STLCC-CPLSMorrison 6122013 Page 1
Centrifuge ndash Floor Bench Mini
VapConcentrator (Detect Instrument Picture to Jump to that section of SOP)
Prepared by Bob Morrison
Instrumentation Specialist
Original June 08 Last revision Nov 2012
Bench Mini
Model 5415D
Bench Micro
Model 5415C
Floor Beckman Avanti
Model J20 XP BRDG
Floor Beckman
Avanti Model J-E FV
Bench Micro 14
CentriVapConcentrator
Labconco BRDG BenchRefrig Z400
BRDG
TOMY mini
BRDG
Beckman micro
lt10K rpms
Centrifugation
Nomogram
RPMs to ldquogrdquo
Convert ldquogrdquo and RPMs
Link to SERVICEREPAIR INFO
Prism micro Spectra-mini
Centrifuge Local Service Center
STLCC-CPLSMorrison 6122013 Page 2
STLCC-CPLSMorrison 6122013 Page 3
Centrifuge Floor Model Beckman Avanti J-20XP BRDG
J series
Electrical 208V
Connector Type
System ID tag 432877 SN JXPO1K16 CAT 368608 Maximum Speed 26 000 rpm
Maximum g-force 82 000 times g
Speed Control plusmn 10 rpm of set speed or 01 whichever is greater
Drive Type Switched Reluctance Drive
Drive Cooling Air-cooled
Centrifuge Warranty 1 year
Drive Warranty 3 years
Refrigeration System Non-CFC non-ozone-depleting refrigerant
Max Capacity Volume 6 000 mL
Single Tube Sample Volume Range 15 ndash 1 000 mL
Ambient Temperature Range 15deg to 40degC
Set Temperature Range -10deg to 40degC
Temperature Control plusmn 2degC of set
User-Settable Programs J-20XPI 30 two-step programs
J-20XP No programming Zonal Capability J-20XPI Yes
J-20XP No Time Setting 1 min to 9959 hours hold or ω2t
(J-20XPI) 1 min to 3 hours hold (J-20XP)
AccelerationDeceleration Rates 12 accel13 decel (J-20XPI)
2 accel3 decel (J-20XP)
Dynamic Rotor Inertia Check Automatic
Heat Output 20 kW (6 900 BTUhr)
Sound Level lt 57 dBa (3 ft from instrument at maximum speed)
Dimensions 865 mm times 876 mm times 711 mm
(341rdquo H times 345rdquo D times 28rdquo W) Weight 264 kg (583 lb)
Power Nameplate 24amp Fuse 20amp 18 per UCLA test runs
Male plug 6-30P
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Floor Avanti J-20XP Control Panel BRDG
STLCC-CPLSMorrison 6122013 Page 4
3 Set dial to match
installed rotor
5 Set Speed
6 Set Time (hrmin)
7 Set Temp
10 Press START will
blink until set values
are reached
11 Press firmly on
doorlid at 4pm
position until rpm
setting is achieved
1 Turn ON MasterSwitch
2 Depress foot pedal to
release door and install
rotor
4 Press SET press
again to see
Actual
readings
Max
Slow
Max
Slow
Off
8 Press ACCEL to
set rate
9 Press DECEL to
set rate
Actual
Set
STOP will blink
until rotor
stops
Door
FRS
Centrifuge Avanti Warning Message DoorLid and FRS
STLCC-CPLSMorrison 6122013 Page 5
Holding down door lid after rpms reach about
4000 generally resolves FRS message
(improves seal of lid and internal vacuum
Door or Lid or FRS (Friction Reduction System) Errors
1) Hold the liddoor down firmly at the 4pm position after pressing START
2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches
3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber
to reduce friction on rotor for higher RPMs
4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal
Centrifuge Avanti DoorLid Errors Next Steps
STLCC-CPLSMorrison 6122013 Page 6
This solenoid may be stuck in
the ldquoopenrdquo position and thus
giving constant ldquodoorlid openrdquo
status to control panel
1 Try cleaning solenoid piston
shaft with q-tip
2 Slide black lock latch tofro
3 Remove and clean piston
shaft with clean cloth DO
NOT LUBRICATE piston as
this can cause overheating
in the solenoid
4 Call ACE Systems for
repairs if this does not work
solenoid may need
replacement or other
problems exist
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Local Service Center
STLCC-CPLSMorrison 6122013 Page 2
STLCC-CPLSMorrison 6122013 Page 3
Centrifuge Floor Model Beckman Avanti J-20XP BRDG
J series
Electrical 208V
Connector Type
System ID tag 432877 SN JXPO1K16 CAT 368608 Maximum Speed 26 000 rpm
Maximum g-force 82 000 times g
Speed Control plusmn 10 rpm of set speed or 01 whichever is greater
Drive Type Switched Reluctance Drive
Drive Cooling Air-cooled
Centrifuge Warranty 1 year
Drive Warranty 3 years
Refrigeration System Non-CFC non-ozone-depleting refrigerant
Max Capacity Volume 6 000 mL
Single Tube Sample Volume Range 15 ndash 1 000 mL
Ambient Temperature Range 15deg to 40degC
Set Temperature Range -10deg to 40degC
Temperature Control plusmn 2degC of set
