Product InformationLoad cell for tensile and compression from 330 kN to 600 kN
PI651_KA330-600.pmd
D2
d2
d1
D1
H
PI 6
51 2
.011
4
Figure: Load cell 600 kN Figure: Drawing of load cell with dimenions
Key Benefits
• through holes for mounting bolts simplify installationand allow a wide range of testing situations to beaccommodated
• hermetically sealed, corrosion-resistant design for highlong-term reliability
Further advantages und features
• Rotational symmetrical measurement body (computeroptimised), thus measurement properties areindependent of the mounting location
• Independent of the stiffness (and so deflection) of theforce transmission
• Insensitive to eccentric force transmission and lateralforces
• No influence through torsional moments
• Low creep error even at abrupt load removal(specimen break)
• The inside of the extensometer is filled with protectivegas, so no reaction with oxygen (from air or moisture)is possible
• All hollows are laser welded, and so a long-time stableand very high consistency will be reached towards theenvironment
• In fatigue tests controlled firmness means quasi-unlimited life of the materials testing machine
• High static overloading gives more protection to theuser (tensile strength 300% nominal force)
Product InformationLoad cell for tensile and compression from 330 kN to 600 kN
All Data at ambient temperature. All rights reserved.
Fmax (tensile/compression) [kN] 3301 3301 2 4001 4001 2 6001 6001 2
Accuracy grade 1 from .. on [kN] 0.66 0.66 0.8 0.8 1.2 1.2Accuracy grade 0.5 from .. on [kN] 3.3 3.3 4.0 4.0 6.0 6.0Nominal measurement path [mm] 0.17 0.17 0.17 0.17 0.17 0.17DimensionsH – Height [mm] 160 160 160 160 160 160D1 – Upper mounting stud dia [mm] 250 250 250 250 250 250d1 – Upper mounting stud dia [mm] 220 220 220 220 220 220D2 – Lower mounting stud dia [mm] 220 220 220 220 220 220d2 – Lower mounting stud dia [mm] 184 184 184 184 184 184
Eccentricity influence / mm 0.02 % 0.02 % 0.02 % 0.02 % 0.02 % 0.02 %Lateral force influenceat 0.1 x Fnom on theload cell upper side 0.02 % 0.02 % 0.02 % 0.02 % 0.02 % 0.02 %Reference temperature [°C] 22 (±1) 22 (±1) 22 (±1) 22 (±1) 22 (±1) 22 (±1)Nominal temperature range [°C] +10 ... +60 +10 ... +60 +10 ... +60 +10 ... +60 +10 ... +60 +10 ... +60Use temperature range [°C] -30 ... +85 -30 ... +85 -30 ... +85 -30 ... +85 -30 ... +85 -30 ... +85Storage temperature range [°C] -50 ... +85 -50 ... +85 -50 ... +85 -50 ... +85 -50 ... +85 -50 ... +85Limit force 150 % 150 % 150 % 150 % 150 % 150 %Static limit lateral force 100 % 100 % 100 % 100 % 100 % 100 %Tensile force 300 % 300 % 300 % 300 % 300 % 300 %Protection system to DIN 40050-1 IP67 IP67 IP67 IP67 IP67 IP67Item no. • 630383 • 089308 • 358453 • 358454 • 358455 • 628347
(BPC-L0330KN.G03) (BPC-L0330KN.T03) (BPC-L0400KN.G03) (BPC-L0400KN.T03) (BPC-L0600KN.G03) (BPC-L0600KN.T03)
1 with calibration certificate according to DIN EN ISO 7500-12 with additional mounting stud M 52 x 1.5 for second test area