Biovac 13ROperating Instructions
2 Biovac 13R
Operating environment of instruments
In order to extend the machine’s service life, please pay attention to the following factors which may damage the machine:
• EnvironmentalConditions;• Effectsofenvironment,includingnaturalultravioletradiation(ie:DirectSunlight)• Erosionandabrasionoftheprotectivecoverandothersafetyparts.
• Indooruse;• Altitude:<2000m;• Ambienttemperature:+5°C~+35°C;• Relativehumidity:<80%;• Operatingvoltage:230V,50/60Hz;• Pleaseensureadequateventilationinareaofuseandappropriatesealingmeasurewhenperformingthe separationofpoisonoussubstancesandpathogenicmicro-organismcells;• Ensurethereisnovibrationand/orastrongdrafttoinfluencetheperformance.• Ensurethereisnodustorflammableandcorrosivegaspresentintheambientair.
Please read this manual carefully when using the instrument for the first time!The maintenance of the instrument should be carried out by the agents authorized by Biovac.
The following materials should not be used in conjunction with this instrument:
• Flammableandexplosivesubstances;• Strongchemicalmaterials;• Toxicorradioactivesubstances,• Pathogenicmicroorganisms
TheBiovac13RHighSpeedRefrigerationBench-topCentrifugecanbemaintainedonlybyqualifiedserviceengineersusingappropriatetools.
Shouldtherebeaproblemnotmentionedinthismanual,pleasecontactaBiovacauthorizedagentforthecorrectprocessingmethods.
OnlyaccessoriesprovidedbyVacutecshouldbeused.Ifotheraccessoriesareused,Vacutecwillnotbeheldresponsibleforanydamageorinjurytoinstrumentoroperatorasaresult.
TheBiovac13RHighSpeedRefrigerationBench-topCentrifugeshouldbeinspectedandmaintainedatstatedintervals.
3Biovac 13R
1. Safety clauses
1.1 Operating requirements Biovac13RHighSpeedRefrigeratedBench-topCentrifugeisdesignedaccordingtopresenttechnical
standards.Thereforeitmustbeusedaccordingtothedesignrequirements.WhentheBiovac13RHighSpeedRefrigeratedBench-topCentrifugeisusedaccordingtothefollowingimproperorinappropriatemethods,equipmentmybedamagedortheoperatorinjured. • Notusedinaccordancewiththedesignrequirements/intendeduse. • Theoperatorandmaintainerhavenotbeentrainedinitscorrectuseandoperation. • Impropermodificationstothedesignwithoutauthorisation. • Failuretoreadtheregulationsforsafeuse.
PersonswhouseormaintaintheBiovac13RHighSpeedRefrigeratedBench-topcentrifugeshouldreadandunderstandrelatedusageandsafetyregulationsinthisOperatingManual.
Thefollowingregulationsshouldbeobeyedtoavoidaccidents: OperatingManualshouldbekeptneartheBiovac13RHighSpeedRefrigeratedBench-topCentrifugeforreferencebyoperators.
Biovac13RHighSpeedRefrigeratedBench-topCentrifugeisdesignedtoseparatebiologiccells,organismsandbacteria.Thedensityofthesampleshouldnotexceed1.2g/cm3atthemaximumspeed.
Duringoperation(whenthesampleisbeingseparatedorrotorsarerunning),operatorsanddangeroussubstancesshouldmaintainadistanceofatleast30cmfromthecentrifuge.Theventilationholeofthecentrifugeshouldnotbecloggedbysubstances.
IfthefollowingsafetyprecautionsarenotobeyedwhenusingtheBiovac13RHighSpeedRefrigeratedBench-topCentrifuge,operatorsorotherpersonsmaybehurtorthecentrifugeandsamplesmaybedamaged:
• Thecentrifugeisnotcorrosionprooforimplosionproof,thereforeitshouldbeusedinacorrosion andimplosionfreeenvironment.
• Thefollowingmaterialsareprohibited: Flammableandexplosivesubstances: Strongchemicalmaterials; • Toxicorradioactivesubstances,orpathogenicmicroorganisms. • Corrosivesubstancesmaydamagematerialsinsidethecentrifugeandtherotors,thereforethey shouldbeputinsidesealedcontainerswhileseparating.
1.2 Precautions • Beforerunning(whenseparatingsamples),appropriaterotorsmustbefixedfirmly; • Duringoperation(whentherotorisrunning)orafterthecentrifugehasbeenstopped(therotoris
stillrunning),openingthelidbyhandisstrictlyforbidden. • AccessoriesmustbeprovidedandapprovedbyVacutec.Somecommonparts,suchasglassand plasticcontainersforseparating,shouldbeVacutecapprovedandbeinaccordancewiththe requirementsofthehighestspeedandmaximumcentrifugalforceofrelatedrotors; • Donotusethecentrifugeorseparatesampleswhilethelidisopen; • Changingmechanicalpartsandelectronicsshouldbecarriedoutbyqualified Vacutecpersonnelonly; • Whenusingthecentrifuge,usetheappropriaterotorsaccordingtoload.DoNotoverload; • Maintenanceshouldbecarriedoutperiodically.See“Cleanandpurify”chapterformore information.
4 Biovac 13R
2. Introduction2.1 Layout of the Biovac 13R High Speed Refrigerated Bench-top Centrifuge
Front and Back view of Centrifuge
Descriptionofabovediagrams: 1. Frontcasingcover 2. ControlpanelanddigitalLCD 3. Lid
4. Observationhole 5. TopPanel 6. Frontventilation 7. Emergencyopen/Forceopen 8. Foot 9. Rearcasingcover 10. Rearventilation 11. PowerjackandOn/Offswitch
2.2 General Biovac13RHighSpeedRefrigeratedBench-topCentrifugewithmultifunctionprotectionintroduces
advancedtechnologysuchasspeedadjustmentviafrequencycontrol,microcomputercontrol,touchbutton,dialsetup,digitalLCD.Threerotorvariationsareavailable(fordetails,see“Table1:Typeofrotorandtechnicalparameters”).Thiscentrifugeisusedfortheseparationofbiologiccells,organismsandbacteria.
2.3 Device structure 2.3.1 The Device Lid
Onlywhenthelidlocksarelockedonboththeleftandrightofthelidmaythecentrifugerun; otherwisethewarningsystemwillsoundanalert(alightwillcomeonandabuzzerwillsound) andthedevicewon’trun.
Toopenthelidpress“”(openbutton)onthecontrolpanel.Oncethelidrisestoacertain height,thelidhingeandspringswillsupportthelid. Ifapowercutoccursortheopenbuttonisnotresponding,thesamplemustberemoved.Referto
Speed/F
S/F
Temp. Time.
