balloon pump

2
Intra-Aortic Balloon Troubleshooting TRIGGERING In order to have consistent and effective pumping, a reliable physiological signal (trigger) is required to initiate an inate/deate cycle. A consistent assist marker  veries reliable trigger recognition by the pump. ECG is the preferred trigger and is the most reliable in general. 1. Sele ct a lead tha t gives t he large st R wave a nd minim izes th e P and T wave s. 2. Avoid bi phasic QR S comple xes. The pu mp may not consistent ly recogn ize its trigger point because there are two slopes to analyze. This could cause  wandering timing or skipped beats. 3. Ensure ECG gai n is adjus ted so tha t the R wave i s tall enou gh to be see n by the trigger algorithm and not so high that the pump is triggering on the P or T  waves also. 4. Regu lar rhythm s should pro duce reg ular pum ping. Arti fact may cau se double triggering or trigger loss. BPW HEIGHT Reects the pressure in the aorta, therefore the plateau pressure on the BPW should be within ±25mmHg of the Peak Diastolic Pressure. BPW WIDTH Is approximately the duration in which the balloon is inated. 1. Ze ro Bas el in e. 2. Bal loon Pre ssu re B ase line . 3. Ra pi d In ation. 4. Pea k Inati on Arti fac t. 5. Pl at eau Press ur e. 6. Rapi d De ati on. 7. De ati on A rti fact . 8. Re tu rn t o Ba se li ne. 9. Dur ati on of B all oon Cyc le. TRIGGER LOSS ALARM 1. T rigge r signal not present or not rec ognizable: a. Check ECG source and AP source. b. Check ECG gain. c. Change trigger mode.  WA NDERING TIMING 1. If irregular rhythm present recomme nd us ing  AFIB trigger for A CAT ® 1 PLUS and KAAT II PLUS ® , ECG with arrhythmia timing for AutoCAT , or R-Wave with arrhythmia timing for TransAct ® . 2. If a regular rhythm is present, the p ump is not see ing a consist ent trigg er signal. a. Check the ECG gain. b. Change ECG lead. c. Change trigger mode. POOR AUGMENTA TION 1. Patient’s stroke volume is greater than the balloon volume. Prophylactic insertion. Patient may be ready to wean from pump. 2. Balloon too small related to patient’s aortic environment. Check balloon pressure waveform’s plateau pressure. 3. Low SVR. 4. Hypovolemia. 5. Hypertension. 6. Balloon positioned too low. 7. Late ination. 8. Partial obstruction of gas lumen, particularly if in conjunction with wide ination artifact on balloon pressure waveform.  ABNORMAL BPW / ALARMS  Wide Ination and/or Deation Artifact Indicates a partial obstruction to the gas lumen of the balloon. May result in poor augmentation and/or poor afterload reduction. May lead to gas loss alarms in higher heart rates when in 1:1 assist ratio. High Pressure / Kinked Line Alarm Generally due to kinked catheter , either internal or external. 1. Repos ition patie nt. Keep affect ed leg straight. Use rolled towel under affected hip to hyperextend hip. 2. App ly sli ght tr act ion to cat het er . 3. Che ck p lac eme nt o f bal loo n, may be too high or low. 4. IAB may be too larg e for pat ien t. a. Recommended to not reduce volume below 2/3. Note: square or rounded plateau pressure. Helium Loss / Gas Loss / Gas Leakage Alarm Not all of the gas that went out to the balloon came back to the pump. 1. Observe for blo od in t ubing . If p resen t, le ave pump off, clamp tubing, notify physician to remove catheter . 2. Ch ec k co nnec ti ons. 3. Ch ec k fo r ki nk s. Note: baseline slightly in negative. High Baseline / Fill Pressure Alarm Indicates too much gas in system. 1. Check for in termittent obstru ction of gas lume n. 2. Over ll. Note: baseline far above zero. 1 2 3 4 5 6 8 7 9 BALLOON PRESSURE WAVEFORM The Balloon Pressure Waveform (BPW) represents helium movement to and from the console and the IAB catheter. It is a calibrated waveform, which allows objective assessment of the safety and effectiveness of counterpulsation. Good Complex Choices Potential Problem Complexes 250 200 150 100 50 0 -50

Upload: raghavendra-prasad

Post on 15-Oct-2015

8 views

Category:

Documents


0 download

DESCRIPTION

iabp

TRANSCRIPT

  • 5/25/2018 Balloon pump

    1/2

    24-Hour Intra-Aortic Balloon Product

    Support Hotline: 800-447-IABP

    Intra-Aortic Balloon Troubleshooting

    TRIGGERING

    In order to have consistent and effective pumping, a reliable physiological signal(trigger) is required to initiate an inflate/deflate cycle. A consistent assist marker

    verifies reliable trigger recognition by the pump.

