AK 200 S and AK 200 ULTRA S systemsSpeci cations
-
AK 200 S
Speci cations subject to change without notice. For further information and operating instructions, please refer to the operators manual.
Quick preparationAutomatic self-testPre-settable treatment parameters
Treatment exibilityPre-settable acetate/bicarbonate uid compositionDirect setting of Na + and HCO3-
Adjustable dialysis uid ow Pro ling of Na +, HCO3- and UFSingle needle/Single pump with low recirculationSingle needle/Double pump with low recirculation Isolated UFHemodia ltration BiCart Select
Blood Volume Sensor
Blood ow controlFlow rate, double needle:0 and 20 to 500 ml/minFlow rate, single needle, single pump: 0 and 20 to 500 ml/min, time or pressure-controlledMean ow, single-needle, double pump: 0 and 10 to 400 ml/min, stroke volume controlled
Blood pressure supervisionArterial pressure: -700 to +750 mmHGSystem pressure: -700 to +750 mmHGVenous pressure: -700 to +750 mmHG
Air detection (Level detector)Method: Ultrasonic detectorDrip chamber size: 22 mm diameter
Heparin administrationFlow rate: 0 to 10 ml/h, programmable stop time, accumulated volume read-out
Water supplyInlet pressure: 100 to 800 kPa (1 to 8 bar)Inlet temperature: 5 to 30CQuality: Fluid should comply with national & international standards, e.g. ANSI/AAMI
Optional modesPro ling: Sodium, bicarbonate and UFHemodia ltration: Single or double-needle
Dialysis uid preparation and monitoringTemperature: Adjustable between 30 to 39 CFlow rate: 300 to 700 ml/minAcetate mode: Na+ range 115 to 160 mmol/lBicarbonate mode: Na+ range 130 to 150 mmol/l HCO3- range 20 to 40 mmol/lpH meter: Range 1.0 to 9.9
Substitution uidSubstitution ow rate: 1.0 to 2.0 l/h 10%
Ultra ltration controlUF volume: Adjustable, 0 to 30 lAccuracy: 50 ml/h or 2.5%UF rate: 0 to 4 l/h
Blood leakage detectionInfrared light
Disinfection and cleaningChemical: Peracetic acid, hypochlorite or formaldehydeHeat: Heat, liquid citric acid or CleanCart
Rinse/drain/decalci cation: Rinse, citric acid or CleanCart C
Power supplyMains voltage: 115/230V AC ( 10%)Frequency: 50 to 60 Hz ( 5Hz)
Power consumptionMax 2250 W at 230 V or 1650 W at 115 VFully prepared for battery backup
Connection of external equipmentConnectors: RS 232 C or RS 422
Dimensions and weightWeight: 81 to 88 kgWidth: Machine 570 mm, stand 573 mmDepth: Machine 610 mm, stand 753 mmHeight: 1152 to 1382 mm in 5 xed positions or 1175 to 1340 mm adjustable
Operating environmentAmbient temperature: 18 to 35 CRelative humidity: 15 to 85%Approx. 2000 meters above sea level
SafetyThe AK 200 S system complies with the following standards:
IEC 60601-1 General requirements for safety, Class 1, type B
IEC 60601-2-16 Particular requirements for safety of haemodialysis, haemodia - ltration and haemo ltration equipment
IEC 60601-2-30 Particular requirements for the safety of automatic cycling indirect blood pressure monitoring equipment
IEC 60601-1-2 Electromagnetic compatibility
CE 0086 This product is CE-marked in accordance with the requirements in EC Council Directive 93/42/EEC of 14 June, 1993, concerning medical devices
HygieneChemical and heat disinfectionautomatic and integratedCleanCart cleaning, decalci cation, disinfectionPre-settable disinfection options, chemical disinfection memoryAutomatic disinfection cycle veri cation
OptionspH meterUltra ltered dialysis uidBattery backupElectrical adjustable standOne or two blood pumpsBlood pressure meterDiascan
Specifications
Features
-
AK 200 ULTRA S
Speci cations subject to change without notice. For further information and operating instructions, please refer to the operators manual.
