Before using the instrument, be sure to read this manual thoroughly. Also, read the manuals of other instruments in this system. Keep the manual where it is easily accessible. Canon CXDI-1 System Digital Radiography Operation Manual

Upload: lyliem

Post on 28-Aug-2018

218 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Before using the instrument, be sure to read this manual thoroughly.Also, read the manuals of other instruments in this system.Keep the manual where it is easily accessible.

Canon CXDI-1 System Digital Radiography

Operation Manual

Page 2: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

PLEASE NOTE

1. Roentgenography, image processing, reading of images, and storage of data must be performed in accordance with the law of the country where the product is being used.Also, the user is responsible for maintaining the privacy of image data.

2. In no event will Canon be liable for direct or indirect consequential damage arising out of the use of this product.Canon will not be liable for loss of image data due to any reason.

• Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries.

• Ethernet is a trademark of Xerox Corporation.• Swiss 721 is a trademark of Bitstream Inc.• Swiss 721 is copyright 1991-1998 Bitstream Inc., and have been licensed to Canon Inc.• TrueType is a trademark of Apple Computer.• Other systems and product names in this manual are the trademarks of each manufacturer that

developed them.• This software is based in part on the work of the Independent JPEG Group.• The word “DEP” used in this manual is an abbreviation of “Display Enhancement Processing”.

© CANON INC. 2009

All rights reserved.Under copyright laws, this manual may not be reproduced, in whole or in part, without the written permission of Canon.

Page 3: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Contents

Starting Up/Shutting Down ............................................ 1

1. Starting Up/Shutting Down the CXDI ....................................31.1 Starting Up the CXDI .................................................................... 3

1.1.1 Turn ON the Power .................................................................................... 31.1.2 Welcome Message .................................................................................... 31.1.3 Exposure Screen Display .......................................................................... 4

1.2 Shutting Down the CXDI ............................................................... 51.2.1 To Shut Down the CXDI ............................................................................ 51.2.2 To Restart the CXDI .................................................................................. 6

Exposure and Output ..................................................... 7

2. Taking an Image ....................................................................92.1 Entering Patient Data .................................................................. 11

2.1.1 Entering Data from the Operation Panel ................................................. 112.1.2 Acquiring Data from a Magnetic Card Reader (optional unit) .................. 142.1.3 Acquiring Data by HIS/RIS or from a Database File ................................ 14

2.2 Selecting and Changing the Exposure Mode .............................. 152.2.1 Selecting an Exposure Mode ................................................................... 152.2.2 Changing Generator Parameters after Exposure Mode Selection .......... 18

2.3 Taking an Image ......................................................................... 202.3.1 Starting Exposure .................................................................................... 202.3.2 Cancelling the Exposure .......................................................................... 222.3.3 Displays after Exposure ........................................................................... 23

2.4 Operations from the Exposure Ready Condition ........................ 252.4.1 Operations from the Exposure Ready Condition ..................................... 252.4.2 Selecting Image Output Destination ........................................................ 27

2.5 END STUDY ............................................................................... 29

3. Output Control (Process Viewer) .........................................31

Viewing the Images ..................................................... 33

4. Browsing Images (Preview Image Screen) .........................354.1 Displaying the Preview Image Screen ........................................ 354.2 Available Operations in the Preview Image Screen .................... 364.3 Selecting Image Output Destination ........................................... 384.4 Rejecting a Bad Image ................................................................ 404.5 Retaking an Image ...................................................................... 41

Page 4: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Contents

5. Viewing Multiple Images (Multi-View Screen) .....................435.1 Displaying the Multi-View Screen ............................................... 435.2 Available Operations on the Multiview Screen ............................ 445.3 Window Operations ..................................................................... 46

5.3.1 Changing the Output Order ..................................................................... 465.3.2 Rejecting a Bad Image ............................................................................ 475.3.3 Checking Images on the Preview Image Screen ..................................... 47

5.4 Multi-Format Output .................................................................... 485.4.1 Choosing Multi-Format before the Exposure ........................................... 485.4.2 Choosing Multi-Format after the Exposure .............................................. 49

6. Browsing Completed Studies (Study List) ...........................536.1 Displaying the Study List ............................................................. 536.2 Customizing the Study List ......................................................... 546.3 Outputting Study Images ............................................................ 566.4 Deleting the Image(s) .................................................................. 58

6.4.1 Deleting All the Images ............................................................................ 586.4.2 Selecting and Deleting Study Images ...................................................... 58

6.5 Reproducing the Image(s) .......................................................... 596.5.1 Resending the Image .............................................................................. 606.5.2 Resending the Study Images .................................................................. 61

Processing the Image .................................................. 63

7. Processing the Image (QA Mode Screen) ...........................657.1 Displaying the QA Screen ........................................................... 657.2 Operations Available on QA Screen ........................................... 667.3 Adjust TRIM and Adjust ROI ....................................................... 68

7.3.1 Trimming the Image ................................................................................. 687.3.2 Changing the Density Adjustment Area ................................................... 69

7.4 LUT ............................................................................................. 707.4.1 AUTO ADJUST. ....................................................................................... 707.4.2 REX ADJUST. ......................................................................................... 73

7.5 EDGE ENHANCE. ...................................................................... 767.6 DEP (Display Enhancement Processing) ................................... 777.7 POSITION ................................................................................... 81

7.7.1 Changing the Trim ................................................................................... 817.7.2 Changing Marks ...................................................................................... 817.7.3 Change Laterality Information ................................................................. 82

7.8 ANNOTATION ............................................................................ 837.8.1 Fixed Annotation ...................................................................................... 837.8.2 Free Annotations ..................................................................................... 85

7.8.2.1 Adding free annotations ............................................................................................ 857.8.2.2 Modifying free annotations ....................................................................................... 88

Page 5: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Contents

7.9 CORRECT PROC. ...................................................................... 907.9.1 Setting the Dosage-Dependent Sharpness Adjustment .......................... 907.9.2 Setting SUPPRESS GRID Processing .................................................... 917.9.3 Setting COLLIMATION MASK Processing Function ............................... 91

7.10 Magnified Image Screen ........................................................... 927.11 Option ....................................................................................... 93

7.11.1 Setting Image Compression .................................................................. 937.11.2 GAMMA Adjustment .............................................................................. 93

CXDI Adjustment ......................................................... 95

8. Calibration ...........................................................................978.1 Setting the Parameters ............................................................... 978.2 Performing Calibration ................................................................ 99

9. Self test .............................................................................1039.1 Self test using a sensor unit with a removable grid unit ............ 1039.2 Self test using the sensor unit with a grid unit which

cannot be removed (CXDI-40G COMPACT) ............................. 106

Troubleshooting ......................................................... 109

10. Before Calling a Service Engineer ..................................11110.1 Fatal Error ............................................................................... 11210.2 Alert ......................................................................................... 11310.3 Error ........................................................................................ 11410.4 Warning ................................................................................... 11910.5 Attention .................................................................................. 122

11. Questions and Answers ..................................................12311.1 Display of Exposure Screen .................................................... 12311.2 Entering/Changing Data .......................................................... 12411.3 Settings ................................................................................... 12511.4 Exposure ................................................................................. 12611.5 Image Output .......................................................................... 127

Page 6: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Contents

Appendix .................................................................... 129

A. Logging in (User Authentication Function) ........................131A.1 User Authentication Function ................................................... 131A.2 Login ......................................................................................... 132A.3 Logout ....................................................................................... 134

A.3.1 Performing Manual Logout .................................................................... 134A.3.2 Auto Logout ........................................................................................... 134

B. Using the Rejection Reason Input Function ......................135B.1 Rejection Reason Input Function ............................................. 135B.2 Operations during Study ........................................................... 136

B.2.1 Rejecting an Image During a Study ...................................................... 136B.2.2 Entering the Rejection Reason during Retake ...................................... 137B.2.3 Changing the Rejection Reason During a Study ................................... 139B.2.4 Cancelling a Rejected Image During a Study ....................................... 140

B.3 Operations after a Study is Completed ..................................... 142B.3.1 Cancelling a Rejected Image and Outputting Again ............................. 142B.3.2 Rejecting an Image and Outputting Again ............................................ 144B.3.3 Changing the Rejection Reason and Outputting Again ......................... 146

C. Using the Sensor Unit Attach/Detach Function ................149C.1 Attach/Detach Function of the Sensor Unit .............................. 149C.2 Detaching the Sensor Unit ....................................................... 150C.3 Connecting a Detached Sensor Unit ........................................ 152

D. Turning the Power of the Power Box On/Off ....................155D.1 Using On/Off for the Power of the Power Box .......................... 155D.2 Turning the Power of the Power Box On/Off ............................ 156

Index .......................................................................... 159

Page 7: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

NotationNotation 1

Buttons that change names when touched are indicated in the text by [Button1]/[Button2].

Example: Two types of gray scale adjustments are available: auto adjustment and REX adjustment, and the type can be selected using [AUTO ADJUST.]/[REX ADJUST.].

Operation: The REX adjustment screen is displayed when [AUTO ADJUST.] is touched, and the auto adjustment screen is displayed when [REX ADJUST.] is touched.

[AUTO ADJUST.] (left) and [REX ADJUST.] (right)

Notation 2Buttons that do not change names when pressed, but switch between normal and highlighted (white character) displays indicate that a function has been enabled.

Example: Each time you touch [EDGE ENHANCEMENT] the button switches between a highlighted display (white characters) and a normal display, edge enhancement adjustments are enabled when highlighted.

Operation: When [EDGE ENHANCEMENT] is highlighted (white characters), edge enhancement adjustments are enabled and can be performed for the image.

Screen appearance when edge enhancement is disabled (left) and when edge enhancement is enabled (right)

Page 8: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Notation

Notation 3Button names are indicated by [Button1]>, [Button2]>, [Button3] and so on when they are touched in order.

Example: On the exposure screen, touch [SYSTEM] > [SETUP MENU] in this order.Operation: Touch [SYSTEM] on the first screen, and then touch [SETUP MENU] on the next screen.

Touch [SYSTEM] (left), and then touch [SETUP MENU] (right).

Notation 4When a more detailed explanation for an operation or setting is presented on another page, the chapter is indicated by (→ page number).Note that when Setup Guide is written, please refer to the separate Setup Guide.

Example: Touch [PRINT SETTING] (→ Setup Guide).Description: Printer settings are provided in detail in the Setup Guide.

Notation 5When an operation or description in a chapter continues from an odd-numbered page to an even-numbered page, “Continued on the following page >>” appears at the bottom right of the odd-numbered page.

Notation 6This operation manual assumes the use of a touch panel.If using a mouse, replace the expression “touch” with “click”.

Page 9: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Starting Up/Shutting Down

1. Starting Up/Shutting Down the CXDI

Page 10: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 2 –

Notes on Starting Up/Shutting Down the CXDI

When Turning OFF the PowerNever turn OFF the power during the shutdown process.This may result in hardware failure or loss of data.

Page 11: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

1. Starting Up/Shutting Down the CXDI

1.1 Starting Up the CXDI1.1.1 Turn ON the Power

Turn ON the power while referring to the hardware manual.

1.1.2 Welcome Message

1 The welcome message appears.

Note, this message will only appear if the check box ( ) next to Show next time was selected the last time the power was turned OFF, or if HELLO DIALOG (→ Setup Guide) is set to [DISPLAY].

Welcome Message Window

2 Close the welcome message.

Touch [CLOSE] to close the welcome message.

When the [LOGIN] screen appearsDepending on the installation setup by a service engineer, the LOGIN screen (→ page 131) may be displayed when the CXDI is started up.Input the user name and password, then touch [OK].

To disable displaying the Welcome message at start-upIf you do not want the welcome message to appear the next time you use the equipment, touch the check box ( ) to clear it.

– 3 –

Page 12: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

1. Starting Up/Shutting Down the CXDI

1.1.3 Exposure Screen DisplayThere are two types of exposure screen: tray type and category type. The exposure screen can be switched depending on the setting (→ Setup Guide).The tray type screen displays the exposure modes that you use most frequently, and the category type screen displays all available exposure modes for each area to be imaged.(An example screen shot is shown below. Note that some items may not appear depending on the user settings and other conditions.)

Tray Type Screen

Category Type Screen

*1 A unique ID issued by the CXDI system (“S:xxx...” and/or “I:yyy...”) is displayed in the message area.

*2 Depending on the installation setup, it may not be displayed.*3 Bucky msec (bucky time) is displayed only with sensor units CXDI-11, CXDI-12 and

CXDI-22.*4 The progress bars are displayed when using a sensor unit which has a direct

Ethernet connection to a PC.

PATIENT buttonOutput indicator

Preview image buttons

Present date and time

MULTI VIEW buttonSending images after exposure

DESTI. PREF. button

CHANGE button

Patient data areaExposure condition parameters (*3) Message area (*1)

Study descriptionbutton

SYSTEM button

Image display area(Density adjustment

area)

Sensor unit button

Exposure modeson a “tray”

[ERR] (Error button)

Progress bar (*4)

PATIENT buttonOutput indicator

Preview image buttons

Present date and time

MULTI VIEW button

Sending images after exposureDESTI. PREF. button

CHANGE button

Patient data areaMessage area (*1)

Study descriptionbutton

Exposure conditions parameters (*3)

SYSTEM button

Sensor unit button

[ERR] (Error button)

Exposure modes on a “tray”Logout (*2)

Category tabs

Progress bar (*4)

Tab scroll button

– 4 –

Page 13: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

1. Starting Up/Shutting Down the CXDI

1.2 Shutting Down the CXDI

NOTE: Never turn off the power during the shutdown process.This may result in hardware failure or loss of data.

NOTE: Be sure to turn off the power once a day, and conduct an inspection based on the procedure in the hardware manual.

1.2.1 To Shut Down the CXDI

1 Display the system screen.

Touch [SYSTEM] on the exposure screen.

2 Select the shutdown method.

Select [SHUTDOWN] or [SHUTDOWN after trans] from the shutdown menu. A window will open and the confirmation message “Are you sure you want to shutdown the system?” will be displayed. Touch [OK].

3 Turn off the power.

A window opens with the message “It is now safe to turn off the power.” Refer to the hardware manual, and turn OFF the power.Depending on the type of hardware used, the power may turn OFF automatically after the message appears.

Difference between [SHUTDOWN] and [SHUTDOWN after trans]Touching [SHUTDOWN] will shut down the system immediately. If images were being output at that time, output will resume after the power is next turned ON. Touching [SHUTDOWN after trans] will shut down the system after output of all the images is complete. However, if an error occurs during output, the system will shut down at that point.

– 5 –

Page 14: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

1. Starting Up/Shutting Down the CXDI

1.2.2 To Restart the CXDIRestart the system if restarting is required after changing settings, etc.

1 Display the System screen.

Touch [SYSTEM] on the exposure screen.

2 Select CXDI RESTART.

Touch [CXDI RESTART]. A window will open and the confirmation message “Do you want to restart the CXDI?” will be displayed. Touch [OK].

– 6 –

Page 15: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Exposure and Output

2. Taking an Image

3. Output Control (Process Viewer)

Page 16: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 8 –

Notes on Exposure and Output

When Using a Magnetic Card Reader (optional unit)1. Advance setting by a service engineer is required to use the magnetic card reader.2. Be sure to insert the magnetic card facing the correct direction, otherwise the magnetic card reader may be

unable to read the data on the card.Also, do not allow cards to become dirty, damaged or deformed, as this may result in reading errors.

3. If patient data is entered without selecting [END STUDY] when an image is present, that image will be automatically output.

When Acquiring Patient Data by HIS/RIS or from a Database File, etc.Advance setting by a service engineer is required.

During Exposure1. Before an exposure, check if the patient name and ID match with ones on the exposure screen.2. The LANMIT sensor in the sensor unit will be gradually depleted if the camera is left in the READY condition.

Therefore, touch [CANCEL] or [END STUDY] to cancel the exposure.3. Never press the exposure button on the x-ray generator while you are performing operations from the

operation unit.This may result in freezing the system, and data may be lost.

4. Never press the exposure button on the x-ray generator when “READY” is not indicated in the message area.

5. After x-ray exposure, do not perform any operation from the operation unit or enter patient data from an external device until the exposure screen, preview image screen, magnified image screen, or QA mode screen is displayed. Doing so may result in freezing the system, and data may be lost.

Note When Selecting [CANCEL]After selecting [CANCEL], do not operate the operation unit until the message “Enter Patient Data” is displayed in the message area.Initiating operation before this message appears may result in an operation error.

When Outputting an ImageIf you wish to use an external storage medium, settings must be changed by the service engineer.

Page 17: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

NOTE: The LANMIT sensor in the sensor unit will be gradually depleted if the camera is left in the READY condition. Therefore, touch [CANCEL] or [END STUDY] to cancel the exposure.

When the CXDI starts up, the exposure screen (tray type screen or category type screen) is displayed.

Exposure Screen (Category Type Screen)

CXDI exposure starts from this exposure screen. Basically, all you have to do to take an image is enter patient data and then select an exposure mode.The taken image will appear and can be checked on the operation unit.

Preview Image Screen

The following pages describe the exposure modes and methods for viewing and outputting images.

