pcra patient controlled regional...
Post on 09-Jul-2020
1 Views
Preview:
TRANSCRIPT
E x p e r t s i n a m b u l a t o r y i n f u s i o n t h e r a p i e s
• A wireless system enables clinicians, nurses and homecareproviders to remotely monitor the infusion therapy throughthe web.
• SMS alert messages can be sent to personnel in charge.
PCRA
Patient Controlled Regional Analgesia
Ready to
Banana bag for Mini RythmicTM Evolution
Epidural, peripheral nerve blocks and wound infiltration analgesia commonly require the use of pre-filled large standard capacity bags. As patients are becoming more mobile in their recovery period it is important to integrate better these large medication bags into the system. In addition, the users require a safe and simple device, without frequent alarms, risk-free programmingand a trustworthy history back up. Clear identification of the infusion route from device to patient is a must in order to avoid potential medication infusion error.Birth delivery analgesia specifically tends to use the “Programmed Automatic Bolus” (PAB) combined with PCA demand facility to optimize pain release during labor.
Micrel is offering the yellow color coded RythmicTM Evolution range, to exceed the expected safety, reliability and optimize the comfort required by medical staff and patient in Regional Analgesia. In addition, RythmicTM Evolution is “Ready to Connect” providing an instant access to the infusion information and pain history through the web for hospital and home care applications
Optimized ambulatory design• Compact and lightweight • Integrate pre-filled bags (50 to 500ml)
Reduced alarm activity• Long battery life (6 days at 5ml/h)• Higher occlusion setting• Automatic restart after occlusion release• Air in line eliminator
Medication error prevention • Protocol Library• Colour coded systems• Non-Luer catheter Surety® connector
Innovative “all in one” safety set• Including spike, anti free flow valve, filter and
2 m extension line (yellow, transparent or green)
Innovations : VAS and PAB• VAS Pain score registration on the pump with
easy historical access• Programmed Automatic Bolus (PAB)
combined with patient bolus demands
FIche_PCRA_2013_Ang 2/06/14 8:23 Page 1
E x p e r t s i n a m b u l a t o r y i n f u s i o n t h e r a p i e s
RythmicTM Evolution and plug and playprinter for events down load
RythmicTM Organiser 501 for upto 500ml external bag
RythmicTM Evolution in bananatransport bag
Power supply and 12 V adaptorMini RythmicTM Evolution
Yellow Spike set SYellow Spike set S with Non-LuerSurety® Connector
Yellow Full set (160 ml and 200ml )Yellow Mini Full set (100ml)
Green Spike set SGreen Spike set S with Non-Luer Surety® Connector
www.micrelmed.com
KM1.EE.162.x KE1.EE.193.x (Non-Luer Surety® Connector)
KM1.EE.161.x (160ml)KE1.EE.190.x (200ml)KE1.EE.167.x (100ml)
KE1.EE.178.xKE1.EE.194.x (Non-Luer Surety® Connector)
Info
rmat
ion
in t
his
docu
men
t is
cor
rect
at
time
of p
rint
ing
ML
044
BRO
14/0
5/20
14 P
CRA
Engl
ish
Contact :info@micrelmed.com
300g,130x75x46 mmwith transparent locker fitting our MiniFull set 100ml custom bag
Included : banana bagOptional, ref : KS5.01.615.xOptional, ref : KS5.04.017.x
KP5.04.261.x (blue) KP5.04.263.x (yellow)
9V disposable cell or internal and rechargeable in 3 h
Standard or yellow colored dedicated to Regional Analgesia
335g,130x133x46 mm with transparent locker,fitting our Full set 160ml/200ml custom bag
Included : belt/shoulder bagOptional, ref : KS5.01.615.xOptional, ref: KS5.01.430.x
KP5.04.125.x (blue)KP5.04.130.x (yellow)
770 g,148 x 249 x 69 mm with transparentlocker fitting standard bags up to 100ml
Optional, ref : KM1.YY.451.xIncludedOptional, ref: KS5.01.525.x
KP5.04.161.x (blue)
840 g,157 x 299 x 82 mm with transparentlocker fitting standard bags up to 500ml
Optional, ref : KM1.YY.418.xIncludedOptional, ref: KS5.01.524.x
KP5.04.144.x (blue)KP5.04.150.x (yellow)
Mini Rythmic™ EvolutionMini Rythmic™ Evolution Rythmic™ EvolutionRythmic™ Evolution Rythmic™ Evolution Rythmic™ Evolution Rythmic™ EvolutionRythmic™ EvolutionOrganiser 101 Organiser 101 Organiser 501Organiser 501
Weight with batteryDimensions
BatteryCarrying bag
Power adaptorPole clampFront panel
Ordering references
Rythmic™ Evolution specificationsDelivery modes
AccuracyFlow rate
Programming units Concentration
Volume to infuseBolus dose
Bolus flow rateLockout intervalSpecial functions
Therapy monitoring
KVOAutomatic restart
after occlusion releaseSafety features
Alarms & warnings
Access code levelsProgramming
History, infusion graph / printout and PC download
Battery lifePower supply
Preventive maintenanceDesign and manufacture
• Continuous • Bolus only • Continuous + bolus • Programmed Automatic Bolus (PAB) with combined PCA+/- 5 %0.1 to 100 ml/hml/mg/µg0.1 to 99.9 mg/ml/0.1 to 999.9 µg/ml1 to 9999 ml0 or from 0.1 to 60ml with 0.1ml increment100 ml/h, adjustable from 50 to 100 ml/h with 10 ml/h increment0 to 999 minutesProtocol libraries (60 protocols per library) - Dose limit / x hours and Bolus limit / x hours - Loading dose - Clinician bolus over-ride - Mandatory priming (enabled, warning only, disabled) - Automatic priming through the pump - Lockable pump keyboardRecording of patient pain scores with visual analogue scale (VAS) - VAS available in the pump history, printable and computer downloadable - Ready to Connect to www.micrelcare.net for therapy monitoring through the web.0.1 to 2.0 ml/h
YesAir in line eliminator - Free flow protection according to IEC 60601-2-24 - Programming within protocol limitsDedicated therapy front face and sets: - yellow for regional analgesia - green for wound infiltration - blue for othersOcclusion (low 0.8 bar, middle 1.2 bar, high 1.6 bar), near end and end of infusion - Low and depleted batteryDose limit, bolus limit - Open mechanism, preventive maintenance warning3Protocol libraries programmed with “RythmicTM Evolution Therapy Manager” computer softwareDirect programming on the pump possible
Yes, including 6500 time stamped events event log - “Plug and play” print facility 6 days at 5 ml/h with bolus90 to 240 V, 50-60 Hz (with optional adaptor KS5.01.615.x)Every 3 years
European Union
CE0120
ISO9001:2008/ ISO13485:2012
FIche_PCRA_2013_Ang 2/06/14 8:24 Page 2