Trust but always Verify:Best practices in spectrophotometric measurement

L S Theobald Lecture15th May 2003

Page 2: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 2

A journey in analytical science

Sourced in classical analysis (ABC)Inspired by qualitative analysisProgressed via colorimetryMatured in spectrophotometryContinued quest for correctness and traceability in spectroscopic measurements

Page 3: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 3

Sourced in classical analysis

ICI General Chemicals DivisionWidnes 1962-1967‘Techniques Course’ Widnes Tech.12 months before you were ‘acceptable’ within the analytical laboratoryClassical gravimetric and volumetric analysis (Over 150 procedures against ‘unknown’ samples from Vogel 3rd

Edition)Trained analyst

Page 4: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 4

Traceability in classical analysisGravimetric analysis

Traceable massesMethod of swings on 2 pan balancesSubstitutionBuoyancy correction

Volumetric analysisCalibration of pipettesCalibration of burettes with drainage correctionCalibration of volumetric flasks

Chemical Analysis –The Working ToolsAnalytical Chemists’ Committee of ICI2nd Edition 1961

Page 5: Best practices in spectrophotometric measurement

Traceability in volumetric analysis

Page 6: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 6

Qualitative analysis

The flame test; low cost atomic emission (Flame and bead tests data have been restored in the 84th Edition, 2003-2004, CRC Handbook of Chemistry and Physics)Calibration of spectroscopes

‘Raies Ultimes’Dispersion curves

Page 7: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 7

Bunsen spectroscope

Page 8: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 8

The eye as a detector

Approximate wavelength range (nm)

Colour of reflected light

Colour of absorbed light

400-456465-482482-487487-493493-498498-530530-559559-571571-576576-580580-587587-597597-617617-780

VioletBlueGreenish-blueBlue-greenBluish-greenGreenYellowish-greenYellow-greenGreenish-yellowYellowYellowish-orangeOrangeReddish-orangeRed

Yellow-greenYellowOrangeRed-orangeRedRed-purpleReddish-purplePurpleVioletBlueBlueGreenish-blueBlue-greenBlue-green

Page 9: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 9

Absorptiometry; the eye as a detector

Nessler tubesIron by thioglycollic acidStandard seriesVisual interpolation

Nessleriser Glass standards‘Permanent’ standardTraceable (?)Samples only

Lovibond tintometer

Page 10: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 10

Nessleriser standards

Page 11: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 11

Absorptiometry; the photocell as a

detector

Hilger Spekker absorptiometerMeasurement of transmittanceCalibration curves and the problem of absorbanceFilter selection

MeasurementBand widthStray light

Page 12: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 12

Hilger Spekker Absorptiometer

Page 13: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 13

Hilger Spekker AbsorptiometerLight from lamp G illuminates 2 photocells A & O via optical nullF & H are heat absorbing filtersM is the wavelength filterI is the variable windowThe sample is in the cell L

Page 14: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 14

Band-pass filters for filter photometers

The chromogenic reagent yields an orange solution (D)

Filter A will be satisfactory

Filters B & C transmit considerable radiation where the solution has little absorption

Page 15: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 15

Loughborough University of Technology

1967 to 1972The Photoelectric Spectrometry Group BulletinBasic dyes as chromogenic reagentsElwell Ward 1971 Society for Analytical Chemistry

Page 16: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 16

Hilger & Watts Uvispeck

Page 17: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 17

Unicam SP 600

Page 18: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 18

Basics of spectrophotometry

Theoretical aspectsNuts & bolts Confidence in the numbers!J R Edisbury (1966)‘Practical hints on absorption spectrometry 200 -800mμ’

Page 19: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 19

Beer Lambert Law Assumptions

The radiation is perfectly monochromaticThere are no uncompensated losses due to scattering or reflectionThe radiation beam strikes the cuvette at normal incidenceThere are no molecular interactions between the absorber and other molecules in solutionThe temperature remains constant

Page 20: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 20

Two More Assumptions

The instrument works properlyThe person operating it is competent

Page 21: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 21

Beer’s Law and Instrument Linearity

00

0

The voltage generated by the detector is a function of the power of the incident radiation falling on it; Beer's law

or log

Combining these two equations we get (

cb

V I

II I e A cbI

V I I e

β

ε

β

ε−

=

⎛ ⎞= = =⎜ ⎟⎝ ⎠

= =

0

)and hence the measured absorbance is

log

and therefore a proportional change in c or b only confirms that is constant and does not confirm unity

cb cboI e

AV cbV

ε β β εβ

εβ

β

− −=

⎛ ⎞ =⎜ ⎟⎝ ⎠

Page 22: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 22

Nuts and bolts

Know your instrumentImportance of optical design

Page 23: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 23

Pye Unicam SP500

Page 24: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 24

Double beam recording design

Page 25: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 25

HP 8450 diode array spectrometer

Page 26: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 26

The rise of spectrophotometry

Photoelectric Spectrometry Group1948Beckman model DU & Unicam SP 500UVSG50th anniversary in 1998

