automated fraction re-analysis does it really make … · automated fraction re-analysis does it...
TRANSCRIPT
Slide 1Chairperson: Ingrid Ginnutt
Automated Fraction Re-Analysis Does it really make sense?
Presented by: Udo Huber
• 1995 PHD in organic chemistry from the University Karlsruhe/Germany• 1996 - 97 Postdoctoral fellow at the University of Hawai’i at Manoa• Since 1997 Application Chemist with HP/Agilent• Since 2000 Senior Application Chemist for the purification system and valve solutions
Slide 2Chairperson: Ingrid Ginnutt
Contents
• IntroductionWhy purification, why re-analysis?Where is my target compound?WorkflowWhat is purity?
• Is my compound pure enough?Possible errors in automated fraction re-analysisThe right point in the workflow to check the purityOther aspects
• A dedicated system for preparative and analytical workSharing the DAD and MSD
Slide 3Chairperson: Ingrid Ginnutt
Why purification at all?
min0 2 4 6 8 10 12 14
mAU
0
50
100
150
200
250
R1 X
O
R2 R3 R1
O
R2
R3+
Target compound
Byproduct
ImpuritiesStarting material
• Target compound goes to activity testing.• Therefore it must be 100 % pure.
Slide 4Chairperson: Ingrid Ginnutt
Why re-analysis?
1. If more then one fraction was collected for the same sample re-analysis has to be done to identify the fractions containing the target compound.
2. Check purity of the target compound to make sure it is pure enough for activity testing.
Slide 5Chairperson: Ingrid Ginnutt
More then one fraction
Why do I get more then one fraction?⇒ Time- or peak-based fraction collection
To identify the right fractions I need:either standard for the compound ⇒ analytical systemor an MSD
Prep system has no MSDif it had an MSD I didn’t get more then one fraction⇒ analytical system
Slide 6Chairperson: Ingrid Ginnutt
Strategies for purification on a system equipped with MSD
min0 1 2 3 4 5 6 7
mAU
0
500
1000
1500
2000
1-Vial
1
1-Vial
21-V
ial 31-V
ial 4
DAD
Peak-based fraction collectionthreshold only (50 mAU)
min0 1 2 3 4 5 6
1-Vial
1
1-Vial
21-V
ial 31-V
ial 4
0500000
100000015000002000000250000030000003500000
MSD
7
Slide 7Chairperson: Ingrid Ginnutt
Strategies for purification on a system equipped with MSD
Target mass: 241 amuTIC
5.2 5.4 5.6 5.8 6 6.2 6.4 6.6
1-Vial
2
EIC (242 [M+H]+)Fraction containing target compound
20000040000060000080000010000001200000140000016000001800000
min
050000100000150000200000250000300000350000400000
5.2 min5.4 5.6 5.8 6 6.2 6.4 6.6
Slide 8Chairperson: Ingrid Ginnutt
Purity check – Sample workflow
1.Sample submitted by chemistSolvent often DMSO/DMF
2.Sample purifiedMobile phase water/acetonitrile
3.Fraction takenMobile phase water/acetonitrile
Slide 9Chairperson: Ingrid Ginnutt
Sample workflow
4.Solvent evaporatedCompound weighed
5.Compound re-dissolvedCertain concentration, solvent often DMSO/DMF
6.Portion taken to activity testing(either DMSO solution, diluted with aqueous buffers or without DMSO)Rest of solution stored
Slide 10Chairperson: Ingrid Ginnutt
Important questions
• Confirm that target compound is going to activity testing, usually done by MS
• Measure purity of compound going to activity testing, must be higher then 90 - 95 %
Activity testing
A representative portion of the sample going to activity testing must be re-analyzed to check the purity!
Slide 11Chairperson: Ingrid Ginnutt
What is purity?
min0 1 2 3 4 5
mAU
020406080
100DAD: 254/8 nm (ref. 360/60 nm)
min0 1 2 3 4 5
mAU
020406080
100 DAD: 204/8 nm (ref. 360/60 nm)
min0 1 2 3 40
200000
600000
1000000MSD
Compound did not ionize!
5
min0 1 2 3 4 5
Compound evaporated!mVELSD
0
50
100
Slide 12Chairperson: Ingrid Ginnutt
Break Number 1
Slide 13Chairperson: Ingrid Ginnutt
Fraction re-analysis –where in the workflow?
Automated fraction re-analysis:Directly from fraction container (step 3)
• Advantage: High level of automation• Disadvantages: ???