– 9 –

Page 18: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

Flow of Operation for Taking Images

2.1 Entering Patient Data

2.2 Selecting and Changing

the Exposure Mode

2.3 Taking an image

Preview Image

2.4 Operations from the Exposure

Ready Condition

2.5 END STUDY

Retake

CANCEL RETAKE

No Display of

Preview Image

CANCEL

END STUDY

Next exposure after

changing patient data

Next exposure without

changing patient data

Display

Preview Image

Exposure

– 10 –

Page 19: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.1 Entering Patient DataPatient data can be entered from the operation unit, a magnetic card reader, or a database file.In addition, 2.1 Entering Patient Data (→ page 11) and 2.2 Selecting and Changing the Exposure Mode (→ page 15) may be performed in any order.

2.1.1 Entering Data from the Operation Panel

1 Display the software keyboard.

Touch [PATIENT] on the exposure screen.

Software Keyboard (Alphabet)

2 Enter the name. (Optional depending on the setting)

Touch the frame to display the cursor and enter the name in 63 or fewer alphanumeric characters.

3 Enter the ID. (Required)

Touch the frame to display the cursor and enter the ID in 32 or fewer alphanumeric characters.

4 Enter the accession number. (Optional depending on the setting)

Touch the frame to display the cursor and enter the accession number in 16 or fewer alphanumeric characters.(Depending on the installation setup by a service engineer, it may not be possible to enter the accession number.)

5 Select patient’s sex. (Optional depending on the setting)

Touch [MALE], [FEMALE] or [OTHER].

Switching the software keyboard inputTouch [NUM] for numerical input, or [ALPHA] for alphabet input. Also, touch [Caps] to select between upper case and lower case alphabet letters.

Continued on the following page >>

– 11 –

Page 20: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

6 Enter the patient’s birth date. (Optional depending on the setting)

Touch the “Year,” “Month,” and “Day” frames, and enter the patient's birth date. When you touch the “Year” frame, you can select either A.D. or LOCAL CALENDAR for the year format.

Software Keyboard (Numeric)

Automatic patient age calculationWhen the birth date is entered, the age of the patient will be calculated and displayed automatically in the Age field if “Automatically Calculate Age” in “PATIENT INFO.” (→ Setup Guide) has been set to [CALC].If the patient is less than one year old, the age will be displayed as 0.

Directly entering the patient’s ageDepending on “PATIENT INFO.” settings (→ Setup Guide), the patient’s age can be entered directly without entering a birth date. Touch the Age field, move the cursor, and enter the patient’s age.Note, however, that if “Automatically Calculate Age” has been set to [DON’T CALC], an inconsistency may arise between the birth date entered and the patient’s age.

– 12 –

Page 21: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

7 Change the study information as necessary.

Touch [STUDY INFO.]. The study information screen appears.Touch the frame for each item to display the software keyboard, change the setting and touch [OK].Note that if the study is ended without saving the changes, the changes will be discarded and the settings will return to the original contents.

Study Information Screen

8 Finalize the patient data.

Touch [OK].However, the exposure screen does not appear in the following cases even if [OK] is selected.• When items set to [ESSENTIAL] in STUDY INFO. have not been entered, or when

[Automatically show study data input screen] has been selected, the study information screen will appear instead.

• When items set to [ESSENTIAL] in ESSENTIAL DATA have not been entered, a warning message will appear.

Study information saved during studyWhen new values other than the study description are entered at the study information screen during the study, the entered values will be automatically saved by touching [OK] and quitting the study information screen.

[STUDY INFO.] display conditions[STUDY INFO.] for choosing study information is displayed when any item in the [STUDY INFO.] screen (→ Setup Guide) is set to [ENABLE] or when the user authentication function (→ Setup Guide) is activated.

To change patient dataTouch [PATIENT] on the exposure screen to correct patient data, or [CANCEL] to clear the data that has been entered.

Continued on the following page >>

– 13 –

Page 22: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

9 Select the Study Description.

Touch the Study Description button located on the left side of the screen. A window will open, so select a study description from the list.

Study Description Window

2.1.2 Acquiring Data from a Magnetic Card Reader (optional unit)Please contact your Canon representative or distributor.

2.1.3 Acquiring Data by HIS/RIS or from a Database FilePlease contact your Canon representative or distributor.

Study Description button displayWhen study description is set to [ENABLE] in [STUDY INFO.] (→ Setup Guide), the study description will be constantly displayed in the exposure screen, or it will appear after the patient data has been entered.

– 14 –

Page 23: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.2 Selecting and Changing the Exposure Mode

2.2.1 Selecting an Exposure Mode

1 Select the sensor unit.

If multiple sensor units can be used, first select a sensor unit by touching that sensor unit button.

2 Touch the required exposure mode button on the tray.

On the category type screen, touch the category tab of the category in which the required exposure mode is found.If there are six or more categories, use or to scroll through them.

Case when exposure mode is selected automaticallyWhen AUTO-PRESS (→ Setup Guide) has been set to [ENABLE], the sensor unit and exposure mode will be selected automatically when the patient data is entered.

Continued on the following page >>

– 15 –

Page 24: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

When the parameter changing screen appears automaticallyWhen the parameter change screen in DISPLAYING PARAMETER CHANGE/QA SCREEN DURING EXPOSURE (→ Setup Guide) is set to [VALID], the parameter changing screen (figure below) will appear. Refer to 2.2.2 Changing Generator Parameters after Exposure Mode Selection (→ page 18) and change the parameters.

When print settings need to be changedThe printing method is registered in exposure mode, so when the printer parameters or other settings have changed, a window will open and the message “Printer settings have changed. Set the printer settings again.” will appear. Touch [OK] to open the print setting window (→ Setup Guide) and change the settings as required.

When a warning message concerning the grid appearsIf the grid of the sensor unit is replaceable, it is possible to register whether to display the grid warning message for each exposure mode when editing the mode (→ Setup Guide).If WARNING MESSAGE ON GRID has been set to [VALID] and the currently installed grid differs from the recommended grid, a window will open (figure below) and the recommended grid name (or YES/NO) will appear. Touch [OK] to replace the grid with the recommended grid, or attach/detach the grid as required.

– 16 –

Page 25: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

3 Check to make sure the system is ready for exposure.

When the system is ready for exposure, the message area turns green and READY will appear in the message area.On the tray type screen, the image display area also turns green.

Exposure Screen (Tray Type)

Changing the Exposure ModeIf you have selected the wrong exposure mode, touch the correct button again to select the correct exposure mode.

– 17 –

Page 26: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.2.2 Changing Generator Parameters after Exposure Mode Selection

1 Display the parameter changing screen.

Touch [CHANGE] on the exposure screen.

Parameter Changing Screen

2 Select the tab.

Touch the tab containing the parameters to be changed.

3 Set the parameters.

Refer to Setup Guide and change the parameters.

4 Save the parameters as necessary.

To save the current conditions as default parameters for the chosen exposure mode, touch [SAVE PARAM].

5 Return to the exposure screen.

Touch [OK].

If the values of the parameters do not change successivelyIf [ENABLE] has been selected in PARAMETER TABLE (→ Setup Guide), only the values set there will be available as options.

Entering a passwordIf a window opens and password entry (→ Setup Guide) is requested, touch the password frame to display the Password screen and enter the password.

– 18 –

Page 27: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

NOTE: If the sensor unit has been attached or removed on the exposure screen after changing parameters and the parameter changing screen, the values of generator parameters that have been changed will be lost and the READY condition of the exposure screen will be canceled.Note, however, that if [SAVE PARAM] has been applied for changed values, values changed will be loaded when selecting the exposure mode.

Changing the exposure parameters on the tray type screenOn the tray type screen, the exposure parameters can also be changed by using and

.They can also be changed by touching the parameter frame and entering the value from the ten key or parameter table.

Parameter Table

– 19 –

Page 28: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.3 Taking an Image2.3.1 Starting Exposure

NOTE: Never press the exposure button on the x-ray generator while you are performing operations from the operation unit.This may result in freezing the system, and data may be lost.

NOTE: Never press the exposure button on the x-ray generator when “READY” is not indicated in the message area.

NOTE: After x-ray exposure, do not enter patient data from the operation unit or from an external device until the exposure screen, preview image screen, magnified image screen, or QA mode screen is displayed.Doing so may result in freezing the system, and data may be lost.

NOTE: If the grid unit is detachable, make sure the patient does not remove it from the sensor unit during exposure.

1 Position the patient.

Position the patient such that the area to be imaged is within the exposure area of the sensor unit.As necessary, take a means to keep the patient still.

2 Press the exposure button.

Press the exposure button of the x-ray generator when the patient is correctly positioned and “READY” is displayed on the touchscreen.A progress bar will be displayed in the message area, and the image will be transferred from the sensor unit.

Operation buffer functionIf using sensor unit which has a direct Ethernet connection to a PC, when the sensor unit button, [END STUDY] or [SEND] is touched while the progress bar is displayed in the message area, that operation will be executed after the image has been completely transferred. In this case, the progress bar will be displayed in the center of the screen.

However, when the exposure mode, [PATIENT] or [CHANGE] is touched while the progress bar is displayed in the message area, these operations can be performed as usual.

– 20 –

Page 29: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

3 Check the image.

The screen will return to the exposure screen and the system will enter the exposure ready condition (→ page 25).However, note that the magnified image screen, QA screen or preview image screen may be displayed depending on the settings.

Magnified image screen (→ page 92)

When AUTO DISPLAY in MAGNIFY SETTING window (→ Setup Guide) is set to [DISPLAY], the magnified image screen will appear.

Magnified Image Screen

QA Mode Screen (→ page 65)

When QA Screen in DISPLAYING PARAMETER CHANGE/QA SCREEN DURING EXPOSURE (→ Setup Guide) is set to [VALID], the QA mode screen will appear.

QA Mode Screen

Continued on the following page >>

– 21 –

Page 30: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

Preview Image Screen (→ page 35)

When Preview image (→ Setup Guide) is set to [DISPLAY], the preview image screen will appear.

Preview Image Screen

2.3.2 Cancelling the ExposureTouch one of the following buttons.

[CANCEL]

The system will return to 2.1 Entering Patient Data (→ page 11).

[END STUDY]

Output of images already taken will start, and the system will return to 2.1 Entering Patient Data (→ page 11).Depending on the installation setup by a service engineer, the IMAGE DESTINATION window may open automatically when you touch [END STUDY]. (→ page 27)

[CANCEL RETAKE] (When retaking an image)

A window will open and the message “Retaking will be canceled. To retake, select [RETAKE] on Preview-screen. To take another image, select an exposure mode.” will be displayed. Touch [OK]. Proceed to 2.4 Operations from the Exposure Ready Condition (→ page 25).

REX checkIt is recommended that the REX (→ page 73) always be checked, as it may be very different from the normal value if you have performed an exposure without changing the settings of the x-ray generator after performing calibration, or when the settings of x-ray generator is wrong.Also, if EXPOSURE REX CHECK (→ Setup Guide) is set to enable, a message will appear when the REX exceeds the range that has been set.

– 22 –

Page 31: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.3.3 Displays after Exposure

If the message QA Process Error or REX Warning appears

An error window may open after exposure due to an improper x-ray dose, etc.Follow the procedure below to perform image processing or to retake the image.

QA Process Error

1 Close the window.

Touch [OK].

2 Display the image.

Refer to 4. Browsing Images (Preview Image Screen) (→ page 35) and display the image.

3 Perform one of the following operations.

Processing the Image

Refer to 7. Processing the Image (QA Mode Screen) (→ page 65) and adjust the parameters.

Retaking the Image

Refer to 4.5 Retaking an Image (→ page 41) and retake the image.

Continued on the following page >>

– 23 –

Page 32: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

If print setting window appears

Change the print settings (→ Setup Guide). Touching changes the display.

– 24 –

Page 33: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.4 Operations from the Exposure Ready Condition

2.4.1 Operations from the Exposure Ready Condition You can perform the following operations.

To continue exposure with the same patient data

Touch and select a sensor unit and an exposure mode.

To output images already taken and continue the exposure

Touch [SEND].However, note that [SEND] is displayed only when SEND AFTER EXPOSURE (→ Setup Guide) is set to [SEND].Depending on the installation setup by a service engineer, the IMAGE DESTINATION window may open automatically when you touch [SEND]. (→ page 27)The image output indicator blinks green during image output.

To output all the images which have not been output yet and end the exposure of the current patient

Touch [END STUDY].Depending on the installation setup by a service engineer, the IMAGE DESTINATION window may open automatically when you touch [END STUDY]. (→ page 27)The image output indicator blinks green during image output.

Retaking an Image

Touch the preview image button and touch [RETAKE] on the preview image screen. (→ page 41)

Continued on the following page >>

– 25 –

Page 34: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

To refer to other images for the current patient

Touch the preview image button to display the preview image screen, or touch [MULTI VIEW] to display the multi-view screen (→ page 43).

Preview Image Screen (left) and Multi-View Screen (4 shots/screen) (right)

To reject a bad image

Touch the preview image button and touch [Reject] on the preview image screen. (→ page 40)

To change the print settings

Touch the preview image button and touch [PRINT SETTING] (→ Setup Guide) on the preview image screen.

To process the image

Touch the preview image button and touch [QA] on the preview image screen. (→ page 65)

– 26 –

Page 35: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.4.2 Selecting Image Output Destination

1 Open the IMAGE DESTINATION window.

Touch [DESTI. PREF.] on the exposure screen.[ON] or [OFF] is displayed for the devices set as the output destinations (→ Setup Guide).

IMAGE DESTINATION Window

2 Select the image output destination.

Touch [ON]/[OFF] next to the name of the destination and select [ON] for the device to send the image.Note that the LOCAL TEMP STORAGE setting is fixed to [ON] and cannot be changed.

3 Set the priority.

Each time [Priority] is touched, the button switches between highlighted display (white characters) and normal display. It is enabled when highlighted.

4 Set the output destination.

Touch [OK]. Close the IMAGE DESTINATION window.

When [DESTI. PREF.] is not displayedDepending on the installation setup by a service engineer, [DESTI. PREF.] may not be displayed.In that case, touch [SEND] or [END STUDY] to open the IMAGE DESTINATION window.

Output priorityWhen there are many studies waiting to be output, the images for the current patient are given priority and output first. However, note that when a study is currently being output, the images for the current patient will be output after all the images for that study have been output.In addition, when multiple images that have been given priority are waiting to be output, these priority images will be output in the order they were taken.

Continued on the following page >>

– 27 –

Page 36: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

Trimming imagesWhen a printer is enabled in the printer setup (→ Setup Guide) but [OFF] is selected for that printer at the IMAGE DESTINATION window, the green film size frame will be displayed in the preview image screen and the image will be trimmed to that film size during output.If you wish to output the entire irradiation field, settings must be changed by the service engineer.

– 28 –

Page 37: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

2.5 END STUDYTouch [END STUDY]. Image output will start and the system will return to waiting for patient data entry status.The image output indicator (upper right) blinks green during image output.

To perform an exposure of the next patient

Enter new patient data. (→ page 11)

To shut down the CXDI

Refer to 1.2 Shutting Down the CXDI (→ page 5), and shut down the CXDI.

Automatic display of the IMAGE DESTINATION windowDepending on the installation setup by a service engineer, the IMAGE DESTINATION window may open automatically when you touch [END STUDY].

– 29 –

Page 38: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

2. Taking an Image

– 30 –

Page 39: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

3. Output Control (Process Viewer)In the process viewer, the output status is displayed in study units, and the output destination and the order of priority can be changed.

1 Display the process viewer.

From the exposure screen, touch [SYSTEM] > [PROCESS VIEWER].

Process Viewer Screen

2 Select a study.

Touch a study. The image output process will stop and [UNSELECT] or [MODIFY] will be displayed. ([MODIFY] will be displayed only when a printer or storage has been selected as the output destination.)To restart the image output process, touch [UNSELECT].

Item Description

State

IDLE Image is waiting to be output.QA Image is being processed.SEND Image is being transferred.* Priority

Time Date and time the first image of the patient was taken.Patient ID Patient IDPatient’s name Patient’s name

Printer

p1 Printer 1 or 3p2 Printer 2 or 4s1 Storage 1 or 3s2 Storage 2 or 4

Continued on the following page >>

– 31 –

Page 40: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

3. Output Control (Process Viewer)

3 Open the CHANGE DESTINATION window.

Touch [MODIFY].[MODIFY] will not be displayed if no output destination has been set.

CHANGE DESTINATION Window

4 Set the output destination.

[ON] will be displayed for devices that have been set as output destinations. Touch [ON] for devices to which the image will not be output to set these to [OFF].However, note that if you select [OFF] and close the window, that device will not be displayed and you will not be able to return it to [ON] the next time the window is opened.

5 Set the priority.

Each time you touch [Priority], the button switches between highlighted display (white characters) and normal display. The selected study is given output priority when this button is highlighted.

6 Close the CHANGE DESTINATION window.

Touch [OK]. A window will open and the confirmation message “Do you want to change the settings?” will be displayed. Touch [OK].

7 Return to the menu screen.

Touch [EXIT].

