KMSCL – Tender No: 02/2010 – Amendment Notice Page 1 of 15
KERALA MEDICAL SERVICES CORPORATIONKERALA MEDICAL SERVICES CORPORATIONKERALA MEDICAL SERVICES CORPORATIONKERALA MEDICAL SERVICES CORPORATION
((((DEPARTMENT OF HEALTH & FAMILY WELFARE, GOVT. OF KERALADEPARTMENT OF HEALTH & FAMILY WELFARE, GOVT. OF KERALADEPARTMENT OF HEALTH & FAMILY WELFARE, GOVT. OF KERALADEPARTMENT OF HEALTH & FAMILY WELFARE, GOVT. OF KERALA))))
BEHIND W&C HOSPITAL, THYCAUD P.OBEHIND W&C HOSPITAL, THYCAUD P.OBEHIND W&C HOSPITAL, THYCAUD P.OBEHIND W&C HOSPITAL, THYCAUD P.O
THIRUVANANTHAPURAMTHIRUVANANTHAPURAMTHIRUVANANTHAPURAMTHIRUVANANTHAPURAM----14141414, KERALA, KERALA, KERALA, KERALA
TELEPHONE NO. 0471TELEPHONE NO. 0471TELEPHONE NO. 0471TELEPHONE NO. 0471----3190808, 4019928, 23373533190808, 4019928, 23373533190808, 4019928, 23373533190808, 4019928, 2337353
Tele. Fax:Tele. Fax:Tele. Fax:Tele. Fax:----0471047104710471----4015522401552240155224015522
No.No.No.No.6910(T02)6910(T02)6910(T02)6910(T02)/BME//BME//BME//BME/KMSCLKMSCLKMSCLKMSCL////2010201020102010 Dated Dated Dated Dated 18181818----00009999----2020202010101010
EQUIPMENT TENDER EQUIPMENT TENDER EQUIPMENT TENDER EQUIPMENT TENDER –––– AMENDMENT AMENDMENT AMENDMENT AMENDMENT ---- NOTICENOTICENOTICENOTICE
Sub : Tender No.02/2010 released on 06/09/2010 – amendments published –
reg
Ref : 1. Discussions in the pre-tender meeting for Anaesthesia equipments
held at 2 p.m and Blood Bank equipments held at 10 a.m on
15/09/2010.
2. Request for amendment submitted by various prospective tenderers
The following amendments are made in respect of the tender clauses noted
herein on the basis of the request from prospective tenderers, expert opinion and
market study.
ANAESTHESIAANAESTHESIAANAESTHESIAANAESTHESIA EQUIPMENTSEQUIPMENTSEQUIPMENTSEQUIPMENTS
GENERAL AMENDMENTSGENERAL AMENDMENTSGENERAL AMENDMENTSGENERAL AMENDMENTS
It was decided to conduct the demonstration of all the Anesthesia equipments
at KMSCL head office.
KMSCL – Tender No: 02/2010 – Amendment Notice Page 2 of 15
SPECIFIC AMENDMENTSSPECIFIC AMENDMENTSSPECIFIC AMENDMENTSSPECIFIC AMENDMENTS
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----119119119119----02/201002/201002/201002/2010 : Suction Apparatus: Suction Apparatus: Suction Apparatus: Suction Apparatus
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
V 5.1.1
Installation & Delivery period
: 8 weeks from the date of
issuance of letter of intent
Installation & Delivery
period : 10 weeks from
the date of issuance of
letter of intent
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----50505050----02/201002/201002/201002/2010 : Fumigation Dispensor: Fumigation Dispensor: Fumigation Dispensor: Fumigation Dispensor
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 6
Should have a stainless
steel construction for the
tank and the bolts,
nuts and washers used
inside the tank.
Should have a tough and
corrosion resistant tank. The
bolts, nuts and washers
used inside the tank should
be stainless steel / brass.
