Page 1: Biomedical Device Design Report

1

Project title: Performance improvement of an Active Ankle

Foot Orthosis (AAFO) using muscle signal feedback.

Tejaswita Patil

NUID: 03922699

Supervisor name: Dr. Carl Nelson

Page 2: Biomedical Device Design Report

2

Abstract

The ability to walk and move around is often taken for granted. Lifestyles are drastically changed when one loses

the ability to get around, not on just a personal level but also on the familial and friends level. It has been shown

that subjects with stroke have decreased walking speed, asymmetry in gait pattern, muscular weakness,

abnormal movement strategies, and spasticity. Individuals who have these abnormalities are more likely to have

risks in falling which may lead into serious physical health issues. Dynamic active-ankle orthosis (AAFO) attempt

to reproduce normal knee motions during the whole gait cycle. In this research experiment, a new type of AAFO

with a series elastic actuator (SEA) mechanism is going to be developed which utilizes muscle nerve impulses

as the control system. This biomedical device will consists of SEA that will power the locomotion to those that

have difficulties in walking/movement in general. This biomedical device will be equipped with electromyography

(EMG) sensors that can sense the muscle impulses. The muscle impulses will send a signal that will then be

read by a 16 bit ADC and amplified for better detection and processing. A controller will then calculate and feed

the output to the motor driver, which in turns will run the motor. The power from motor will then be transmitted

through the SEA, thus moving the ankle.

Page 3: Biomedical Device Design Report

3

Introduction:

The ability to walk and move around is often taken for granted. Lifestyles are drastically changed when one loses

the ability to get around, not on just a personal level but also on the familial and friends level. Moving around,

yet alone being able to stand, can be lost in numerous different ways, but stroke is the second leading cause of

loss of mobility [1-2]. It has been shown that subjects with stroke have decreased walking speed, asymmetry in

gait pattern, muscular weakness, abnormal movement strategies, and spasticity [3-5]. Individuals who have

these abnormalities are more likely to have risks in falling and may lead into serious physical health issues. Not

only do people who experience stroke and lose functions in movement, but these individuals develop energy

inefficient patterns of walking, if they are still able to walk [4].

To aid in the movement to those that have loss mobility from stroke, there are currently numerous medical

devices that enhance mobility. These devices are collectively called foot orthosis, which may be sub categorized

into ankle foot orthosis (AFO) or knee-ankle-foot orthosis (KAFO). In general, a FO brace is designed to aid in

the control of biomechanical alignment, correct or accommodate deformity, protect and support an injury, assist

rehabilitation, reduce pain, increase mobility of the foot, as well as increase independence of the patient.

To date, several types of AFOs have been developed: passive AFOs, semi-dynamic KAFOs (SDAFOs), and

dynamic AFOs (sometimes referred to as active) [7]. Passive AFOs work without any power source and are the

most primitive type of AFOs. SDAFOs lock the knee joint during the stance phase. Dynamic AFOs attempt to

reproduce normal knee motions during the whole gait cycle. There are two types of dynamic AFOs that have

been reported in the literature. The first one is activated by using a pneumatic system and the second one which

uses a spring mechanism. Both systems are bulky and are controlled through complex control systems that limit

their application as assistive devices.

Dynamic AFOs have been recommended for this group of subjects [6-8]. However, the effect of various AFO on

the performance of these subjects during walking depends on the structure of orthosis and their ankle mechanism

[6-9]. In previous studies, Gok et al. showed that the gait of hemiplegic subjects improved more with metal

orthosis compared to plastic AFO, due to its better stabilization during stance phase

A common problem of stroke subjects during walking is contributed to the ankle joint plantar flexion power. Ankle

joint plantar flexion power is decreased significantly compared to normal subjects [5]. Various attempts have

been done to restore plantar flexion moment in subjects with stroke by use of externally powered orthosis [8, 11-

19]. The idea of incorporating active ankle mechanism in design of AFO has been fantasized since 1980.

It was recently shown that patients who lack complete muscular control can learn to control slow cortical potential

to operate e.g. an electronic spelling device [1-20 Charles]. It has been shown that motor imagery can modify

Rolandic mu and central beta rhythms in a task-specific way [2-21] and therefore, mental imagination of certain

movements can serve to produce self-induced variations of the electroencephalogram (EEG) [22-25]. Our goal

is to develop a lightweight powered orthosis for the human lower limb that could comfortably provide plantar

flexion and dorsiflexion torque during walking.

Page 4: Biomedical Device Design Report

4

In the first design of this orthosis, hydraulic and pneumatic actuators were used to power the ankle joint. Up to

now various approaches have been used to restore the power of ankle joint including: Series elastic actuator

(SEA), magneto rheological fluid, passive pneumatic element, frictional clutch, oil damper, artificial pneumatic

muscle, shape memory alloys [26-32]. On top of all above mentioned mechanisms, SEA provides more

advantages, due to the low controlled impedance, good control bandwidth, shock tolerance, low friction.

However, most of the studies that have looked at these mechanisms neglect precise control system according

to natural body function. As in, patients require to control the movement of their ankle by themselves rather than

operating it by separate computer and operator.

The project that our group wants to focus on is to improve an assisting device for patients with disabled or semi-

disabled ankle movement. It is generally referred to as “Drop foot”. “Drop foot” is the inability of an individual to

lift their foot because of reduced or no muscle activity around their ankle. In this research experiment a new type

of AFO with SEA mechanism and EMG sensor is going to be developed that utilized muscle nerve impulses as

the control system. The feasibility and validation of this new design will be investigated in this study.

