© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 1
Trust but always Verify:Best practices in spectrophotometric measurement
L S Theobald Lecture15th May 2003
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 2
A journey in analytical science
Sourced in classical analysis (ABC)Inspired by qualitative analysisProgressed via colorimetryMatured in spectrophotometryContinued quest for correctness and traceability in spectroscopic measurements
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 3
Sourced in classical analysis
ICI General Chemicals DivisionWidnes 1962-1967‘Techniques Course’ Widnes Tech.12 months before you were ‘acceptable’ within the analytical laboratoryClassical gravimetric and volumetric analysis (Over 150 procedures against ‘unknown’ samples from Vogel 3rd
Edition)Trained analyst
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 4
Traceability in classical analysisGravimetric analysis
Traceable massesMethod of swings on 2 pan balancesSubstitutionBuoyancy correction
Volumetric analysisCalibration of pipettesCalibration of burettes with drainage correctionCalibration of volumetric flasks
Chemical Analysis –The Working ToolsAnalytical Chemists’ Committee of ICI2nd Edition 1961
Traceability in volumetric analysis
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 6
Qualitative analysis
The flame test; low cost atomic emission (Flame and bead tests data have been restored in the 84th Edition, 2003-2004, CRC Handbook of Chemistry and Physics)Calibration of spectroscopes
‘Raies Ultimes’Dispersion curves
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 7
Bunsen spectroscope
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 8
The eye as a detector
Approximate wavelength range (nm)
Colour of reflected light
Colour of absorbed light
400-456465-482482-487487-493493-498498-530530-559559-571571-576576-580580-587587-597597-617617-780
VioletBlueGreenish-blueBlue-greenBluish-greenGreenYellowish-greenYellow-greenGreenish-yellowYellowYellowish-orangeOrangeReddish-orangeRed
Yellow-greenYellowOrangeRed-orangeRedRed-purpleReddish-purplePurpleVioletBlueBlueGreenish-blueBlue-greenBlue-green
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 9
Absorptiometry; the eye as a detector
Nessler tubesIron by thioglycollic acidStandard seriesVisual interpolation
Nessleriser Glass standards‘Permanent’ standardTraceable (?)Samples only
Lovibond tintometer
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 10
Nessleriser standards
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 11
Absorptiometry; the photocell as a
detector
Hilger Spekker absorptiometerMeasurement of transmittanceCalibration curves and the problem of absorbanceFilter selection
MeasurementBand widthStray light
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 12
Hilger Spekker Absorptiometer
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 13
Hilger Spekker AbsorptiometerLight from lamp G illuminates 2 photocells A & O via optical nullF & H are heat absorbing filtersM is the wavelength filterI is the variable windowThe sample is in the cell L
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 14
Band-pass filters for filter photometers
The chromogenic reagent yields an orange solution (D)
Filter A will be satisfactory
Filters B & C transmit considerable radiation where the solution has little absorption
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 15
Loughborough University of Technology
1967 to 1972The Photoelectric Spectrometry Group BulletinBasic dyes as chromogenic reagentsElwell Ward 1971 Society for Analytical Chemistry
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 16
Hilger & Watts Uvispeck
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 17
Unicam SP 600
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 18
Basics of spectrophotometry
Theoretical aspectsNuts & bolts Confidence in the numbers!J R Edisbury (1966)‘Practical hints on absorption spectrometry 200 -800mμ’
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 19
Beer Lambert Law Assumptions
The radiation is perfectly monochromaticThere are no uncompensated losses due to scattering or reflectionThe radiation beam strikes the cuvette at normal incidenceThere are no molecular interactions between the absorber and other molecules in solutionThe temperature remains constant
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 20
Two More Assumptions
The instrument works properlyThe person operating it is competent
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 21
Beer’s Law and Instrument Linearity
00
0
The voltage generated by the detector is a function of the power of the incident radiation falling on it; Beer's law
or log
Combining these two equations we get (
cb
V I
II I e A cbI
V I I e
β
ε
β
ε−
=
⎛ ⎞= = =⎜ ⎟⎝ ⎠
= =
0
)and hence the measured absorbance is
log
and therefore a proportional change in c or b only confirms that is constant and does not confirm unity
cb cboI e
AV cbV
ε β β εβ
εβ
β
− −=
⎛ ⎞ =⎜ ⎟⎝ ⎠
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 22
Nuts and bolts
Know your instrumentImportance of optical design
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 23
Pye Unicam SP500
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 24
Double beam recording design
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 25
HP 8450 diode array spectrometer
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 26
