Biovac 13ROperating Instructions

Page 2: Biovac 13R Operating Instructions - Vacutec

2 Biovac 13R

Operating environment of instruments

In order to extend the machine’s service life, please pay attention to the following factors which may damage the machine:

• EnvironmentalConditions;• Effectsofenvironment,includingnaturalultravioletradiation(ie:DirectSunlight)• Erosionandabrasionoftheprotectivecoverandothersafetyparts.

• Indooruse;• Altitude:<2000m;• Ambienttemperature:+5°C~+35°C;• Relativehumidity:<80%;• Operatingvoltage:230V,50/60Hz;• Pleaseensureadequateventilationinareaofuseandappropriatesealingmeasurewhenperformingthe separationofpoisonoussubstancesandpathogenicmicro-organismcells;• Ensurethereisnovibrationand/orastrongdrafttoinfluencetheperformance.• Ensurethereisnodustorflammableandcorrosivegaspresentintheambientair.

Please read this manual carefully when using the instrument for the first time!The maintenance of the instrument should be carried out by the agents authorized by Biovac.

The following materials should not be used in conjunction with this instrument:

• Flammableandexplosivesubstances;• Strongchemicalmaterials;• Toxicorradioactivesubstances,• Pathogenicmicroorganisms

TheBiovac13RHighSpeedRefrigerationBench-topCentrifugecanbemaintainedonlybyqualifiedserviceengineersusingappropriatetools.

Shouldtherebeaproblemnotmentionedinthismanual,pleasecontactaBiovacauthorizedagentforthecorrectprocessingmethods.

OnlyaccessoriesprovidedbyVacutecshouldbeused.Ifotheraccessoriesareused,Vacutecwillnotbeheldresponsibleforanydamageorinjurytoinstrumentoroperatorasaresult.

TheBiovac13RHighSpeedRefrigerationBench-topCentrifugeshouldbeinspectedandmaintainedatstatedintervals.

Page 3: Biovac 13R Operating Instructions - Vacutec

3Biovac 13R

1. Safety clauses

1.1 Operating requirements Biovac13RHighSpeedRefrigeratedBench-topCentrifugeisdesignedaccordingtopresenttechnical

standards.Thereforeitmustbeusedaccordingtothedesignrequirements.WhentheBiovac13RHighSpeedRefrigeratedBench-topCentrifugeisusedaccordingtothefollowingimproperorinappropriatemethods,equipmentmybedamagedortheoperatorinjured. • Notusedinaccordancewiththedesignrequirements/intendeduse. • Theoperatorandmaintainerhavenotbeentrainedinitscorrectuseandoperation. • Impropermodificationstothedesignwithoutauthorisation. • Failuretoreadtheregulationsforsafeuse.

PersonswhouseormaintaintheBiovac13RHighSpeedRefrigeratedBench-topcentrifugeshouldreadandunderstandrelatedusageandsafetyregulationsinthisOperatingManual.

Thefollowingregulationsshouldbeobeyedtoavoidaccidents: OperatingManualshouldbekeptneartheBiovac13RHighSpeedRefrigeratedBench-topCentrifugeforreferencebyoperators.

Biovac13RHighSpeedRefrigeratedBench-topCentrifugeisdesignedtoseparatebiologiccells,organismsandbacteria.Thedensityofthesampleshouldnotexceed1.2g/cm3atthemaximumspeed.

Duringoperation(whenthesampleisbeingseparatedorrotorsarerunning),operatorsanddangeroussubstancesshouldmaintainadistanceofatleast30cmfromthecentrifuge.Theventilationholeofthecentrifugeshouldnotbecloggedbysubstances.

IfthefollowingsafetyprecautionsarenotobeyedwhenusingtheBiovac13RHighSpeedRefrigeratedBench-topCentrifuge,operatorsorotherpersonsmaybehurtorthecentrifugeandsamplesmaybedamaged:

• Thecentrifugeisnotcorrosionprooforimplosionproof,thereforeitshouldbeusedinacorrosion andimplosionfreeenvironment.

• Thefollowingmaterialsareprohibited: Flammableandexplosivesubstances: Strongchemicalmaterials; • Toxicorradioactivesubstances,orpathogenicmicroorganisms. • Corrosivesubstancesmaydamagematerialsinsidethecentrifugeandtherotors,thereforethey shouldbeputinsidesealedcontainerswhileseparating.

1.2 Precautions • Beforerunning(whenseparatingsamples),appropriaterotorsmustbefixedfirmly; • Duringoperation(whentherotorisrunning)orafterthecentrifugehasbeenstopped(therotoris

stillrunning),openingthelidbyhandisstrictlyforbidden. • AccessoriesmustbeprovidedandapprovedbyVacutec.Somecommonparts,suchasglassand plasticcontainersforseparating,shouldbeVacutecapprovedandbeinaccordancewiththe requirementsofthehighestspeedandmaximumcentrifugalforceofrelatedrotors; • Donotusethecentrifugeorseparatesampleswhilethelidisopen; • Changingmechanicalpartsandelectronicsshouldbecarriedoutbyqualified Vacutecpersonnelonly; • Whenusingthecentrifuge,usetheappropriaterotorsaccordingtoload.DoNotoverload; • Maintenanceshouldbecarriedoutperiodically.See“Cleanandpurify”chapterformore information.

Page 4: Biovac 13R Operating Instructions - Vacutec

4 Biovac 13R

2. Introduction2.1 Layout of the Biovac 13R High Speed Refrigerated Bench-top Centrifuge

Front and Back view of Centrifuge

Descriptionofabovediagrams: 1. Frontcasingcover 2. ControlpanelanddigitalLCD 3. Lid

4. Observationhole 5. TopPanel 6. Frontventilation 7. Emergencyopen/Forceopen 8. Foot 9. Rearcasingcover 10. Rearventilation 11. PowerjackandOn/Offswitch

2.2 General Biovac13RHighSpeedRefrigeratedBench-topCentrifugewithmultifunctionprotectionintroduces

advancedtechnologysuchasspeedadjustmentviafrequencycontrol,microcomputercontrol,touchbutton,dialsetup,digitalLCD.Threerotorvariationsareavailable(fordetails,see“Table1:Typeofrotorandtechnicalparameters”).Thiscentrifugeisusedfortheseparationofbiologiccells,organismsandbacteria.

2.3 Device structure 2.3.1 The Device Lid

Onlywhenthelidlocksarelockedonboththeleftandrightofthelidmaythecentrifugerun; otherwisethewarningsystemwillsoundanalert(alightwillcomeonandabuzzerwillsound) andthedevicewon’trun.

Toopenthelidpress“”(openbutton)onthecontrolpanel.Oncethelidrisestoacertain height,thelidhingeandspringswillsupportthelid. Ifapowercutoccursortheopenbuttonisnotresponding,thesamplemustberemoved.Referto

Speed/F

S/F

Temp. Time.

