assay name - this is a title style_+d - siemens€¦  · web viewalways store the test strips in...

21
Insert hospital name here Insert department name here Moderately Complex Prothrombin Time (PT/INR) Testing and Quality Control Procedure Xprecia Stride™ Analyzer Prepared By: Date : Approved By: Date : Effective Date: Discontinued Date: (retain this procedure for at least two years) Supersedes an Earlier Procedure: (Y or N) Earlier Procedure Discontinuance Date: The medical/laboratory director or the director’s designee should review all copies of this procedure at least once a year. The director should keep a log of the copies being maintained. Reviewed By: Date: Comments: Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-02 1 of 21

Upload: others

Post on 09-Feb-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

Moderately Complex Prothrombin Time (PT/INR) Testing and Quality Control ProcedureXprecia Stride™ Analyzer

Prepared By: Date:

Approved By: Date:

Effective Date:

Discontinued Date: (retain this procedure for at least two years)

Supersedes an Earlier Procedure: (Y or N)

Earlier Procedure Discontinuance Date:

The medical/laboratory director or the director’s designee should review all copies of this procedure at least once a year. The director should keep a log of the copies being maintained.

Reviewed By: Date: Comments:

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-021 of 16

Page 2: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

Title:Prothrombin Time (PT/INR) Testing on Xprecia Stride™ Analyzer

Principle of the TestQuick1 reported the first PT test in 1935. It has become one of the most useful tests for evaluating the extrinsic and common pathways of the coagulation process. The application for this test is a method for monitoring oral anticoagulant drug therapy, such as warfarin. Warfarin is a vitamin K antagonist that reduces the activity of Factors II (prothrombin), VII, IX, and X.2

Warfarin is prescribed to prevent blood clots from forming or growing larger in blood or blood vessels. When using anticoagulation medication, patients have to stay within a specific therapeutic range, as determined by their doctor. The doctor needs to monitor the warfarin activity to ensure that the medication dosage is correct. To monitor the activity, the doctor orders a PT test. A PT test is a blood test that measures the time it takes for blood to clot, reporting results using International Normalized Ratio (INR). The INR was developed to standardize the PT results. The analyzer measures PT values in whole blood. To begin testing, the contact end of the test strip is inserted into the test strip port on the analyzer. Then, a blood sample is applied to the test strip target area. The blood sample is automatically drawn by capillary action into the reaction chamber of the strip where the blood mixes with reagents and activates the coagulation cascade. When the analyzer senses that the blood has clotted, the testing stops. A PT response is determined and appears on the analyzer as INR.

Pre-analytic Process

Specimen Collection and Handling

Specimen CollectionImportant: Within 15 seconds of sticking the fingertip, apply the drop of blood to the test strip target area.

BIOHAZARDAll products or objects that come in contact with human or animal body fluids should be handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear facial protection, gloves, and protective clothing.

PREPARING THE PATIENT SAMPLES Prepare the patient to collect a blood sample by following your facility’s standard procedure. For example, clean the finger with an alcohol wipe, or have the patient wash his or her hands in

warm, soapy water. Dry the fingertip before taking a blood sample. Only use fresh whole blood, from a finger-stick source, with test strips.

COLLECTING A FINGER-STICK BLOOD SAMPLE

2 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 3: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

INSTRUCTIONS To stick the finger, firmly place the lancet against the finger and press the lancet trigger. Gently squeeze from the base of the finger to form a round drop of blood. If the blood smears or runs, wipe it off with a tissue and gently squeeze another round drop of

blood. (Requirement) The drop of blood should be about the same size as the test strip target area (a

minimum of 6 μL in volume). Low sample volume will cause an error message.

Specimen Storage and Stabilityn/a – a whole blood finger stick sample required and must be tested with 15 seconds.

Specimen Rejection CriteriaAdd your laboratory-specific rejection criteria here

Specimen Referral CriteriaAdd your laboratory-specific referral criteria here

Reagents

Storage and Stability

Always store the test strips as packaged and use within the expiration date printed on the test strip vial.