    ECG is the preferred trigger and is the most reliable in general.

    1. Select a lead that gives the largest R wave and minimizes the P and T waves.2. Avoid biphasic QRS complexes. The pump may not consistently recognize its

    trigger point because there are two slopes to analyze. This could causewandering timing or skipped beats.

    3. Ensure ECG gain is adjusted so that the R wave is tall enough to be seen bythe trigger algorithm and not so high that the pump is triggering on the P or T

    waves also.

    4. Regular rhythms should produce regular pumping. Artifact may cause doubletriggering or trigger loss.

    BPW HEIGHTReflects the pressure in the aorta, therefore theplateau pressure on the BPW should be within25mmHg of the Peak Diastolic Pressure.

    BPW WIDTHIs approximately the duration in which theballoon is inflated.

    1. Zero Baseline.

    2. Balloon Pressure Baseline.

    3. Rapid Inflation.

    4. Peak Inflation Artifact.

    5. Plateau Pressure.

    6. Rapid Deflation.

    7. Deflation Artifact.

    8. Return to Baseline.

    9. Duration of Balloon Cycle.

    TRIGGER LOSS ALARM

    1. Trigger signal not present or not recognizable:

    a. Check ECG source and AP source.

    b. Check ECG gain.

    c. Change trigger mode.

    WANDERING TIMING

    1. If irregular rhythm present recommend usingAFIB trigger for ACAT1 PLUS and KAAT II PLUS, ECG with arrhythmiatiming for AutoCAT, or R-Wave with arrhythmia timing for TransAct.

    2. If a regular rhythm is present, the pump is not seeing a consistent trigger signal.

    a. Check the ECG gain.

    b. Change ECG lead.

    c. Change trigger mode.

    POOR AUGMENTATION

    1. Patients stroke volume is greater than the balloon volume. Prophylacticinsertion. Patient may be ready to wean from pump.

    2. Balloon too small related to patients aortic environment. Check balloon

    pressure waveforms plateau pressure.3. Low SVR.

    4. Hypovolemia.

    5. Hypertension.

    6. Balloon positioned too low.

    7. Late inflation.

    8. Partial obstruction of gas lumen, particularly if in conjunction with wideinflation artifact on balloon pressure waveform.

    ABNORMAL BPW / ALARMS

    Wide Inflation and/or Deflation ArtifactIndicates a partial obstruction to the gas lumenof the balloon. May result in poor augmentationand/or poor afterload reduction. May lead togas loss alarms in higher heart rates when in1:1 assist ratio.

    High Pressure / Kinked Line Alarm

    Generally due to kinked catheter, either internalor external.

    1. Reposition patient. Keep affected legstraight. Use rolled towel under affected hipto hyperextend hip.

    2. Apply slight traction to catheter.

    3. Check placement of balloon,may be too high or low.

    4. IAB may be too large for patient.

    a. Recommended to not reduce volume below 2/3.

    Note: square or rounded plateau pressure.

    Helium Loss / Gas Loss / Gas Leakage AlarmNot all of the gas that went out to the balloon cameback to the pump.

    1. Observe for blood in tubing. If present, leavepump off, clamp tubing, notify physician toremove catheter.

    2. Check connections.

    3. Check for kinks.

    Note: baseline slightly in negative.

    High Baseline / Fill Pressure Alarm

    Indicates too much gas in system.

    1. Check for intermittent obstruction of gas lumen.

    2. Overfill.

    Note: baseline far above zero.

    1

    2

    3

    4

    5

    6

    8

    7

    9

    BALLOON PRESSURE WAVEFORM

    The Balloon Pressure Waveform (BPW) represents helium movement to and from the console and the IAB catheter. It is a calibrated waveform, which allows objective

    assessment of the safety and effectiveness of counterpulsation.

    2002 Arrow International, Inc. All rights reserved.

    Printed in Czech Republic

    IBT-P 1/02 8M

    2400 Bernville RoadReading, PA 19605

    610-378-0131

    Good Complex Choices

    Potential Problem Complexes

    250

    200

    150

    100

    50

    0

    -50

  • 5/25/2018 Balloon pump

    2/2