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----03030303----02/2010 : Anaesthesia Machine02/2010 : Anaesthesia Machine02/2010 : Anaesthesia Machine02/2010 : Anaesthesia Machine
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing Existing Existing Existing clauseclauseclauseclause Amended clauseAmended clauseAmended clauseAmended clause
V 5.1.1
Installation & Delivery
period : 8 weeks from the
date of issuance of letter
of intent
Installation & Delivery
period : 10 weeks from
the date of issuance of
letter of intent
Appendix
I 2
Gas cylinder (pin indexed)
yokes with sliding stainless
steel clamping bars for
easy handling.
Gas cylinder (pin
indexed) yokes with
sliding stainless steel
/sturdy clamping bars for
easy handling.
Appendix
I 4
Regulator two each for O2
and N2O with output
pressure 4.22kg/cm2.
N2Oregulator is activated
only when oxygen
regulators are pressurized.
Regulator two each for
O2 and N2O with output
pressure 4.22kg/cm2.
N2Oregulator is activated
only when minimum
oxygen on flow.
KMSCL – Tender No: 02/2010 – Amendment Notice Page 3 of 15
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing Existing Existing Existing clauseclauseclauseclause Amended clauseAmended clauseAmended clauseAmended clause
Appendix
I 12
Non-return cum pressure
relief valve when pressure
exceeds 200cmof H2O
Non-return cum pressure
relief valve when pressure
exceeds 120cmof H2O
Appendix
I 17
Should have safety
certificate from a
competent authority CE /
FDA (US) / STQC CB
certificate / STQC S
certificate or valid detailed
electrical and functional
safety test report from
ERTL. Copy of the
certificate / test report shall
be produced along with the
technical bid.
Should have safety
certificate from a
competent authority CE /
FDA (US) / STQC CB
certificate / STQC S
certificate or valid detailed
electrical and functional
safety test report from
ERTL / ISI (IS.13485).
Copy of the certificate /
test report shall be
produced along with the
technical bid.
Appendix
I 19
The table top made up of
stainless steel size approx
550x 350mm.
The table top made up of
stainless steel / chemical
resistant fiber
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----135135135135----02/201002/201002/201002/2010 : Ventilator ICU: Ventilator ICU: Ventilator ICU: Ventilator ICU
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
V 5.2.5
The tenderers who
have a minimum
annual turnover of `5
Crores every year
during the last three
completed financial
years.
The tenderers who have a
minimum average annual
turnover of ` 3 Crores during
the last three completed
financial years.
Appendix I III Nebulizer - Complete
set. Nebulizer – Ultrasonic one
complete set
Appendix I III
Patient circuit (Adult
reusable) - 1 complete
set.
Patient circuit (Adult reusable) -
2 complete set.
Patient circuit (Paediatric
reusable) - 1 complete set. The
rate for paediatric set has to be
offered separately and the same
KMSCL – Tender No: 02/2010 – Amendment Notice Page 4 of 15
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
will be ordered wherever
necessary. The price of
paediatric set offered will be
considered for bid evaluation.
Appendix I IV
Monitor with ELD/TFT
(10” or higher size)
graphical display for
real time simultaneous
display of two
waveforms.
Monitor with LCD/TFT (10” or
higher size) graphical display for
real time simultaneous display of
two waveforms. Should display
minimum 3 graphs and 2 loops
and may not simultaneously.
Appendix I V
1. Gas delivery system
by sound less
compressor integrated
with the unit.
2. It should also be
operable with
compressed Air 45 to
60 psi.
1. Gas delivery system by sound
less in built compressor /
external integrated compressor
with the unit.
2. In case of compressor failure
it should also be operable with
compressed air / oxygen supply
of 45 to 60 psi.
Appendix I VI
Internal battery
(maintenance free)
with 1 hour minimum
operating time.
Internal battery (maintenance
free) with 1 hour minimum
operating time for the
ventilator
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix
I II Nil
Trigger (Non Invasive Ventilation) -
0 to 5 cms H2O from baseline
flow/Pressure
Appendix
I VIII Nil
1. Automatic leak compensation
(Non Invasive Ventilation) during
triggering and cycling – 50 lpm
or more.