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
5Biovac 13R
chapter7.1foremergencyopeninginstructions. Whiletherotorsarerunningand/orthepowersupplyison,DONOTopenthelidmanually.
2.3.2 The Chamber Thechambermaintainsastableworkingtemperature.
2.3.3 The Rotor Fordetailssee“Table1:Typeofrotorandtechnicalparameter”.Therotorrotatesata predeterminedspeedtocreatetherelativecentrifugalforcerequiredtoseparatevarioussamples. Thecentrifugalforceisseveralthousandtimesmorethang(accelerationofearthgravity),itis therefore important to use and maintain the rotor carefully and safely!
2.3.4 The Refrigeration System Therefrigerationsystemisusedtocreatealow-temperatureoperatingenvironmentforthe centrifugechamber.Thetemperatureisautomaticallycontrolled.Thecentrifugeshouldbe operatedinanenvironmentwithaconstantambienttemperaturebetween5ºCto35ºCandan ambienthumidity≤80%.Optimumoperatingconditionscanbeachievedwithadequate ventilationaroundthecentrifugeinaroomwithambienttemperature controlledbyair-conditioning. Aheatsinkisinstalledinthefrontofthemachinewhereairflowandcirculationisessential.Any dustshouldberemovedfromtheradiatorwithabrushorcompressedaironceeverythree monthstoensureeffectiveheatdissipation.Thedigitaltemperaturesensorinstalledatthebottom ofthechamberisusedtomonitorthetemperatureinthechambercontinuously.
Thetemperatureinthechambermayvaryfromthatofthesamplesatfirst,andtake aperiodtoequalise.
2.3.5 The Control System The control system allows the setting of the rotor speed, running time, temperature and control of displaysystemandwarningsystemoftheentiremachine.Toensureproperrunningofthe machineandthesafetyoftheoperator,DONOTopenthefrontcasingofthedevice,asthemain controlcircuitboardandotherelectronicsareinstalledontothefrontcoverofthedevice.
2.3.5 The Display System Thedisplaysystemcanshoweverysettingparametersimultaneously,andcantrackanychange inrealtime.Inaddition,itdisplaysrelativesystemerrorsandwarnings.
2.3.5 The Warning System Thewarningsystemwillalerttheuserof“overspeed”,deviceoverheating,lidwarning,etc. Analertisindicatedbyaflashing,red,warninglightonthecontrolpanelandabuzzerwillsound simultaneously.Ifrunning,thedevicewillbestoppedautomaticallyandtheuserwillbeunable tostartasequenceagainuntiltheerrorhasbeenresolved.NOTE:usermaypressthe“Esc”button onthecontrolpaneltosilencethewarningbuzzer.
2.4 Safety Protection Device
TheBiovac13RHighSpeedRefrigeratedBench-topCentrifugehasaseriesofsafetyprotectiondevices:
• Themountingrackismadeofsteelplatingandtheprotectionsteelringismadeofsteelcasting.Theinternalbladderoftheinnerchamberisstainlesssteel.ThefrontframeismadeofABS.Thelidisexplosionproof.Thereisaviewportinthecenterofthelid.
6 Biovac 13R
Onlywhenthecentrifugeisonandtherotorhascompletelystopped,willthe“”buttonopenthelid.
• OverspeedStop Thewarningbuzzerissoundedwhentherotorspeedexceedsthesettingspeedby200r/min.Therotorstopsautomaticallyifthespeedexceedsthemaximumratedspeedby250r/min.Waituntiltherotorstopscompletely,thenopenthelid.Restartoncetheerrorhasbeenresolved.
• Openingisstrictlyforbiddenduringoperation. If“”isaccidentlypressedduringoperation,thedevicewon’trespond,however,thewarningalarmis
indicated.
• Openinginthecaseofanemergency Iftheelectricityisofforthereisanurgentproblemwiththerotor,youmayopenthelidmanually
whenitisnotpossibletodosobypressingthe“”button(ENSURETHEROTORHASSTOPPEDCOMPLETELYFIRST,TOAVOIDINJURY).
DONOTUSETHEEMERGENCYOPENINGFACILITYUNNECESSARILY!
2.5 Technical Specifications
2.3.1 Technical Data
2.6 Settling the Centrifuge
2.6.2 Thisdeviceshouldbeputonaswing-outtable-boardofappropriaterigidity,andbefreefrom vibrationandawayfromanydirectheatsourceorsunlight.
2.6.2 Thereshouldbeaspaceof10-15cmaroundthedevicetoallowforheatemissionfromthe condenser.
2.6.3 Aftermounting,adjustthe4feetlevelstoensuretherequiredsupportonthetable-board.
2.6.4 RequiredPoweroutputforthepowersupply:AC230V,50/60Hz.
THEDEVICESHOULDALWAYSBEEARTHED.
2.7 Configuration
Thefollowingaccessoriesaresuppliedwiththedevice.(Seepackinglistfordetails)
Youshouldinspectthecontentswhenreceivingproductsandunpacking.Ifthereissomethingwrong,pleasecontactVacutec.
Speed/F
S/F
Temp. Time.
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
Speed/F
S/F
Temp. Time.
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
Speed/F
S/F
Temp. Time.
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
7Biovac 13R
3 Available optional accessories
ThestandardconfigurationoftheBiovac13RHighSpeedRefrigeratedBench-topCentrifugeincludes: 1xCentrifugeandaNo.#2201rotor.
Accessoriesalsoavailableonenquiryare: RotorsNo.#2202and#2203(Fordetails,see“Table1:Typeofrotorandtechnicalparameters”)
Table1:Typeofrotorandtechnicalparameters
TypeofRotor
Rotor Number
Angleoftheangle rotor
(º)
Rated/Maxspeed (r/min)
Maximum capacity (ml)
Max centrifugal radius(cm)
Rated/Maximum centrifugal
acceleration (xg)
AnglerotorFS15.02 #2201 45º 13,800/
15000 24x2.0 8.2 17,459/20627
AnglerotorFS15.002 #2202 45º 13,800/
15000 64x0.2 7.2 15,330/18112
AnglerotorFS15.05 #2203 45º 13,800/
15000 16x5.0 8.9 18,949/22388
8 Biovac 13R
4.Preparationsbeforeuse4.1Transportationandinstallation Biovac13RHighSpeedRefrigeratedBench-topCentrifugeistransportedinspecial-purposecartonwith
protectionmaterialsforbufferinginside.Afteropening,theprotectionmaterialsforbufferingshouldberemoved.
• Thenetweightofthedeviceisabout47kg,thereforeitisnecessarytoliftthedevicefrombothsides(theleftandright)tokeepitsbalance.Thefrontcovershouldnotbepressedonorusedforlifting.