1

1

2

2

3

3

4

4

5

5

6

6

7

7

8

8

9

Page 5: Biovac 13R Operating Instructions - Vacutec

5Biovac 13R

chapter7.1foremergencyopeninginstructions. Whiletherotorsarerunningand/orthepowersupplyison,DONOTopenthelidmanually.

2.3.2 The Chamber Thechambermaintainsastableworkingtemperature.

2.3.3 The Rotor Fordetailssee“Table1:Typeofrotorandtechnicalparameter”.Therotorrotatesata predeterminedspeedtocreatetherelativecentrifugalforcerequiredtoseparatevarioussamples. Thecentrifugalforceisseveralthousandtimesmorethang(accelerationofearthgravity),itis therefore important to use and maintain the rotor carefully and safely!

2.3.4 The Refrigeration System Therefrigerationsystemisusedtocreatealow-temperatureoperatingenvironmentforthe centrifugechamber.Thetemperatureisautomaticallycontrolled.Thecentrifugeshouldbe operatedinanenvironmentwithaconstantambienttemperaturebetween5ºCto35ºCandan ambienthumidity≤80%.Optimumoperatingconditionscanbeachievedwithadequate ventilationaroundthecentrifugeinaroomwithambienttemperature controlledbyair-conditioning. Aheatsinkisinstalledinthefrontofthemachinewhereairflowandcirculationisessential.Any dustshouldberemovedfromtheradiatorwithabrushorcompressedaironceeverythree monthstoensureeffectiveheatdissipation.Thedigitaltemperaturesensorinstalledatthebottom ofthechamberisusedtomonitorthetemperatureinthechambercontinuously.

Thetemperatureinthechambermayvaryfromthatofthesamplesatfirst,andtake aperiodtoequalise.

2.3.5 The Control System The control system allows the setting of the rotor speed, running time, temperature and control of displaysystemandwarningsystemoftheentiremachine.Toensureproperrunningofthe machineandthesafetyoftheoperator,DONOTopenthefrontcasingofthedevice,asthemain controlcircuitboardandotherelectronicsareinstalledontothefrontcoverofthedevice.

2.3.5 The Display System Thedisplaysystemcanshoweverysettingparametersimultaneously,andcantrackanychange inrealtime.Inaddition,itdisplaysrelativesystemerrorsandwarnings.

2.3.5 The Warning System Thewarningsystemwillalerttheuserof“overspeed”,deviceoverheating,lidwarning,etc. Analertisindicatedbyaflashing,red,warninglightonthecontrolpanelandabuzzerwillsound simultaneously.Ifrunning,thedevicewillbestoppedautomaticallyandtheuserwillbeunable tostartasequenceagainuntiltheerrorhasbeenresolved.NOTE:usermaypressthe“Esc”button onthecontrolpaneltosilencethewarningbuzzer.

2.4 Safety Protection Device

TheBiovac13RHighSpeedRefrigeratedBench-topCentrifugehasaseriesofsafetyprotectiondevices:

• Themountingrackismadeofsteelplatingandtheprotectionsteelringismadeofsteelcasting.Theinternalbladderoftheinnerchamberisstainlesssteel.ThefrontframeismadeofABS.Thelidisexplosionproof.Thereisaviewportinthecenterofthelid.

Page 6: Biovac 13R Operating Instructions - Vacutec

6 Biovac 13R

Onlywhenthecentrifugeisonandtherotorhascompletelystopped,willthe“”buttonopenthelid.

• OverspeedStop Thewarningbuzzerissoundedwhentherotorspeedexceedsthesettingspeedby200r/min.Therotorstopsautomaticallyifthespeedexceedsthemaximumratedspeedby250r/min.Waituntiltherotorstopscompletely,thenopenthelid.Restartoncetheerrorhasbeenresolved.

• Openingisstrictlyforbiddenduringoperation. If“”isaccidentlypressedduringoperation,thedevicewon’trespond,however,thewarningalarmis

indicated.

• Openinginthecaseofanemergency Iftheelectricityisofforthereisanurgentproblemwiththerotor,youmayopenthelidmanually

whenitisnotpossibletodosobypressingthe“”button(ENSURETHEROTORHASSTOPPEDCOMPLETELYFIRST,TOAVOIDINJURY).

DONOTUSETHEEMERGENCYOPENINGFACILITYUNNECESSARILY!

2.5 Technical Specifications

2.3.1 Technical Data

2.6 Settling the Centrifuge

2.6.2 Thisdeviceshouldbeputonaswing-outtable-boardofappropriaterigidity,andbefreefrom vibrationandawayfromanydirectheatsourceorsunlight.

2.6.2 Thereshouldbeaspaceof10-15cmaroundthedevicetoallowforheatemissionfromthe condenser.

2.6.3 Aftermounting,adjustthe4feetlevelstoensuretherequiredsupportonthetable-board.

2.6.4 RequiredPoweroutputforthepowersupply:AC230V,50/60Hz.

THEDEVICESHOULDALWAYSBEEARTHED.

2.7 Configuration

Thefollowingaccessoriesaresuppliedwiththedevice.(Seepackinglistfordetails)

Youshouldinspectthecontentswhenreceivingproductsandunpacking.Ifthereissomethingwrong,pleasecontactVacutec.

Speed/F

S/F

Temp. Time.

1

1

2

2

3

3

4

4

5

5

6

6

7

7

8

8

9

Speed/F

S/F

Temp. Time.

1

1

2

2

3

3

4

4

5

5

6

6

7

7

8

8

9

Speed/F

S/F

Temp. Time.

1

1

2

2

3

3

4

4

5

5

6

6

7

7

8

8

9

Page 7: Biovac 13R Operating Instructions - Vacutec

7Biovac 13R

3 Available optional accessories

ThestandardconfigurationoftheBiovac13RHighSpeedRefrigeratedBench-topCentrifugeincludes: 1xCentrifugeandaNo.#2201rotor.

Accessoriesalsoavailableonenquiryare: RotorsNo.#2202and#2203(Fordetails,see“Table1:Typeofrotorandtechnicalparameters”)

Table1:Typeofrotorandtechnicalparameters

TypeofRotor

Rotor Number

Angleoftheangle rotor

(º)

Rated/Maxspeed (r/min)

Maximum capacity (ml)

Max centrifugal radius(cm)

Rated/Maximum centrifugal

acceleration (xg)

AnglerotorFS15.02 #2201 45º 13,800/

15000 24x2.0 8.2 17,459/20627

AnglerotorFS15.002 #2202 45º 13,800/

15000 64x0.2 7.2 15,330/18112

AnglerotorFS15.05 #2203 45º 13,800/

15000 16x5.0 8.9 18,949/22388

Page 8: Biovac 13R Operating Instructions - Vacutec

8 Biovac 13R

4.Preparationsbeforeuse4.1Transportationandinstallation Biovac13RHighSpeedRefrigeratedBench-topCentrifugeistransportedinspecial-purposecartonwith

protectionmaterialsforbufferinginside.Afteropening,theprotectionmaterialsforbufferingshouldberemoved.