User-Settable Programs J-20XPI 30 two-step programs
J-20XP No programming Zonal Capability J-20XPI Yes
J-20XP No Time Setting 1 min to 9959 hours hold or ω2t
(J-20XPI) 1 min to 3 hours hold (J-20XP)
AccelerationDeceleration Rates 12 accel13 decel (J-20XPI)
2 accel3 decel (J-20XP)
Dynamic Rotor Inertia Check Automatic
Heat Output 20 kW (6 900 BTUhr)
Sound Level lt 57 dBa (3 ft from instrument at maximum speed)
Dimensions 865 mm times 876 mm times 711 mm
(341rdquo H times 345rdquo D times 28rdquo W) Weight 264 kg (583 lb)
Power Nameplate 24amp Fuse 20amp 18 per UCLA test runs
Male plug 6-30P
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Floor Avanti J-20XP Control Panel BRDG
STLCC-CPLSMorrison 6122013 Page 4
3 Set dial to match
installed rotor
5 Set Speed
6 Set Time (hrmin)
7 Set Temp
10 Press START will
blink until set values
are reached
11 Press firmly on
doorlid at 4pm
position until rpm
setting is achieved
1 Turn ON MasterSwitch
2 Depress foot pedal to
release door and install
rotor
4 Press SET press
again to see
Actual
readings
Max
Slow
Max
Slow
Off
8 Press ACCEL to
set rate
9 Press DECEL to
set rate
Actual
Set
STOP will blink
until rotor
stops
Door
FRS
Centrifuge Avanti Warning Message DoorLid and FRS
STLCC-CPLSMorrison 6122013 Page 5
Holding down door lid after rpms reach about
4000 generally resolves FRS message
(improves seal of lid and internal vacuum
Door or Lid or FRS (Friction Reduction System) Errors
1) Hold the liddoor down firmly at the 4pm position after pressing START
2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches
3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber
to reduce friction on rotor for higher RPMs
4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal
Centrifuge Avanti DoorLid Errors Next Steps
STLCC-CPLSMorrison 6122013 Page 6
This solenoid may be stuck in
the ldquoopenrdquo position and thus
giving constant ldquodoorlid openrdquo
status to control panel
1 Try cleaning solenoid piston
shaft with q-tip
2 Slide black lock latch tofro
3 Remove and clean piston
shaft with clean cloth DO
NOT LUBRICATE piston as
this can cause overheating
in the solenoid
4 Call ACE Systems for
repairs if this does not work
solenoid may need
replacement or other
problems exist
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 3
Centrifuge Floor Model Beckman Avanti J-20XP BRDG
J series
Electrical 208V
Connector Type
System ID tag 432877 SN JXPO1K16 CAT 368608 Maximum Speed 26 000 rpm
Maximum g-force 82 000 times g
Speed Control plusmn 10 rpm of set speed or 01 whichever is greater
Drive Type Switched Reluctance Drive
Drive Cooling Air-cooled
Centrifuge Warranty 1 year
Drive Warranty 3 years
Refrigeration System Non-CFC non-ozone-depleting refrigerant
Max Capacity Volume 6 000 mL
Single Tube Sample Volume Range 15 ndash 1 000 mL
Ambient Temperature Range 15deg to 40degC
Set Temperature Range -10deg to 40degC
Temperature Control plusmn 2degC of set
User-Settable Programs J-20XPI 30 two-step programs
J-20XP No programming Zonal Capability J-20XPI Yes
J-20XP No Time Setting 1 min to 9959 hours hold or ω2t
(J-20XPI) 1 min to 3 hours hold (J-20XP)
AccelerationDeceleration Rates 12 accel13 decel (J-20XPI)
2 accel3 decel (J-20XP)
Dynamic Rotor Inertia Check Automatic
Heat Output 20 kW (6 900 BTUhr)
Sound Level lt 57 dBa (3 ft from instrument at maximum speed)
Dimensions 865 mm times 876 mm times 711 mm
(341rdquo H times 345rdquo D times 28rdquo W) Weight 264 kg (583 lb)
Power Nameplate 24amp Fuse 20amp 18 per UCLA test runs
Male plug 6-30P
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Floor Avanti J-20XP Control Panel BRDG
STLCC-CPLSMorrison 6122013 Page 4
3 Set dial to match
installed rotor
5 Set Speed
6 Set Time (hrmin)
7 Set Temp
10 Press START will
blink until set values
are reached
11 Press firmly on
doorlid at 4pm
position until rpm
setting is achieved
1 Turn ON MasterSwitch
2 Depress foot pedal to
release door and install
rotor
4 Press SET press
again to see
Actual
readings
Max
Slow
Max
Slow
Off
8 Press ACCEL to
set rate
9 Press DECEL to
set rate
Actual
Set
STOP will blink
until rotor
stops
Door
FRS
Centrifuge Avanti Warning Message DoorLid and FRS
STLCC-CPLSMorrison 6122013 Page 5
Holding down door lid after rpms reach about
4000 generally resolves FRS message
(improves seal of lid and internal vacuum
Door or Lid or FRS (Friction Reduction System) Errors