Unopened Storage: 5–30°C (41–86°F) @ up to 75% RH, use before the 24 month lot expiration date. Opened Storage: 5–30°C (41–86°F) @ up to 75% RH, use within 2 months after first opening vial

IMPORTANT: Always store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes of when you remove it from the test strip vial. Never use bent, scratched, or damaged test strips. Discard expired test strip vials.

IngredientsXprecia Stride test strips contain the reagent Dade® Innovin®, which is a preparation of purified recombinant human tissue factor (0.1–0.2 mg/L) combined with synthetic phospholipids (≤ 30 mg/L), calcium chloride (1.2–2.0 g/L), and stabilizers (1.5–2.5 g/L).

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-023 of 16

Page 4: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

Reagents Special Preparation and HandlingReagent Test Strips

WARNINGS AND CAUTIONS

Don’t take test strips internally or drink control solutions. Always store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes of when you remove it from the test strip vial. Never use bent, scratched, or damaged test strips. Discard expired test strip vials. Don’t scan a test strip barcode and then use a different test strip from another vial. Each test strip is single-use only; never perform a second test using the same test strip. Don’t touch or move the test strip after you apply the drop of blood. Don’t move the test strip

during the test. Never add more blood to the test strip after the test has begun.

Calibration

Calibration IntervalBefore using a new lot of reagent test strips, analyzer will prompt user to scan the test strip vial barcode to automatically add the lot calibration information.

Calibration ProcedureWhen a new lot of test strips are detected, you will be prompted by the analyzer to scan the test strip lot number using the integrated barcode reader.

INSTRUCTIONS

1. Hold the analyzer 10 cm (4 inches) from the barcode.

2. Aim the barcode reader at the barcode on the test strip vial.

3. Tap the SCAN Icon.

Tip: An audible tone sounds, a check mark displays, and the screen changes when the barcode is accepted

The analyzer stores at least 2 vial calibration information sets to allow the “Scan Vial” step to be omitted if the same test strip lot is used in subsequent tests. The calibration information set is retained between Xprecia Stride analyzer power on/off cycles. The calibration information stored in the analyzer correlates the information from the barcode on the vial and the information from the test strip barcode.

The logic internal to the analyzer determines the expiration date of the test strip and prevents completion of the test process if the date has expired.

Quality Control (QC)For detailed QC procedural information, refer to the Xprecia Stride Analyzer User Guide and Xprecia PT/INR Instructions for Use.

4 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 5: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

Electronic QC Checks

When the analyzer is first turned on, the instrument performs a series of electronic, signal, software and memory integrity checks, as well as ensuring there is sufficient battery voltage to operate the Xprecia Stride Analyzer. The key tests during this phase are the Heater/ Thermistor check along with the Strip Port Hardware check. These are part of overall Electronics Integrity Check. Failure to pass any of these Power On Tests will prevent further operation of the analyzer. Upon the start of a test, the analyzer performs a self-test to verify it's operational. The analyzer electrically monitors for the presence of a test strip. Upon test strip insertion, the analyzer electrically connects to the test strip electrodes to acquire test data. It also conducts test strip integrity checks to validate test strip content.

Understanding the LQC test

Always perform QC tests in accordance with local, state, and federal guidelines. QC tests help maintain regulatory compliance requirements, as applicable to your facility. Use control solutions to perform quality control checks on the Xprecia Stride and the test strips to ensure they are functioning correctly.

About LQC

The Xprecia System PT Control kit contains assayed liquid quality controls (LQC) for the assessment of precision and analytical bias in the normal (Xprecia System PT Control 1) and therapeutic (Xprecia System PT Control 2) range for the International Normalized Ratio (INR) to be used with the Xprecia System PT/INR Strips.

The method-dependent assigned values and ranges for each lot of PT Control 1 and PT Control 2 appear on each vial as a barcode to be read by the Xprecia Stride Analyzer. When the barcode on the control vial is scanned, the assigned range for each lot-specific level of LQC can be read on the analyzer screen display as an INR value. The assigned ranges are set as ± 0.2 INR from the mean assigned value for PT Control 1 and ± 0.6 INR from the mean assigned value for PT Control 2. The assigned ranges are based on 2SD of the systems total variability.