• Thenetweightofthedeviceisabout47kg,thereforeitisnecessarytoliftthedevicefrombothsides(theleftandright)tokeepitsbalance.Thefrontcovershouldnotbepressedonorusedforlifting.

• Keeplevelandavoidvibrationwhentransporting• Pleaseusespecialpackingcasesduringlong-distancetransportation.Keeplevelandhandlewith

care!

4.2ProperplacementoftheCentrifuge Biovac13RHighSpeedRefrigeratedBench-topCentrifugeisforindooruse,andtheworkareashould

meetthefollowingrequirements:

• Asafetyclearanceof30cmmustbemaintainedaroundthecentrifuge.Whilerunning,thereshouldbenoindividualorhazardoussubstanceincloseproximitytothecentrifuge.

• Theadapterortabletopshouldbesolidtoavoidvibration;ifmobileracksortrolliesareused,theyshouldbeequippedwithlockingdevicestoensurethesafeoperationofthecentrifuge.

• Ifthecentrifugeisplacedbesidethewallorinacorner,thereshouldbeaspaceof10cmbetweenthebackofthecentrifugeandthewalland15cmbetweeneachsideandthewalltoallowsufficientventilationandheatemission.

• Thecentrifugeshouldbeplacedawayfromwindowstoavoidtheheatfromdirectsunlight.• Thefoursupportinglegsofthecentrifugeshouldstandevenlyona“swing-out”table-board.• Thecentrifugeshouldbeputinacleanroomwithaconstantambienttemperaturebetween5ºCand

35ºCandanambienthumidityoflessthan80%

4.3Fixedplacementofequipment Afterthecentrifugehasbeenplaced,don’tmoveitfrequently.Alwaysensurethecentrifugestandslevel

onthetable-board.Theshouldbenovibration.

Toavoiddamagetotherotoraxis,pleaseremovetherotorwhenmovingthecentrifuge.

4.4Properconnectionofelectricsupply

Page 9: Biovac 13R Operating Instructions - Vacutec

9Biovac 13R

Theeachoftheelectricplugconnectedtothecentrifugemustbeingoodcondition.Thevoltageandfrequencyofthepowersupplyshouldbeinaccordancewiththerequirementsstatedinthetechnicaldescriptionorspecificationslabelledonthenameplateofthecentrifuge.Thepowerjack/switchisattherearofthemachine,“I”foronand“O”foroff.Onlyusethepowercableincludedwiththemachine.

Neverpress“▷/☐”or“QTemp”buttonimmediatelyafterstartup;thedoormustbeopenedfirsttoremovetherotorfixingexpandedpolyethylene(EPE)androtor,andthenthechamberandrotormustbeinspectedandcleanedbeforerunning.

5. Direction for use and operation5.1 Control panel and display interface

ShownasFig.6,controlpanelisdescribedinbriefasfollows:

1-Datadisplaywindow 2-Cancel/Exitbutton 3-Quickaccessbuttonofcommonparameters 4-Menubutton 5-Quicktemperaturebutton 6-Dooropeningbutton 7-Inchingbutton(quickstart;pressthebuttonandrotorspeedwillquicklyincreasetoitsmaximum allowablespeed;releasethebuttonandrotorwillquicklystop) 8-Start/Stopbutton 9-Dial(Selectandmodifysettingparameters;actsas“Enter”buttonwhenpressedin) 10-Speedandcentrifugalforcesettingbutton 11-Curvesettingbutton 12-Temperaturesettingbutton 13-Timesettingbutton

Page 10: Biovac 13R Operating Instructions - Vacutec

10 Biovac 13R

ShownasFig.7,maininterfaceisdescribedinbriefasfollows:

1. Thetoprowisthestatepromptarea.“”indicatesdoorisinlockedstate;“”indicatesdoorisin unlockedstate;“”indicatesthebuzzerisinnormalworkingstate.

2. Thesecondrowisthespeed/centrifugalforcedisplayarea:theleftismaindatadisplayandthe rightiscorrespondingauxiliarydatadisplayafterconversion.Whentherotorspins,real-timespeed isdisplayed;whentherotorstops,“0”isdisplayed.Whencentrifugalforceisauxiliarydataandthe rotorisnotidentified,displayedcentrifugalforceis“-----”.

3. Thethirdrowisrunningtimeandchambertemperaturedisplayarea:whentherotorisnotrunning, displayedrunningtimeis“00:00”;whentherotorisrunningthedisplayedvalueistheremaining runtime.Displayedtemperatureisreal-timechambertemperaturevalue.

4. Thefourthrowissetparameter/alarmmessagedisplayarea:whenthesystemisinnormaloperation thedisplaysare:SetSpeed/Centrifugalforce;Setruntime;Settemperature;SetcurveandCurrent identifiedrotornumbersuccessivelyfromlefttoright.Whenasystemerroroccurs,therelevant errormessageisdisplayed.

5.2 Start Connectthepowercablesuppliedwiththemachineintothepowersocketontherearpanelofthe

machineandconnecttheotherendintomainssupply.ThepowersupplyusedforthemachineisAC230V,50/60Hz.Oncepluggedin,turnonthepowerswitchnexttothepowersocket.Thebuzzerwillsoundandthesoftwareversionwillbedisplayed.Oncetheself-checkiscompleted,itwillenterthemaininterface(asshowninFig.7),thenextoperationcannowbeconducted.

Page 11: Biovac 13R Operating Instructions - Vacutec

11Biovac 13R

5.3 Open Whilethepowerswitchisturnedonandtherotorhascompletelystopped,press“”button(door-

openbutton)onthecontrolpaneland“”willappearinthedisplaywindow;thelidwillautomaticallyopenviaspringaction,andthelidneedstoberaisedtherestofthewayandleanedbackwardagainsttheframe;theinnerchamberwillnowbevisibleandtheiconrepresentingdoorstatusonthemaininterfacewillbedisplayedasunlocked.

ATTENTION:whenthedoorisopened,thelidwillautomaticallybounceuptoacertainheight,keepclearofthelidduringthisoperationtoavoidinjury.Inthecaseofanerror,wherethedoorcannotbeopenedautomaticallyandthesamplesinthechamberneedtoberemoved,thedoorcanbeopenedmanuallybythemethodstatedinchapter7.1

5.4 Close Presstheliddownuntilthedoorhookontheliddrivesthelockingdeviceandyouhearthelockclick;

whentheicondisplaysas“”,itindicatesthatthedoorhasbeencorrectlyclosedandlocked. LidAdjustment:whenclosingthelid,ifyoufeelthelidistootightortooloose,unscrewthelocknutonthedoorhookwitha17mmsocketwrenchandthenadjustthedoorhookappropriately;turnclockwisetotightenandanti-clockwisetoloosen.Aftercompletingtheadjustment,tightenthelocknutagain.