1) Hold the liddoor down firmly at the 4pm position after pressing START
2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches
3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber
to reduce friction on rotor for higher RPMs
4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal
Centrifuge Avanti DoorLid Errors Next Steps
STLCC-CPLSMorrison 6122013 Page 6
This solenoid may be stuck in
the ldquoopenrdquo position and thus
giving constant ldquodoorlid openrdquo
status to control panel
1 Try cleaning solenoid piston
shaft with q-tip
2 Slide black lock latch tofro
3 Remove and clean piston
shaft with clean cloth DO
NOT LUBRICATE piston as
this can cause overheating
in the solenoid
4 Call ACE Systems for
repairs if this does not work
solenoid may need
replacement or other
problems exist
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP Control Panel BRDG
STLCC-CPLSMorrison 6122013 Page 4
3 Set dial to match
installed rotor
5 Set Speed
6 Set Time (hrmin)
7 Set Temp
10 Press START will
blink until set values
are reached
11 Press firmly on
doorlid at 4pm
position until rpm
setting is achieved
1 Turn ON MasterSwitch
2 Depress foot pedal to
release door and install
rotor
4 Press SET press
again to see
Actual
readings
Max
Slow
Max
Slow
Off
8 Press ACCEL to
set rate
9 Press DECEL to
set rate
Actual
Set
STOP will blink
until rotor
stops
Door
FRS
Centrifuge Avanti Warning Message DoorLid and FRS
STLCC-CPLSMorrison 6122013 Page 5
Holding down door lid after rpms reach about
4000 generally resolves FRS message
(improves seal of lid and internal vacuum
Door or Lid or FRS (Friction Reduction System) Errors
1) Hold the liddoor down firmly at the 4pm position after pressing START
2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches
3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber
to reduce friction on rotor for higher RPMs
4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal
Centrifuge Avanti DoorLid Errors Next Steps
STLCC-CPLSMorrison 6122013 Page 6
This solenoid may be stuck in
the ldquoopenrdquo position and thus
giving constant ldquodoorlid openrdquo
status to control panel
1 Try cleaning solenoid piston
shaft with q-tip
2 Slide black lock latch tofro
3 Remove and clean piston
shaft with clean cloth DO
NOT LUBRICATE piston as
this can cause overheating
in the solenoid
4 Call ACE Systems for
repairs if this does not work
solenoid may need
replacement or other
problems exist
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Avanti Warning Message DoorLid and FRS
STLCC-CPLSMorrison 6122013 Page 5
Holding down door lid after rpms reach about
4000 generally resolves FRS message
(improves seal of lid and internal vacuum
Door or Lid or FRS (Friction Reduction System) Errors
1) Hold the liddoor down firmly at the 4pm position after pressing START
2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches
3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber
to reduce friction on rotor for higher RPMs
4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal
Centrifuge Avanti DoorLid Errors Next Steps
STLCC-CPLSMorrison 6122013 Page 6
This solenoid may be stuck in
the ldquoopenrdquo position and thus
giving constant ldquodoorlid openrdquo
status to control panel
1 Try cleaning solenoid piston
shaft with q-tip
2 Slide black lock latch tofro
3 Remove and clean piston
shaft with clean cloth DO
NOT LUBRICATE piston as
this can cause overheating
in the solenoid
4 Call ACE Systems for
repairs if this does not work
solenoid may need
replacement or other
problems exist
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Avanti DoorLid Errors Next Steps
STLCC-CPLSMorrison 6122013 Page 6
This solenoid may be stuck in
the ldquoopenrdquo position and thus
giving constant ldquodoorlid openrdquo
status to control panel
1 Try cleaning solenoid piston
shaft with q-tip
2 Slide black lock latch tofro
3 Remove and clean piston
shaft with clean cloth DO
NOT LUBRICATE piston as
this can cause overheating
in the solenoid
4 Call ACE Systems for
repairs if this does not work
solenoid may need
replacement or other
problems exist
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti ACE second fix Door Sensor
STLCC-CPLSMorrison 6122013 Page 7
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Swinging-Bucket Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
