Breathing Air Quality Analyser
The AIRQUAL-1 has been designed to fully comply with the EN12021 legislation for the testing of contaminants within compressed breathing air lines.
The AIRQUAL-1 allows for a convenient ‘on the spot’ measurement of contaminants such as oil mist and Carbon Monoxide to ensure users receive clean and healthy breathing air when using air fed masks.
Tel: 01908 506819 | Email: info@elitemeasurement.com | www.elitemeasurement.com
Product Overview
Elite Measurement Solutions, Unit 14 Cromwell Business Centre, Howard Way, Newport Pagnell, Buckinghamshire, MK16 9QS
Material Solutions
Measurement Solutions
Main features This comprehensive test kit is compact, easy to operate and can be used to indicate the level of contamination, as well as, operating flow, temperature and pressure. It is supplied complete with oil mist, water vapour, CO and CO2 test tubes and real-time Oxygen analyser to allow immediate multiple contaminant testing.
Basic featuresSupplied in a robust, lightweight (4Kgs) and easily transportable Aluminium case the AIRQUAL-1 is rated to work on lines with a pressure of up to 10 Bar.
The kit includes:
• Calibration flow meters• ¼” female BSP connection• Stop Watch• Complete set of Gastec tubes of client’s choice• Tube tip breaker
Benefits• Low cost tubes only £13 per test• No additional equipment required• On-site customer calibration function• No down time• No batteries required• Produced from Aluminium and Stainless Steel - Ideal for
the shipping industry
SpecificationUsage specification Maximum inlet pressure: 10 bar g (145 psi g)Analysis operating pressure: 1 bar g 14.5 psi g) Maximum inlet temperature: 40°C (104°F)Minimum inlet temperature: 5°C (59°F)Flow accuracy: ±4% outletAir flow rate range at outlet: 30 - 2500 cc/min Hose connections: 1⁄4” BSP Female Approved detector tubes: Gastec Airtec tubes (or other
tubes subject to calibration)
Elite Measurement Solutions, Unit 14 Cromwell Business Centre, Howard Way, Newport Pagnell, Buckinghamshire, MK16 9QS
Breathing Air Quality Analyser
For further information on this product or other products that are available please call
Email: info@elitemeasurement.com www.elitemeasurement.com
01908 506819
Compact & easy to use system
StandardsSubstance Standards (BS EN12021)Oxygen 21% (+/-1%)Carbon Dioxide Not more than 500ppmCarbon Monoxide Not more than 15ppmOil Mist/Vapour Not more than 0.5mg/m³Odour/Taste Without significant odour or tasteWater (Liquid) There should be no free water
Water (Vapour) Air for compressed air line breathing apparatus shall have a dewpoint sufficiently low to prevent condensation & freezing. Where apparatus is used and stored at a known temperature pressure dewpoint shall be at least 5°C below the likely lowest temperature. Where conditions of usage and storage of the air is not known the pressure dewpoint shall not exceed -11°C.
Contaminant Part Number Range Sample Time
Carbon Monoxide GAS1A 5-50ppm 3 mins
Carbon Dioxide GAS2A 250-3000ppm 5 mins
Carbon Dioxide GAS2AG 200-3000ppm 1.5 mins
Water Vapour GAS6AH 500-5000ppm 1 mins
Water Vapour GAS6A 30-80mg/m3 10 mins
Oil Mist GAS1096A 0.3-1.5mg/m3 60 mins
Oil Mist GAS1096AD 0.2-5.0mg/m3 20 mins
Nitrogen Oxides GAS11A 0.06-0.7ppm 2 mins
Oxygen 18100050-3 0-300% Instant