Specifications
Quick preparationAutomatic self-testPre-settable treatment parameters
Treatment exibilityOnline HDF with pre- or post-dilution Online HF with pre- or post-dilution All modes with ultra-pure dialysis uid Online preparation of substitution uid for HF and HDFAdjustable dialysis uid ow 300-700 ml/minPro ling of Na +, HCO3- and UFDesigned for the BiCart and BiCart Select system SN/SP and SN/DP with minimum recirculationBlood Volume Sensor
QuicAutoPre-
TreaO On Al On foAd 30P D BSN reB D
HygieneChemical and heat disinfectionautomatic and integratedCleanCart cleaning, decalci cation, disinfectionChemical disinfection memory
OptionspH meterBattery backupElectrical adjustable standBlood pressure meterDiascan
Features
Blood ow controlFlow rate, double needle: 0 and 20 to 500 ml/min (HD) 0 and 20 to 700 ml/min (HDF/HF)Flow rate, single needle, single pump: 0 and 20 to 500 ml/min, time- or pressure-controlledMean ow, single-needle, double pump: 0 and 10 to 400 ml/min, stroke volume controlled
Blood pressure supervisionArterial pressure: -700 to +750 mmHGSystem pressure: -700 to +750 mmHGVenous pressure: -700 to +750 mmHG
Air detection (Level detector)Method: Ultrasonic detectorDrip chamber size: 22 mm diameter
Heparin administrationFlow rate: 0 to 10 ml/h, programmable stop time, accumulated volume read-out
Water supplyInlet pressure: 120 to 800 kPa (1 to 8 bar)Inlet temperature: 5 to 30CQuality: Fluid should comply with national & international standards, e.g. ANSI/AAMI
Optional modesPro ling: Sodium, bicarbonate and UFHemodia ltration: Single or double-needle
Dialysis uid preparation and monitoringTemperature: Adjustable between 30 to 39 CFlow rate: 300 to 700 ml/minAcetate mode: Na+ range 115 to 160 mmol/lBicarbonate mode: Na+ range 130 to 150 mmol/l, HCO3- 20 to 40 mmol/lpH meter: Range 1.0 to 9.9
Substitution uidSubstitution ow rate: 0.3 to 27 l/hSubstitution uid volume: 0 to 150 litres
Ultra ltration controlUF volume: Adjustable, 0 to 30 lAccuracy: 50ml/h or 2.5%UF rate: 0 to 4 l/h
Blood leakage detectionInfrared light
Disinfection and cleaningChemical: Peracetic acid, hypochlorite or formaldehydeHeat: Heat, liquid citric acid or CleanCart Rinse/drain/decalci cation: Rinse, citric acid or CleanCart C
Power supplyMains voltage: 115/230V AC ( 10%)Frequency: 50 to 60 Hz ( 5Hz)
Power consumption:Max 2250 W at 230 V Fully prepared for battery backup
Connection of external equipmentConnectors: RS 232 C or RS 422
Dimensions and weightWeight: 81 to 88 kgWidth: Machine 570 mm, stand 573 mmDepth: Machine 610 mm, stand 753 mmHeight: 1152 to 1382 mm in 5 xed positions or 1175 to 1340 mm adjustable
Operating environmentAmbient temperature: 18 to 35 CRelative humidity: 15 to 85%Approx. 2000 meters above sea level
SafetyThe AK 200 ULTRA S system complies with the following standards:
IEC 60601-1 General requirements for safety, Class 1, type B
IEC 60601-2-16 Particular requirements for safety of hemodialysis, hemodia ltration and hemo ltration equipment
IEC 60601-2-30 Particular requirements for the safety of automatic cycling indirect blood pressure monitoring equipment
IEC 60601-1-2 Electromagnetic compatibility
CE 0086 This product is CE-marked in accordance with the requirements in EC Council Directive 93/42/EEC of 14 June, 1993, concerning medical devices
-
Gambro Lundia AB
Box 10101, 220 10 Lund
Sweden
Tel +46 46 16 90 00
Fax + 46 46 16 96 97
www.gambro.com
HC
EN
2391
_4
200
9.04
. Gam
bro
Lund
ia A
BAK 200 ULTRA S is a trademark of Gambro Lundia AB, registered in the European Union.