13 000 26 500 1841 6 x 50 mL 300 mL
29 x 105 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG
STLCC-CPLSMorrison 6122013 Page 8
JS131 SN 01U-3617
Link to Beckman Avanti Rotor JS-131 - pdf
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772
STLCC-CPLSMorrison 6122013 Page 9
JLA-16250 SN 01U-1772
Link to Beckman Avanti Rotor JLA-16250 - pdf
Fixed-Angle Rotor Aluminum
Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)
Max RPM Max g k Factor Number of Tubes Rotor Capacity
16 000 38 420 1 060 6 250 mL 15 L
62 120 mm
25 times 55 in
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540
STLCC-CPLSMorrison 6122013 Page 10
JA 2550 SN 01U-2540
Link to Beckman Avanti Rotor JA-2550 - pdf
Fixed-Angle Rotor Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
25 000 75 600 418 8 x 50 mL 400 mL
29 x 104 mm
1125 x 4 in
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617
STLCC-CPLSMorrison 6122013 Page 11
Swinging-Bucket Rotor Anodized Aluminum
Max RPM Max g k Factor Number of Tubes Rotor Capacity
4300 4220 11 800 4 x 750 mL 3 L
96 x 130 mm
38 x 52 in
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG
STLCC-CPLSMorrison 6122013 Page 12
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG
STLCC-CPLSMorrison 6122013 Page 13
Microplus Plate Carrier
No 362394 MicroPlus Carrier Assembly Includes carrier base and
rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of
MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)
Link to Beckman Avanti Rotor JS-43 - pdf
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Avanti J Series Emergency Access to Rotor
(note This info from J25 Model and may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 14
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Avanti J-XP
Emergency Access to
Rotor Steps 5-8 (note This info from J25 Model and
may not be same for J-XP20)
STLCC-CPLSMorrison 6122013 Page 15
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Avanti
Warnings Rotor
Speed Door Sys
STLCC-CPLSMorrison 6122013 Page 16
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor BRDG Avanti Maintenance
STLCC-CPLSMorrison 6122013 Page 17
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifugation Nomogram RPMs to ldquogrdquo
STLCC-CPLSMorrison 6122013 Page 18
Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a
rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2
g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)
Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges
and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor and Bench Adapters
STLCC-CPLSMorrison 6122013 Page 19
392075 Adapters 15ml
Conical 17mm diameter
Marathon Microtube Adapters
15ml to 5ml
Marathon Fisher 04-974-
122B
Corning 8445
Marathon Fisher 004-974-
011A
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Floor Avanti J-20XP Lubricants BRDG
STLCC-CPLSMorrison 6122013 Page 20
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 21
Centrifuge Floor Models Beckman Avanti J-E FV
Specifications Depth 800 cm (315 in)
Height 914 cm (36 in) with door open 1524 cm (60 in)
Width 635 cm (25 in)
Weight 2674 kg (589 lb)
Approximate Acceleration TimeTwo profiles--MAX (full acceleration
to set speed) and SLOW
Approximate Deceleration TimeThree profiles--MAX SLOW and
OFF (coast)
Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz
187to229216to264 VAC 24 A
Friction Reduction System (FRS) 190 mm (75 in) Hg
Maximum Heat Dissipation into Room under Steady-State
Conditions 20 kW (6900 Btuhr)
Noise Level 1 Meter in Front of Centrifuge lt 64 dBA
Set Speed Actual rotor speed plusmn 50 rpm of set speed
Set Temperature-10deg C to +40deg C in 1deg C increments
Set Time to 99 hr 59 min or continuous (hold)
Speed Range 300 to 21000 rpm
Temperature Control plusmn 2deg C of set temperature (after
equilibration)
Ambient Temperature Range 15deg C to 35deg C
Clearances Required 76 cm (30 in) sides 160 cm (625 in)
rear
Humidity Restrictions lt 95 (noncondensing)
Pollution Degree 2
Time -- Actual Display Indicates run time remainingdisplays
elapsed time
Finish Control panel silicone elastomer Top surface urethane
paint Other surfaces general purpose paint
Link to Beckman Avanti J-E Brochure ndash pdf
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Bench Refrigerated Z400K MIDSCILabnet
STLCC-CPLSMorrison 6122013 Page 22
Z400K Refrigerated Universal Centrifuge
