abx microses 60 - visoleonbase.visole.net/base/abx_micros_es_60_-2017-05-02-14... · 2017. 5....

2
ABX Micros ES 60 60 samples per hour (OT Version) CBC + 3 DIFF (18 parameters) Large color touch screen Comprehensive quality control program (optional external printer in closed tube and open tube versions) Integrated ticket printer . Full data print-out . PLT, RBC, and WBC curves (optional bottle version) Unique pack system . Innovative ABX patent . Ready to use . Optimized packaging . Simplify your lab organisation *Available in closed tube version (optional external barcode reader for open tube version) Integrated barcode reader* Secured identification (patient, QC and reagent lot numbers) USB connection . Archiving patients and QC results . Downloading QC target values A new generaon hematology analyzer based on the Micr os concept, universally r ecognized for its r eliability, r obustness and high quality r esults > ABX Micro-sampling Only 10μl of whole blood per analysis > ABX Liquid valves Precise volume and reliability > ABX stepper motor vacuum pump Noiseless running (no-compressor) ABX Micros Concept “Zero maintenance” system

Upload: others

Post on 11-Feb-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

  • ABX Micros ES 60

    • 60 samples per hour (OT Version)

    • CBC + 3 DIFF (18 parameters)

    • Large color touch screen

    • Comprehensive quality control program

    (optional external printer in closed tube and open tube versions)

    Integrated ticket printer. Full data print-out. PLT, RBC, and WBC curves

    (optional bottle version)

    Unique pack system. Innovative ABX patent. Ready to use . Optimized packaging . Simplify your lab organisation

    *Available in closed tube version (optional external barcode reader for open tube version)

    Integrated barcode reader*Secured identification (patient,QC and reagent lot numbers)

    USB connection. Archiving patients and QC results . Downloading QC target values

    A new generation hematology analyzer basedon the Micros concept, universally recognizedfor its reliability, robustness and high qualityresults

    > ABX Micro-samplingOnly 10µl of whole blood per analysis

    >

    ABX Liquid valvesPrecise volume and reliability

    >

    ABX stepper motor vacuum pumpNoiseless running (no-compressor)

    ABX Micros Concept“Zero maintenance” system

  • More Powerful, More Intuitive, More Secured, More Connectivity... The perfect combination of intuitive use and innovative technologies

    • User friendly interface with rapid access to all menus and virtual keyboard• User profile management with different access levels• Flexible data exchange for patient results and quality control• Multi-connection capability (network and peripherals)

    Full ConnectivityIntegrated Technologies

    > SoftwareJavaTM, Linux Operating System, HTML

    > Hardware: 7 connections portsRS232 (2), USB (3), PS2 (1), RJ45 (1)

    > ABX and ASTM protocols available Barcode reader

    ML

    Comprehensive quality control program3 simultaneous activated blood control levels

    Levey-Jennings graphics XB management

    Upload / Download control information Complete traceability in compliance withaccreditation standards

    (users, reagents, controls, maintenance...)

    Automatic system monitoringReal-time instrument status with help messages

    Automatic reagent test count down Automatic daily maintenance and operational alerts

    Customized reporting format

    Easy interpretation Control all the parameters from the main screen Access to detailed results (normal ranges & curves)

    1000 patients results capacityUnlimited patient report archiving

    METHODS & TECHNOLOGIES

    RBC/PLT Detection PrinciplesMethod ImpedanceRuby diameter 50 µmDepression of count 200 mbDuration of count 2 (or 3) x 6 secondsDilution ratio 1/15 000

    HGB MeasurementMethod PhotometryWavelength 550 nmDilution ratio 1/250

    HCT MeasurementMethod Numeric integration

    WBC MeasurementMethod ImpedanceRuby diameter 80 µmDepression of count 200 mbDuration of count 2 (or 3) x 6 secondsDilution ratio 1/260

    3 Part DifferentiationMethod Impedance + specific lyse action

    MCV, MCH, MCHC, RDW, PCT*, PDW*Calculation

    SPECIFICATIONS

    Throughput*:OT: 60 samples/hourCT: 50 samples/hourSound Pressure Level:< 60 dBaOperating Temperature & Humidity:16 - 30°C (61- 86°F) room temperature.Max relative humidity 85% up to 30°CSpecimen Volume:CBC + 3 part DIFF 10 µLReagents:- ABX Minipack LMG- ABX Minidil LMG- ABX Minilyse LMG- ABX Miniclean- ABX Alphalyse 360 (0,36L)

    PARAMETERS & PERFORMANCE DATA

    18 Parameters:WBC RBC MCH PLTLYM# & LYM% HGB MCHC MPV MON# & MON% HCT RDW PCT* GRA# & GRA% MCV PDW*

    Linearity:Parameter Linearity Limit Visible RangeWBC 0 - 100 x 103/mm3 100 - 150 x 103/mm3

    RBC 0 - 8 x 106/mm3 8 - 18 x 106/mm3

    HGB 0 - 26 g/dL 26 - 30 g/dLHCT 0 - 80 % 80 - 90%PLT (Whole blood) 0 - 2200 x 103/mm3 2200 - 6000 x 103/mm3

    PLT (PLT concentrate) 0 - 4000 x 103/mm3 4000 - 6000 x 103/mm3

    CERTIFICATIONEN 61326: 2001EN 61326-2-6: 2006IEC 61010-1: 2001IEC 61010-2-81: 2001 / A1: 2003IEC 61010-2-101: 2002UL61010-1: 2004C22.2 N°61010-1: 2004In Vitro Diagnostics Directive: 98/79/ECcTUVus mark

    SOFTWARE8.4” LCD screen with Inverter Card256,000 colors, 640 x 480 pixel resolutionTouch screen technologySecure digital memory card: 512 Mb58mm thermal printer with rapid paper loading

    ABX MicrosES60Hematology Analyzer