© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 1

Trust but always Verify:Best practices in spectrophotometric measurement

L S Theobald Lecture15th May 2003

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 2

A journey in analytical science

Sourced in classical analysis (ABC)Inspired by qualitative analysisProgressed via colorimetryMatured in spectrophotometryContinued quest for correctness and traceability in spectroscopic measurements

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 3

Sourced in classical analysis

ICI General Chemicals DivisionWidnes 1962-1967‘Techniques Course’ Widnes Tech.12 months before you were ‘acceptable’ within the analytical laboratoryClassical gravimetric and volumetric analysis (Over 150 procedures against ‘unknown’ samples from Vogel 3rd

Edition)Trained analyst

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 4

Traceability in classical analysisGravimetric analysis

Traceable massesMethod of swings on 2 pan balancesSubstitutionBuoyancy correction

Volumetric analysisCalibration of pipettesCalibration of burettes with drainage correctionCalibration of volumetric flasks

Chemical Analysis –The Working ToolsAnalytical Chemists’ Committee of ICI2nd Edition 1961

Traceability in volumetric analysis

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 6

Qualitative analysis

The flame test; low cost atomic emission (Flame and bead tests data have been restored in the 84th Edition, 2003-2004, CRC Handbook of Chemistry and Physics)Calibration of spectroscopes

‘Raies Ultimes’Dispersion curves

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 7

Bunsen spectroscope

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 8

The eye as a detector

Approximate wavelength range (nm)

Colour of reflected light

Colour of absorbed light

400-456465-482482-487487-493493-498498-530530-559559-571571-576576-580580-587587-597597-617617-780

VioletBlueGreenish-blueBlue-greenBluish-greenGreenYellowish-greenYellow-greenGreenish-yellowYellowYellowish-orangeOrangeReddish-orangeRed

Yellow-greenYellowOrangeRed-orangeRedRed-purpleReddish-purplePurpleVioletBlueBlueGreenish-blueBlue-greenBlue-green

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 9

Absorptiometry; the eye as a detector

Nessler tubesIron by thioglycollic acidStandard seriesVisual interpolation

Nessleriser Glass standards‘Permanent’ standardTraceable (?)Samples only

Lovibond tintometer

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 10

Nessleriser standards

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 11

Absorptiometry; the photocell as a

detector

Hilger Spekker absorptiometerMeasurement of transmittanceCalibration curves and the problem of absorbanceFilter selection

MeasurementBand widthStray light

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 12

Hilger Spekker Absorptiometer

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 13

Hilger Spekker AbsorptiometerLight from lamp G illuminates 2 photocells A & O via optical nullF & H are heat absorbing filtersM is the wavelength filterI is the variable windowThe sample is in the cell L

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 14

Band-pass filters for filter photometers

The chromogenic reagent yields an orange solution (D)

Filter A will be satisfactory

Filters B & C transmit considerable radiation where the solution has little absorption

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 15

Loughborough University of Technology

1967 to 1972The Photoelectric Spectrometry Group BulletinBasic dyes as chromogenic reagentsElwell Ward 1971 Society for Analytical Chemistry

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 16

Hilger & Watts Uvispeck

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 17

Unicam SP 600

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 18

Basics of spectrophotometry

Theoretical aspectsNuts & bolts Confidence in the numbers!J R Edisbury (1966)‘Practical hints on absorption spectrometry 200 -800mμ’

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 19

Beer Lambert Law Assumptions

The radiation is perfectly monochromaticThere are no uncompensated losses due to scattering or reflectionThe radiation beam strikes the cuvette at normal incidenceThere are no molecular interactions between the absorber and other molecules in solutionThe temperature remains constant

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 20

Two More Assumptions

The instrument works properlyThe person operating it is competent

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 21

Beer’s Law and Instrument Linearity

00

0

The voltage generated by the detector is a function of the power of the incident radiation falling on it; Beer's law

or log

Combining these two equations we get (

cb

V I

II I e A cbI

V I I e

β

ε

β

ε−

=

⎛ ⎞= = =⎜ ⎟⎝ ⎠

= =

0

)and hence the measured absorbance is

log

and therefore a proportional change in c or b only confirms that is constant and does not confirm unity

cb cboI e

AV cbV

ε β β εβ

εβ

β

− −=

⎛ ⎞ =⎜ ⎟⎝ ⎠

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 22

Nuts and bolts

Know your instrumentImportance of optical design

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 23

Pye Unicam SP500

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 24

Double beam recording design

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 25

HP 8450 diode array spectrometer

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 26

The rise of spectrophotometry

Photoelectric Spectrometry Group1948Beckman model DU & Unicam SP 500UVSG50th anniversary in 1998