Page 5: Biomedical Device Design Report

5

Product Specifications :

The product is required to produce plantar flexor torque and

dorsiflexor torque. These values depend on the weight and size

of the patient. From the previous researches we can conclude

that it will require peak plantarflexor torque of 70 Nm and peak

dorsiflexor torque of 38 Nm [33].

Another feature of walking is gait. Gait is a cyclical pattern of

leg and foot movement that creates locomotion. Gait is

commonly discussed in terms of a percentage of a single gait

cycle. A gait cycle is defined for a single leg and begins with

the initial contact of the foot with the ground or ‘heel strike’;

the conclusion of cycle occurs as the same foot makes a

second ‘heel strike’. To illustrate a typical pattern of gait,

consider the illustration of the ankle complex during stance

phase of a single cycle of gait, figure 1 and the kinematics

and kinetics of a normal ankle, figure 2. Notice that in figure

2, peak ankle moment occurs at roughly 45% of the gait cycle

and at a normalized value of -1.25 Nm/kg. The negative sign

represents the physiological direction of the plantarflexing

ankle complex. The foot rotates downwards to push off from

the ground. At the point at which the peak moment occurs,

the ankle angle begins a rapid descent to its lowest overall

value of -24 degrees at 60% of the gait cycle. The region of

gait approximately between 45% and 60% of the gait cycle

is known as ‘push off’. At the conclusion of ‘push off’, now

considered ‘toe off’, the leg initiates ‘swing’ and the foot is then positioned for the next ‘heel strike’.

Electronic Parts Specifications:

1. 16 ADC and Amplifier: ADS1115 16-Bit ADC - 4 Channel with Programmable Gain Amplifier

Description: For microcontrollers without an analog-to-digital converter or when you want a higher-precision

ADC, the ADS1115 provides 16-bit precision at 860 samples/second over I2C. The chip can be configured as 4

single-ended input channels, or two differential channels. As a nice bonus, it also includes a programmable gain

amplifier, up to x16, to help boost up smaller single/differential signals to the full range. This ADC was chosen

because it can run from 2V to 5V power/logic, can measure a large range of signals and its super easy to use.

It is a great general purpose 16 bit converter.

2. SEA and Motor

Figure1. Stance phase of a single gait cycle. 60-100% of gait is the swing phase, not shown.

Figure 1 Normal Ankle Gait: Kinematics and Kinetics [34].

Page 6: Biomedical Device Design Report

6

Description: The Series Elastic Actuator (SEA) was developed for the current version of this device. SEA includes

a coupling, metal plates, four compression springs, eight bushings, one ball screw, one ball nut, and four guide

rails the compression springs were used in this design to minimize the effect of backlash, to isolate motor from

shock loads and torque ripple. The ball screw and the ball nut convert rotations of a motor into linear motion.

The SEA generates enough force for the movement of ankle based on the control signal of the motor. The SEA

consists of a 100W AC servo motor with maximum speed of 5000 rpm (XML-SA01A, LS Mecapion). (FIG)

3. Controller and Motor Driver:

Description: A master controller (PCI1710 and PCI 1723, Advantech) is used to detect the signal from the

ADS1115 16-Bit ADC and produced output for the Industrial motor driver (XDLL740A001, LS Mecapion) which

in turns rotate the motor with required speed and torque.

4. EMG Sensor: Muscle Sensor v3 (SpurkFun)

Description: Measuring muscle activity by detecting its electric potential, referred to as electromyography (EMG).

This sensor will measure the filtered and rectified electrical activity of a muscle; outputting 0-Vs Volts depending

the amount of activity in the selected muscle, where Vs signifies the voltage of the power source.

Parameter Min TYP Max

Power Supply Voltage (Vs) ±3V ±5V ±30V

Gain Setting, Gain = 207*(X /1 kΩ) 0.01 Ω (0.002x) 50 kΩ (10,350x) 100 kΩ (20,700x)

Output Signal Voltage (Rectified & Smoothed) 0V -- +Vs

Differential Input Voltage 0 mV 2‐ 5mV +Vs/Gain

5. Sensor Pads: Sensor Cable - Electrode Pads (3 connector) [CAB-12970ROH]

Description: This is a simple three conductor sensor cable with electrode pad leads. These cables are 24" long

and feature a 3.5mm audio jack connector on one end with snap style receptacles for biomedical sensor pads.

Each cable has a red/blue/black set. FIG

Review of Existing Products

An example of a passive AFO that is currently out on the market is make by Townsend, Whose design is

composed of the following: 1 3/8 inch wide solid core graphite uprights and 1 3/8 inch of 2 inch wide solid core

graphite bands with radius edges, polypro foot plate trimmed to metatarsal joint heads with reinforced stirrups,

removable &; replaceable soft interface AK & BK liners, knee joint covers protective coating removable &;

replaceable soft interface strap pads. A figure of Townsends® standard KAFO is shown in figure 3. Townsend

Premier Series AFO says that it will maximize patient functional mobility. The patient will also achieve the

secondary benefit of wearing a brace that enhances their comfort and lifestyle. Townsend’s solid core graphite

shell designs are combined with patented knee and ankle hinge technology. Resulting in rigid, durable bracing

solution that achieves more effective and consistent control. Ultra-light, extremely low profile, and cosmetically

Page 7: Biomedical Device Design Report

7

appealing, these optimize patient compliance and satisfaction. Townsends® KAFO works in

the way that the patient must insert their leg into the device and then secure the device with

the provided straps. The device wearer will then have aided support in their movements. The

patient’s movement will be restricted in a manner that will strengthen their current way of

walking, in order to correct it to a more normal walking pattern. This device has its downfall

because some patients may not be able to move their lower part of their body easily, thus

will need more aid in movement. Thus, we may need to explore more dynamic AFOs.

