STLCC-CPLS:Morrison 6/12/2013 Page 1 Centrifuge Floor, Bench, Mini, Vap/Concentrator (Detect Instrument Picture to Jump to that section of SOP) Prepared by: Bob Morrison Instrumentation Specialist Original June 08, Last revision Nov 2012 Bench Mini Model 5415D Bench Micro, Model 5415C Floor Beckman Avanti Model J20 XP, @BRDG Floor Beckman Avanti Model J-E, @FV Bench Micro 14 CentriVap/Concentrator Labconco, @BRDG Bench,Refrig , Z400 @ BRDG TOMY, mini @BRDG Beckman micro <10K rpms Convert “g” and RPMs Link to SERVICE/REPAIR INFO Prism, micro Spectra-mini

Upload: others

Post on 09-Feb-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

STLCC-CPLSMorrison 6122013 Page 1

Centrifuge ndash Floor Bench Mini

VapConcentrator (Detect Instrument Picture to Jump to that section of SOP)

Prepared by Bob Morrison

Instrumentation Specialist

Original June 08 Last revision Nov 2012

Bench Mini

Model 5415D

Bench Micro

Model 5415C

Floor Beckman Avanti

Model J20 XP BRDG

Floor Beckman

Avanti Model J-E FV

Bench Micro 14

CentriVapConcentrator

Labconco BRDG BenchRefrig Z400

BRDG

TOMY mini

BRDG

Beckman micro

lt10K rpms

Centrifugation

Nomogram

RPMs to ldquogrdquo

Convert ldquogrdquo and RPMs

Link to SERVICEREPAIR INFO

Prism micro Spectra-mini

Centrifuge Local Service Center

STLCC-CPLSMorrison 6122013 Page 2

STLCC-CPLSMorrison 6122013 Page 3

Centrifuge Floor Model Beckman Avanti J-20XP BRDG

J series

Electrical 208V

Connector Type

System ID tag 432877 SN JXPO1K16 CAT 368608 Maximum Speed 26 000 rpm

Maximum g-force 82 000 times g

Speed Control plusmn 10 rpm of set speed or 01 whichever is greater

Drive Type Switched Reluctance Drive

Drive Cooling Air-cooled

Centrifuge Warranty 1 year

Drive Warranty 3 years

Refrigeration System Non-CFC non-ozone-depleting refrigerant

Max Capacity Volume 6 000 mL

Single Tube Sample Volume Range 15 ndash 1 000 mL

Ambient Temperature Range 15deg to 40degC

Set Temperature Range -10deg to 40degC

Temperature Control plusmn 2degC of set

User-Settable Programs J-20XPI 30 two-step programs

J-20XP No programming Zonal Capability J-20XPI Yes

J-20XP No Time Setting 1 min to 9959 hours hold or ω2t

(J-20XPI) 1 min to 3 hours hold (J-20XP)

AccelerationDeceleration Rates 12 accel13 decel (J-20XPI)

2 accel3 decel (J-20XP)

Dynamic Rotor Inertia Check Automatic

Heat Output 20 kW (6 900 BTUhr)

Sound Level lt 57 dBa (3 ft from instrument at maximum speed)

Dimensions 865 mm times 876 mm times 711 mm

(341rdquo H times 345rdquo D times 28rdquo W) Weight 264 kg (583 lb)

Power Nameplate 24amp Fuse 20amp 18 per UCLA test runs

Male plug 6-30P

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Floor Avanti J-20XP Control Panel BRDG