Whenclosingthedoor,anappropriateforceshouldbeappliedtopresstheliddowntoavoiddamage to the door lock.

5.5 Installation/Dismounting of the rotor Ifinappropriaterotorandadapterareused,itwillcauseanunsatisfactoryseparationeffectand

could also damage the centrifuge.

TherotorusedmustbeaproductconfirmedandappointedbyVacutec,otherwiseitmaypotentiallycauseafault.ThespecificationandmodelofalltypesofrotorsconfirmedbyVacutecareprovidedinthismanual(fordetailsonspecificationandmodel,seeTable1“RotorTypeandTechnicalParameters”).

Wheninstallingrotors,usethe5mmhexagonspannersuppliedalongwiththemachine. Installation steps for rotors are as follows:

• Turnonpowerswitchandwaitforself-checkcompletion. • Press“”buttonandopenthelidofthecentrifuge;makesurethechamberisclean andimpurityfree;

• Cleanexternalsurfaceofmotorshaftandinspectthesealringonthemotorshaftensuring itisintact; • Preparetherotoryouwanttouse;holdtherotorwithbothhandsandalignthecenterholeofthe rotorwiththemotorshaft,andthenverticallyloweritontotheshaft; • Screwonthelocknutbyhandandthenusethe5mminnerhexagonspannertotighten thenutclockwise; • Rotorinstallationiscompleted.

Beforeuse,especiallybetweenprolongedperiodsoftime,theseatingoftherotorshouldbeinspectedforanychanges;whennecessary,thelocknutshouldbere-tightenedtoensuretherotorissecure.

Page 12: Biovac 13R Operating Instructions - Vacutec

12 Biovac 13R

5.6Maintenanceofmotorsupportshaft Maintenancestepsofmotorsupportshaft: • Turnonpowerswitchandwaitforself-checktocomplete • Press“”buttonandopenthelidofthecentrifuge;thendismounttherotorusing5mminner

hexagonspanner.NOTE:tightenclockwiseandloosenanti-clockwise. • Removedust,impuritiesandseparatedsampleresidueinthecentrifugechamberensuring

itisclean. • Cleantheconicalsectionandthescrewthread(showninFig.8)ontherotatingshaft,ensuringno

dirtordustremains;thenevenlyapplythelubricatinggreasesuppliedalongwiththemachine.

5.7 Calculation of rotor load Calculationofmaximumload:

Biovac13Rdeskcentrifugewillgenerateahighcentrifugalforceduringhigh-speedrunning;eachrotorisdesignedtohaveenoughmechanicalstrengthtowithstandthisforcewhenworkingattheratedmaximumspeed,buttoavoidaccidentorinjury,therotorloadmustnotexceeditsmaximumratedload.

Whenseparatingsamples,thesamplesarefirstplacedintoacontainerandthenplacedtogetherintoa rotor, the sum total of the container and sample must not exceed maximum rated load of the rotor, otherwiseyoumustreducethesampleweightorcalculatepermissiblerunningspeedofrotor(NPERM)toensuretherotorloaddoesnotexceeditsmaximumratedload.

Calculationmethodofpermissiblerunningspeed(NPERM)ofrotorisasfollows: NPERM=NmaxX(maximumpermissibleload÷actualload)0.5

Page 13: Biovac 13R Operating Instructions - Vacutec

13Biovac 13R

Where: Nmax=maximumratedspeed; Maximumpermissibleload=maximumpermissibleload; Actualload=actualload

Don’tusetherotorunderoverloadedconditions,asthismayresultinseveredamageorinjurytotheequipment or operator.

5.8 Sample filing in centrifugal container Whenthecentrifugeruns,ifbalancedcorrectly,separationzoneofsampleswillnotinterferewitheach

otherduetotheabsenceofvibration,resultinginanimprovedcentrifugaleffect.Therefore,samplesshouldbeplacedintothecentrifugalcontainerassymmetricallyaspossibletoachievethebestbalanceduringthesequence.

Carefullyinspectthecentrifugalcontainer(centrifugetube,etc)beingused,ensuringitisinaccordancewiththeallowablemaximumratedacceleration(centrifugalforce);alwaysusethelowestrunning speed possible.

Pay strict adherence to the service life of the centrifugal container in use; especially when frequently runningundermaximumpermissibleload(speedandtemperature). Ifdamaged,itshouldbereplacedimmediately!

5.9 Safe use of rotor • Beforetherotorruns,sampleshouldbeloadedcorrectlyandsymmetrically.

• Whenexchangingarotor;usethe5mmspannertoloosenthelocknutanti-clockwiseand removetherotor. • Neverstartthemachinebeforethelocknuthasbeentightenedontotherotatingshaft. • Ifthecentrifugewillberunrepeatedly,therotorlocknutmustbeinspectedaftereachruntoensure itisstilltight.Ifthelocknutisloose,itmustbetightenedbeforeproceedingwiththenextrun. • Centrifugaltesttubesdonotneedtobeloadedsimultaneously,butdoneedto beloadedsymmetrically.

5.10 Parameter setting ATTENTION:

• Beforesettingparametersviathecontrolpanel,therotormustbecorrectlyinstalledon therotatingshaft. • Buttonsaretouchsensitive.Simplypresstherelevantbuttonfortherequiredoperation. • Iftheparametersareinsertedincorrectly,theycanbereset.

Page 14: Biovac 13R Operating Instructions - Vacutec

14 Biovac 13R

Turnthedialwhileinthestandbyinterfacetoentertheparametersettinginterface;continuetoturnthedialtoswitchbetweentheparametersthatneedtobeset;whentherespectiveparameteritemisselected,agreenborderwillflickeraroundit.Afterselectingtherequiredparameter,pressthedialintoconfirmyourselectionandentertherequiredparameter.

5.10.1 Setting of Speed:

SettingrangeofspeedofBiovac13Rcentrifuge:minimumspeed:300rpm;maximumspeed:13800rpm.

Description: 1. Whileinthestandbyinterface,turntehdialuntilthespeedsettingzoneflickersasshowninFig.10.

Press[S/F]buttontoquicklyselecttheappropriateparametersetting. 2. Pressthedialtoconfirmandenterthespeedsettinginterface;onceintheinterface,press[S/F] buttontoswitchbetweenspeedselectionandcentrifugalforceselection.Theselectedparameter itemhasaflashinggreenborderindicatingthatthenumericalvaluecanbemodifiedandset;non- selecteditemsaregreyedout,shownasFig.10.Iftherotorisnotidentified,centrifugalforce displayis“-----”. 3. Turnthedialtochangethenumericalvalue.Speedissetbysteppedvariableadjustment:change invalueis100ateachturnofthedial.Iftherotorhasbeenidentified,thecorrespondingvaluewill beautomaticallyadjustedrelativetothevalueinsertedforthedesiredspeed. 4. Adjusttotherequiredspeedvalueandpressthedialtosavethesetting;youcanalsopressthe“Esc” buttontocancelthechangeandreturntothestandbyinterface.

