Page 1: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

A Guide To Using

Model 1502 Torqo

By Mesa Laboratories, Inc.

Page 2: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Contents

Overview 7

Introduction ...............................................................................................................................7 What’s Inside.............................................................................................................................7 About This Manual ....................................................................................................................8

Chapter 1: Unpacking and Inspection.........................................................................8 Chapter 2: System Power Up......................................................................................8 Chapter 3: Bottle and Cap Tooling.............................................................................8 Chapter 4: Understanding the Control Panel ..............................................................8 Chapter 5: Quick Start ................................................................................................8 Chapter 6: Test Modes................................................................................................8 Chapter 7: Programming a Removal Test ..................................................................8 Chapter 8: Programming an Incremental Test............................................................9 Chapter 9: Programming a Bridge Test ......................................................................9 Chapter 10: Programming the Batch Mode ................................................................9 Chapter 11: Testing in Batch Mode............................................................................9 Chapter 12: Performing a Strip Torque Test ..............................................................9 Chapter 13: Performing a Life Test ............................................................................9 Chapter 14: Retesting a Sample..................................................................................9 Chapter 15: Logging Test Results ..............................................................................9 Chapter 16: Transmitting Test Results .....................................................................10 Chapter 17: Using Machine Numbers ......................................................................10 Chapter 18: Clearing Memory..................................................................................10 Chapter 19: Using Lot Numbers...............................................................................10 Chapter 20: Using Specification Limits ...................................................................10 Chapter 21: Using the Torqo Printer ........................................................................10 Chapter 22: Setting the Date and Time.....................................................................10 Chapter 23: Programming the Min Torque .............................................................10 Chapter 24: Gold Bottle Calibration Check..............................................................11 Chapter 25: System Calibration................................................................................11 Chapter 26: Child Resistant Cap Option ..................................................................11 Chapter 27: Key lock Option....................................................................................11 Chapter 28: Multi Speed Option...............................................................................11 Chapter 29: Error Messages .....................................................................................11 Chapter 30: Trouble Shooting ..................................................................................11 Chapter 31: Maintenance and Cleaning....................................................................11

Conventions .............................................................................................................................12

Unpacking And Inspection 13

What’s Inside...........................................................................................................................13 Unpacking and Inspection .......................................................................................................13 Checklist for Items supplied with the system ..........................................................................14

System Power Up 16

A Guide To Using Model 1502 Torqo Contents � i

Page 3: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

What’s Inside...........................................................................................................................16 Connecting the Drive Tower to the Readout ...........................................................................16 Powering the Readout Instrument............................................................................................17 Powering the Drive Tower.......................................................................................................17 Initializing The System............................................................................................................18 Powering the Optional Printer .................................................................................................19

Bottle and Cap Tooling 20

What’s Inside...........................................................................................................................20 Installing Tooling For Holding The Bottle ..............................................................................20 Installing Tooling For Griping The Cap ..................................................................................21 About the 1506A Beverage Bottle Clamp ...............................................................................22

Specifications ............................................................................................................22 1506A Beverage Bottle Clamp Nomenclature ..........................................................22

Adjusting the 1506A Beverage Bottle Clamp..........................................................................23 About The 1614 Universal Vise ..............................................................................................25

Specifications ............................................................................................................25 Vise Jaws For Rectangular Bottles ............................................................................26

Adjusting The 1614 Universal Vise.........................................................................................26 About The Chuck Height Adjuster ..........................................................................................27 Positioning The Chuck With The Height Adjuster ..................................................................27

Understanding the Control Panel 28

What's Inside............................................................................................................................28 Control Panel Nomenclature....................................................................................................28

Torque Display ..........................................................................................................29 LED Indicators ..........................................................................................................29 Blinking Metric LED.................................................................................................30 Keypad ......................................................................................................................30

Quick Start 31

What’s Inside...........................................................................................................................31 Performing A Test ...................................................................................................................31

Zeroing the Display ...................................................................................................31

Test Modes 34

What’s Inside...........................................................................................................................34 Description Of The Testing Modes..........................................................................................34

Removal Torque Test Mode......................................................................................35 Removal & Incremental Test Mode ..........................................................................35 Removal & Bridge Test Mode...................................................................................35 Reverse Ratchet Test Mode.......................................................................................35 Batch Test Mode .......................................................................................................36

Selecting A Test Mode ............................................................................................................36 Can not Change Modes .............................................................................................36

Programming a Removal Test 38

What’s Inside...........................................................................................................................38 Test Features............................................................................................................................38 Setting the Re-application Torque ...........................................................................................39 Programming to Leave the Cap Loose.....................................................................................39

A Guide To Using Model 1502 Torqo Contents � ii

Page 4: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming an Incremental Test 41

What’s Inside...........................................................................................................................41 Test Features............................................................................................................................41 Programming the Degrees to Drive .........................................................................................42

Programming a Bridge Test 43

What’s Inside...........................................................................................................................43 Test Features............................................................................................................................43 Programming the Degrees to Drive .........................................................................................44

Programming the Batch Mode 45

What’s Inside...........................................................................................................................45 Test Features............................................................................................................................45 Selecting Batch Mode..............................................................................................................46 Programming the Number of Samples in a Batch....................................................................47 Retesting a Sample in Batch Mode..........................................................................................47 Programming a Skip in Batch Mode........................................................................................48 Erasing Data from Memory in Batch Mode.............................................................................48

Testing in Batch Mode 49

What’s Inside...........................................................................................................................49 Using Batch Mode ...................................................................................................................49 Exiting Batch Mode.................................................................................................................54 Starting a New Batch Project...................................................................................................54

Performing a Strip Torque Test 55

What's Inside............................................................................................................................55 Strip Torque Test Description.................................................................................................. 55 Programming a Strip Torque Test............................................................................................55

Performing a Life Test 57

What's Inside............................................................................................................................57 Life Test Description ...............................................................................................................57 Programming a Life Test .........................................................................................................57

Retesting a Sample 59

What's Inside............................................................................................................................59 Retesting a Sample ..................................................................................................................59

Logging Test Results 60

What’s Inside...........................................................................................................................60 Manually Saving Data .............................................................................................................60 Automatically Saving Data...................................................................................................... 61

Transmitting Test Results 62

What’s Inside...........................................................................................................................62 Output Port ..............................................................................................................................62 Connector Requirements..........................................................................................................62

Mating Connectors ....................................................................................................62

A Guide To Using Model 1502 Torqo Contents � iii

Page 5: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Pin Out.......................................................................................................................63 Communication Protocol .........................................................................................................63 Data Format Information .........................................................................................................63 Transmitting Data Automatically ............................................................................................65

Transmitting End of Test Results Automatically ......................................................65 Transmitting and Saving Automatically....................................................................65 Transmitting Live Data to a PC.................................................................................66 Transmitting Live Data to a PC and Saving ..............................................................66 Transmitting Automatically in Batch Mode ..............................................................67

Transmitting Manually ............................................................................................................67

Using Machine Numbers 68

What’s Inside...........................................................................................................................68 Machine Number Description..................................................................................................68 Entering Machine Numbers .....................................................................................................69

Inputting a Machine Number.....................................................................................69 Printing a List of the Machine Numbers in use .........................................................70

Clearing Memory 71

What's Inside............................................................................................................................71 Erasing Data for a Specific Machine .......................................................................................71 Erasing all Memory .................................................................................................................71

Using Lot Numbers 72

What's Inside............................................................................................................................72 Lot Number Description ..........................................................................................................72 Entering a Lot Number ............................................................................................................73

Keying in the Lot Number.........................................................................................73 Deleting a Lot Number..............................................................................................73

Using Specification Limits 74

What's Inside............................................................................................................................74 Limits Description ...................................................................................................................74 Entering High and Low Limits ................................................................................................75

Using The Torqo Printer 77

What's Inside............................................................................................................................77 Print Features ...........................................................................................................................77 Sample Printout .......................................................................................................................78 Printing Test Results................................................................................................................79 Printing the Current Machine Configuration ...........................................................................79 Printing the Current Machines Report .....................................................................................81

Viewing Test Results 82

What's Inside............................................................................................................................82 Viewing End of Test Results ...................................................................................................82 View Summary Test Results....................................................................................................82

Standard Deviation Formula......................................................................................83

Setting the Date and Time 84

A Guide To Using Model 1502 Torqo Contents � iv

Page 6: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

What's Inside............................................................................................................................84 Changing the Date and Time ...................................................................................................84

Programming the Min Torque 86

What's Inside............................................................................................................................86 Min Torque Description...........................................................................................................86 Inputting a Min Torque............................................................................................................87

Gold Bottle Calibration Check 88

What's Inside............................................................................................................................88 Gold Bottle Description ...........................................................................................................88 Using the Gold Bottle ..............................................................................................................89

System Calibration 90

What's Inside............................................................................................................................90 General Calibration Information..............................................................................................90

Calibration Frequency ...............................................................................................90 Required Calibration Equipment .............................................................................................90

Calibration Beam.......................................................................................................91 Calibration Fixture ....................................................................................................91 Calibration Weight Set ..............................................................................................91 Calibration Weight Chart ..........................................................................................92

Selecting a Calibration to be Used...........................................................................................92 Performing a Calibration Verification .....................................................................................93

Installing the Calibration Fixture...............................................................................93 Installing the Calibration Beam.................................................................................94 Creating a Calibration Verification Chart..................................................................94 Calibration Verification Chart ...................................................................................95

Performing a User Calibration.................................................................................................98 Calibration Methodology...........................................................................................98 Entering the Calibration Mode ..................................................................................98 Erasing all Previous User Calibrations ......................................................................99 Zeroing the Torqo in the Calibration Mode...............................................................99 Entering Calibration Torques ..................................................................................100 System Report .........................................................................................................105

Child Resistant Cap Option 107

What's Inside..........................................................................................................................107 About the Child Resistant Option ..........................................................................................107 Connecting the Air Supply ....................................................................................................107 Adjusting the Down Force.....................................................................................................108 Down Force Mechanism Nomenclature ................................................................................109 Adjusting the Down Force Foot.............................................................................................109 Operating the Child Resistant Mechanism.............................................................................110

Key lock Option 111

What's Inside..........................................................................................................................111 Key lock Option Description .................................................................................................111 Using the Key lock Option ....................................................................................................111

Multi-Speed Option 112

A Guide To Using Model 1502 Torqo Contents � v

Page 7: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

What's Inside..........................................................................................................................112 Multi-Speed Option Description............................................................................................112 Using the Multi-Speed Option...............................................................................................113

Error Messages 114

What's Inside..........................................................................................................................114 Description of Error Messages ..............................................................................................114

Trouble Shooting 118

What's Inside..........................................................................................................................118 Contacting Mesa Labs............................................................................................................118 Trouble Shooting the System.................................................................................................118 Restoring the Factory Defaults ..............................................................................................119

Performing a Soft Reset...........................................................................................119 Re-initializing the System......................................................................................................119

Blinking LED's ........................................................................................................119 Performing a Hard Reset .........................................................................................119

Erratic System Behavior ........................................................................................................120 L's on the Display ....................................................................................................120 Checking the Memory Chip ....................................................................................120

Torqo Drive Problems ...........................................................................................................121 Checking the Power.................................................................................................121 Checking the Fuse ...................................................................................................121 Checking the Drive Belt ..........................................................................................122 Replacing the Belt ...................................................................................................122

Determining the Belt Size......................................................................................................123 Checking the DC Power ........................................................................................................123 Checking for the Clock Pulses...............................................................................................123

Maintenance and Cleaning 125

What's Inside..........................................................................................................................125 General Information...............................................................................................................125 General Maintenance .............................................................................................................125 Cleaning the Tooling .............................................................................................................126

Glossary of Terms 127

Index 128

A Guide To Using Model 1502 Torqo Contents � vi

Page 8: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Overview

Introduction

Mesa Labs is an industry leader in the manufacture of precision

torque measuring equipment. Over forty years of experience

in the torque-testing field is reflected in every automated

system.