– 32 –

Page 41: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Viewing the Images

4. Browsing Images (Preview Image Screen)

5. Viewing Multiple Images (Multi-View Screen)

6. Browsing Completed Studies (Study List)

Page 42: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 34 –

Notes on Viewing Images

Notes on the Study List1. When referencing local temporary storage, the list of studies stored in the specified device will be displayed.

Due to the storage capacity of the device, images of studies that are not displayed in the list are automatically deleted and cannot be reproduced.

2. Taking an image immediately after executing [RESEND] or [RESEND ALL] places a heavy load on the hardware, and may interfere with the exposure. Therefore, [RESEND] and [RESEND ALL] should be executed only after all the exposures have been finished if possible.

3. Once images are deleted, they cannot be restored.

Page 43: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

4.1 Displaying the Preview Image ScreenFollow the procedure below to display the preview image screen.

Preview Image Screen

To display the QA mode screen automatically after the exposure

From the exposure screen, touch [SYSTEM] > [SETUP MENU] > [CUSTOMIZE DISPLAY]. On the CUSTOMIZE DISPLAY screen (→ Setup Guide), select [DISPLAY] for PREVIEW IMAGE.

Selecting an image from the exposure screen

Touch a preview image button.However, note that [Reject], [PRINT SETTING], [QA] and [RETAKE] cannot be selected for images that were output by touching [SEND].

Number of the currently displayed image out of the total number of images

Preview image buttons

– 35 –

Page 44: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

4.2 Available Operations in the Preview Image Screen

Changing image output destination

Touch [DESTI. PREF.]. (→ page 38) If [DESTI. PREF.] does not appear, touch [OK] and return to the exposure screen, then touch [SEND] or [END STUDY] to open the IMAGE DESTINATION window.

To reject a bad image

Touch [Reject]. (→ page 40)

To change the print settings

Touch [PRINT SETTING] (→ Setup Guide).However, [PRINT SETTING] will not be displayed unless a printer has been set as the output destination.

To process the image

Touch [QA]. (→ page 65)

To retaking an image

Touch [RETAKE]. (→ page 41)

To return to the exposure screen

Touch [OK]. The system enters the ready condition.

To view several images of the current patient

Touch [MULTI VIEW]. (→ page 43)

A green frame indicating the film size or irradiation field will be displayed.

REX (→ page 73)

Film size Orientation of film P: Portrait L: Landscape

Multi-format layout (→ page 48)

Fixed ratio (%)Displayed when fixed ratio printing has been selected (→ Setup Guide)

The following contents will be displayed when the printing mode has been set.

Series description

– 36 –

Page 45: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

Selecting an image

Touch the preview image button to display the preview image screen.The preview image buttons have the following functions.

To change a series description

Touch the frame to display the software keyboard and enter the series description in 63 or fewer alphanumeric characters.A series description can be entered for each image.If an image is retaken after changing the series description, the description registered in the exposure mode will be used.

Button Description

or The previous or the next image will be displayed.

or The first or the last image will be displayed.

AEC (Automatic Exposure Control) areaThe AEC (→ Setup Guide) setting range can be displayed on the preview image screen if the setup has been performed by the service engineer.When the AEC display is enabled, [Disp. AEC] will appear on the preview image screen. The setting range indication (yellow frames on the image) can be made to appear and disappear by touching this button.If the optional generator communication module has been used, the effective area will be displayed as yellow squares according to the data sent from the x-ray generator.

– 37 –

Page 46: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

4.3 Selecting Image Output DestinationDestination of images can provisionally be changed in a unit of study.

1 Display the preview image screen.

Preview Image Screen

2 Open the IMAGE DESTINATION window.

Touch [DESTI. PREF.].

IMAGE DESTINATION Window

3 Set the priority.

Each time you touch [Priority] the button switches between highlighted display (white characters) and normal display. The selected study is given output priority when this button is highlighted.

When [DESTI. PREF.] is not displayedDepending on the installation setup by a service engineer, [DESTI. PREF.] may not be displayed. In these cases, touch [OK] to close the preview image screen, then touch [END STUDY] at the exposure screen. The IMAGE DESTINATION window opens and you can select the image output destination.

– 38 –

Page 47: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

4 Set the output destination.

Touch [ON]/[OFF] and select [ON] for the device to which the image is to be output.

5 Close the IMAGE DESTINATION window.

Touch [OK].

– 39 –

Page 48: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

4.4 Rejecting a Bad Image

NOTE: The rejection reason input function may be available for certain service engineer settings.For details on operation of this function, see B. Using the Rejection Reason Input Function (→ page 135).

When an image has been rejected, the output destination is limited to local temporary storage.

1 Display the preview image screen.

Preview Image Screen

2 Reject the image.

Each time you touch [Reject], the button switches between highlighted display (white characters) and normal display. The image is rejected when this button is highlighted.An image that has been rejected cannot be retaken and print settings cannot be made. In addition, a red mark will be displayed on the exposure mode name indicating that it has been rejected.

Rejected Image

– 40 –

Page 49: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

4.5 Retaking an Image

NOTE: The rejection reason input function may be available for certain service engineer settings.For details on operation of this function, see B. Using the Rejection Reason Input Function (→ page 135).

If you are dissatisfied with the current image, it can be retaken using the same exposure mode.

1 Display the preview image screen.

Preview Image Screen

2 Select [RETAKE].

Touch [RETAKE]. A window will open and the confirmation message “The original image will be rejected. Do you want to retake the image?” will be displayed. Touch [OK]. The exposure screen appears and the system enters the ready condition with the same exposure mode selected as that used for the current image.You cannot select a different exposure mode for retaking the image.

3 Retake the image.

The image displayed in step 1 will be rejected.

Notes on Retaking ImagesDepending on the installation setup by a service engineer, the window may not be displayed.

Continued on the following page >>

– 41 –

Page 50: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

4. Browsing Images (Preview Image Screen)

Retaking and rejecting imagesWhen an image is retaken multiple times in succession, all images that are retaken will be rejected, so only the last image can be output (valid image). The example below shows the flow of operation when an image is retaken twice.

Flow of operation when an image is retaken twice

An invalid image can be changed to a valid image from the preview image screen. However, because only one image can be chosen as valid, the image that was previously valid will automatically be invalidated when another image is validated. In the example below, the lower image shows a case in which the second image taken in the above figure is changed to the valid image.

Changing the valid image

When you touch [Reject] for a rejected image on the preview image screen and make that image the valid image, a window will open and the confirmation message “The No. n image [Exposure Mode] will be rejected. (“n” is the exposure number of the valid image.)” will be displayed. Touch [OK].Also, you can invalidate all the valid images from the preview image screen.

When image processing parameters have been changed before retaking theimageWhen parameters are changed on the QA screen after exposure and then [RETAKE] is touched to retake an image, the parameters used are as follows.• Changes made to the print setting, mark (except “ROTATE” and “FLIP”), and

“Marker” parameters are valid.• For other parameters, those that applied when the first image was taken will be

used.

Valid

image

Valid

image

Valid

image

First

image

taken

First

image

taken

First

image

taken

Second

image

taken

Second

image

taken

Third

image

taken

First retake Second retake

Invalid

image

Invalid

image

Invalid

image

Retake Retake

Valid

image

First

image

taken

Second

image

taken

Third

image

taken

Invalid

image

Invalid

image

First

image

taken

Second

image

taken

Third

image

taken

Valid

image

Invalid

image

Invalid

image

Before the change After validating the second image

The case when the

second image

taken in the above

figure is changed

to the valid image.

– 42 –

Page 51: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

5.1 Displaying the Multi-View ScreenFollow the procedure below to display the multi-view screen.

1 Select multi-format (number of images) on the multi-view screen before viewing the images.

From the exposure screen, touch [SYSTEM] > [SETUP MENU] > [CUSTOMIZE DISPLAY] to display the CUSTOMIZE DISPLAY screen (→ Setup Guide), then set the number of images to be displayed on the preview image screen to [4 SHOTS/1] or [9 SHOTS/1].

2 Display the Multi-View screen.

Touch [MULTI VIEW] on one of the exposure screen, preview image screen, or CALL IMAGES screen (select [CALL IMAGES] in the study list).

Multi-View Screen (4 shots/screen)

– 43 –

Page 52: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

5.2 Available Operations on the Multiview Screen

Multi-View Screen (4 shots/screen)

To change the display of print preview images

Touch or located below the print preview.The print preview image will be displayed when a printer for which output is enabled is set.

To change the image

Touch or located on the upper right side of the multi-view screen.

To exit the multi-view screen

Touch [EXIT]. The system returns to the screen displayed before entering the multi-view screen.

Display DescriptionPrinter name This displays the output printer name.Print information Film size, film orientation, multi-format layout are displayed.

Print preview image

This displays the image that was cut to the film size shown in the print information.Rejected images are not displayed.The display order is the printer setting order (Printer 1 and Printer 3 first, Printer 2 and Printer 4 last), and this becomes the output number order.

Number of films This displays the film number currently displayed/total films to be output.

Output information

This displays the printer number/output number.For example, 1-2 means the second output to Printer 1.However, 0 is displayed when no printer is selected as a destination.Rejected images are indicated by an × on the output number.

Print previewimage

Output information

Print information

Printer name

Number of films

– 44 –

Page 53: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

To change the output number of an image

Directly touch the image whose output number is to be changed to open the window. (→ page 46)

To reject a bad image

Directly touch the image to be rejected to open the window. (→ page 47)

To display a preview image

Directly touch the image to open the window. (→ page 47)

Images output using SEND AFTER EXPOSUREImages that have been output using SEND AFTER EXPOSURE (→ Setup Guide) will appear in sepia tone, and the window will not open even if these images are touched directly.

– 45 –

Page 54: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

5.3 Window OperationsWhen an image on the right side of the screen is touched directly, the window will open and various operations can be performed.

Open the window.

5.3.1 Changing the Output OrderChange the output number.The images will be output to output destinations other than a printer in the order they were taken.

1 Open the window.

Directly touch an image on the right side of the screen to open the window.

2 Change the output number.

Change the number using or .

3 Determine the output number.

Touch [OK].The image output number will change, as will the output number of the remaining images.

Output number

– 46 –

Page 55: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

5.3.2 Rejecting a Bad ImageWhen an image has been rejected, the output destination is limited to local temporary storage.

1 Open the window.

Directly touch an image on the right side of the screen to open the window.

2 Reject the image.

Touch [Reject]. An “X” will appear in place of the output number of the image, and the image will be set to rejected.When an image is rejected, the output order of that image cannot be changed, but the output order of the other images will change.However, when an image that was rejected as a result of retaking is made the valid image, a window will open and the confirmation message “The No.n image [Exposure Mode] will be rejected. (“n” is the exposure number of the valid image.)” will be displayed. Touch [OK].

Reject

5.3.3 Checking Images on the Preview Image Screen

1 Open the window.

Directly touch an image on the right side of the screen to open the window.

2 Display the preview image screen.

Touch [Preview]. This changes to the preview image screen or CALL IMAGES screen (during image playback from the study list).

– 47 –

Page 56: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

5.4 Multi-Format OutputWhen the same multi-format is selected for successive images, multiple images can be printed on a single film.The multi-format is registered as a print setting (→ Setup Guide) in exposure mode. The method for changing this setting is described below.

5.4.1 Choosing Multi-Format before the Exposure

1 Display parameter changing screen.

Select exposure mode, then touch [CHANGE] on the exposure screen.

2 Open the print setting window.

Touch [PRINT SETTING] on the parameter changing screen.

3 Set the multi-format.

Touch the multi-format button on the print setting window to open the multi-format window, and set the format.

Multi-Format Window

4 Take multiple images using the set multi-format.

Take multiple images using the set multi-format.To take images using a different exposure mode, set the multi-format again in the same manner.

– 48 –

Page 57: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

5 Display the print preview.

Touch [MULTI VIEW].The figure below shows an example in which a 14×17 portrait image has been made into a 4-image exposure using the 2x1 multi-format. On the print preview screen of this example, the images for output numbers 1 and 2 are cut off and assigned to a single film.In addition, a preview of the output number 3 and 4 images can be displayed by touching

below the print preview.

Multi-Format

5.4.2 Choosing Multi-Format after the ExposureThe method for changing the print setting after the exposure is described below.

1 Display the preview image screen or the CALL IMAGES screen.

Display the preview image screen (→ page 35) or the CALL IMAGES screen (→ page 59).

2 Open the print setting window.

Touch [PRINT SETTING].

3-image layoutIf the 2x2 multi-format is selected for a three-image layout, the third image is located in the center of the second row.

Continued on the following page >>

– 49 –

Page 58: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

3 Set the multi-format.

Touch the multi-format button on the print setting window to open the multi-format window, and set the format.

4 Close the print setting window.

Touch [OK].

5 Display the print preview.

Touch [MULTI VIEW].

When images cannot be printed in multi-formatImages cannot be printed in multi-format in the following cases:

1. When a different multi-format has been setIf images that were set for different multi-format are included in the continuous images in a single study, the images from different multi-format is assigned to a different film.In the example in the figure below, when all four images can be printed in multi-format, the number of divisions for the images is set at 2x2. In this case, four images are assigned to one film.In comparison, when only the first two images can be printed in multi-format, the number of divisions is set to 2x2 for the first and second image, but the number of divisions for the third image is set to 1x1. As a result, only the first and second images are assigned to one film.

2. When a different reduction ratio has been selectedIf a different reduction ratio is included in fixed ratio printing, the images with different reduction ratio are assigned to different film even if the same multi-format is set.

3. When a different film size or film orientation has been selectedChange the print settings.

4. When a different printer has been selected as the output destinationSelect the same printer for all the images.

5. When an image taken by the CXDI-31 is includedImages taken by the CXDI-31 are assigned to different films.

All four images can be printed in multi-format

Only the first two images can be printed in multi-format

– 50 –

Page 59: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

Multi-format when images are resent from the study listWhen only one image will be printed, touching [RESEND] on the CALL IMAGES screen automatically assigns images to one film and prints if the same multi-format is set for consecutive images.In the example in the figure below, if image 1 is reproduced and printed when the study has five images, only image 1 is printed. However, if any of images 2 to 5 are reproduced and printed, images 2 to 5 are all assigned to one film at 2x2 and output.

Furthermore, if LOCAL TEMP STORAGE, STORAGE or EXTERNAL DISK STORAGE has also been selected as the destination, images 2 through 5 will be output to it.

Image1

Multi-format

1x1

Image2

Multi-format

2x2

Image3

Multi-format

2x2

Image4

Multi-format

2x2

Image5

Multi-format

2x2

RESEND RESEND

Image1

Multi-format

1x1

Image

2

Image

3

Image

4

Image

5

When only image1 is resent,

only that image will be output

to the printer.

When any one of images 2 through 5 is resent,

images 2 through 5 will be output to a printer.

– 51 –

Page 60: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

5. Viewing Multiple Images (Multi-View Screen)

– 52 –

Page 61: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

6.1 Displaying the Study ListImages stored in the CXDI can be displayed or output in the study list.Before displaying the study list, see Setup Guide and set [SINGLE] or [MULTI] for SELECT STUDY LIST.

1 Display the study list.

From the exposure screen, touch [SYSTEM] > [STUDY LIST].

Study List Screen (When MULTI has been selected)

2 Select the location where the image is stored

If the images are saved in multiple locations, click [Storage destination folder] (the currently selected folder name is displayed) to select it from the list.Touch the name of the folder to be referenced to display the studies stored in that folder.

3 Select the study.

If there are more than 12 studies, use and to scroll through the list.In addition, touch the display items to rearrange the display order.

External disk storageAdvance setting by a service engineer is required to use the external disk storage.

Storagedestination folder

Display items

– 53 –

Page 62: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

6.2 Customizing the Study List

1 Open the COLUMN SETTING window.

Touch [COLUMN SETTING].

COLUMN SETTING Window

2 Set the display.

Add the items displayed in the study list to the Used Column.

3 Close the COLUMN SETTING window.

Touch [OK]. This updates the list.

Setup ProcedureTo add an item to the study list

Select an item in the Unused Column and touch [Add] to move that item to the Used Column.

To delete an item from the study list

Select an item in the Used Column and touch [Delete] to move that item to the Unused Column.

To change the position of an item in the list

Select an item in the Used Column and touch [Up] or [Down] to change the display order.Items will be displayed from the left side of the study list in order from the smallest number.

– 54 –

Page 63: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

Explanation of displayFor the exposure mode, the body part (DICOM) set as the exposure mode for the first image contained in a study will be displayed. See the table below for Previous Trans Result, Cnt of Success Trans and Trans Result (device name).

A Trans Result: Successful

×: A transfer request was sent, but the transfer failed.–: A transfer request was not sent.

B Cnt of Success TransHowever, 9 or more times will be indicated by “*”.

C Transfer count displayHowever, 99 or more times will be indicated by “*”.

Item

Previous Trans Result

Cnt of Success Trans

Trans Result (device name)

A

B

A, C

Description

If a list in an external storage medium is

displayed on [Storage destination folder], all

results will be displayed as “-” mark.

– 55 –

Page 64: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

6.3 Outputting Study ImagesOne or multiple studies can be selected.

1 When selecting multiple studies, you can rearrange the order.

When multiple studies have been selected, the study results are output in order from the top of the list. Therefore, touch the display item and rearrange the studies in the desired order before selecting the studies.

