bueno tension superficial buenisimo

Upload: kristian-dasa

Post on 05-Apr-2018

223 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    1/15

    MEASUREMENT OF INTERFACIAL TENSIONIN FLUID-FLUID SYSTEMS

    J. DrelichCh. FangC.L. WhiteMichigan Technological University, Houghton, Michigan

    INTRODUCTION

    For more than a century, a variety of techniques have been

    used to measure interfacial tensions between immisci-

    ble fluid phases. A recent monograph by Rusanov and

    Prokhorov (1) provides a broad review of the technicalliterature on the interfacial tension techniques with de-

    tailed discussion of the theoretical bases and instrumenta-

    tion. Additional valuable sources of information on the in-

    terfacial tension measurement methods include selected

    chapters in Refs. 25. In this article, we present a very

    brief overview of the most common techniques used in

    interfacial tension measurements. The reader is encou-

    raged to explore Refs. 15 and references therein for fur-

    ther details.

    This article is organized as follows. Classical Inter-

    facial Tension Measurement Methods reviews the me-

    thods that are used in surface chemistry laboratories. A

    short comparison of these techniques is presented at the

    end of the section. This comparison has been prepared toguide a selection of the experimental method for mea-

    surements of interfacial tension in liquid-fluid systems,

    including systems with surfactants, viscous liquids, or mol-

    ten metals. Many of the industrial operations involve the

    liquid-fluid interfaces, for which the composition is cons-

    tantly refreshed and does not reach equilibrium. The im-

    portance of such dynamic interfacial tensions is increa-

    singly recognized to be essential to the understanding and

    control of interfacial processes in multiphase, multicom-

    ponent systems. Dynamic Interfacial Tension Measure-

    ments discusses a freshly created interface. In Measure-

    ment of Ultralow Interfacial Tension an example of:

    when the value of interfacial tension is significantly less

    than 1 mN/m is discussed. Ultralow interfacial tensions

    are common in the fluid systems of advanced tech-

    nologies of liquid-liquid emulsification processes when

    effective surfactant solutions are used. Finally, in Mic-

    rotensiomery, we discuss the methods of interfacial

    tension measurements that have been applied (or have

    potential to be applied) to microinterfaces of microdrop-

    lets. Fundamental research on the interfacial properties of

    nanomaterials (materials and particles with microstructur-

    al features on the micrometer or nanometer scale) and

    droplets of micrometer-sized or nanometer-sized dimen-sion will be an important challenge in the rapidly deve-

    loping field of nanotechnology.

    CLASSICAL INTERFACIAL TENSION

    MEASUREMENT METHODS

    Fig. 1 shows a classification of common interfacial ten-

    sion measurement methods, both classical and modern.

    Group I represents examples of techniques commonly

    used for direct measure of the interfacial tension with a

    microbalance. The techniques in group II are those in

    which interfacial tension can be determined from direct

    measurement of capillary pressure. Analysis of equilib-

    rium between capillary and gravity forces is employed in

    the techniques of groups III and IV. Group III techniques

    rely on the balance between surface tension forces and a

    variable volume of liquid, whereas Group IV techniques

    fix the volume of a liquid drop and measure the distortion

    of that drop under the influence of gravity. Group V in-

    cludes techniques where the shapes of fluid drops are dis-

    torted by centrifugal forces and are used to measure ultra-

    low interfacial tensions.

    Group I: Direct Measurement

    Using a Microbalance

    Interfacial tension at fluid-fluid interfaces is a reflection

    of the excess energy associated with unsaturated inter-

    molecular interactions at the interface. This excess energy

    tends to drive interfaces to adopt geometries that mini-

    mize the interfacial area, and this tendency can be inter-

    preted as a physical force per unit length (i.e., a tension)

    3152 Encyclopedia of Surface and Colloid ScienceCopyright D 2002 by Marcel Dekker, Inc. All rights reserved.

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    2/15

    applied in the plane of the interface. The excess energy

    per unit area (E/A) is numerically equal to this force per

    unit length (F/L), which is numerically equal to the in-

    terfacial tension (g).

    To directly measure interfacial tensions using a mic-

    robalance, a plate, ring, rod, or other probe of simple shapeis brought into contact with the interface. If the probe is

    completely wetted by one of the liquids, this liquid will

    adhere to the probe and climb as the result of capillary

    force, increasing the interfacial area and leading to a force

    tending to pull the probe toward the plane of the interface

    (Figs. 2 and 3). This restoring force is directly related to

    the interfacial tension and can be measured by a micro-balance. The force (F) acting along the three-phase contact

    Fig. 1 Classification of techniques for interfacial tension measurements that are discussed in this article.

    Fig. 2 A schematic of the Wilhelmy plate method. Fig. 3 Illustration of the ring method.

    M

    Measurement of Interfacial Tension in Fluid-Fluid System 3153

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    3/15

    line is exactly equal to the weight of the liquid meniscus

    standing above the plane of the fluid-fluid interface. This

    force, measured by the microbalance, is used to calculate

    the interfacial tension:

    g Fp cosy

    1

    where p is the perimeter of the three-phase contact line and

    y is the contact angle measured for the liquid meniscus in

    contact with the object surface.

    The two principal techniques used for direct measure-

    ment of interfacial tension using the microbalance are

    Wilhelmy plate and du Nouy ring methods. The Wilhelmy

    plate technique is used in both static and detachment

    modes, whereas du Nouy ring technique is strictly a de-

    tachment technique. In the static measurement, the plate

    remains in contact with liquid during the entire cycle ofinterfacial tension measurement. If the instrument ope-

    rates in the detachment mode, the interfacial tension is

    measured by measuring the force required to separate the

    ring or plate from contact with the interface.

    Wilhelmy plate technique

    A vertical thin plate is used in this technique (Fig. 2) (6).

    The commercial plates are made of roughened platinum-

    iridium alloy or platinum. The metal plate must be cleaned

    from organic contaminants by an organic solvent and then

    flamed before the experiment. Both roughening and clean-

    ing of the plate surface are used to maintain good wetting

    of the plate by the test liquid. It should be noted that

    materials other than platinum or platinum-iridium alloy,

    such as glass, mica, and steel (1, 5) have also been used.

    Nevertheless, good wetting of the test liquid to the plate is

    always necessary. The use of plates made of material other

    than metal is a must requirement in the case of certain

    liquids such as during the measurements of the interfacial

    tension between a heavy nonpolar liquid (i.e., carbon tet-

    rachloride) and immisible, but lighter, polar liquid (i.e.,

    water). For such systems, the plate should be hydrophobic.

    Several polymers, especially fluorinated polymers, can be

    used for this purpose. Adsorption and self-assembling of

    organic amines on the surface of the platinum plate could

    also be a solution to this problem.

