Body Imaging at 7T?: Perspectives in 2012
Neil M. Rofsky, MD, FACR, FSCBTMR, FISMRM
![Page 2: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/2.jpg)
In 1999, did you think you would use a 3T scanner for body imaging?
• 1) I was confident of this
• 2) I was unsure
• 3) I was very skeptical
• 4) I thought it was impossible
![Page 3: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/3.jpg)
How would you characterize your current use of 3T for body imaging?
• 1) Interchangeably with 1.5T
• 2) Sparingly
• 3) Not at all
• 4) No access to a 3T in my practice
![Page 4: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/4.jpg)
Starfruit at 7T
![Page 5: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/5.jpg)
7T
• Unlike conventional 1.5T and 3T MR systems using a large volume built-in transmit body RF coil, 7 Tesla MRI at this point uses local transmit/receive RF coils
Umutlu,et al. Eur Radiol 2011; 21:841–849
![Page 6: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/6.jpg)
Why 7T?
• SNR advantage
• Chemical separation
• Spectro
• Multipoint Dixon
• Na-23, O-17, P-31
• Unique contrast
• Susceptibility – brain, DWI?
• Fxnal blood flow
![Page 7: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/7.jpg)
1H NMR spectra of subcutaneous fat (left) and tibial bone marrow (right) from a 26 year-old healthy male at 7 Tesla (7T).
Ren J et al. J. Lipid Res. 2008;49:2055-2062
![Page 8: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/8.jpg)
So what?
• The ability to resolve protons adjacent to double bonds allows noninvasive estimation of the fatty acid composition of adipose tissue.
• Potential as an internal standard to quantify intramuscular lipids
• New insights into insulin sensitivity, diabetes, and obesity
![Page 9: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/9.jpg)
Challenges: Images
• Artifacts
• Tissue susceptibility
• Chemical shift
• Signal heterogeneities
• Radiofrequency (RF) wavelength
• SAR
• Flip angle adjustments
• Echo train imaging
![Page 10: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/10.jpg)
7T, 1H frequency = 298 MHz λ
• the radio frequency (RF) ��s in body
tissues are ≤ the object of interest resulting in non-uniform excitation and receive patterns due to complex RF field interferences.
300 cm
![Page 11: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/11.jpg)
T2WI (TSE) at 7T
Umutlu,et al. Eur Radiol (2011) 21:841–849
![Page 12: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/12.jpg)
16 channel transmit/receive array
Ladd et al. 2010 IEEE International Symposium on Biomedical Imaging Pages: 572-572
What # is optimal???? $$$$
![Page 13: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/13.jpg)
7T Abdomen: T2WI
![Page 14: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/14.jpg)
T2W, echo-train imaging: FSE, TSE
• Mult refocusing pulses
• SAR intensive
• Mitigate?
�� FAs
• Novel sequence design
![Page 15: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/15.jpg)
T2W, echo-train imaging: FSE, TSE
1/100th reduction in SAR
Sarkar, et al. Radiology 2011; 259: 550-557.
![Page 16: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/16.jpg)
T2W, echo-train imaging: FSE, TSE
1/100th reduction in SAR
Sarkar, et al. Radiology 2011; 259: 550-557.
*
![Page 17: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/17.jpg)
Renal imaging at 7T (flat contrast)
T1WI GRE in phase T1WI GRE opposed phase
Umutlu,et al. Eur Radiol 2011; 21:841–849
![Page 18: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/18.jpg)
Renal imaging at 7T (flat contrast)
T1WI GRE in phase T1WI GRE opposed phase
Umutlu,et al. Eur Radiol 2011; 21:841–849
add-on system for RF shimming integrated on the small signal side of the MR system
![Page 19: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/19.jpg)
Renal imaging at 7T 2D FLASH
Umutlu,et al. Eur Radiol; 2011:841–849
L S
![Page 20: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/20.jpg)
Prostate at 7T: Radiative Surface Coil
0.5 mm x 0.5 mm x 3 mm Raaijmakers et al. Mag Res in Medicine 66:1488–1497 (2011)
3T ERC 0.7x 0.7 x 2mm
![Page 21: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/21.jpg)
7T: Patient acceptance
Mean grades < 3 (“very well tolerable”) except for exam duration at 7T.
Theyson, et al. Magn Reson Mater Phy (2008) 21:63–72
![Page 22: Body Imaging at 7T? - Society of Computed Body ......In 1999, did you think you would use a 3T scanner for body imaging? • 1) I was confident of this • 2) I was unsure • 3) I](https://reader034.vdocuments.us/reader034/viewer/2022042219/5ec5d01a4a29781b3c1abf6e/html5/thumbnails/22.jpg)
Conclusions
• Not useful for clinical in 2012
• Impossible??
• Niche applications??
• Stay tuned…..