The dynamic AFO that we’d like to talk about is a design that was proposed by Cullel et al.

whose group developed an electrically controlled dynamic AFO with compression spring

actuators to duplicate the normal knee stiffness pattern during the whole

walking gait cycle and to mimic the quadriceps and hamstrings muscle group

behavior [34]. The knee actuator includes: a stance control (SC) stiff spring,

a swing control soft spring, a solenoid, a control circuit, a shaft that both springs are wrapped

around, an upper block connected to the thigh segment, and a lower

block attached to the two compression springs. This can be seen in the

figure 4.

Technology gap to fill

As mentioned in our review of existing products, we wish to improve

current products our proposed dynamic AFO is innovated and

revolutionary due to the fact that we will incorporate SEA and EMG

sensors in order to obtain the muscles electrical impulses generated

from the patient's muscles. We believe that this method will allow

independence and comfort. Our proposed device out classes current

state-of- the-art dynamic AFOS because the current dynamic AFOs require

ones to be connected to a computer in order to receive the information. Our

rationale for using EMG sensors with our dynamic AFO is to better simulate natural motion, as in there is not a

computer connected to a person who does not suffer from orthosis problems.

Functional Decomposition

From our functional decomposition, as seen in Appendix A, we see that the main functional goal is to improve

mobility for the user. From there we broke our main function into three sub-functions which are: to have the

desired mechanical properties, to have desired electrical attributes, and to mimic natural locomotion. We chose

this breakdown because our biomedical device has two integral components being the base component of the

device (SEA) and the EMG sensors. The base of the biomedical device must have the desired mechanical

properties, namely: being able to support load and provide proper torque. To further expand on supporting load,

our biomedical device must be able to support load from the entire weight of the individual, the knee, and the

ankle. This is critical, since the biomedical device we are proposing is intended to relieve some of the load, thus

allowing easier movement for the patient.

Figure 2: Generic KAFO

Figure 4: Stance Control diagram

Page 8: Biomedical Device Design Report

8

The next sub function from our main function that we want to discuss is to have the desired electrical properties.

This sub function is critical in our biomedical device, because our device will consists of EMG sensors. In order

for our biomedical device to work efficiently, we must have good signals, and more specifically we must be able

to process the signals and sense the muscle nerve impulses. Going back to having to the desired electrical

properties, we must be able to provide the proper electrical power to our device, which in turns allow regulation

of current to our motor; this all should be shown on a display.

Finally, the third sub function from our main function, improve mobility, is to mimic natural locomotion. This is a

key sub function. Because our biomedical device is supposed to aid in movement. It would be more desirable to

have a device that can let you “not stand out,” thus mimicking natural motion to alleviate the “Drop foot,” would

be best. To mimic natural motion, our biomedical device will have to be able to move the foot and ankle at

variable speeds and have flexibility.

Morphology Chart

Appendix B shows a well-organized brainstorming approach for designing our proposed AAFO. The main

purpose of this table is to create certain number of concepts and methods that can be achieved in the physical

domain. According to the proposed morphological chart for this biomedical device, there are three columns of

concepts and one column for sub functions. In the first column of this chart a list of sub functions has been written

and three concepts of design have been specified for each sub function. For instance, mechanical power

transmission is one of the most significant functions that can directly affect efficiency of device. There are three

different concepts of design for power transmission including ball screw, pneumatic cylinders, and cable or pulley

system. According to many factors like friction, weight, and dimension, designers should find the best option for

current device. As mentioned before, AAFOs must work with minimum amount of energy and should be as small

and light as possible. Consequently, using reasonable combination of concepts for different functions, three or

more designs will be created that must be evaluated using Pugh matrix with certain methods which will be

described later.

Interpretation of Pugh Matrix

As indicated in our appendix we have two iteration of our Pugh matrix. In order to analyze the Pugh matrix that

we constructed, we take the information gained from our morphology chart and compare the different concepts

to each other. Giving a +1 for a better design, 0 for an equivalent design, and -1 for a worse design when

compared to a given concept. With that being said, we can take column one (D1) as our datum and compare it

to column two (D2) & three (D3) and compare row by row we will to determine which design is best. It is good

to note that by default, the D1 summation will be zero (since we are using D1 as the datum). We thus compare

D2 and D3 with D1 as the datum, row by row.

We see that when looking at the sub function, support load from the knee, D2 scored a -1 and D3 scored +1.

We see that when looking at the sub function, support body weight, D2 scored a 0 and D3 scored -1. We see

that when looking at the sub function, support load from the ankle, D2 scored a 0 and D3 scored -1. An exhaustive

and explicit scoring can be found in the Appendix. We wish to summarize this and report that the D2 has a final

score of -2 and D3 has a final score of -10.