Hermie Labortechnik GmbH (Germany)
74cm wide 34cm deep 58cm high
Weight 79kg approx 182lbs
Max speed 13500 rpm
Max RCF 17310g
8amp 120V 960watts 60 db noise level
Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf
Main ONOFF
Toggle
Plastic cap
emergency or
power out lid open
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 23
Centrifuge Bench Models General Operating Instructions
1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open
2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position
3 Choose the proper rotor for the operation and tube sizes involved
4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary
5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure
6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage
7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage
8 Close rotor door and set timers if desired
9 Push the start button to engage the timers Hold down the start button to
manually continue centrifuge operations
Link to Beckman Avanti J-25 Manual
(Not J-20XP but has similar features)
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Z400 Controls Note Follow order below
STLCC-CPLSMorrison 6122013 Page 24
1 Press this button to release lid Dial
settings prior to a run cannot be made
if the lid is locked
Green LED is off when rotor is stopped
and lid can be opened
3 Rotate to Preset the RPMs
or RCF
2 Press to toggle
from RPMs to RCF
6 Press lid firmly down to
Locked position then press
start
5 Set Temperature
4 Adjust time 1- 60 minutes
For continuous rotate until
ldquo- -rdquo is displayed
3 Set Fast
or Slow
breaking
QUICKshort
runs press
and hold this
button after
setting other
values PRESS and HOLD to adjust
settings during operations
(change rpms temp time)
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Bench Z400 Bucket Configuration
STLCC-CPLSMorrison 6122013 Page 25
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Notes Bench Z400 Refrigerated BRDG
bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature
bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before
pressing this button expired time will show on display
bull Press STOP button for any emergency to interrupt the run
bull Take care to balance all loads carriers and tubes and contents
bull DO NOT mix plate carriers with buckets use all four of a single type on any spin
bull Lubrication of bucket swing points with vaseline may be periodically required
bull Never use cleaning agents with pH outside of the 6-8 range
bull Emergency Lid Release
ndash Switch off and unplug the unit
ndash Remove the white plastic cap in the left side wall near the top of the unit
ndash Pull on the red cord inside to release the lid
bull Error Messages (See Manual for other messages not listed)
ndash Error 1 Incorrect balance
ndash Error 2 Permanent imbalance service required
ndash Error 10 Overtemperature in rotor chamber (gt 50C)
ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required
ndash Error 20 No rotor identified sensor defective or improper rotors used
ndash Error 21 Start bit missing (magnet lost)
ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid
ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and
restore power connection to outlet
STLCC-CPLSMorrison 6122013 Page 26
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Z400K Rotor and Carrier Inventory BRDG
STLCC-CPLSMorrison 6122013 Page 27
C0383-75A Buckets For
12 x 15ml wo lid
C0383-75C Buckets For
7x50ml Conical Tubes wo lid
Swing Out Rotor
MIDSCI CAT C0383-75 Manf 22086 V01
Max Speed= 3500 rpm Max RCF x g = 2300
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
Microplate Carrier for
4 x 500ml Rotor Hermle
MIDSCI Cat C0383-75H
Manf 625020
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Bench Z400 Fixed Rotors
STLCC-CPLSMorrison 6122013 Page 28
Fixed-Angle Rotor 44x1520ml
MIDSCI Cat C0252-9 Manf 22088 V01
Max Speed = 13500 RPM Max RCF x g 17523
(On order Oct 2011)
Fixed-Angle Rotor 6x50ml Conical
MIDSCI CAT C0300-97 Manf 22097
Max Speed 6000 RPMs Max RCF xg 3863
Hotlink to Z400K Hermle Accessories and Rotors
Brochure pdf
C0232-8B
Adapter for 15ml Conical
in 50ml Rotor 6pk
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Mini BRDG TOMY Bench
STLCC-CPLSMorrison 6122013 Page 29
LidDoor Release
Timer Set
RPMs Set
Rotation will
begin
immediately if
ldquoonrdquo
Actual RPMs
Max = 15K OnOff Toggle