2. Should provide reusable mask of
medium size – 1no, large size – 1
no.
3. Back-up ventilation.
Appendix
I IX Nil
Replacement guarantee should be
provided for flow sensors and
oxygen sensor / analyser for the
KMSCL – Tender No: 02/2010 – Amendment Notice Page 5 of 15
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
entire 3 years warranty period and
also the rate offered for CMC
should include the replacement
guarantee for flow sensors and
oxygen sensor / analyser.
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----62626262----02/201002/201002/201002/2010 : Infusion Pump: Infusion Pump: Infusion Pump: Infusion Pump
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 8
Should have a LCD
display with backlight
and graphical display of
infusion
Should have a LED /
LCD display with
backlight and display
of infusion
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----75757575----02/201002/201002/201002/2010 : Multiparameter Monitor: Multiparameter Monitor: Multiparameter Monitor: Multiparameter Monitor
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
V 5.2.5
The tenderers who have a
minimum annual turnover
of ` 2 Crores every year
during the last three
completed financial years.
The tenderers who have a
minimum average annual
turnover of ` 2 Crores for
the last three completed
financial years.
Appendix I 3
Should have TFT display
with at least 8 inches or
higher with 4 wave forms
and numeric display
simultaneously. The
waveforms should be
user selectable.
Should have TFT display
with at least 10 inches or
higher with 3/4 wave forms
and numeric display
simultaneously. The
waveforms should be user
selectable.
Appendix I 10.3 Disposable SPO2 probes
for neonatal use.
Disposable SPO2 probes for
neonatal use – 1 no
KMSCL – Tender No: 02/2010 – Amendment Notice Page 6 of 15
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----121121121121----02/201002/201002/201002/2010 : Syringe pump: Syringe pump: Syringe pump: Syringe pump
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 2 Should have large format
LCD/TFT display
Should have large format
LED/LCD/TFT display
Appendix I 3
Should have automatic
syringe loading with
automatic syringe
recognition.
Should have manual /
automatic syringe loading
with automatic syringe
recognition
Appendix I 11
Should have a
rechargeable battery with
back up time of
minimum 3 hours.
Should have a rechargeable
battery with back up time of
minimum 2 hours.
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----150150150150----02/201002/201002/201002/2010 : Anesthesia pump: Anesthesia pump: Anesthesia pump: Anesthesia pump
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 4
Battery life: more than
10 hours Battery life: 6 – 8 hours
Appendix I 10 Menu driven display Menu driven display –LCD
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----163163163163----02/2010 : NIBP monitor02/2010 : NIBP monitor02/2010 : NIBP monitor02/2010 : NIBP monitor
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 8
Should have minimum
350 readings storage
capacity
Should have minimum 50
readings storage capacity
Appendix I 7 Pulse range should be
20-250 pulse/min.
Pulse range should be 30-
200 pulse/min or more.
Appendix I 6 Measurement range
should be 0-300 mmHg.
Measurement range should
be 0-230 mmHg or more.
Appendix I 5 Auto mode (5-120 min.
int) Stat (5 min cont)
Auto mode (1-60 min. int or
more) Stat (5 min cont)
KMSCL – Tender No: 02/2010 – Amendment Notice Page 7 of 15
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----02020202----02/201002/201002/201002/2010 : Anaesthesia Workstation: Anaesthesia Workstation: Anaesthesia Workstation: Anaesthesia Workstation
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 3
Should provide with
adult and pediatric
silicon rubber tubing
breathing circuit
Should provide with adult
and pediatric reusable and
autoclavable light weight
tubing breathing circuit
Appendix I 7
Should have display for
set parameters and
graphical display for
measured parameters
Should have display of at
least 6 inches for set
parameters and graphical
display of measured
parameters
Appendix I 10
Gas cylinder (pin
indexed) yokes with
sliding stainless steel
clamping bars for easy
handling
Gas cylinder (pin indexed)
yokes with sturdy clamping
bars for easy handling
Appendix I 11
Two Pin index yokes for
connecting cylinders
each for O2, N2O and
air
Two Pin index yokes for
connecting cylinders each
for O2, N2O and air though
pipeline
Appendix I 12
Regulator two each for
O2 and N2O with output
pressure 4.22kg/cm2.