    In the Wilhelmy plate method, the plate is put in a fixedposition relative to the horizontal surface of the liquid

    (Fig. 2). Then, the force (F) vertically acting on the plate

    by the liquid meniscus is measured by using a microba-

    lance. The force applied to the plate is equal to the weight

    of the liquid meniscus uplifted over the horizontal surface.

    By measuring this force, the interfacial tension can be

    calculated by using Eq. 1 where p = 2(L+ t). Modern

    instruments use plates of standard dimensions so that

    measurements of the plate size and its weight are not re-

    quired. Adsorption of organic compounds from the labo-

    ratory environment or test solutions can be a major source

    of experimental error when measuring surface tensions

    using the Wilhelmy plate method.

    Du Nouy ring method

    In this method, the interfacial tension relates to the force

    required to pull a wire ring off the interface (Fig. 3) (7, 8).

    As in the case of Wilhelmy plate, the ring is usually made

    up of platinum or platinum-iridium alloy of a radius (R)

    of 2 3 cm. The radius (r) of the wire ranges from 1/30 to

    1/60 of that of the ring (9).

    Again, Eq. 1 describes in general the calculation pro-

    cedure of the technique. Here, the perimeter (p) of the

    three-phase contact line is equal to twice the circumfe-

    rence of the ring: p = 4pR. Because additional volume ofliquid is lifted during the detachment of the ring from the

    interface, a correction factor ( f) is required in Eq. 1 (8):

    g Fp cos y

    f 2

    The correction factor varies from about 0.75 to 1.05 and

    depends on the dimensions of the ring (R, r), its surface

    wettability (y), and difference in fluid density (Dr). The

    tabulated f values in relation to R/r (for y = 0) can be

    found in Ref. 8, and also calculated from the following

    approximate equation (10):

    f 0:7259:075

    10

    4F

    p3DrgR3 1:679r

    R 0:04534 1=2

    3

    The application range of Eq. 3 is: 0.045 DrgR3/F 7.5. The maximum force is measured by the microbalance

    (F) and corresponds to detachment of the ring from the

    interface. The F value is measured experimentally and

    then Eq. 3 is used to calculate the correction factor f. The

    interfacial tension is then calculated from Eq. 2. The in-

    terfacial tension reading made by modern computerized

    instrumentation does not require separate calculation of f,

    since its calculation is incorporated in the software.

    The high-accuracy measurements with the ring method

    require that the plane of the ring remain parallel to theinterface. The major error in this technique is caused by

    deformation of the ring, which is a very delicate probe and

    subject to inadvertent deformation during handling and

    cleaning. It is also important that perfect wettability of the

    ring surface by the denser fluid be maintained (y = 0). If

    perfect wetting is not achieved, additional correction of the

    instrument reading is needed. Poor wetting of ring by the

    3154 Measurement of Interfacial Tension in Fluid-Fluid System

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    4/15

    denser fluid makes the measurement of interfacial tension

    impossible to carry out. In the case of special measure-

    ments requiring homemade rings, very large rings should

    be avoided to avoid the small value of the correction factor

    (see Eq. 3). If all of the necessary experimental precau-

    tions are observed, this method can guarantee higher ac-

    curacy than any other detachment method.

    Group II: Measurement of Capillary Pressure

    Interfacial tension is defined as the work required to create

    a unit area of interface at a constant temperature, pressure,

    and chemical potential. Because it is always positive for

    interfaces between immiscible phases, interfacial tension

    always tends to decrease the area of interface. This ten-

    dency gives rise to a pressure difference between fluids on

    either side of a curved interface, with the higher pressure

    on the concave side of the interface. This pressure dif-

    ference results in phenomena such as a capillary rise,

    bubble and drop formation, etc. A formula describing the

    pressure difference (DP) across the curved interface is

    known as the Young-Laplace equation (11, 12):

    DP g 1R1

    1R2

    4

    where R1 and R2 are the radii of curvature.

    The pressure difference across a curved interface (DP)

    can be measured in a number of ways (e.g., using a pres-

    sure sensor or observing a capillary rise) and then be used

    to calculate g if the radii of curvature are known. The most

    common and probably one of the oldest methods in this

    group of interfacial tension measurement techniques is a

    maximum bubble pressure method that is briefly described

    in the next paragraph. Modification of the maximum bub-

    ble pressure method based on a continuous measurement

    of varying pressure during growing bubble or drop is now

    a basic technique in examination of dynamic (not equi-

    librated) interfacial tension and is further discussed in

    Dynamic Interfacial Tension Measurements.

    Maximum bubble pressure

    This method is based on measuring the maximum pressure

    (p) to force a gas bubble out of a capillary into a liquid

    (Fig. 4) (13, 14). The measured pressure is the sum of

    capillary pressure (DP) caused by the interfacial tension

    and the hydrostatic pressure (rAghA) caused by the liquid

    column above the orifice of the capillary:

    DP p rAghA 5This pressure can be expressed as the height (h) of thecolumn of an imaginary liquid of density (Dr = rArB):

    h DPDrg

    6

    Sugden (13) derived an expression to relate h with the

    Laplace capillary constant a = 2g/(Drg) and the bubble

    meniscus:

    r

    X r

    b r

    a

    zcb

    b2

    1=27

    where X = a2/h, b = 2b2/a2, zc is the height of the bub-

    ble, and b is the curvature radius at the apex (lowest point

    of the bubble). Then he tabulated the minimum values of

    Fig. 4 Maximum bubble pressure method. (A) A sequence illustrating the shape of bubble at three different stages of bubble growth.

    (B) Relationship between pressure inside the bubble and radius of the bubble.

    M

    Measurement of Interfacial Tension in Fluid-Fluid System 3155

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    5/15

    X/r as dependent of a given value of r/a within the range

    0 < r/a 1.5. Using this table, the surface tension can becalculated by following an iteration procedure.

    Direct and easier, but a little less accurate, calculation

    of the interfacial tension can be done by using the fol-

    lowing equation (5):

    g DPr2

    1 2rDrg3DP

    rDrg2

    6DP2

    !8

    As further discussed in the next section; the maximum

    bubble and drop pressure technique or its modifications

    have been very useful in studying the dynamic interfacial

    tensions. This technique has also been attractive to exa-

    mination of surface tension for molten metals (2).

    Group III: Analysis of the Balance Between

    Capillary and Gravity Forces

    Methods based on analysis of capillary effects, other than

    the shape of a drop or meniscus, such as capillary rise and

    drop volume or weight, are among the oldest surface ten-

    sion measurement methods in use. A variety of modern

    instruments, usually fully automated and computerized

    (groups I, II, and IV), have replaced these methods in most

    laboratories. We provide a short review of two techniques

    that still might be attractive to researchers who have limi-

    ted access to modern instrumentation.