STLCC-CPLSMorrison 6122013 Page 4

3 Set dial to match

installed rotor

5 Set Speed

6 Set Time (hrmin)

7 Set Temp

10 Press START will

blink until set values

are reached

11 Press firmly on

doorlid at 4pm

position until rpm

setting is achieved

1 Turn ON MasterSwitch

2 Depress foot pedal to

release door and install

rotor

4 Press SET press

again to see

Actual

readings

Max

Slow

Max

Slow

Off

8 Press ACCEL to

set rate

9 Press DECEL to

set rate

Actual

Set

STOP will blink

until rotor

stops

Door

FRS

Centrifuge Avanti Warning Message DoorLid and FRS

STLCC-CPLSMorrison 6122013 Page 5

Holding down door lid after rpms reach about

4000 generally resolves FRS message

(improves seal of lid and internal vacuum

Door or Lid or FRS (Friction Reduction System) Errors

1) Hold the liddoor down firmly at the 4pm position after pressing START

2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches

3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber

to reduce friction on rotor for higher RPMs

4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal

Centrifuge Avanti DoorLid Errors Next Steps

STLCC-CPLSMorrison 6122013 Page 6

This solenoid may be stuck in

the ldquoopenrdquo position and thus

giving constant ldquodoorlid openrdquo

status to control panel

1 Try cleaning solenoid piston

shaft with q-tip

2 Slide black lock latch tofro

3 Remove and clean piston

shaft with clean cloth DO

NOT LUBRICATE piston as

this can cause overheating

in the solenoid

4 Call ACE Systems for

repairs if this does not work

solenoid may need

replacement or other

problems exist

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Local Service Center

STLCC-CPLSMorrison 6122013 Page 2

STLCC-CPLSMorrison 6122013 Page 3

Centrifuge Floor Model Beckman Avanti J-20XP BRDG

J series

Electrical 208V

Connector Type

System ID tag 432877 SN JXPO1K16 CAT 368608 Maximum Speed 26 000 rpm

Maximum g-force 82 000 times g

Speed Control plusmn 10 rpm of set speed or 01 whichever is greater

Drive Type Switched Reluctance Drive

Drive Cooling Air-cooled

Centrifuge Warranty 1 year

Drive Warranty 3 years

Refrigeration System Non-CFC non-ozone-depleting refrigerant

Max Capacity Volume 6 000 mL

Single Tube Sample Volume Range 15 ndash 1 000 mL

Ambient Temperature Range 15deg to 40degC

Set Temperature Range -10deg to 40degC

Temperature Control plusmn 2degC of set

User-Settable Programs J-20XPI 30 two-step programs

J-20XP No programming Zonal Capability J-20XPI Yes

J-20XP No Time Setting 1 min to 9959 hours hold or ω2t

(J-20XPI) 1 min to 3 hours hold (J-20XP)

AccelerationDeceleration Rates 12 accel13 decel (J-20XPI)

2 accel3 decel (J-20XP)

Dynamic Rotor Inertia Check Automatic

Heat Output 20 kW (6 900 BTUhr)

Sound Level lt 57 dBa (3 ft from instrument at maximum speed)

Dimensions 865 mm times 876 mm times 711 mm

(341rdquo H times 345rdquo D times 28rdquo W) Weight 264 kg (583 lb)

Power Nameplate 24amp Fuse 20amp 18 per UCLA test runs

Male plug 6-30P

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Floor Avanti J-20XP Control Panel BRDG

STLCC-CPLSMorrison 6122013 Page 4

3 Set dial to match

installed rotor

5 Set Speed

6 Set Time (hrmin)

7 Set Temp

10 Press START will

blink until set values

are reached

11 Press firmly on

doorlid at 4pm

position until rpm

setting is achieved

1 Turn ON MasterSwitch

2 Depress foot pedal to

release door and install

rotor

4 Press SET press

again to see

Actual

readings

Max

Slow

Max

Slow

Off

8 Press ACCEL to

set rate

9 Press DECEL to

set rate

Actual

Set

STOP will blink

until rotor

stops

Door

FRS

Centrifuge Avanti Warning Message DoorLid and FRS

STLCC-CPLSMorrison 6122013 Page 5

Holding down door lid after rpms reach about

4000 generally resolves FRS message

(improves seal of lid and internal vacuum

Door or Lid or FRS (Friction Reduction System) Errors

1) Hold the liddoor down firmly at the 4pm position after pressing START

2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches

3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber

to reduce friction on rotor for higher RPMs

4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal

Centrifuge Avanti DoorLid Errors Next Steps

STLCC-CPLSMorrison 6122013 Page 6

This solenoid may be stuck in

the ldquoopenrdquo position and thus

giving constant ldquodoorlid openrdquo

status to control panel

1 Try cleaning solenoid piston

shaft with q-tip

2 Slide black lock latch tofro

3 Remove and clean piston

shaft with clean cloth DO

NOT LUBRICATE piston as

this can cause overheating

in the solenoid

4 Call ACE Systems for

repairs if this does not work

solenoid may need

replacement or other

problems exist

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 3

Centrifuge Floor Model Beckman Avanti J-20XP BRDG

J series

Electrical 208V

Connector Type

System ID tag 432877 SN JXPO1K16 CAT 368608 Maximum Speed 26 000 rpm

Maximum g-force 82 000 times g

Speed Control plusmn 10 rpm of set speed or 01 whichever is greater

Drive Type Switched Reluctance Drive

Drive Cooling Air-cooled

Centrifuge Warranty 1 year

Drive Warranty 3 years

Refrigeration System Non-CFC non-ozone-depleting refrigerant

Max Capacity Volume 6 000 mL

Single Tube Sample Volume Range 15 ndash 1 000 mL

Ambient Temperature Range 15deg to 40degC

Set Temperature Range -10deg to 40degC

Temperature Control plusmn 2degC of set

User-Settable Programs J-20XPI 30 two-step programs

J-20XP No programming Zonal Capability J-20XPI Yes

J-20XP No Time Setting 1 min to 9959 hours hold or ω2t

(J-20XPI) 1 min to 3 hours hold (J-20XP)

AccelerationDeceleration Rates 12 accel13 decel (J-20XPI)

2 accel3 decel (J-20XP)

Dynamic Rotor Inertia Check Automatic

Heat Output 20 kW (6 900 BTUhr)

Sound Level lt 57 dBa (3 ft from instrument at maximum speed)

