anaesthesia machine 1
TRANSCRIPT
The Anaesthesia Machine - 1
Presented by- Dr. Vineet Chowdhary
Moderator- Dr. Avnish Bharadwaj
NO EQUIPMENT IS MORE INTIMATELY ASSOCIATED WITH THE PRACTICE OF
ANAESTHESIOLOGY THAN THE ANAESTHESIA MACHINE
Anaesthesia WorkstationAn anesthesia workstation integrates most of the components necessary for administration of anesthesia into one unitIt is a device which delivers a precisely-known but variable gas mixture, including anaesthetizing and life-sustaining gases. Consists of:The anesthesia machineVentilatorBreathing systemScavenging system MonitorsAdded to this may be drug delivery systems, suction equipment, and a data management system
HistoryThe original concept of Boyle's machine was invented by the British anaesthetist H.E.G. Boyle in 19171920 – A vapourizing bottle is incorporated
to the machine.1926 – A 2nd vaporizing bottle and by-pass
controls are incorporated.1930 – A Plunger device is added to the
vaporizing bottle.1933 – A dry-bobbin type of flowmeter is
introduced.1937 – Rotameters displayed dry-bobbin
type of flowmeters
Types of Anesthesia MachineIntermittent-Gas flows only during inspiration
Egs: Entonox apparatus, Mackessons apparatus
Continuous-Gas flows both during inspiration and expiration. Egs :
Boyle Machine Forregar Dragger
Standards for Anesthesia Machines and Workstations
Standards for anesthesia machines and workstations provide guidelines to manufacturers regarding their minimum performance, design characteristics, and safety requirements. During the past 2 decades, the progression of anesthesia machine standards has been as follows:
1979: American National Standards Institute 1988: American Society for Testing and Materials, 1994: ASTM F1161-94 (reapproved in 1994 and
discontinued in 2000) 2005: International Electrical Commission (IEC) 2005: ASTM (reapproved)F1850 European standard is EN740
Contd….
To comply with the 2005 ASTM F1850-00 standard, newly manufactured workstations must have monitors that measure the following parameters:
continuous breathing system pressure, exhaled tidal volume, ventilatory CO2 concentration, anesthetic vapor concentration, inspired oxygen concentration, oxygen supply pressure, arterial hemoglobin oxygen saturation arterial blood pressure, and continuous electrocardiogram.
Contd…
The anesthesia workstation must have a prioritized alarm system that groups the alarms into three categories: high, medium, and low.
These monitors and alarms may be enabled automatically and made to function by turning on the anesthesia workstation, or the monitors and alarms can be enabled manually and made functional by following a pre-use checklist.
Basic Schematics
System components
Electrical Pneumatic
1.Master Switch2.Power Failure Indicator3.Reserve Power4.Electrical Outlet5.Circuit Breakers6.Data Communication Port
1.High Pressure System2.Intermediate System3.Low Pressure System
Electrical ComponentsMaster Switch
Master (main power) switch activates both the pneumatic and electrical functions . On most machines, when the master switch is in the OFF position, the only electrical components that are active are the battery charger and the electrical outlets
Standby position - allows the system to be powered up quickly
Computer-driven machines should be turned OFF and restarted with a full checkout at least every 24 hours.
STANDBY mode is not used for an extended period.
Most machines are equipped with a visual and/or audible indicator to alert the anesthesia provider to the loss of mains power .The machine will usually give an indication when mains power is lost.
Power Failure Indicator
Reserve PowerBackup source of power for the occasional outage is necessary. The anesthesia provider should check the battery status during the preuse checkout procedure.While some older anesthesia machines used replaceable batteries, most new machines use rechargeable batteries.It usually takes a number of hours to fully recharge a battery after it has completely discharged.
Electrical Outlets Most modern anesthesia machines have electrical
outlets. These are intended to power monitors and other
devices. As a general rule, these outlets should only be used
for anesthesia monitors. Other appliances should be connected directly to
mains power.
Next to each outlet is a circuit breaker.
There are circuit breakers for both the anesthesia machine and the outlets .When a circuit breaker is activated, the electrical load should be reduced and the circuit breaker reset
Circuit Breakers
Data Communication Ports Most modern anesthesia machines have
data communications ports. These are used to communicate between
the anesthesia machine, monitors, and the data management system
Gases are supplied under tremendous pressure for the convenience of storage and transport.
The anaesthesia machine receives medical gases from a gas supply; controls the flow of desired gases reducing their pressure, to a safe level.