    Capillary rise method

    The basis for the capillary rise method is to measure the

    height h of the meniscus in a round glass tube having the

    known inner radius r, as shown in Fig. 5 (2, 15). For

    small-diameter tubes (i.e., r < < h) the shape of the me-

    niscus is spherical, and the surface tension can be calcu-

    lated by using the following equation:

    g Drghr2cosy

    9

    Since the glass tubes are easy to clean with acids, bases,

    and organic solvents, and because many of the liquids

    perfectly wet the glass surface, the cosy term in the above

    equation will often equal unity.

    If the shape of the meniscus is not spherical, Eq. 9

    should be replaced with (15):

    g 12Drgrh 1 r

    3h 0:1288 r2

    h2 0:1312 r3

    h3

    10

    The capillary rise method can be one of the most ac-

    curate techniques used to make surface tension measure-

    ments. Technical problems with the technique are related

    to fabrication of a uniform bare capillary tube and precise

    determination of its inside diameter. In addition, the ca-

    pillary rise method is not very convenient for measuring

    the interfacial tension between two liquids.

    Drop volume or weight

    In this method, the weight or volume of a drop falling froma capillary with a radius r is measured (Fig. 6) (16, 17).

    The weight (W) of the drop falling off the capillary cor-

    relates with the interfacial tension through the following

    equation (5):

    W VDrg 2prgf rffiffiffiffiV

    3p

    11

    Fig. 5 Illustration of the capillary rise method.

    Fig. 6 Schematic illustration of drop volume or weight method.

    3156 Measurement of Interfacial Tension in Fluid-Fluid System

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    6/15

    where Vis the drop volume, ris the radius of the capillary,

    and f is the correction factor required because only a

    portion of the drop volume is released from the capillary

    during detachment (2). The correction factor is a function

    of r/V1/3, and this correlation was empirically determined

    and tabulated by Harkins and Brown (17). It can also be

    calculated from the empirical function (5):

    frffiffiffiffiV

    3p

    0:167 0:193 rffiffiffiffi

    V3p

    :0489 rffiffiffiffi

    V3p

    2

    0:0496 rffiffiffiffiV

    3p

    312

    Because of small volume of each drop, many drops need

    to be collected for the accurate measurement of drop

    weight or volume. In modern instrumentation, the volume

    of liquid and the number of droplets released from thecapillary can be determined very precisely and thus the

    weight or volume of the individual drop is not difficult to

    calculate (18).

    Capillaries used in the drop weight or volume tech-

    niques are usually made of glass; however, metal capil-

    laries are also used on occasion (1). Glass is wetted by

    many liquids, is transparent, and is relatively easy to clean.

    Capillary tubes specifically fabricated for routine inter-

    facial tension measurements are now difficult to purchase

    in the U.S. market; however, glass capillaries can be pro-

    duced relatively easily in glass workshops.

    The measurements of interfacial tension with the drop

    weight or volume technique are very simple but, unfortu-

    nately, sensitive to vibrations on the other side. Vibrations

    of the apparatus can cause premature separation of the drop

    from the end of the capillary before the drop reaches the

    critical size. In addition, the measurements in multicom-

    ponent solutions when adsorption occurs might not reflect

    equilibrium saturation of the solutes at the interface.

    Group IV: Analysis of

    Gravity-Distorted Drops

    Interfacial tension causes interfaces to behave as elastic

    membranes that always tend to compress the liquid. In the

    absence of other forces (e.g., in zero gravity), the liquid

    surface has a natural tendency to form spherical shapes to

    minimize the interfacial area per unit volume of liquid and

    thus, to minimize the excess energy of the interface. The

    shape of an interface in a gravitational field (Fig. 7) de-

    pends on the competition between the capillary and gra-

    vitational forces and can be described by the Bashforth-

    Adams equation (19):

    gsinf

    x 1

    R1

    2g

    b Drgz 13

    Eq. 13 is often expressed in a dimensionless form as:

    sinf

    x=b 1

    R1=b 2 Drgb

    2

    g

    z

    b14

    where g is the interfacial tension; Dr = AB equalsthe difference in density of fluids; R1 is the radius of cur-

    vature; x is the radius of rotation of point S around the z

    axis; f is the angle ofR2 vector with the axis of symmetry;

    b is the radius of curvature at the apex of the curvature; and

    g is the acceleration due to gravity. Fig. 7 shows the de-

    tails of drop geometry.

    The techniques of curved interface shape analysis are

    particularly attractive to researchers because they do not

    require advanced instrumentation. The experimental setup

    requires a camera with a low-magnification lens to record

    the shape of the drop. The interfacial tension can be easily

    calculated from the dimensions of the pendant drop, ses-

    sile drop, or liquid meniscus taken from the photographic

    picture and by using numerical solutions to the above

    equations. Modern instruments, however, use image anal-

    ysis software whose role is to match the entire drop profile

    to the best fit of the theoretical curve (e.g., the Bashforth-

    Adams equation) describing the shape of the drop (14).These advances significantly improved the precision of

    the techniques and reduced the time of the measurement,

    providing an opportunity for examination of the interface

    aging process. Probably the most advanced software, axi-

    symmetric drop shape analysis, was introduced by Neu-

    mann and co-workers (20). Since advanced instrumen-

    tation is not always available to researcher, a brief reviewFig. 7 Definition of dimensions and coordinates describing the

    sessile drop.

    M

    Measurement of Interfacial Tension in Fluid-Fluid System 3157

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    7/15

    of classical approaches of interfacial tension determination

    from the shape of the interface seems to be appropriate.

    Pendant drop method

    In a simple method, two parameters of the pendant drop

    that should be experimentally determined are the equa-

    torial diameter D and the diameter d at the distance D

    from the top of the drop (Fig. 8) (21, 22). The interfacial

    tension is then calculated from the following equation (2,

    21, 22):

    g DrgD2

    H15

    The shape dependent parameter (H) depends on a value of

    the shape factor S = d/D. Tables including the set of

    1/Hvs. Svalues are available in several references (1, 22).

    The values of 1/H can also be calculated from the fol-

    lowing empirical formula (23):

    1

    H B4

    Sa B3S3 B2S2 B1S B0 16

    where Bi (i = 0, 1, 2, 3, 4) and a are empirical constants

    for a certain range of S, which are shown in Table 1.