Page 27: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 27

UVSG monographs

1. Standards in Absorption SpectrometryEd. C. Burgess & A. Knowles, 1981

2. Standards in Fluorescence SpectrometryEd. J. N. Miller, 1981

3. Practical Absorption SpectrometryEd. A. Knowles & C. Burgess 1984

4. UV Spectroscopy; Techniques, instrumentation, data handlingEd. B.J. Clark, T. Frost & M.A. Russell, 1993

5. Standards and Best Practice in Absorption SpectrometryEd. C. Burgess & T. Frost, 1999

Page 28: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 28

UVSG/CORM conferences

1. Advances in Standards and Methodology in SpectrophotometryEd. C. Burgess & K.D. Mielenz, 1987

2. Spectrophotometry, Luminescence and Colour;Science and ComplianceEd. C. Burgess & D.G. Jones, 1995

3. Optical Spectrometry: Applications and Instrumentation into the 21st Century1998

Page 29: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 29

Spectrophotometry in a regulated environment

Glaxo Laboratories Ltd 1974Monitoring the performance of UV-visible spectrophotometers;UV Spectrometry Group Bulletin 5 77, 1997

Control chartsStandards for accuracy and precision

Ho(III) & Sm(III)Thompson’s solutionSap Green solutions

Page 30: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 30

Ho(III) as a wavelength standard

Page 31: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 31

Qualification and calibration of

spectrophotometers

Basis for measurementWavelength scale

Operational rangeAccuracy & precision

Photometric scaleOperational rangeAccuracy & precision (Linearity)Noise and drift

ResolutionStray light

Page 32: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 32

Traceable standards

Page 33: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 33

Starna RM-RE

Page 34: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 34

Long term photometric stability

How reliable is the instrument over its working life?How stable is the standard?How stable is the standardising laboratory?

Page 35: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 35

HP 8450 life time photometric accuracy

• NPL calibratedneutral density filter

• 546nm checked daily• Nine years

(110 months) • 2161 data points• Monthly averages• 95% CLs

Page 36: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 36

Potassium dichromate

• 2 maxima/minimapairs

• One close to thevisible

• One in ‘bomb alley’ of UV region

• Easy to prepare(60mg in 0.005Msulphuric acid)

Page 37: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 37

HP 8450 photometric accuracy

• acid dichromate• Four years

(47 months) • Duplicates• Satisfactory

performance• Reading slightly low at

350nm• No correlation max/min

Page 38: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 38

HP 8450 photometric accuracy

• acid dichromate• Four years

(47 months) • Duplicates• Less satisfactory

performance in UV• Correlation max/min

Page 39: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 39

HP 8450 photometric accuracy

• acid dichromate• Four years

(47 months) • Duplicates• Maxima to maxima• No correlation

Page 40: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 40

Technology transfer standards

StableNon hazardousUV & Visible rangeEasy to use

Page 41: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 41

Sap Green 1979

Page 42: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 42

Sap Green Linearity

Page 43: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 43

Comparison of scanning and DAS spectrophotometers 1990

7 DAS (HP8450 & HP8452)6 Scanning instruments (PE Lambda 5, 6 and 15)All labs had the ‘same’ solution of Sap Green

Page 44: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 44

Collaborative trials with Sap Green

Page 45: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 45

Collaborative trials with Sap Green

Page 46: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 46

Spectrophotometric methods

The instrument is OK but what about the method?

Optimum conditionsBest precisionBest accuracy

Page 47: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 47

Precision and noise

10

0.4343log

however is a function of

c Tc

T

s sRSDc T Ts T

= =

The relative standard deviation of concentration

Three main sources;1. Electronics, which is a constant (Johnson Noise)

2. Detector, photon statistics (Shot noise)

3. Cuvette positioning errors and source flicker

1Ts k=

22Ts k T T= +

3Ts k T=

Page 48: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 48

Precision and noise

Page 49: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 49

Relative errors;HP 8450 diode array spectrometer

Page 50: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 50

Accurate values of absorbance

Setting standard valuesMinimising biasLess out of specification resultsForgotten lessons

Ringbom-Ayres plotsVFG method

Page 51: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 51

Vandenbelt, Forsyth and GarrettInd. Eng. Chem. (Analytical Edition) 1945

• variation of molar extinction coefficient with absorbance

• Impact on standard values

• Beckman DU working absorbance range 0.5 to 1.9 for broad bands

Page 52: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 52

VFG plot

Page 53: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 53

In conclusion

Spectrophotometry is alive if not wellStandards are becoming more importantTrust your instruments and your methods BUT always verifyThe journey continues…

Page 54: Best practices in spectrophotometric measurement

© Burgess Consultancy 2003Web Site: www.burgessconsultancy.comeMail: [email protected] 54

NIR wavelength for the future?