Mesa’s commitment to supplying state of the art products

has resulted in the development of the Torqo and its

patented measuring system.

The Model 1502 Torqo with minimal maintenance will

provide the user with years of trouble free service.

What’s Inside

This manual is designed to provide the user with a step-by-

step procedure from receipt of the system through operation,

maintenance and calibration.

In addition to this manual, a set of quick reference cards is

supplied with each Torqo.

A Guide To Using Model 1502 Torqo Overview � 7

Page 9: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

About This Manual

For convenience, this manual is divided into a number of

chapters. A brief description of each chapter follows.

Chapter 1: Unpacking and Inspection

This chapter is a guide to insure that the Torqo is received in

proper condition and that all necessary components have

been located.

Chapter 2: System Power Up

This chapter contains the required information for

electrically connecting the system components and then

powering up the system.

Chapter 3: Bottle and Cap Tooling

This chapter contains the required information for installing

and adjusting the tooling that will be used for testing.

Chapter 4: Understanding the Control Panel

This chapter describes the Front Panel Display, Indicator

LEDs and Keypad.

Chapter 5: Quick Start

This chapter contains a step-by-step procedure for

performing a test.

Chapter 6: Test Modes

This chapter describes the available test modes that can be

selected and a step-by-step procedure for selecting one.

Chapter 7: Programming a Removal Test

This chapter describes the features that are applicable to a

removal torque test and a button-by-button procedure for

programming these features.

A Guide To Using Model 1502 Torqo Overview � 8

Page 10: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Chapter 8: Programming an Incremental Test

This chapter describes the features that are applicable to an

Incremental torque test and a button-by-button procedure for

programming these features.

Chapter 9: Programming a Bridge Test

This chapter describes the features that are applicable to a

Bridge torque test and a button-by-button procedure for

programming these features.

Chapter 10: Programming the Batch Mode

This chapter describes the features that are applicable to the

Batch Mode and a button-by-button procedure for

programming these features.

Chapter 11: Testing in Batch Mode

This chapter provides a button-by-button, step-by-step

description of how to test in Batch Mode.

Chapter 12: Performing a Strip Torque Test

This chapter provides a button-by-button description of how

to program and run a strip torque test.

Chapter 13: Performing a Life Test

This chapter provides a description and step-by-step

procedure for programming and performing a life test.

Chapter 14: Retesting a Sample

This chapter describes the procedure for retesting a sample

and storing the test results in an existing file location.

Chapter 15: Logging Test Results

This chapter provides a description and the step-by-step

procedure for logging test results.

A Guide To Using Model 1502 Torqo Overview � 9

Page 11: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Chapter 16: Transmitting Test Results

This chapter provides a description of the output port,

connector requirements, communication protocol and

procedure for transmitting test results both manually and

automatically.

Chapter 17: Using Machine Numbers

This chapter describes the use of multiple machine numbers

for storing both test setups and test results in memory.

Chapter 18: Clearing Memory

This chapter describes in detail the procedure for erasing

data from memory.

Chapter 19: Using Lot Numbers

This chapter describes the procedure for inputting and

deleting lot numbers.

Chapter 20: Using Specification Limits

This chapter describes the procedure for inputting user

provided pass / fail limits.

Chapter 21: Using the Torqo Printer

This chapter describes the many features that the printer

provides and the print commands for using these features.

Chapter 22: Setting the Date and Time

This chapter describes the procedure for setting the date and

time.

Chapter 23: Programming the Min Torque

This chapter provides a button-by-button procedure for

programming the min torque.

A Guide To Using Model 1502 Torqo Overview � 10

Page 12: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Chapter 24: Gold Bottle Calibration Check

This chapter describes the Gold Bottle and the procedure for

using it to verify the calibration of the Torqo.

Chapter 25: System Calibration

This chapter is intended to acquaint the user with the

required Calibration Equipment, the Calibration Verification

Process, the Process for performing a User Calibration and

the procedure for Selecting the calibration to be used.

Chapter 26: Child Resistant Cap Option

This section is indented to acquaint the user with the

features and use of this option.

Chapter 27: Key lock Option

This chapter describes the features and use of the key lock

option.

Chapter 28: Multi Speed Option

This chapter describes the use of the Multi Speed Option.

Chapter 29: Error Messages

This chapter describes the Error Messages that may be

displayed when programming and testing.

Chapter 30: Trouble Shooting

This chapter is provided to help the user solve some of the

minor problems that may occur with a Torqo.

Chapter 31: Maintenance and Cleaning

This chapter describes the recommended maintenance that

should be performed on the Torqo.

A Guide To Using Model 1502 Torqo Overview � 11

Page 13: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Conventions

The following conventions are used in this manual.

A Guide To Using Model 1502 Torqo Overview � 12

Page 14: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Unpacking And Inspection

What’s Inside

This chapter provides the user with a checklist for the

receipt of a new Model 1502 Torqo.

Unpacking and Inspection

Your Torque meter was carefully inspected, both electrically

and mechanically before shipment. Upon receiving this

system, carefully unpack all items from the shipping

container and check for any signs of damage that may have

occurred during shipment.

Immediately report any shipping damage to the shipping agent

Retain and use the original packing materials in case re-

shipment is necessary.

A Guide To Using Model 1502 Torqo Unpacking And Inspection � 13

Page 15: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Checklist for Items supplied with the system

The following items are shipped with all Torqo's.

AC adaptor with attached cable and jack.

Input power cable for drive tower.

Interface cable for connecting the control instrument to the

drive tower.

Quick reference cards.

A Guide To Using Model 1502 Torqo Unpacking And Inspection � 14

Page 16: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Operator's manual.

When calibration equipment is purchased with the system,

the certificates of calibration for the beam and weights are

included behind the front cover of the manual.

A Certificate of Calibration for your new Torqo is included

behind the front cover of the manual.

A Guide To Using Model 1502 Torqo Unpacking And Inspection � 15

Page 17: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

System Power Up

What’s Inside

This chapter contains the required information for powering

up the Model 1502 Torqo.

Connecting the Drive Tower to the Readout

Plug the Interface Cable into the socket, marked Drive,

on the rear panel of the control instrument, and then plug the

other end of the cable into the socket, marked Drive on the

rear panel of the drive tower.

This cable is not direction sensitive.

A Guide To Using Model 1502 Torqo System Power Up � 16

Page 18: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Powering the Readout Instrument

The Readout is energized by first inserting the Jack on the

end of the AC adaptor cable into the socket, marked Power,

on the rear panel of the instrument and then plugging the

transformer into an AC outlet.

Now by actuating the switch marked ON/OFF on this panel

the system will turn on.

The front panel torque display should now be lit.

Powering the Drive Tower

The drive is energized by first inserting the Input Power Cable into the receptacle, marked AC INPUT, on the rear

panel of the drive tower and then plugging the other end of

the cable into an AC Outlet.

Now by actuating the switch on this panel the drive will be

on and ready for use.

A Guide To Using Model 1502 Torqo System Power Up � 17

Page 19: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

The Torqo is supplied for use with 110 VAC 60 HZ.

Consult the factory or your local sales representative for

operation with other voltages.

The use of a surge protector with the Torqo is recommended as a protective measure against electrical noise.

It can also serve as a convenient way to turn the system on

and off with a single switch.

Initializing The System

This step is only necessary when the Torqo is powered-up for the first time after shipment.

Now with the Torqo turned on press

[ Total ] then [ Mode ]

This will initialize the system in the factory default mode

and apply the factory default settings.

Now press

[ Total ] then [ Clear ]

This will clear all previously stored data from memory.

Completing the steps above will initialize the Torqo and

leave it in the Ready State.

A Guide To Using Model 1502 Torqo System Power Up � 18

Page 20: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Powering the Optional Printer

The printer should be powered up in accordance with the

manual that is supplied with the printer.

All Mesa supplied printers have been pre-tested with a

Torqo for functionality.

An interface cable for connecting the printer to the Torqo is

supplied when this option is purchased.

This cable should be attached to the printer and then

plugged into the connector marked RS-232 on the rear

panel of the instrument.

A Guide To Using Model 1502 Torqo System Power Up � 19

Page 21: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Bottle and Cap Tooling

What’s Inside

This chapter contains the required information for installing

and adjusting the tooling that will be used for holding the

bottle, griping the cap and positioning the chuck.

The tooling pictured here may not necessarily be the tooling

shipped with your system however the descriptions for

installing this tooling are still applicable for most other

tooling packages.

Installing Tooling For Holding The Bottle

Screws are provided for holding the tooling in place.

All Mesa supplied tooling packages are designed to be mounted on top of the Spindle Cap shown in this picture.

(Qty 2 x .250-20 x 1" Skt Hd Cap Screws)

(Wrench size 3/16 " Hex.)

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 20

Page 22: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

The Model 1614 Universal Pin vise is secured to the Torqo

by carefully centering it on the shaft that protrudes through

the spindle cap and then installing and tightening the Cap

Screws that are supplied with the vise.

The Model 1506A Beverage Bottle clamp is secured to the

Torqo by carefully centering it on the shaft that protrudes

through the spindle cap and then installing and tightening

the Cap Screws that are supplied with the clamp.