    The pendant drop technique, as other interfacial tension

    measurement techniques, requires extreme cleanliness to

    obtain good quality and reproducible results. Here, the

    needle used for hanging the drop should be well cleaned

    and the climbing of the interface over the outer surface of

    the needle should be avoided. Needles made of stainless

    steel or glass that are relatively easy to clean with acids,

    bases, and organic solvents are most often used in surface

    chemistry laboratories. It is recommended that needles

    with a diameter that is less than 0.5 D be used (21). The

    diameter of the needle should not be too small, however,

    because this reduces the value of dand, consequently, the

    precision of interfacial tension determination.

    Sessile drop method

    This method is based on the analysis of the profile of the

    drop sitting on a solid substrate, as shown in Fig. 7 (24,

    25). It is recommended that substrates used in sessile dropmeasurements be poorly wetted by the drop, i.e., they

    should have a contact angle larger than 90 degrees. In a

    simple experimental approach, one first needs to locate

    the equator of the drop, and then measure the height from

    the top of the drop to its equator (ze). For a very large

    sessile drop, an analytical expression for the interfacial

    tension is as follows (5):

    g Drgz2

    e

    217

    From a practical point of view, it is often difficult to pre-

    cisely locate the equator of the drop and measure ze for

    many drops. Although the large drop is almost flat, loc-

    ating the top of the drop is sometimes experimentally

    difficult. It should be recognized, however, that large

    drops are not required if tabulated dependencies of drop

    shape parameters, based on the Bashford-Adams analysis

    (Eq. 14) are used (19, 25).

    Practical Comments

    Most of the techniques reviewed in this section have been

    commercialized. Table 2 summarizes the accuracy, com-

    Fig. 8 Pendant drop.

    Table 1 Empirical constants for Eq. 16

    Range of S A B4 B3 B2 B1 B0

    0.401 0.46 2.56651 0.32720 0 0.97553 0.84059 0.18069

    0.46 0.59 2.59725 0.31968 0 0.46898 0.50059 0.13261

    0.59 0.68 2.62435 0.31522 0 0.11714 0.15756 0.05285

    0.68 0.90 2.64267 0.31345 0 0.09155 0.14701 0.05877

    0.90 1.00 2.84636 0.30715 0.69116 1.08315 0.18341 0.20970

    3158 Measurement of Interfacial Tension in Fluid-Fluid System

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    8/15

    mercial availability, and suitability of these techniques forvarious types of fluid-fluid systems. The accuracy of most

    of these techniques for pure liquid-gas systems is about

    0.1 mN/m. The capillary rise technique is capable of

    significantly better accuracy than the others in Table 2.

    Most of the interfacial tension measurement techniques

    listed in Table 2, including the capillary rise method, have

    been successfully applied to liquid-liquid systems. Re-

    duced accuracy of the detachment du Nouy ring method

    is associated with difficulties in calibrating the weight of

    the ring immersed in the less dense liquid. The same

    problem can be expected in the interfacial tension mea-

    surements using the Wilhelmy plate instrument. All tech-

    niques in Table 2 yield reduced accuracy when applied to

    liquid-liquid interfaces or when one or both of the liquids

    is viscous.

    The measurements with viscous liquids are always dif-

    ficult to carry out due to problems with handling the

    liquid, injection of a liquid sample of the required volume

    into the instrument, lowvelocity liquid flow, and long

    time viscous effects during deformation of the interface.

    Two techniques are particularly recommended to examine

    the surface tension of viscous liquids: the Wilhelmy plate

    (not a detachment option!) and the sessile drop methods.

    In both techniques, samples can be equilibrated for se-

    veral hours before the measurements are taken.

    It is important to recognize that any interfacial ten-

    sion measurement can be strongly influenced by inter-facially active solutes or impurities that are accidentally

    introduced into the fluid-fluid system or present on solid

    surfaces that act as part of the measurement system (1,

    2). Any solid surfaces that make contact with liquids

    (e.g., plates, rings, and capillary tubes) must be careful-

    ly cleaned prior to making measurements. Furthermore,

    some interfacially active contaminants can be introduced

    from the skin or breath of laboratory workers. Finally, it isimportant to note that interfacial tensions are influenced

    by temperature, which should be controlled and reported

    for all measurements.

    When using techniques that depend on a known wet-

    tability of a solid probe by one of the liquids, surface active

    solutes (whether present intentionally or as impurities) can

    seriously influence the interfacial tension measurements.

    Interfacially active solutes in a liquid phase can adsorb not

    only on the fluid-fluid interfaces but on the liquid-solid

    interfaces as well. This adsorption will affect wettability

    of the solid surface (the cosy term in Eqs. 1 and 2) and,

    therefore, influence the measured result. In principle, such

    effects can be eliminated by employing solid probes of

    alternate materials to which the solute does not adsorb, but

    plates and rings are not usually available in a wide range

    of alternative materials.

    Even when adsorption to solid-liquid interfaces is not a

    problem, it is important to allow fluid-fluid interfaces to

    achieve equilibrium before making a measurement. When

    equilibrium is achieved rapidly (within a few seconds),

    the drop volume technique may be suitable. If equilibra-

    tion requires longer times, sessile drop, Wilhelmy plate,

    or other quasi-static techniques may be more appropriate.

    This consideration applies to any fluid-fluid system in

    which kinetically limited processes (adsorption, viscous

    flow, etc.) take place.

    Because of the relatively high temperatures involvedand their reactivity with many gases and solids, surface

    tension measurements on liquid metals pose special chal-

    lenges. The four principal techniques that have been em-

    ployed are the maximum bubble pressure method, the

    sessile drop method, the drop volume or weight method,

    and the pendant drop method (26, 27). It is important that

    measurements on liquid metals be carried out in inert

    Table 2 Accuracy and suitability of classic techniques used in interfacial tension measurements

    Method

    Accuracy

    mN=m

    Suitability

    for surfactant

    solutions

    Suitability

    for two-liquid

    systems

    Suitability for

    viscous liquids

    Suitability for

    melted metals

    Commercial

    availability

    Wilhelmy plate $0.1 Limited Good Very good Not recommended YesDu Nouy ring $0.1 Limited Reduced accuracy Not recommended Not recommended YesMaximum bubble 0.1 0.3 Very good Very good Not recommended Yes Yes

    pressure

    Capillary rise

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    9/15

    gas environment to avoid reactions with gases and other

    phases. Oxygen and other reactive gases are known to exert

    strong effects on the surface tension of selected metals,

    even when present in parts per million concentration (27).

    DYNAMIC INTERFACIAL

    TENSION MEASUREMENTS

    In fluid-fluid systems containing interfacially active so-

    lutes, a freshly created interface will not generally be in

    compositional equilibrium with the two immiscible fluids

    it separates. It is only after solute redistribution from one

    or both phases (i.e., adsorption) has occurred that this

    interface will achieve its equilibrium state. It is sometimes

    important to measure the interfacial tension of freshly

    created interfaces, and such measurements yield what is

    known as dynamic surface tension. A detailed reviewof experimental techniques, theoretical background, and

    literature on the measurements of dynamic interfacial ten-

    sions was recently published by Dukhin et al. (3). Another

    valuable source of analysis of adsorption at the interface

    and dynamic interfacial tension is the book published by

    Joos (28).