• Keeplevelandavoidvibrationwhentransporting• Pleaseusespecialpackingcasesduringlong-distancetransportation.Keeplevelandhandlewith
care!
4.2ProperplacementoftheCentrifuge Biovac13RHighSpeedRefrigeratedBench-topCentrifugeisforindooruse,andtheworkareashould
meetthefollowingrequirements:
• Asafetyclearanceof30cmmustbemaintainedaroundthecentrifuge.Whilerunning,thereshouldbenoindividualorhazardoussubstanceincloseproximitytothecentrifuge.
• Theadapterortabletopshouldbesolidtoavoidvibration;ifmobileracksortrolliesareused,theyshouldbeequippedwithlockingdevicestoensurethesafeoperationofthecentrifuge.
• Ifthecentrifugeisplacedbesidethewallorinacorner,thereshouldbeaspaceof10cmbetweenthebackofthecentrifugeandthewalland15cmbetweeneachsideandthewalltoallowsufficientventilationandheatemission.
• Thecentrifugeshouldbeplacedawayfromwindowstoavoidtheheatfromdirectsunlight.• Thefoursupportinglegsofthecentrifugeshouldstandevenlyona“swing-out”table-board.• Thecentrifugeshouldbeputinacleanroomwithaconstantambienttemperaturebetween5ºCand
35ºCandanambienthumidityoflessthan80%
4.3Fixedplacementofequipment Afterthecentrifugehasbeenplaced,don’tmoveitfrequently.Alwaysensurethecentrifugestandslevel
onthetable-board.Theshouldbenovibration.
Toavoiddamagetotherotoraxis,pleaseremovetherotorwhenmovingthecentrifuge.
4.4Properconnectionofelectricsupply
9Biovac 13R
Theeachoftheelectricplugconnectedtothecentrifugemustbeingoodcondition.Thevoltageandfrequencyofthepowersupplyshouldbeinaccordancewiththerequirementsstatedinthetechnicaldescriptionorspecificationslabelledonthenameplateofthecentrifuge.Thepowerjack/switchisattherearofthemachine,“I”foronand“O”foroff.Onlyusethepowercableincludedwiththemachine.
Neverpress“▷/☐”or“QTemp”buttonimmediatelyafterstartup;thedoormustbeopenedfirsttoremovetherotorfixingexpandedpolyethylene(EPE)androtor,andthenthechamberandrotormustbeinspectedandcleanedbeforerunning.
5. Direction for use and operation5.1 Control panel and display interface
ShownasFig.6,controlpanelisdescribedinbriefasfollows:
1-Datadisplaywindow 2-Cancel/Exitbutton 3-Quickaccessbuttonofcommonparameters 4-Menubutton 5-Quicktemperaturebutton 6-Dooropeningbutton 7-Inchingbutton(quickstart;pressthebuttonandrotorspeedwillquicklyincreasetoitsmaximum allowablespeed;releasethebuttonandrotorwillquicklystop) 8-Start/Stopbutton 9-Dial(Selectandmodifysettingparameters;actsas“Enter”buttonwhenpressedin) 10-Speedandcentrifugalforcesettingbutton 11-Curvesettingbutton 12-Temperaturesettingbutton 13-Timesettingbutton
10 Biovac 13R
ShownasFig.7,maininterfaceisdescribedinbriefasfollows:
1. Thetoprowisthestatepromptarea.“”indicatesdoorisinlockedstate;“”indicatesdoorisin unlockedstate;“”indicatesthebuzzerisinnormalworkingstate.
2. Thesecondrowisthespeed/centrifugalforcedisplayarea:theleftismaindatadisplayandthe rightiscorrespondingauxiliarydatadisplayafterconversion.Whentherotorspins,real-timespeed isdisplayed;whentherotorstops,“0”isdisplayed.Whencentrifugalforceisauxiliarydataandthe rotorisnotidentified,displayedcentrifugalforceis“-----”.
3. Thethirdrowisrunningtimeandchambertemperaturedisplayarea:whentherotorisnotrunning, displayedrunningtimeis“00:00”;whentherotorisrunningthedisplayedvalueistheremaining runtime.Displayedtemperatureisreal-timechambertemperaturevalue.
4. Thefourthrowissetparameter/alarmmessagedisplayarea:whenthesystemisinnormaloperation thedisplaysare:SetSpeed/Centrifugalforce;Setruntime;Settemperature;SetcurveandCurrent identifiedrotornumbersuccessivelyfromlefttoright.Whenasystemerroroccurs,therelevant errormessageisdisplayed.
5.2 Start Connectthepowercablesuppliedwiththemachineintothepowersocketontherearpanelofthe
machineandconnecttheotherendintomainssupply.ThepowersupplyusedforthemachineisAC230V,50/60Hz.Oncepluggedin,turnonthepowerswitchnexttothepowersocket.Thebuzzerwillsoundandthesoftwareversionwillbedisplayed.Oncetheself-checkiscompleted,itwillenterthemaininterface(asshowninFig.7),thenextoperationcannowbeconducted.
11Biovac 13R
5.3 Open Whilethepowerswitchisturnedonandtherotorhascompletelystopped,press“”button(door-
openbutton)onthecontrolpaneland“”willappearinthedisplaywindow;thelidwillautomaticallyopenviaspringaction,andthelidneedstoberaisedtherestofthewayandleanedbackwardagainsttheframe;theinnerchamberwillnowbevisibleandtheiconrepresentingdoorstatusonthemaininterfacewillbedisplayedasunlocked.
ATTENTION:whenthedoorisopened,thelidwillautomaticallybounceuptoacertainheight,keepclearofthelidduringthisoperationtoavoidinjury.Inthecaseofanerror,wherethedoorcannotbeopenedautomaticallyandthesamplesinthechamberneedtoberemoved,thedoorcanbeopenedmanuallybythemethodstatedinchapter7.1
5.4 Close Presstheliddownuntilthedoorhookontheliddrivesthelockingdeviceandyouhearthelockclick;
whentheicondisplaysas“”,itindicatesthatthedoorhasbeencorrectlyclosedandlocked. LidAdjustment:whenclosingthelid,ifyoufeelthelidistootightortooloose,unscrewthelocknutonthedoorhookwitha17mmsocketwrenchandthenadjustthedoorhookappropriately;turnclockwisetotightenandanti-clockwisetoloosen.Aftercompletingtheadjustment,tightenthelocknutagain.
Whenclosingthedoor,anappropriateforceshouldbeappliedtopresstheliddowntoavoiddamage to the door lock.