Requirement

Don't allow control solution to leak into the test strip port, as it may damage the Xprecia Stride Analyzer.

LQC MaterialsXprecia Stride PT Control Kit – SMN 10873633 (4 vials of PT 1, 4 vials of PT 2, 8 vials of CaCl2 Diluent).

Add your laboratory-specific QC materials here

QC FrequencyAdd your laboratory-specific QC frequency here

You must perform QC tests:

at the start of each shift and with every new lot, new shipment or as required by local, state, and federal or national

regulations.

Refer to the Xprecia System PT Controls Instructions For Use for more information on quality controls.

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-025 of 16

Page 6: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

QC Preparation

INSTRUCTIONS:

1. Have the test strip vial and the vial lot information on the vial available. 2. Ensure the bottle of control solution and the lot information are available. 3. Use one transfer pipette to combine the entire volume of 1 vial of diluent (CaCl2) into 1 control vial. 4. Mix carefully, by swirling the bottle using a circular motion, to completely dissolve all of the control plasma inside. Don't shake in order to avoid foam formation. 5. Close the bottle and allow to stand for at least 5 minutes at 15–25°C (59–77°F). 6. Gently swirl the bottle again prior to use.

Important:

Only use the control solution manufactured by Siemens to verify the performance of the analyzer.

– Retain the transfer pipette for use during the application of control solution to a test strip.

– Don’t use the control solution after the expiration date on the bottle.

– To ensure proper results, make sure you perform a quality control test using Xprecia PT Control 1 and then repeat the test using Xprecia PT Control 2.

Troubleshooting Out-of-Range QC ValuesIf the LQC is out of range, verify the test results and redo the test with a new strip and confirm the results.

Analytic Process

Instrument Operation and System Description

Intended use

The Xprecia Stride™ Coagulation System, which includes the Xprecia Stride™ Coagulation Analyzer and the Xprecia™ System PT/INR Strips, is intended for use by professional healthcare providers to provide an INR (International Normalized Ratio) based on a prothrombin time (PT) response for the monitoring of oral anticoagulation therapy with warfarin, a vitamin K antagonist. The Xprecia Stride™ Coagulation Analyzer is intended to be used with only the Xprecia™ System PT/INR Strips and the Xprecia™ System PT Controls. The analyzer uses fresh capillary (fingerstick) whole blood applied to an Xprecia™ System PT/INR Strip. It is intended for in vitro diagnostic use at the point-of-care. Xprecia™ System PT/INR Strips are for use with only the Xprecia Stride™ Coagulation Analyzer for PT/INR determinations by professional healthcare providers. This product is for in vitro diagnostic use. The Xprecia Stride™ Coagulation System is intended for use in patients 18 years of age and older. Patients must be stabilized (> 6 weeks) on warfarin therapy. The Xprecia Stride™ Coagulation System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy.

General warnings and precautions

‒Always follow the safety procedures and precautions listed throughout this guide when using Xprecia Stride analyzer.

6 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 7: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

‒All parts of analyzer are potentially infectious and capable of transmitting blood-borne pathogens between patients and healthcare professionals.

‒You must disinfect the device after each patient use. You can only use the Xprecia Stride for testing patients when all standard precautions and the recommended cleaning and disinfection procedures in this guide are followed.

‒Only use auto-disabling, single-use lancing devices with Xprecia Stride.

Refer to the following general safety reference materials for further information:

‒Guideline for Isolation Precautions: Preventing Transmission of Infectious Agents in Healthcare Settings 2007, found at http://www.cdc.gov/hicpac/2007IP/2007isolationPrecautions.html. ‒Protection of Laboratory Workers From Occupationally Acquired Infections; Approved Guideline-Third Edition Clinical and Laboratory Standards Institute (CLSI) M29-A3.

‒FDA Public Health Notification: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication (2010), found at http://www.fda.gov/medicaldevices/safety/ alertsandnotices/ucm224025.html.

‒CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Blood-borne Pathogens (2010), found at ttp://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html.

‒Hematocrit range of 22–52% doesn't significantly affect test results.