Installing Tooling For Griping The Cap

All Mesa supplied cap chucks are mounted on the drive

shaft by tightening a clamp collar on the top of the chuck.

(Wrench size 5/32" Hex)

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 21

Page 23: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

About the 1506A Beverage Bottle Clamp

The Model 1506A was designed as a quick acting universal

holding device that can be easily adjusted to accommodate

many different size bottles.

Specifications

�� Max Bottle Diameter: 6.5 inches

�� Max Bottle Height: 13.5 inches

�� Minimum Bottle Height: 4.0 inches

1506A Beverage Bottle Clamp Nomenclature

Height Adjusting

Lock Knob

Ring Support

Plate

Clamp Handle

Rubber

Clamp

Ring

Rubber

Pad

A-21954

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 22

Page 24: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Adjusting the 1506A Beverage Bottle Clamp

This clamp is easily adjusted for a specific bottle size by

following the procedure described below.

Step 1:

Loosen the Height Adjusting Knobs and raise the Ring

Support Plate to a position greater than the height of the

bottle to be tested and slightly tighten one knob. Now place

the test sample on the base of the fixture.

Step 2:

Now loosen the knob, lower the Ring Support Plate to a

position slightly above the sample and slightly tighten one

knob again. Now engage the cap chuck to aid in centering

the bottle.

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 23

Page 25: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Step 3:

Now with both height-adjusting knobs loose, lower the Ring

Support Plate until the Rubber Clamp Ring is resting on the

bottle. Raise the 2 Clamp Handles so that they are at an

angle of approximately 15° off horizontal and then tighten

both Knobs.

15°

Step 4:

Push Clamp Handles down to secure a bottle and then raise,

to the latched position, for loading the next sample.

The clamping action should not distort the shape of the bottle.

If this occurs, loosen the Height Adjusting Knobs and reduce the 15° angle.

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 24

Page 26: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

About The 1614 Universal Vise

The Model 1614 Vise was designed as a universal holding

device that can be easily adjusted to accommodate both

round and square bottles.

Specifications

�� Min Bottle Diameter: .62 inches

�� Max Bottle Diameter: 6.0 inches

Pin Positioning Chart

The chart below shows the diameter range for each pin

position.

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 25

Page 27: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Vise Jaws For Rectangular Bottles

Special vise jaws are frequently supplied for rectangular and

unusual shaped bottles. These special jaws are required in

order to insure that the bottle is centered under the cap

chuck.

See example below.

Adjusting The 1614 Universal Vise

When pins are used for clamping the test sample, the vise

should be tightened just enough to hold the bottle but not so

tight as to distort the shape of the bottle.

By noting the position of the handle crank and returning to

this position when changing samples, the user can rapidly

load and unload the vise.

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 26

Page 28: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

About The Chuck Height Adjuster

The Height Adjuster was designed to position the cap

chuck at a convenient height for loading and unloading test

samples while providing for the minimum amount of

vertical chuck motion for the operator.

Positioning The Chuck With The Height Adjuster

Loosen the Height Adjuster Thumb Screw and position the

cap chuck approximately 1" above the bottle cap when using

the 1506A beverage clamp or 1" above the position that

allows the test bottle to be loaded into the 1614 vise or other

tooling systems. (See picture below)

Height

Adjuster

Thumb

Screw

Now between tests just lift the cap chuck and let the

magnetic height adjuster hold it in position

A Guide To Using Model 1502 Torqo Bottle and Cap Tooling � 27

Page 29: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Understanding the Control Panel

What's Inside

This chapter contains a description of the Control Panel,

LED Indicators, Keyboard and how they function in the

system.

Control Panel Nomenclature

The titles that are shown on the picture below will be used

throughout this manual.

LED Indicators

Keypad Torque Display

A Guide To Using Model 1502 Torqo Understanding the Control Panel � 28

Page 30: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Torque Display

This four-digit display is used to report test results and

provide a communication link between the user and the

software.

LED Indicators

The six LED indicators on the control panel are used as

status indicators for the system.

At a glance the user can determine what test mode is

selected, what unit of measure is selected and whether or not

the system is ready to perform a test.

The Removal, Removal & Incremental, Removal & Bridge,

Batch and Reverse Ratchet LED's serve only one purpose

and that is to indicate what has been selected in the Current Machine. (See Using Machine Numbers)

The Metric LED serves two purposes.

First it serves as an indicator when lit, that the unit of

measure is Metric not English.

Secondly it serves as an end of test status indicator.

A Guide To Using Model 1502 Torqo Understanding the Control Panel � 29

Page 31: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Blinking Metric LED

This LED will blink when a test has been completed

indicating that the data that is currently displayed must

either be saved or discarded before another test can be

performed. (See Logging Test Results)

Keypad

The Torqo keypad was designed with a two-color overlay to

make it as user friendly as possible.

It serves as the primary communication link between the

user and the system software.

The picture to the left shows the keypad with a typical three

function key high-lighted.

The number keys are used for inputting numerical values

when programming the Torqo.

The words printed in Black above the numbers are the

primary key functions and are used when the system is in the

Ready State.

The words printed in Red beneath the numbers are the

secondary key functions and are accessed by pressing the

[Mode] key first.

Do not use sharp objects to depress the keys on the Keypad.

Damage to the Keypad will occur.

A Guide To Using Model 1502 Torqo Understanding the Control Panel � 30

Page 32: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Quick Start

What’s Inside

This chapter contains a button-by-button, step-by-step

procedure to help the user become familiar with the

operation of the Torqo by actually performing a Removal

Torque Test.

Performing A Test

With the system turned on and initialized as described in the

System Power Up section of this manual the front panel

should appear as shown.

By default, the Removal Torque Test Mode will be selected.

The torque display should now read zero.

Zeroing the Display

If the display does not read zero, press [Clear] before

proceeding with this sample test sequence.

The display must never be zeroed when the cap chuck is connected to a bottle in the holding fixture.

This action can result in an offset that will affect the measurements.

A Guide To Using Model 1502 Torqo Quick Start � 31

Page 33: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Now load the bottle to be tested into the holding fixture and

clamp the bottle in accordance with the procedure described

in the Bottle and Cap Tooling section of this manual.

Encage the cap chuck by gently pressing the drive extension

rod down until the chuck engages the cap.

Now press [Open]

Observe the display. The torque will increase until the

system detects that the cap has released. The drive will then

reverse and re-apply the cap to a preset torque value.

The re-application torque is normally set by the user and the

procedure for setting it is described in the section of this

manual titled Setting the Re-application Torque in the

chapter titled Programming a Removal Test.

The Re-application Torque used in this sample test was set

by default when the system was initialized. The chart below

shows the default values for the different size Torqo's.

Mdl #

Max. Torque

(In-lbs)

Application Torque

(25%)

1502-10 10 2.50

1502-20 20 5.00

1502-30 30 7.50

1502-40 40 10.00

1502-50 50 12.50

1502-70 70 17.50

1502-100 100 25.00

A Guide To Using Model 1502 Torqo Quick Start � 32

Page 34: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

After re-applying the cap, the drive will relax to zero and

the removal torque will be displayed. (See example)

Note the Orange Metric LED is Blinking. This indicates

that the test has been completed.

Before another test can be started the results of the last test

must either be discarded by pressing [Clear] or saved in

memory by pressing [Next].

For this Quick Start example press [Next] and this will save

the data in memory for printing in the next step.

Now press [Print] then [1], this will produce a hard copy of

the stored test results when a printer is connected to the

system.

The Torqo is programmed to produce a number of different

printed reports, refer to the section titled Using the Torqo Printer for more information on this subject.

A Guide To Using Model 1502 Torqo Quick Start � 33

Page 35: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Test Modes

What’s Inside

This chapter contains a description of the test modes that can

be selected and a step-by-step procedure for selecting one.

Description Of The Testing Modes

Provisions have been made for a number of special test

modes in the standard Torqo.

�� Removal Torque Test Mode

�� Removal & Incremental Torque Test Mode

�� Removal & Bridge Torque Test Mode

�� Reverse Ratchet Torque Test Mode

�� Batch Test Mode

A chapter is devoted to the programming of each of these

modes however a brief description of each of these modes

can be found below.

A Guide To Using Model 1502 Torqo Test Modes � 34

Page 36: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Removal Torque Test Mode

This mode is used to measure the Removal Torque on

conventional threaded bottles and caps.

Rotating the cap in the counter-clockwise direction until the

cap releases performs the test.

Removal & Incremental Test Mode

This mode is used to measure the removal torque as well as

the Incremental Torque in a single test cycle.

Rotating the cap in the counter-clockwise direction until the

cap releases and then re-applying the cap to a position that is

6 degrees past the initial starting position performs this test.

Removal & Bridge Test Mode

This mode is used to measure the removal torque as well as

the Bridge Torque in a single test cycle.

Rotating the cap in the counter-clockwise direction until the

cap releases and then continuing to drive until the Bridges

on the tamper evident band are broken performs this test.

Reverse Ratchet Test Mode

This mode is used to measure the torque required to rotate a

child resistant closure in the free running state. (i.e. No

vertical down force)

This test is performed by rotating a child resistant cap in the

counter-clockwise direction for 2 complete revolutions

while looking for and storing the highest measured torque

value.

A Guide To Using Model 1502 Torqo Test Modes � 35

Page 37: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Batch Test Mode

This test mode enables the user to keep track of the removal

torque measurements by capping head number.

Selecting A Test Mode

A test mode can be selected by following the procedure

described below.

Press [Mode]

At this time a Blinking LED on the front panel will indicate

the current mode of operation.

Now by continuing to press the [Mode] key the Blinking

LED will advance from one mode to another.

When the desired test mode is illuminated press the [Next] key to terminate the selection process and return to the ready

state.

Can not Change Modes

If the system does not accept the newly selected Mode the

following condition exists.

Data has already been stored in memory for this machine

number and therefore the mode can not be changed.

The Torqo software does not allow the user to combine data from different types of tests in the same data file. (i.e. Same Machine Number)

A Guide To Using Model 1502 Torqo Test Modes � 36

Page 38: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

There are two 2 remedies for this problem.

1. Clear memory for the current machine number.

(See Chapter 17: Clearing Memory)

2. Change machine number and then select the desired

mode. (See Chapter 18: Using Machine Numbers)

A Guide To Using Model 1502 Torqo Test Modes � 37

Page 39: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming a Removal Test

What’s Inside

This chapter describes the features that are applicable to a

removal torque test and a button-by-button procedure for

programming these features.

