systems.

63. Utility Specifications - 5536 34" Utility Console - Left-Side Delivery

Fold Line

82 86.0142.00 Rev T

A-dec

Right

This p

Figure

Dental Furniture Pre-Installation Guide

-Side Delivery

roduct configuration supports the use of 541/545 delivery systems

64. Utility Specifications - 5536 34" Utility Console - Right-Side Delivery

Fold Line

86.0142 83

PREFERENCE SLIMLINEcifications 5537 18" Utility Console

553

This p

Figure

.00 Rev T

Utility Spe

7 18" Utility Console

roduct configuration supports the use of 541/545 delivery systems.

65. Utility Specifications - 5537 18" Utility Console

84 86.0142.00 Rev T

A-dec

553

Left-

This p

Figure

Fold Line

Dental Furniture Pre-Installation Guide

7 34" Utility Console

Side Delivery

roduct configuration supports the use of 541/545 delivery systems.

66. Utility Specifications - 5537 34" Utility Console - Left-Side Delivery

86.0142 85

PREFERENCE SLIMLINEtions 5537 34" Utility Console

Right

This p

Figure

.00 Rev T

Utility Specifica

-Side Delivery

roduct configuration supports the use of 541/545 delivery systems.

67. Utility Specifications - 5537 34" Utility Console - Right-Side Delivery

Fold Line

86 86.0142.00 Rev T

A-dec

563

Figure

Dental Furniture Pre-Installation Guide

6 18" Utility Console Wall-Mount

68. Utility Specifications - 5636 18" Utility Console Wall-Mount

NOTE No plumbing available through the floor directly under the cabinet. Refer to Optional Template “A” for location of optional utilities. In cases where wall plumbing is not an option, refer to Optional Template “B.”

27 1/4” [692 mm] (distance from floor to bottom of cleat)

86.0142 87

PREFERENCE SLIMLINE 5636 18" Utility Console Wall-Mount

Optio

You c llow the placement of utilities within the wa and routing the necessary utilities throug

Figure

.00 Rev T

Utility Specifications

nal Utilities

an use an optional cutout in the bottom of the cabinet when the wall is made of material that does not all. In these applications, the dental dealership and/or contractor assumes all responsibility for spottingh the floor and into the optional cutout in the bottom of the cabinet.

69. Utility Specifications - 5636 18" Utility Console Wall-Mount - optional utilities

88 86.0142.00 Rev T

A-dec

563

Left-

Figure

27 1/4” [692 mm] (distance from floor to bottom of cleat)

Dental Furniture Pre-Installation Guide

6 34" Utility Console

Side Delivery

70. Utility Specifications - 5636 34" Utility Console - Left-Side Delivery

NOTE No plumbing available through the floor directly under the cabinet. Refer to Optional Template “A” for location of optional utilities. In cases where wall plumbing is not an option, refer to Optional Template “B.”

86.0142 89

PREFERENCE SLIMLINEcifications 5636 34" Utility Console

OptioYou c llow the placement of utilities within the wa and routing the necessary utilities throug

Figure

.00 Rev T

Utility Spe

nal Utilitiesan use an optional cutout in the bottom of the cabinet when the wall is made of material that does not all. In these applications, the dental dealership and/or contractor assumes all responsibility for spottingh the floor and into the optional cutout in the bottom of the cabinet.

71. Utility Specifications - 5636 34" Utility Console - Left-Side Delivery, optional utilities

90 86.0142.00 Rev T

A-dec

Right

Figure

Dental Furniture Pre-Installation Guide

-Side Delivery

72. Utility Specifications - 5636 34" Utility Console - Right-Side Delivery

NOTE No plumbing available through the floor directly under the cabinet. Refer to Optional Template “A” for wall plumbing location of optional utilities. In cases where wall plumbing is not an option, refer to Optional Template “B.”

27 1/4” [692 mm] (distance from floor to bottom of cleat)

86.0142 91

PREFERENCE SLIMLINEcifications 5636 34" Utility Console

OptioFigure

.00 Rev T

Utility Spe

nal Utilities73. Utility Specifications - 5636 34" Utility Console - Right-Side Delivery, optional utilities

92 86.0142.00 Rev T

A-dec

F U

Listedcorrespfull-sizservice

Table 1

Temp Part Number

5536 86.0907.00

5537 86.0910.00

5636 86.0908.00

5636 86.0909.00

Dental Furniture Pre-Installation Guide

L L - S I Z E T E M P L A T E R E F E R E N C E

below are the titles for each full-size template and the onding page numbers showing a miniature version. To order a e template, have the part number and contact A-dec customer at 1.800.547.1883 or 1.503.538.7478.

5. Full-Size Templates - Preference Slimline

late Title Page Number

- 18" and 34" Utility Console with Utilities page 80 through page 82

- 18" and 34" Utility Console with Utilities page 83 through page 85

- 18" Utility Console with Utilities page 86 through page 87

- 34" Utility Console with Utilities page 88 through page 91

 

86.0142.00 Rev TCopyright 2012 A-dec Inc.

All rights reserved.

A-dec Headquarters2601 Crestview DriveNewberg, OR 97132 USATel: 1.800.547.1883 Within USA/CanadaTel: 1.503.538.7478 Outside USA/CanadaFax:1.503.538.0276www.a-dec.com

International Distribution Centers

A-dec United KingdomTel: 0800 ADECUK (233285) Within UKTel: +44 (0) 24 7635 0901 Outside UKwww.a-dec.co.uk

A-dec AustraliaTel: 1.800.225.010 Within AustraliaTel: +61 (0)2 8332 4000 Outside Australiawww.a-dec.com.au