    Table 3 provides a characteristic time range for the

    selected interfacial tension measurement techniques. Of

    different techniques already discussed in this article, the

    capillary rise method is not recommended for dynamic

    interfacial tension measurements. The techniques dis-

    cussed in the next sections are not very suitable for exa-

    mination of dynamic effects at interfaces either.

    Interfacial tension changes that occur over time in-

    tervals of at least several seconds (and continue over se-

    veral minutes, hours, or days) can be studied by using

    most of the classical techniques discussed in the previous

    section. For example, Fig. 9 shows the results of inter-

    facial tension relaxation between bitumen and water of

    varying pH value recorded with the Wilhelmy plate ins-

    trument (30). In this bitumen-water system, the dynamic

    character of the interfacial tension is caused by diffusion

    of natural surfactants from the bitumen to interface and

    aqueous phase, and surfactant reaction with ions dissolvedin water (31).

    As emphasized in the previous section, examination of

    the interfacial tension for surfactant solutions using clas-

    sical techniques should be carried out with caution. Sur-

    factants often adsorb on the solid surfaces of equipment

    used in measurements and change the wetting character-

    Table 3 Characteristic time range for common interfacial tension measurement techniques

    Method Time rangea Comments

    Wilhelmy plate >10 s Some of the surfactants might alter the wetting propertiesof the plate, causing the change of measurement

    conditions (possible source of error)

    Du Nouy ring >30 s Same as above

    Pendant drop >10 s Strongly surface active chemicals might cause the release

    of pending drop before completion of the measurement

    Sessile drop >10 s Some of the surfactants might alter the wetting properties

    of a solid support substantially changing the shape of

    the sessile drop

    Drop volume/weight 1 s 20 min Hydrodynamic effects associated with releasing liquid

    volume and circulation of liquid inside the drop

    sometimes significantly reduce the accuracy of the

    interfacial tension measurements

    Maximum bubble pressure 1 ms 100 s Difficulties with determination of the real surface age and

    problems with hydrodynamic effects in the vicinity

    of interfaceGrowing drop/bubble >10 msb Not available commercially

    Oscillating jetc 1 10 ms Not available commercially

    Pulsating bubblec 5 ms 0.2 s Not available commercially

    aBased on Dukhin et al. (3).bIt is claimed by MacLeod and Radke (29) that the dynamics interfacial tension can be measured for several hours, although Ref. 3 specifies the upper

    limit as 600 s.cMethods not discussed in this article.

    3160 Measurement of Interfacial Tension in Fluid-Fluid System

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    10/15

    istics of a solid surface. This effect usually causes expe-

    rimental problems in the Wilhelmy plate, du Nouy ring,

    and sessile drop techniques (see comments in Table 3).

    Short-time interfacial tension and wetting effects play

    important roles in highvolume industrial processes such

    as froth flotation of particles and droplets, detergency,

    foam or froth generation, and stability (3). In these pro-

    cesses, dynamic interfacial tensions become more crucial

    to the success of the technology than the equilibrium (or

    near-equilibrium) interfacial tension. This issue has been

    emphasized in the literature (3, 32, 33), but straightforward

    links between dynamic interfacial tensions and (e.g., pro-

    cess efficiencies) have not yet been well established. This

    research area is just evolving, and the continued funda-

    mental research will probably establish a better connection

    of dynamic interfacial phenomena with practical needs.

    Four basic techniques for measurements of the dynamic

    interfacial tension at short intervals include the maximum

    bubble pressure, growing drop (bubble), oscillating jet,

    and pulsating bubble methods (Table 3). Neither the os-

    cillating jet technique nor the pulsating bubble technique

    is discussed in this article. Bases of these techniques are

    provided in Refs. 13 and references therein.

    The maximum bubble (drop) pressure method and its

    modifications have been the most popular techniques used

    in research conducted in recent years. Apparatus for ma-

    king these measurements are also available commercially.

    The maximum bubble pressure technique was briefly des-

    cribed in Classical Interfacial Tension Measurement Me-

    thods. Following is a short description of a modification

    of the maximum bubble pressure technique that is suitablefor dynamic surface tension measurements.

    Growing Drop (Bubble) Method

    Modern instrumentation permits the pressure inside a bub-

    ble or drop to be precisely and continuously measured as it

    forms and detaches from the end of a capillary (3, 29, 34,

    35). The geometry of a drop or bubble can also be mo-

    nitored during growth and detachment by using advanced

    videographic equipment. This ability to simultaneously

    monitor both pressure and geometry (size and shape) of

    bubbles or drops as they form allows dynamic interfacial

    tensions to be evaluated over a range of growth rates.

    Furthermore, the same experimental approach allows mea-

    surement of surface tensions using approaches normally

    applied to systems in (or near) equilibrium, such as the

    drop volume and maximum pressure drop techniques.

    Fig. 10 describes an experimental approach used by

    MacLeod and Radke (29) for measurement of dynamic

    interfacial tension using the growing drop technique. In

    Fig. 9 Dynamic interfacial tension measured between bitumen

    and water at 60C using the Wilhelmy plate technique. The pH

    values for water at the beginning (t = 0) and at the end of the

    measurements (t = 90 min) are listed in the figure legend. (The

    results are from Ref. 30.)

    Fig. 10 Schematic of the growing drop apparatus used by MacLeod and Radke. (From Ref. 29.)

    M

    Measurement of Interfacial Tension in Fluid-Fluid System 3161

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    11/15

    this apparatus, a liquid drop or gas bubble is formed and

    released from a capillary by using a precise micropump to

    carefully control the growth rate of the drop or bubble. A

    pressure transducer is used to simultaneously monitor and

    record the internal pressure in the drop or bubble, while

    its size and shape are recorded by using a video camera.

    These experiments can be carried out for a range of flow

    rates ranging from near equilibrium growth (very low

    flow rates) to highly nonequilibrium growth conditions

    (very rapid bubble or drop growth).