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 27

UVSG monographs

1. Standards in Absorption SpectrometryEd. C. Burgess & A. Knowles, 1981

2. Standards in Fluorescence SpectrometryEd. J. N. Miller, 1981

3. Practical Absorption SpectrometryEd. A. Knowles & C. Burgess 1984

4. UV Spectroscopy; Techniques, instrumentation, data handlingEd. B.J. Clark, T. Frost & M.A. Russell, 1993

5. Standards and Best Practice in Absorption SpectrometryEd. C. Burgess & T. Frost, 1999

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 28

UVSG/CORM conferences

1. Advances in Standards and Methodology in SpectrophotometryEd. C. Burgess & K.D. Mielenz, 1987

2. Spectrophotometry, Luminescence and Colour;Science and ComplianceEd. C. Burgess & D.G. Jones, 1995

3. Optical Spectrometry: Applications and Instrumentation into the 21st Century1998

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 29

Spectrophotometry in a regulated environment

Glaxo Laboratories Ltd 1974Monitoring the performance of UV-visible spectrophotometers;UV Spectrometry Group Bulletin 5 77, 1997

Control chartsStandards for accuracy and precision

Ho(III) & Sm(III)Thompson’s solutionSap Green solutions

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 30

Ho(III) as a wavelength standard

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 31

Qualification and calibration of

spectrophotometers

Basis for measurementWavelength scale

Operational rangeAccuracy & precision

Photometric scaleOperational rangeAccuracy & precision (Linearity)Noise and drift

ResolutionStray light

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 32

Traceable standards

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 33

Starna RM-RE

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 34

Long term photometric stability

How reliable is the instrument over its working life?How stable is the standard?How stable is the standardising laboratory?

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 35

HP 8450 life time photometric accuracy

• NPL calibratedneutral density filter

• 546nm checked daily• Nine years

(110 months) • 2161 data points• Monthly averages• 95% CLs

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 36

Potassium dichromate

• 2 maxima/minimapairs

• One close to thevisible

• One in ‘bomb alley’ of UV region

• Easy to prepare(60mg in 0.005Msulphuric acid)

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 37

HP 8450 photometric accuracy

• acid dichromate• Four years

(47 months) • Duplicates• Satisfactory

performance• Reading slightly low at

350nm• No correlation max/min

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 38

HP 8450 photometric accuracy

• acid dichromate• Four years

(47 months) • Duplicates• Less satisfactory

performance in UV• Correlation max/min

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 39

HP 8450 photometric accuracy

• acid dichromate• Four years

(47 months) • Duplicates• Maxima to maxima• No correlation

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 40

Technology transfer standards

StableNon hazardousUV & Visible rangeEasy to use

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 41

Sap Green 1979

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 42

Sap Green Linearity

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 43

Comparison of scanning and DAS spectrophotometers 1990

7 DAS (HP8450 & HP8452)6 Scanning instruments (PE Lambda 5, 6 and 15)All labs had the ‘same’ solution of Sap Green

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 44

Collaborative trials with Sap Green

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 45

Collaborative trials with Sap Green

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 46

Spectrophotometric methods

The instrument is OK but what about the method?

Optimum conditionsBest precisionBest accuracy

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 47

Precision and noise

10

0.4343log

however is a function of

c Tc

T

s sRSDc T Ts T

= =

The relative standard deviation of concentration

Three main sources;1. Electronics, which is a constant (Johnson Noise)

2. Detector, photon statistics (Shot noise)

3. Cuvette positioning errors and source flicker

1Ts k=

22Ts k T T= +

3Ts k T=

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 48

Precision and noise

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 49

Relative errors;HP 8450 diode array spectrometer

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 50

Accurate values of absorbance

Setting standard valuesMinimising biasLess out of specification resultsForgotten lessons

Ringbom-Ayres plotsVFG method

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 51

Vandenbelt, Forsyth and GarrettInd. Eng. Chem. (Analytical Edition) 1945

• variation of molar extinction coefficient with absorbance

• Impact on standard values

• Beckman DU working absorbance range 0.5 to 1.9 for broad bands

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 52

VFG plot

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 53

In conclusion

Spectrophotometry is alive if not wellStandards are becoming more importantTrust your instruments and your methods BUT always verifyThe journey continues…

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 54

NIR wavelength for the future?