Page 9: Biomedical Device Design Report

9

We now do another iteration of the Pugh matrix analysis by making D2 the datum and compare row by row to

D1 and D3. Again, the explicit scoring can be found in the Appendix and we will simply report that D1 has a final

score of 9 and D3 has final score of -11. Since we see that D1>D2>D3 in the first iteration and D1>D3 in the

second iteration, we can skip the work for the third iteration of the Pugh matrix (setting D3 as the datum) and

come to the conclusion that D1 is best.

QFD

Quality function deployment (QFD) is a technique where one can find out how to strongly influence the quality of

a product, subassembly, or component. One of the most important outcomes of the QFD matrix is contained in

areas 7-8. Area 8 is a scaled version of area 7, where all the entries sum to 1 (or 100%). Thus area 8 represents

the relative importance (in percent) of each DP. This information can be used to focus the design efforts on those

areas which have the greatest capacity to increase quality as measured by FR outcomes.

With that being said, we see from our QFD matrix in Appendix G, the most important DP are the sensors. This

makes sense because the product is improved based on the addition and improvement of the sensor system.

The value in the improvement column for the good sensitivity / filtering is assigned as 2.9 out of 3.0 as this is a

considerable improvement of the sensor system. Finally the sensors becomes the most important DP in the

technical evaluation section with 19% relative importance. The second most important DP is the user interface

in this new design as a portable interface with LCD display has been added to the existing device replacing the

stationary PC based monitoring system. This gives it the relative importance of 13%. Apart from Sensors and

user interface, material type, motor power and surface coating are the most influential DPs that has higher

impacts on the customer needs. Material type is important for the mechanical properties, motor power is the key

source of actuation and surface coating is important for the user comfort, sensor placement and wear resistance.

We replaced some metal parts (Steel/Al alloy) with carbon fiber composite in the new design. This enables us to

reduce the weight while retaining the strength.

Ease of use, light-weight, wear resistance, endurance for cyclic load, robust design, reusability, good response

, biocompatibility are our most important design considerations as the device is supposed to be used every day

and in uncontrolled environment. With the EMG sensor, user interface and the additional electronic components

the new design is much effective, robust and user friendly.

Modeling of Parts/Design

One of the most important parts of this project is designing mechanical components, mechanisms and support

components like plastic orthosis. The following figures show decomposition of parts along with assembled

design. Since weight plays an important role in design of the orthosis, we as designers must optimize the design

to minimize the use of materials while l decreasing weight.

Therefore, weight will be considerably decreased which will in turn remove stress free parts in specific

components (four horizontal plates in the SEA). Design of plastic orthosis for supporting leg, ankle and foot must

customize for the specific patient. This customizable orthosis must be done using 3D imaging of foot... The

Page 10: Biomedical Device Design Report

10

minimum thickness plastic plate can be found using preliminary finite element modeling of plastic orthosis under

static and dynamic loading. In this design, the thickness is 5mm.

Materials

To choose the best materials for our biomedical device, we must look at the performance index for that specific

materials function. The first component that we want to look at of our biomedical device is the shaft. The objective

function for the shaft is to minimize the weight of the shaft and maximizing the yield strength. With that being

said, we looked at numerous types of metals, as indicated by our Ashby diagrams in Appendix C.

The most optimal material is then stainless steel, martensitic, AISI 440B, wrought, tempered at 316°C. Here we

can see that the stresses are induced due to tension, compression and torsion. Referring to the performance

indices on Table 7.2 located in “A Primer on Engineering Design of Biomedical Devices “by Dr. Carl Nelson,

where common performance indices for different types of materials, in various failure modes are mentioned, the

index for a shaft in bending = σ/ρ.

Looking at the Ashby diagram created by CES EduPack 2015 software (Granta Design, Cambridge, UK) and by

setting the slope of the line equal to one (since it is on a log scale) we are able to find materials that are

considered better or worse for the performance index we chose.. Since we need to maximize the performance

index, we shift the line towards the materials on the upper side of the diagram. When doing this, we see numerous

different materials that can be a “good” choice. Here we choose stainless steel, because stainless steel is readily

available and is economical (which was also shown in Gok et al. as mentioned earlier). Some advantages of

using stainless steel include: high strength, good wear resistance, and ductile energy absorption. Some of the

disadvantages include: corrosion and high density.

Figure 5: Exploded view of parts Figure 6: 3D Models

Page 11: Biomedical Device Design Report

11

The second material we want to look at is the foot plate. We need to minimize the weight of the foot plate while

maintaining its yield strength. The foot plate undergoes tension, compression and torsion. Similar to the process

as mentioned above, our group chose polypropylene as the material to be used for the footplate. The Ashby

Diagram for foot plate casting can be found in Appendix A, where we chose polypropylene to be the best material.

The third part we want to look at is the Lead screw (ball screw). Stresses induced in the lead screw are due to

tension, compression and torsion. Since this screw will be very frequently used as the force is applied by the

motor to the shaft through this screw, we need the screw to be resistant to torsion and bending. We use AISI

1055 alloy steel as the material with similar reasons to why we used AISI 440B.

The next material selection that we want to examine is for the joints. Depending on which joint, we choose

different types of materials. For one of the joints we chose to use 6061 aluminum for the joints (similar type of

material has been used in Patoglu et al [39]. Aluminum is used in this application because of its good strength

and hardness, corrosion resistance, ease of fabrication and other advantages. It is good to note that similar

materials like 7075-T6- aluminum and 303 stainless steel have been used in existing FO from Becker

Orthopedics for the connection joints [40].