Dimensions 865 mm times 876 mm times 711 mm

(341rdquo H times 345rdquo D times 28rdquo W) Weight 264 kg (583 lb)

Power Nameplate 24amp Fuse 20amp 18 per UCLA test runs

Male plug 6-30P

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Floor Avanti J-20XP Control Panel BRDG

STLCC-CPLSMorrison 6122013 Page 4

3 Set dial to match

installed rotor

5 Set Speed

6 Set Time (hrmin)

7 Set Temp

10 Press START will

blink until set values

are reached

11 Press firmly on

doorlid at 4pm

position until rpm

setting is achieved

1 Turn ON MasterSwitch

2 Depress foot pedal to

release door and install

rotor

4 Press SET press

again to see

Actual

readings

Max

Slow

Max

Slow

Off

8 Press ACCEL to

set rate

9 Press DECEL to

set rate

Actual

Set

STOP will blink

until rotor

stops

Door

FRS

Centrifuge Avanti Warning Message DoorLid and FRS

STLCC-CPLSMorrison 6122013 Page 5

Holding down door lid after rpms reach about

4000 generally resolves FRS message

(improves seal of lid and internal vacuum

Door or Lid or FRS (Friction Reduction System) Errors

1) Hold the liddoor down firmly at the 4pm position after pressing START

2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches

3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber

to reduce friction on rotor for higher RPMs

4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal

Centrifuge Avanti DoorLid Errors Next Steps

STLCC-CPLSMorrison 6122013 Page 6

This solenoid may be stuck in

the ldquoopenrdquo position and thus

giving constant ldquodoorlid openrdquo

status to control panel

1 Try cleaning solenoid piston

shaft with q-tip

2 Slide black lock latch tofro

3 Remove and clean piston

shaft with clean cloth DO

NOT LUBRICATE piston as

this can cause overheating

in the solenoid

4 Call ACE Systems for

repairs if this does not work

solenoid may need

replacement or other

problems exist

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP Control Panel BRDG

STLCC-CPLSMorrison 6122013 Page 4

3 Set dial to match

installed rotor

5 Set Speed

6 Set Time (hrmin)

7 Set Temp

10 Press START will

blink until set values

are reached

11 Press firmly on

doorlid at 4pm

position until rpm

setting is achieved

1 Turn ON MasterSwitch

2 Depress foot pedal to

release door and install

rotor

4 Press SET press

again to see

Actual

readings

Max

Slow

Max

Slow

Off

8 Press ACCEL to

set rate

9 Press DECEL to

set rate

Actual

Set

STOP will blink

until rotor

stops

Door

FRS

Centrifuge Avanti Warning Message DoorLid and FRS

STLCC-CPLSMorrison 6122013 Page 5

Holding down door lid after rpms reach about

4000 generally resolves FRS message

(improves seal of lid and internal vacuum

Door or Lid or FRS (Friction Reduction System) Errors

1) Hold the liddoor down firmly at the 4pm position after pressing START

2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches

3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber

to reduce friction on rotor for higher RPMs

4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal

Centrifuge Avanti DoorLid Errors Next Steps

STLCC-CPLSMorrison 6122013 Page 6

This solenoid may be stuck in

the ldquoopenrdquo position and thus

giving constant ldquodoorlid openrdquo

status to control panel

1 Try cleaning solenoid piston

shaft with q-tip

2 Slide black lock latch tofro

3 Remove and clean piston

shaft with clean cloth DO

NOT LUBRICATE piston as

this can cause overheating

in the solenoid

4 Call ACE Systems for

repairs if this does not work

solenoid may need

replacement or other

problems exist

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Avanti Warning Message DoorLid and FRS

STLCC-CPLSMorrison 6122013 Page 5

Holding down door lid after rpms reach about

4000 generally resolves FRS message

(improves seal of lid and internal vacuum

Door or Lid or FRS (Friction Reduction System) Errors

1) Hold the liddoor down firmly at the 4pm position after pressing START

2) If Door error use cotton tipped applicator to wipeclean rectangular electrodes inside both lid latches

3) Maintain pressure at 4pm position until past 5000rpms when centrifuge attempts to remove air in chamber

to reduce friction on rotor for higher RPMs

4) For long runs place a heavy rotor at 4pm position to keep liddoor firmly closed for vacuum seal

Centrifuge Avanti DoorLid Errors Next Steps

STLCC-CPLSMorrison 6122013 Page 6

This solenoid may be stuck in

the ldquoopenrdquo position and thus

giving constant ldquodoorlid openrdquo

status to control panel

1 Try cleaning solenoid piston

shaft with q-tip

2 Slide black lock latch tofro

3 Remove and clean piston

shaft with clean cloth DO

NOT LUBRICATE piston as

this can cause overheating

in the solenoid

4 Call ACE Systems for

repairs if this does not work

solenoid may need

replacement or other

problems exist

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Avanti DoorLid Errors Next Steps