    Fig. 11 shows selected results of MacLeod and Radke

    (29) for growth of 0.25 mM aqueous decanol droplets for

    a range of droplet growth rates between 5 and 100 mm3/

    min. The plots of interfacial tension vs. time show ini-

    tially increasing, reaching a maximum, and then decrea-

    sing as a function of time. The positive slope of the g vs. t

    curves for t < 1 results from the depletion of decanol

    adsorption due to rapid expansion (stretching) of theinterface when the bubble is small (see Fig. 4A). As the

    drop geometry proceeds beyond hemispherical shape

    (Fig. 4C), the relative rate of surface area growth decrea-

    ses, allowing decanol from the bulk liquid to replenish the

    surface adsorption. The decrease in interfacial tension as-

    sociated with this increase in adsorption yields the negative

    slope observed in the g vs. tcurves for t > 1. As expected,

    the largest dynamic interfacial tensions, approaching those

    for pure water, are observed for droplets formed at the

    highest capillary flow rates. Conversely, the lowest values

    (approaching the equilibrium interfacial tension for this

    solution) were observed for droplets growing at very slow

    rates. At drop formation times approaching 100 s, inter-

    facial tensions for drops formed at all flow rates approach

    the equilibrium value for the 0.25 mM aqueous 1-decanol

    solution.

    The zero time for each measurement in Fig. 11 cor-

    responds to the moment of detachment for the previous

    drop. Data points for a given capillary flow rate in Fig. 11

    start at the time where the drop reaches a hemispherical

    shape (Fig. 4B), which also corresponds to the point at

    which the interfacial tension can be evaluated by using the

    maximum pressure method. Interfacial tension measure-

    ments using the drop-volume method are, of course, de-

    termined at the point where the drop detaches from the

    capillary and correspond to the last data point in thecorresponding curve for the growing drop method. These

    two techniques, therefore, provide snapshots of the in-

    terfacial tension values in dynamic systems, and they

    bracket the family of curves determined by the growing

    drop method. The growing drop method provides the very

    important advantage of continuously recording the inter-

    facial tension throughout the drop formation and allows

    the competing kinetic effects of interfacial stretching and

    solute transport for adsorption to be explored.

    MEASUREMENT OF ULTRALOW

    INTERFACIAL TENSION

    Recovery of petroleum using tertiary oil recovery tech-

    nology; cleaning of solid surfaces from dirt, grease, and

    oil; formulation of stable emulsions; in situ remediation

    of oil-contaminated soil with surfactant solutions; and

    other applications often rely on lowering the interfacial

    tension between immiscible liquids to ultralow values

    (much less than 1 mN/m) using surfactants. The measure-

    ments of such low interfacial tensions are extremely dif-

    ficult to perform with classical interfacial tension mea-

    surement methods reviewed in Classical Interfacial

    Tension Measurement Methods or the dynamic tech-

    niques discussed in the previous section (e.g., see the ac-

    curacy values for methods shown in Table 2). For themeasurements of ultralow interfacial tensions, the spin-

    ning drop technique has been developed at both laboratory

    and commercial scales (3639). The basis of this tech-

    nique is discussed in the following paragraph. An ad-

    ditional method designed for measurements of ultralow

    interfacial tensions was proposed by Lucassen (40) and

    is based on the analysis of the shape of the drop suspended

    Fig. 11 Dynamic surface tension of 0.25 mM aqueous decanol

    solution droplets growing in air at 23C. Based on the experi-

    mental data presented by MacLeod and Radke. (From Ref. 29.)

    3162 Measurement of Interfacial Tension in Fluid-Fluid System

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    12/15

    in liquid with a density gradient. The need for a strict

    control of liquid density limits this latter technique to rela-

    tively few applications.

    Spinning Drop Technique

    This technique relies on the fact that gravitational ac-

    celeration has little effect on the shape of a fluid drop

    suspended in a liquid, when drop and the liquid are con-

    tained in a horizontal tube spun about its longitudinal axis(Fig. 12) (36, 37). At low rotational velocities (o), the

    fluid drop will take on an ellipsoidal shape, but when o is

    sufficiently large, it will become cylindrical. Under this

    latter condition, the radius (r) of the cylindrical drop is

    determined by the interfacial tension, the density diffe-

    rence (Dr) between the drop and the surrounding fluid,

    and the rotational velocity of the drop. As the result, the

    interfacial tension is calculated from the following equa-

    tion (38):

    g 14

    r3Dro2 18

    The spinning drop method has been very successful inexamination of ultralow interfacial tensions down to 106

    mN/m (39). For example, Fig. 13 shows the interfacial

    tension values for octane drops suspended in an aqueous

    phase saturated with ethoxylated alcohols (41). As shown

    in Fig. 13, the interfacial tension for the octane-water-

    CnE4 system varied from about 1 to 104 mN/m and de-

    pended on both temperature in the system and length of

    the hydrocarbon chain in the chemical structure of etho-

    xylated alcohol (Ref. 41 provides additional examples).

    The minimum in the interfacial tension vs. the temperature

    curve coincides with the phase inversion temperature, at

    which both hydrophilic and oleophilic natures of surfac-

    tant are in balance.

    MICROTENSIOMETRY

    Criminology, biology, and pharmaceutical processing are

    among a number of fields in which material quantities of

    interest may be too small to apply conventional tensio-

    metric techniques. Furthermore, the developing field of

    nanotechnology promises to yield novel structures on a

    nanometer scale, the interfaces of which are sure to be the

    subject of considerable study. The study of interfaces on

    the nanometer scale can be very important because they

    may exhibit properties that are very different from their

    macroscopic counterparts. For example, knowledge of

    partition and adsorption of surfactants and other interfa-

    cially active components of two-liquid systems is often

    crucial for the control of emulsion formulation and sta-

    bility. It has recently been demonstrated that interfacial

    tension of microscopic droplets may differ significantly

    from the interfacial tension of macroscopic drops for the

    same surfactant solutions (42). This difference arises from

    dissimilar partitioning of surfactants into two immiscible

    liquids and is due to the effect of an enlarged interfacial

    area. In this regard, it becomes critical to conduct the study

    of interfacial phenomena, including the measurements ofinterfacial tension, on a length scale that is characteristic

    of the individual elements of the emulsion.

    Microtensiometry is the study of interfaces on very

    small particles and in finely dispersed systems. The micro-

    pipette technique, discussed first, is closely related to tech-

    niques depending on pressure differentials across curved

    interfaces that have already been presented. Techniques

    based on atomic force microscopy (AFM) offer the pos-

    sibility of imaging nanometer-sized particles and directly

    measuring their interactions. AFM-based techniques are

    relatively new and still very much under development.

    Fig. 12 Schematic of the rotating drop method.

    Fig. 13 The effect of temperature on interfacial tension mea-

    sured between octane and aqueous phase saturated with CnE4.

    Graph based on the experimental data presented in Ref. 41.

    M

    Measurement of Interfacial Tension in Fluid-Fluid System 3163

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    13/15

    Micropipette Technique (4345)

    The micropipette technique was recently developed to

    directly measure interfacial tensions of micrometer-sized

    droplets. The technique was first used in examination of

    vesicles (43, 44) and next liquid droplets (42, 45). In this

    technique, the droplet is first captured at the tip of the glassmicropipette and then sucked into the pipette (Fig. 14A).