Test Features

This test is designed to measure the peak torque required to

rotate the cap in the counter-clockwise direction. At the

conclusion of the test the removal torque value will be

displayed and can be stored in memory.

The following features can be programmed as part of a

removal torque test.

1. Re-applying the cap to a specified torque after

measuring the removal torque.

( Setting the Re-Application Torque )

2. Driving the cap completely off after measuring the

removal torque.

( Programming to Leave the Cap Loose )

A Guide To Using Model 1502 Torqo Programming a Removal Test � 38

Page 40: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Remember to select the Removal Test Mode before attempting to program the features for this test.

Setting the Re-application Torque

This is the torque that the cap will be re-applied to after the

Removal Torque Test has been performed.

Press [Set-TQ]

The current re-application torque value will be displayed.

Now key in the desired value and then press [Next] to

accept this value.

The cap will not be re-applied when zero is entered.

Programming to Leave the Cap Loose

This feature enables the system to drive the cap for a

programmed distance after a Removal Torque Test has been

performed.

By inputting a number in degrees that is equivalent to the

degrees required to remove the cap, will leave the cap ready

to be lifted off.

Press [Set-TQ]

The current re-application torque value will be displayed.

A Guide To Using Model 1502 Torqo Programming a Removal Test � 39

Page 41: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Press [Clear]

The display will now read 00.00.

Press [Next]

The system is now programmed not to re-apply the cap.

Now Press [Mode] then [Deg]

The display will now show the currently programmed

distance in degrees to drive.

Key in the desired number of degrees and then press [Next] to accept this value and [Next] again to return to the Ready State.

The [Set-Tq], Re-application Torque, is the torque

value that will be used by the system when the [Close] button is depressed.

This feature enables the system to be used for applying caps

to a very precise torque value.

A Guide To Using Model 1502 Torqo Programming a Removal Test � 40

Page 42: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming an Incremental Test

What’s Inside

This chapter describes the features that are applicable to an

Incremental torque test and a button-by-button procedure for

programming these features.

Test Features

This test is designed to measure the peak torque required to

rotate the cap in the counter-clockwise direction and then re-

apply the cap to a position that is 6 degrees past the initial

starting position. Both the removal torque and the

incremental torque can be displayed and / or stored in

memory at the conclusion of a test.

The following features can be programmed as part of an

incremental torque test.

1. Re-apply the cap to a specified position.

( Programming the Degrees to Drive )

A Guide To Using Model 1502 Torqo Programming an Incremental Test � 41

Page 43: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Remember to select the Incremental Test Mode before attempting to program the features for this test.

Programming the Degrees to Drive

This feature enables the system to re-apply the cap to a

specified position past the initial starting point by inputting

the degrees to drive.

Now Press [Mode] then [Deg]

The display will now show the currently programmed

distance in degrees to drive.

Remember the industry standard for the Incremental test is 6°.

Now key in the desired number of degrees and then press

[Next] to accept this value and [Next] again to return to the

ready to test state.

A Guide To Using Model 1502 Torqo Programming an Incremental Test � 42

Page 44: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming a Bridge Test

What’s Inside

This chapter describes the features that are applicable to a

Bridge torque test and a button-by-button procedure for

programming these features.

Test Features

This test is designed to measure the peak torque required to

rotate the cap in the counter-clockwise direction and then

continue to drive until the bridges on the tamper evident

band break. Both the removal torque and the bridge torque

can be displayed at the conclusion of a test.

The following features can be programmed as part of a

bridge torque test.

1. Distance to drive in order to break the bridges.

( Programming the Degrees to Drive )

A Guide To Using Model 1502 Torqo Programming a Bridge Test � 43

Page 45: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Remember to select the Bridge Torque Test Mode before attempting to program the features for this test.

Programming the Degrees to Drive

This feature enables the system to drive a specified distance

after the removal torque measurement in order to perform

the bridge torque measurement.

Press [Mode] then [Deg]

The display will now show the currently programmed

distance in degrees to drive. (The default value is 270°.)

Now key in the desired number of degrees and then press

[Next] to accept this value and [Next] again to return to the

ready to test state.

A Guide To Using Model 1502 Torqo Programming a Bridge Test � 44

Page 46: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming the Batch Mode

What’s Inside

This chapter describes the features that are applicable to the

Batch Mode and a button-by-button procedure for

programming these features.

Test Features

This test mode enables the user to store the measurements

from a series of removal torque tests by capping head

number and produce statistical reports for evaluating the

performance of each capping head

The Torqo relies on the user to test the samples in the proper order. (i.e. Head 1, Head 2, Head 3, etc.)

1. Programming the number of samples in a batch.

(i.e. Number of capping heads)

2. Retesting a sample.

3. Programming a skip.

4. Erasing the data from memory.

A Guide To Using Model 1502 Torqo Programming the Batch Mode � 45

Page 47: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Selecting Batch Mode

The following button-by-button procedure is used for

selecting the Batch mode.

Press [Mode] then [Batch] then [Next].

The Torqo is now in Batch Mode and the Removal LED and

the Batch LED on the Control Panel will be lit.

When Batch Mode is selected the machine number will be set to Zero. This is the only number that can be used for this mode and therefore only one batch mode can be programmed in the system.

For further explanation of machine numbers refer to the

section of this manual titled Using Machine Numbers.

Remember to select the Batch Test Mode before attempting to program the features for this test.

A Guide To Using Model 1502 Torqo Programming the Batch Mode � 46

Page 48: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming the Number of Samples in a Batch

This feature enables the user to input the number of samples

(i.e. Capping Heads) in a batch.

Press [Mode] then [Batch].

The display will now show the default number 10.

Press [Clear] and then key in the desired number and then

press [Next] to accept this number and return to the Ready

State.

Retesting a Sample in Batch Mode

This feature enables the user to replace a sample in memory

that has been tested incorrectly.

Remember if a value needs to be corrected, it must be done before the next batch is started.

Press [Count]

This will display the total count in the current batch.

Now press [Batch] and then enter the number of the sample

to be rerun and then press [Next].

To exit this feature, once commenced, without performing a

retest press [Clear] then [Next].

A Guide To Using Model 1502 Torqo Programming the Batch Mode � 47

Page 49: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming a Skip in Batch Mode

This feature enables the user to skip a sample when testing

in the batch mode.

Press [Count]

This will display the sample number of the last sample that

was tested.

Now press [Batch] and then enter the number of the sample

to be skipped and then press [Next].

Example: If sample 3 has a damaged cap then after testing sample 2 it may be desirable to skip 3 and go to sample 4.

Press [Count] then [Batch] then [3] then [Next] and

then [Open] to skip sample 3 and test sample 4.

To exit this feature, once commenced, without performing a

skip press [Clear] then [Next].

Erasing Data from Memory in Batch Mode

The following procedure enables the user to erase only the

data in the Batch Mode memory.

Press [Mach]

This will display the current machine number, which is

always Zero when in Batch Mode.

Now press [Display] to erase the data.

To exit after pressing [Mach] without erasing memory press

[Next].

A Guide To Using Model 1502 Torqo Programming the Batch Mode � 48

Page 50: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Testing in Batch Mode

What’s Inside

This chapter provides a button-by-button, step-by-step

description of how to test in Batch Mode.

For purposes of explanation, the following example will be

used.

This example is designed to simulate an hourly testing

schedule from a 4 head capper. (i.e. Test 1 bottle per head

per hour.)

Using Batch Mode

Now with the system in the Removal Test Mode,

Press [Mode] then [Batch] then [Next].

The Torqo is now in Batch Mode and in the Ready State.

However before the testing process can begin, the size of the

batch (i.e. Number of Capping Heads) must be entered.

Press [Mode] then [Batch] then [Clear] then [4] for the

number of capping heads in this example and then [Next]. The number 4 will now remain in memory until changed.

Now press [Mode] then [Batch] then [Next] to start the

batch and store the current time as the name for the first

batch.

A Guide To Using Model 1502 Torqo Testing in Batch Mode � 49

Page 51: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Now test the first 4 samples in numerical order (i.e. Head #

1, 2, 3 and 4) and save the data at the conclusion of each test

by pressing [Next].

When all four samples have been tested and the results

saved, the Batch LED will blink indicating that the batch is

complete.

The software will not permit a new batch to be started until the current batch is complete.

If [Mode] [Batch] Next] is pressed the Error Message

[E18] will be displayed.

[E18] = Current Batch not complete

When the Batch LED is blinking, the software will not permit the testing of another sample.

At this time it is advisable to print the test results for the

current Batch and check for mistakes or out of specification

results. Press [Print] then [Batch] to print the report as

shown below.

Mesa TORQO

Serial #: 00-0000 Date: 01/01/01 08:05

Units: lb-in

Batch Machine:

Current Batch

Batch started: 08:00

----------------------------------------

Head 1: 10.14

Head 2: 10.11

Head 3: 09.57

Head 4: 09.86

<<<<<<< End of Data>>>>>>>

A Guide To Using Model 1502 Torqo Testing in Batch Mode � 50

Page 52: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

If a value is found that needs to be corrected, it must be

corrected before the next batch is started.

The procedure for this is described in the previous chapter

under the heading of Retesting a Sample in Batch Mode.

To commence the next batch press [Mode] then [Batch] then [Next] and once again the current time will be used to

name the new batch.

Now test the next four samples making sure to maintain the

1,2,3,4 order.

Once again print and check the results before continuing.

Observe the Batch Start time, it has now changed from

08:00 to 09:00.

Mesa TORQO

Serial #: 00-0000 Date: 01/01/01 09:05

Units: lb-in

Batch Machine:

Current Batch

Batch started: 09:00

----------------------------------------

Head 1: 10.20

Head 2: 10.15

Head 3: 09.77

Head 4: 09.94

<<<<<<< End of Data>>>>>>>

Now repeat the above-described process for the 10:00, 11:00

and 12:00 o'clock tests.

Five batches with four samples per batch will now be stored

in memory.

The following reports are available for analyzing the

performance of the four capping heads.

A Guide To Using Model 1502 Torqo Testing in Batch Mode � 51

Page 53: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

[Print] then [1] = Prints data only.

[Print] then [2] = Prints data with summary.

[Print] then [3] = Prints data with summary and graphs.

[Print] then [4] = Prints summary only.

[Print] then [5] = Prints summary with graphs.

[Print] then [6] = Prints graphs only.

[Print] then [7] = Prints grand summary of all batches

and all heads.

[Total] then [Print] then [Batch] = Prints a list of all

batches.