STLCC-CPLSMorrison 6122013 Page 6

This solenoid may be stuck in

the ldquoopenrdquo position and thus

giving constant ldquodoorlid openrdquo

status to control panel

1 Try cleaning solenoid piston

shaft with q-tip

2 Slide black lock latch tofro

3 Remove and clean piston

shaft with clean cloth DO

NOT LUBRICATE piston as

this can cause overheating

in the solenoid

4 Call ACE Systems for

repairs if this does not work

solenoid may need

replacement or other

problems exist

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti ACE second fix Door Sensor

STLCC-CPLSMorrison 6122013 Page 7

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Swinging-Bucket Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

13 000 26 500 1841 6 x 50 mL 300 mL

29 x 105 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS131 SN 01U-3617BRDG

STLCC-CPLSMorrison 6122013 Page 8

JS131 SN 01U-3617

Link to Beckman Avanti Rotor JS-131 - pdf

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP Rotor JLA-16250 SN 01U-1772

STLCC-CPLSMorrison 6122013 Page 9

JLA-16250 SN 01U-1772

Link to Beckman Avanti Rotor JLA-16250 - pdf

Fixed-Angle Rotor Aluminum

Lightweight alternative to conventional 6 250 rotors empty rotor weight is 103 kg (227 lb)

Max RPM Max g k Factor Number of Tubes Rotor Capacity

16 000 38 420 1 060 6 250 mL 15 L

62 120 mm

25 times 55 in

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP Rotor JA 2550 SN 01U-2540

STLCC-CPLSMorrison 6122013 Page 10

JA 2550 SN 01U-2540

Link to Beckman Avanti Rotor JA-2550 - pdf

Fixed-Angle Rotor Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

25 000 75 600 418 8 x 50 mL 400 mL

29 x 104 mm

1125 x 4 in

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP Rotor JS 43 SN 01U-617

STLCC-CPLSMorrison 6122013 Page 11

Swinging-Bucket Rotor Anodized Aluminum

Max RPM Max g k Factor Number of Tubes Rotor Capacity

4300 4220 11 800 4 x 750 mL 3 L

96 x 130 mm

38 x 52 in

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP JS43 Rotor with Microplus Plate Carrier BRDG

STLCC-CPLSMorrison 6122013 Page 12

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP JS 43 Rotor w Microplus carrier BRDG

STLCC-CPLSMorrison 6122013 Page 13

Microplus Plate Carrier

No 362394 MicroPlus Carrier Assembly Includes carrier base and

rubber pad to cushion plates Each carrier can hold up to 3 microplates1 deep-wellsquare well plate 1 rack of

MiniTubes or other labware in 96- well format Set of 2 Maximum allowable speed is 3250 rpm (1924 x g)

Link to Beckman Avanti Rotor JS-43 - pdf

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Avanti J Series Emergency Access to Rotor

(note This info from J25 Model and may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 14

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Avanti J-XP

Emergency Access to

Rotor Steps 5-8 (note This info from J25 Model and

may not be same for J-XP20)

STLCC-CPLSMorrison 6122013 Page 15

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Avanti

Warnings Rotor

Speed Door Sys

STLCC-CPLSMorrison 6122013 Page 16

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor BRDG Avanti Maintenance

STLCC-CPLSMorrison 6122013 Page 17

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifugation Nomogram RPMs to ldquogrdquo

STLCC-CPLSMorrison 6122013 Page 18

Protocols for centrifugation typically specify the amount of acceleration to be applied to the sample rather than specifying a

rotational speed such as revolutions per minute g = RCF = 000001118 times r times N2

g = Relative centrifuge force r = rotational radius (cm) N = rotating speed (revolutions per minute rmin)

Hot link to web site calculator for ldquogrdquo and RPMs for various Centrifuges

and rotors helliphttpwwwcurrentprotocolscomtoolsg-forcerpm-conversion-tool

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor and Bench Adapters

STLCC-CPLSMorrison 6122013 Page 19

392075 Adapters 15ml

Conical 17mm diameter

Marathon Microtube Adapters

15ml to 5ml

Marathon Fisher 04-974-

122B

Corning 8445

Marathon Fisher 004-974-

011A

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Floor Avanti J-20XP Lubricants BRDG

STLCC-CPLSMorrison 6122013 Page 20

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 21

Centrifuge Floor Models Beckman Avanti J-E FV

Specifications Depth 800 cm (315 in)

Height 914 cm (36 in) with door open 1524 cm (60 in)

Width 635 cm (25 in)

Weight 2674 kg (589 lb)

Approximate Acceleration TimeTwo profiles--MAX (full acceleration

to set speed) and SLOW

Approximate Deceleration TimeThree profiles--MAX SLOW and

OFF (coast)

Electrical Requirements 50 Hz 180 to 253 VAC 24 A 60 Hz

187to229216to264 VAC 24 A

Friction Reduction System (FRS) 190 mm (75 in) Hg

Maximum Heat Dissipation into Room under Steady-State

Conditions 20 kW (6900 Btuhr)