Slide 14Chairperson: Ingrid Ginnutt
Possible errors in when doing automated fraction re-analysis (1)
23
4
fill
timemin0 1 2 3 4 5 6 7 8 9
mAU
0
500
1000
1500
2000
2500
3000
3500
Slide 15Chairperson: Ingrid Ginnutt
Possible errors in when doing automated fraction re-analysis (1)
min0 1 2 3 4 5 6 7 8 9
mAU
0
500
1000
1500
2000
2500
3000
3500
Red
Pink
Yellow
No mixing when vial is filled!
Blue
Slide 16Chairperson: Ingrid Ginnutt
Possible errors in when doing automated fraction re-analysis (1)
T = 0 h T = 6 h T = 24 h
Flow = 1 mL/min
Flow = 4.5 mL/min
Flow = 20 mL/min
T = 96 h
Slide 17Chairperson: Ingrid Ginnutt
Possible errors in when doing automated fraction re-analysis (1)
Portion drawn for re-analysis. Is this a representative sample?
Slide 18Chairperson: Ingrid Ginnutt
Possible errors in when doing automated fraction re-analysis (2)
Portion drawn for re-analysis. Is this a representative sample?
Sample was submitted in DMSO
Mobile phase is water/ACN
Slide 19Chairperson: Ingrid Ginnutt
Possible errors in when doing automated fraction re-analysis (3)
Sample decomposed while evaporated
Dissolved in
DMSOEvaporation
Portion taken to activity testing
Portion drawn for re-analysis. Is this a representative sample?
Do you do compound confirmation on a no-hit?
Slide 20Chairperson: Ingrid Ginnutt
What is the solution?
What can I What can I do???do???
Slide 21Chairperson: Ingrid Ginnutt
Take a second sample from the solution that goes to activity testing!
• Let the purification system do what it was designed for!• Use an analytical system to confirm compound and determine purity.
Portion taken to activity testing
Portion drawn for re-analysis. Is this a representative sample?
Slide 22Chairperson: Ingrid Ginnutt
Does this approach solve the problems?
Fraction mixed before re-analysis
Compounds dissolved before re-analysis
Decomposition discovered during re-analysis
Slide 23Chairperson: Ingrid Ginnutt
Other aspects
• Using a combined inject/collector seriously compromises the pluming of the system and leads to additional broadening and therefore loss of recovery. At […] they see 90% recovery with our system and 60% with others
• You make a very good, in fact the best, analytical system already which is optimised for that specific job. All the labs interested have [Agilent] 1100 [Series] systems available to them for such use.
• A separate analytical system means our expensive Prep system is not tied up doing routine analysis as the real prep runs start to build up in a queue -Not cost effective.
• Performing re-analysis off line means you can specifically select which fractions you would like to re-analyse rather than all collected fractions being re-analysed. This eliminates unneeded analytical runs which are both time consuming and wasteful.
Slide 24Chairperson: Ingrid Ginnutt
Break Number 2
Slide 25Chairperson: Ingrid Ginnutt
A single system for purification and analytical work
• Two separated systemsabsolutely no compromises regarding flow pathspreparative AND analytical work in parallel
• Two systems sharing a single MSD and DADcan be operated with ChemStation only (rev. A.10.01), purification software and with Easy Access Plusnot possible to do analytical and preparative work in parallelpreparative autosampler and pumps used for analytical work (delay volume)
Recommended for:Preparative flow rates up to 40 mL/min (1 inch columns)Injection volumes < 900 µLAnalytical flow rates above 0.8 - 1 mL/min (4.6 mm columns)Further details will be described in an Application Note
Slide 26Chairperson: Ingrid Ginnutt
The Agilent active splitter
HPLC Flow
ActiveSplitter
MS Flow
Slide 27Chairperson: Ingrid Ginnutt
The Agilent active splitter
• Agilent active splitter provides control of the transfer of material from one flow stream to another.
• Rotor seal includes three selectable aliquot volumes (22 nL, 100 nL, or 300 nL). • Gating frequency (0.25-2.0 Hz) and aliquot volume determines the rate of mass
transfer from source stream to destination stream.
Slide 28Chairperson: Ingrid Ginnutt
Two systems sharing a single MSD and DAD – preparative work
Splitter on
Waste
Waste
Plug
Slide 29Chairperson: Ingrid Ginnutt
Two systems sharing a single MSD and DAD – analytical work
Splitter on
Waste
Waste
Plug
Slide 30Chairperson: Ingrid Ginnutt
Two systems sharing a single MSD and DAD – direct injection
Splitter on
Waste
Waste
Plug
Slide 31Chairperson: Ingrid Ginnutt
Automated fraction re-analysis –does it really make sense?
• IntroductionWhy re-analysis, where is my target compound and what is purity?
• Possible errors in automated fraction re-analysisConcentration gradientCrystallizationDecompositionHow to avoid those problems?
• A dedicated system for preparative and analytical workSharing the DAD and MSD