pcra patient controlled regional...

2
Experts in ambulatory infusion therapies • A wireless system enables clinicians, nurses and homecare providers to remotely monitor the infusion therapy through the web. • SMS alert messages can be sent to personnel in charge. PCRA Patient Controlled Regional Analgesia Ready to Banana bag for Mini Rythmic TM Evolution Epidural, peripheral nerve blocks and wound infiltration analgesia commonly require the use of pre-filled large standard capacity bags. As patients are becoming more mobile in their recovery period it is important to integrate better these large medication bags into the system. In addition, the users require a safe and simple device, without frequent alarms, risk-free programming and a trustworthy history back up. Clear identification of the infusion route from device to patient is a must in order to avoid potential medication infusion error. Birth delivery analgesia specifically tends to use the “Programmed Automatic Bolus” (PAB) combined with PCA demand facility to optimize pain release during labor. Micrel is offering the yellow color coded Rythmic TM Evolution range, to exceed the expected safety, reliability and optimize the comfort required by medical staff and patient in Regional Analgesia. In addition, Rythmic TM Evolution is “Ready to Connect” providing an instant access to the infusion information and pain history through the web for hospital and home care applications Optimized ambulatory design • Compact and lightweight • Integrate pre-filled bags (50 to 500ml) Reduced alarm activity • Long battery life (6 days at 5ml/h) • Higher occlusion setting • Automatic restart after occlusion release • Air in line eliminator Medication error prevention • Protocol Library • Colour coded systems • Non-Luer catheter Surety ® connector Innovative “all in one” safety set • Including spike, anti free flow valve, filter and 2 m extension line (yellow, transparent or green) Innovations : VAS and PAB • VAS Pain score registration on the pump with easy historical access • Programmed Automatic Bolus (PAB) combined with patient bolus demands

Upload: others

Post on 09-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PCRA Patient Controlled Regional Analgesiaoncomedical.ch/editors/tiny_mce/plugins/moxiemanager/...Experts in ambulatory infusion therapies RythmicTM Evolution and plug and play printer

E x p e r t s i n a m b u l a t o r y i n f u s i o n t h e r a p i e s

• A wireless system enables clinicians, nurses and homecareproviders to remotely monitor the infusion therapy throughthe web.

• SMS alert messages can be sent to personnel in charge.

PCRA

Patient Controlled Regional Analgesia

Ready to

Banana bag for Mini RythmicTM Evolution

Epidural, peripheral nerve blocks and wound infiltration analgesia commonly require the use of pre-filled large standard capacity bags. As patients are becoming more mobile in their recovery period it is important to integrate better these large medication bags into the system. In addition, the users require a safe and simple device, without frequent alarms, risk-free programmingand a trustworthy history back up. Clear identification of the infusion route from device to patient is a must in order to avoid potential medication infusion error.Birth delivery analgesia specifically tends to use the “Programmed Automatic Bolus” (PAB) combined with PCA demand facility to optimize pain release during labor.

Micrel is offering the yellow color coded RythmicTM Evolution range, to exceed the expected safety, reliability and optimize the comfort required by medical staff and patient in Regional Analgesia. In addition, RythmicTM Evolution is “Ready to Connect” providing an instant access to the infusion information and pain history through the web for hospital and home care applications

Optimized ambulatory design• Compact and lightweight • Integrate pre-filled bags (50 to 500ml)

Reduced alarm activity• Long battery life (6 days at 5ml/h)• Higher occlusion setting• Automatic restart after occlusion release• Air in line eliminator

Medication error prevention • Protocol Library• Colour coded systems• Non-Luer catheter Surety® connector

Innovative “all in one” safety set• Including spike, anti free flow valve, filter and

2 m extension line (yellow, transparent or green)

Innovations : VAS and PAB• VAS Pain score registration on the pump with

easy historical access• Programmed Automatic Bolus (PAB)

combined with patient bolus demands

FIche_PCRA_2013_Ang 2/06/14 8:23 Page 1

Page 2: PCRA Patient Controlled Regional Analgesiaoncomedical.ch/editors/tiny_mce/plugins/moxiemanager/...Experts in ambulatory infusion therapies RythmicTM Evolution and plug and play printer

E x p e r t s i n a m b u l a t o r y i n f u s i o n t h e r a p i e s

RythmicTM Evolution and plug and playprinter for events down load

RythmicTM Organiser 501 for upto 500ml external bag

RythmicTM Evolution in bananatransport bag

Power supply and 12 V adaptorMini RythmicTM Evolution

Yellow Spike set SYellow Spike set S with Non-LuerSurety® Connector

Yellow Full set (160 ml and 200ml )Yellow Mini Full set (100ml)