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Microfuge 11 BRDG Beckmann
STLCC-CPLSMorrison 6122013 Page 30
LidDoor Release
Timer Set Outer
white dial Must
be gt0 to start any
rotations
Set RPMs
Max is lt 10K
(Push button of
time dial pops
out when rpms
exceed10k) OnOff Toggle
Start switch
User after RPMs
and timegt0 is set
Push Button (center)
to activate timed run
or reset device for new
rpm settings
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 31
Centrifuge Micro Eppendorf 5415C Analog Controls
OnOff lower right
side
Timer gt 1 minute
Timer lt 1 minute
Push button to start timing
or hold down to continue wo
timer
Door to rotor Will
not open if timers
are set
Rotor assembly Clockwise
to tighten CCW to release
The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that
reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total
spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The
5415C centrifuge features microprocessor control of operating functions which permits
operation for very short spin times LED pilot light cycles on and off at 1 second intervals
for counting The Eppendorf also features a safety switch which prevents starting the
centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in
operation
Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Micro VWR Micro1816 Digital Controls
STLCC-CPLSMorrison 6122013 Page 32
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Micro Fisher Scientific- Micro14 Micro7
STLCC-CPLSMorrison 6122013 Page 33
Models Micro14 and Micro7
Values are set with the membrane keypad and displayed digitally
Speed can be set and displayed in rpm or rcf
Brushless maintenance free motors for fast acceleration and
smooth operation
complete with removable autoclavable rotors
Specifications
Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute
Size About 8rdquo x8rdquo x 6rdquo high
Requires Power Adapter Total Power Electronics Model TPE-
110V
Optional outputs vary with pins on using device max is 32V at
35A
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Microcentrifuge Labnet PRISM C2500
STLCC-CPLSMorrison 6122013 Page 34
Speed Range 500 to 15000 RPM
Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml
Timer 05 to 99 minutes or continuous Quick button for momentary
Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm
Set Time Depress membrane button LED
blinks use dial to set value
RPM or RCF Depress membrane button wait
for LED to blink use dial to set value
START Push DOWN on dial to START rotor
STOP Push down again on dial to Stop rotor
Set Lo or HI Accel depress RPM button for gt5
sec to toggle between these two options
Set Sound or No-Sound depress TIME button
gt 5 sec to toggle between these two options
OnOff toggle
Rear panel low
C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips
Mounting
1 Remove standard rotor lid
2 Unscrew standard rotor retaining nut
3 Install adapter on top of standard rotor
4 Push down on adapter and secure with rotor retaining nub
Hotlink to Labnet Prism Manual
C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Labnet Prism Parts and Accessories
STLCC-CPLSMorrison 6122013 Page 35
Labnet Tech Services (labnetwebcorningcom)
Labnet Tech Services (labnetwebcorningcom)
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Prism Labnet RA Number Form
STLCC-CPLSMorrison 6122013 Page 36
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge
Bench
Prism
Repair
Ticket March
2013
STLCC-CPLSMorrison 6122013 Page 37
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Microcentrifuge PRISM C2500 Wrench
STLCC-CPLSMorrison 6122013 Page 38
Rotor Wrench looks
like this
Rotor Wrench Backup
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Mini Labnet Spectrafuge
STLCC-CPLSMorrison 6122013 Page 39
Maximum speedRCF 6000rpm
Maximum capacity 2000g
Strip tube rotor 2 x 02 ml strips
Standard tube rotor 6 x 1520 ml
Dimensions (W x D x H) 5 x 5 x 44 inches
Weight 1 lb045 kg
Electrical 120V~ 60 Hz
Warranty 12 Months
OnOff Toggle switch
Start Rotor will start when lid is
closed if toggle is in on position
Stop Depress tab on front of
rotor lid then raise lid with
handfingers
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 40
bull Acrylic lid bull Lid latch with safety sensor
bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber
bull 300-watt heater
bull Microprocessor-controlled programming for run time from