For the current example press [Print] then [2], this will

produce the example shown below.

Mesa TORQO

Serial #: 00-0000 Date: 01/01/01 12:05

Units: lb-in

Batch Machine:

Batch Mode

5 batches, 4 heads per batch

======= ( test type ) =======

Removal torque

Head # 1

----------------------------------------

08:00 : 10.14

09:00 : 10.20

10:00 : 10.32

11:00 : 10.08

12:00 : 10.47

**************************

Summary

**************************

Low: 10.08

Average: 10.24

High: 10.47

Std dev: 00.16

Range: 00.39

A Guide To Using Model 1502 Torqo Testing in Batch Mode � 52

Page 54: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Head # 2

----------------------------------------

08:00 : 10.11

09:00 : 10.15

10:00 : 10.30

11:00 : 10.08

12:00 : 10.46

**************************

Summary

**************************

Low: 10.08

Average: 10.22

High: 10.46

Std dev: 00.16

Range: 00.38

Head # 3

----------------------------------------

08:00 : 09.57

09:00 : 09.77

10:00 : 10.06

11:00 : 10.08

12:00 : 09.96

**************************

Summary

**************************

Low: 09.57

Average: 09.88

High: 10.08

Std dev: 00.22

Range: 00.51

Head # 4

----------------------------------------

08:00 : 09.86

09:00 : 09.94

10:00 : 10.22

11:00 : 09.96

12:00 : 10.16

**************************

Summary

**************************

Low: 09.86

Average: 10.02

High: 10.22

Std dev: 00.15

Range: 00.36

<<<<<<< End of Data >>>>>>>

A Guide To Using Model 1502 Torqo Testing in Batch Mode � 53

Page 55: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Exiting Batch Mode

Batch Mode maybe exited and returned to at anytime.

The software will retain the current position in a batch when exiting and returning.

For instance, if batch mode is exited after testing 2 samples

in a batch of four and then returned to, the software will

return the system to the ready to test state for sample 3.

To exit Batch Mode press [Mach]

This will display the current machine number, which is

always Zero, when in Batch Mode.

Now press [1] then [Next] to exit and return to machine

number 1.

To return to the Batch Mode press [Mode] then [Batch] then [Next].

Starting a New Batch Project

To exit the current Batch Mode Project and start another, the

data stored in memory for the first project must be erased.

Press [Mach]

This will display the current machine number, which is

always Zero, when in Batch Mode.

Now press [Display] to clear the data from memory.

Now follow the procedure described at the beginning of this

chapter to start a new batch.

A Guide To Using Model 1502 Torqo Testing in Batch Mode � 54

Page 56: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Performing a Strip Torque Test

What's Inside

This chapter provides a description and the step-by-step

procedure for performing a Strip Torque Test.

Strip Torque Test Description

This test is designed to measure the torque required to strip

the threats in a closure system.

The test is performed by driving the cap in the close

direction until the threats in the cap or on the bottle fail.

Programming a Strip Torque Test

The following procedure will program and then perform the

test.

The Torqo must be in the Removal Mode in order to program and run a Strip Torque Test.

A Guide To Using Model 1502 Torqo Performing a Strip Torque Test � 55

Page 57: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Press [Set Tq] and then input the maximum torque capacity

of the Torqo and then press [Next].

The torque capacity of the system can be found as part of the Model Number on the system label.

(i.e. Mdl 1502-30 has a capacity of 30 in-lbs)

The Torqo is now programmed to perform a Strip Torque

test.

Place the sample to be tested into the bottle holding tooling,

engage the cap chuck and press [Close].

The drive will now rotate the cap until the threads yield or

the capacity of the system is reached. If the threats yield, the

torque at the yield point will be displayed at the end of the

test cycle. If the capacity of the system is reached before the

threats yield, the Torqo will display the error message E 2.

The error message E 2 is displayed whenever the maximum capacity of the system is reached.

(For more information on error messages please refer to the

chapter titled Error Messages.)

A Guide To Using Model 1502 Torqo Performing a Strip Torque Test � 56

Page 58: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Performing a Life Test

What's Inside

This chapter provides a description and the step-by-step

procedure for programming and performing a life test.

Life Test Description

A life test will continuously test a sample for a

programmable number of repetitions.

The test is performed by first performing a removal

measurement and then re-applying the cap to a user

programmable torque value.

Programming a Life Test

The following procedure will program and then perform the

test.

The Torqo must be in the Removal Mode in order to program and run a Life Test.

A Guide To Using Model 1502 Torqo Performing a Life Test � 57

Page 59: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

With Removal Mode selected, program the application

torque by pressing [Set-TQ] and then keying in the desired

value. Press [Next] to terminate the selecting process.

Place the sample to be tested into the bottle holding tooling

and engage the cap chuck.

Now press [Mode] then [Display].

The default value of 100 will be displayed.

Press [Clear] and key in the desired number.

Now when [Next] is pressed to accept this number the test

will commence and execute the programmed number of

repetitions.

The test result from each test can be stored in memory by

using the features described in the chapter titled Logging

Test Results.

A Guide To Using Model 1502 Torqo Performing a Life Test � 58

Page 60: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Retesting a Sample

What's Inside

This chapter provides a step-by-step procedure for retesting

a sample in a data file.

Retesting a Sample

Retesting a sample is primarily used when the operator

discovers an obvious problem that may have influenced the

results of a test after the results have been stored in memory.

To replace this sample in memory, press [Count], then

[Batch], then enter the number of the sample to be replaced

and press [Next].

Now test the replacement sample.

The results from this test will replace the old results when

the [Next] key is pressed.

A Guide To Using Model 1502 Torqo Retesting a Sample � 59

Page 61: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Logging Test Results

What’s Inside

This chapter provides a description and the step-by-step

procedure for logging test results.

Manually Saving Data

At the conclusion of a test the torque read out will display

the measured torque.

The Orange Metric LED will blink indicating that the test is complete. At this time, the test results must either be saved or discarded.

Press [Next] to save a measurement or [Clear] to discard it.

The Orange Metric LED will now stop blinking.

A Guide To Using Model 1502 Torqo Logging Test Results � 60

Page 62: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Automatically Saving Data

The Torqo can be programmed to automatically save test

results at the end of a test.

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 1 and then

press [Next] to complete the selection process.

The Torqo will now automatically save the test results at the

conclusion of a test.

A Guide To Using Model 1502 Torqo Logging Test Results � 61

Page 63: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Transmitting Test Results

What’s Inside

This chapter provides a description of the output port,

connector requirements, communication protocol and

procedure for transmitting test results automatically.

Output Port

The Torqo is supplied with an RS-232 serial communication

port for connecting to a printer or personnel computer.

Connector Requirements

The connector is a female, 9 pin Sub-D style (AMP #

7455201-1).

Mating Connectors

Amphenol = 17-301-1

3M = 8209-6060

Amp = 745203-3

A Guide To Using Model 1502 Torqo Transmitting Test Results � 62

Page 64: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Pin Out

01 = GND (Safety)

02 = XMIT

03 = RCV

05 = GND

Communication Protocol

Data is transmitted in normal ASCII format with the

following specifications.

Baud Rate: 9600

Data Bits: 8

Stop Bits: 1

Parity: No

Xon / Xoff: Yes

Data Format Information

When transmitting test results the following data strings are

used for the different test modes.

Removal Torque

Date_Time_Units_Torque_Test Type_<CR><LF>

Example:

01/01/00_09:00_lb-in_10.80_removal_<CR><LF>

A Guide To Using Model 1502 Torqo Transmitting Test Results � 63

Page 65: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Incremental Torque

Date_Time_Units_Removal Torque_Test Type_Incremental

Torque<CR><LF>

Example:

01/01/00_09:00_lb-in_10.80_inc_12.00<CR><LF>

Bridge Torque

Date_Time_Units_Removal Torque_Test Type_Bridge

Torque<CR><LF>

Example:

01/01/00_09:00_lb-in_10.80_brg_1.15<CR><LF>

Reverse Ratchet Torque

Date_Time_Units_Torque_Test Type_<CR><LF>

Example:

01/01/00_09:00_lb-in_10.80_rrt_<CR><LF>

Batch Mode With Head Number

Head No._Units_Torque_Test Type_<CR><LF>

Example:

Head 1_lb-in_10.80_removal_<CR><LF>

A Guide To Using Model 1502 Torqo Transmitting Test Results � 64

Page 66: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Transmitting Data Automatically

The Torqo can be programmed to automatically transmit test

results at the end of a test or live data during the testing

process.

Transmitting End of Test Results Automatically

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 2 and then

press [Next] to complete the selection process.

The Torqo will now automatically transmit the test results at

the conclusion of a test.

Transmitting and Saving Automatically

This feature provides for the transmission of the end of test

results while also saving the data in Torqo memory.

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 3 and then

press [Next] to complete the selection process.

The Torqo will now automatically transmit and save the test

results at the conclusion of a test.

A Guide To Using Model 1502 Torqo Transmitting Test Results � 65

Page 67: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Transmitting Live Data to a PC

This feature requires Mesa's PC software to receive the data into the PC.

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 4 and then

press [Next] to complete the selection process.

The Torqo will now automatically transmit the measured

torque as the test is being performed.

Transmitting Live Data to a PC and Saving

This feature will transmit live data to a PC and save the end

of test results in Torqo memory.

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 5 and then

press [Next] to complete the selection process.

The Torqo will now automatically transmit the measured

torque as the test is being performed and then save the end

of test results in the Torqo memory.

A Guide To Using Model 1502 Torqo Transmitting Test Results � 66

Page 68: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Transmitting Automatically in Batch Mode

This feature will automatically transmit the end of test

results with the capping head number.

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 6 and then

press [Next] to complete the selection process.

The Torqo will now automatically transmit as described

above.

Transmitting Manually

This feature will transmit the end of test results when the

[Next] key is pressed.

Press [Mode] then [Xmit]

The current status of this feature will now be indicated on

the display. The default state is 0.

Continue to press [Xmit] until the display reads 7 and then

press [Next] to complete the selection process.

The Torqo will now transmit as described above.

A Guide To Using Model 1502 Torqo Transmitting Test Results � 67

Page 69: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Using Machine Numbers

What’s Inside

This chapter describes the use of multiple machine numbers

for storing both test setups and test results in memory.

This feature enables the Torqo to serve multiple operators or

multiple packaging lines.

Machine Number Description

A machine number is a user-defined number for a test setup

and data storage file. Any number from 1 to 9999 maybe

used.