So the pressure inside a source ( cylinder or pipeline ) must be brought to a certain level before it can be used for the purpose of ventilation.
And it needs to be supplied in a constant pressure, otherwise the flow meter would need continous adjustment.
.
This is achieved by bringing down the pressure of a gas supply in a graded manner with the help of three pressure reducing zones .
Thus the pneumatic part of the machine can be conveniently divided into three parts- high ,intermediate and low pressure systems
• Consists of:Consists of:– Hanger YolkHanger Yolk – Check valve Check valve
– Cylinder Cylinder Pressure Pressure Indicator Indicator (Gauge)(Gauge)
– Pressure Pressure Reducing Reducing Device Device (Regulator)(Regulator)
• Usually not Usually not used, unless used, unless pipeline gas pipeline gas supply is offsupply is off
Hanger Yoke Assembly The Hanger yoke assembly
1) Orients and supports the cylinder
2) Provides a gas-tight seal
3) Ensures uni-directional gas flow
The workstation standard recommends that there be at least one yoke each for oxygen and nitrous oxide.
If the machine is likely to be used in locations that do not have piped gases, it is advisable to have a double yoke, especially for oxygen.
BODY It is threaded into the frame of the machine.
It provides support for the cylinder(s).
Commonly the swinging gate type is used.
When a cylinder is mounted onto or removed from a yoke, the hinged part can be swung to side.
RETAINING SCREW It is threaded into the distal end of the yoke.
Tightening the screw presses the outlet of the cylinder valve against the washer and the nipple so that a gas tight seal is obtained.
The cylinder is then supported by the retaining screw, the nipple, and the index pins.
The conical point of the retaining screw is shaped to fit the conical depression on the cylinder valve.
NippleNipple
It is a part of the yoke It is a part of the yoke through which the gas enters through which the gas enters the machine.the machine.
It fits into the port of It fits into the port of the cylinder valve.the cylinder valve.
If it is damaged, it may be If it is damaged, it may be impossible to obtain a tight impossible to obtain a tight seal with the cylinder seal with the cylinder valvevalve..
These are situated below the nipple.
These help to prevent mounting of incorrect cylinder to yoke.
The holes into which the pins are fitted must be of a specific depth.
If they extend too far into the body of the yoke, it may be possible to mount a incorrect cylinder.
INDEX PINS
Bodok seal-cylinders are fitted with yoke with a sealing washer called BODOK SEAL
-it is made up of non combustible material and has a metal periphery which make it long lasting.
-it should be less than 2.4mm thick prior to compression.
-only one seal should be use between the valve & yoke
Caution!!The wrong Cylinder may be fitted by:
1.Using extra sealing washers2.Removing of/ Wearing of the Index Pins3.Placing an inverted gas cylinder
FilterIt is used to prevent particulate matter from entering the machine.
It is to be placed between the cylinder and the pressure reducing device.
FILTER
Placing cylinder in yoke
Placing a Cylinder in a Yoke1.Cylinder valves and yokes not be contaminated with oil or grease2. Persons placing a cylinder in a yoke should always wash their hands first3. Pin Index Safety System pins are present4.Retract the retaining screw5. The washer is placed over the nipple6.The cylinder is supported by the foot and guided into place manually 7. The port on the cylinder valve is guided over the nipple
and the index pins engaged in the appropriate holes8. The retaining screw is tightened 9. Do not insert the screw in the safety relief device10. Make certain that the cylinder is full and that there is no leak
CHECK VALVE ASSEMBLY
It allows gas from a cylinder to enter the machine but prevents gas from exiting the machine when there is no cylinder in the yoke.
It allows an empty cylinder to be replaced with a full one without having to turn off the `in–use` cylinder.
Prevents transfer of gas from one cylinder to the other with a lower pressure in a double yoke.
It consists of a plunger that slides away from the side of the greater pressure.
It is not designed to act as a permanent seal for empty yoke and may allow small amount of gas to escape.
As soon as a cylinder is exhausted it should be replaced by a full one or a dummy plug.
