![Page 1: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/1.jpg)
www.postnova.com .www.postnova.com . 02.06.2014 1
Integration of Postnova Field Flow Fractionation Systems with Agilent’sICP-MS 7900 and 8800
Trevor [email protected]
![Page 2: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/2.jpg)
www.postnova.com .www.postnova.com .
FFF Principle
02.06.2014 2
Separation Mechanism
• Separation in a narrow ribbon-like channel• Laminar flow inside the channel• External field perpendicular to the solvent flow
![Page 3: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/3.jpg)
www.postnova.com .www.postnova.com .
There are Multiple versions of FFF
02.06.2014 3
RJ2
![Page 4: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/4.jpg)
Slide 3
RJ2 Rainer Jünger - Postnova Analytics, 12/14/2012
![Page 5: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/5.jpg)
www.postnova.com .www.postnova.com .
FFF Size of Analyte
02.06.2014 4
Separation Range
Online Detection: Agilent 7900-8800 ICP-MS & Postnova Light Scattering
![Page 6: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/6.jpg)
www.postnova.com .
The Postnova FFF Platform
02.06.2014 5
![Page 7: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/7.jpg)
www.postnova.com .www.postnova.com . 02.06.2014 6
AF4
Asymmetrical Flow Field Flow Fractionation
Universal Separator
![Page 8: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/8.jpg)
www.postnova.com .www.postnova.com .
Asymmetric Flow FFF
Applications• Peptides• Proteins• Antibodies• Virus• Liposomes• Latex Bead• Nano Particles• Synthetic / natural
Polymers
02.06.2014 7
AF2000 AT / MT Series
• Universal Separator for Proteins, Polymers and Nanoparticles ambient to mid temp
![Page 9: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/9.jpg)
www.postnova.com .www.postnova.com .
Flow FFF
02.06.2014 8
Asymmetric Flow FFF Principle
• Hydraulic Pressure Gradient Field (Cross-Flow) for Separation• Separation based on Size• Channel at Ambient, Mid or High Temperature
![Page 10: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/10.jpg)
www.postnova.com .www.postnova.com .
Patented Cross Flow Controller
Cross Flow Controller will always deliver the exact cross flow every run
Hollow FiberMicro ChannelAnalytical ChannelPreparative Channel
No Valves!
02.06.2014 9
![Page 11: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/11.jpg)
www.postnova.com .
Asymmetric Flow FFFStep 1• Sample is injected directly into the
flow stream• A second pump provides focus flow• Cross flow is achieved by a precise
syringe pump• Constant flow to the detectors, no valves
Step 2• Sample is focused to narrow band• Improved resolution and sample washing• Very low concentrations can be enriched
Step 3 • Focus flow stops and main pump elutes
the sample• Cross flow syringe pump varies• Detector flow stays constant
02.06.2014 10
![Page 12: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/12.jpg)
www.postnova.com .www.postnova.com .
Asymmetric Flow FFF
02.06.2014 11
Smart Stream Splitting
• Major portion of flow in the channel is containing no sample• Minor portions of the flow is carrying the sample above the membrane• Splitting of those two stream via 2 outlets at the channel end• Only the sample containing sub-stream is guided to the detector• Sample dilution is reduced at channel outlet• Increase of sensitivity by factor 5 or more depending on conditions
![Page 13: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/13.jpg)
www.postnova.com .www.postnova.com . 02.06.2014 12
HDL Lipoprotein Preparation for Mass Spectroscopy
Smart Stream Splitting
Enrichment of HDL by a factor of 2 using Smart Stream Splitting
![Page 14: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/14.jpg)
www.postnova.com .www.postnova.com . 02.06.2014 13
AF2000 MultiFlow FFF Channels
Asymmetric Flow FFF - Principle
![Page 15: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/15.jpg)
www.postnova.com .www.postnova.com .
