ams clinic4 turbo pump failure
DESCRIPTION
AMS Clinic4 Turbo Pump Failure. Yu Jun Leong (Griffin Lab, Rice University). Problem (HR- ToF -AMS). Turbo pumps failing frequently (5/year) Latest P5 failure was severe:. More on P5 (March 2012). Exchange pump Takes longer to start Vanes locked up after crash - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/1.jpg)
AMS Clinic4Turbo Pump Failure
Yu Jun Leong (Griffin Lab, Rice University)
![Page 2: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/2.jpg)
Problem (HR-ToF-AMS)
• Turbo pumps failing frequently (5/year)• Latest P5 failure was severe:
![Page 3: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/3.jpg)
![Page 4: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/4.jpg)
More on P5 (March 2012)
• Exchange pump• Takes longer to start• Vanes locked up after crash• Scratches - detection cylinder
![Page 5: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/5.jpg)
Questions
• Are we causing these frequent failures?– How best to diagnose?
• P5 – Agilent representative:– Beyond bearing failure– We/controller drove pump > rated rotation– Additional charge for damage
![Page 6: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/6.jpg)
Appendix: Repair Log
Location Date Problem Action
Rice 3/1/2011 P6 failed Replaced with Exchange unit from Agilent-Varian Inc.
Rice 9/22/2011 P3 failed Replaced with Exchange unit from Agilent-Varian Inc.
Rice 11/8/2011 P4 failed Replaced with Exchange unit from Agilent-Varian Inc.
Rice 12/4/2011 P4 failed Replaced with Exchange unit from Agilent-Varian Inc.
Rice 3/20/2012 P5 failed (severe) Replaced with Exchange unit from Agilent-Varian Inc.
![Page 7: AMS Clinic4 Turbo Pump Failure](https://reader036.vdocuments.us/reader036/viewer/2022082517/56812d29550346895d922552/html5/thumbnails/7.jpg)
Appendix: Stats
• Latest speed and currentPump Hz Amps
P2 963 449
P3 1350 332
P4 1350 302
P5 963 287
P6 1350 300