In The Name Of God
Mohammad Javad Hosseinishahi
Solid-Phase Microextraction in Targeted and Nontargeted Analysis:
Displacement and Desorption EffectsSanja Risticevic and Janusz Pawliszyn
Supervisor: Dr.M.Saraji
Introduction Experimental section
Results and discussionConclusionsReference
This nonexhaustive and environmentally friendly technique integrates sampling, extraction, concentration, and sample introduction into a simple solvent-free procedureSPME involves the use of a fibre coated with an extracting phase, that can be a liquid (Polymer) or a solid (Sorbent)
Solid-phase microextraction (SPME) was introduced in 1990
After extraction, the SPME fibre is transferred to the injection port of separating instruments, such as a GC
Introduction
volume of the extraction phase
the extracted amount at equilibrium
fiber coating/sample matrix distribution constant of the analyte
Initial concentration of analyte in the sample
Introduction
volume of the extraction phase
fiber constant
fiber coating/sample matrix distribution constant of the analyte
Introduction
A) Chemicals and Materials
B) Sample Preparation
C) SPME Procedure
D) Instrument
Experimental section
The automated SPME holder and10 and 20 mL screw cap vials were purchased from Supelco
HPLC-grade methanol and acetone
Analyte standards were all of >97% purity (except for >95% purity for heptanal, nonanal, citral
isomers, farnesol isomers, dodecanal, tridecanal, and linalool)
Commercial SPME fiber assemblies in 23-gauge needle sizes and automated formats (100 μm PDMS), 85
μm (PA), 60 μm (CW) [metal], 65 μm PDMS/ (DVB), 85 μm (CAR)/PDMS, 50/30 μm DVB/CAR/PDMS,
and 16 μm of Carbopack Z/PDMS [metal]) were obtained from Supelco
Chemicals and Materials
250 ml NaCl
250 ml Water
100 g of apple tissue
Sample Preparation3ml
SPME Procedure
The aqueous extraction standards and apple samples were analyzed fresh by HS-SPME at 30 °C using a 500 rpm agitation speed and 60 min extraction time
t test was used to determine whether extraction equilibrium was reached for a particular component.
To compare DVB/CAR/PDMS and PDMS/DVB coatings interms of extraction kinetics and desorption efficiency, extraction time profiles were performed with 5, 30, 60, and 120 min points .
Interanalyte displacements were examined by analyzing aqueous calibration standards with 30 and 60 min extraction times and apple samples with 1, 5, 15, 30, 60, 120, and 180 min extraction times. All samples were analyzed at 30 °C
Instrumental
GC × GC/ToFMS
Primary dimension columns employed
Secondary dimension columns employed
Details:
)30 m × 0.25 mm i.d. × 0.25 μm(
(1.15 m × 0.10 mm i.d. × 0.10 μm)
A) Trends in Coating Selectivity and SensitivityB) Desorption Efficiency of Commercial CoatingsC) Determination of the Linear Dynamic Range and Interanalyte Displacements
Results and Discussions
PDMS, PDMS/DVB, and DVB/CAR/PDMS coatings extracted all of the investigated metabolites
PDMS
Trends in Coating Selectivity and Sensitivity
PDMS/DVBDVB/CAR/PDMS
Trends in Coating Selectivity and Sensitivity
while only 47, 50, 44 and 39 peaks were extracted with PA, CAR/PDMS, CW, andCarbopack Z/PDMS coatings, respectively
PACAR/PDMSCWCarbopack Z/PDMS
Trends in Coating Selectivity and Sensitivity
PA
Some nonpolar components such as: a) octane b) nonane
Moderately polar such as:a) 2-hexanoneb) Hexanalc) ethyl butanoate
Not detected
CW Nonpolar analytes octane to undecane
Moderately polar Analytes such as:a) 2-hexanone b) Hexanalc) ethylbutanoated) α-pinenee) eucalyptol
Not detected
Carbopack Z/PDMS
This extraction phase was found to be unsuitable for the extraction of small molecular weight analytes
Pore size is 100 Å
Trends in Coating Selectivity and Sensitivity
Fiber constants are reported for PDMS, PA, DVB/CAR/PDMS, and PDMS/DVB coatings, since the implementation of these coatings in spiked water analysis resulted insatisfactory reproducibility.