Capabilities of FFF-ICP-MS
02.06.2014 14
FFF:• MS enables to detect even very low
concentrations• Various elements can be traced
simultaneously• Element concentration & ratio distribution
ICP-MS:• Size distribution of particles or
macromolecules• Size information by Light scattering, from
FFF-Theory• Separation of size fractions for further
analysis
FFF-ICP-MS allows size-based elemental distributions
![Page 16: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/16.jpg)
www.postnova.com .www.postnova.com .
FFF Coupled with ICP-MS
02.06.2014 15
Size Separation – Particle Size – Element CompositionSize Separation – Particle Size – Element Composition
![Page 17: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/17.jpg)
www.postnova.com .www.postnova.com . 02.06.2014 16
BSA – Bovine Serum Albumin – Aggregates
0
1
2
3
4
0 2 4 6 8 10 12 14 16Retention Time [min]
Det
ecto
r Res
pons
e [U
V/m
V]
BSA monomer
Dimer
TrimerTetramer?
AF2000 – UV
![Page 18: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/18.jpg)
www.postnova.com .www.postnova.com .
AF4-Separation of Aquatic Colloids
Element distribution across size distribution from AF4-ICP-MS • Varying distributions of the
different elements were found
02.06.2014 17
ENVIRO: Hasselhöv et. al., Göteborg, Sweden
![Page 19: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/19.jpg)
www.postnova.com .www.postnova.com .
AF4-ICP-MS Analysis of Nano Products
02.06.2014 18
FOOD/COSMETIC: Laborda et. al., Zaragozza, Spain• Silver Nanoparticles of different size and silver-content can be characterized
by AF4-ICP-MS
![Page 20: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/20.jpg)
www.postnova.com .www.postnova.com .
4 Nanoparticle Mix:Peak 1: Rg = 6.3 nmPeak 2: Rg = 17 nmPeak 3: Rg = 33 nmPeak 4: Rg = 136 nm
02.06.2014 19
Latex Nanoparticle Separation in 0.01 % SDS
• System: AF2000 MT coupled with PN3621 MALS• Conditions: 25°C, 20 μL injection volume, c = 1,2 mg/mL (total conc.
of mix)
AF2000 – MALS
![Page 21: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/21.jpg)
www.postnova.com .www.postnova.com .
AF2000 - UV
02.06.2014 20
Polystyrene Latex Nanoparticle Mix
UV S
ignal
254 n
m
Retention Time [min]
![Page 22: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/22.jpg)
www.postnova.com . 21
The CHALLENGE: Making polystyrene particles visible to ICP-MS
• Field-Flow Fractionation system can fractionate particles according to their size (hydrodynamic diameter)
• Coupled to ICP-MS can provide element/size distributions• Most size calibrants are polymers
• Polystyrene spheres often used• Don’t contain metal ions and cannot be detected with ICP-MS
• Therefore other detectors (UV, MALS, DLS) need to be coupled before the ICP-MS to obtain both size and elemental distribution
![Page 23: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/23.jpg)
www.postnova.com . 22
Detection of Sulfur-Coated PolystereneParticles by FFF-ICP-MS
• Polystyrene (PS) spheres can agglomerate and surface-charged particles are often produced to prevent it
• Surface Sulfonated-PS particles used as calibrants
• S present on the surface of nanoparticles
• The nanoparticles are in µg/g range and S is in low ng/g
• S not detectable in nanoparticles (surface) with Agilent 7700 (Xemode) due to much poorer LODs in comparison with those of the 8800 ICP-MS (Reaction gas: O2)
![Page 24: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/24.jpg)
www.postnova.com . 23
FFF-ICP- 8800 MS detection of S-coated PS nanoparticles (100 and 200 nm) via S detection
0
1
2
3
4
5
6
7
8
9
0 20 40 60 80 100
Time [min]
Norm
aliz
ed U
V si
gnal
0
10
20
30
40
50
60
70
80
ICP-
MS
sign
al [k
cps]
UV 254 nm UV 236 nm S32 -> 48 S34 -> 50
H2 + O2 mode: 4.5 ml/min H2 and 45% of O2
Heidi Goenag Infante -LGCTeddington,Middlesex TW11 0LY, UK
![Page 25: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/25.jpg)
www.postnova.com .www.postnova.com . 02.06.2014 24
Centrifugal Field Flow Fractionation
Nanoparticle Separator
![Page 26: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/26.jpg)
www.postnova.com .www.postnova.com .