Page 15: Biovac 13R Operating Instructions - Vacutec

15Biovac 13R

5.10.2 Setting of the centrifugal force:

SettingrangeofcentrifugalforceofBiovac13Rcentrifugerotor:minimumcentrifugalforce:100xg;maximumcentrifugalforce:22380g.

Settingrotorspeedcanbechangedbysettingthe“centrifugalforce”value;thespeedandcentrifugalforcewillnotcorrespondintheirrelationuntilaftertherotorhasbeenidentified(namelytherotornumber),centrifugalforceandrotorspeedwillthenconvertinrelationtoeachotherandthecurrentcentrifugalforcewillbedisplayed.

Settingmethod(asper5.10.1SettingofSpeed):

Calculation of centrifugal force Relative centrifugal force is usually thousands of times more than terrestrial gravity and it is the unit usedtomeasuretheefficiencyofdifferentkindsofinstrumentsseparatingorprecipitatingarticles.Calculationofcentrifugalforceisrelatedtocentrifugalspeedandcentrifugalradius;centrifugalforceiscalculatedbytheequationbelow: RCF=11.18X(n/1000)2Xr Where: r=centrifugalradius(cm) n=centrifugalspeed(rpm)

Attention: maximum centrifugal force value is only related to maximum speed and maximum centrifugal radius. The centrifugal force value “X g” displayed in the interface is calculated according to current rotor speed and maximum centrifugal radius.

Page 16: Biovac 13R Operating Instructions - Vacutec

16 Biovac 13R

5.10.3 Setting of speed-up and speed-down curves:

Speed-upandSpeed-downtimecontrolofBiovac13Rcentrifuge:3groupsofspeed-upcurvesand3groupsofspeed-downcurvesareprovidedrespectivelyaccordingtodifferentrotorsandtheycanbefreelyarrangedingroups.3#curveismaximumspeedand1#curveisminimumspeed.

Settingmethod(asper5.10.1SettingofSpeed): 1. Turnthedialwhileinthestandbyinterfacetoentertheparametersettinginterface. 2. Inthecurvesettingmenuinterface,thecurvesettingboxisdisplayedbyagreenflashingborder. 3. Orsimplypress[Curve]buttontoquicklyselectthatparametersetting.

4. Turnthedialtochangethevalueofthecurvesetting 5. Thefirstrowofcurvesettingmenuisthespeed-upcurveandthesecondrowisthespeed-down curve.Pressthe[Curve]buttontoswitchbetweenthesetwooption.Adjusttothepreferedcurve andpressthedialtosavethesetting;orpressthe“Esc”buttontocancelthechangeandreturnto standbyinterface.

5.10.4 Setting of temperature:

Page 17: Biovac 13R Operating Instructions - Vacutec

17Biovac 13R

ChambertemperaturecontrolofBiovac13Rcentrifuge:minimumtemperatureaccessiblevarieswitheachtypeofrotoratdifferentspeeds.Settingrangeoftemperatureforthecentrifugeis-10~40ºC. Settingmethod(asper5.10.1SettingofSpeed) 1. Turnthedialwhileinthestandbyinterfacetoentertehparametersettinginterface. 2. Inthetemperaturesettingmenuinterface,whenthetemperaturesettingboxisselectedisdisplaysa flashinggreenborder. 3. Orsimplypress[Temp]buttontoquicklyselectthatparametersetting. 4. Turnthedialtochangethevalueofthetemperaturesetting. 5. Thesecondrowintemperaturesettingmenuisaswitchingdisplayindicatingchambertemperature andtubetemperature(displayedsampletemperatureisestimatedvalue,notactualvalue).“Chamber temp”ischambertemperatureand“tubetemp(simulation)”issampletemperature.Press[Temp] buttontoswitchbetweenthetwomodes.Adjusttotherequiredtemperatureanddisplaymode andpressthedialtosavethesetting;orpress“Esc”buttontocancelthechangeandreturnto standbyinterface.

5.10.5 Setting of running time:

SettingrangeofruntimeofBiovac13Rcentrifugerotor:1min~99h59min. Inthetimesetting,hour[H]andminute[M]canbesetrespectively. Settingmethod(asper5.10.1SettingofSpeed) 1. Turnthedialwhileinstandbyinterfacetoentertheparametersettinginterface.Onceinthetime settinginterface,tehtimesettingboxisdisplayedasselected,indicatedbyaflashinggreenborder. 2. Settingsequenceoftime:minute->hour->saveandexit.Thewhitenumericalvalueiscurrent adjustedparameter.Oncethevaluehasbeenadjusted,pressthedialtoselectthehoursetting.Press [Time]buttontoquicklyselectparametersetting. 3. Turnthedialtoswitchbetweenthevalues. 4. Thesecondrowoftimesettingmenuisruntimemodeoption;“countdown@start”iscountdown atstartuptime;“countdown@sets/f ”iscountdownaftercurrentspeedreachesthesetspeedor presetcentrifugalforce.Press[Time]buttontoswitchbetweenthetwomodes. 5. Adjusttherequiredtimeanddisplaymodeandpressthedialtosavethesetting;orpress“Esc” buttontocancelthechangeandreturntostandbyinterface.

Page 18: Biovac 13R Operating Instructions - Vacutec

18 Biovac 13R

5.10.6 Reading and storage of user setting data

Biovac13Rcentrifugeprovides3groupsofsettingdatasavedforusers.Userscansaveandreadtheirmostcommonparametersettings. Settingmethod(asper5.10.1SettingofSpeed): 1. Press[P1]/[P2]/[P3]buttonwhileinthestandbyinterfacetoquicklyenterthe parametersettinginterface. 2. Turnthedialinthesettingmenuinterfacetoswitchthe“Read/Save”function;theselecteditemis indicatedbyaflashinggreenborder. 3. Whenswitching“Read/Save”,correspondingdataisdisplayedintheupperparameterdisplaybox. Forexample,whenselecting“loadfrompreset”,tehdatadisplayedintheupperdisplayboxis thedatathattheuserhaspreviouslysaved;pressthedialandthedataisreadouttocovercurrent setparametervalue;whenselecting“savetopreset”,thedatadisplayedintheupperdisplayboxis thecurrentdata;presstehdialtosavethedata.Press[P1]/[P2]/[P3]toswitchbetweenthe 3groupsofdata. 4. Adjusttherequiredtimeandpressthedialtosavethesetting;orpress“Esc”buttontocancelthe changeandreturntostandbyinterface.

5.10.7 Parameter change during the run process

Duringtherunofthecentrifuge,thespeed,temperature,time,centrifugalforceandotherparameterscanbechanged;oncenewparametersareentered,displayscreenreturnstomaininterfaceandthenewparameterscomeintoforce.