2 Select a study image to send.

Select a study/studies in the list. [RESEND ALL] and [DELETE] will appear.When multiple studies can be selected, [UNSELECT] and [RANGE] will also appear.To select a series of successive studies, touch [RANGE] to highlight the button (white characters).To cancel all of the selections, touch [UNSELECT].

3 RESEND ALL window appears

Touch [RESEND ALL].

RESEND ALL Window

4 Select a device for sending the image.

Touch [ON]/[OFF] next to the destination name and select [ON] for the device for sending the image.[ON]/[OFF] will not be displayed for devices with invalid settings (→ Setup Guide).

5 Set the priority.

Each time you touch [Priority] the button switches between highlighted display (white characters) and normal display. The selected study is given output priority when this button is highlighted.

6 Output the images.

Touch [RESEND ALL].

– 56 –

Page 65: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

If a message appearsIf the output location is different from that at exposure when the study is resent, a window opens with a message.In this window, touch [Cancel] to close the window, and then go to output location settings (→ Setup Guide) to check the settings.

Message Examples

– 57 –

Page 66: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

6.4 Deleting the Image(s)

NOTE: Once images are deleted, they cannot be restored.

6.4.1 Deleting All the Images[DELETE ALL] will be displayed only when no study has been selected.

1 Select [DELETE ALL].

Touch [DELETE ALL]. A window will open and the confirmation message “Are you sure you want to Delete ALL?” will be displayed. Touch [OK].

6.4.2 Selecting and Deleting Study Images

1 Select the study.

Touch a study to be deleted. [RESEND ALL] and [DELETE] will appear.

2 Select [DELETE].

Touch [DELETE]. A window will open and the confirmation message “Do you want to delete it?” will be displayed. Touch [OK].

Entering passwordIf a window opens and password entry (→ Setup Guide) is requested, touch the password frame to display the password screen and enter the password.

To select multiple studiesWhen multiple studies can be selected, [UNSELECT] and [RANGE] will also appear.To select a series of successive studies, touch [RANGE] to highlight the button (white characters).To cancel all of the selections, touch [UNSELECT].

Entering passwordIf a window opens and password entry (→ Setup Guide) is requested, touch the password frame to display the password screen and enter the password.

– 58 –

Page 67: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

6.5 Reproducing the Image(s)Touch the required study, and touch [CALL IMAGES].When multiple studies have been selected, [CALL IMAGES] will not function even if touched.

CALL IMAGES Screen

To change patient data

Touch [PATIENT] to open the patient data entry screen. (→ page 11)

To select an image

When a study contains more than one image, touch the preview image buttons to select the image.The preview image buttons have the following functions.

To display the multi-view screen

Touch [MULTI VIEW] to switch to the multi-view screen. (→ page 43)

To display the QA screen

Touch [QA] to switch to the QA screen. (→ page 65)

To change the print settings

Touch [PRINT SETTING] (→ Setup Guide) to open the print setting window.[PRINT SETTING] will not be displayed unless a printer has been set as the output destination.

To resend only the displayed image

Touch [RESEND]. (→ page 60)If multi-format has been selected for that image, note that multiple images may be output. (→ page 48)

Button Function

or The previous or the next image will be displayed.

or The first or the last image will be displayed.

Preview image buttons

Continued on the following page >>

– 59 –

Page 68: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

To resend all the images contained in a study

Touch [RESEND ALL]. (→ page 61)

To return to the study list

Touch [RETURN TO STUDY LIST].

6.5.1 Resending the ImageFollow the procedure below to resend the displayed image.

1 Open the RESEND window.

Touch [RESEND].

RESEND Window

2 Select the device for sending the image.

Touch [ON]/[OFF] next to the destination name and select [ON] for the device for sending the image.[ON]/[OFF] will not be displayed for devices with invalid settings (→ Setup Guide).

3 Set the priority.

Each time you touch [Priority], the button switches between highlighted display (white characters) and normal display. The selected study is given output priority when this button is highlighted.

4 Resend the images.

Touch [RESEND].

– 60 –

Page 69: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

6.5.2 Resending the Study ImagesFollow the procedure below to resend all the images contained in a study.

1 Open the RESEND ALL window.

Touch [RESEND ALL].

RESEND ALL Window

2 Select the device for sending the image.

Touch [ON]/[OFF] next to the name of the destination and select [ON] for the device for sending the image.However, [ON]/[OFF] will not be displayed for devices that have not been set as the output destination (→ Setup Guide).

3 Set the priority.

Each time you touch [Priority], the button switches between highlighted display (white characters) and normal display. The selected study is given output priority when this button is highlighted.

4 Resend the study.

Touch [RESEND ALL].

If all images contained in a study are rejectedIf you touch [RESEND] or [RESEND ALL] after all images contained in a study were rejected (→ page 40), a window will open and the confirmation message “All the selected images are the rejected images. They can only be sent to the internal storage. Do you want to continue?” will be displayed.If you touch [OK], the IMAGE DESTINATION window will open and you will be able to make the [ON]/[OFF] setting to select internal storage as the output destination.

– 61 –

Page 70: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

6. Browsing Completed Studies (Study List)

– 62 –

Page 71: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Processing the Image

7. Processing the Image (QA Mode Screen)

Page 72: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 64 –

Notes on Image Processing

Notes on the Edge Enhancement Process and DEPWhen the parameters of the edge enhancement process or DEP are set to excessively high values, artifacts may occur and the useful information in the image may be degraded.If you think the image looks as if it has been over-processed, set the parameters to lower values, or void the process.

Page 73: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)In the CXDI series, image processing is called QA (Quality Assurance) processing, and the image can be optimized by adjusting the parameters on the QA screen.

7.1 Displaying the QA ScreenYou can access the QA screen in one of the following ways:

QA Mode Screen

To set the QA mode screen to appear automatically after the exposure

From the exposure screen, select [SYSTEM] > [SETUP MENU] > [SYSTEM SETUP], and in the DISPLAYING PARAMETER CHANGE/QA SCREEN DURING EXPOSURE, set the QA Screen to [VALID].

To perform QA processing after the exposure

Touch [QA] on the preview image screen (→ page 35).

To perform QA processing after completing the study

Load the image from the study list (→ page 53), and then touch [QA].

– 65 –

Page 74: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.2 Operations Available on QA Screen

NOTE: After performing QA processing and touching [OK] to exit the QA screen, it may not be possible to return to parameters even after touching [RESET].

Touching or at the bottom of the QA screen switches the display options at the bottom between the commands [RESET] and [OK] and the commands [OPTION], [SAVE PARAM], Frequency Processing ([Standard Frequency Processing], [MLT(F) Frequency Processing], [MLT(M) Frequency Processing], and [MLT(S) Frequency Processing]).

LUT Tab

To set the image processing parameters

Touch the tabs to set the parameters.The parameters cannot be changed for images loaded from external storage media in the study list.

To adjust the trimming area or density adjustment area (→ page 68)

Touch [Adjust TRIM]/[Adjust ROI] to select the range.

To display a magnified image (→ page 92)

Touch [MAG] or [QA-MAG].

To adjust the gamma correction of the operation unit (→ page 93)

Touch [OPTION]

To return the parameters to their original values

Touch [RESET].

– 66 –

Page 75: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

To save the changed parameters

Touch [SAVE PARAM].The edited parameters can be saved as the default for the selected exposure mode.[SAVE PARAM] is not displayed after the called image is processed from the study list.

To use the image processing software

To use the image processing software, touch [STD. FREQUENCY] to open the window, and then select [MLT(F) FREQUENCY], [MLT(M) FREQUENCY], or [MLT(S) FREQUENCY].

NOTE: Advance setting by a service engineer is required to use [MLT(F) FREQUENCY], or [MLT(M) FREQUENCY].

For the operating procedures of the image processing software, refer to the operation manual supplied with the software.

– 67 –

Page 76: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.3 Adjust TRIM and Adjust ROI

7.3.1 Trimming the ImageImages can be trimmed manually.Selecting the POSITION tab enables fine adjustment of the position using the buttons. (→ page 81)

To trim the image

1 Set the film size.

To output to a printer, touch [PRINT SETTING] from the preview image screen or CALL IMAGES screen, and set the film size before adjusting the trimming.If the film size is changed after adjusting the trimming, the trimming area becomes the film size.

2 Select [ADJUST TRIM].

Touch [ADJUST TRIM]/[ADJUST ROI] and select [ADJUST TRIM].

3 Set the trimming area.

Touch two points on the image.The new trimming area is indicated by a red line.

Limitations on trimming areasThe following limitations apply to trimming areas.• After a printer is set, no settings can be made that exceed the printable size.• If an area of less than 8 pixels (approx. 1 cm) is set for the vertical and horizontal

direction of the displayed image, a window will open and the message “The cropped area is too small. Select a larger area!” will appear. Touch [OK] to close the window and change the trimming area.

– 68 –

Page 77: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

4 Exit the screen.

Touch [OK].

7.3.2 Changing the Density Adjustment AreaThe density adjustment area (→ Setup Guide) can be changed manually in gray scale adjustment mode (→ page 70) if [AUTO ADJUST.].

Density Adjustment Area

1 Select [ADJUST ROI].

Touch [ADJUST TRIM]/[ADJUST ROI] and select [ADJUST ROI].

2 Set the density adjustment area.

Touch two points on the image. The area you have selected will be indicated by a reddish purple frame.The reference point is changed, and the density of the entire image changes.

3 Exit the screen.

Touch [OK].

AnnotationsWhen you change the image trimming range, annotations (fixed or free) may be out of the image.

Changing the gray scaleWhen [AUTO ADJUST.] is touched to change to [ADJUST REX], [ADJUST TRIM]/[ADJUST ROI] is fixed at [ADJUST TRIM] and the density adjustment area cannot be changed.

– 69 –

Page 78: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.4 LUTThe gray scale curve used to display the image is modified using the lookup table (LUT).Any changes in these parameters are immediately reflected in the displayed image.Two types of gray scale adjustments are available: auto adjustment and REX adjustment (Reached EXposure index), and the type can be selected using [AUTO ADJUST.]/[REX ADJUST.].

AUTO ADJUST. (left) and REX ADJUST. (right)

7.4.1 AUTO ADJUST.The gray scale is adjusted based on the Fixed ROI (→ Setup Guide) or analysis value calculated from the Automated ROI.See Setup Guide for details about the ROI.

1 Select AUTO ADJUST.

Touch [AUTO ADJUST.]/[REX ADJUST.] and select [AUTO ADJUST.].

LUT graphThe relation between pixel value and density is shown in this graph.

Density

Pixel

Value

– 70 –

Page 79: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

2 Select the characteristic gray scale curve.

Touch to open the window and select.

When adjusting the REXIf the REX is adjusted and then [AUTO ADJUST.] is selected, a window is opened with the message “Any REX value settings will be lost, OK?” To change, touch [OK].

Table 7.4.1A: Gray scale Curves and Their Characteristics

Name Features

CHEST This is an S-shape curve for examining the chest.

LINEAR This is a linear lookup table.

INV. LINEAR This lookup table reverses the image density.

STAN. CONTRAST This is a standard S-shape curve. It is usually used for examining bones.

BONE #1 While maintaining contrast in a low density area, soft tissue can be observed in the high density area.

BONE #2A low density area can be seen almost the same as in BONE #1. However, soft tissue in the high density area can be observed more clearly.

CHEST #2 This is the version of the CHEST gray scale curve that supports high contrast.

BONE #3 This is the version of the BONE #1 gray scale curve that supports high contrast.

BONE #4 This is the version of the BONE #2 gray scale curve that supports high contrast.

Continued on the following page >>

– 71 –

Page 80: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

3 Set the corresponding density.

Touch or to select a setting from 1 to 29.The displayed value on the left is the saved value, and the value on the right is the modified value.The analysis value obtained by the Automated ROI or the Fixed ROI will be converted to a density corresponding to the setting. The following table is the density when the printer’s LUT is linear, the maximum density is 3.2 D, and the minimum density is 0.2 D.

4 Set the contrast.

Touch or to select a setting from 1 to 29.The displayed value on the left is the saved value, and the value on the right is the modified value.Higher values result in higher contrast, and lower values result in lower contrast.

Table 7.4.1B: Setting Values – DensitySetting 1 2 3 4 5 6 7 8 9 10

Density 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2

Setting 11 12 13 14 15 16 17 18 19 20

Density 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2.0 2.1 2.2

Setting 21 22 23 24 25 26 27 28 29

Density 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3.0 3.1

BrightnessBy adjusting the brightness, the density of the whole area can be changed. For example, a high brightness setting will shift the LUT graph toward the left, raising the density of the whole image, as illustrated in the right.Similarly, a low brightness setting will shift the LUT graph to the right, lowering the density of the whole image.

ContrastAdjusting the contrast setting as shown below changes the contrast of the whole image.

Density

Pixel

Value

UP

DOWN

Analysis Value

BRIGHTNESS

Density

Pixel

ValueAnalysis Value

UP

DOWN

DOWN

UP

– 72 –

Page 81: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.4.2 REX ADJUST.Use of the same REX (Reached EXposure index) will cause the density of the image to change with the X-ray dose, as in conventional film/screen systems.

1 Select REX ADJUST.

Touch [AUTO ADJUST.]/[REX ADJUST.] and select [REX ADJUST.].

2 Select the characteristic gray scale curve.

Touch to open the window and select the gray scale curve.

Table 7.4.2: Gray scale Curves and Their Characteristics

Name Features

CHEST This is an S-shape curve for examining the chest.

LINEAR This is a linear lookup table.

INV. LINEAR This lookup table reverses the image density.

STAN. CONTRAST This is a standard S-shape curve. It is usually used for examining bones.

BONE #1 While maintaining contrast in a low density area, soft tissue can be observed in the high density area.

Continued on the following page >>

– 73 –

Page 82: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

3 Set the REX.

Touch or to select a setting from1 to 16383.Or, touch the frame and enter the value from the ten key window.The larger the number, the lighter the entire image. The smaller the number, the darker the entire image.

BONE #2The low density area can be seen almost the same as in BONE #1. However, soft tissue in a high density area can be observed more clearly.

CHEST #2 This is the version of the CHEST gray scale curve that supports high contrast.

BONE #3 This is the version of the BONE #1 gray scale curve that supports high contrast.

BONE #4 This is the version of the BONE #2 gray scale curve that supports high contrast.

REX usage examplesTo observe over time a patient whose image has been taken before, take the image using the same exposure conditions, contrast, and gray scale curve as before.You can also observe under the same conditions by setting the gray scale to REX ADJUST. and setting the REX to the same value as before.

Table 7.4.2: Gray scale Curves and Their Characteristics

Name Features

– 74 –

Page 83: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

REXWhen REX is adjusted, the LUT graph shifts in the pixel value direction (right figure).

An enlarged graph is shown below.

This graph depicts an REX for a chest front exposure.The name of the quadrants, from the upper right in counterclockwise direction, are as follows: the first quadrant, the second quadrant, the third quadrant and the fourth quadrant. The first quadrant is the gray scale curve showing the relationship between pixel and density. The third quadrant is the sensor characteristics curve, showing the relationship between the X-ray dose that has reached the sensor and the sensor output value. The fourth quadrant is the logarithm conversion curve, which converts the sensor output value (proportional to the X-ray dose) into a pixel value proportional to the logarithm of X-ray dose. Only the right half of the graph (the first and the fourth quadrants) will be explained here to simplify the explanation.

First, an explanation of the algorithm that determines the gray scale curveThe CXDI selects the maximum pixel value of the upper lung area as the analysis value for chest PA exposure. In the above graph, the analysis value is 1220. Suppose 1.9 D has been specified as the density at the analysis value position. The gray scale curve shifts in the horizontal direction. The amount by which it shifts is determined by the position where the arrow from the analysis value intersects with the arrow from the density.

Next, an explanation of the algorithm that determines the REX.The REX is determined by using the gray scale curve. The position where the arrow from the gray scale curve at reference density 0.75 D (fixed) intersects with the logarithm conversion curve is the REX. Thus, the sensor output value corresponding to the area expressed as density 0.75 D is the REX.0.75 D is the density when the printer LUT is linear, the maximum density is 3.2 D, and the minimum density is 0.2 D.

Density

Pixel

Value

DOWN UP

REX

Extracted analysis

value positionDensity

X-Ray dose

Density = 1.9 D

Low X-ray dose

(High sensitivity)

High X-ray dose

(Low sensitivity)

Gray scale curve

(First quadrant)

Pixel value

(12 bit log)

Analysis value = 1220

Density = 1.9 D

(Second quadrant)Reference density = 0.75 D

REX = 375

(Fourth quadrant)(Third quadrant)

Analysis value = 1220

Sensor output

(14 bit linear)

Logarithm conversion curveSensor Characteristics Curve

– 75 –

Page 84: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.5 EDGE ENHANCE.

NOTE: When parameters are set to excessively high values, artifacts may be generated or effective image information may be degraded.If you think the image looks as if it has been over-processed, set the parameters to lower values, or void the process.

The edge enhancement process emphasizes the edges of the image and sharpens the image.However, it is difficult to check the effect of changes by modifying the parameters in the QA screen.

EDGE ENHANCE Screen

1 Enable edge enhancement.

Touch and highlight (white character) [EDGE ENHANCEMENT].