    The interfacial tension is calculated from the minimum

    pressure, at which the droplet extends a hemispherical pro-

    trusion into the pipette, and by using the Laplace Eq. 4

    in the following form (42):

    Dp 2g 1Rp

    1Ro

    19

    where Rp is the inner radius of the pipette and Ro is the

    radius of the exterior segment of the droplet; the dimen-

    sions of the pipettes internal diameter must be smaller

    than the diameter of the droplet.In the conventional technique shown in Fig. 14A, the

    large pressure difference required to draw the droplet into

    the pipette when the droplet does wet or adhere to the pi-

    pette surfaces, is a limitation. To avoid this limitation, a

    two-pipette technique, with a separation force applied bet-

    ween the pipettes to deform the droplet, has been used as

    shown in Fig. 14B (45). In the two-pipette technique, the

    separation force between the pipettes must also be mea-

    sured, and the interfacial tension is calculated from the

    forcedrop deformation relation. Table 4 presents the re-sults of interfacial tension measurements for waterorga-

    nic liquid systems determined with this technique.

    Atomic Force Microscopy

    The application of AFM permits roughness, heterogeneity,

    and interaction forces to be studied at submicroscopic

    scales that may extend down to molecular sizes. Among

    the most popular applications of the AFM are studies of

    interactions between substrates and colloidal particles. The

    AFM has not been established as a technique to measure

    interfacial tensions directly: however, it appears to have

    great potential for such measurements at the microscopic

    and submicroscopic levels. The technique is currently be-

    ing evaluated as a tool for measuring the wetting properties

    of colloidal particles (48).

    Fig. 14 Illustration of micropipette techniques. (A) Based on analysis of pressure differences required to suck a microdroplet into the

    pipette. (B) Based on the examination of the forcedrop deformation relation.

    Fig. 15 Schematic of atomic force microscopy and its ap-

    plication to measurements of surface tension for microdroplets.

    Table 4 Results of interfacial tension measurements for the

    organic solvent microdroplets in water obtained with the mic-

    ropipette technique and their comparison to the similar results

    measured for bulk two-liquid systems

    System

    Interfacial tension

    measured for

    microdroplets

    [mN/m)

    Interfacial tension

    measured for

    macrosystem

    [mN/m)

    Water-ethyl acetate 6.8 0.6 6.8

    Water-chloroform 30.9 1.6 31.6

    Water-benzene 33.7 0.7 34.1

    Water-toluene 36.4 0.8 36.1

    Water-hepto 40.3 1.1 39.6

    (From Ref. 45.)

    3164 Measurement of Interfacial Tension in Fluid-Fluid System

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    14/15

    AFM is a scanning probe technique based on measuring

    interaction forces between a cantilever tip and a specimen

    (Fig. 15). The force measurement is based on measuring

    the deflection of the cantilever, which has a known spring

    constant. The cantilever deflection is detected by the ref-

    lection of a laser beam as shown in the figure. Movement

    of the specimen under the cantilever tip (both in the ho-

    rizontal plane for scanning, and vertically for force mea-

    surements) is controlled very precisely by a piezoelectric

    specimen stage (reverse systems with the piezoelectric

    stage attached to the cantilever holder are also in use).

    Interaction forces as small as 1pN (1012 N) can be mea-

    sured between the probe tip and the specimen.

    Fig. 15 shows an approach for measuring the in-

    terfacial tension between a probe tip and a microscopic

    drop of liquid. The capillarity forces exerted on the tip by

    the liquid can be measured as it is inserted into the drop

    and as it is withdrawn and detaches from the drop. Cal-culation of surface tensions from these force-distance

    curves will depend on the shape of the probe tip, but the

    equations should be similar to those for classical macro-

    detachment techniques.

    A major challenge in using the AFM technique to mea-

    sure interfacial tensions is fabrication of appropriate probe

    tips. Cylindrical tubes and spheres offer simple geometries

    and may present the best near-term options. Carbon nano-

    tubes have recently been used as AFM tips, and may offer

    another promising approach (49).

    REFERENCES

    1. Rusanov, A.I.; Prokhorov, V.A. Interfacial Tensiometry;

    Elsevier: Amsterdam, 1996.

    2. Adamson, A.W.; Gast, A.P. Physical Chemistry of Surfa-

    ces, 6th Ed.; John Wiley & Sons, Inc.: New York, 1997.

    3. Dukhin, S.S.; Kretzschmar, G.; Miller, R. Dynamics of

    Adorption at Liquid Interfaces: Theory, Experiment, Appli-

    cation; Elsevier: Amsterdam, 1995.

    4. Hiemenz, P.C.; Rajagopalan, R. Principles of Colloid and

    Surface Chemistry, 3rd Ed.; Marcel Dekker, Inc.: New

    York, 1997.

    5. Sonntag, H. Koloidy; PWN: Warszawa, 1982.

    6. Wilhelmy, L. Ueber die abhangigkeit der capillaritats-

    constanten des alkohols von substanz und gestalt des

    benetzten fasten korpers. Ann. Phys. Chem. 1863, 4 (29),177217.

    7. Lecomte du Nouy, P. A new apparatus for measuring sur-

    face tension. J. Gen. Physiol. 1919, 1, 521524.

    8. Harkins, W.D.; Jordan, H.F. A method for determination of

    surface and interfacial tension from the maximum pull on a

    ring. J. Am. Chem. Soc. 1930, 52, 17511772.

    9. Vold, R.D.; Vold, M.J. Colloid and Interface Chemistry;

    Addison-Wesley Publishing Co.: London, 1983.

    10. Zuidema, H.H.; Waters, G.W. Ring method for the de-

    termination of interfacial tension. Ind. Eng. Chem. Anal.

    Ed. 1941, 13, 312313.

    11. Young, T. Miscellaneous Works; Peacock, G., Ed.; J. Mur-

    ray: London, 1855; Vol. I, 418.12. de Laplace, P.S. Mechanique Celeste. Supplement to Book

    10; 1806.

    13. Sugden, S. The determination of surface tension from the

    maximum pressure in bubbles. J. Chem. Soc. 1922, 121,

    858866.

    14. Rehbinder, P.A. Dependence of surface activity and sur-

    face tension of solutions upon temperature and concentra-

    tion. Z. Phys. Chem. 1924, 111, 447464.

    15. Lord Rayleigh, OM.; F.R.S. On the theory of the capil-

    lary tube. Proc. R. Soc. London, Ser. A 1916, 92, 184

    195.