5.5 Installation/Dismounting of the rotor Ifinappropriaterotorandadapterareused,itwillcauseanunsatisfactoryseparationeffectand
could also damage the centrifuge.
TherotorusedmustbeaproductconfirmedandappointedbyVacutec,otherwiseitmaypotentiallycauseafault.ThespecificationandmodelofalltypesofrotorsconfirmedbyVacutecareprovidedinthismanual(fordetailsonspecificationandmodel,seeTable1“RotorTypeandTechnicalParameters”).
Wheninstallingrotors,usethe5mmhexagonspannersuppliedalongwiththemachine. Installation steps for rotors are as follows:
• Turnonpowerswitchandwaitforself-checkcompletion. • Press“”buttonandopenthelidofthecentrifuge;makesurethechamberisclean andimpurityfree;
• Cleanexternalsurfaceofmotorshaftandinspectthesealringonthemotorshaftensuring itisintact; • Preparetherotoryouwanttouse;holdtherotorwithbothhandsandalignthecenterholeofthe rotorwiththemotorshaft,andthenverticallyloweritontotheshaft; • Screwonthelocknutbyhandandthenusethe5mminnerhexagonspannertotighten thenutclockwise; • Rotorinstallationiscompleted.
Beforeuse,especiallybetweenprolongedperiodsoftime,theseatingoftherotorshouldbeinspectedforanychanges;whennecessary,thelocknutshouldbere-tightenedtoensuretherotorissecure.
12 Biovac 13R
5.6Maintenanceofmotorsupportshaft Maintenancestepsofmotorsupportshaft: • Turnonpowerswitchandwaitforself-checktocomplete • Press“”buttonandopenthelidofthecentrifuge;thendismounttherotorusing5mminner
hexagonspanner.NOTE:tightenclockwiseandloosenanti-clockwise. • Removedust,impuritiesandseparatedsampleresidueinthecentrifugechamberensuring
itisclean. • Cleantheconicalsectionandthescrewthread(showninFig.8)ontherotatingshaft,ensuringno
dirtordustremains;thenevenlyapplythelubricatinggreasesuppliedalongwiththemachine.
5.7 Calculation of rotor load Calculationofmaximumload:
Biovac13Rdeskcentrifugewillgenerateahighcentrifugalforceduringhigh-speedrunning;eachrotorisdesignedtohaveenoughmechanicalstrengthtowithstandthisforcewhenworkingattheratedmaximumspeed,buttoavoidaccidentorinjury,therotorloadmustnotexceeditsmaximumratedload.
Whenseparatingsamples,thesamplesarefirstplacedintoacontainerandthenplacedtogetherintoa rotor, the sum total of the container and sample must not exceed maximum rated load of the rotor, otherwiseyoumustreducethesampleweightorcalculatepermissiblerunningspeedofrotor(NPERM)toensuretherotorloaddoesnotexceeditsmaximumratedload.
Calculationmethodofpermissiblerunningspeed(NPERM)ofrotorisasfollows: NPERM=NmaxX(maximumpermissibleload÷actualload)0.5
13Biovac 13R
Where: Nmax=maximumratedspeed; Maximumpermissibleload=maximumpermissibleload; Actualload=actualload
Don’tusetherotorunderoverloadedconditions,asthismayresultinseveredamageorinjurytotheequipment or operator.
5.8 Sample filing in centrifugal container Whenthecentrifugeruns,ifbalancedcorrectly,separationzoneofsampleswillnotinterferewitheach
otherduetotheabsenceofvibration,resultinginanimprovedcentrifugaleffect.Therefore,samplesshouldbeplacedintothecentrifugalcontainerassymmetricallyaspossibletoachievethebestbalanceduringthesequence.
Carefullyinspectthecentrifugalcontainer(centrifugetube,etc)beingused,ensuringitisinaccordancewiththeallowablemaximumratedacceleration(centrifugalforce);alwaysusethelowestrunning speed possible.
Pay strict adherence to the service life of the centrifugal container in use; especially when frequently runningundermaximumpermissibleload(speedandtemperature). Ifdamaged,itshouldbereplacedimmediately!
5.9 Safe use of rotor • Beforetherotorruns,sampleshouldbeloadedcorrectlyandsymmetrically.
• Whenexchangingarotor;usethe5mmspannertoloosenthelocknutanti-clockwiseand removetherotor. • Neverstartthemachinebeforethelocknuthasbeentightenedontotherotatingshaft. • Ifthecentrifugewillberunrepeatedly,therotorlocknutmustbeinspectedaftereachruntoensure itisstilltight.Ifthelocknutisloose,itmustbetightenedbeforeproceedingwiththenextrun. • Centrifugaltesttubesdonotneedtobeloadedsimultaneously,butdoneedto beloadedsymmetrically.
5.10 Parameter setting ATTENTION:
• Beforesettingparametersviathecontrolpanel,therotormustbecorrectlyinstalledon therotatingshaft. • Buttonsaretouchsensitive.Simplypresstherelevantbuttonfortherequiredoperation. • Iftheparametersareinsertedincorrectly,theycanbereset.
14 Biovac 13R
Turnthedialwhileinthestandbyinterfacetoentertheparametersettinginterface;continuetoturnthedialtoswitchbetweentheparametersthatneedtobeset;whentherespectiveparameteritemisselected,agreenborderwillflickeraroundit.Afterselectingtherequiredparameter,pressthedialintoconfirmyourselectionandentertherequiredparameter.
5.10.1 Setting of Speed:
SettingrangeofspeedofBiovac13Rcentrifuge:minimumspeed:300rpm;maximumspeed:13800rpm.
Description: 1. Whileinthestandbyinterface,turntehdialuntilthespeedsettingzoneflickersasshowninFig.10.
Press[S/F]buttontoquicklyselecttheappropriateparametersetting. 2. Pressthedialtoconfirmandenterthespeedsettinginterface;onceintheinterface,press[S/F] buttontoswitchbetweenspeedselectionandcentrifugalforceselection.Theselectedparameter itemhasaflashinggreenborderindicatingthatthenumericalvaluecanbemodifiedandset;non- selecteditemsaregreyedout,shownasFig.10.Iftherotorisnotidentified,centrifugalforce displayis“-----”. 3. Turnthedialtochangethenumericalvalue.Speedissetbysteppedvariableadjustment:change invalueis100ateachturnofthedial.Iftherotorhasbeenidentified,thecorrespondingvaluewill beautomaticallyadjustedrelativetothevalueinsertedforthedesiredspeed. 4. Adjusttotherequiredspeedvalueandpressthedialtosavethesetting;youcanalsopressthe“Esc” buttontocancelthechangeandreturntothestandbyinterface.