2 Set the degree of enhancement.

Touch or to select a setting from 1 to 11.The displayed value on the left is the saved value, and the value on the right is the modified value.The effect is proportional to the setting value.Higher values result in higher emphasis, and lower values result in lower emphasis.

3 Set the enhancement frequency.

Touch or to select a setting from 1 to 11.The displayed value on the left is the saved value, and the value on the right is the modified value.The larger the value, the clearer the definition of detailed (high frequency) structures, such as bone trabeculae. Low frequency values result in clearer definition of larger (low frequency) structures, such as soft tissue.

The settings and corresponding frequencies are shown below:

Table 7.5.1: Setting Values – Enhancement FrequencySetting 1 2 3 4 5 6 7 8 9 10 11

FREQUENCY(lp/mm) 0.06 0.08 0.11 0.15 0.20 0.27 0.36 0.47 0.69 0.81 1.28

– 76 –

Page 85: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.6 DEP (Display Enhancement Processing)

NOTE: When parameters are set to excessively high values, artifacts may be generated or effective image information may be degraded.If you think the image looks as if it has been over-processed, set the parameters to lower values, or void the process.

NOTE: If you wish to adjust both the LUT (→ page 70) and the DEP, void the DEP by touching the DEP button in DEP mode so the word is not highlighted. Adjust the LUT of the main area of interest in LUT mode first. Then, adjust the area of lesser interest with DEP parameters.

DEP widens the range that can be viewed on film or monitor by narrowing the pixel value range of areas of lesser interest without changing the view of the microstructure of the area of main interest.The effect of changes by modifying the parameters can be checked in the QA screen.

DEP Screen

1 Enable DEP.

Touch and highlight (white character) [DEP].

2 Set the high density effect.

Touch or to select a setting from +10 to -10.The value on the left is the value analyzed automatically (DEP analysis value). The value on the right is the effectivity that will be added to the analyzed value on the left side (table below).The larger the number, the stronger the processing.However, too large a value can make the parts of the image not covered by the patient very light, resulting in a poorer image.

Table 7.6.1: Setting Values – Corresponding Efficacy (high density)Setting -10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0

Efficacy 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00

Setting +1 +2 +3 +4 +5 +6 +7 +8 +9 +10

Efficacy -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 -0.35 -0.40 -0.45 -0.50

Continued on the following page >>

– 77 –

Page 86: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

Suppose the analyzed value is 0.30. If the setting is +1, the DEP will be 0.25. If the setting is -1, it will be 0.35.If the analyzed value is 1.00, it is the same as if the DEP had not been performed.The effect is limited to the 0 to 1 range.

3 Set the starting level for high density.

Touch or to select a setting from +10 to -10.The value on the left is the value analyzed automatically (DEP analysis value). The value on the right is the efficacy that will be added to the analyzed value on the left side (table below).The larger the number, the narrower the effective area.

Suppose the starting level is 2400. If the setting is +1, the starting level will be 2500. If the setting is -1, the level will be 2300.

Efficacy of high densityWhen the efficacy of high density is increased, the range of pixels will narrow after the image is processed at a higher level than the compression starting level.

Table 7.6.1B: Setting Values – Starting Level Changes (high density)Setting -10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0

Starting Level

Change-1000 -900 -800 -700 -600 -500 -400 -300 -200 -100 0

Setting +1 +2 +3 +4 +5 +6 +7 +8 +9 +10

Starting Level

Change100 200 300 400 500 600 700 800 900 1000

Starting level for high densityWhen the starting level is high as shown at the right, the effective range (shaded part) will narrow.

Pixel value before the process

Pixel value after the process

Compression starting level

0.35 (Setting -1)

0.3 (Setting 0)

0.25 (Setting +1)

Distribution frequency

Pixel

value

Distribution frequency

Pixel

value

Efficacy range

Starting level low

Efficacy range

Starting level high

– 78 –

Page 87: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

4 Set the low density effect.

Touch or to select a setting from +10 to -10.Although large values will improve the visibility of low density regions of the image, the use of too large a value may result in an increase of the graininess of the image.The value on the left is the value analyzed automatically (DEP analysis value). The value on the right is the efficacy that will be added to the analyzed value on the left side (table below).The larger the number, the stronger the processing.

Suppose the analyzed value is 0.50. If the setting is +1, the DEP will be 0.45. If the setting is -1, it will be 0.55.If the analyzed value is 1.00, it is the same as if the DEP had not been performed.The effect is limited to the 0 to 1 range.

5 Set the starting level of low density.

Touch or to select a setting from +10 to -10.The value on the left is the value analyzed automatically (DEP analysis value). The value on the right is the efficacy that will be added to the analyzed value on the left side (table below).A larger starting level results in a wider effect range, and a smaller starting level results in a narrower range.

Suppose the starting level is 1500. If the setting is +1, the starting level will be 1600. If the setting is -1, the level will be 1400.

Table 7.6.1C: Setting Values – Corresponding Efficacy (low density)Setting -10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0

Efficacy 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00

Setting +1 +2 +3 +4 +5 +6 +7 +8 +9 +10

Efficacy -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 -0.35 -0.40 -0.45 -0.50

Efficacy of low densityWhen the efficacy of low density is increased, the range of pixels will narrow after the image is processed at a lower level than the compression starting level.

Table 7.6.1D: Setting Values – Starting Level Changes (low density)Setting -10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0

Starting Level

Change-1000 -900 -800 -700 -600 -500 -400 -300 -200 -100 0

Setting +1 +2 +3 +4 +5 +6 +7 +8 +9 +10

Starting Level

Change100 200 300 400 500 600 700 800 900 1000

Pixel value before the process

Pixel value after the process

Compression starting level

0.45 (Setting +1)

0.50 (Setting 0)

0.55 (Setting -1)

Continued on the following page >>

– 79 –

Page 88: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

Starting level of low densityAs shown at the right, a smaller starting level for low density results in a narrower effect range (shaded part).

Distribution frequency

Pixel

value

Distribution frequency

Pixel

value

Efficacy range

Efficacy range

Starting level low

Starting level high

– 80 –

Page 89: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.7 POSITIONYou can trim the image and perform geometric transformations here.

POSITION Screen

7.7.1 Changing the Trim

1 Select the option to be adjusted.

Touch [ADJUST TRIM/ADJUST ROI] to select the range.

2 Set the range.

If [ADJUST TRIM] was selected, set the trimming area (→ page 68), and if [ADJUST ROI] was selected, adjust the density adjustment area (→ page 69).

3 Move the frame.

Touch , , or to move the frame. A red frame is moved if [ADJUST TRIM] was selected, and a red-purple frame is moved if [ADJUST ROI] was selected.

7.7.2 Changing Marks

1 Select the mark name.

Touch to select from the list.

2 Enable the mark.

Touch and highlight (white character) [View]. When View is enabled, Annotation and DICOM Output are enabled.When Link Setting (→ Setup Guide) is set to [ENABLE], Link is displayed.

Mark list window

Orientation button

Continued on the following page >>

– 81 –

Page 90: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

3 Set the orientation.

Touch the orientation button to open the orientation window and make the setting.

4 Close the ORIENTATION window

Touch [OK].

7.7.3 Change Laterality Information

1 Flips the image horizontally.

Touch and highlight (white character) [Marker].When Marker is enabled, Annotation is enabled and , , and can be selected. The selected contents are reflected to the Annotation and DICOM Output.When the Link setting (→ Setup Guide) is set to [ENABLE], Link is displayed.

2 Select laterality.

Touch the required button ( , , or ). The annotation view is switched.When Link is displayed, the display position of the linked annotation is also switched automatically.

Display/Button Description

Indicates the orientation of the image using when viewing a patient facing the CXDI from the x-ray generator side.

FLIPFlips the image vertically.

Flips the image horizontally.

ROTATE

Rotates the image 90° counterclockwise.

Rotates the image 180°.

Rotates the image 90° clockwise.

Link Setting and orientationIf the Link Setting has been set to [ENABLE], the changes will not be reflected even if you touch [SAVE PARAM] after changing the orientation.

Free annotationWhen the image has been rotated or flipped, free annotations may be outside the image.

Button Laterality Information Annotation Display

Left L

Right R

UnpairedNo display

Both Left and Right

– 82 –

Page 91: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.8 ANNOTATIONEnter annotations.See Setup Guide for details about the settings.

7.8.1 Fixed AnnotationText can be added to any selected location on the image.

ANNOTATION Screen (when free annotation is activated)

1 Open the DETAIL SETUP window.

Touch [FIXED ANNOTATION] (when free annotation is activated) or [DETAIL].

DETAIL SETUP Window

2 Select the annotation setting value.

Touch [Use System Settings]/[Don’t Use System Settings] ([Use System Settings]/[Use Saved Image Setting] in the case of a study list), and select the setting value.

Continued on the following page >>

– 83 –

Page 92: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

3 Enter annotation settings.

Set Position and Justification for the annotation.When the display position of the enabled link annotation is changed, the right and left sides are automatically switched.

4 Enter multi text.

Touch on the window and select a multi text from the preset list, or touch the multi text frame to enter text from the displayed software keyboard.For display of multi text annotations, see Setup Guide.

Multi Text List Window

5 Close the window.

Touch [OK].

Link Setting and annotationIf the Link Setting has been set to [ENABLE], link annotations will not be reflected even if you touch [SAVE PARAM] after changing the position or justification.

– 84 –

Page 93: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.8.2 Free Annotations

7.8.2.1 Adding free annotationsYou can add arrows and text at optional positions in the image.

1 Open the FREE ANNOTATION window.

Touch [FREE ANNOTATION].

FREE ANNOTATION Window

Software keyboard displayWhen “Automatic display of keyboard” is enabled through the Optional Setting under the Annotation Setup (→ Setup Guide), [ADD TEXT (KEYBD)] and [ADD TEXT (LIST)] will be displayed in the FREE ANNOTATION window.Touch [ADD TEXT (KEYBD)] to display the software keyboard, or [ADD TEXT (LIST)] to display the TEXT window.By touching these buttons, step 5 and step 6 are reversed.

TEXT Window

List

Continued on the following page >>

– 85 –

Page 94: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

2 Add an arrow.

Touch [ADD ARROW] to open the ADD ARROW window, and then touch two points on the image (start point and end point). A red arrow will be displayed.

ADD ARROW Window

3 Adjust the arrow.

Adjust the POSITION and WIDTH.

4 Close the ADD ARROW window.

Touch [OK].

5 Determine the position of text to be added.

Touch [ADD TEXT] to open the ADD TEXT window, and then touch one point on the image. A red “+” will be displayed, determining the text position.

ADD TEXT Window

Display Description

POSITION Touch , , , or to move the arrow.

WIDTH Touch the button to open the WIDTH window, and select the thickness from the list.

– 86 –

Page 95: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

6 Adjust the text.

Adjust the position, text, and font.

7 Close the ADD TEXT window.

Touch [OK].

8 Close the FREE ANNOTATION window.

Touch [EXIT].

Display Operation

POSITION Touch , , , or to move the text.

TEXT

Touch the inside of the text field to input text from the software keyboard. When selecting text from the list, touch to open the TEXT window, and then select.

TEXT Window

FONT

Touch the button to open the FONT window, and then touch to select the font and size from the list.

Continued on the following page >>

– 87 –

Page 96: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.8.2.2 Modifying free annotations

1 Select the arrow or text.

Select the arrow or text from the list.

2 Open the MODIFY window.

Touch [MODIFY]. The MODIFY ARROW window or MODIFY TEXT window will be displayed.

MODIFY ARROW Window (left) and MODIFY TEXT Window (right)

When a free annotation has gone out of the imageWhen a free annotation has gone out of an image due to trimming, rotating or flipping the image, or when an arrow or text extends out of the image when inputting a free annotation, >>> will be displayed on the left side of the list in the FREE ANNOTATION window.If you attempt to exit the QA screen with a free annotation extending from the image, a window will open and the message “Free annotation is out of frame. Are you sure you want to close QA Screen?” will be displayed.In this case, touch [CANCEL] to close the window, then modify the free annotation so that it is in the image.

– 88 –

Page 97: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

3 Change the settings.

To change the arrow, see step 3 under 7.8.2.1 Adding free annotations (→ page 85), and to change the text, see step 6 under 7.8.2.1 Adding free annotations (→ page 85).

– 89 –

Page 98: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.9 CORRECT PROC.In this screen, the dosage-dependent sharpness of the image can be set, the stripes caused by the fixed grid can be decreased, and the collimation mask can be set.

CORRECT PROC. Screen

7.9.1 Setting the Dosage-Dependent Sharpness Adjustment

1 Enable the dosage-dependent sharpness.

Touch and highlight (white character) [D. DEPEND MTF. IMP].

2 Set the type.

Touch to open the window and select the body part.

Type Button Window

3 Set the intensity as required.

Touch or to select a setting from 0 to 3.The displayed value on the left is the saved value, and the value on the right is the modified value.Larger values provide more sharpness.

– 90 –

Page 99: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.9.2 Setting SUPPRESS GRID ProcessingYou can use the SUPPRESS GRID function to decrease the stripes on the image caused by using fixed grid.For details on the SUPPRESS GRID function, see Setup Guide.

7.9.3 Setting COLLIMATION MASK Processing Function

NOTE: Advance setting by a service engineer is required to use the COLLIMATION MASK processing function.

The white areas around the periphery of the image are blackened out.Each time you touch [COLLIMATION MASK], the button switches between highlighted display (white characters) and normal display, and it is enabled when highlighted.

Without COLLIMATION MASK Processing (left) With COLLIMATION MASK Processing (right)

– 91 –

Page 100: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.10 Magnified Image ScreenThis displays a magnified image.

Magnified Image Screen

1 Display a magnified image.

Touch [MAG.] or [QA-MAG.].

2 Magnify the image.

Touch the part on the image that you want to magnify.To return to the original status, touch [RESET].

3 Return to the QA screen.

Touch [CLOSE].

Button Description

[MAG.] The image is displayed after LUT processing is completed.

[QA-MAG.]The image is displayed after all image processing is completed. The display processing time is longer than when [MAG.] is selected.

Display on high-definition monitorsThe system can be set by a service engineer to display images on a high-definition monitor instead of on the magnify screen.If this is done, steps 2 and 3 cannot be performed.

– 92 –

Page 101: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

7.11 Option

NOTE: Advance setting by a service engineer is required to use the JPEG SETUP.

Image compression can be set and the gamma can be adjusted.Touch [OPTION] to open the option window. If [OPTION] is not displayed, touch so that the button is displayed.

Option Window

7.11.1 Setting Image CompressionThe image is compressed when saved to storage or to an external storage medium.For details on the storage procedure, see Setup Guide.

7.11.2 GAMMA Adjustment

1 Open the GAMMA Adjustment window.

Touch for gamma adjustment.

GAMMA Adjustment Window

Entering a passwordIf a window opens and password entry (→ Setup Guide) is requested, touch the password frame to display the password screen and enter the password.

Continued on the following page >>

– 93 –

Page 102: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

7. Processing the Image (QA Mode Screen)

2 Enter the gamma value.

Touch the inside of the frame in the preview image (image displayed in the preview image screen or QA screen) or magnified image (image displayed in the magnify image screen) to display the numeric input keys, and then enter the gamma value in a range from 1.00 to 2.50.

3 Close the GAMMA Adjustment window.

Touch [OK]. This closes the window and saves the entered values as system setting values.

– 94 –

Page 103: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

CXDI Adjustment

8. Calibration

9. Self test

Page 104: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 96 –

Notes on CXDI Adjustment

Before Calibration1. The temperature in the room when performing calibration must be the same as during an ordinary exposure.

Otherwise, you will not be able to obtain a uniform image.2. Wait for the sensor unit to stabilize before calibrating the instrument (approximately 30 minutes after turning

ON the power).3. Do not use a AEC (Automatic Exposure Control).4. A more uniform image may be obtained by setting the tube voltage and exposure time to values used

frequently during the actual exposures.5. Uniformity of image may be improved when the exposure is performed many times.

Four or more exposures should be done.6. If the sensor unit’s grid is detachable, also refer to the hardware operating manual.7. When a multiple number of sensors are used, perform the calibration for each of the sensors.8. When a single sensor is set for a multiple number of sensor units, perform the calibration for each of the

units.9. With the CXDI-40G COMPACT, do not remove the grid.

When calibration is to be performed, ask the service technician to do this job.

During Calibration1. Align the system as for an ordinary exposure, such as when the whole imaging area of the sensor unit will

be exposed.Although calibration can be performed even by irradiating part of the sensor unit, you will not be able to obtain a normal image for the areas that have not been irradiated by x-rays.

2. Do not put any lead mark on the sensor unit, or place any object between the sensor unit and the X-ray tube.Such an object will appear on all the pictures.

3. If the message “Calibration Error.” “Too Much/Little Dose. Or collimator is used.” is displayed, perform calibration again after adjusting the X-ray exposure condition.

During Self Test1. Perform the exposure using the same settings as calibration for the exposure parameter settings and grid

usage. With the CXDI-40G COMPACT, proceed to undertake the self test while leaving the grid in place.2. In some cases, an error may be detected in TEST1 due to unintended exposure of the ion chamber or photo

timer inside the sensor unit. This is not a malfunction, and you may continue usage as before.3. Align the system as for an ordinary exposure, such as when the whole imaging area of the sensor unit will

be exposed.If a partial exposure occurs, an error may be detected.