This number appears in the header section of all printed

reports. (See sample below)

Mesa TORQO

Serial #: 00-0000 Date: 01/01/01 09:05

Units: lb-in

Machine #: 1

A Guide To Using Model 1502 Torqo Using Machine Numbers � 68

Page 70: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Entering Machine Numbers

The default machine number is 0001. All setup features and

all test results will be stored with this machine number until

a different number is assigned.

Inputting a Machine Number

Press [Mach]

The current machine number will be displayed.

Now key in the desired number and press [Next].

The newly entered number is now the current machine and

any new setup features will be stored with this number as

well as any measurements that are performed and

subsequently saved.

If a Gold Bottle is frequently used to insure the calibration of the Torqo, assigning a machine number for this purpose can both save time and avoid setup errors.

Select a number and then input the setup features for testing

a Gold Bottle as described in the chapter titled Gold Bottle Calibration Check.

Now just select the assigned number when testing the Gold

Bottle.

By simply changing machine number data can be stored in

different files.

The Torqo can store up to 9 different machine numbers.

A Guide To Using Model 1502 Torqo Using Machine Numbers � 69

Page 71: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Erasing Both the Data and the Machine Number

This feature enables the user to remove from memory all

that is associated with a specific machine number as well as

the number.

Press [Total] then [Mach] then [Display]

Printing a List of the Machine Numbers in use

Press [Print] then [Mach]

A list of the currently used numbers will now be printed.

A Guide To Using Model 1502 Torqo Using Machine Numbers � 70

Page 72: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Clearing Memory

What's Inside

This chapter describes the procedure for erasing data for a

specific machine number and erasing all memory.

Erasing Data for a Specific Machine

This feature enables the user to erase the stored test results

for a specific machine number without affecting the setup

features or any information that is stored under other

machine numbers.

Press [Mach]

This will display the currently selected machine number. If

this is not the machine number for the data that is to be

erased, enter the desired number and then press [Display].

The data will be erased but the machine number with its

setup features will remain in memory.

Erasing all Memory

This feature enables the user to clear all memory.

All data, setup information and machine numbers will be

erased by this procedure.

Press [Total] then [Clear]

A Guide To Using Model 1502 Torqo Clearing Memory � 71

Page 73: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Using Lot Numbers

What's Inside

This chapter provides a button-by-button description of the

procedure for inputting and deleting lot numbers.

Lot Number Description

A lot number is a user-entered number for the identification

of the test results. A number with 4, 8 or 12 digits can be

entered.

This number appears in the header section of all printed

reports. (See sample below)

Mesa TORQO

Serial #: 00-0000 Date: 01/01/01 09:05

Lot number: 123456789123

Units: lb-in

Machine #: 1

A Guide To Using Model 1502 Torqo Using Lot Numbers � 72

Page 74: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Entering a Lot Number

The default state of the machine does not have a lot number

programmed. A lot number can be entered for each machine

number and it will be reported on all printouts for that

machine number until the programmed number is deleted.

Keying in the Lot Number

Press [Mode] then [Lot]

The current status of this feature will now be indicated on

the display. The default value is 0000. Now enter the first

four digits of the desired number. If only four digits are to be

used press [Next] twice to return to the Ready State,

however if four more digits are required press [Display] and then enter the next four digits. Press [Display] again if

four more digits are to be entered or [Next] twice to

terminate the programming process.

Deleting a Lot Number

Using the procedure described above, replace all digits with

the number 0 and then press the [Next] key twice to delete

the lot number and return to the ready state.

A Guide To Using Model 1502 Torqo Using Lot Numbers � 73

Page 75: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Using Specification Limits

What's Inside

This chapter provides a button-by-button procedure for

inputting user provided pass / fail specifications.

Limits Description

Both high and low limits can be programmed into the Torqo

for indicating the acceptability of the samples being tested.

Limits can be inputted for all testing modes and all machine

numbers.

When a failure is detected, the torque display will blink the

measured value instead of a continuous display.

A failure will be printed in red and a left arrow (<) will

appear in the margin adjacent to the failed data.

Both the Test Mode and the Machine Number must be selected before entering Limits.

A Guide To Using Model 1502 Torqo Using Specification Limits � 74

Page 76: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Entering High and Low Limits

For this example the Torqo will be programmed with limits

for a Removal and Bridge Torque Test.

First the system must be in the Removal Mode to program

the limits for the removal portion of the test and then

changed to Removal and Bridge Mode to program the limits

for the bridge portion of the test.

When the system is in the Removal Test mode the front

panel will appear as shown with the Removal LED lit.

Now press [Mode] then [Limits]

The torque display will now flash an L and then an H indicating that either the [Low] or [High] key must be

selected.

When a selection is made the currently programmed limit

will be displayed.

The default state of the machine sets the low limit to zero

and the high limit to the maximum torque capacity of the

system.

For this example press [High].

Now press [Clear] and then enter the desired high limit and

then press [Next] twice to enter this value.

A Guide To Using Model 1502 Torqo Using Specification Limits � 75

Page 77: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Now repeat the above procedure to input the low limit. Press

[Mode], then [Limits], then [Low], then [Clear], then

enter the desired value and press [Next] twice.

To program the limits for the Bridge portion of this test the

test mode must be changed to Removal and Bridge and the

front panel will appear as shown with the Bridge LED lit.

(See Selecting a Test Mode.)

Now repeat the procedure used above for programming the

high and low limits for the removal test.

A Guide To Using Model 1502 Torqo Using Specification Limits � 76

Page 78: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Using The Torqo Printer

What's Inside

This describes chapter describes the many features that the

printer provides and the print commands for using these

features.

Print Features

The following list of printed reports can be obtained from

the Torqo.

�� Header Information

�� Numerical Test Data

�� Summary Data

�� Graphical Data

X - Charts

R – Charts

Histograms

�� Current Machine Configuration

�� List of Currently Used Machine Numbers

�� Configuration and Calibration Data

A Guide To Using Model 1502 Torqo Using The Torqo Printer � 77

Page 79: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Sample Printout

The sample Printout below depicts the major features of a

printed report.

Header

Data

Summary

Graphs

Mesa TORQO

Serial #: 00-0000 Date: 01/01/01 12:05

Units: lb-in

Machine #: 1

======= ( test type ) =======

Removal torque

Samples 4

Sample 1: 09.78

Sample 2: 11.92

Sample 3: 08.76

Sample 4: 08.26

**************************

Summary

**************************

Low: 08.26

Average: 09.68

High: 11.92

Std dev: 01.40

Range: 03.66

X-CHART – open

Group size: 1 ( 4 Samples)

Min:08.00 Avg:09.68 Max:12.00

0 Sub-group number 20

Histogram-Open

Minimum: 0 Maximum: 2

08.00 12.00

A Guide To Using Model 1502 Torqo Using The Torqo Printer � 78

Page 80: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Printing Test Results

The Header shown in the sample printout is always printed

as a means of identifying when and where the report was

generated.

The following list of Print Commands enables the user to

select the Report that best suits the application.

To commence the printing process, press [Print] and then

observe the Torque Display. The display will now scroll

from 1 to 9 indicating that a second key must be selected in

order to print.

Press [Print] then [1] for Data only.

Press [Print] then [2] for Data with Summary.

Press [Print] then [3] for Data with Summary and Graph.

Press [Print] then [4] for Summary only.

Press [Print] then [5] for Summary and Graph.

Press [Print] then [6] for Graph only.

Press [Print] then [7] for Grand Summary in Batch Mode.

Printing the Current Machine Configuration

The Machine Configuration Report contains all the setup

features that have been programmed for the current machine.

It is a good idea to print and save a hard copy of all setups for operator reference.

Printing the setup and comparing it with a saved reference copy at the start of a shift can avoid unnecessary problems.

A Guide To Using Model 1502 Torqo Using The Torqo Printer � 79

Page 81: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Press [Mach] then [Print].

The printer will now produce a Machine Configuration

Report similar to the sample shown below.

Mesa TORQO

Serial #: 00-0000

Date: 01/01/01 12:05

Units: lb-in

Machine #: 1

REMOVAL torque

ENGLISH units

MANUAL LOG

Removal:

Low Limit: 00.00 lb-in

High Limit: 30.00 lb-in

Count: 50

Min TQ: 010.0 %

Minimum torque: 03.00 lb-in

SetTQ: 07.50 lb-in

The sample report above indicates that the current Machine

Number [# 1] is setup to perform a Removal Torque Test.

The test results will be displayed in English Units of

measure rather than Metric.

The test results will be Manually Logged. ( i.e. The test

results will be saved in memory by pressing the next key at

the end of a test.)

The Low Limit for acceptance testing is 00.00 lb-in and

the High Limit is 30.00 lb-in.

The Count (i.e. PK-Cnt) is set to 50. This is the default

value for the removal mode and should not be changed.

The Minimum Torque is set at 10%, which is 03.00 lb-in for the current example.

The Re-application torque (Set-TQ) is set at 07.50 lb-in.

A Guide To Using Model 1502 Torqo Using The Torqo Printer � 80

Page 82: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Printing the Current Machines Report

This feature enables the user to obtain a list of the currently

programmed machine numbers.

Press [Mach] then [Print].

The printer will now produce a Machine Configuration

Report similar to the sample shown below.

Mesa TORQO

Serial #: 00-0000

Date: 01/01/01 12:05

Units: lb-in

Machine Numbers:

1

7

3

5

A Guide To Using Model 1502 Torqo Using The Torqo Printer � 81

Page 83: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Viewing Test Results

What's Inside

This chapter describes the features that are provided for

viewing the end of test results and the summary test results.

Viewing End of Test Results

At the conclusion of a Removal Torque test the measured

Torque will automatically be displayed, however when a

Removal & Incremental or Removal & Bridge Test is

performed there are two measurements to be displayed.

Once again the Removal Torque will be displayed

automatically and to view the second measurement the user

must press [Display]. The display key can be used to toggle

between the two values.

View Summary Test Results

The summary data can be viewed when the system is in the

Ready State by pressing the appropriate key. In a Mode

where two tests are performed, the selected key will act as a

toggle in the same fashion as the Display key in Viewing End of Test Results.

A Guide To Using Model 1502 Torqo Viewing Test Results � 82

Page 84: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Press [Low] to view the lowest Removal Torque value for

the current machine.

Press [Avg] to view the average Removal Torque value for

the current machine.

Press [High] to view the highest Removal Torque value for

the current machine.

Press [Range] to view the range of the Removal Torque for

the current machine.

Press [Std-Dev] to view the Standard Deviation value for

the current machine.