In the worst-case scenario, for the PA coating, RSD ranged from 0.9% for linalool to 25.4% for ethyl stearate
This result is attributed to the difference in the average sizes of the micropore diameter between Carboxen 1006 in DVB/CAR/PDMS and CAR/PDMS coatings and DVB in PDMS/DVB and DVB/CAR/PDMS coatings (12 and 16 Å, respectively).
Trends in Coating Selectivity and Sensitivity
The trends in extraction efficiencies obtained with PDMS/DVB and CAR/PDMS coatings are also illustrated in Figure
Contour plots of extracted ion chromatograms (the x- and y-axes represent the retention times in the first and second dimensions) corresponding to 60 min HS-SPME extraction of an aqueous sample spiked with 52 metabolites and obtained with (A) PDMS/DVB and (B) CAR/PDMS coatings.
S1 TablePage 2 - 3
Trends in Coating Selectivity and Sensitivity
Likewise, since the ability of an adsorbent to extract a particular analyte is strongly dependent on the average size of the micropore diameter, KfsVf improvements of 11-fold for 2-pentanol and 4-fold for α-pinene, 1-pentanol, e2-hexanone, and 2-hexanol were detected when DVB /CAR /PDMS was compared to PDMS/DVB
A – < 100 g/molB – between 100 and 120 g/mol
C – between 180 and 215 g/mol
The largest improvement in extraction sensitivity was achieved with DVB /CAR /PDMS for analytes having molecular weights of <185 g/mol.
Above this molecular weight threshold, the performance of the two coatings was similar across the hydrophobicity and volatility range
PDMS/DVB performed slightly better only for the latest eluting members of each homologous series.
Desorption Efficiency of Commercial Coatings
Memory Effect For DVB /CAR /PDMS
& CAR/PDSM
Desorption Efficiency of Commercial Coatings
5, 30, 60 and 120 min extraction A – 1-nonanol, B – 1-undecanol, C – 1-pentadecanol and D-1-heptadecanol
Determination of linear dynamic range (LDR, 9-point calibration curve, each point run in triplicates) and method repeatability for actual spiked metabolite concentrations employed in coating evaluation study for experimental design involving DVB/CAR/PDMS coating and 60 min HS-SPME extraction
S3 TablePage 15-16
Determination of the Linear Dynamic Range and Interanalyte Displacements
Figure 2. SPME calibration curves for 2-pentanol (A; 60 and 30 min extraction times employed) and 2-heptadecanone (B) for aqueous samples spiked with 52 metabolites and analyzed with DVB /CAR /PDMS fiber coating at 30 °C
Determination of the Linear Dynamic Range and Interanalyte Displacements
Determination of the Linear Dynamic Range and Interanalyte Displacements
Extraction time (1−180 min) uptakes of major components in the apple matrix exhibiting high HS-SPME sensitivity and selectivity: (A) ethyl butanoate, (B) ethyl 2-methylbutanoate, (C) hexyl hexanoate.
Determination of the Linear Dynamic Range and Interanalyte Displacements
DVB/CAR/PDMS extraction time profiles of nonane (plot A), nonanal (plot B) and 1-nonanol (plot C)
Determination of the Linear Dynamic Range and Interanalyte Displacements
HS-SPME extraction time profiles of representative low S/N and low KfsVf polar compounds in the apple matrix: (A) (2Z)-2-penten-1-ol and (B) (3Z)-3-hexen-1-ol
Determination of the Linear Dynamic Range and Interanalyte Displacements
HS-SPME extraction time profiles of selected compounds in apple homogenate for which the occurrence of interanalyte displacement was detected: (A) 1-methoxybutane, (B) 2-methylpropanol
Determination of the Linear Dynamic Range and Interanalyte Displacements
HS-SPME extraction time profiles of selected compounds in apple homogenate for which the occurrence of interanalyte displacement wasdetected: (A) 1-methoxybutane, (B) 2-methylpropanol
Conclusions
The wide volatility, hydrophobicity, polarity, and molecular weight ranges of compounds considered in this systematic study allowed for a comprehensive evaluation of performance characteristics of commercially available SPME−GC fiber coatings in terms of extraction sensitivity, desorption efficiency, and feasibility in complex sample analysis
The investigation of complex mixtures with DVB/CAR/PDMS revealed that interanalyte displacements are infrequent.