N2O regulator is
activated only when
oxygen regulators are
pressurized
Regulator two each for O2
and N2O. Nitrous oxide
should be activated only
with oxygen on flow
Appendix I 21
Non-return cum pressure
relief valve when
pressure exceeds
200cmof H2O.
Non-return cum pressure
relief valve when pressure
exceeds 120cmof H2O.
Appendix I 22
Should have change
over from open circuit to
closed circuit and vice
versa.
Should have only one
common gas outlet
Appendix I 24 Circle absorber with
corrugated reusable
Circle absorber with
corrugated reusable
KMSCL – Tender No: 02/2010 – Amendment Notice Page 8 of 15
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
breathing circuit for
closed circuit system with
each unit. It should be
autoclavable.
breathing circuit for closed
circuit system with each
unit. It should be
autoclavable. It should be
with ventilator selector
switch and circle on/off
switch
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----105105105105----02/201002/201002/201002/2010 : Pulse Oxymeter: Pulse Oxymeter: Pulse Oxymeter: Pulse Oxymeter
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 1
Should have
plethismographic wave
form with numeric
display for SPO2 and
Heart rate on TFT
display.
Should have
plethismographic wave form
with numeric display for
SPO2 and Heart rate on
LCD / TFT display.
Appendix I 6
Should provide with
adult reusable finger
probe.
Should provide with adult
reusable finger probe with
technology from standard
reputed companies.
KMSCL – Tender No: 02/2010 – Amendment Notice Page 9 of 15
BLOOD BANKBLOOD BANKBLOOD BANKBLOOD BANK EQUIPMENTSEQUIPMENTSEQUIPMENTSEQUIPMENTS
SPECIFIC AMENDMENTSSPECIFIC AMENDMENTSSPECIFIC AMENDMENTSSPECIFIC AMENDMENTS
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----36363636----02/2010 02/2010 02/2010 02/2010 :::: Donor couchDonor couchDonor couchDonor couch
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
II 2.2.20
The decision on the mode of
verification of specification of
the equipments will be
decided after the pre-tender
meeting by the corporation
on the basis of expert advice.
Demonstration shall
be arranged at third
party location and the
Tender Inviting
Authority reserves the
right to finalise the
location.
Appendix
I 2
Should have movable arm
rest
Should have movable
arm rest (swinging up
& down movements)
Appendix
I 4
Should have corded remote
control for initiating the
movements.
Should have keys for
initiating the
movements.
Appendix
I 9
Should have a lifting capacity
of 150 kg
Should have a lifting
capacity of 120 kg
Appendix
I 13
Should have safety certificate
from a competent authority
CE / FDA (US) / STQC CB
certificate / STQC S certificate
or valid detailed electrical and
functional safety test report
from ERTL. Copy of the
certificate / test report shall be
produced along with the
technical bid.
Clause deleted
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix
I 15 Nil
Should work in ambient
temperature of 10 to 400 C
KMSCL – Tender No: 02/2010 – Amendment Notice Page 10 of 15
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----128128128128----02/2010 : 02/2010 : 02/2010 : 02/2010 : Tube sealerTube sealerTube sealerTube sealer
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
II 2.2.20
The decision on the mode of
verification of specification of
the equipments will be
decided after the pre-tender
meeting by the corporation
on the basis of expert advice.
Demonstration shall
be arranged at KMSCL
head office.
Appendix
I 1
Should be a portable type
radio frequency sealing
system.
Should be a bench
top type radio
frequency sealing
system.
Appendix
I 16
Should have safety certificate
from a competent authority
CE / FDA (US) / STQC CB
certificate / STQC S
certificate or valid detailed
electrical and functional
safety test report from ERTL.
