Page 1: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

1

Copyright 2010 ITRI工業技術研究院

Agilent 35670A Operator’s Guide

Different types of measurement.

• Rotating Machinery

• Structures(FRF)

• Sound/Acoustics

• Spectra and Networks

• Control Systems

Page 2: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

2

Copyright 2010 ITRI工業技術研究院

• HP 35670A (Front Panel)

Page 3: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

3

Copyright 2010 ITRI工業技術研究院

Spectrum Analyzer

• Initialize the analyzer- [Preset]• [Inst Mode]-FFT Analysis, channels• [Meas Data] [Active Trace]- selecting which type of data you w

ant to display ,PWR SPEC,LIN SPEC,TIME CHANNEL(1),• [Freq] –span measurements, resolution, record length• [Input] – Measurement parameters affect the way the analyzer take

s input data, RANGE,FRONT END CH1 SET UP, XDCR UNIT CH1 SET UP

grounded/floating input coupling ac/dc

anti-alias filter on/off autoranging on/off

A-weight filter on/off ICP power supply on/off

Sensitivity on/off

Page 4: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

4

Copyright 2010 ITRI工業技術研究院

• [Window]- Selecting a windowing function, HANN, FLAT TOP, U

NIFORM,FORCE EXPO, FORCE EXPO SETUP

• [Trigger]- Selecting a triggering function, Free Run Trigger, External Trigger, Channel Trigger, Source Trigger, GPIB Trigger

• [Avg]- average on/off, number averages, average type,fast avg on/off, update rate, repeat on/off, overlap %, overload rej on/off, average preview off/manual/time

• [Source]-source on/off, level, dc offset,

Page 5: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

5

Copyright 2010 ITRI工業技術研究院

• [Trace Coord]-Linear Magnitude, Log Magnitude, db Magnitude, phase, wrapped phase, x units,Y units

• [Analysis]- define function, define constant, data edit, curve fit, synthsis

• [Scale]- autoscale on/off, match Y scale, axes scale markers

• [Disp Format]- single, upper/lower, front/back, upper/big lower, quad, measurement state, waterfall setup

• [Marker]- marker on/off, coupled on/off, peak trk on/off

• [Marker Fctn]- harmonic marker, band marker, sideband marker , freq & damping

Page 6: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

6

Copyright 2010 ITRI工業技術研究院

Measuring Rotating Machinery

1. Initialize the analyzer.• Press [Preset] [DO PRESET].• Press [Inst Mode] [FFT ANALYSIS].• Press [CHANNELS 1 2] (or [CHANNELS 1 2 4]) Press [Meas Data], then pr

ess [CHANNEL 1 2] (or [CHANNEL 1 2 3 4]) to highlight the channel whose transducer you will calibrate.

• Press [PWR SPEC CHANNEL x].2. Match the active trace’s display unit to the calibrator’s output unit.• Press [Input], then press [CHANNEL 1 2] (or [CHANNEL 1 2 3 4]) to highlig

ht the channel whose transducer you will calibrate.• Press [XDCR UNIT CHx SETUP] [XDCR UNIT LABEL], then press the softk

ey corresponding to your calibrator’s output unit. • Press [Input] [XDCR UNIT CHx SETUP], then press [XDCR UNIT ON OFF]

to highlight ON.• Press [Trace Coord] [Y UNITS] [EU].• Press [AMPLITUDE PK PP RMS].

Page 7: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

7

Copyright 2010 ITRI工業技術研究院

3. Connect the transducer’s output cable to the analyzer as described at the beginning of this chapter, then secure the transducer to the calibrator.

4. Specify the measurement parameters.• Press [Freq] [SPAN], then press the down-arrow key if you want to reduce t

he frequency span.• Press [Window] [FLAT TOP] to maximize amplitude accuracy.5. Press [Scale], then press [AUTOSCALE ON OFF] to highlight ON.6. Turn on the calibrator, then press [Start].7. After the display has been updated at least once, move the marker to the cal

ibrator’s output frequency and enter calibrator’s output amplitude.8. Press [Marker] [MARKER TO PEAK].• Press [Input] [XDCR UNIT CHx SETUP] [CAL VALUE AT MARKER], then t

ype the calibrator’s specified output amplitude and press [ENTER].

Page 8: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

8

Copyright 2010 ITRI工業技術研究院

Measuring Structures(FRF)

Page 9: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

9

Copyright 2010 ITRI工業技術研究院

Page 10: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

10

Copyright 2010 ITRI工業技術研究院

1. Initialize the analyzer.• Press [Preset] [DO PRESET].• Press [Active Trace] [A B].• Press [Meas Data] [FREQ RESP 2/1].2. Connect the device-under-test (DUT) as shown in the illustration below.3. Specify the measurement parameters.• Press [Freq] [START] <number> <unit>, then press [STOP] <number> <unit>.• Press [Input].• Press [CHANNEL 1 2 3 4] to highlight 1.• Press [CHANNEL 1 RANGE] <number> <unit>.• Press [CHANNEL 1 2 3 4] to highlight 2.• Press [CHANNEL 2 RANGE] <number> <unit>.• Press [Window] [FORCE EXPO].• Press [FORC EXPO SETUP].• Press [CHANNEL 1 FORC EXPO] to highlight FORC.• Press [FORCE WIDTH] <number> <unit>.• Press [CHANNEL 2 FORC EXPO] to highlight EXPO.• Press [EXPO DECAY] <number> <unit>.

Page 11: Copyright 2010 ITRI 工業技術研究院 1 Agilent 35670A Operator’s Guide Different types of measurement. Rotating Machinery Structures(FRF) Sound/Acoustics Spectra

11

Copyright 2010 ITRI工業技術研究院

4. Specify the triggering and averaging parameters.• Press [Trigger] [CHANNEL 1 2 3 4] to highlight 1.• Press [TRIGGER SETUP] [CHANNEL LEVEL] <number> [PERCENT %]• Press [CHANNEL 1 DELAY] [+/-] <number> <unit>.• Press [CHANNEL 1 2 3 4] to highlight 2.• Press [CHANNEL 2 DELAY] [+/-] <number> <unit>.• Press [Avg], then press [AVERAGE ON OFF] to highlight ON.5. Configure the display.• Press [Trace Coord] [X-AXIS LIN LOG] to highlight LIN.• Press [Active Trace] [B].• Press [Trace Coord] [PHASE].• Press [Active Trace] [A B].• Press [Scale], then press[AUTOSCALE ON OFF] to highlight ON.6. Measure the DUT.• Press [Start].• Slowly tap the device 10 times with the impact hammer at or near the accele

rometer.