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-027 of 16

Page 8: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

Daily Start-up and Maintenance

Running LQC Control Tests

INSTRUCTION:1. On the Home screen, tap QC icon.

2. Scan the lot information on the control solution bottle.3. To scan the test strip barcode, aim the barcode reader at the barcode on the test strip, then tap

SCAN. 4. Tip: An audible tone sounds, a check mark displays, and the screen changes when the barcode is

accepted. 5. Insert the test strip into the test strip port.

6. If required, scan the lot information on the test strip vial. 7. Place the analyzer on a level surface.

8. After the analyzer prepares for the QC test, apply the reconstituted control solution. a. Horizontally position the transfer pipette so that the tip is almost touching the front edge of

the test strip.b. Gently squeeze the pipette base containing the control solution to apply some (a minimum

of 6 μL) to the test strip target area. Capillary action draws the control solution into the test strip target area.

c. The Xprecia Stride sounds an audible tone when the test strip target area contains enough control solution.

d. Be careful to not overfill the test strip target area. After the test finishes, read the result on screen.

9. Press the Test Strip Eject button to discard the test strip according to your facility’s biohazard control policies. Tip When ejecting a used test strip; always point the Xprecia Stride analyzer down facing your biohazard container before you press the Test Strip Eject button.

10.Follow the instructions in the user manual for cleaning and disinfecting the entire exterior surface of the analyzer, and the test strip port protective cap, with a Siemens recommended germicidal wipe. Requirement You must clean and disinfect the device after each test.

11.Remove your gloves, thoroughly wash and dry your hands, and put on a new pair of gloves before performing a patient test.

IMPORTANTDon't use any non-recommended germicidal wipes to clean the analyzer, as they will damage the exterior.

Performing the PT/INR Test Preparing the patient samples for testing Prepare the patient to collect a blood sample by following your facility’s standard procedure. For example, clean the finger with an alcohol wipe, or have the patient wash his or her hands in warm, soapy water.

Dry the fingertip before taking a blood sample.

Only use fresh whole blood from a capillary (fingerstick) source with test strips.

8 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 9: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

IMPORTANT

– Within 15 seconds of sticking the fingertip, apply the drop of blood to the test strip target area.– Don’t add more blood to the test strip once the test has begun. – Don’t touch the test strip while the test is in progress. – Discard expired test strip vials.

INSTRUCTIONS

1. To turn the Xprecia Stride Analyzer on, press power on button.2. If you are prompted, enter an Operator ID (OID). 3. On the Home screen, tap the Patient Test Icon.4. If you are prompted, enter a Patient ID (PID). 5. Open the test strip vial and remove 1 test strip. 6. Immediately close the vial. Make sure the vial cap seals tightly. 7. To scan the test strip barcode, aim the barcode reader at the barcode on the test strip, then tap

SCAN .

Tip An audible tone sounds, a check mark displays, and the screen changes when the barcode is accepted.

8. Gently, but firmly, insert the test strip with the printed side up into the test strip port until it stops.9. (Optional) If the analyzer requests that you scan the test strip vial, aim the barcode reader at the

barcode on the test strip vial, then tap SCAN icon. 10. The Wait screen displays as the analyzer warms the test strip until it reaches operating temperature,

approximately 30 seconds.11. After the Apply Sample screen displays, prepare and apply the patient blood sample. See the

following: ‒Collecting a fingerstick blood sample on page 38. The test begins when the sample is drawn into the test strip by capillary action.

RequirementBe sure to fill the entire target area with blood. Don’t overfill the target area.

12. After the Test in Progress screen displays, don’t touch the test strip or add more blood. The INR value, time, and date display when the test is complete.

Tip You can set the analyzer down to tend to the patient while the test is in progress.

13. Press the Test Strip Eject button to discard the test strip according to your facility’s biohazard control policies.

Requirement When ejecting a used test strip, always point the analyzer down facing your biohazard container before you press the Test Strip Eject button.

14. Discard the lancet according to your facility’s biohazard control policies. 15. Follow the instructions in the User Guide to clean the entire exterior surface of the analyzer, and the

test strip port protective cap, with a Siemens recommended germicidal wipe. Requirement You must clean and disinfect the device after each test.