    16. Tate, T. On the magnitude of a drop of liquid formed un-

    der different circumstance. Phil. Mag. 1864, 27, 176

    180.

    17. Harkins, W.D.; Brown, F.E. The determination of surface

    tension (free surface energy), and the weight of falling

    drops: the surface tension of water and benzene by the

    capillary height method. J. Am. Chem. Soc. 1919, 41,

    499525.

    18. Miller, R.; Hofmann, A.; Schano, K.H.; Halbig, A.; Hart-

    mann, R. Measurement of surface and interfacial tensions

    with an automatic drop-volume tensiometer. Parfuem. Kos-

    met. 1992, 73, 390397.

    19. Bashforth, F.; Adams, J.C. An Attempt to test the Theory of

    Capillary Action; Cambridge University Press: London,

    1883.

    20. Lahooti, S.; del Rio, O.I.; Neumann, A.W.; Cheng, P.

    Axisymetric Drop Shape Analysis (ADSA). Applied Sur-

    face Thermodynamics; Neumann, A.W., Spelt, J.K., Eds.;

    Marcel Dekker, Inc.: New York, 1996; 441507.

    21. Andreas, J.M.; Hauser, E.A.; Tucker, W.B. Boundary ten-

    sion by pendent drops. J. Phys. Chem. 1938, 42, 1001

    1019.

    22. Stauffer, C.E. The measurement of surface tension by the

    pendent drop technique. J. Phys. Chem. 1965, 69, 1933

    1938.

    23. Misak, M.D. Equations for determining 1/H versus S

    values in computer calculations of interfacial tension by

    pendent drop method. J. Colloid Interface Sci. 1968, 27,

    141142.

    24. Quincke, G. Ueber die capillaritats-constanten des queck-

    silbers. Ann. Phys. Chem. 1858, 4 (15), 148.

    25. Padday, J.F. The profiles of axially symmetric menisci.

    Trans. R. Soc. London 1971, A269, 265293.26. Lang, G. Surface Tension of Liquid Elements. Handbook of

    Chemistry and Physics, 72nd Ed.; Lide, D.R., Ed.; CRC

    Press: Boca Raton, 1992; 133145, Chapter 4.

    27. Eustathopoulos, N.; Nicholas, M.G.; Drevet, B. Wettabi-

    lity at High Temperatures; Pergamon: Amsterdam, 1999;

    Chapter 4.

    28. Joos, P. Dynamic Surface Phenomena; VSP: Zeist, The

    Netherlands, 1999.

    M

    Measurement of Interfacial Tension in Fluid-Fluid System 3165

  • 8/2/2019 Bueno Tension Superficial Buenisimo

    15/15

    29. MacLeod, C.A.; Radke, C.J. A growing drop technique for

    measuring dynamic interfacial tension. J. Colloid Interface

    Sci. 1993, 160, 435448.

    30. Drelich, J. The Role of Wetting Phenomena in the Hot

    Water Process for Bitumen Recovery from Tar Sand.Ph.D. Dissertation, The University of Utah: Salt Lake

    City, Utah, USA, 1993.

    31. Rudin, J.; Wasan, D.T. Mechanisms for lowering of in-

    terfacial tension in alkali/acidic oil systems. Colloids Surf.

    1992, 68, 8194.

    32. Rosen, M.J. Surfactants and Interfacial Phenomena;

    Wiley-Interscience: New York, 1976.

    33. Dynamic Properties of Interfaces and Association Struc-

    tures; Pollai, V., Shah, D.O., Eds.; ACS Press: Champaign,

    Illinois, 1996.

    34. Passerone, A.; Liggieri, L.; Rando, N.; Ravera, F.; Ricci,

    E. A new experimental method for the measurement of the

    interfacial tension between immiscible fluids at zero bond

    number. J. Colloid Interface Sci. 1991, 146, 152162.

    35. Horozov, T.; Arnaudov, L. A novel fast technique for mea-

    suring dynamic surface and interfacial tension of surfactant

    solutions at constant interfacial area. J. Colloid Interface

    Sci. 1999, 219, 99 109.

    36. Vonnegut, B. Rotating bubble method for the determina-

    tion of surface and interfacial tensions. Rev. Sci. Instr.

    1942, 13, 6 16.

    37. Princen, H.M.; Zia, I.Y.Z.; Mason, S.G. Measurement of

    interfacial tension from the shape of a rotating drop. J.

    Colloid Interface Sci. 1967, 23, 99 107.

    38. Couper, A.; Newton, R.; Nunn, C. A simple derivation of

    Vonneguts equation for the determination of interfacial

    tension by the spinning drop technique. Colloid Polym.

    Sci. 1983, 261, 371372.

    39. Manual of the Spinning Drop Tensiometer Site 04; Kruss

    GmbH: Hamburg, Germany, 1995.

    40. Lucassen, J. The shape of an oil droplet suspended in an

    aqueous solution with density gradient. J. Colloid Interface

    Sci. 1979, 70, 355365.41. Sottmann, T.; Strey, R. Ultralow interfacial tensions in

    water-n-alkane-surfactant systems. J. Chem. Phys. 1997,

    106, 86068615.

    42. Yeung, A.; Dabros, T.; Masliyah, J. Does equilibrium

    interfacial tension depend on method of measurement? J.

    Colloid Interface Sci. 1998, 208, 241247.

    43. Evans, E.; Skalak, R. Mechanics and Thermodynamics of

    Biomembranes; CRC Press, Inc.: Boca Raton, Florida,

    1980.

    44. Evans, E.; Needham, D. Physical properties of surfactant

    bilayer membranes: thermal transitions, elasticity, rigidity,

    cohesion, and colloidal interactions. J. Phys. Chem. 1987,

    91, 42194228.

    45. Moran, K.; Yeung, A.; Masliyah, J. Measuring interfacial

    tensions of micrometer-sized droplets: a novel microme-

    chanical technique. Langmuir 1999, 15, 84978504.

    46. Sarid, D. Scanning Force Microscopy; Oxford University

    Press: New York, 1991.

    47. Wiesendanger, R. Scanning Probe Microscopy and Spec-

    troscopy: Methods and Applications; Cambridge Univer-

    sity Press: Cambridge, Massachusettes, 1994.

    48. Ecke, S.; Preuss, M.; Butt, H.-J. Microsphere tensiometry to

    measure advancing and receding contact angles on indi-

    vidual particles. J. Adhesion Sci. Technol. 1999, 13, 1181

    1191.

    49. Dai, H.; Hafner, J.H.; Rinzler, A.G.; Colbert, D.T.; Sma-

    ley, R.E. Nanotubes as nanoprobes in scanning probe mic-

    roscopy. Nature 1996, 384, 147151.

    3166 Measurement of Interfacial Tension in Fluid-Fluid System