To fabricate our biomedical device we must use pins to constrain some movement of the ankle joint and provide

a moment. Failure in pins occur due to shear primarily and bending. Since pins are relatively small part, we

chose to use standardized pins to follow the rules of DFA. As stainless steel pins are known for their strength

and availability, we use stainless steel pins in this application.

We will require fasteners to attach different parts in the SEA. Thus, fasteners should be strong enough to bear

the load of the motor and the person. Again, for the sake of DFA and availability, we purchase standardized

fasteners from the market.

Finally we perform material selection of the Actuator. We want the actuator to be as light as possible since it is

attached near the calf and can induce a lot of stress while walking. We use a Series Elastic Actuator for this

purpose.

The last material we want to look at is really more of the device that is composed multiple components and is

the heart of our biomedical device. The SEA, which was developed at the MIT Leg Laboratory, is used to power

the Active Ankle Foot Orthosis (AAFO) [41].

This actuator can be used to control the motion of the ankle in the sagittal plane. The SEA consists of a brushless

DC motor in series with a set of springs to provide series elasticity (Pratt and Williamson, 1995). By measuring

the deflection of the springs, the SEA provides force control through a position sensor. The deflection of the

springs is measured by a linear potentiometer sampled at 1000 Hz. This measurement is passed through a first

order filter with a cutoff frequency of 50 Hz. It is then numerically differentiated and passed through another first

order filter with a cutoff frequency of 8 Hz. The SEA is controlled through an internal feedback loop with a

Proportional-Derivative (PD) controller with gains of 10 and 15, respectively.

The advantages of the SEA is that it has low impedance, the motor is isolated from shock loads and the effects

of backlash, torque ripple and friction are filtered (by the spring). The SEA allows the implementation of any

Page 12: Biomedical Device Design Report

12

virtual, torsion mechanical element around the ankle. Specifically, the AAFO will implement a virtual linear torsion

spring or spring-damper system with varying values of stiffness and damping to accommodate the user.

DFM/DFA

For our device we want to use as least amount of as parts, while still achieving all of our functional requirements.

Therefore, we choose to remove extra parts or combine multiple parts to create multifunctional parts (ie: ball

screw system). Are medical device has many prefabricated and standardized parts readily available for purchase

(ie: AC servo motor, ball screw system, fasteners and vertical shafts).In our medical dice, the thickness of the

horizontal metal plates are standard to minimize manufacturing process. Our vertical shafts have standard

diameter, but must be cut based the patient. We acknowledge this in our DFM/DFA analysis, but nothing can

really be done since it is in our device’s nature to be formed for a specific patient. We also must customize the

four vertical plates for our device. This can be done by wire cutting or laser cutting, however these methods are

relatively expensive

The last guideline that ensure the most efficient design for manufacturing is to avoid secondary operations. Only

some of the parts that we need in our device require secondary operations, since most of our parts are standard

components. Most of the parts especially in SEA are symmetric that do not have very tight tolerances.

As seen in Appendix E, our DFA analysis yields that foot plate>motor>SEA> transmission>pins fasteners

>EMG> interface>battery has an assembly index of 358 while motor>SEA> transmission>pins>

fasteners>battery>EMG sensor>interface >Foot plate has an assembly index of 374. Thus, we conclude that the

first assembly method is a better avenue than the latter.

FMEA Analysis

FMEA is a systematic, proactive method for evaluating a process to identify where and how it might fail and to

assess the relative impact of different failures, in order to identify the parts of the process that are in need of

change. FMEA consists of identifying what failures might occur, predicting what the effect on the overall system

would be, and taking or recommending steps to prevent the failures. Three factors are considered for ranking

each failure mode: severity of the failure, likelihood of the failure occurring, and likelihood that the failure would

be detected. Ultimately, a risk priority number (RPN) is calculated for each failure mode:

RPN = severity × occurrence × detection. Different failures modes are observed which can be seen in Appendix

F.

Here we considered numerous types of failures. Failures such as: bad motor driver, no power being sent, and

low power are considered. An exhaustive list of failure modes and their corresponding RPN can be seen in

Appendix F. To be brief, we will mention the top two RPNs and their corresponding failure modes. The highest

RPN (120) that we calculated was due to failure of the motor. This makes sense because without the motor, the

AAFO will not function at all (be more of a heavy weight). The second highest RPN (100) is when there is a

failing in charging. This makes sense due to the fact that one of the key sub functions that we defined earlier is

to have proper electrical properties. Thus, not charge implies that there is no proper electrical properties. Some

Page 13: Biomedical Device Design Report

13

possible avenues to correct our current issues could be incorporating an energy efficient battery, yet this may

increase in weight which is something we wish to avoid.