16. Remove your gloves, thoroughly wash and dry your hands, and put on a new pair of gloves before performing another patient test.

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-029 of 16

Page 10: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

IMPORTANT Don't use any non-recommended germicidal wipes to clean Stride, as they will damage the exterior.

Entering Patient ID (PID)INSTRUCTIONS

1. Tap the Settings Icon.

2. Tap Administrator Icon.

3. Tap Patent ID Icon.

4. Tap Patient Test to allow Patient ID (PID) scanning for patient tests.

5. 5. Tap STAT Test to allow an operator to skip Patient ID (PID) scanning for patient tests.

6. Tap to choose the Entry Type: ‒ for barcode reader ‒ for keypad

7. Tap Minimum.

8. Tap (-) or (+) to enter the minimum number of characters for the PID.

9. Tap Maximum.

10. Tap (-) or (+) to enter the maximum number of characters for the PID.

11. To accept your changes and return to the previous screen, tap Return Arrow .

10 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 11: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

Removing the Test Strip

1. Press the Test Strip Eject button to discard the test strip according to your facility’s biohazard control policies.

Requirement When ejecting a used test strip, always point Stride down facing your biohazard container before you press the Test Strip Eject button.

2. Discard the lancet according to your facility’s biohazard control policies.

3. Follow the instructions in the User Manual to clean the entire exterior surface of the analyzer, and the test strip port protective cap, with a Siemens recommended germicidal wipe.

Requirement You must clean and disinfect the device after each test.

4. Remove your gloves, thoroughly wash and dry your hands, and put on a new pair of gloves before performing another patient test.

IMPORTANT Don't use any non-recommended germicidal wipes to clean Stride, as they will damage the exterior.

Post-analytic Process

Reporting ResultsResults display as the International Normalized Ratio (INR). Desired INR values may vary depending upon the clinical practice and test methodologies, and the optimum therapeutic range for this method should be established by each user.2 Normal values can vary and each site should establish its own reference interval. Based on test samples from 120 non-anticoagulated subjects, the normal INR range for the Xprecia Stride PT/INR was 0.9–1.1.

PT/INR results may be affected due to variations in the type of reagents and/or instruments.Each lot of Xprecia System PT/INR test strips is calibrated to a reference lot of human recombinant thromboplastin traceable to the World Health Organization International Reference Preparation.3

Note Patients with long clotting times (>4.5 INR) are outside the reporting range. Confirm results by using an alternative test method. Contact the patient’s physician.

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-0211 of 16

Page 12: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

UNUSUAL RESULTSIf an unexpected result is obtained or if you are concerned that the result doesn’t match the clinical symptoms or patient history, perform a QC test using the Xprecia System PT Controls (as described in the Xprecia System PT Controls Instructions for Use) to confirm the device is working correctly, and then repeat the patient test.

If the results are confirmed, more in-depth testing may be necessary. INR results inconsistent with the patient’s clinical presentation may be an indication of improper test strip storage, interference with certain drugs, or changes in the patient’s diet.

Reference IntervalAdd your laboratory-specific expected values here

Critical ValuesAdd your laboratory-specific critical values here

Reporting Protocol for Critical ValuesAdd your laboratory-specific protocol for reporting results here

Units for Reporting ResultsThe system reports Prothrombin Time in INR.

Acceptable ResultsAdd your laboratory-specific acceptance criteria here

Patient test results are acceptable and may be reported when:

Corrective Action Add your laboratory-specific repeat and corrective action protocol here

Patient test results must be repeated and corrective action taken when:

Procedure NotesCalculationsAll measurements and calculations are performed automatically by the Xprecia Stride Analyzer, and the screen displays the PT results in INR.

12 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 13: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

DisposalDispose of hazardous or biologically contaminated materials according to the practices of your institution. Discard all materials in a safe and acceptable manner, and in compliance with all federal, state, and local requirements.