Copy of the certificate / test
report shall be produced
along with the technical bid.
Clause deleted
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----15151515----02/202/202/202/2010 : Blood bank refrigerator 010 : Blood bank refrigerator 010 : Blood bank refrigerator 010 : Blood bank refrigerator (Various types)
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
III 3.1.4
Earnest Money Deposit:
Blood Bank Refrigerator
(165L): Rs. 16,200
Blood Bank Refrigerator
(300L): Rs. 2,600
Total: Rs. 18,800 /-
Earnest Money Deposit:
Blood Bank Refrigerator
(165L): Rs. 16,200
Blood Bank Refrigerator
(300L): Rs. 26,000
Total: Rs. 42,200 /-
II 2.2.20
The decision on the mode
of verification of
specification of the
equipments will be
decided after the pre-
Demonstration shall be
arranged at third party
location and the Tender
Inviting Authority reserves
the right to finalise the
KMSCL – Tender No: 02/2010 – Amendment Notice Page 11 of 15
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
tender meeting by the
corporation on the basis of
expert advice.
location.
V 5.1.3 CMC/AMC period :
5 years
CMC/AMC period : 3
years
V 5.1.4
Frequency of visits to all
User Institution concerned
during Warranty/CMC or
AMC : One visit every
three months (4 visits in a
year)
Frequency of visits to all
User Institution
concerned during
Warranty/CMC or AMC :
One visit every six
months (2 visits in a
year)
Appendix I 1.11
Alarm system should be
incorporated with battery
backup for minimum 4hrs.
Alarm system should be
incorporated with battery
backup for minimum 2hrs.
Appendix I 1.13
Should have a CFC free,
Urethane foam insulation
(80-90mm) to protect
cabinet from ambient
temperature fluctuations
Should have a CFC free,
Urethane foam insulation
(50-90mm) to protect
cabinet from ambient
temperature fluctuations
Appendix I 1.17
Should have safety
certificate from a competent
authority CE / FDA (US) /
STQC CB certificate / STQC
S certificate or valid detailed
electrical and functional
safety test report from ERTL.
Copy of the certificate / test
report shall be produced
along with the technical bid.
Clause deleted
Appendix I 2.1
Capacity : Should be able
to accommodate 300
standard blood bags for
each of 450 ml capacity
Capacity : Should be
able to accommodate
120 numbers standard
blood bags for each of
350 ml capacity
Appendix I 2.3
Should have provision for
air condition
Should have provision
for air circulation
KMSCL – Tender No: 02/2010 – Amendment Notice Page 12 of 15
Section Section Section Section
No.No.No.No. Clause No.Clause No.Clause No.Clause No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix
I 1.19 Nil
Should be able to accommodate
60 numbers standard blood bags
for each of 350 ml capacity
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----16161616----02/2010 02/2010 02/2010 02/2010 :::: Blood Collection MonitorBlood Collection MonitorBlood Collection MonitorBlood Collection Monitor
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
II 2.2.20
The decision on the
mode of verification of
specification of the
equipments will be
decided after the pre-
tender meeting by the
corporation on the basis
of expert advice.
Demonstration shall be
arranged at KMSCL head
office.
V 5.1.1
Installation & Delivery
period : 6 weeks from the
date of issuance of letter
of intent
Installation & Delivery
period : 8 weeks from the
date of issuance of letter of
intent
Appendix I 7
Should have volume
setting ranges from 5ml
to 500ml in steps of 5ml,
Automatic storage and
recall of set volume.
Should have volume
setting ranges from 50ml to
500ml in steps of 5ml,
Automatic storage and
recall of set volume.