15Biovac 13R
5.10.2 Setting of the centrifugal force:
SettingrangeofcentrifugalforceofBiovac13Rcentrifugerotor:minimumcentrifugalforce:100xg;maximumcentrifugalforce:22380g.
Settingrotorspeedcanbechangedbysettingthe“centrifugalforce”value;thespeedandcentrifugalforcewillnotcorrespondintheirrelationuntilaftertherotorhasbeenidentified(namelytherotornumber),centrifugalforceandrotorspeedwillthenconvertinrelationtoeachotherandthecurrentcentrifugalforcewillbedisplayed.
Settingmethod(asper5.10.1SettingofSpeed):
Calculation of centrifugal force Relative centrifugal force is usually thousands of times more than terrestrial gravity and it is the unit usedtomeasuretheefficiencyofdifferentkindsofinstrumentsseparatingorprecipitatingarticles.Calculationofcentrifugalforceisrelatedtocentrifugalspeedandcentrifugalradius;centrifugalforceiscalculatedbytheequationbelow: RCF=11.18X(n/1000)2Xr Where: r=centrifugalradius(cm) n=centrifugalspeed(rpm)
Attention: maximum centrifugal force value is only related to maximum speed and maximum centrifugal radius. The centrifugal force value “X g” displayed in the interface is calculated according to current rotor speed and maximum centrifugal radius.
16 Biovac 13R
5.10.3 Setting of speed-up and speed-down curves:
Speed-upandSpeed-downtimecontrolofBiovac13Rcentrifuge:3groupsofspeed-upcurvesand3groupsofspeed-downcurvesareprovidedrespectivelyaccordingtodifferentrotorsandtheycanbefreelyarrangedingroups.3#curveismaximumspeedand1#curveisminimumspeed.
Settingmethod(asper5.10.1SettingofSpeed): 1. Turnthedialwhileinthestandbyinterfacetoentertheparametersettinginterface. 2. Inthecurvesettingmenuinterface,thecurvesettingboxisdisplayedbyagreenflashingborder. 3. Orsimplypress[Curve]buttontoquicklyselectthatparametersetting.
4. Turnthedialtochangethevalueofthecurvesetting 5. Thefirstrowofcurvesettingmenuisthespeed-upcurveandthesecondrowisthespeed-down curve.Pressthe[Curve]buttontoswitchbetweenthesetwooption.Adjusttothepreferedcurve andpressthedialtosavethesetting;orpressthe“Esc”buttontocancelthechangeandreturnto standbyinterface.
5.10.4 Setting of temperature:
17Biovac 13R
ChambertemperaturecontrolofBiovac13Rcentrifuge:minimumtemperatureaccessiblevarieswitheachtypeofrotoratdifferentspeeds.Settingrangeoftemperatureforthecentrifugeis-10~40ºC. Settingmethod(asper5.10.1SettingofSpeed) 1. Turnthedialwhileinthestandbyinterfacetoentertehparametersettinginterface. 2. Inthetemperaturesettingmenuinterface,whenthetemperaturesettingboxisselectedisdisplaysa flashinggreenborder. 3. Orsimplypress[Temp]buttontoquicklyselectthatparametersetting. 4. Turnthedialtochangethevalueofthetemperaturesetting. 5. Thesecondrowintemperaturesettingmenuisaswitchingdisplayindicatingchambertemperature andtubetemperature(displayedsampletemperatureisestimatedvalue,notactualvalue).“Chamber temp”ischambertemperatureand“tubetemp(simulation)”issampletemperature.Press[Temp] buttontoswitchbetweenthetwomodes.Adjusttotherequiredtemperatureanddisplaymode andpressthedialtosavethesetting;orpress“Esc”buttontocancelthechangeandreturnto standbyinterface.
5.10.5 Setting of running time:
SettingrangeofruntimeofBiovac13Rcentrifugerotor:1min~99h59min. Inthetimesetting,hour[H]andminute[M]canbesetrespectively. Settingmethod(asper5.10.1SettingofSpeed) 1. Turnthedialwhileinstandbyinterfacetoentertheparametersettinginterface.Onceinthetime settinginterface,tehtimesettingboxisdisplayedasselected,indicatedbyaflashinggreenborder. 2. Settingsequenceoftime:minute->hour->saveandexit.Thewhitenumericalvalueiscurrent adjustedparameter.Oncethevaluehasbeenadjusted,pressthedialtoselectthehoursetting.Press [Time]buttontoquicklyselectparametersetting. 3. Turnthedialtoswitchbetweenthevalues. 4. Thesecondrowoftimesettingmenuisruntimemodeoption;“countdown@start”iscountdown atstartuptime;“countdown@sets/f ”iscountdownaftercurrentspeedreachesthesetspeedor presetcentrifugalforce.Press[Time]buttontoswitchbetweenthetwomodes. 5. Adjusttherequiredtimeanddisplaymodeandpressthedialtosavethesetting;orpress“Esc” buttontocancelthechangeandreturntostandbyinterface.
18 Biovac 13R
5.10.6 Reading and storage of user setting data
Biovac13Rcentrifugeprovides3groupsofsettingdatasavedforusers.Userscansaveandreadtheirmostcommonparametersettings. Settingmethod(asper5.10.1SettingofSpeed): 1. Press[P1]/[P2]/[P3]buttonwhileinthestandbyinterfacetoquicklyenterthe parametersettinginterface. 2. Turnthedialinthesettingmenuinterfacetoswitchthe“Read/Save”function;theselecteditemis indicatedbyaflashinggreenborder. 3. Whenswitching“Read/Save”,correspondingdataisdisplayedintheupperparameterdisplaybox. Forexample,whenselecting“loadfrompreset”,tehdatadisplayedintheupperdisplayboxis thedatathattheuserhaspreviouslysaved;pressthedialandthedataisreadouttocovercurrent setparametervalue;whenselecting“savetopreset”,thedatadisplayedintheupperdisplayboxis thecurrentdata;presstehdialtosavethedata.Press[P1]/[P2]/[P3]toswitchbetweenthe 3groupsofdata. 4. Adjusttherequiredtimeandpressthedialtosavethesetting;orpress“Esc”buttontocancelthe changeandreturntostandbyinterface.
5.10.7 Parameter change during the run process
Duringtherunofthecentrifuge,thespeed,temperature,time,centrifugalforceandotherparameterscanbechanged;oncenewparametersareentered,displayscreenreturnstomaininterfaceandthenewparameterscomeintoforce.
19Biovac 13R
5.11 Start centrifugal separation Oncethesettingofuserparametersiscomplete;rotorshavebeeninstalled;thesampleforseparation
hasbeencorrectlyinserted;thelidofthemachinehasbeencorrectlyclosed,thenthemachinecanbestartedtobegintheseparation.
