Audiogram+: GN Resound proprietary fitting rule Ole Dyrlund GN ReSound Audiological Research Copenhagen

Upload: others

Post on 01-Apr-2021

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Audiogram+:GN Resound proprietary fitting rule

Ole Dyrlund

GN ReSound Audiological Research Copenhagen

Page 2: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Loudness normalization - Principle

Page 3: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Background for Audiogram+

! Audiogram+ is a loudness normalization rule adapted to the 14 band compression system in Canta based on clinical trials. � Allen J.B., Hall J.L., Jeng P.S. (1990)

! Audiogram+ calculates gain targets at the 11 audiometric frequencies for 50 and 80 dB SPL inputs based on air conduction thresholds and a number of additional factors.

! Gain targets provide an optimal trade-off between speech perception and listening comfort

Page 4: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Speech perception with Audiogram+

Speech-in-Noise Test Results

020406080

100

CIC (n=18) ITE/BTE (n=33) PBTE (n=21)

Model

Perc

ent

Cor

rect

UnaidedOwnCanta-P1Canta-P2

�Own� aids are a selection of digital and programmable instruments

Page 5: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

14 Band Compression system

Page 6: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Effect of input signal for 14 band system

Broadband vs. sine input for same overall level

-25-20

-15-10

-50

0 5 10 15

Band number

Diff

eren

ce re

lativ

e to

sin

e (d

B)

White noise vs. sineinputSpeech shaped vs. Sine input

At 1kHz broadband input is 9 dB lower than sine input

Page 7: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Broad band vs. sine band input

Insertion gain dB

Frequency

50 dB Input80 dB Input

Broad band InputSine Input

Page 8: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Audiogram+: Additional factors taken into account

! Severe Hearing Losses! Ski-slope Hearing losses! First Time Users! Air-Bone Gap! UCL! Age! Previous type of amplification

Page 9: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Accuracy of Audiogram+

�Very accurate prediction of average desiredamplification for mild, moderate and severelosses

�Standard deviation twice as high for severehearing losses, meaning more fine-tuning isneeded for these losses.

Page 10: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Prescribed versus fine-tuned gain (mild to moderate HL)

-15

-10

-5

0

5

10

15

20

100 1000 10000

Frequency (Hz)

Des

ired

extr

a ga

in (d

B)

Mean differences and standard deviation (SD) between Audiogram+ prescribed gain targets and the final gain setting preferred by test subjects. Blue line: 80 dB inputs, blue dotted line: corresponding Standard Deviation. Black line: 50 dB inputs, black dotted line: corresponding Standard Deviation.

Page 11: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

-15

-10

-5

0

5

10

15

20

100 1000 10000

Frequency (Hz)

Des

ired

extr

a ga

in (d

B)

Prescribed versus fine-tuned gain (severe HL)

Mean differences and standard deviation (SD) between Audiogram+ prescribed gain targets and the final gain setting preferred by test subjects. Blue line: 80 dB inputs, blue dotted line: corresponding Standard Deviation. Black line: 50 dB inputs, black dotted line: corresponding Standard Deviation.

Page 12: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Can prescription accuracy be increased for severe and profound losses?

Page 13: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Design

! Fitting data from several test sites were divided in two groups according to previous use of linear amplification or fast acting WDRC

! Target prescriptions calculated by the original Audiogram+ algorithm were compared with the gain preferred by the users

! Comparison was carried out at 0.25, 0.5, 1, 2, 4 and 6 kHz.

Page 14: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Test subjects

! 59 users of fast acting WRDC! 36 users of linear hearing aids! All users binaurally fitted! All experienced users of hearing aids

Page 15: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Previous nonlinear users, 0.5 kHz

y = 0.0749x - 0.1417

y = 0.0932x - 1.4607

-15

-10

-5

0

5

10

15

20

25

30

35

40

45

0 20 40 60 80 100 120

Hearing loss (dB HL)

Extr

a ga

in (d

B)

G(80)G(50)Linear (G(50))Linear (G(80))

Page 16: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Previous linear users, 0.5 kHz

y = 0.2034x - 4.2815

y = 0.2929x - 7.5298

-15

-10

-5

0

5

10

15

20

25

30

35

40

45

0 20 40 60 80 100 120

Hearing loss (dB HL)

Extr

a ga

in (d

B)

G(80)G(50)Linear (G(50))Linear (G(80))

Page 17: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

0

10

20

30

40

50

60

70

80

90

100

100 1000 10000

0

5

10

15

20

25

30

35

40

100 1000 10000

T80 (Lin)T50 (Lin)T80 (Nonlin)T50 (Nonlin)

New rule: Example

Page 18: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Accuracy of improved fitting rule

G(80) G(50)HTL 250 500 1 2 4 6 All 250 500 1 2 4 6 All

Linear <65 Diff. -1 -2 -1 0 1 1 -1 -1 -1 -1 0 1 2 -1SD 6 3 4 4 5 3 4 4 4 4 4 5 6 5

>65 Diff. 1 2 0 -1 0 0 0 1 1 -2 -2 -1 -2 -1SD 11 9 7 8 9 9 9 11 6 4 5 6 7 6

Nonlinear <65 Diff. 0 0 0 0 1 0 0 1 0 0 0 0 -1 0SD 4 4 4 5 4 3 4 4 4 4 5 4 4 4

>65 Diff. 2 1 0 0 1 0 0 -1 1 -1 0 0 -1 0SD 7 6 6 6 6 6 6 7 7 5 6 6 6 6

Previous <65 Diff. -1 0 0 0 -2 -2 -1 -1 0 1 1 -1 -1 0combined SD 4 4 4 5 5 4 4 5 4 4 4 5 5 4rue >65 Diff. 1 0 1 2 4 4 2 0 0 0 0 1 2 1

SD 11 9 8 10 10 10 10 9 8 5 7 9 9 8

Mean difference and SD between preferred fitting and prescription:

Page 19: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Accuracy of improved fitting rule

0123456789

10

SD dB

80 dB Input 50 dB Input

>65 Linear >65 Non-Linear >65 Combined <65 All

Page 20: Audiogram+: GN Resound proprietary fitting rulemedi.uni-oldenburg.de/download/paper/Entwickler-Forum... · 2009. 3. 9. · Background for Audiogram+! Audiogram+ is a loudness normalization

Conclusion

! Audiogram+ is tailored to the Resound compression system (channels, time constants)

! Audiogram+ gives an optimal trade-off between speech perception and listening comfort and a high prescription accuracy.

! Tested in a large number of clinical trials ! For severe hearing losses use of separate

prescriptions for previous users of linear and nonlinear amplification improves the prescription accuracy.