1 to 999 minutes heat from OFF to HI (100deg C) in one degree
increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms
bull LCD for display of program number actual and set point temperature
heater time remaining and run time remaining
bull Control panel with RUNSTOP button PREHEAT button
set point SELECT button increasedecrease buttons and
3 Quick-Start buttons
bull Centrifugal motion is provided by a brushless motor capable of
achieving up to 1725 rpm
bull Quick-Stop System to stop vacuum pump open valve to bleed
air into the chamber and stop the rotor within seconds after time
expires or RUNSTOP button is pressed
bull Built-in diaphragm pump achieves vacuum of 296 Hg and has
displacement capacity of 34 liters per minute All wetted parts
are Teflon or Teflon-coated
bull Automatic vacuum release valve
bull Built-in vacuum delay allows rotor to achieve speed before
applying vacuum
bull Audible alarm to signal completion of set point run time
bull Two 177 milliliter glass traps
bull Vacuum port with quick disconnect fitting for 14 ID tubing
bull Includes DNA Rotor 7462900
bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug
bull 230 volt model has an 8-foot cord with NEMA 6-15 plug
bull ETL and ETL-C listing on 115 volt 60 Hz model
bull CE mark on 230 volt 5060 Hz model
bull Overall dimensions with closed lid 159 wide x 261 deep x
112 high (404 x 663 x 284 cm)
bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)
Link to Labconco CentriVap Info ndash pdf
Centrifuge Bench Labconco VapConcentrator BRDG
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Labconco Centri-Vap Concentrator
STLCC-CPLSMorrison 6122013 Page 41
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 42
Centrifuge Bench Labconco VapConcentratorControl Panel
Updown set
values
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 43
Centrifugation Centrifuge Classes
Mini-Micro
- 14
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 44
Centrifugation Differential Density Size
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
STLCC-CPLSMorrison 6122013 Page 45
Centrifugation Rotor Types
ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang
vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out
to a horizontal position This rotor is particularly useful when samples are to be resolved
in density gradients The longer pathlength permits better separation of individual particle
types from a mixture However this rotor is relatively inefficient for pelleting Also care
must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient
materials that can precipitate
ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity
When the rotor begins to rotate the solution in the tubes reorients This rotor type is
most commonly used for pelleting applications Examples include pelleting bacteria
yeast and other mammalian cells It is also useful for isopycnic separations of
macromolecules such as nucleic acids
ndash In vertical rotors sample tubes are held in vertical position during rotation This type of
rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)
separations due to the short pathlength Applications include plasmid DNA RNA and
lipoprotein isolations
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Micro IEC Micromax
STLCC-CPLSMorrison 6122013 Page 46
Model MicromaxMicromax RF
The IEC MicromaxMicromax RF delivers superior separating power in
less time Micromax reaches forces up to 21000 xg at speeds of 15000
rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than
15 seconds
MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg
at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)
microtubes IEC Micromaxs brushless motor is designed for longevity
and requires no maintenance All controls are microprocessor based for
added reliability
includes a choice of g-force or rpm speed control gentle and fast
acceleration and braking rates audible end of run signal as well as
momentary spin and time hold selections A glove friendly membrane
control panel with digital display makes it easy to clean and to operate
Repeat runs are as simple as touching a single membrane switch
Less Noise Less Space and Less Time Other less powerful
microcentrifuges are noisy distractions but the IEC Micromax operates
at a quiet 55 BA at full speed so you can work while Micromax works