Method LimitationsLIMITATIONSThe PT/INR assay measures results in patients on warfarin therapy. Don’t use this assay to analyze patients on heparin therapy. In vitro studies showed analyzer tests results are not affected by heparin levels up to 3 U/mL, low molecular weight heparin up to 2 IU anti factor Xa activity/mL, ascorbic acid levels up to 1.5 mg/dL, triglycerides up to 3270 mg/dL, hemoglobin up to 200 mg/dL, acetaminophen up to 20 mg/dL, Oritavancin up to 5 mg/L, Fondaparinux up to 2.5 mg/L, Clopidogrel concentration up to 40 mg/L, Daptomycin concentration up to 300 mg/L, uric acid up to 24 mg/dL , unconjugated bilirubin up to 20 mg/dL, and conjugated bilirubin up to 29 mg/dL.

PT/INR results may be affected by many commonly administered drugs and further studies should be made to determine the source of unexpected abnormal results. If the presence of these drugs are known or suspected, consider monitoring patients for signs and symptoms of bleeding.

Hematocrit range between 22–52% doesn’t significantly affect test results.

Reportable RangeReporting Range: 0.8–4.5 INR. Results outside this range will not be displayed.

Refer patients with INR >4.5 to laboratory testing.

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-0213 of 16

Page 14: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

Adding an Operator

1. Tap Settings Icon.

2. Tap Administrator Icon.

3. Tap the Lock Icon.

4. Tap Enable to require a password to use the analyzer.

5. Tap Validate to compare the password against a list of valid passwords stored on the analyzer.

6. Tap to choose the Entry Type: for barcode reader or for keypad

7. Tap Minimum.

8. Tap (-) or (+) to enter the minimum number of characters for the login.

9. Tap Maximum.

10. Tap (-) or (+) to enter the minimum number of characters for the login.

11. To accept your changes and return to the previous screen, tap .

DMS Software for managing data

The Xprecia Data Management Software (DMS) provides a simple and easy way to transfer data from the analyzer. Once a number of test results have been collected over a designated period of time, they can be uploaded into DMS. From there, test results from the DMS can be exported for further evaluation.

Requirement A computer running DMS is necessary to configure and upload results from the analyzer. To install DMS and start uploading data, insert the DMS installation media into your computer and follow the installation instructions when prompted. Then plug the USB cable from your computer into the Xprecia Stride and use DMS to export the test results.

Using DMS, you can: ‒Set up and manage operators ‒Configure and upgrade Xprecia Stride devices ‒Search and export both patient test results and QC test results ‒View error and fault log information on the Xprecia Stride analyzers

IMPORTANTWhen connected, the Xprecia DMS overwrites any analyzer user settings and updates the Operator ID (OID) list with the latest updates from DMS to the Xprecia Stride Analyzer.

14 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08

Page 15: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert hospital name here Insert department name here

Equipment and Supplies

ReferencesREFERENCES1. Quick A J, Stanly-Brown M, and Bancroft F W: A study of the coagulation defect in hemophilia and in jaundice. Am. J. Med. Sci 190: 501-511, 1935.2. Hirsh J, Dalen J E, Anderson DR, Poller L, Bussey H I, Ansell, J, Deykin D: Oral anticoagulants: mechanism of action, clinical effectiveness, and optimal therapeutic range. Chest 119: 8S-21S, 2001.3. WHO Expert Committee on Biological Standardization. Thirty-third Report. Geneva, World Health Organization, 1983 (WHO Technical Report Series, No. 687).

Technical AssistanceSiemens Healthcare Diagnostics Technical Care Center: 1-877-229-3711, Prompt 14, 1

Customer Service: 1-800-255-3232

Serial Number:

Customer Account Number:

Xprecia Stride PT/INR Technical Procedure, Rev. 1.0, 2017-0215 of 16

Page 16: Assay Name - This is a title style_+D - Siemens€¦  · Web viewAlways store the test strips in the original test strip vial with the cap closed. Use the test strip within 5 minutes

Insert department name here Insert hospital name here

Trademark InformationXprecia Stride is a trademark of Siemens Healthcare Diagnostics

© 2017 Siemens Healthcare Diagnostics

16 of 16 DCA HbA1c Technical Procedure, Rev. 3.0, 2012-08