Page 1: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Monique Hom

Sr. Clinical Education Manager HHS

AVAPS-AE

Auto-titration Mode of Noninvasive Ventilation

Page 2: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

AVAPS-AE: Why do we need it?

3

Page 3: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

AVAPS-AE

AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed

to better treat Respiratory Insufficiency patients (OHS, COPD and NMD)

in the hospital and homecare environments

• Proven performance of AVAPS

– Maintains targeted Tidal volume

• Auto EPAP

– Maintains patent upper airway at comfortable pressure

• Auto backup rate

– Applies an Auto backup rate near a patient’s resting rate

4

Ralph T
Typewritten Text
Neuromuscular Diseases (NMD)
Ralph T
Typewritten Text
Ralph T
Line
Page 4: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Theory of Operation

Upper

Airway

Resistance

Device

Pressure

P

search crit

P therapy

P search opt

5

Illustration courtesy of Krames Medical Illustration.

AVAPS-AE: Auto EPAP Proactive Analysis

Page 5: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Popt – Optimal Pressure Search

(High Pressure Search)

EP

AP

P therapy

P therapy

Critical Pressure Searches

(Low Pressure Search)

P

therapy

EP

AP

P therapy

Popt

Pcrit

AVAPS-AE: Auto EPAP proactive analysis

P ther

P opt

Pre

ssure

Resistance

P crit

6

Page 6: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Auto EPAP today

Patient type

Treatment / Device Pressure support

requirements

OSA CPAP/REMstar Auto Low to none

Cardiac

Servo-Ventilation / BiPAP

AutoSV

Dynamically applied

pressure support

Respiratory insufficiency

AVAPS-AE / Trilogy,

BiPAP A40

High levels of pressure

support

7

Page 7: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Current PRI Auto EPAP/CPAP

Flatness Roundness Peak Shape

AVAPS-AE

Flo

w

Pre

ssu

re

Forced oscillation technique (FOT) at 5 Hz, 1

cm amplitude during EPAP

Comparison of Auto EPAP methods

8

Page 8: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Benefits of FOT

• FOT improves detection of obstructed airways at higher pressure support levels

– Not affected by high levels of pressure support

– FOT measurements taken at end exhalation (10 breaths)

Increase in

EPAP by 1

cm H2O s

9

Page 9: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

FOT – Patent vs. Obstructed Airway

Patent/Normal airway

Resulting flow (higher resistance)

Resulting flow (low resistance)

Obstructed airway

11

Page 10: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

EPAP

Target Vt

IPAP PS min

PS max

AVAPS-AE Maintaining Tidal volume and Airway Patency

EPAP min

EPAP max

Resistance

12

Page 11: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Auto Back-up rate

• Auto backup rate is near resting rate

• Comfortable assistance when needed

• No manual adjustments (auto-default

setting)

Auto Backup Rate

Auto backup rate combined with the tidal volume assurance of AVAPS

provides a minimum level of ventilation

13

Page 12: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

Auto Backup rate M

ovin

g b

reath

ra

te

Minimum auto

backup rate

Spontaneous breaths

Time 50 spontaneous breaths

10

Calculated

spontaneous

rate

Target auto

backup rate

Targeted auto backup rate is

2 bpm below avg.

spontaneous rate

Bp

m

14

Page 13: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

AVAPS-AE Settings

AVAPS-AE mode is available in Trilogy Aura and BiPAP A40:

AVAPS rate From 0.5 to 5 cmH2)/min in 0.5cmH2O increments

Tidal volume 200 – 1500ml Maximum pressure 6 – 40 cmH20

Pressure support Max 2 – 36 cmH2O Pressure support Min 2 – 36 cmH2O

EPAP Max pressure 4 – 25 cmH2O

EPAP Min pressure 4 – 25 cmH2O Breath rate Auto/0 – 40BPM

Page 14: AVAPS-AE AE.pdf · AVAPS-AE is an Auto-titration Mode of Noninvasive ventilation designed to better treat Respiratory Insufficiency patients (OHS, COPD and NMD) in the hospital and

Confidential Sector, MMMM dd, yyyy, Reference

AVAPS-AE suggested settings

OHS* COPD – OSA*

Vt Target 8 -10 ml/Kg of ideal

bodyweight

8 ml/Kg of ideal

bodyweight

Max P 35 cmH2O 30 cmH20

PS Max 19- 35 cmH2O 30 cmH2O

PS Min 14 -19 cmH2O 12 cmH2O

EPAP Max 14 cmH20 14 cm H2O

EPAP Min 4 cmH2O 4 cm H2O

Breath Rate Auto Auto

AVAPS Rate 2 N/A

* AVAPS-AE protocol Dr. N. Hart, Dr. P. Murphy, Lane Fox Respiratory unit, St. Thomas’ Hospital London UK

* AVAPS-AE Multi Center Trial protocol l,Prof Jean François MUIR, France

Pmax has to be the sum of EPAP min and PS max

Pmax = EPAP min + PS Max