Only 12 inches wide this machine requires less bench space than the
majority of microcentrifuges (particularly refrigerated) with comparable
sample capacity and power It takes less than 15 seconds to attain
21000 xg or to brake allowing you to process more samples per day
Unique Polypropylene Microtube Rotors Virtually impervious to chemical
attack and may be repeatedly autoclaved without damage New sealed
rotor provides aerosol containment and can be removed from the
Micromax (seal intact) and opened under an environmental hood
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Avanti J-20XP Rotor ___________BRDG
STLCC-CPLSMorrison 6122013 Page 47
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG
STLCC-CPLSMorrison 6122013 Page 48
Fisher
Part 51144
Plastic Lid
Fisher Part 04-976-012 3-99 Rotor
Red= Off
Green=On
Button Control Panel
amp Lid Release
Manufacturer Fisher Scientific
Model 21000R
Weight 160 lbs
Measurements 29 12 in x 19in x 15in
Electrical 120v 60Hz
Safety
Rotor containment ring around guard bowl offers
protection against rotor failure
Rotor reconition system that prevents each unit
from being programmed above max speed
Sensors inside rotor chamber monitor for
balance and alert you if an imbalance condition
is detected
Soft closing lid that locks when shut and cannot
be reopened unless indicated on the control
panel
Brushless motor is virtually maintenance free
Specifications
Capacity- 4x 250ml
Maximum tube size- 608 mm x 125 mm long
Max speed - 15000 rpm
Maximum Force- 21000xG
Speed Control - variable
Temperature range- -9c to +40c
Timer - Range 0 to 99min 59sec with continous
run Resolution 1 sec
Link to Fisher Sci 21000R
Operator Manual hellip pdf
Note Nut must be tightened with wrench to fully lock
rotor on spindle Finger tight is not good enough
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Bench Refr FisherMarathon 21000R Control
Panel
STLCC-CPLSMorrison 6122013 Page 49
Link to Fisher Sci 21000R
Operator Manual hellip pdf Press and use
keypad to set
speed then
ENTER
Press and use
keypad to set
run time then
ENTER
Press and use
keypad to set
run Temp then
ENTER Press MANUAL to skip use of Stored
Programs The device has automatic
rotor identification if standard rotors
are used
Press and lift lid
simultaneously Error Codes will be displayed here See list next slide
Program if not
in manual mode
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor
04-976-012BRDG
STLCC-CPLSMorrison 6122013 Page 50
Fisher 3-99 04-976-012
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL
04-976-009 BRDG
STLCC-CPLSMorrison 6122013 Page 51
Fisher 04-976-009
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL
with Adapter Tubes as required to ensure snug fit of Sample in rotor
STLCC-CPLSMorrison 6122013 Page 52
Fisher 04-976-009
Hint You can make adapters for 5mL sample
tubes by cutting off a 15mL tube just below the
cap location
You can make further reductions for a 2mL
sample tube by cutting down 5mL tubes just
below the cap location and inserting into the
5mL cutoff made above (three layers of tubes)
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters
04-976-006 BRDG
STLCC-CPLSMorrison 6122013 Page 53
TEC 12-98 51138 269OG
TEC 6561
Fisher 04-976-006 11-98
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Marathon Rotor Options 4 x 250ML buckets
STLCC-CPLSMorrison 6122013 Page 54
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrigure Marathon Rotor Options Microplate Microtube
STLCC-CPLSMorrison 6122013 Page 55
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Marathon Rotor Options Fixed Angle
STLCC-CPLSMorrison 6122013 Page 56
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Marathon Refr Error Codes
STLCC-CPLSMorrison 6122013 Page 57
Ch hd A Program mode has been set that does not
match the rotor Switch to Manual Mode and set all
Speed Time Temp settings
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Emergency Open Locked Lid or Power Out
STLCC-CPLSMorrison 6122013 Page 58
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments
Centrifuge Marathon Fuses inside case
STLCC-CPLSMorrison 6122013 Page 59
25A 250V fuse located on wall PC board
25A 250V fuse located under black shield
(remove with pop-up screws on side) on
PC board
Remove front display door by
screw on bottom then lift and
pull off Note location of any
loose wiring attachments