Page 14: Biomedical Device Design Report

14

Appendix A: Functional Decomposition

Page 15: Biomedical Device Design Report

15

Appendix B: Morphology Chart

Sub Function Concept 1 Concept 2 Concept 3

Support Knee Plastic Leg Orthosis Metal Structure Carbon Fibre Orthosis

Support body weight Composit Structure Metal Structure Glass

Support load for ankle Motor Pneumatic Systems Spring loaded system

Transmit power Ball Screw Pneumatic Cylinders Cable & Pulley

Provide eletrical power Wall Socket Battery IC Engine

Display device status LCD Screen LED Display LED bulbs

Regulate current to motor MOSFET Driver BJT Driver Relay Based Driver

Process Signals ADC with high resolution and amplifierADC with higher resolution Amplifier

Sense muscle/nerve impulseEMG Sensor Force Sensor Limit Switches

Move leg up/down SEA(Seres Elastic Actuator) Use of Cam Pump

Move/turn ankle SEA(Seres Elastic Actuator) Pnumatic System Spur gear

Have flexibility Elastic Plastics Super elastic metals Rubber

Have variable speeds Electronic Controller Change in the dimension of mechanical partsUse brakes

Page 16: Biomedical Device Design Report

16

Appendix C: Ashby Diagrams

Shaft

Plate

Page 17: Biomedical Device Design Report

17

Joints

Foot plate

Page 18: Biomedical Device Design Report

18

Appendix D: Pugh Matrices

Iter

ati

on

1It

era

tio

n 2

Cri

teri

on

D

esi

gn 1

De

sign

2D

esi

gn 3

Cri

teri

on

D

esi

gn 1

De

sign

2D

esi

gn 3

Sup

po

rt K

nee

dat

um

-11

Sup

po

rt K

nee

1d

atu

m1

Sup

po

rt b

od

y w

eigh

td

atu

m0

-1Su

pp

ort

bo

dy

wei

ght

0d

atu

m-1

Sup

po

rt lo

ad f

or

ankl

ed

atu

m0

-1Su

pp

ort

load

fo

r an

kle

1d

atu

m-1

Tran

smit

po

wer

dat

um

1-1

Tran

smit

po

wer

0d

atu

m-1

Pro

vid

e el

etri

cal p

ow

erd

atu

m0

-1P

rovi

de

elet

rica

l po

wer

1d

atu

m-1

Dis

pla

y d

evic

e st

atu

sd

atu

m0

-1D

isp

lay

dev

ice

stat

us

0d

atu

m-1

Reg

ula

te c

urr

ent

to m

oto

rd

atu

m0

-1R

egu

late

cu

rren

t to

mo

tor

1d

atu

m-1

Pro

cess

Sig

nal

sd

atu

m0

-1P

roce

ss S

ign

als

1d

atu

m-1

Sen

se m

usc

le/n

erve

imp

uls

ed

atu

m-1

-1Se

nse

mu

scle

/ner

ve im

pu

lse

1d

atu

m-1

Mo

ve le

g u

p/d

ow

nd

atu

m-1

-1M

ove

leg

up

/do

wn

1d

atu

m-1

Mo

ve/t

urn

an

kle

dat

um

-1-1

Mo

ve/t

urn

an

kle

1d

atu

m-1

Ha

ve f

lexi

bili

tyd

atu

m1

0H

ave

fle

xib

ility

0d

atu

m-1

Ha

ve v

aria

ble

sp

eed

sd

atu

m-1

-1H

ave

var

iab

le s

pee

ds

1d

atu

m-1

0-3

-10

90

-11

Page 19: Biomedical Device Design Report

19

Appendix E: DMA/DFA

Pa

rt#

Pa

rts

Ass

e

mb

ly

Pa

rt

Re

qu

ire

d

AL

= a

lignm

ent

diffi

cult

OB

= o

bstr

ucte

d

NTD

= n

ot t

op d

own

RE

S =

res

ista

nce

HIP

= h

old

in p

lace

Sm

all

Tang

led

Fle

xibl

e

No

end

No

inse

rt

Hea

vy

or t

ools

A

LO

BN

TDR

ES

HIP

FA

STE

N

<12

mm

(+

1)(+

2)Tw

ist

(+1)

<2m

m (

+2)

(+2)

(+2)

(+2)

(+1)

if c

lear

(+2)

(+2)

(+2)

(+2)

(+2)

(+2)

Scr

ew (

+3)

Foo

t P

late

22

22

22

41

16

mot

or2

22

22

22

24

120

Ser

ies

elas

tic a

ctua

tor

22

22

22

24

118

tran

smis

sion

22

22

22

41

16

pins

12

22

23

1619

2

fast

ener

s1

22

210

70

EM

G S

enso

r2

22

318

inte

rface

21

2

batt

ery

22

22

24

456

To

tal

3835

2 0

mot

or2

22

41

10

SE

A2

22

22

41

14

tran

smis

sion

22

22

24

114

pins

1

22

22

316

192

fast

ener

s1

22

210

70

batt

ery

22

22

24

456

EM

G s

enso

r2

22

318

inte

rface

1

0

Foo

t pl

ate

22

22

22

41

16

To

tal

3837

4

Goa

l: m

inim

ize

asse

mbl

y in

dex

and

part

s co

unt

Par

t re

quire

d if:

rela

tive

mot

ion

Par

ts r

educ

tion:

diffe

rent

mat

eria

l

(dis

)ass

embl

y im

poss

ible

with

out

Re

trie

veH

an

dle

Inse

rt

Page 20: Biomedical Device Design Report

20

Appendix F: FMEA

Page 21: Biomedical Device Design Report

21

Appendix G: QFD

Page 22: Biomedical Device Design Report

22

References

[1]Azarpazhooh MR, Shahripour RB, Kapral MK, Mokhber N, Shoeibi A, Farzadfard MT, et al. Incidence of first

ever stroke during Hajj ceremony. BMC Neurol. 2013;13(1):193.

[2]Bjorn-Mortensen K, Lynggaard F, Pedersen ML. Incidence of Greenlandic stroke-survivors in Greenland: a

2-year cross-sectional study. Int J Circumpolar Health. 2013;72:22626.