Centrifugal FFF – Principle
Content• FFF Principle• Systems• Applications• Summary
02.06.2014 25
Separation Principle
• Gravity Separation Field up to 2.500 g• Size Separation Range: Particles 5 nm – 100 µm• Separation based on Size and Density
![Page 27: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/27.jpg)
www.postnova.com .www.postnova.com .
GIT Innovation Award
02.06.2014 26
CF2000 has been the winning instrument at Analytical Instrumentation and Software categoryCF2000 has been the winning instrument at Analytical Instrumentation and Software category
RJ1
![Page 28: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/28.jpg)
Slide 26
RJ1 Rainer Jünger - Postnova Analytics, 12/14/2012
![Page 29: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/29.jpg)
www.postnova.com .www.postnova.com .
CF2000 – Centrifugal FFF
• Postnova Centrifugal FFF• Eluent 0,05 % NovaChem100• Power Decay Gradient 3500 rpm• Postnova PN3000 MALS• Sample Concentration:• Relaxation Time:
02.06.2014 27
Polystyrene Nano-Micro Particles
![Page 30: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/30.jpg)
www.postnova.com .www.postnova.com .
0
2
4
6
8
10
12
14
0 10 20 30 40 50 60 70 80
DetectorS
igna
l [V]
Retention time [min]
__ UV
10 nm
30 nm60 nm
CF2000 – Centrifugal FFF
• Postnova Centrifugal FFF• Eluent 0,05% NovaChem100• Power Decay Gradient 3500 rpm• Postnova PN3211 UV @ 530 nm• Sample Concentration:• Relaxation Time:
02.06.2014 28
Gold Nanoparticle High Resolution Separation
![Page 31: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/31.jpg)
www.postnova.com .www.postnova.com .
CF2000 – Centrifugal FFF
02.06.2014 29
Gold Nano Rods Separation• 0,05% NovaChem100• 90° LS Signal MALS • Concentration:• Relax. Time:
Axial = 10 nmLength = 35 nm
Peak 1 – „Rods“ Peak 2 – „Eggs“Axial = 25 nmLength = 35 nm
Peak 1 „Rods“
Peak 2 „Eggs“
![Page 32: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/32.jpg)
www.postnova.com .www.postnova.com .
CF2000 – Centrifugal FFF
02.06.2014 30
Gold / Silver Nanoparticles
2 0 2 5 3 0 3 5 4 00 . 0
0 . 2
0 . 4
0 . 6
0 . 8
1 . 0
A u D = 2 0 n m
A g D = 2 0 n m
Inte
nsity
R e t e n t i o n T i m e [ m i n ]
9 0 ° L ig h t S c a t t e r i n g S ig n a l U V @ 5 3 0 n m
Ag
Au
• Centrifugal FFF allows separation of particles with the same size according to differences in density
![Page 33: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/33.jpg)
www.postnova.com .www.postnova.com .
CF2000 – Centrifugal FFF
02.06.2014 31
Silica Coated Gold Nanoparticles• 0,05% NovaChem100• 90° LS Signal MALS • Concentration:• Relax. Time:
Fraction Size (nm)1 98-1082 154-1683 123-1294 173-1875 277-364
![Page 34: ASTS-2014 Postnova Integration Field Flow Fractionation ... · Interfacing Postnova FFF systems PN 9050 Interface box, will provide and automated system with the capability to divert](https://reader034.vdocuments.us/reader034/viewer/2022042405/5f1ca00b361f7960b6673642/html5/thumbnails/34.jpg)
www.postnova.com .
Interfacing Postnova FFF systemsPN 9050 Interface box, will provide and automated system with the
capability to divert flow from the plasma in the event the plasma goes out.
Acidifying dosing pumps are available and can be fully integrated
PN9020 provides all triggers and data collection required to unify the two systems
AF2000 is completely metal free.
Postnova and Agilent are in a VAR partnership