The following formula is used to compute the standard

deviation.

Standard Deviation Formula

This formula computes the mean value (μ) and the

standard deviation (�x) of the measured values (xi) for n

points.

A Guide To Using Model 1502 Torqo Viewing Test Results � 83

Page 85: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Setting the Date and Time

What's Inside

This chapter describes the step-by-step procedure for

changing the date and time.

Changing the Date and Time

The date and time are printed as part of the Header in all

reports and therefore it is important to have it correct.

Press [Mode] then [Time]

Key in the month and then press [Next].

Key in the day and then press [Next].

A Guide To Using Model 1502 Torqo Setting the Date and Time � 84

Page 86: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Key in the year and then press [Next].

Key in the hour and then press [Next].

Key in the minute and then press [Next] twice.

The Setting Procedure is now complete and the software has

returned the system to the Ready State.

A Guide To Using Model 1502 Torqo Setting the Date and Time � 85

Page 87: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Programming the Min Torque

What's Inside

This chapter contains a description of the Min Torque and a

button-by-button procedure for programming the Min

torque.

Min Torque Description

The Min Torque is a value of torque that must be reached

during a test in order for the system to consider the torque

reading valid.

This feature is provided as a method of preventing the

system from recording an erroneous torque during a removal

torque test.

Erroneous torques can be generated from slippage in the

tooling due to the mechanical tolerances of the test sample

and from improper use of the Torqo.

This value is programmed as a percent of the full-scale

torque of the system.

The default value for min torque is 10% of full-scale torque.

10% was selected as a reasonable value because most

measuring devices should not be used consistently in the

bottom 10% of their range. However the torqo, unlike

analog systems, does not lose its accuracy at the low end of

its range and therefore with proper attention to the cap and

bottle tooling measurements can be made below 10%.

A Guide To Using Model 1502 Torqo Programming the Min Torque � 86

Page 88: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

During the testing process, the Torqo looks for a torque that

is greater than the Min Torque.

The system will drive 180° in search of this value.

If a value greater than Min Torque is not found then the

error message E 3 will be displayed.

Inputting a Min Torque

The Min Torque can be changed by the following procedure.

Press [Mode] then [Min-TQ]

The current status of this feature will now be indicated on

the display. The default value is 010.0.

Key in the desired value and press [Next] twice to accept

the entry and return to the Ready State.

A Guide To Using Model 1502 Torqo Programming the Min Torque � 87

Page 89: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Gold Bottle Calibration Check

What's Inside

This chapter describes the Gold Bottle and the procedure for

using it to verify the calibration of the Torqo.

Gold Bottle Description

The Model 1612 Gold Bottle is used to quickly verify the

calibration and operational repeatability of the Model 1502

series torqo's. This bottle is pre-set at the factory to a

customer specified torque value.

Each Gold Bottle is labeled with its exact torque rating.

Since the cap torque of this bottle is the torque produced by

a permanent magnet assembly, the value will always be the

same. Therefore when tested on the Torqo, the measured

value should always repeat unless something has been

damaged.

The bottle is normally supplied with the customers cap so

that the customers Torqo Cap Chuck can be used in the

testing process.

These Gold Bottles are readily available and their torque can

be certified on an annual basis.

A Guide To Using Model 1502 Torqo Gold Bottle Calibration Check � 88

Page 90: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Using the Gold Bottle

The Gold Bottle must be tested in the Removal Mode and

since the cap cannot be removed, it therefore cannot be re-

applied.

The following procedure must be used when testing it.

Press [Mode] repeatedly until the Removal LED indicator is

blinking and then press [Next] to terminate the selection

process.

Press [Set-TQ] and key in the value 00.50 and then press

[Next] to terminate the input process.

Insert the Gold Bottle into the bottle holding tooling.

Press [Clear] to zero the display.

Now engage the Cap Chuck.

Press [Open]

Observe the Removal Torque when the Metric LED

indicator is flashing.

The torque reading on the display should now match the

reading on Gold Bottle Label.

The allowable variation is computed by adding together the tolerance on the bottle and the +/- .5% FS tolerance of the machine.

For example:

If the Gold Bottle has a Torque Setting of 12.00 +/- .2 in-lbs

and the Torqo is a 30 in-lb machine then the allowable

variation is +/- .35 in-lbs. (i.e. .2 +.15 = .35)

A Guide To Using Model 1502 Torqo Gold Bottle Calibration Check � 89

Page 91: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

System Calibration

What's Inside

This chapter is intended to acquaint the user with the

required Calibration Equipment, the Calibration Verification

Process, the process for performing a User Calibration and

the procedure for Selecting the Calibration to be used.

General Calibration Information

The Torqo is supplied with a factory Calibration that is

burned into read-only memory. This is done so that the user

will always have a backup calibration when an incorrect user

calibration is performed.

Calibration Frequency

Checking the calibration of the Torqo at least twice a year is recommended when the system is used frequently.

Required Calibration Equipment

The following list of equipment is required to perform a

Torqo Calibration. When ordering this equipment please

provide the Model Number and Serial Number of the

machine that the equipment will be used on.

A Guide To Using Model 1502 Torqo System Calibration � 90

Page 92: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

1. Calibration Beam (Disk)

2. Calibration Fixture

3. Calibration Weight Set

Calibration Beam

The calibration beam is a disk with a diameter of precisely

8.000 inches. Each disk is serialized and the actual diameter

is stamped on the bottom. The beam is supplied as an

assembly of the following parts.

1. Calibration Disk (C-19935)

2. Calibration Cables (A-21013)

Calibration Fixture

The calibration fixture is supplied as an assembly of the

following parts.

1. Calibration Bracket (D-19936)

2. Cable Pulleys (B-19870)

3. Retain Rings (N5100-25)

Calibration Weight Set

The calibration weight set consists of three weights that are

selected to provide a torque of 25%, 50% and 75% of the

full-scale torque of the Torqo.

(See chart below)

A Guide To Using Model 1502 Torqo System Calibration � 91

Page 93: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Calibration Weight Chart

The required weight in lbs for producing a specific torque

can be computed by dividing the desired torque value by 4,

since 4 inches is the Radius of the calibration beam.

This chart shows the weights that Mesa supplies when a

calibration kit is purchased. The 100% value is obtained by

combining the 25% and 75% weights

Calibration Weights (lbs) Torque

Capacity 5

in-lb

10

in-lb

20

in-lb

30

in-lb

40

in-lb

50

in-lb

70

in-lb

100

in-lb

25% .3125 .625 1.250 1.875 2.500 3.125 4.375 6.250

50% .6250 1.250 2.500 3.750 5.000 6.250 8.750 12.500

75% .9375 1.875 3.750 5.625 7.500 9.375 13.125 18.750

100% 1.250 2.500 5.000 7.500 10.000 12.500 17.500 25.000

Selecting a Calibration to be Used

The currently selected calibration can be determined and

then changed by the following procedure.

Press [Mode] then [Mach] and observe the display.

The currently selected calibration identification number will

be displayed. (See table below)

�� 0 = No correction in use. (For Factory use only)

�� 1 = Factory Calibration in use.

�� 2 = User Calibration in use.

�� 3 = User Calibration in use and latched.

A Guide To Using Model 1502 Torqo System Calibration � 92

Page 94: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

After performing a user Calibration, the number (3) will be automatically selected.

With the currently selected number displayed, Press the

[Limits] key to page through the setting numbers.

When the desired number is displayed, Press [Next] twice

to terminate the selection process.

Performing a Calibration Verification

This procedure is used to verify that the Torqo is in

calibration and to produce a hard copy of the measured

values when a Torqo Printer is connected to the system.

The first step in this procedure is to install the Calibration

Fixture on the Torqo.

Installing the Calibration Fixture

The calibration fixture mounts on the front of the Torqo as

shown in the picture.

To install the fixture, loosen the two 5/16 Socket Head Cap

Screws that can be accessed from under the base plate and

then insert the fixture between the base plate and the Torqo

cover. Now re-tighten the bolts.

(Wrench Size ¼ " Hex)

The next step is to install the calibration beam.

A Guide To Using Model 1502 Torqo System Calibration � 93

Page 95: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Installing the Calibration Beam

The beam is installed by first removing any bottle holding

tooling and then carefully placing the beam on top of the spindle cap.

Screws are provided for holding the beam in place.

(Qty 2 x .250-20 x 1" Skt Hd Cap Screws)

(Wrench size 3/16 " Hex.)

Creating a Calibration Verification Chart

With a Torqo Printer connected, a Calibration Verification

Chart can be created as the calibration check is being

performed.

If the system is not connected to a printer, the measured values should be manually recorded for analysis at the conclusion of this procedure.

Press [Mode] then [Cal] to commence the process. With the

depression of the cal button the following information will

be printed. This Header not only identifies the Torqo but

also provides information about which calibration is

currently selected.

Mesa TORQO

Serial #: 00-0000

Date: 01/01/01 12:05

Units: lb-in

Factory

Calibration Chart

A Guide To Using Model 1502 Torqo System Calibration � 94

Page 96: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Now hang the 25% weight on the left hand calibration cable.

This will apply the torque in the counter clockwise direction.

(i.e. Opening Direction)

Observe the torque display and when the reading becomes

steady press [Print] to send the data to the printer.

Repeat this procedure for each weight in both the opening

and closing direction and then press the [Next] key twice to

return to the ready state.

Upon completion of this procedure, a calibration chart

similar to the one pictured below will have been created.

Calibration Verification Chart

Mesa TORQO

Serial #: 00-0000

Date: 01/01/01 12:05

Units: lb-in

Factory

Calibration Chart

01/01/01 12:05 Open: 05.00 lb-in

01/01/01 12:06 Open: 10.00 lb-in

01/01/01 12:07 Open: 15.00 lb-in

01/01/01 12:08 Open: 20.00 lb-in

01/01/01 12:09 Close: 05.00 lb-in

01/01/01 12:10 Close: 10.00 lb-in

01/01/01 12:11 Close: 15.00 lb-in

01/01/01 12:12 Close: 20.00 lb-in

Now compare the values in the above chart with the

theoretical values that can be found in the tables below.