In order to minimize losses –
Yokes should not be left vacant for extended periods
An empty cylinder should be replaced as soon as possible
An yoke plug can be used to prevent gas leak or
An empty cylinder can be left behind after closing the
valve
A Bourdon tube is a hollow metal tube(copper alloy) bent into a curve, then sealed on one side and linked to a clock like mechanism
Safety features in Cylinder Pressure Indicator
Gauge is usually color coded.Name and symbol of gas are written over dial.If bourdon tube ruptures gas is vented from back side Gauges are angled and placed in such a way that it can be easily read by anesthetist.Instructions like “use no oil’’ “open the valve slowly’’ are written on the gauge
Electronic Cylinder Pressure IndicatorLight emitting diodes(LED’S)in electronic pressure gauge indicateCylinder valve is close –Dark colorCylinder valve is open –
Pressure adequate –Green Pressure inadequate-Red
PRESSURE REDUCING DEVICE (REGULATOR)
The pressure in a cylinder varies. The anesthesia machine is fitted with devices (reducing valves, regulators, reducing regulators, reduction valves, regulator valves) to maintain constant flow with changing supply pressure.
These reduce the high and variable pressure found in a cylinder to a lower (40 to 48 psig, 272 to 336 kPa) and more constant pressure suitable for use in an anesthesia machine.
The machine standard requires reducing devices for each gas supplied to the machine from cylinders.
Physical Principle- A large pressure acting over a small area is balanced by a small pressure over a larger area
Pressure regulators have safety relief valvesIf due to any reason there is build up of pressure in pressure regulator then the safety valve blow off at a set pressure of 525 k pa(70psi)
Safety features on pressure regulator
INTERMEDIATE PRESSURE SYSTEMBegins at the regulated cylinder supply source at 45 psig includes the pipeline sources at 50 to 55 psig and extends to the flow control valve.
Consists of:Pipeline inlet connectionsPipeline pressure indicatorsPipingGas power outletMaster switchOxygen pressure failure devicesOxygen flushAdditional reducing devicesFlow control valves
Check valve
MASTER SWITCH (PNEUMATIC COMPONENT ) The pneumatic portion of the master switch is located
in the intermediate pressure system downstream of the inlets for the cylinder and pipeline supplies
The oxygen flush is usually independent of this switch.
The master switch may be a totally electronic switch that when activated controls the various pneumatic components in the anesthesia machine.
When the master switch is turned off ,the pressure in the intermediate system will drop to zero
PIPELINE INLET CONNECTIONS It is the entry point for gases from the pipelines. The anesthesia workstation standard requires pipeline
inlet connections for oxygen and nitrous oxide. Most machines also have an inlet connector for air. These inlets are fitted with threaded non
interchangeable Diameter Index Safety System (DISS) fittings
A unidirectional (check) valve prevents reversed gas flow from the machine into the piping system
Each pipeline inlet is required to have a filter with a pore size of 100μm or less. The filter may become clogged, resulting in a reduction in gas flow.
PIPELINE INLET CONNECTIONS
PIPELINE PRESSURE INDICATORS Indicators to monitor the pipeline pressure of each
gas are required by the anesthesia workstation standard.
They are usually found on a panel on the front of the machine and may be color coded for the gases that they monitor
The workstation standard requires that the indicator be on the pipeline side of the check valve in the pipeline inlet.
If the indicator is on the pipeline side of the check valve, it will monitor pipeline pressure only. If the hose is disconnected or improperly connected, it will read “0” even if a cylinder valve is open
If the indicator were on the machine (downstream) side of the check valve, it would not give a true indication of the pipeline supply pressure unless the cylinder valves were closed. If a cylinder valve is open and the pipeline supply fails, there will be no change in the pressure on the indicator until the cylinder is nearly empty.
Pipeline pressure indicators should always be checked before the machine is used. The pressure should be between 50 and 55 psig (345 and 380 kPa). The indicators should be scanned repeatedly during use.
Piping is used to connect components inside the machineIt must be able to withstand four times the intended service pressureLeaks between the pipeline inlet or cylinder pressure reducing system and the flow control valve not exceed 25 mL/minuteIf the yoke and pressure reducing system are included, the leakage may not exceed 150 mL/minute.
PIPING
Some machines have a gas selector switch that prevents air and nitrous oxide from being used together.
GAS SELECTOR SWITCH
GAS POWER OUTLETOne or more gas power (auxiliary gas) outlets may be present on an anesthesia machine. It may serve as the source of driving gas for the anesthesia ventilator or to supply gas for a jet ventilator. Either oxygen or air may be used.The ventilator is an integral part of the modern machine and the breathing system and is connected to the ventilator with internal piping. Therefore, the power outlet is not found in many anesthesia machines today.
OXYGEN PRESSURE FAILURE DEVICES
One of the most serious mishaps that occurred with early machines was depletion of the oxygen supply (usually from a cylinder) without the user awareness.