Noise Level 1 Meter in Front of Centrifuge lt 64 dBA

Set Speed Actual rotor speed plusmn 50 rpm of set speed

Set Temperature-10deg C to +40deg C in 1deg C increments

Set Time to 99 hr 59 min or continuous (hold)

Speed Range 300 to 21000 rpm

Temperature Control plusmn 2deg C of set temperature (after

equilibration)

Ambient Temperature Range 15deg C to 35deg C

Clearances Required 76 cm (30 in) sides 160 cm (625 in)

rear

Humidity Restrictions lt 95 (noncondensing)

Pollution Degree 2

Time -- Actual Display Indicates run time remainingdisplays

elapsed time

Finish Control panel silicone elastomer Top surface urethane

paint Other surfaces general purpose paint

Link to Beckman Avanti J-E Brochure ndash pdf

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Bench Refrigerated Z400K MIDSCILabnet

STLCC-CPLSMorrison 6122013 Page 22

Z400K Refrigerated Universal Centrifuge

Hermie Labortechnik GmbH (Germany)

74cm wide 34cm deep 58cm high

Weight 79kg approx 182lbs

Max speed 13500 rpm

Max RCF 17310g

8amp 120V 960watts 60 db noise level

Hotlink to Z400 Instruction Manual (22 pgs)hellip Pdf

Main ONOFF

Toggle

Plastic cap

emergency or

power out lid open

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 23

Centrifuge Bench Models General Operating Instructions

1 The device must be powered on before it will release the door to the rotor assembly for loading specimens Do not force the door open

2 DO NOT SET TIMERS until you are completely ready to run Once set they must run down to zero (0) before the door to the rotor can be opened Do not force the timers to zero position

3 Choose the proper rotor for the operation and tube sizes involved

4 To load open the rotor assembly door and loosen the cap turning counterclockwise (CCW) then loosen the rotor knob if necessary

5 After loading secure rotor internally with the cap knob turning clockwise (CW) until secure

6 Inspect tubes before use make sure they do not have hairline cracks or other signs of damage Use only tubes made of materials that suit the RPM and RCF requirements for your usage

7 Ensure that the tubes fit snugly in rotor positions Use Tube Adapters when appropriate Loose tubes may vibrate at high RPMs and cause cracking or breakage

8 Close rotor door and set timers if desired

9 Push the start button to engage the timers Hold down the start button to

manually continue centrifuge operations

Link to Beckman Avanti J-25 Manual

(Not J-20XP but has similar features)

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Z400 Controls Note Follow order below

STLCC-CPLSMorrison 6122013 Page 24

1 Press this button to release lid Dial

settings prior to a run cannot be made

if the lid is locked

Green LED is off when rotor is stopped

and lid can be opened

3 Rotate to Preset the RPMs

or RCF

2 Press to toggle

from RPMs to RCF

6 Press lid firmly down to

Locked position then press

start

5 Set Temperature

4 Adjust time 1- 60 minutes

For continuous rotate until

ldquo- -rdquo is displayed

3 Set Fast

or Slow

breaking

QUICKshort

runs press

and hold this

button after

setting other

values PRESS and HOLD to adjust

settings during operations

(change rpms temp time)

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Bench Z400 Bucket Configuration

STLCC-CPLSMorrison 6122013 Page 25

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Notes Bench Z400 Refrigerated BRDG

bull Pre-cooling is recommended run 10-15 minutes at 20 full RPMs with desired temperature

bull Use the QUICK button for short runs lt 60 seconds but be sure to set control variables before

pressing this button expired time will show on display

bull Press STOP button for any emergency to interrupt the run

bull Take care to balance all loads carriers and tubes and contents

bull DO NOT mix plate carriers with buckets use all four of a single type on any spin

bull Lubrication of bucket swing points with vaseline may be periodically required

bull Never use cleaning agents with pH outside of the 6-8 range

bull Emergency Lid Release

ndash Switch off and unplug the unit

ndash Remove the white plastic cap in the left side wall near the top of the unit

ndash Pull on the red cord inside to release the lid

bull Error Messages (See Manual for other messages not listed)

ndash Error 1 Incorrect balance

ndash Error 2 Permanent imbalance service required

ndash Error 10 Overtemperature in rotor chamber (gt 50C)

ndash Error 11 Temperature sensor overheated or too low solenoid valve broken service required

ndash Error 20 No rotor identified sensor defective or improper rotors used

ndash Error 21 Start bit missing (magnet lost)

ndash Error no 1-49 Forced stop rotor will be brakes with preset speed reset error by opening lid

ndash Error no 50-99 Emergency stop rotor stops brakeless Reset message by unplug and

restore power connection to outlet

STLCC-CPLSMorrison 6122013 Page 26

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Z400K Rotor and Carrier Inventory BRDG