Attention: Indication whether the lid of the machine is closed correctly is displayed on the main interface via the door lock status icon.
5.11.1 Press “ “ to start
1. Pressthestartbutton““onthecontrolpanelandtherotorwillfirstbeidentifiedata lowspeed;onceidentified,thecorrectsetspeedorcentrifugalforcewillbeenforced. 2. Therotorwillbeacceleratedtothesetspeedaccordingtothesetspeed-upcurve. 3. Runtimeisdisplayedby“countdown”methodduringtherunprocess;whenruntimedisplay reacheszero,namely“00:00”,thecentrifugestopsandtheoperationiscomplete 4. Chambertemperaturequicklydropsatthestartofthecompressor;oncethechamberreachestheset temperature,thecompressorswitchesoff. 5. Real-timespeedandcentrifugalforcearedisplayedduringtherun;displayedcentrifugalforceis acquiredviaautomaticconversioninrelationtotherotordetectedbythecentrifuge. 6. Ifrotorspeedorcentrifugalforceexceedsitsmaximumpermissiblevalueorsetvalueduring therun,analarmwillsound;whenmotortemperatureprotectionistriggeredorcompressorover voltageprotectiontriggers,analarmwillsoundandprompttheusertomakethe relevantprocessingadjustment.
Attention:fordetailsonalarmpromptmessagecontent,see“Troubleshooting”. 7. Thelidofthemachinemustnotbeopenedwhiletherotorisstillturning.
5.11.2 Press “ “ button to start
Pressandhold““buttonandthesystemwillidentifytherotor;onceidentified,therotorspeedwillquicklyincreasetopermissiblemaximumspeedonthemachine;release““buttonandtherotorwillquicklyslowuntilitstopscompletely.
5.12 Stop Separation
Oncethecentrifugebeginsitsrun,itcanbestoppedbythetwomethodsbelow: 1. Undernormalconditions,theruntimeofthecentrifugehasbeenpreset,andwillautomaticallystop
aftercountingdowntozero.
2. Ifyouwanttostopthecentrifugeduringtheprocessofanormalrun,pressthestopbutton““onthecontrolpanelatanytime,andthecentrifugewillstoprunningimmediately.TheCountdowninthedisplaywindowindicatingtheruntimewillstopsimultaneously.
Onlyoncetherotorspeeddropstozero,i.e.speeddisplaywindowdisplays“0rpm”andcentrifugalforcedisplaywindowdisplays“0xg”,youcanpresstheopenbutton““onthecontrolpanelandthelidwillpopopenviaaspringaction,thelidcanthenbeopenedbyraisingitbyhand;thedoorlockstatusicononthemaininterfacedisplaysthelockasopen,theseparatedsampleinthecentrifugecannowberemoved.
20 Biovac 13R
5.13 Temperature Control
5.13.1 Pre-Refrigeration Function
Whenthelidofthecentrifugeisclosedandpriortherunningofthepresetprogramme,oraftercentrifugalseparationiscomplete,thecentrifugewillenter“pre-refrigerationmode”accordingtothesettemperature.Inthisway,itcanreachtherequiredtemperaturepriorcentrifugalseparation;andmaintaintheappropriatetemperatureafterseparationtoprotectsamples.Ifambienttemperatureorchambertemperatureishigherthanthatofthesample,usersareadvisedtopre-refrigerateandtheninsertsamplesforcentrifugalseparation,otherwisethe“TEMPOVER40ºC”warningmayappear.
Attention:Pre-Refrigerationfunctionwillstopafter1hour;Itisresetbyopeningandclosingthedooragainorduetopowerfailure.
5.13.2 Quick temperature function
Thecentrifugehasaquicktemperaturefunction.Firstsetthetargettemperatureparameterandexpectedtimedurationandmakesuretherotoriscorrectlyinstalledandlocked;closethelidandthenpressthe<Qtemp>buttontostartthequicktemperaturefunction,andtherotorwillspin.
Ifsettemperatureishigherthanchambertemperature,thecompressordoesnotstartandheatisgeneratedbyfrictionbetweenthehigh-speedrunningrotorandair,toquicklyreachthetargettemperature.
Ifsettemperatureislowerthanthechambertemperature,therotorrunsatthespecifiedspeed;thecompressorwillstartandtherefrigerationsystemstartstoquicklyreducethechambertemperaturetoreachthetargettemperature.
Afterthetemperaturereachesthesetvalue,therotorcontinuesrunning,butmainparameter<Qtemp>ondisplayscreenwillflickerandsimultaneouslybuzzersoundsintermittentlytoremindtheuser
21Biovac 13R
thatquicktemperatureiscomplete.Pressthestopbuttonandthequicktemperaturefunctionwillbeterminatedandtherefrigerationandrotorwillstop.
5.13.3 Temperature control in the separation process
Whenthecentrifugeisstarted,thetemperaturecontrolsystemwillrunforaperiodoftimeuntiltherequiredtemperatureisreached.Thetemperatureinthecentrifugalchamberwillthenbestablewithin±2ºCrangeofsettemperature.
5.14 Running record
In order to ensure normal operation of the machine and extend the service life of the machine, and accumulateinformationaboutthemachineoperationtoassistinthediagnosisoffaultsoraccidents,pleasecarefullykeeparecordofeachrun(onebookforeachmachine).
Belowistherecommendedoperationrecordtable:
RECORD TABLE
Date of test
Name of sample
Weight of monotube (including sample)
Rotor No.
Rotory
speedTime Temperature Working
condition Operator
22 Biovac 13R
6. Maintenance
6.1 Cleaning and sterilization
Usershavetheresponsibilitytocleanandpurifythecentrifugewhendangeroussubstancesspilloverintothemachine.Usersshouldcleanandpurifyitaccordingtothemethodrecommededbythemanufacturerstoensurethatthemachineisnotdamaged.
Inappropriate detergents and incorrects methods of cleaning may do harm to the centrifuge or internal parts. Cleaning and sterilization should be carried out according to operating procedures described in the operating and technical instructions.
6.1.1 Implementation of cleaning and sterilization
Turnoffthepowerandremovethepowercablebeforecleaning!
Thecentrifugeshell,rotorchamber,rotorandcentrifugalcontainersusedandsoonshouldbecleanedandmaintainedregularly;thuscontaminationcanbeclearedavoidingdamagetothepartsusedandapollutedenvironment.Ifyouarenotclearonwhichcleaningagentsshouldbeusedforcleaningthecentrifuge,pleasecontactVacutec.