Page 19: Biovac 13R Operating Instructions - Vacutec

19Biovac 13R

5.11 Start centrifugal separation Oncethesettingofuserparametersiscomplete;rotorshavebeeninstalled;thesampleforseparation

hasbeencorrectlyinserted;thelidofthemachinehasbeencorrectlyclosed,thenthemachinecanbestartedtobegintheseparation.

Attention: Indication whether the lid of the machine is closed correctly is displayed on the main interface via the door lock status icon.

5.11.1 Press “ “ to start

1. Pressthestartbutton““onthecontrolpanelandtherotorwillfirstbeidentifiedata lowspeed;onceidentified,thecorrectsetspeedorcentrifugalforcewillbeenforced. 2. Therotorwillbeacceleratedtothesetspeedaccordingtothesetspeed-upcurve. 3. Runtimeisdisplayedby“countdown”methodduringtherunprocess;whenruntimedisplay reacheszero,namely“00:00”,thecentrifugestopsandtheoperationiscomplete 4. Chambertemperaturequicklydropsatthestartofthecompressor;oncethechamberreachestheset temperature,thecompressorswitchesoff. 5. Real-timespeedandcentrifugalforcearedisplayedduringtherun;displayedcentrifugalforceis acquiredviaautomaticconversioninrelationtotherotordetectedbythecentrifuge. 6. Ifrotorspeedorcentrifugalforceexceedsitsmaximumpermissiblevalueorsetvalueduring therun,analarmwillsound;whenmotortemperatureprotectionistriggeredorcompressorover voltageprotectiontriggers,analarmwillsoundandprompttheusertomakethe relevantprocessingadjustment.

Attention:fordetailsonalarmpromptmessagecontent,see“Troubleshooting”. 7. Thelidofthemachinemustnotbeopenedwhiletherotorisstillturning.

5.11.2 Press “ “ button to start

Pressandhold““buttonandthesystemwillidentifytherotor;onceidentified,therotorspeedwillquicklyincreasetopermissiblemaximumspeedonthemachine;release““buttonandtherotorwillquicklyslowuntilitstopscompletely.

5.12 Stop Separation

Oncethecentrifugebeginsitsrun,itcanbestoppedbythetwomethodsbelow: 1. Undernormalconditions,theruntimeofthecentrifugehasbeenpreset,andwillautomaticallystop

aftercountingdowntozero.

2. Ifyouwanttostopthecentrifugeduringtheprocessofanormalrun,pressthestopbutton““onthecontrolpanelatanytime,andthecentrifugewillstoprunningimmediately.TheCountdowninthedisplaywindowindicatingtheruntimewillstopsimultaneously.

Onlyoncetherotorspeeddropstozero,i.e.speeddisplaywindowdisplays“0rpm”andcentrifugalforcedisplaywindowdisplays“0xg”,youcanpresstheopenbutton““onthecontrolpanelandthelidwillpopopenviaaspringaction,thelidcanthenbeopenedbyraisingitbyhand;thedoorlockstatusicononthemaininterfacedisplaysthelockasopen,theseparatedsampleinthecentrifugecannowberemoved.

Page 20: Biovac 13R Operating Instructions - Vacutec

20 Biovac 13R

5.13 Temperature Control

5.13.1 Pre-Refrigeration Function

Whenthelidofthecentrifugeisclosedandpriortherunningofthepresetprogramme,oraftercentrifugalseparationiscomplete,thecentrifugewillenter“pre-refrigerationmode”accordingtothesettemperature.Inthisway,itcanreachtherequiredtemperaturepriorcentrifugalseparation;andmaintaintheappropriatetemperatureafterseparationtoprotectsamples.Ifambienttemperatureorchambertemperatureishigherthanthatofthesample,usersareadvisedtopre-refrigerateandtheninsertsamplesforcentrifugalseparation,otherwisethe“TEMPOVER40ºC”warningmayappear.

Attention:Pre-Refrigerationfunctionwillstopafter1hour;Itisresetbyopeningandclosingthedooragainorduetopowerfailure.

5.13.2 Quick temperature function

Thecentrifugehasaquicktemperaturefunction.Firstsetthetargettemperatureparameterandexpectedtimedurationandmakesuretherotoriscorrectlyinstalledandlocked;closethelidandthenpressthe<Qtemp>buttontostartthequicktemperaturefunction,andtherotorwillspin.

Ifsettemperatureishigherthanchambertemperature,thecompressordoesnotstartandheatisgeneratedbyfrictionbetweenthehigh-speedrunningrotorandair,toquicklyreachthetargettemperature.

Ifsettemperatureislowerthanthechambertemperature,therotorrunsatthespecifiedspeed;thecompressorwillstartandtherefrigerationsystemstartstoquicklyreducethechambertemperaturetoreachthetargettemperature.

Afterthetemperaturereachesthesetvalue,therotorcontinuesrunning,butmainparameter<Qtemp>ondisplayscreenwillflickerandsimultaneouslybuzzersoundsintermittentlytoremindtheuser

Page 21: Biovac 13R Operating Instructions - Vacutec

21Biovac 13R

thatquicktemperatureiscomplete.Pressthestopbuttonandthequicktemperaturefunctionwillbeterminatedandtherefrigerationandrotorwillstop.

5.13.3 Temperature control in the separation process

Whenthecentrifugeisstarted,thetemperaturecontrolsystemwillrunforaperiodoftimeuntiltherequiredtemperatureisreached.Thetemperatureinthecentrifugalchamberwillthenbestablewithin±2ºCrangeofsettemperature.

5.14 Running record

In order to ensure normal operation of the machine and extend the service life of the machine, and accumulateinformationaboutthemachineoperationtoassistinthediagnosisoffaultsoraccidents,pleasecarefullykeeparecordofeachrun(onebookforeachmachine).

Belowistherecommendedoperationrecordtable:

RECORD TABLE

Date of test

Name of sample

Weight of monotube (including sample)

Rotor No.

Rotory

speedTime Temperature Working

condition Operator

Page 22: Biovac 13R Operating Instructions - Vacutec

22 Biovac 13R

6. Maintenance

6.1 Cleaning and sterilization

Usershavetheresponsibilitytocleanandpurifythecentrifugewhendangeroussubstancesspilloverintothemachine.Usersshouldcleanandpurifyitaccordingtothemethodrecommededbythemanufacturerstoensurethatthemachineisnotdamaged.

Inappropriate detergents and incorrects methods of cleaning may do harm to the centrifuge or internal parts. Cleaning and sterilization should be carried out according to operating procedures described in the operating and technical instructions.

6.1.1 Implementation of cleaning and sterilization

Turnoffthepowerandremovethepowercablebeforecleaning!

Thecentrifugeshell,rotorchamber,rotorandcentrifugalcontainersusedandsoonshouldbecleanedandmaintainedregularly;thuscontaminationcanbeclearedavoidingdamagetothepartsusedandapollutedenvironment.Ifyouarenotclearonwhichcleaningagentsshouldbeusedforcleaningthecentrifuge,pleasecontactVacutec.