STLCC-CPLSMorrison 6122013 Page 27

C0383-75A Buckets For

12 x 15ml wo lid

C0383-75C Buckets For

7x50ml Conical Tubes wo lid

Swing Out Rotor

MIDSCI CAT C0383-75 Manf 22086 V01

Max Speed= 3500 rpm Max RCF x g = 2300

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

Microplate Carrier for

4 x 500ml Rotor Hermle

MIDSCI Cat C0383-75H

Manf 625020

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Bench Z400 Fixed Rotors

STLCC-CPLSMorrison 6122013 Page 28

Fixed-Angle Rotor 44x1520ml

MIDSCI Cat C0252-9 Manf 22088 V01

Max Speed = 13500 RPM Max RCF x g 17523

(On order Oct 2011)

Fixed-Angle Rotor 6x50ml Conical

MIDSCI CAT C0300-97 Manf 22097

Max Speed 6000 RPMs Max RCF xg 3863

Hotlink to Z400K Hermle Accessories and Rotors

Brochure pdf

C0232-8B

Adapter for 15ml Conical

in 50ml Rotor 6pk

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Mini BRDG TOMY Bench

STLCC-CPLSMorrison 6122013 Page 29

LidDoor Release

Timer Set

RPMs Set

Rotation will

begin

immediately if

ldquoonrdquo

Actual RPMs

Max = 15K OnOff Toggle

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Microfuge 11 BRDG Beckmann

STLCC-CPLSMorrison 6122013 Page 30

LidDoor Release

Timer Set Outer

white dial Must

be gt0 to start any

rotations

Set RPMs

Max is lt 10K

(Push button of

time dial pops

out when rpms

exceed10k) OnOff Toggle

Start switch

User after RPMs

and timegt0 is set

Push Button (center)

to activate timed run

or reset device for new

rpm settings

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 31

Centrifuge Micro Eppendorf 5415C Analog Controls

OnOff lower right

side

Timer gt 1 minute

Timer lt 1 minute

Push button to start timing

or hold down to continue wo

timer

Door to rotor Will

not open if timers

are set

Rotor assembly Clockwise

to tighten CCW to release

The Eppendorf 5415C micro centrifuge is a compact quiet running centrifuge that

reaches a maximum speed of 14000rpm and RCF of 16000g within ten seconds Total

spin time is controlled by built-in 30-minute timer that automatically shuts off the unit The

5415C centrifuge features microprocessor control of operating functions which permits

operation for very short spin times LED pilot light cycles on and off at 1 second intervals

for counting The Eppendorf also features a safety switch which prevents starting the

centrifuge with lid open and safety lock which prevents opening of lid while centrifuge is in

operation

Includes 18 place Rotor inside Model Eppendorf F-45-18-11 5402 5415

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Micro VWR Micro1816 Digital Controls

STLCC-CPLSMorrison 6122013 Page 32

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Micro Fisher Scientific- Micro14 Micro7

STLCC-CPLSMorrison 6122013 Page 33

Models Micro14 and Micro7

Values are set with the membrane keypad and displayed digitally

Speed can be set and displayed in rpm or rcf

Brushless maintenance free motors for fast acceleration and

smooth operation

complete with removable autoclavable rotors

Specifications

Model 1418 Capacity 18 x 15 ml Max RCF =14000 =g Timer 99 minute

Size About 8rdquo x8rdquo x 6rdquo high

Requires Power Adapter Total Power Electronics Model TPE-

110V

Optional outputs vary with pins on using device max is 32V at

35A

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Microcentrifuge Labnet PRISM C2500

STLCC-CPLSMorrison 6122013 Page 34

Speed Range 500 to 15000 RPM

Maximum RCF 21200 xg Maximum Capacity 24 x 1520 ml

Timer 05 to 99 minutes or continuous Quick button for momentary

Dimensions (WxDxH) 945 x 138 x 748 in 24 x 35 x 19 cm

Set Time Depress membrane button LED

blinks use dial to set value

RPM or RCF Depress membrane button wait

for LED to blink use dial to set value

START Push DOWN on dial to START rotor

STOP Push down again on dial to Stop rotor

Set Lo or HI Accel depress RPM button for gt5

sec to toggle between these two options

Set Sound or No-Sound depress TIME button

gt 5 sec to toggle between these two options

OnOff toggle

Rear panel low

C2400-SS StripSpin Adapter for 24D 02 mL tubes and strips

Mounting

1 Remove standard rotor lid

2 Unscrew standard rotor retaining nut

3 Install adapter on top of standard rotor

4 Push down on adapter and secure with rotor retaining nub

Hotlink to Labnet Prism Manual

pdf

C2400-RC1 Rotor Plastic Cover for 24 Place Order thru MIDSCI

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Labnet Prism Parts and Accessories

STLCC-CPLSMorrison 6122013 Page 35

Labnet Tech Services (labnetwebcorningcom)

Labnet Tech Services (labnetwebcorningcom)