• Organicsolventsareforbidden,astheycandegreasethebearingsofthemotor; • Liquid,especiallyorganicsolventsshouldnotcomeincontactwiththesupportingshaftandball
bearingswhenbeingcleaned; • Ifthereisfrostorathinicelayer,theyshouldbewipedawayoncetheyreverttowater.
6.1.2 Implementation of high-pressure sterilization
• Rotorsusedbythiscentrifugecanendurehigh-pressuresterilization.Temperature:121ºCwith apressureof250kPa,maintainedforatleast10minutes.Pleaserefertotheinstructionsofferedby manufacturersvariousseparatingcontainers.
• Alwaysconfirmwhetheritispermissibletosterilizerotorsandcontainerswithhigh-pressure sterilizationmethod.Thetemperatureanddurationofhigh-pressuresterilizationshouldbestrictly controlledaccordingtothedifferentrotorsandcontainersusedforseparating;
• Dontusearotororcontainerforseparatingiftheyarecorrodedordamaged.
6.1.3 Maintenance
6.1.3.1 Donotbringtherotorintocontactwithsharpobjects.Whenmovingordisassemblingthe centrifuge,youshouldtakecarenottodamagetherotor,asscratchesandindentationsmaycause cracksduringoperation.
6.1.3.2 Regularlycheckforspots,groovesorminutecracksintherotor(especiallyatthebottomofthetest tube).Ifanyarefound,stopusingthemimmediatelyandcontactVacutec.
6.1.3.3 Therotorshouldbecleanedonceeveryweekundernormaloperation.Rinsetherotorimmediately ifitsisusedtoseparatesalinesolutionorothercorrosivesamples.Ifliquidissprinkledontothe rotor;blotitupandrinsethepartimmediately.
23Biovac 13R
6.1.3.4 Pleasecleanwithspongesorcottonclothdippedinneutralcleaningsolution,thenrinsethe cleaningsolutionoffwithdistilledwater.Don’tsprinkleorspraywaterontotherotor;becauseit maycausetherotortocorrodeduetoresidualwater.Inverttherotorandallowittodry.
6.1.3.5 Dustthecoolingplatesofthecondensatoronceeverythreemonths.
6.1.3.6 Checktheprotectivecover
6.1.3.7 Inspectionofrotorpartsandsafetyconditions.
6.1.3.8 Confirmthedeviceisadequatelyearthed.
6.1.3.9 Removeanywaterorcondensationfoundinthechamberimmediately.
6.1.3.10 Turnofftheelectricityanddisconnectthepowercablebeforedisassembingthefrontcasingofthe device.Attention: This should be carried out by a trained Vacutec Technician only.
6.1.3.11 OnlysparesprovidedbyVacuteccanbeusedinthisdevice.
6.1.3.12 Turnofftheelectricitywheneverthecentrifugeisn’tinuse.
6.1.3.13 TransportationandStorage Thismachineisaprecisioninstrument.Water,vibration,tiltingandinvertedplacementarestrictly
forbiddenduringtransportationandstorage.
Please note the protection indicators on the packaging!
Fragile This Side UP Keep Dry Do Not Tilt
7. Trouble shooting
7.1 To open lid in case of emergency
Undernormalconditions,thefunctiontoautoopenthelidwillbedeactivatedwhentheelectricityisoffortheoperationtoopenthelidisdisabledbyaccident.However,samplesinthechambermayneedtoberemoved.Inthiscase,youmayopenthelidofthecentrifugemanuallytotakesamplesout.
Attention:Thisisforemergencyonly!
Intheeventofapowerfailure,thebrakewillnotengage,thereforetherotorwillcontinuetospinforaprolongedperiod.Pleasewaitforittostopcompletely(observethroughtheobservationholeattehtopofthelid)beforeattemptingtoopenthelidmanually.
Firstturnoffthepowerswitch;Anemergencydooropeningholeisfoundontheleftpanelofthemachineatthetopfrontcorner(shownas2inthebelowfigure)
24 Biovac 13R
Ascrewdriverwithadiameteroflessthan4mmoranyotherhardtool(shownas1inthebelow figure)canbeinsertedintotheholeandpushedfirmlytoreleasethedoorlock.Oncethelidsprings up,thelidcanthenbeopenedbyhand. Afterthelidhasbeenopened,themanualopenerwillautomaticallybereset.Thedoorcanonce againbeusedundernormaloperationonceelectricalpowerisrestored. Attention:thismethodistobeusedonlyinanemergencyandnotasaregularoccurance.
7.2 Fault alarm Message
Thefollowingtableslisttehalarmmessagesthatthemachinedisplaysandtherelevanterror explanationsandtroubleshootingsolutionstoassisttheuserinclearinganerror.
Ifusersareunabletocleartheerror,orthealarmmessagethatthemachinedisplaysisnotlistedin thefollowingtables,usersshouldimmediatelycontactsupportpersonnelatVacutec.
Fault Alarm Message Table
Fault ErrorexplanationandtroubleshootingsolutionsStartandnodisplayor sudden no display
1.Thereisnoelectricityorthecableisnotconnectedcorrectly.2.Inspectpowersocketandconnectionwhethertheyareingoodcondition;inspectpowersocketwhetherithaspower.3.Inspectpowerswitchandensurethereisgoodcontact4.Iftheerrorcannotbecleared,pleasecontactVacutec’ssupportpersonnel.
The Machine suddenly stops during a run
1.Exceededmaximumratedspeedofrotor:Oncerotorspeedexceedsitsratedspeed250r/min,overspeedalarmwillbesounded.Themachinemustbestoppedtoresetthespeed;2.Exceededthesetspeedofrotor:3.InvertermotoroverheatedorCompressoroverheated,thepowersupplyoftheinverterandcompressorinthemachinewillshutoff,sothemachinestopsrunning.4.Chamberovertemperature:Checksettemperaturewhetheritiswithin40ºC;checkpaneldisplaytemperaturewhetheritexceeds40ºC.Ifitistheformer,settemperatureshouldbelowerthan40ºC.Ifitisthelatter,chambershallbecooledtoroomtemperature.Butinanyevent,themachineshallstopcompletelyfirstandthenberestartedaccordingtotherequirement;5.Ifkeyboardpanelhasnodisplay,pleaseinspectthepowersupplysystemofthemachine;6.Voltagemaypossiblybetoolow,inspectpowersupplyvoltagewhetheritmeetstherequirement.
Thelidcannotbeopened
1.Therotorhasnotstoppedrunningcompeletlythereforethedoorwillnoopen.2.Openthedoormanuallyoncetherotorhasstoppedspinning);3.Iftheerrorcannotberesolved,pleasecontactVacutec’ssupportpersonnel.