[3]Jørgensen HS, Nakayama H, Raaschou HO, Olsen TS. Recovery of walking function in stroke patients: the

Copenhagen Stroke Study. Archives of physical medicine and rehabilitation. 1995;76(1):27- 32.

[4]Danielsson A, Carin W, Katharina S. Comparison of energy cost of walking with and without a carbon

composite ankle foot orthosis in stroke subjects. 13th ISPO World Congress; 10-15 May; Leipzig,

Germany2010.

[5]Chen G, Patten C, Kothari D, Zajac F. Gait differences between individuals with post-stroke hemiparesis and

non-disabled controls at matched speeds. Gait & posture. 2005;22(1):51-6.

[6]Slijper A, Danielsson A, Willen C. Ambulatory Function and Perception of Confidence in Persons with Stroke

with a Custom-Made Hinged versus a Standard Ankle Foot Orthosis. Rehabil Res Pract. 2012;2012:206495.

[7] Feng, T. Hefzy, MS., Elahnina M. State of the Art Review of Knee–Ankle–Foot Orthoses. Annals of

Biomedical Engineering, Vol. 43, No. 2, 2015 ( 2015) 427–44.

[8]De Wit DC, Buurke JH, Nijlant JM, Ijzerman MJ, Hermens HJ. The effect of an ankle-foot orthosis on walking

ability in chronic stroke patients: a randomized controlled trial. Clin Rehabil. 2004 Aug;18(5):550-7.

[9]Blaya JA, Herr H. Adaptive control of a variable-impedance ankle-foot orthosis to assist dropfoot gait. Neural

Systems and Rehabilitation Engineering, IEEE Transactions on. 2004;12(1):24-31.

[10]Gök H, Küçükdeveci A, Altinkaynak H, Yavuzer G, Ergin S. Effects of ankle-foot orthoses on hemiparetic

gait. Clinical rehabilitation. 2003;17(2):137-9.

[11]Ruthenberg BJ, Wasylewski NA, Beard JE. An experimental device for investigating the force and power

requirements of a powered gait orthosis. J Rehabil Res Dev. 1997 Apr;34(2):203-13.

[12]Yano H, Kaneko S, Nakazawa K, Yamamoto SI, Bettoh A. A new concept of dynamic orthosis for

paraplegia: the weight bearing control (WBC) orthosis. Prosthet Orthot Int. [Research Support, Non-U.S.Gov't].

1997 Dec;21(3):222-8.

[13]Blaya JA, Herr H. Adaptive control of a variable-impedance ankle-foot orthosis to assist dropfoot gait. IEEE

Trans Neural Syst Rehabil Eng. 2004 Mar;12(1):24-31.

[14]Noel M, Cantin B, Lambert S, Gosselin CM, Bouyer LJ. An electrohydraulic actuated ankle foot orthosis to

generate force fields and to test proprioceptive reflexes during human walking. IEEE Trans Neural Syst

Rehabil Eng. 2008 Aug;16(4):390-9.

[15]Xu R, Jiang N, Mrachacz-Kersting N, Lin C, Asin Prieto G, Moreno JC, et al. A closed-loop braincomputer

interface triggering an active ankle-foot orthosis for inducing cortical neural plasticity. IEEE Trans Biomed Eng.

2014 Jul;61(7):2092-101.

[16]Esfahani, Ehsan Tarkesh. Developing an active ankle foot orthosis based on shape memory alloys. Diss.

The University of Toledo, 2007.

[17]Yamamoto S, Ibayashi S, Fuchi M, Yasui T. Immediate-term effects of use of an ankle-foot

orthosis with an oil damper on the gait of stroke patients when walking without the device. Prosthet Orthot Int.

2014 Jan 27.

[18]Ohata K, Yasui T, Tsuboyama T, Ichihashi N. Effects of an ankle-foot orthosis with oil damper on

muscle activity in adults after stroke. Gait Posture. 2011 Jan;33(1):102-7.

Page 23: Biomedical Device Design Report

23

[19]Yamamoto S, Hagiwara A, Mizobe T, Yokoyama O, Yasui T. Development of an ankle-foot

orthosis with an oil damper. Prosthet Orthot Int. 2005 Dec;29(3):209-19.

[20]Tarkesh, Ehsan, Mohammad H. Elahinia, and Mohamed Samir Hefzy. "Developing an Active Ankle Foot

Orthosis Based on Shape Memory Alloy Actuators." ASME 2007 Summer Bioengineering Conference.

American Society of Mechanical Engineers, 2007.

[21] Birbaumer, N., Ghanayim, N., Hinterberger, T., Iversen, I., Kotchoubey, B., Kuebler, A., Perelmouter, J.,

Taub, E. and Flor, H., A spelling device for the paralysed, Nature, 398 (1999) 297-298

[22] Pfurtscheller, G. and Neuper, C., Motor imagery activates primary sensorimotor area in humans, Neurosci.

Lett., 239 (1997) 65-68

[23] Lauer, R.T., Peckham, P.H. and Kilgore, K.L., EEG based control of a hand grasp neuroprosthesis,

NeuroReport, 10 (1999) 1-5.

[24] Neuper, C., Schloegl, A. and Pfurtscheller, G., Enhancement of left-right sensorimotor EEG differences

during feedback regulated motor imagery, J. Clin. Neurophysiology., 16 (1999) 373-382

[25] Pfurtscheller, G., Neuper, C., SchloÈ gl, A. and Lugger, K., Separability of EEG signals recorded during

right and left motor imagery using adaptive autoregressive parameters, IEEE Trans. Rehab. Eng., 6 (1998)

316-325.