The rise of spectrophotometry
Photoelectric Spectrometry Group1948Beckman model DU & Unicam SP 500UVSG50th anniversary in 1998
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 27
UVSG monographs
1. Standards in Absorption SpectrometryEd. C. Burgess & A. Knowles, 1981
2. Standards in Fluorescence SpectrometryEd. J. N. Miller, 1981
3. Practical Absorption SpectrometryEd. A. Knowles & C. Burgess 1984
4. UV Spectroscopy; Techniques, instrumentation, data handlingEd. B.J. Clark, T. Frost & M.A. Russell, 1993
5. Standards and Best Practice in Absorption SpectrometryEd. C. Burgess & T. Frost, 1999
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 28
UVSG/CORM conferences
1. Advances in Standards and Methodology in SpectrophotometryEd. C. Burgess & K.D. Mielenz, 1987
2. Spectrophotometry, Luminescence and Colour;Science and ComplianceEd. C. Burgess & D.G. Jones, 1995
3. Optical Spectrometry: Applications and Instrumentation into the 21st Century1998
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 29
Spectrophotometry in a regulated environment
Glaxo Laboratories Ltd 1974Monitoring the performance of UV-visible spectrophotometers;UV Spectrometry Group Bulletin 5 77, 1997
Control chartsStandards for accuracy and precision
Ho(III) & Sm(III)Thompson’s solutionSap Green solutions
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 30
Ho(III) as a wavelength standard
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 31
Qualification and calibration of
spectrophotometers
Basis for measurementWavelength scale
Operational rangeAccuracy & precision
Photometric scaleOperational rangeAccuracy & precision (Linearity)Noise and drift
ResolutionStray light
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 32
Traceable standards
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 33
Starna RM-RE
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 34
Long term photometric stability
How reliable is the instrument over its working life?How stable is the standard?How stable is the standardising laboratory?
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 35
HP 8450 life time photometric accuracy
• NPL calibratedneutral density filter
• 546nm checked daily• Nine years
(110 months) • 2161 data points• Monthly averages• 95% CLs
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 36
Potassium dichromate
• 2 maxima/minimapairs
• One close to thevisible
• One in ‘bomb alley’ of UV region
• Easy to prepare(60mg in 0.005Msulphuric acid)
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 37
HP 8450 photometric accuracy
• acid dichromate• Four years
(47 months) • Duplicates• Satisfactory
performance• Reading slightly low at
350nm• No correlation max/min
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 38
HP 8450 photometric accuracy
• acid dichromate• Four years
(47 months) • Duplicates• Less satisfactory
performance in UV• Correlation max/min
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 39
HP 8450 photometric accuracy
• acid dichromate• Four years
(47 months) • Duplicates• Maxima to maxima• No correlation
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 40
Technology transfer standards
StableNon hazardousUV & Visible rangeEasy to use
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 41
Sap Green 1979
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 42
Sap Green Linearity
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 43
Comparison of scanning and DAS spectrophotometers 1990
7 DAS (HP8450 & HP8452)6 Scanning instruments (PE Lambda 5, 6 and 15)All labs had the ‘same’ solution of Sap Green
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 44
Collaborative trials with Sap Green
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 45
Collaborative trials with Sap Green
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 46
Spectrophotometric methods
The instrument is OK but what about the method?
Optimum conditionsBest precisionBest accuracy
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 47
Precision and noise
10
0.4343log
however is a function of
c Tc
T
s sRSDc T Ts T
= =
The relative standard deviation of concentration
Three main sources;1. Electronics, which is a constant (Johnson Noise)
2. Detector, photon statistics (Shot noise)
3. Cuvette positioning errors and source flicker
1Ts k=
22Ts k T T= +
3Ts k T=
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 48
Precision and noise
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 49
Relative errors;HP 8450 diode array spectrometer
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 50
Accurate values of absorbance
Setting standard valuesMinimising biasLess out of specification resultsForgotten lessons
Ringbom-Ayres plotsVFG method
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 51
Vandenbelt, Forsyth and GarrettInd. Eng. Chem. (Analytical Edition) 1945
• variation of molar extinction coefficient with absorbance
• Impact on standard values
• Beckman DU working absorbance range 0.5 to 1.9 for broad bands
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 52
VFG plot
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 53
In conclusion
Spectrophotometry is alive if not wellStandards are becoming more importantTrust your instruments and your methods BUT always verifyThe journey continues…
© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: chris@burgessconsultancy.com 54
NIR wavelength for the future?