25Biovac 13R
The machine vibrationisexcessive
1.Therotorisexceedingcriticalspeedandsomevibrationofthemachineisnormal;2.Inspecttherotorwhetheritislockedinplace.3.Inspectthesymmetryofrotorloadandensurethemachineislevel.4.Ensuretherotorisinstalledproperly;5.Inspectdriveshaftandrotateitbyhand;ifdriveshaftdoesnotrotatesmoothly,driveshaftormotormaypossiblyhaveafault.Iftheerrorcannotbecleared,pleasecontactVacutec’ssupportpersonnel.
Turnonpowerswitch,buttheindicatorlightondisplaypanelisoff.
1.Powersupplyisnotconnected;inspectthedistributionpowersupply.2.Iftheproblemcannotberesolved,pleasecontactVacutec’ssupportpersonnel.
Operating panel displaysabnormally
Itmaypossiblybecausedbygridinterference.Pleaseshutdownandwaitaminute,thenrestartit,anddisplayshouldbenormal.
Pressstartbutton,butrotor does not run
1.Shutdownandthenrestart2.Iftheerrorstillcannotbecleared,pleasecontactVacutec’ssupportpersonnel.
Refrigeration fails or the machine does not reach the preset temperature
1.Ambienttemperatureistoohigh;reduceambienttemperaturetobelow35ºC;2.Temperaturesettingisincorrect;resettheparameter.3.Iftheerrorcannotberesolved,contactVacutec’ssupportpersonnel.
Themachinegivesoffsmoke&oraburntodour
1.Shutoffpowersupply;2.Rotatedriveshaftbyhand;ifdriveshaftcannotberotatedsmoothly,itmayhaveafault.3.Iftheerrorcannotberesolved,contactVacutec’ssupportpersonnel.
Forfaultsnotmentionedabove
PleasecontactVacutec’ssupportpersonnelforassistance.
7.3 Safety protection
7.3.1 Automatic Stop protection function
• Safetyprotectionfor“Overspeed”:Iftherotationspeedoftherotoris250rpmabovetehratedspeed, theautomaticstopprotectionfunctionfor“Overmaxspeed”willcomeintoeffect.Theprompt message“OVERMAXSPEED”appearsandthebuzzersoundssimultaneously.
• Safetyprotectionfor“Overheat”:Ifthetemperatureinthechamberexceeds40ºC,theautomatic
stopfunctionfor“Overheat”willcomeintoeffect.Thepromptmessage“tempover40ºC”appears andthebuzzersoundssimultaneously.
• Whenruntimecompletes,thedevicestopsautomatically,andtheruntimeisdisplayedas“00:00”
26 Biovac 13R
• Procedureinthecaseofabnormaloperation:Ifthereisabnormalvibration,noiseorapeculiarsmell,shutdownthethedevice,switchofftheelectricity,removeanysamplesanddoaninspection.PleasecontactVacutecifthecausecannotbefound.
8. Technical Data
Function/Parameter Technicalindicators
Operating environment
•Environmentfreeofdustandvolatile orcorrosivegases;novibrationorairflowrestrictions to hinder the performance
•Altitude:≤2000m•Relativehumidity:≤80%
Ambienttemperature •+5ºC–+35ºCAppliedvoltage •AC230V,50/60HzRated power •480VA
Typesoffusesandparameter•F1,F2:250VAC,10A;•F3,F5:250VAC,0.5A;•F4,F6:250VAC,2A
Settimerange •1min-99h59minSettemprange •-10-40ºCSetspeedrange •300-13800rpm(canextendto15000rpm)Rated/Maximumcentrifugalforce(RCFmax) •17,459xg/22,388xgAscendingvelocity •Max(highest)ascendingvelocity,<5minDescendingvelocity •Minimum(lowest)descendingvelocity,<5min
Refrigerationeffect
•Whentemperatureofthecentrifugalchamberisequaltothatoftheambienttemperature,±15ºC-20ºC,atemperatureof-5ºCinsidethechambershouldnotbemaintainedforlongerthan30minat the rated speed
Completemachinevibration •vibrationofthecentrifugeshouldn’texceed0.1mm
Precisionofsustainedrotaryspeed •±1%
Timeaccuracycentrifuge •Timeaccuracyofcentrifugedoesnotexceed±1%
Noise(atthemaximumspeed) •Lessthan58dBatthemaximumspeedExternaldimensions(device) •525mm(L)*300mm(W)*350mm(H)Netweight •47kgExternaldimensions(Packaging) •660mm(L)*420mm(W)*630mm(H)
27Biovac 13R
Roto
r par
amet
erVe
sselparam
eter
adap
ter
RotorN
o.(Rated/M
axsp
eed)
Roto
r typ
e(centrifu
geang
le)
Capacity
(ml)
Botto
m
figure
dim
ensio
ns(m
m)
leng
th(m
m)
Num
/se
tNam
eOrderNo.
Num
/se
t
Max
ce
ntrif
ugal
ra
dius
(cm)
Max
cent
rifug
al
forc
e(xg)
#2201
(13800/1500rpm)
angl
e ro
tor
(45º)
1.5/2
ø11
4724
none
8.2
17459/20627
0.5
18ºcon
icø8
3224
1x0.5m
lad
apte
rADP1
303-001
247.2
15330/18112
0.2
18ºcon
icø6.5
2224
1x0.2m
lad
apte
rADP1
303-002
246.4
13626/16099
Roto
r par
amet
erVe
sselparam
eter
adap
ter
RotorN
o.(Rated/M
axsp
eed)
Roto
r typ
e(centrifu
geang
le)
Capacity
(ml)
Botto
m
figure
dim
ensio
ns(m
m)
leng
th(m
m)
Num
/se
tNam
eOrderNo.
Num
/se
t
Max
ce
ntrif
ugal
ra
dius
(cm)
Max
cent
rifug
al
forc
e(xg)
#2202
(13800/1500rpm)
angl
e ro
tor
(45º)
8x0.2
18ºcon
icø6.5
228
none
6.8
15330/18112
0.2
18ºcon
icø6.5
2264
none
6.8
15330/18112
Roto
r par
amet
erVe
sselparam
eter
adap
ter
RotorN
o.(Rated/M
axsp
eed)
Roto
r typ
e(centrifu
geang
le)
Capacity
(ml)
Botto
m
figure
dim
ensio
ns(m
m)
leng
th(m
m)
Num
/se
tNam
eOrderNo.
Num
/se
t
Max
ce
ntrif
ugal
ra
dius
(cm)
Max
cent
rifug
al
forc
e(xg)
#2203
(13800/1500rpm)
angl
e ro
tor
(45º)
16x5
ø13.5
5216
none
8.9
18949/22388
28 Biovac 13R