• Organicsolventsareforbidden,astheycandegreasethebearingsofthemotor; • Liquid,especiallyorganicsolventsshouldnotcomeincontactwiththesupportingshaftandball

bearingswhenbeingcleaned; • Ifthereisfrostorathinicelayer,theyshouldbewipedawayoncetheyreverttowater.

6.1.2 Implementation of high-pressure sterilization

• Rotorsusedbythiscentrifugecanendurehigh-pressuresterilization.Temperature:121ºCwith apressureof250kPa,maintainedforatleast10minutes.Pleaserefertotheinstructionsofferedby manufacturersvariousseparatingcontainers.

• Alwaysconfirmwhetheritispermissibletosterilizerotorsandcontainerswithhigh-pressure sterilizationmethod.Thetemperatureanddurationofhigh-pressuresterilizationshouldbestrictly controlledaccordingtothedifferentrotorsandcontainersusedforseparating;

• Dontusearotororcontainerforseparatingiftheyarecorrodedordamaged.

6.1.3 Maintenance

6.1.3.1 Donotbringtherotorintocontactwithsharpobjects.Whenmovingordisassemblingthe centrifuge,youshouldtakecarenottodamagetherotor,asscratchesandindentationsmaycause cracksduringoperation.

6.1.3.2 Regularlycheckforspots,groovesorminutecracksintherotor(especiallyatthebottomofthetest tube).Ifanyarefound,stopusingthemimmediatelyandcontactVacutec.

6.1.3.3 Therotorshouldbecleanedonceeveryweekundernormaloperation.Rinsetherotorimmediately ifitsisusedtoseparatesalinesolutionorothercorrosivesamples.Ifliquidissprinkledontothe rotor;blotitupandrinsethepartimmediately.

Page 23: Biovac 13R Operating Instructions - Vacutec

23Biovac 13R

6.1.3.4 Pleasecleanwithspongesorcottonclothdippedinneutralcleaningsolution,thenrinsethe cleaningsolutionoffwithdistilledwater.Don’tsprinkleorspraywaterontotherotor;becauseit maycausetherotortocorrodeduetoresidualwater.Inverttherotorandallowittodry.

6.1.3.5 Dustthecoolingplatesofthecondensatoronceeverythreemonths.

6.1.3.6 Checktheprotectivecover

6.1.3.7 Inspectionofrotorpartsandsafetyconditions.

6.1.3.8 Confirmthedeviceisadequatelyearthed.

6.1.3.9 Removeanywaterorcondensationfoundinthechamberimmediately.

6.1.3.10 Turnofftheelectricityanddisconnectthepowercablebeforedisassembingthefrontcasingofthe device.Attention: This should be carried out by a trained Vacutec Technician only.

6.1.3.11 OnlysparesprovidedbyVacuteccanbeusedinthisdevice.

6.1.3.12 Turnofftheelectricitywheneverthecentrifugeisn’tinuse.

6.1.3.13 TransportationandStorage Thismachineisaprecisioninstrument.Water,vibration,tiltingandinvertedplacementarestrictly

forbiddenduringtransportationandstorage.

Please note the protection indicators on the packaging!

Fragile This Side UP Keep Dry Do Not Tilt

7. Trouble shooting

7.1 To open lid in case of emergency

Undernormalconditions,thefunctiontoautoopenthelidwillbedeactivatedwhentheelectricityisoffortheoperationtoopenthelidisdisabledbyaccident.However,samplesinthechambermayneedtoberemoved.Inthiscase,youmayopenthelidofthecentrifugemanuallytotakesamplesout.

Attention:Thisisforemergencyonly!

Intheeventofapowerfailure,thebrakewillnotengage,thereforetherotorwillcontinuetospinforaprolongedperiod.Pleasewaitforittostopcompletely(observethroughtheobservationholeattehtopofthelid)beforeattemptingtoopenthelidmanually.

Firstturnoffthepowerswitch;Anemergencydooropeningholeisfoundontheleftpanelofthemachineatthetopfrontcorner(shownas2inthebelowfigure)

Page 24: Biovac 13R Operating Instructions - Vacutec

24 Biovac 13R

Ascrewdriverwithadiameteroflessthan4mmoranyotherhardtool(shownas1inthebelow figure)canbeinsertedintotheholeandpushedfirmlytoreleasethedoorlock.Oncethelidsprings up,thelidcanthenbeopenedbyhand. Afterthelidhasbeenopened,themanualopenerwillautomaticallybereset.Thedoorcanonce againbeusedundernormaloperationonceelectricalpowerisrestored. Attention:thismethodistobeusedonlyinanemergencyandnotasaregularoccurance.

7.2 Fault alarm Message

Thefollowingtableslisttehalarmmessagesthatthemachinedisplaysandtherelevanterror explanationsandtroubleshootingsolutionstoassisttheuserinclearinganerror.

Ifusersareunabletocleartheerror,orthealarmmessagethatthemachinedisplaysisnotlistedin thefollowingtables,usersshouldimmediatelycontactsupportpersonnelatVacutec.

Fault Alarm Message Table

Fault ErrorexplanationandtroubleshootingsolutionsStartandnodisplayor sudden no display

1.Thereisnoelectricityorthecableisnotconnectedcorrectly.2.Inspectpowersocketandconnectionwhethertheyareingoodcondition;inspectpowersocketwhetherithaspower.3.Inspectpowerswitchandensurethereisgoodcontact4.Iftheerrorcannotbecleared,pleasecontactVacutec’ssupportpersonnel.

The Machine suddenly stops during a run

1.Exceededmaximumratedspeedofrotor:Oncerotorspeedexceedsitsratedspeed250r/min,overspeedalarmwillbesounded.Themachinemustbestoppedtoresetthespeed;2.Exceededthesetspeedofrotor:3.InvertermotoroverheatedorCompressoroverheated,thepowersupplyoftheinverterandcompressorinthemachinewillshutoff,sothemachinestopsrunning.4.Chamberovertemperature:Checksettemperaturewhetheritiswithin40ºC;checkpaneldisplaytemperaturewhetheritexceeds40ºC.Ifitistheformer,settemperatureshouldbelowerthan40ºC.Ifitisthelatter,chambershallbecooledtoroomtemperature.Butinanyevent,themachineshallstopcompletelyfirstandthenberestartedaccordingtotherequirement;5.Ifkeyboardpanelhasnodisplay,pleaseinspectthepowersupplysystemofthemachine;6.Voltagemaypossiblybetoolow,inspectpowersupplyvoltagewhetheritmeetstherequirement.

Thelidcannotbeopened

1.Therotorhasnotstoppedrunningcompeletlythereforethedoorwillnoopen.2.Openthedoormanuallyoncetherotorhasstoppedspinning);3.Iftheerrorcannotberesolved,pleasecontactVacutec’ssupportpersonnel.