the results obtained in this study clearly demonstrate that nonpolar high Henry’s constant compounds that have high Kfs are not displaced
The only analytes that were affected by interanalyte competition for adsorption sites included small molecular weight and early-eluting compounds with medium to high polarity and low fiber constants
(1) Arthur, C. L.; Pawliszyn, J. Anal. Chem. 1990, 62, 2145−2148.(2) Pawliszyn, J. In Handbook of Solid Phase Microextraction;Pawliszyn, J., Ed.; Chemical Industry Press: Beijing, China, 2009; pp. 54-1(3) Risticevic, S.; Niri, V. H.; Vuckovic, D.; Pawliszyn, J. Anal. Bioanal. Chem. 2009, 393, 781−795.(4) Risticevic, S.; Chen, Y.; Kudlejova, L.; Vatinno, R.; Baltensperger, B.; Stuff, J. R.; Hein, D.; Pawliszyn, J. Nat. Protoc. 2010, 5, 162−176.(5) Pawliszyn, J. In Solid Phase Microextraction: Theory and Practice; Pawliszyn, J., Ed.; Wiley-VCH: New York, 1997; pp 11−96.(6) Baltussen, E.; Cramers, C. A.; Sandra, P. J. F. Anal. Bioanal. Chem. 2-3 , 373 , 2002
)7 (Nongonierma, A.; Cayot, P.; Quere, J. L.; Springett, M.; Voilley,A. Food Rev. Int. 2006, 22, 51−94.
)8 (Musteata, F. M.; Pawliszyn, J. J. Proteome Res. 2005, 4, 789−800.)9 (Gunther, C. S.; Matich, A. J.; Marsh, K. B.; Nicolau, L. Food Res. Int. 2011, 44, 1331−1338.
)10 (Robinson, A. L.; Boss, P. K.; Heymann, H.; Solomon, P. S.; Trengove, R. D. J. Chromatogr., A 2011, 1218, 504−517.
)11 (Harrigan, G. G.; Martino-Catt, S.; Glenn, K. C. Metabolomics. 272-2759 , 3 , 2007 )12 (Fiehn, O.; Wohlgemuth, G.; Scholz, M.; Kind, T.; Lee, D. Y;.
Lu, Y.; Moon, S.; Nikolau, B. Plant J. 2008, 53, 691−704.)13 (Lοpez, R.; Lapena, A. C.; Cacho, J.; Ferreira, V. J. Chromatogr., A. 15-8 , 1143 ,2007
References
)14 (Murray, R. A. Anal. Chem. 2001, 73, 1646−1649.)15 (Mestres, M.; Sala, C.; Marti, M. P.; Busto, O.; Guasch, J. J.Chromatogr., A 1999,
835, 137−144.(16) Gorecki, T.; Yu, X.; Pawliszyn, J. Analyst 1999, 124, 643−649.(17) Gorecki, T.; Martos, P.; Pawliszyn, J. Anal. Chem. 1998, 70, 19−27.(18) Zhang, Z.; Pawliszyn, J. Anal. Chem. 1993, 65, 1843−1852.(19) Psillakis, E.; Yiantzi, E.; Sanchez-Prado, L.; Kalogerakis, N. Anal. Chim. Acta. 2012, 742, 30−36.(20) Sukola, K.; Koziel, J.; Augusto, F.; Pawliszyn, J. Anal. Chem.2001,73,13-18.(21) Koziel, J.; Jia, M.; Pawliszyn, J. Anal. Chem. 2000, 72, 5178−5186.(22) Frank, D. C.; Owen, C. M.; Patterson, J. Lebensm.-Wiss. Technol.2004,37,139-154. (23) Contini, M.; Esti, M. Food Chem. 2006, 94, 143−150.(24) Cavalli, J.-F.; Fernandez, X.; Lizzani-Cuvelier, L.; Loiseau, A.-M.J. Agric. Food Chem. 2003, 51, 7709−7716.(25) Mestres, M.; Busto, O.; Guasch, J. J. Chromatogr., A 1998, 808,211-218(26) Harynuk, J.; Gοrecki, T. J. Chromatogr., A 2003, 1019, 53−63.
Thanks a lot for attention