[26] Wolpaw, J.R. and McFarland, D.J., Multichannel EEG-based brain-computer communication,

Electroenceph. clin. Neurophysiology., 90 (1994) 444-449.

[27] http://www.ai.mit.edu/research/abstracts/abstracts2001/bio-machines/05blaya.pdf

[28] Hwang, Sungjae, et al. &quot;Development of an active ankle foot orthosis for the prevention of foot drop

and toe drag.&quot; Biomedical and Pharmaceutica Engineering, 2006. ICBPE 2006. International Conference

on. IEEE, 2006.

[29]. Herr, Hugh, Joaquin Blaya, and Gill A. Pratt. &quot;Active ankle foot orthosis.&quot; U.S. Patent No.

8,376,971. 19 Feb. 2013.

[30] J. Furusho, T. Kikuchi, M. Tokuda, T. Kakehashi, K. Ikeda, S. Morimoto, Y. Hashimoto, H. Tomiyama, A.

Nakagawa, and Y. Akazawa, “Development of shear type compact MR brake for the intelligent ankle-foot

orthosis and its control; research and development in NEDO for practical application of human support robot,”

in Proc. IEEE 10th Int. Conf. Rehabil. Robot., Noordwijk, The Netherlands, 2007, pp. 89–94. 5. W. Svensson

and U. Holmberg, “Ankle-foot- orthosis control in inclinations and stairs,” in Proc. IEEE Int. Conf. Robot.,

Autom. Mechatronics, Chengdu, China, 2008, pp. 301–306

[31] Shorter, Kenneth Alex, et al. &quot;Technologies for powered ankle-foot orthotic systems: Possibilities and

challenges.&quot; Mechatronics, IEEE/ASME Transactions on 18.1 (2013): 337-347. 7. J. A. Blaya and H.

Herr, “Adaptive control of a variable-impedance anklefoot orthosis to assist drop-foot gait,” IEEE Trans. Neural

Syst. Rehabil. Eng., vol. 12, no. 1, pp. 24–31, Mar. 2004. 8. S. Hwang, J. Kim, and Y. Kim, “Development of an

active ankle-foot orthosis for hemiplegic patients,” presented at the 1st Int. Convention Rehabil. Eng. Assistive

Technol.: Conjunction 1st Tan Tock Seng Hospital Neurorehabil. Meet., Singapore, 2007.

[32] K. W. Hollander, R. Ilg, T. G. Sugar, and D. Herring, “An efficient robotic tendon for gait assistance,” J.

Biomech. Eng., vol. 128, pp. 788–792, Oct. 2006. 10. J. Hitt, A. M. Oymagil, T. Sugar, K. Hollander, A.

Boehler, and J. Fleeger, “Dynamically controlled ankle-foot orthosis (DCO) with regenerative kinetics:

Incrementally attaining user portability,” in Proc. IEEE Int. Conf. Robot. Autom., Apr. 2007, pp. 1541–1546. 11.

A. W. Boehler, K. W. Hollander, T. G. Sugar, and D. Shin, “Design, implementation and test results of a robust

control method for a powered ankle foot orthosis (AFO),” in Proc. IEEE Int. Conf. Robot. Autom., 2008, pp.

2025–2030.

Page 24: Biomedical Device Design Report

24

[33] E. T. Esfahani, “Developing an active ankle foot orthosis based on shape memory alloys,” M.S. thesis,

University of Toledo, Toledo, OH, 2007, p. 161

[34] Y. Bar-Cohen, “Electro-active polymers: Current capabilities and challenges,” in Proc. SPIE Smart Struct.

Mater. Symp., San Diego, CA, 2002,

[35] S. Yamamoto, A. Hagiwara, T. Mizobe, O. Yokoyama, and T. Yasui, “Development of an ankle-foot

orthosis with an oil damper,” Prosthetics Orthotics Int., vol. 29, pp. 209–219, 2005.

[36] R. Chin, E. Loth, and E. Hsiao-Wecksler, “Fluid-power harvesting by pneumatic bellow during human gait,”

presented at the 8th Int. Symp. Fluid Power, ASME Fluids Eng. Conf., Jacksonville, FL, 2008.

[37] Ferris, D.P., J.M. Czerniecki, and B. Hannaford, An ankle-foot orthosis powered by artificial pneumatic

muscles. Journal of Applied Biomechanics, 2005. 21(2): p. 189-197.

[38] Carl Nelson, A Primer on Engineering Design of Biomedical Devices, University of Nebraska Lincoln,

Second edition, 2013

[39]Aykut Cihan Satici, Ahmetcan Erdogan, Volkan Patoglu, Design of a Reconfigurable Ankle Rehabilitation

Robot and Its Use for the Estimation of the Ankle Impedance, 2009 IEEE 11th International Conference on

Rehabilitation Robotics Kyoto International Conference Center, Japan, June 23-26, 2009

[40]http://www.beckerorthopedic.com/Catalog/AnkleComponentsCatalog.pdf

[41]Joaquin Blaya, Dava Newman & Hugh Herr, Active Ankle Foot Orthoses (AAFO), Artificial Intelligence

Laboratory Massachusetts Institute of Technology Cambridge, Massachusetts