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Prism Labnet RA Number Form

STLCC-CPLSMorrison 6122013 Page 36

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge

Bench

Prism

Repair

Ticket March

2013

STLCC-CPLSMorrison 6122013 Page 37

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Microcentrifuge PRISM C2500 Wrench

STLCC-CPLSMorrison 6122013 Page 38

Rotor Wrench looks

like this

Rotor Wrench Backup

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Mini Labnet Spectrafuge

STLCC-CPLSMorrison 6122013 Page 39

Maximum speedRCF 6000rpm

Maximum capacity 2000g

Strip tube rotor 2 x 02 ml strips

Standard tube rotor 6 x 1520 ml

Dimensions (W x D x H) 5 x 5 x 44 inches

Weight 1 lb045 kg

Electrical 120V~ 60 Hz

Warranty 12 Months

OnOff Toggle switch

Start Rotor will start when lid is

closed if toggle is in on position

Stop Depress tab on front of

rotor lid then raise lid with

handfingers

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 40

bull Acrylic lid bull Lid latch with safety sensor

bull Phenol-free lid gasket bull Epoxy-coated aluminum chamber

bull 300-watt heater

bull Microprocessor-controlled programming for run time from

1 to 999 minutes heat from OFF to HI (100deg C) in one degree

increments and heat time from 1 to 999 minutes Memory stores 1 to 9 pgms

bull LCD for display of program number actual and set point temperature

heater time remaining and run time remaining

bull Control panel with RUNSTOP button PREHEAT button

set point SELECT button increasedecrease buttons and

3 Quick-Start buttons

bull Centrifugal motion is provided by a brushless motor capable of

achieving up to 1725 rpm

bull Quick-Stop System to stop vacuum pump open valve to bleed

air into the chamber and stop the rotor within seconds after time

expires or RUNSTOP button is pressed

bull Built-in diaphragm pump achieves vacuum of 296 Hg and has

displacement capacity of 34 liters per minute All wetted parts

are Teflon or Teflon-coated

bull Automatic vacuum release valve

bull Built-in vacuum delay allows rotor to achieve speed before

applying vacuum

bull Audible alarm to signal completion of set point run time

bull Two 177 milliliter glass traps

bull Vacuum port with quick disconnect fitting for 14 ID tubing

bull Includes DNA Rotor 7462900

bull 115 volt model has a 65-foot 3-wire cord with NEMA 5-15 plug

bull 230 volt model has an 8-foot cord with NEMA 6-15 plug

bull ETL and ETL-C listing on 115 volt 60 Hz model

bull CE mark on 230 volt 5060 Hz model

bull Overall dimensions with closed lid 159 wide x 261 deep x

112 high (404 x 663 x 284 cm)

bull Actual weight 69 lbs (31 kg) Shipping weight 75 lbs (34 kg)

Link to Labconco CentriVap Info ndash pdf

Centrifuge Bench Labconco VapConcentrator BRDG

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Labconco Centri-Vap Concentrator

STLCC-CPLSMorrison 6122013 Page 41

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 42

Centrifuge Bench Labconco VapConcentratorControl Panel

Updown set

values

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 43

Centrifugation Centrifuge Classes

Mini-Micro

- 14

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 44

Centrifugation Differential Density Size

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

STLCC-CPLSMorrison 6122013 Page 45

Centrifugation Rotor Types

ndash In swinging bucket rotors the sample tubes are loaded into individual buckets that hang

vertically while the rotor is at rest When the rotor begins to rotate the buckets swing out

to a horizontal position This rotor is particularly useful when samples are to be resolved

in density gradients The longer pathlength permits better separation of individual particle

types from a mixture However this rotor is relatively inefficient for pelleting Also care

must be taken to avoid ldquopoint loadsrdquo caused by spinning CsCl or other dense gradient

materials that can precipitate

ndash In fixed-angle rotors the sample tubes are held fixed at the angle of the rotor cavity

When the rotor begins to rotate the solution in the tubes reorients This rotor type is

most commonly used for pelleting applications Examples include pelleting bacteria

yeast and other mammalian cells It is also useful for isopycnic separations of

macromolecules such as nucleic acids

ndash In vertical rotors sample tubes are held in vertical position during rotation This type of

rotor is not suitable for pelleting applications but is most efficient for isopycnic (density)

separations due to the short pathlength Applications include plasmid DNA RNA and

lipoprotein isolations

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Micro IEC Micromax

STLCC-CPLSMorrison 6122013 Page 46

Model MicromaxMicromax RF

The IEC MicromaxMicromax RF delivers superior separating power in

less time Micromax reaches forces up to 21000 xg at speeds of 15000

rpm on 24 (15-20 mL) samples or 48 (05 mL) microtubes in less than

15 seconds

MicromaxMicromaxRF microcentrifuge produces forces up to 21000 xg

at speeds of 15000 rpm on 24 (15-20 mL) samples or 48 (05 mL)