Page 25: Biovac 13R Operating Instructions - Vacutec

25Biovac 13R

The machine vibrationisexcessive

1.Therotorisexceedingcriticalspeedandsomevibrationofthemachineisnormal;2.Inspecttherotorwhetheritislockedinplace.3.Inspectthesymmetryofrotorloadandensurethemachineislevel.4.Ensuretherotorisinstalledproperly;5.Inspectdriveshaftandrotateitbyhand;ifdriveshaftdoesnotrotatesmoothly,driveshaftormotormaypossiblyhaveafault.Iftheerrorcannotbecleared,pleasecontactVacutec’ssupportpersonnel.

Turnonpowerswitch,buttheindicatorlightondisplaypanelisoff.

1.Powersupplyisnotconnected;inspectthedistributionpowersupply.2.Iftheproblemcannotberesolved,pleasecontactVacutec’ssupportpersonnel.

Operating panel displaysabnormally

Itmaypossiblybecausedbygridinterference.Pleaseshutdownandwaitaminute,thenrestartit,anddisplayshouldbenormal.

Pressstartbutton,butrotor does not run

1.Shutdownandthenrestart2.Iftheerrorstillcannotbecleared,pleasecontactVacutec’ssupportpersonnel.

Refrigeration fails or the machine does not reach the preset temperature

1.Ambienttemperatureistoohigh;reduceambienttemperaturetobelow35ºC;2.Temperaturesettingisincorrect;resettheparameter.3.Iftheerrorcannotberesolved,contactVacutec’ssupportpersonnel.

Themachinegivesoffsmoke&oraburntodour

1.Shutoffpowersupply;2.Rotatedriveshaftbyhand;ifdriveshaftcannotberotatedsmoothly,itmayhaveafault.3.Iftheerrorcannotberesolved,contactVacutec’ssupportpersonnel.

Forfaultsnotmentionedabove

PleasecontactVacutec’ssupportpersonnelforassistance.

7.3 Safety protection

7.3.1 Automatic Stop protection function

• Safetyprotectionfor“Overspeed”:Iftherotationspeedoftherotoris250rpmabovetehratedspeed, theautomaticstopprotectionfunctionfor“Overmaxspeed”willcomeintoeffect.Theprompt message“OVERMAXSPEED”appearsandthebuzzersoundssimultaneously.

• Safetyprotectionfor“Overheat”:Ifthetemperatureinthechamberexceeds40ºC,theautomatic

stopfunctionfor“Overheat”willcomeintoeffect.Thepromptmessage“tempover40ºC”appears andthebuzzersoundssimultaneously.

• Whenruntimecompletes,thedevicestopsautomatically,andtheruntimeisdisplayedas“00:00”

Page 26: Biovac 13R Operating Instructions - Vacutec

26 Biovac 13R

• Procedureinthecaseofabnormaloperation:Ifthereisabnormalvibration,noiseorapeculiarsmell,shutdownthethedevice,switchofftheelectricity,removeanysamplesanddoaninspection.PleasecontactVacutecifthecausecannotbefound.

8. Technical Data

Function/Parameter Technicalindicators

Operating environment

•Environmentfreeofdustandvolatile orcorrosivegases;novibrationorairflowrestrictions to hinder the performance

•Altitude:≤2000m•Relativehumidity:≤80%

Ambienttemperature •+5ºC–+35ºCAppliedvoltage •AC230V,50/60HzRated power •480VA

Typesoffusesandparameter•F1,F2:250VAC,10A;•F3,F5:250VAC,0.5A;•F4,F6:250VAC,2A

Settimerange •1min-99h59minSettemprange •-10-40ºCSetspeedrange •300-13800rpm(canextendto15000rpm)Rated/Maximumcentrifugalforce(RCFmax) •17,459xg/22,388xgAscendingvelocity •Max(highest)ascendingvelocity,<5minDescendingvelocity •Minimum(lowest)descendingvelocity,<5min

Refrigerationeffect

•Whentemperatureofthecentrifugalchamberisequaltothatoftheambienttemperature,±15ºC-20ºC,atemperatureof-5ºCinsidethechambershouldnotbemaintainedforlongerthan30minat the rated speed

Completemachinevibration •vibrationofthecentrifugeshouldn’texceed0.1mm

Precisionofsustainedrotaryspeed •±1%

Timeaccuracycentrifuge •Timeaccuracyofcentrifugedoesnotexceed±1%

Noise(atthemaximumspeed) •Lessthan58dBatthemaximumspeedExternaldimensions(device) •525mm(L)*300mm(W)*350mm(H)Netweight •47kgExternaldimensions(Packaging) •660mm(L)*420mm(W)*630mm(H)

Page 27: Biovac 13R Operating Instructions - Vacutec

27Biovac 13R

Roto

r par

amet

erVe

sselparam

eter

adap

ter

RotorN

o.(Rated/M

axsp

eed)

Roto

r typ

e(centrifu

geang

le)

Capacity

(ml)

Botto

m

figure

dim

ensio

ns(m

m)

leng

th(m

m)

Num

/se

tNam

eOrderNo.

Num

/se

t

Max

ce

ntrif

ugal

ra

dius

(cm)

Max

cent

rifug

al

forc

e(xg)

#2201

(13800/1500rpm)

angl

e ro

tor

(45º)

1.5/2

ø11

4724

none

8.2

17459/20627

0.5

18ºcon

icø8

3224

1x0.5m

lad

apte

rADP1

303-001

247.2

15330/18112

0.2

18ºcon

icø6.5

2224

1x0.2m

lad

apte

rADP1

303-002

246.4

13626/16099

Roto

r par

amet

erVe

sselparam

eter

adap

ter

RotorN

o.(Rated/M

axsp

eed)

Roto

r typ

e(centrifu

geang

le)

Capacity

(ml)

Botto

m

figure

dim

ensio

ns(m

m)

leng

th(m

m)

Num

/se

tNam

eOrderNo.

Num

/se

t

Max

ce

ntrif

ugal

ra

dius

(cm)

Max

cent

rifug

al

forc

e(xg)

#2202

(13800/1500rpm)

angl

e ro

tor

(45º)

8x0.2

18ºcon

icø6.5

228

none

6.8

15330/18112

0.2

18ºcon

icø6.5

2264

none

6.8

15330/18112

Roto

r par

amet

erVe

sselparam

eter

adap

ter

RotorN

o.(Rated/M

axsp

eed)

Roto

r typ

e(centrifu

geang

le)

Capacity

(ml)

Botto

m

figure

dim

ensio

ns(m

m)

leng

th(m

m)

Num

/se

tNam

eOrderNo.

Num

/se

t

Max

ce

ntrif

ugal

ra

dius

(cm)

Max

cent

rifug

al

forc

e(xg)

#2203

(13800/1500rpm)

angl

e ro

tor

(45º)

16x5

ø13.5

5216

none

8.9

18949/22388