Appendix I 10
Should have the
following alarm
indications
a. LCD&LED indication
and audible alarm for
debit flow when flow
rate goes below 20
ml/mt or high flow
rate above 180
ml/mt.
b. LCD & LED
indications and
Should have the following
alarm indications
a. LCD/LED indication
and audible alarm for
debit flow when flow
rate goes below 20
ml/mt or high flow rate
above 180 ml/mt.
b. LCD / LED indications
and audible alarm at
the end of collection
c. LCD / LED indications
KMSCL – Tender No: 02/2010 – Amendment Notice Page 13 of 15
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
audible alarm at the
end of collection
c. LCD & LED
indications & audible
alarm during power
failure, LED blinking
when battery low.
d. LCD&LED
indications and
audible alarm during
power failure
& audible alarm during
power failure, LED
blinking when battery
low.
d. LCD/LED indications
and audible alarm
during power failure
Appendix I 11
Equipment should have
international/national
standard certificate for
safety and quality.
Clause deleted
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----153153153153----02/2010 : Cryobath02/2010 : Cryobath02/2010 : Cryobath02/2010 : Cryobath
Section Section Section Section
No.No.No.No.
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
II 2.2.20
The decision on the mode of
verification of specification
of the equipments will be
decided after the pre-tender
meeting by the corporation
on the basis of expert
advice.
Demonstration shall be
arranged at KMSCL head
office.
V 5.1.3 CMC/AMC period :
5 years
CMC/AMC period : 3
years
Appendix I 2 Programmable temp. range
: 30C to 560C
Programmable temp.
range 30C to 400C
Appendix I 5
Time taken for one process
:60-120 minutes
(for plasma bags stored at -
400C)
Time taken for one
process : 5 to 6 hours
(for plasma bags stored
at - 400C)
KMSCL – Tender No: 02/2010 – Amendment Notice Page 14 of 15
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----154154154154----02/2010 : Cryofuge02/2010 : Cryofuge02/2010 : Cryofuge02/2010 : Cryofuge
Section Section Section Section
No.No.No.No.
ClausClausClausClause e e e
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
II 2.2.20
The decision on the
mode of verification of
specification of the
equipments will be
decided after the pre-
tender meeting by the
corporation on the basis
of expert advice.
Demonstration shall be
arranged at third party
location and the Tender
Inviting Authority reserves
the right to finalise the
location.
V 5.1.3 CMC/AMC period :
7 years CMC/AMC period : 5 years
Appendix I
Max. Capacity : 6 x 1000
ml bottles or 12 blood
bags (550ml
each)
Max. Capacity : 12 blood
bags (550ml
each)
Appendix I Rotors and Buckets for
Cryofuge : Round bucket Clause deleted
TENDER NO. TENDER NO. TENDER NO. TENDER NO. EQPTEQPTEQPTEQPT----164164164164----02/2010 : Plasma thawing bath02/2010 : Plasma thawing bath02/2010 : Plasma thawing bath02/2010 : Plasma thawing bath
Section Section Section Section
NoNoNoNo....
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
II 2.2.20
The decision on the
mode of verification of
specification of the
equipments will be
decided after the pre-
tender meeting by the
corporation on the basis
of expert advice.
Demonstration shall be
arranged at third party
location and the Tender
Inviting Authority reserves
the right to finalise the
location.
Appendix I 3
Should have two
separate basket
assemblies with built-in
fingers for securely
holding the plasma bags
of all size.
Should have separate slot
for holding the plasma bags
of all size.
KMSCL – Tender No: 02/2010 – Amendment Notice Page 15 of 15
Section Section Section Section
NoNoNoNo....
Clause Clause Clause Clause
No.No.No.No. Existing clauseExisting clauseExisting clauseExisting clause Amended clauseAmended clauseAmended clauseAmended clause
Appendix I 4
Should give an alarm
when the plasma bags
are thawed
Clause deleted
Appendix I 7
Should have audio visual
over-temperature alarm
system
Should have audio / visual
over-temperature alarm
system
Appendix I 8
Should have a deep
thawing chamber with a
stirrer
Should have a deep
thawing chamber with a
stirrer / pump
Appendix I 9
Should have a system to
drain the chamber within
3 minutes
Should have a system to
drain the chamber 3 to 5
minutes
Sd/-
MANAGING DIRECTORMANAGING DIRECTORMANAGING DIRECTORMANAGING DIRECTOR