4. Do not put any lead mark on the sensor unit, or place any object between the sensor unit and the X-ray tube.If an object obstructs exposure, an error may be detected.

Page 105: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

8. CalibrationCalibration is used to adjust the X-ray sensitivity of the sensor unit.Perform calibration in the following cases.1. When a warning message is displayed2. When the exposure conditions are changed significantly3. When the density of the image does not appear as desired

8.1 Setting the Parameters

1 Display the calibration screen.

From the exposure screen, touch [SYSTEM] > [CALIBRATION].

Calibration Screen

Display DescriptionLast Calibration The date and time of the last calibration exposuretimes The number of times that calibration exposures will be made.

Bucky msec Displayed when the sensor unit using the movable grid is selected.

Exposure Parameters

Starting from the top, the parameters are displayed in the following order: X-ray tube voltage, X-ray tube current, exposure time, mAs value, and tube-sensor distance.

Last Grid Grid name used previously. It is not displayed for sensor units CXDI-11 and CXDI-12.

Timeout for warning The time until a warning is displayed for performing the next calibration.

Memo Enter a memo as required.

Grid The grid name currently being used. It is not displayed for sensor units CXDI-11 and CXDI-12.

Sensor unit button

Exposureparameters

Continued on the following page >>

-97-

Page 106: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

8. Calibration

2 Select the sensor unit.

If multiple sensor units can be used, first select a sensor unit by touching that sensor unit button.

3 Remove the grid unit.

If a sensor unit with removable grid unit is selected in step 2, remove the grid unit.If [START] is touched while the grid unit is attached, it will display an error, and calibration will not be performed.

4 Set times (number of exposures).

Set the number using or .Or, touch the frame and enter the value from the ten key window.

5 Select the bucky time.

Touch to select from the list.This setting can only be made when using a movable grid.Calibration will not be performed properly unless the X-ray irradiation time is within the bucky time.

6 Set the exposure parameters.

Set the number using or .Or, touch the frame and enter the value from the ten key window or parameter table.

7 Set the timeout for warning.

Set the period from the completion of calibration until a warning is displayed to perform the next calibration. Touch [hour]/[day]/[week] to select the unit of time.Touch or to set the display duration.Or, touch the frame and enter the value from the ten key window.A warning will not be displayed if the time is set to 0.

8 Enter a memo as required.

Touch the frame to display the software keyboard and enter the memo in 64 or fewer alphanumeric characters.

Warning on calibrationThe warning on calibration will be displayed as a yellow [ERR] mark blinking on the lower right side of the message area.Although exposures can be performed even when the warning is displayed, you must remember to calibrate the instrument later, as the warning will disappear when another error occurs.

– 98 –

Page 107: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

8. Calibration

8.2 Performing Calibration

1 Start the calibration.

After setting the parameters on the calibration screen, touch [START].The unit is ready to perform an x-ray exposure.

Preparing for Exposure Screen

2 Press the exposure button.

Press the exposure button on the x-ray generator. Data processing starts.If data processing succeeds and two or more exposures are set, it is possible to proceed with the next exposure.Be sure to perform calibration exposures only for the number of set exposures.

Data Processing Success

Continued on the following page >>

– 99 –

Page 108: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

8. Calibration

3 Exposure is completed.

When the last exposure is completed, the message “Please wait.” is displayed, and data processing is started.

When data processing results in an errorA window is opened, and an error message is displayed.After checking the error details, touch [OK] to close the window.Touch [EXIT] to cancel calibration, or [RETAKE] to perform calibration again.

When the grid stripes have been detected in the course of the calibration ofthe CXDI-40G COMPACTA window is opened, and an error message is displayed.After checking the error details, touch [OK] to close the window.To cancel calibration, touch [EXIT]; to perform calibration again, check that the grid has been removed, and then touch [RETAKE].

– 100 –

Page 109: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

8. Calibration

4 After the calibration, result will appear.

When calibration is successfully completed, the message “Calibration has succeeded.” is displayed. Touch [CHECK] to return to the initial screen.

Calibration Successful

5 Change the exposure parameters.

If the settings of the x-ray generator were changed when starting calibration, change exposure parameters other than number of exposures and bucky time.The changed parameters will be saved when you touch [EXIT] or change the sensor unit.

6 Return to the menu screen.

Touch [EXIT].

7 Return the x-ray generator to the settings for normal exposure.

Be sure to change the settings of the x-ray generator to those for normal exposures.

When calibration has failedA window is opened, and an error message is displayed.After checking the error details, touch [OK] to close the window.Touch [EXIT] to cancel calibration, or [RETAKE] to perform calibration again.

REX checkIf exposure is performed with the settings for calibration exposure, the REX will display a value differing significantly from normal exposure, and a normal image will not be obtained.It is recommended that you always check the REX.

– 101 –

Page 110: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

8. Calibration

– 102 –

Page 111: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

9. Self testThis procedure checks the image data transfer path of the CXDI system and individual sensor performance within the sensor unit.

9.1 Self test using a sensor unit with a removable grid unit

1 This displays the Self Test Screen.

From the exposure screen, touch [SYSTEM] > [SELF TEST].

Self Test Screen

2 Select the sensor unit.

If multiple sensor units can be used, first select a sensor unit by touching that sensor unit button.

3 Remove the grid unit.

If a sensor unit with removable grid unit is selected in step 2, remove the grid unit.If [TEST START] is touched while the grid unit is attached, it will display an error, and the self test cannot be performed.

Continued on the following page >>

– 103 –

Page 112: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

9. Self test

4 Start the self test.

Touch [TEST START].The exposure parameters for the last calibration are displayed. Set the exposure parameters of the x-ray generator to match the displayed parameters.

Ready State

5 Press the exposure button.

Press the exposure button on the x-ray generator. The self test is started.The self test consists of five parts from TEST1 to TEST5. The self test is finished when TEST5 is completed.When the test is successfully completed, Passed. is displayed for TEST1 to TEST5.

– 104 –

Page 113: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

9. Self test

6 Return to the menu screen.

Touch [EXIT].

If the test failsA window opens, and an error code is displayed.After recording the error code, touch [OK] to close the window. Make a note of the error code and contact a Canon representative or distributor.

– 105 –

Page 114: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

9. Self test

9.2 Self test using the sensor unit with a grid unit which cannot be removed (CXDI-40G COMPACT)

Take the following steps for the self test using the CXDI-40G COMPACT.Steps which are not described here are the same as for a sensor unit with a removable grid unit. Refer to the steps described in 9.1 Self test using a sensor unit with a removable grid unit (see page 103).

1 Display the Self Test Screen.

From the exposure screen, touch [SYSTEM] > [SELF TEST].

Self Test Screen

2 Start the self test.

Without removing the grid, touch [TEST START].Although the exposure parameters for the self test can be changed, as a basic rule, their optimal values were set by the service technician at the time when the product was installed.Simply go ahead and perform the self test.

– 106 –

Page 115: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

9. Self test

3 Press the exposure button.

4 Return to the menu screen.

When the exposure parameters are to be changedWhen the exposure parameters are to be changed, touch [EXP. COND.].The Exposure Parameter Screen is displayed. Set the optimum parameters.

Exposure Parameter Screen

– 107 –

Page 116: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

9. Self test

– 108 –

Page 117: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Troubleshooting

10. Before Calling a Service Engineer

11. Questions and Answers

Page 118: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 110 –

Troubleshooting Notes

Remedies1. If an error continues to occur despite performing the remedies described here, please contact a Canon

representative or distributor.Before making your inquiry, be sure to make a note of the type of error that occurred, its number and details, and then explain the situation in as much detail as possible.

2. If an error occurs immediately after taking an image with PREVIEW IMAGE set to [DISPLAY] on the CUSTOMIZE DISPLAY screen (→ Setup Guide), do not touch [ERR] displayed in the lower right side of the message area until the preview image appears.

Page 119: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service EngineerIf a problem occurs, an error window is displayed or [ERR] is displayed in the bottom right message area.The error type and details are displayed in the error window as shown below. Use the information shown in the tables on the following pages to take the appropriate measure.If there is more than one remedy, select one.

Type of Error

Type of error Screen DisplayFatal Error The entire screen blinks in red and an error window opens.

Alert An error window opens, and a red [ERR] warning blinks in the lower right corner of the message area.

Error An error window opens, and a red [ERR] warning is displayed in the lower right corner of the message area.

WarningA yellow [ERR] warning blinks in the lower right corner of the message area.Touch [ERR] to open the error window.

AttentionA yellow [ERR] warning is displayed in the lower right corner of the message area.Touch [ERR] to open the error window.

Touch [ERR] to open the window.

Type of error

Type and number

Touch [CLOSE] toclose the window.

Description

– 111 –

Page 120: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

10.1 Fatal ErrorError Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

-100Sensor Unit: SelfCheck Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.If [SHUTDOWN] is displayed, touch it. The system will automatically shut down.

When these errors occur, note the error code and contact a Canon representative or distributor.

Detected an Error.

-101Capture Board SelfCheck Error

Detected Capture Board Error.

-102Capture Device Open Error

Failed to open capture device.

-103

READ Timeout Error

Failed to transmit data.Do not disconnect sensor during data transmission.

-104Requested I/O pending error

Failed to access the device.

-105CXDCAP.INI Open Error

Install CXDCAP.INI file.

-106Capture Board SelfCheck Error

Detected Capture Board Error.

-107Power Supply Error

Shuts down after 10 seconds.

-522File Open Error: (file name)

File is not found.

– 112 –

Page 121: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

10.2 AlertError Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

-4

System Info Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Driver info is updated. If there is the trouble of sensor, the sensor power unit may not be supplied.

-5

System Info Error

Capture device info is updated. If there is the trouble of sensor, the sensor power unit may not be supplied.

-6

System Info Error

A/D board info is updated. If there is the trouble of sensor, the sensor power unit may not be supplied.

-7

System Info Error

LANMIT info is updated. If there is the trouble of sensor, the sensor power unit may not be supplied.

-8System Self Check Error: (Contents)

Call a service engineer and let him/her know the error code.

-115Capture Board Error

Detected Capture Board Error.

-116Sensor Unit: Self Check Error

Detected an Error.

-119

Image Error Touch [OK] to close the window.Perform operations according to the messages displayed.

Note the error code and contact a Canon representative or distributor.

An error occurred in the image. Call a service person. Check the image in the Preview-screen and also in the high-resolution monitor or by using [MAG.] or [QA-MAG.] in the QA screen. Also, turn the power box off and on again to restart the CXDI.

-546

Internal Storage Limit Warning:

Touch [OK] to close the window.The amount of available internal storage space has dropped below (current settings). Images currently in storage commitment or that failed in storage commitment will be automatically deleted.

-547

Internal Storage Limit Warning: Touch [OK] to close the window and then see 6.4 Deleting the Image(s) (→ page 58), and delete the image.The amount of available internal storage space has dropped below (current

settings). Check the contents of the study list.

-5002OPU-CCR RPC Error Touch [OK] to close the window, and then see

1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

OPU-CCR RPC Error.

-6001OPU-CCR RPC Error

Operation has been canceled.

– 113 –

Page 122: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

10.3 ErrorError Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

-2Insufficient tmp disk space

Touch [OK] to close the window, and then see 6.4 Deleting the Image(s) (→ page 58), and delete the image.

Output the saved data or erase the backed up data.

-3Insufficient que disk space

Output the saved data or erase the backed up data.

-120Sensor Unit: SelfCheck Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Detected an Error.

-121Sensor Unit: Sequence Error

Sequence Error. Call a service engineer.

-122Sensor Unit: Control Error

Detected Control Error. Call a service engineer.

-123Sensor Unit: PLD Error

Detected PLD Error. Call a service engineer.

-124

Sensor Unit: Grid Error • This cuts off control of the movable grid. The grid lines appear over the image, and the image quality cannot be guaranteed. To continue with the exposure anyway, touch [OK].

• Touch [CANCEL] to re-initialize the grid unit.

If you select [OK], the system will separate the Grid component, then change to force-Grid mode. In this case, the image quality cannot be guaranteed. If you select [CANCEL], Grid component will be re-initialize.

-125Sensor Unit: Grid Error Touch [OK] to close the window, and then

check the image. Perform exposure again if there are any abnormalities.Because of the Grid error, this image may have a quality problem.

-126

Sensor Unit: Temperature ErrorTouch [OK] to close the window and try again after a while.The internal temperature of the Sensor Unit exceeded the temperature limit.

Cannot capture images until the temperature goes below the limit.

-127

Sensor Unit

Touch [OK] to close the window and lock the slide adapter (optional unit) or the grid unit.

Sensor Unit cannot become ready for exposure.Check the following part:(message)

-128

Sensor Unit: Signal Line Error Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Problem has occurred between sensor unit and EO box. Call a service engineer.

-129Sensor Unit: Data Transmit Error Touch [OK] to close the window. As image

quality will not be guaranteed, take an image again.Failed to transmit the data.

-130Capture Board Error Touch [OK] to close the window, and then see

1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Detected Capture Board Error.

-131Capture Board Control Error

Detected Capture Board Control Error. Call a service engineer.

– 114 –

Page 123: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

-132Capture Board Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Capture Board Error. Call a service engineer.

-134Capture Board: Sequence Error

Failed to synchronize with the sensor unit. Call a service engineer.

-135Capture Board: Sequence Error

X-ray monitor in sensor unit failed to detect the beginning of irradiation.

-136Capture Board: Data Transmit Error

Failed to transmit the data. Call a service engineer.

-137Capture Board: Control Line Error

Control line error has occurred in capture board. Call a service engineer.

-138Sensor Unit: Sensor Analogue Power Supply Error

Sensor Analogue Power Supply was detected. Call a service engineer.

-139

Grid ID Readout Error Touch [OK] to close the window. Exposure can be continued.When these errors occur, note the error code and contact a Canon representative or distributor.

Grid ID readout error has been detected.

-140

Acquisition Parameter Error Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Detected Illegal Parameter Error.

-141LUT Write Error

Touch [OK] to close the window, and then see 8. Calibration (→ page 97) to perform calibration.

Quit the exam and perform calibration.

-143Calibration Table Error

Perform Calibration.

-144Sensor Unit Configuration Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Incorrect Sensor Unit Configuration. Call a service engineer.

-145Sensor Unit: Firmware Write Error

The firmware is incorrect.

-146

Capture Device Access Error

Detected Communication ErrorEither an error occurred in the sensor, or the sensor power may be turned off.

-147

Capture Device Access Error If sensor power is turned off, turn the sensor power on and try performing the operation again.If sensor power is on, note the error code and contact a Canon representative or distributor.

Detected Communication ErrorEither an error occurred in the sensor, or the sensor power may be turned off.

-148Sensor Unregistered Error. Touch [OK] to close the window, and then

connect a registered sensor.The connected sensor cannot be used because it is unregistered.

Error Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

Continued on the following page >>

– 115 –

Page 124: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

-149

Sensor Connection Error Touch [OK] to close the window. Then, turn off the power of unused sensor unit(s).Turn on only one sensor unit when multiple sensor units share the same body part.

Multiple sensors sharing the same BodyPart cannot be used at the same time. Turn on the power for one sensor only that will be used.

-302Calibration Error

Touch [OK] to close the window. Adjust exposure dose (mAs) or focal distance and take an image again.

Too Much Dose. Or collimator is used.

-303Calibration Error

Too Little Dose. Or collimator is used.

-304Calibration Error

• Touch [OK] to close the window and repeat the exposure.

• Please contact Canon if the irradiation field aperture is not being used.

Invalid Data. Or collimator is used.

-305Calibration Error

Collimator may be used.

-306

QA Process Error Touch [OK] to close the window, and then see 7. Processing the Image (QA Mode Screen) (→ page 65) to manually perform QA processing.

Image Analysis Error. Adjust image process parameter manually.

-307

Calibration Error Check that the grid has been removed, and then retake.When calibration is to be performed for the CXDI-40G COMPACT, ask the service technician to do this job.

Periodic noise is detected in acquisition data. It is necessary to remove the grid before calibration.

-310

REX Warning

Touch [OK] to close the window and adjust the dose.

REX is (detected value). It is below lower limit (setting values). Check generator setting and exposure conditions. If exposure conditions were correct, adjust IP parameters.

-311

REX Warning

REX is (detected value). It is over maximum limitation (setting values). Check generator setting and exposure conditions. If exposure conditions were correct, adjust IP parameters.

-402Incomplete End Error Touch [OK] to close the window and repeat the

exposure.The last image in the last study was not saved completely.

-403

Incomplete End Error• Touch [OK] to recover.• Touch [CANCEL] to delete the image.Do you want store captured images into internal storage in order to restore

them later?

-509External Storage Error Touch [OK]. Change the attribute of the storage

medium and try again.Selected directory or disk for storage can only be read.

-512External Storage Error Touch [OK] to close the window and check the

setting of the external storage medium.Cannot recognize the external storage. Set the disk properly and retry.

-513Media Control Error Touch [OK] to close the window and try again

after a while.Operation has been canceled because the external storage or directory is busy.