Green Spike set SGreen Spike set S with Non-Luer Surety® Connector

www.micrelmed.com

KM1.EE.162.x KE1.EE.193.x (Non-Luer Surety® Connector)

KM1.EE.161.x (160ml)KE1.EE.190.x (200ml)KE1.EE.167.x (100ml)

KE1.EE.178.xKE1.EE.194.x (Non-Luer Surety® Connector)

Info

rmat

ion

in t

his

docu

men

t is

cor

rect

at

time

of p

rint

ing

ML

044

BRO

14/0

5/20

14 P

CRA

Engl

ish

Contact :[email protected]

300g,130x75x46 mmwith transparent locker fitting our MiniFull set 100ml custom bag

Included : banana bagOptional, ref : KS5.01.615.xOptional, ref : KS5.04.017.x

KP5.04.261.x (blue) KP5.04.263.x (yellow)

9V disposable cell or internal and rechargeable in 3 h

Standard or yellow colored dedicated to Regional Analgesia

335g,130x133x46 mm with transparent locker,fitting our Full set 160ml/200ml custom bag

Included : belt/shoulder bagOptional, ref : KS5.01.615.xOptional, ref: KS5.01.430.x

KP5.04.125.x (blue)KP5.04.130.x (yellow)

770 g,148 x 249 x 69 mm with transparentlocker fitting standard bags up to 100ml

Optional, ref : KM1.YY.451.xIncludedOptional, ref: KS5.01.525.x

KP5.04.161.x (blue)

840 g,157 x 299 x 82 mm with transparentlocker fitting standard bags up to 500ml

Optional, ref : KM1.YY.418.xIncludedOptional, ref: KS5.01.524.x

KP5.04.144.x (blue)KP5.04.150.x (yellow)

Mini Rythmic™ EvolutionMini Rythmic™ Evolution Rythmic™ EvolutionRythmic™ Evolution Rythmic™ Evolution Rythmic™ Evolution Rythmic™ EvolutionRythmic™ EvolutionOrganiser 101 Organiser 101 Organiser 501Organiser 501

Weight with batteryDimensions

BatteryCarrying bag

Power adaptorPole clampFront panel

Ordering references

Rythmic™ Evolution specificationsDelivery modes

AccuracyFlow rate

Programming units Concentration

Volume to infuseBolus dose

Bolus flow rateLockout intervalSpecial functions

Therapy monitoring

KVOAutomatic restart

after occlusion releaseSafety features

Alarms & warnings

Access code levelsProgramming

History, infusion graph / printout and PC download

Battery lifePower supply

Preventive maintenanceDesign and manufacture

• Continuous • Bolus only • Continuous + bolus • Programmed Automatic Bolus (PAB) with combined PCA+/- 5 %0.1 to 100 ml/hml/mg/µg0.1 to 99.9 mg/ml/0.1 to 999.9 µg/ml1 to 9999 ml0 or from 0.1 to 60ml with 0.1ml increment100 ml/h, adjustable from 50 to 100 ml/h with 10 ml/h increment0 to 999 minutesProtocol libraries (60 protocols per library) - Dose limit / x hours and Bolus limit / x hours - Loading dose - Clinician bolus over-ride - Mandatory priming (enabled, warning only, disabled) - Automatic priming through the pump - Lockable pump keyboardRecording of patient pain scores with visual analogue scale (VAS) - VAS available in the pump history, printable and computer downloadable - Ready to Connect to www.micrelcare.net for therapy monitoring through the web.0.1 to 2.0 ml/h

YesAir in line eliminator - Free flow protection according to IEC 60601-2-24 - Programming within protocol limitsDedicated therapy front face and sets: - yellow for regional analgesia - green for wound infiltration - blue for othersOcclusion (low 0.8 bar, middle 1.2 bar, high 1.6 bar), near end and end of infusion - Low and depleted batteryDose limit, bolus limit - Open mechanism, preventive maintenance warning3Protocol libraries programmed with “RythmicTM Evolution Therapy Manager” computer softwareDirect programming on the pump possible

Yes, including 6500 time stamped events event log - “Plug and play” print facility 6 days at 5 ml/h with bolus90 to 240 V, 50-60 Hz (with optional adaptor KS5.01.615.x)Every 3 years

European Union

CE0120

ISO9001:2008/ ISO13485:2012

FIche_PCRA_2013_Ang 2/06/14 8:24 Page 2