microtubes IEC Micromaxs brushless motor is designed for longevity

and requires no maintenance All controls are microprocessor based for

added reliability

includes a choice of g-force or rpm speed control gentle and fast

acceleration and braking rates audible end of run signal as well as

momentary spin and time hold selections A glove friendly membrane

control panel with digital display makes it easy to clean and to operate

Repeat runs are as simple as touching a single membrane switch

Less Noise Less Space and Less Time Other less powerful

microcentrifuges are noisy distractions but the IEC Micromax operates

at a quiet 55 BA at full speed so you can work while Micromax works

Only 12 inches wide this machine requires less bench space than the

majority of microcentrifuges (particularly refrigerated) with comparable

sample capacity and power It takes less than 15 seconds to attain

21000 xg or to brake allowing you to process more samples per day

Unique Polypropylene Microtube Rotors Virtually impervious to chemical

attack and may be repeatedly autoclaved without damage New sealed

rotor provides aerosol containment and can be removed from the

Micromax (seal intact) and opened under an environmental hood

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Avanti J-20XP Rotor ___________BRDG

STLCC-CPLSMorrison 6122013 Page 47

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Bench Refrigerated FischerMarathon 21000R BRDG

STLCC-CPLSMorrison 6122013 Page 48

Fisher

Part 51144

Plastic Lid

Fisher Part 04-976-012 3-99 Rotor

Red= Off

Green=On

Button Control Panel

amp Lid Release

Manufacturer Fisher Scientific

Model 21000R

Weight 160 lbs

Measurements 29 12 in x 19in x 15in

Electrical 120v 60Hz

Safety

Rotor containment ring around guard bowl offers

protection against rotor failure

Rotor reconition system that prevents each unit

from being programmed above max speed

Sensors inside rotor chamber monitor for

balance and alert you if an imbalance condition

is detected

Soft closing lid that locks when shut and cannot

be reopened unless indicated on the control

panel

Brushless motor is virtually maintenance free

Specifications

Capacity- 4x 250ml

Maximum tube size- 608 mm x 125 mm long

Max speed - 15000 rpm

Maximum Force- 21000xG

Speed Control - variable

Temperature range- -9c to +40c

Timer - Range 0 to 99min 59sec with continous

run Resolution 1 sec

Link to Fisher Sci 21000R

Operator Manual hellip pdf

Note Nut must be tightened with wrench to fully lock

rotor on spindle Finger tight is not good enough

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Bench Refr FisherMarathon 21000R Control

Panel

STLCC-CPLSMorrison 6122013 Page 49

Link to Fisher Sci 21000R

Operator Manual hellip pdf Press and use

keypad to set

speed then

ENTER

Press and use

keypad to set

run time then

ENTER

Press and use

keypad to set

run Temp then

ENTER Press MANUAL to skip use of Stored

Programs The device has automatic

rotor identification if standard rotors

are used

Press and lift lid

simultaneously Error Codes will be displayed here See list next slide

Program if not

in manual mode

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Fischer-Marathon 8 x 15 ml x 28mL Fixed Angle Rotor

04-976-012BRDG

STLCC-CPLSMorrison 6122013 Page 50

Fisher 3-99 04-976-012

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Fisher Marathon Microtube Rotor 24 x 15mL x 5mL

04-976-009 BRDG

STLCC-CPLSMorrison 6122013 Page 51

Fisher 04-976-009

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Prep for CEQ Use Microtube Rotor 24 x 15mL

with Adapter Tubes as required to ensure snug fit of Sample in rotor

STLCC-CPLSMorrison 6122013 Page 52

Fisher 04-976-009

Hint You can make adapters for 5mL sample

tubes by cutting off a 15mL tube just below the

cap location

You can make further reductions for a 2mL

sample tube by cutting down 5mL tubes just

below the cap location and inserting into the

5mL cutoff made above (three layers of tubes)

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Marathon Swinging Bucket 4 x 250mL Plus Adapters

04-976-006 BRDG

STLCC-CPLSMorrison 6122013 Page 53

TEC 12-98 51138 269OG

TEC 6561

Fisher 04-976-006 11-98

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Marathon Rotor Options 4 x 250ML buckets

STLCC-CPLSMorrison 6122013 Page 54

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrigure Marathon Rotor Options Microplate Microtube

STLCC-CPLSMorrison 6122013 Page 55

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Marathon Rotor Options Fixed Angle

STLCC-CPLSMorrison 6122013 Page 56

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Marathon Refr Error Codes

STLCC-CPLSMorrison 6122013 Page 57

Ch hd A Program mode has been set that does not

match the rotor Switch to Manual Mode and set all

Speed Time Temp settings

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Emergency Open Locked Lid or Power Out

STLCC-CPLSMorrison 6122013 Page 58

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments

Centrifuge Marathon Fuses inside case

STLCC-CPLSMorrison 6122013 Page 59

25A 250V fuse located on wall PC board

25A 250V fuse located under black shield

(remove with pop-up screws on side) on

PC board

Remove front display door by

screw on bottom then lift and

pull off Note location of any

loose wiring attachments