The result was delivery of 100% anesthetic gas. Numerous inventions have been devised to prevent this
Oxygen Failure Safety Device
Oxygen Supply Failure Alarm
Pressure Sensor Shut-off Valve: Datex OhmedaPressure Sensor Shut-off Valve: Datex Ohmeda
Operates in a threshold manner: either open or shutOxygen pressure moves the piston and pin upward and the valve opens for N2OWhen pressure of oxygen falls below preset value, force of the valve return spring completely closes the valve
Oxygen Failure Protection Device: DragerBased on a proportioning principle rather than a threshold principlePressure of N2O falls in Propotion of decrease of Oxygen.Total cutoff seen at <12psig.Seat nozzle assembly connected to a spring loaded conical tapered piston
Oxygen Supply Failure Alarm
ASTM standard specifies that whenever the oxygen supply pressure falls below a certain threshold (usually 30 psig), alarm must get activated within 5 seconds. It should not be possible to disable this alarmThey aid in preventing hypoxia caused by problems occurring upstream in the machine circuitry (disconnected oxygen hose, low oxygen pressure in the pipeline, and depletion of oxygen cylinders)
CAUTION!!
These devices do not offer total protection against a hypoxic mixture being delivered, because they do not prevent anesthetic gas from flowing if there is no flow of oxygen.
Equipment problems (such as leaks) or operator errors (such as a closed or partially closed oxygen flow control valve) that occur downstream are not prevented by these devices.
They do not guard against accidents from crossovers in the pipeline system or a cylinder containing the wrong
gas.
Second-stage Pressure RegulatorSecond-stage Pressure RegulatorSome machines have pressure regulators in the intermediate pressure system just upstream of the flow indicatorsReduce the pressure further to around 26 psi (177 kPa) for nitrous oxide and 14 psi (95 kPa) for oxygenThe purpose of this pressure regulator is to eliminate fluctuations in pressure supplied to the flow indicatorsBy reducing the pressures below the normal fluctuation range, the flow will remain more constant. Not all anesthesia machines are equipped with this device.
OXYGEN FLUSH The oxygen flush (oxygen bypass, emergency oxygen
bypass) receives oxygen from the pipeline inlet or cylinder pressure regulator and directs a high unmetered flow directly to the common gas outlet.
It is commonly labeled “02+.” On most anesthesia machines, the oxygen flush can be
activated regardless of whether the master switch is turned ON or OFF.
A flow between 35 and 75 L/minute must be delivered. The button is commonly recessed or placed in a collar
to prevent accidental activation.
It consists of a button and stem It consists of a button and stem connected to a spring loaded connected to a spring loaded ball .The ball is in contact with the ball .The ball is in contact with the seat .When the button is depressed, seat .When the button is depressed, the ball is forced away from the the ball is forced away from the seat , allowing the oxygen to flow to seat , allowing the oxygen to flow to the outlet. A spring opposing the the outlet. A spring opposing the ball will close the valve when the ball will close the valve when the button is not depressed . Delivers button is not depressed . Delivers oxygen at 60 psig.oxygen at 60 psig.
Reported hazards associated with the oxygen flush include
Accidental activation – causing oxygen-enriched gas mixture, anaesthetic dilution
The flush valve stuck in the ON position Barotrauma and awareness during anesthesia Internal leakage
The anesthesia workstation standard requires that the connection of the flush valve delivery line to the common gas outlet be designed so that activation does not increase or decrease the pressure at the vaporizer outlet by more than 10 kPa or increase the vapor output by more than 20%.
FLOW ADJUSTMENT CONTROL Controls flow of gas through it’s associated indicator by manual adjustment of a variable orificeCurrent standard requires that there be only one flow control for each gas. It must be adjusted or identifiable with its flow indicator
CONTROL KNOBCONTROL KNOBTouch and colour coded
Joined to stem
Large enough to be turned easily
ROTATORY STYLE KNOBSROTATORY STYLE KNOBSOxygen knob- Fluted Profile, as large/ larger than any other gas knobKnobs turned counter clockwise- increase flowKnobs turned clockwise- decrease flowAll other knobs should be roundOxygen knobs must look and feel different than other knobsThey should operate smoothlyKnob should not be over turned during closure because further tightening may damage the pin/seat
CAUTION!!CAUTION!!Loose or worn knobs may respond to light touch or accidental brushingLeakage through open flow control valvesInability to turn the knobFailure to allow adequate gas flow