Error Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

– 116 –

Page 125: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

-514

External Storage Error Touch [OK] to close the window, finish the study, and output the image.Also, see Setup Guide, and check that the maximum storage space of the external storage media is set to an appropriate setting.

The exposed images cannot be stored because the size of folder partition is smaller than the exposed images. Use a new disk, erase the folder contents or enlarge the folder size and retry.

-517

Internal Storage ErrorTouch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Exposed image cannot be stored, because the capacity of internal storage memory is insufficient.

-518

Internal Storage Error

Exposed image cannot be stored, because the capacity of internal storage memory is insufficient.

-519

External Storage ErrorTouch [OK]. Replace the external disk and try again.Capacity of external storage memory is insufficient. Retry after making a space

in the disk.

-520

Internal Storage Error Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Exposed image cannot be stored, because the capacity of internal storage memory is insufficient.

-521Queue File Open Error

Queue file is not found.

-533

Reject Info Output Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Selected directory for reject list can only be read. Reject info cannot be saved on the file.

-535

Reject Info Output Error

Selected directory for reject list does not exist, or attribution is incorrect. Reject info cannot be saved on the file.

-537

Reject Info Output Error

The file format for reject list includes incorrect strings. Reject info cannot be saved on the file.

-544Storage Commitment Result Error

Receipt of the storage commitment failed.

-545Storage Commitment Result Error

Storage commitment failed.

-549

External Storage Error (folder name)

An error has occurred in DICOMDIR. Files may have been destroyed. Call a service person. The DICOMDIR functions will remain canceled until the system is restarted.

-599

File Open Error Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

File is not found.

Error Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

Continued on the following page >>

– 117 –

Page 126: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

-601Resend Error

Touch [OK] to close the window.File may have been automatically deleted.

-5100Sensor Unit: Detect Error Touch [OK] and shut down the system (see 1.2

Shutting Down the CXDI (→ page 5)). Check the cable connections.

If the cause cannot be found, note the error code and contact a Canon representative or distributor.

Check the sensor power unit or cable connection.

-5101

Sensor Unit: Detect Error

Check the sensor unit cable connection. If you want to recover the system, shut it down.

-5102Grid Error Touch [OK] to close the window and remove

the grid from the sensor unit.Remove Grid for this exposure.

-5202

Grid ID ErrorTouch [OK] to close the window and check the grid in the sensor unit.Replace with the recommended grid (Grid Name).

Installed grid: (Grid Name).

-5203Grid ID Error Touch [OK] to close the window and remove

the grid from the sensor unit.This is a none-grid mode. Remove the grid.

-5206Grid ID Error

Touch [OK] to close the window and attach the grid to the sensor unit.

This is a fixed-grid mode. Attach the grid.

-5207Grid ID Error

This is a moving-grid mode. Attach the grid.

Error Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

– 118 –

Page 127: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

10.4 WarningError Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

-9

SMART Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

An error is detected in HDD.It's possible that the HDD(1 or 2) will break down. Call a service person.

-10

HDD Communication Error

Cannot communicate with HDD.Mirroring function will be disabled since the HDD(1 or 2) is not available. Call a service person.

-11

SMART Error

An error is detected in HDD.It's possible that the HDD(1 or 2) will break down and the data will be lost. Call a service person.

-150Sensor Unit: SelfCheck Error

Detected an Error.

-151Sensor Unit: Flash ROM Error

Detected Flash ROM Error. Call a service engineer.

-152Sensor Unit: Sequence Error Touch [OK] to close the window. As image

quality cannot be guaranteed, take an image again.This image may have a problem.

-153

Sensor Unit: Temperature Warning Touch [OK] to close the dialogue window. Lower the temperature of the room where the sensor unit is installed, or wait until the temperature in the sensor unit drops.

The internal temperature of the Sensor Unit is too high!Do not leave the Sensor Unit in READY condition.

-160Capture Board Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Detected Capture Board Error.

-161Capture Board Error

Capture Board Error.

-162Capture Board: Sequence Error

X-ray monitor in sensor unit failed to detect the beginning of irradiation.

-163Capture Board: Sequence Error

X-ray monitor in sensor unit failed to detect the end of irradiation.

-164Image quality attention Touch [OK] to close the window. As image

quality cannot be guaranteed, take an image again.CXDI has stopped irradiation. Check image.

-502

DICOM Connect Error • After confirming that the network is normal, touch [RETRY].

• Touch [CANCEL] to close the window, and then see Setup Guide and set the parameters.

• If running the software offline, check that the output status of the process viewer is [IDLE]. If output status is [SEND], change it to [IDLE].

Cannot connect to target. Check network or port number setting.(The following message is added when transferring all data at once.)If operating offline, press the Close button and check the output status of the Process Viewer.

-503DICOM Transfer Error

Error has occurred during association.

Continued on the following page >>

– 119 –

Page 128: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

-504

DICOM Parameter Error • Touch [RETRY].• Touch [CANCEL] to close the window, and

then see Setup Guide and set the parameters.

DICOM transfer Parameter is wrong.

-505

DICOM Response Error • Check that the printer or storage medium is normal. Then touch [RETRY].

• Touch [CANCEL] to close the window and then check the settings.

Receiving device reported an error in DICOM response. Check receiving device.

-507

DICOM Printer Status Error • After eliminating the printer error, touch [RETRY].

• Touch [CANCEL] to cancel output.CXDI has aborted DICOM transmission because of a FAILURE in DICOM printer status. Check receiving device.

-509External Storage Error

• Replace the disk with a writable disk and touch [RETRY].

• Touch [CANCEL] to cancel save.• Contact the system administrator.

Selected directory or disk for storage can only be read.

-510

External Storage Error

Selected directory or disk for storage does not exist. Or image cannot be saved on external storage, because the external storage or directory is not properly set.

-512External Storage Error

Cannot recognize the external storage. Set the disk properly and retry.

-514

External Storage Error

The exposed images cannot be stored because the size of folder partition is smaller than the exposed images. Use a new disk, erase the folder contents or enlarge the folder size and retry.

-515Internal Storage Error Touch [OK] to close the window, and then see

1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Selected directory for storage can only be read.

-516Internal Storage Error

Selected directory for storage does not exist, or attribution is incorrect.

-517

Internal Storage Error Touch [RETRY] after the green mark disappears from the message area of the exposure screen.

Exposed image cannot be stored, because the capacity of internal storage memory is insufficient.

-518

Internal Storage Error • Touch [RETRY] to store the image on the internal storage again.

• To cancel saving, touch [CANCEL].Exposed image cannot be stored, because the capacity of internal storage memory is insufficient.

-519

External Storage Error • Replace the disk with a writable disk and touch [RETRY].

• Touch [CANCEL] to cancel save.• Contact the system administrator.

Capacity of external storage memory is insufficient. Retry after making a space in the disk.

-520

Internal Storage Error • Touch [RETRY] to store the image on the internal storage again.

• To cancel save, touch [CANCEL].Exposed image cannot be stored, because the capacity of internal storage memory is insufficient.

-528

DICOM Storage Commitment Warning

Touch [OK] to close the window.Processing has not been performed, although parameters of commitment have been specified. Check the execution path of commitment.

Error Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

– 120 –

Page 129: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

-533

Reject Info Output Error

Touch [OK] to close the window, and then see 1.2 Shutting Down the CXDI (→ page 5), and turn off the CXDI.

When these errors occur, note the error code and contact a Canon representative or distributor.

Selected directory for reject list can only be read. Reject info cannot be saved on the file.

-535

Reject Info Output Error

Selected directory for reject list does not exist, or attribution is incorrect. Reject info cannot be saved on the file.

-537

Reject Info Output Error

The file format for reject list includes incorrect strings. Reject info cannot be saved on the file.

-539Reject Info Output Error • Touch [RETRY] to store the image on the

storage again.• To cancel save, touch [CANCEL].Error has occurred on saving reject info on the file. Do you retry it?

-540

DICOM Connect Error

• After confirming the storage is normal, touch [RETRY].

• Touch [CANCEL] to close the window and then check the settings.

Transmission of the storage commitment request failed. Connection to output equipment is not possible. Check the network or port number settings. Retry?

-541

DICOM Transfer Error

Transmission of the storage commitment request failed. An error occurred during transmission with the output equipment. Retry?

-542

DICOM Parameter Error

Transmission of the storage commitment request failed. The DICOM output parameter settings are invalid. Retry?

-543

DICOM Response Error

Transmission of the storage commitment request failed. The DICOM response reported by the receiving equipment was an error. Check the receiving equipment. Retry?

-548

Storage Commitment Request Error

Touch [OK] to close the window.The storage transfer was canceled, and so the storage commitment request was not issued.

-6000

Calibration Error Touch [CLOSE] to close the window, and then see 8. Calibration (→ page 97) to perform the calibration.

(Sensor unit ID: 1 to 4): (Elapsed time) have passed since the last calibration. Perform calibration.

Error Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

– 121 –

Page 130: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

10. Before Calling a Service Engineer

10.5 AttentionError Code

Error Message (Upper box: Error type Lower box: Error description) Remedy

-506

DICOM Response Warning

Touch [OK] to close the window and check the settings of output destination.

Although transfer finished properly, receiving device reported a warning in DICOM response. Check receiving device.

-508

DICOM Printer Status Warning

Although transfer finished properly, receiving device reported a warning in DICOM printer status. Check receiving device.

-511External Storage Attention Touch [OK] to close the window. Replace the

external disk before it reaches full capacity.Remaining space in external storage is less than (setting values)%.

-550Internal Counter Mismatch Error (folder name)

Touch [OK] to close the window.The setting file was restored since there was a mismatch in the internal counter.

– 122 –

Page 131: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

11. Questions and Answers

11.1 Display of Exposure Screen

Q1-1 The exposure screen does not appear on the touchscreen.

A1-1

Reset the system according to the display on the touchscreen.Contact a Canon representative or distributor if the problem cannot be resolved even after resetting the system several times.• When the Microsoft Windows XP Professional screen is displayed

Select [start] > [Turn Off Computer] > [Turn Off].• When the exposure screen appears completely black or completely blue without showing anything else

Refer to the hardware manual, and reset the system.

Q1-2Nothing is displayed on the touchscreen even when the power is turned ON, or the screen has frozen with the message Checking NV RAM....

A1-2 Refer to the hardware manual, and reset the system.

Q1-3 No operations can be performed from the touchscreen.

A1-3Operation from the touchscreen will not be accepted while the system is performing a process or when the system has crashed. Wait for about one minute and try again.If it still doesn’t work, refer to the hardware manual, and reset the system.

Q1-4 What is the green mark blinking at the upper right side of the exposure screen?

A1-4 It indicates that the image is being transferred or the image is being processed. It disappears after the process ends.

Q1-5 What are the S and I displayed in the message area of the exposure screen?

A1-5 They are the original IDs issued by the system for DICOM.

Q1-6What are the small green rectangles in the part where the preview image will be displayed on a tray type screen after an exposure?

A1-6 The rectangle shows the Fixed ROI area.For details about the Fixed ROI area, see Setup Guide.

Q1-7 There is no study description button on the exposure screen.

A1-7 Set the Study Description to [ENABLE]. (→ Setup Guide)

Q1-8 The new exposure mode button that has been made is not displayed on the exposure screen.

A1-8 On the tray type screen, the exposure mode will not be displayed unless it has been POSITION IN TRAY (→ Setup Guide).

Q1-9 The print preview image is not displayed properly on the multiview screen.

A1-9 Refer to 5.4.2 Choosing Multi-Format after the Exposure (→ page 49) and then set the printer for output.

– 123 –

Page 132: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

11. Questions and Answers

11.2 Entering/Changing Data

Q2-1 There are exposure parameters that are not displayed on the exposure screen.

A2-1You can select the exposure parameters to be displayed from the SYSTEM SETUP screen (→ Setup Guide).Parameters set to off are not displayed.

Q2-2 Do you have to enter all the data on a patient?

A2-2 An error occurs if any of the required fields (→ Setup Guide) are not completed.The required fields can be selected.

Q2-3 How can patient data be changed after input?

A2-3 Touch [PATIENT]. Now the data can be changed. The [OK] at the bottom right of the software keyboard changes to [CHANGE]. To exit, touch [CHANGE].

Q2-4 How can exposure parameters be changed on the category type screen?

A2-4 Unlike the tray type screen, they cannot be changed directly from the category type screen. Touch [CHANGE] to display the screen for changing parameters.

Q2-5Although the exposure parameters have been changed, they returned to the previous settings after an exposure was performed.

A2-5 Unless you touch [SAVE PARAM] and save the parameters after changing them (→ page 18), the changed parameters will return to the previous settings after an exposure has been taken.

Q2-6 There is a parameter that cannot be changed during the exposure operation.

A2-6 For msec and mAs, only the one that has been selected on the SYSTEM SETUP screen can be changed.Selection can be made in the Exposure Condition Display (→ Setup Guide).

Q2-7 Exposure parameters cannot be changed freely.

A2-7If [ENABLE] has been selected for PARAMETER TABLE (→ Setup Guide) on the SYSTEM SETUP screen, only the preset values in the exposure parameter table will be displayed. Setting to [DISABLE] allows setting of any user-selected value.

– 124 –

Page 133: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

11. Questions and Answers

11.3 Settings

Q3-1 A new exposure mode button cannot be made.

A3-1To create a new exposure method, copy the exposure method used before and then edit it. (→ Setup Guide)This eliminates the need to enter the parameters from the beginning.

Q3-2 What is a bucky msec?

A3-2 It is a measure of the time for linking the movable grid with the X-ray irradiation. When it is not set properly, vertical lines may appear on the image.

Q3-3 What is the difference between the fixed ROI and phototimer?

A3-3 In the Fixed ROI settings, the adjustment range for LUT processing (→ page 70) is set. By setting the phototimer, the position of the phototimer installed in the CXDI is set.

Q3-4 What is an X-ray monitor?

A3-4 It is used to detect the completion of X-ray irradiation and judge the timing for image generation.Normally, this cannot be changed. (→ Setup Guide)

– 125 –

Page 134: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

11. Questions and Answers

11.4 Exposure

Q4-1 The message QA Process Error has appeared.

A4-1There are several causes. Typical examples and remedies are shown below:• You have pressed the wrong exposure mode button. Select the correct exposure mode.• Adjust the REX.

Q4-2 The density of the image cannot be adjusted as desired.

A4-2

There are several causes. Typical examples and remedies are shown below:• Check the QA processing parameters. (→ page 65)• Check the IP PARAMETER for the exposure method. (→ Setup Guide)• Perform calibration. (→ page 97)• Execute SELF TEST. (→ page 103)

Q4-3 The image is completely white or black.

A4-3

There are several causes. Typical examples and remedies are shown below:• The X-ray dose may be wrong, or the CXDI system may not be synchronizing with the x-ray generator.

Check the settings of the x-ray generator.• You have pressed the wrong exposure mode button. Select the correct exposure mode.• Adjust the REX.

Q4-4 Why is an exposure mode chosen automatically after entering a patient’s data?

A4-4 Because AUTO-PRESS (→ Setup Guide) is set to ENABLE on the SYSTEM SETUP screen.

– 126 –

Page 135: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

11. Questions and Answers

11.5 Image Output

Q5-1 The image cannot be output to the printer, etc.

A5-1

In some cases, it is difficult to determine the cause. Check the following items:1. Check the image output status in the Process Viewer.

It takes time when there are many images waiting to be output.If you are in a hurry, touch [MODIFY] to display the CHANGE DESTINATION window, and then touch [Priority] to set to Priority Output. (→ page 31)

2. Check that the output destination is selected. (→ Setup Guide)If exposure finishes with nothing selected, no image is output. In this case, after the output destination is selected, the image in the Study List (→ page 53) can be loaded and output.

3. Check that there is no problem with the printer, etc. where the image is going to be output.4. Check that there is no problem in the network.

Q5-2 Part of the printed image was missing.

A5-2Check the printer settings. (→ Setup Guide)In handling, if [LIFESIZE] is selected and a large value is set for the Margin ratio, a large portion of the image may be cut away. Redo the settings.

Q5-3 Can an image that has already been output be sent to another destination?

A5-3 If the study is displayed in the Study List (→ page 53), the image can be resent after it is loaded.

– 127 –

Page 136: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

11. Questions and Answers

– 128 –

Page 137: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Appendix

A. Logging in (User Authentication Function)

B. Using the Rejection Reason Input Function

C. Using the Sensor Unit Attach/Detach Function

D. Turning the Power of the Power Box On/Off

Page 138: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

– 130 –

Notes on Setting the User Authentication Function

Note on User Authentication Function1. Advance setting by a service engineer is required to use the user authentication function.2. Only logged in users can use the CXDI when the user authentication function is set.

A user name and password are required to log in to the CXDI when the user authentication function is set.

Note on Rejection Reason Input FunctionAdvance setting by a service engineer is reguired to use the rejection reason input function.

Page 139: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

A. Logging in (User Authentication Function)A.1 User Authentication Function

Advance setting by a service engineer is required to use the user authentication function.When this function is set, the login screen is displayed, and the exposure screen will not be displayed unless the correct user name and password are entered.For details about user registration, see Setup Guide.

Login Screen

[SHUTDOWN] display[SHUTDOWN] will be displayed only in the initial Login screen after starting up the CXDI.When you touch [SHUTDOWN], a window will open and the confirmation message “Are you sure you want to shutdown the system?” will appear. Touch [OK] to shut down the system.

– 131 –

Page 140: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

A. Logging in (User Authentication Function)

A.2 LoginEnter the user name and password registered to the CXDI to log in.

1 Display the User Name Input screen.

Touch the box for the User Name in the login screen.

User Name Input Screen

2 Enter the user name

Enter the user name, then touch [OK]. The Password Input screen is displayed.

Password Input Screen

3 Enter the password

Enter the password, then touch [OK].After the user name and password registered to the CXDI are entered, the exposure screen is displayed.

– 132 –

Page 141: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

A. Logging in (User Authentication Function)

Logout button displayAfter logging in, a logout button is displayed at the bottom left of the exposure screen.The logout button displays the currently logged in user name.

Logout Button

– 133 –

Page 142: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

A. Logging in (User Authentication Function)

A.3 LogoutLogout can be executed by two methods: touching the logout button, or automatically logging out using a timer (Auto Logout).

A.3.1 Performing Manual LogoutTouch the logout button in the exposure screen. A window will open and the confirmation message “Are you sure you want to log out?” will be displayed. (If a study is in progress, the following message is displayed: “Now performing examination! Are you sure you want to log out?”) Touch [OK].

Logout Confirmation Screen

A.3.2 Auto LogoutIf the Auto Logout function is enabled, a logout is performed automatically if no screen operations are performed for a certain period of time.For details about Auto Logout settings, see Setup Guide.

Conditions when the Auto Logout function will not activateThe Auto Logout function will not activate under the following conditions.• At the User Management screen, User Properties screen, or Password Input screen• When [SETUP MENU], [EDIT EXPOSURE MODE], [CALIBRATION], or

[SELF TEST] is selected and something is set at the SYSTEM SETUP screen• When a message window is open• When an error window is open• When an image is called from the Study List

– 134 –

Page 143: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

B.1 Rejection Reason Input FunctionAdvanced setting by a service engineer is required to use the rejection reason input function. When this function is set, the user must input the reason when an image is rejected (→ page 40) during exposure.The entered rejection reason is output to a CSV (Comma Separated Value) format file (except when reproduced from the study image list).

Rejection Reason Input Window

CSV format file outputInformation other than the rejection reason can be output to a CSV format file if the setup has been performed by the service engineer.For details, please contact a Canon representative or distributer.

– 135 –

Page 144: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

B.2 Operations during Study

B.2.1 Rejecting an Image During a StudyDuring a study, the image is rejected in the preview image screen or multiview screen.

1 Display the preview image screen.

Touch a preview image button on the exposure screen.The image can be displayed immediately after exposure for certain settings.

Preview Image Screen

2 Display the the image to be rejected.

Touch a preview image button.When displaying from the multiview screen, touch [MULTI VIEW].

3 Display the rejection reason input window.

Touch [Reject].When the multiview screen is displayed, directly touch the image to open the window, and then touch [Reject].

Rejection Reason Input Window

– 136 –

Page 145: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

4 Select or enter the rejection reason.

Select the rejection reason from the Reject Reason List.If the reason you want to enter is not found in the Reject Reason List, touch the Reject Reason input box to display the keyboard, and then enter the reason.

5 Reject the image.

Touch [OK].When the preview image screen is displayed, the rejection reason is displayed on the image.It is not displayed in the multiview screen.

B.2.2 Entering the Rejection Reason during RetakeWhen retaking an exposure, the image selected for retake always becomes a reject image.

1 Display the preview image screen.

Touch a preview image button on the exposure screen.The image can be displayed immediately after exposure for certain settings.

Preview Image Screen

2 Display the image that will be retaken.

Touch a preview image button.

Rejection reasonIf the image was already rejected in the past, its rejection reason is entered.

Continued on the following page >>

– 137 –

Page 146: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

3 Select [RETAKE].

Touch [RETAKE]. A window will open and the confirmation message “The original image will be ‘rejected’. Do you want to retake the image?” will be displayed. (This message may not appear depending on the setup performed by the service engineer.) Touch [OK] to retake the image. The rejection reason input window is opened.

Rejection Reason Input Window

4 Select or enter the rejection reason.

Select the rejection reason from the Reject Reason List.If the reason you want to enter is not found in the Reject Reason List, touch the Reject Reason input box to display the keyboard, and enter the reason.

5 Reject the image.

Touch [OK]. The exposure screen appears and the system enters the ready condition with the same exposure mode selected as that used for the current image.You cannot select a different exposure mode for retaking the image.

6 Retake the image.

The image displayed in Step 2 will be rejected.

– 138 –

Page 147: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

B.2.3 Changing the Rejection Reason During a StudyThe rejection reason can be changed in the preview image screen or multiview screen.

1 Display the preview image screen.

Touch a preview image button on the exposure screen.The image can be displayed immediately after exposure for certain settings.

Preview Image Screen

2 Display the image whose rejection reason you will change.

Touch a preview image button.To select from the multivew screen, touch [MULTI VIEW].

3 Display the process selection dialog box.

Touch [Reject].When the multiview screen is displayed, directly touch the image to open the window, and then touch [Reject].

Process Selection Dialog Box

4 Display the rejection reason input window.

Touch [CHANGE REJECT REASON].The current rejection reason is displayed in the Reject Reason input box.

Continued on the following page >>

– 139 –

Page 148: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

5 Select or enter the rejection reason.

Select the rejection reason from the Reject Reason List.If the reason you want to enter is not found in the Reject Reason List, touch the Reject Reason input box to display the keyboard, and then enter the reason.

6 Change the rejection reason.

Touch [OK].When the preview image screen is displayed, the rejection reason is displayed on the image.It is not displayed in the multiview screen.

B.2.4 Cancelling a Rejected Image During a StudyCancel rejection of the image.

1 Display the preview image screen.

Touch a preview image button on the exposure screen.

Preview Image Screen

2 Display the image whose rejection you want to cancel.

Touch a preview image button.When displaying from the multiview screen, touch [MULTI VIEW].

– 140 –

Page 149: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

3 The process selection dialog box is displayed.

Touch [Reject].When the multiview screen is displayed, directly touch the image to open the window, and then touch [Reject].

Process Selection Dialog Box

4 Cancel the rejection.

Touch [VALIDATE IMAGE].

Cancelling rejection of an image that has been retakenWhen rejection of a retaken image is cancelled, the other image is rejected, and the rejection reason input window is opened.Enter the rejection reason for the image to be rejected.

– 141 –

Page 150: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

B.3 Operations after a Study is Completed

NOTE: Changes in the CALL IMAGES screen are not saved. Therefore, to save the changes, resend the image using image RESEND or study RESEND ALL.

B.3.1 Cancelling a Rejected Image and Outputting AgainThis describes how to play back an image from the study list, cancel the rejection, and then resend the image.

1 Display the study list.

From the exposure screen, touch [SYSTEM] > [STUDY LIST].

Displaying the Study List

2 Display the CALL IMAGES screen.

Touch the required study, and touch [CALL IMAGES].

CALL IMAGES Screen

– 142 –

Page 151: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

3 Display the image whose rejection you want to cancel.

Touch a preview image button.When displaying from the multiview screen, touch [MULTI VIEW].

4 Display the process selection dialog box.

Touch [Reject].When the multiview screen is displayed, directly touch the image to open the window, and then touch [Reject].

Process Selection Dialog Box

5 Cancel the rejection.

Touch [VALIDATE IMAGE].

6 Output the images.

Touch [RESEND] or [RESEND ALL]. The RESEND window or RESEND ALL window is opened. Set the output destination to [ON], and then touch [RESEND] or [RESEND ALL].

Cancelling rejection of an image that has been retakenIn contrast to rejection during exposure, even if rejection of an image that was rejected during retake is cancelled, the other image is not rejected.

– 143 –

Page 152: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

B.3.2 Rejecting an Image and Outputting AgainThis describes how to play back images from the study image list, set them to reject images, and output them again. In this operation, rejected images can be output to temporary internal storage only.

1 Display the study list.

From the exposure screen, touch [SYSTEM] > [STUDY LIST].

Displaying the Study List

2 Display the CALL IMAGES screen.

Touch the required study, and touch [CALL IMAGES].

CALL IMAGES Screen

3 Display the image to be rejected.

Touch a preview image button.When displaying from the multiview screen, touch [MULTI VIEW].

– 144 –

Page 153: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

4 Display the rejection reason input window.

Touch [Reject].When the multiview screen is displayed, directly touch the image to open the window, and then touch [Reject].

Rejection Reason Input Window

5 Select or enter the rejection reason.

Select the rejection reason from the Reject Reason List.If the reason you want to enter is not found in the Reject Reason List, touch the Reject Reason input box to display the keyboard, and then enter the reason.

6 Reject the image.

Touch [OK].

7 Output the images.

Touch [RESEND] or [RESEND ALL]. The RESEND window or RESEND ALL window is opened. Set LOCAL TEMP STORAGE to [ON], and then touch [RESEND] or [RESEND ALL].

– 145 –

Page 154: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

B.3.3 Changing the Rejection Reason and Outputting Again

NOTE: The rejected image is output only temporarily to local temporary storage. Therefore, after changing the rejection reason, be sure to select local temporary storage as the output destination.

1 Display the study list.

From the exposure screen, touch [SYSTEM] > [STUDY LIST].

Displaying the Study List

2 Display the Preview Image Screen.

Touch the image in the study list, and then touch [CALL IMAGES].

Preview Image Screen

3 Display the image whose rejection reason you will change.

Touch a preview image button.When displaying from the multiview screen, touch [MULTI VIEW].

– 146 –

Page 155: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

4 Display the process selection dialog box.

Touch [Reject].When the multiview screen is displayed, directly touch the image to open the window, and then touch [Reject].

Process Selection Dialog Box

5 Display the rejection reason input window.

Touch [CHANGE REJECT REASON].

6 Select or enter the rejection reason.

Select the rejection reason from the Reject Reason List.If the reason you want to enter is not found in the Reject Reason List, touch the Reject Reason input box to display the keyboard, and then enter the reason.

7 Change the rejection reason.

Touch [OK].

8 Output the images.

Touch [RESEND] or [RESEND ALL]. The RESEND window or RESEND ALL window is opened. Set LOCAL TEMP STORAGE to [ON], and then touch [RESEND] or [RESEND ALL].

– 147 –

Page 156: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

B. Using the Rejection Reason Input Function

– 148 –

Page 157: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

C. Using the Sensor Unit Attach/Detach Function

NOTE: This function can be used only with sensor units equipped with an attach/detach function.Be sure to also refer to the user’s manuals for both the sensor unit and power box on the use of this function.

The procedure below is an example of operations when using the category type exposure screen.

C.1 Attach/Detach Function of the Sensor UnitThe sensor unit can be attached/detached while the system is running.The procedure for attaching and detaching the sensor unit is simple.

NOTE: To attach/detach the sensor unit, always turn off the power of the power box first.

NOTE: Do not remove the connector or disconnect the power of the power box while image data is being transferred. Image data that has been taken will be lost.

NOTE: Always attach/detach the sensor unit from the exposure screen. Attaching/detaching the sensor unit while on another screen may result in malfunction or loss of data.

NOTE: If a dialog box is being displayed on the exposure screen, close the dialog box first and then perform attachment/detachment.

NOTE: If the sensor unit is detached during READY status, READY status will be canceled, and the selected exposure mode and parameters set on the parameter changing screen will become disabled. To continue taking exposures, select the exposure mode and make settings again (→ page 15) after connecting the detached sensor unit.

– 149 –

Page 158: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

C. Using the Sensor Unit Attach/Detach Function

C.2 Detaching the Sensor Unit

1 Display the exposure screen.

2 Turn off the power of the power box.

Power off will be automatically detected and a disconnection dialog box will be displayed.

3 Remove the connector.

Disconnect the sensor cable relay connector of the sensor unit.

– 150 –

Page 159: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

C. Using the Sensor Unit Attach/Detach Function

4 Complete detaching the sensor unit.

Touch [OK] on the dialog box.The sensor unit icon will change to indicate the sensor unit is not connected.

– 151 –

Page 160: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

C. Using the Sensor Unit Attach/Detach Function

C.3 Connecting a Detached Sensor Unit

1 Display the exposure screen.

2 Connect the detached sensor unit.

Connect the sensor cable relay connector of the sensor unit.

3 Turn on the power of the power box.

Power on is automatically detected and a connection dialog box is displayed.

– 152 –

Page 161: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

C. Using the Sensor Unit Attach/Detach Function

4 Complete connecting the sensor unit.

Touch [OK] on the dialog box.The sensor unit icon display will change.

If connecting an unregistered sensor unitThe message “Sensor Unregistered Error” is displayed if a sensor unit having serial number that has not already been registered is connected. After confirmation, touch [OK] to close the window and connect a sensor unit having a registered serial number.

– 153 –

Page 162: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

C. Using the Sensor Unit Attach/Detach Function

– 154 –

Page 163: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

D. Turning the Power of the Power Box On/Off

NOTE: This operation is only available with the sensor unit which has a direct Ethernet connection to a PC.Be sure to refer to the user’s manuals for both the sensor unit and the power box.

The procedure below is an example of operations when using the category type exposure screen.

D.1 Using On/Off for the Power of the Power BoxThe power of the power box (or remote switch) can be turned on/off while the system is running.With a mobile system, power consumption can be reduced by turning the power off while the system is being moved.

NOTE: Do not turn off the power of the power box while transferring image data. Image data that has been taken will be lost.

NOTE: Always turn the power of the power box on/off from on the exposure screen.Turning the power of the power box on/off while on another screen may result in malfunction or loss of data.

NOTE: If a dialog box is being displayed on the exposure screen, close the dialog box first and then perform on/off.

NOTE: If the power box is turned off during READY status, READY status will be canceled, and the selected exposure mode and parameters set on the parameter changing screen will become disabled. To continue taking exposures, select the exposure mode and make settings again (→ page 15) after the power is turned on.

– 155 –

Page 164: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

D. Turning the Power of the Power Box On/Off

D.2 Turning the Power of the Power Box On/Off

1 Display the exposure screen.

2 Turn off the power of the power box.

Power off is automatically detected and a disconnection dialog box is displayed.

– 156 –

Page 165: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

D. Turning the Power of the Power Box On/Off

3 Close the disconnection dialog box.

Touch [OK].The sensor unit icon will change to indicate the sensor unit is not connected.With a mobile system, power consumption can be reduced by putting the system in this state while it is being moved.

4 Turn on the power of the power box.

Power on will be detected and a connection dialog box will be displayed.

Continued on the following page >>

– 157 –

Page 166: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

D. Turning the Power of the Power Box On/Off

5 Close the connection dialog box.

Touch [OK].The exposure screen will be displayed.

– 158 –

Page 167: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Index

AA.D. 12Accession No. 11ADJUST ROI 69ADJUST TRIM 68AEC 37ALPHA 11ANNOTATION 83AUTO ADJUST. 70Auto Logout 134AUTO-PRESS 15

BBirth Date 12Brightness 72bucky time 4, 98

CCalibration 97CALL IMAGES 59CANCEL 8Category type Screen 4Cnt of Success Trans 55COLLIMATION MASK 91COLUMN SETTING 54Contrast 72CORRECT PROC. 90CXDI RESTART 6

Ddatabase file 8, 14DELETE 58Density adjustment area 66, 69DESTI. PREF. 27, 36, 38

Eerror window 111Exposure Screen 4

FFixed Annotation 83Free Annotation 85FREQUENCY 76

GGray-scale adjustment 70

HHIS 8, 14

IID 11

LLOCAL CALENDER 12LOCAL TEMP STORAGE 61Login 132Logout 134LUT 70

MMag 92Magnetic card reader 14Magnified image screen 21, 92Marker 82Memo 98Message area 17MULTI VIEW 50Multi-format 49, 50Multi-View Screen 43

NName 11NUM 11Number of exposures 96

OOutput order 46

PPOSITION 81Preview image buttons 4, 37, 59Preview Image Screen 22, 35, 47PRINT SETTING 36, 49, 59Priority 27, 38, 56, 60, 61Process Viewer Screen 31

QQA 65QA Mode Screen 21, 65QA-Mag 92

RRecommended grid 16Reject 26, 36, 40, 45, 47, 61Rejection Reason Input Function 135RESEND 60RESEND ALL Window 56, 61RESET 66

– 159 –

Page 168: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change

Index

RETAKE 23, 25, 36, 41, 42REX 22, 36, 70REX ADJUST. 73RIS 8, 14

SSAVE PARAM 18, 67Self test 103Sex 11SHUTDOWN 5Study description 4, 14STUDY INFO. 13Study List 53

TTimeout for warning 98Transfer 55Tray type Screen 4, 9

UUser Authentication Function 131

VView 81

WWelcome Message Window 3

YYear 12

– 160 –

Page 169: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change
Page 170: Canon CXDI-1 System Digital Radiography Operation …downloads.canon.com/medical/manual_CXDI_V7.2_L-IE-4135I.pdf · Canon CXDI-1 System Digital Radiography ... Buttons that change