A Guide To Using Model 1502 Torqo System Calibration � 95

Page 97: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Model 1502-5 ( Torque Capacity 5 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 1.250 .5 .025

50 2.500 .5 .025

75 3.750 .5 .025

100 5.000 .5 .025

Model 1502-10 ( Torque Capacity 10 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 2.50 .5 .050

50 5.00 .5 .050

75 7.50 .5 .050

100 10.00 .5 .050

Model 1502-20 ( Torque Capacity 20 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 5.000 .5 .10

50 10.00 .5 .10

75 15.00 .5 .10

100 20.00 .5 .10

Model 1502-30 ( Torque Capacity 30 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 7.500 .5 .15

50 15.00 .5 .15

75 22.50 .5 .15

100 30.00 .5 .15

A Guide To Using Model 1502 Torqo System Calibration � 96

Page 98: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Model 1502-40 ( Torque Capacity 40 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 10.00 .5 .20

50 20.00 .5 .20

75 30.00 .5 .20

100 40.00 .5 .20

Model 1502-50 ( Torque Capacity 50 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 12.50 .5 .25

50 25.00 .5 .25

75 37.50 .5 .25

100 50.00 .5 .25

Model 1502-70 ( Torque Capacity 70 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 17.50 .5 .35

50 35.00 .5 .35

75 52.50 .5 .35

100 70.00 .5 .35

Model 1502-100 ( Torque Capacity 100 lb-in )

Torque in

% Capacity

Torque in

lb-in

Tolerance in

+/- % FS

Tolerance

+/- lb-in

25 25.0 .5 .5

50 50.0 .5 .5

75 75.0 .5 .5

100 100.0 .5 .5

All readings should be within the specified tolerance range.

A Guide To Using Model 1502 Torqo System Calibration � 97

Page 99: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Performing a User Calibration

Once again the calibration fixture and beam must be

installed on the Torqo before the calibration procedure can

begin.

Calibration Methodology

The calibration procedure consists of entering on the display

a known value of torque that will be applied by the weight

and beam. Then applying this value by hanging the weight

on the calibration cable. At this time the software will

compute the difference between the entered value and the

value applied by the weight. This difference will become the

correction value for this specific point and will be stored in

memory. Once all the points ( i.e. 25%, 50%, 75% and

100% in both the clockwise and counter-clockwise

direction) have been measured the software will compute a

correction curve to be applied during the measuring process.

Entering the Calibration Mode

The Calibration Mode is entered by first turning the

instrument off and then re-powering the system with the

[Cal] key depressed

The front panel will now appear as pictured.

The word HELP indicates that the system is in the calibrate

mode and ready for user input.

A Guide To Using Model 1502 Torqo System Calibration � 98

Page 100: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Before commencing the calibration procedure all previous

user calibrations should be erased and the system zeroed.

Erasing all Previous User Calibrations

Press [Mode], this will erase all previous calibrations and

return to the HELP screen.

Zeroing the Torqo in the Calibration Mode

Press [Cal] then [Next]; this will automatically zero the

system prior to commencing the calibration. Once again the

software will return to the HELP screen.

The following calibration procedure will begin by inputting

25%, then 50%, then 75% and finally 100% in the clockwise

direction and then repeating the procedure in the counter-

clockwise direction.

Care must be taken when entering values to insure that they have been keyed in correctly.

A Guide To Using Model 1502 Torqo System Calibration � 99

Page 101: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Entering Calibration Torques

Now key in the following numbers [0] [2] [5] [0].

The first 0 indicates that the direction will be clockwise.

The 25.0 indicates that the applied torque will be 25%.

Now press [Next]

Observe the display, when the next key was selected above

the decimal point was removed indicating that the torque

load can now be applied.

Place the 25% weight on the right hand calibration cable and

the 0250 count on the display will decrease to a small

number. (i.e. Usually less than 0050) Now when the display

becomes steady press [Next] and the value will be stored.

The display should now have returned to the HELP screen

and the 25% weight should be removed.

Now key in the following numbers [0] [5] [0] [0].

The first 0 indicates that the direction will be clockwise.

The 50.0 indicates that the applied torque will be 50%.

Now press [Next]

Once again the decimal point will be gone and now the 50%

weight should be placed on the right hand cable. The 0500

count on the display will decrease usually to a value less

than 0050. Now when the displayed reading becomes steady

press [Next].

A Guide To Using Model 1502 Torqo System Calibration � 100

Page 102: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

The display should now have returned to the HELP screen

and the 50% weight should be removed.

Now key in the following numbers [0] [7] [5] [0].

The first 0 indicates that the direction will be clockwise.

The 75.0 indicates that the applied torque will be 75%.

Now press [Next]

Once again the decimal point will be gone and now the 75%

weight should be placed on the right hand cable. The 0750

count on the display will decrease usually to a value less

than 0050. Now when the displayed reading becomes steady

press [Next].

The display should now have returned to the HELP screen

and the 75% weight should be removed.

After calibrating at 25%, 50% and 75% in the clockwise

direction, key in the following numbers [2] [0] [0] [0].

The 2 indicates that the direction will be clockwise.

The 00.0 indicates that the applied torque will be 100%.

Now press [Next].

Once again the decimal point will be gone and now the

100% weight should be placed on the right hand cable. The

2000 count on the display will decrease usually to a value

less than 2050. Now when the displayed reading becomes

steady press [Next].

The 100% weight value is obtained by combining the 25%

weight and the 75% weight.

A Guide To Using Model 1502 Torqo System Calibration � 101

Page 103: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Now key in the following numbers [1] [2] [5] [0].

The 1 indicates that the direction will be counter clockwise.

The 25.0 indicates that the applied torque will be 25%.

Now press [Next]

Place the 25% weight on the left hand calibration cable and

the 1250 count on the display will decrease to a small

number. (i.e. Usually less than 1050) Now when the display

becomes steady press [Next] and the value will be stored.

The display should now have returned to the HELP screen

and the 25% weight should be removed.

Now key in the following numbers [1] [5] [0] [0].

The 1 indicates that the direction will be counter clockwise.

The 50.0 indicates that the applied torque will be 50%.

Now press [Next]

Once again the decimal point will be gone and now the 50%

weight should be placed on the left hand cable. The 1500

count on the display will decrease usually to a value less

than 1050. Now when the displayed reading becomes steady

press [Next].

The display should now have returned to the HELP screen

and the 50% weight should be removed.

A Guide To Using Model 1502 Torqo System Calibration � 102

Page 104: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Now key in the following numbers [1] [7] [5] [0].

The 1 indicates that the direction will be counter clockwise.

The 75.0 indicates that the applied torque will be 75%.

Now press [Next]

Once again the decimal point will be gone and now the 75%

weight should be placed on the left hand cable. The 1750

count on the display will decrease usually to a value less

than 1050. Now when the displayed reading becomes steady

press [Next].

The display should now have returned to the HELP screen

and the 75% weight should be removed.

Now key in the following numbers [3] [0] [0] [0].

The 3 indicates that the direction will be counter clockwise.

The 00.0 indicates that the applied torque will be 100%.

Now press [Next].

Once again the decimal point will be gone and now the

100% weight should be placed on the left hand cable. The

3000 count on the display will decrease usually to a value

less than 3050. Now when the displayed reading becomes

steady press [Next].

The display should now have returned to the HELP screen

and the 100% weight should be removed.

Press [Next] twice to terminate this procedure and return to

the ready state. As the calibration procedure is terminated,

the software computes the correction curve for this user

calibration.

A Guide To Using Model 1502 Torqo System Calibration � 103

Page 105: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

At this time it is a good idea to printout a report of the

complete system setup and calibration information.

This report shows both the factory calibration information

and the user calibration information.

Observe the actual correction values that are shown in this

report.

The new correction values should not be significantly different than the factory calibration values.

(See sample System Report below)

A Guide To Using Model 1502 Torqo System Calibration � 104

Page 106: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

System Report

Calibration/setup data:

Current date: 01/01/01 12:00

Calibrate date: 05/05/00

Software version: 20.0

Serial number: 00-0000

Printer type: Epson

Torque capacity: 30 lb-in/34.56 kg-cm

Report header:

Mesa TORQO

Configuration switches…

1 2 3 4 5 6 7 8

on off on off off on off off

Available drive speeds: 1 RPM

Pass/Fail criteria:

Mach #: 1 Close torque: 07.50

% Minimum torque: 10.00

Minimum torque: 3.00

Test type: Remove inc brg rrt

Count: 50 30 50 3

High: 30.00 30.00 30.00 30.00

Low: 00.00 00.00 00.00 00.00

Deg: 0 6 270 720

Beam deflection calibration

CW CCW

Factory User Factory User

7FD4: 0 0 8042: 0 0

7FD5: 1 0 8043: 1 0

7FD6 2 1 8044: 2 1

Values Omitted for Sample

8005: 25 24 8073: 20 24

8006: 26 24 8074: 20 24

8007: 26 25 8075: 21 25

8008: 26 27 8076: 21 25

8009: 27 28 8077: 22 26

800A: 27 29 8078: 22 26

800B: 26 29 8079: 21 26

800C: 25 28 807A: 20 24

800D: 25 27 807B: 20 23

800E: 24 25 807C: 19 21

800F: 23 24 807D: 19 19

Values Omitted for Sample

803A: 1 5 80AD: 1 1

8040: 1 5 80AE: 0 0

8041: 0 4 80AF: 0 0

25% Point

50% Point

75% Point

A Guide To Using Model 1502 Torqo System Calibration � 105

Page 107: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

At this time, it is a good idea to produce a Calibration Verification Chart with the New User Calibration.

This chart is clearly identified as a User Calibration Chart

and should be maintained with the system records until

another User Calibration is performed.

Mesa TORQO

Serial #: 00-0000

Date: 01/01/01 12:05

Units: lb-in

User

Calibration Chart

01/01/01 12:05 Open: 05.00 lb-in

01/01/01 12:06 Open: 10.00 lb-in

01/01/01 12:07 Open: 15.00 lb-in

01/01/01 12:08 Open: 20.00 lb-in

01/01/01 12:09 Close: 05.00 lb-in

01/01/01 12:10 Close: 10.00 lb-in

01/01/01 12:11 Close: 15.00 lb-in

01/01/01 12:12 Close: 20.00 lb-in

A Guide To Using Model 1502 Torqo System Calibration � 106

Page 108: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Child Resistant Cap Option

What's Inside

This section is indented to acquaint the user with the

features and use of this option.

About the Child Resistant Option

This option enables the Torqo to provide a pre-settable down

force for push and turn caps.

An air supply with the ability to provide a minimum of 30

PSI is required.

Once connected to the air, the down force can be set to the

required force for a given application.

Connecting the Air Supply

The air supply connects to the air filter on the rear of the

tower.

This connection is a � NPT thread.

Both the air filter and regulator are designed for a maximum

of 100 PSI however the Force Gauge is limited to 30 PSI.

A Guide To Using Model 1502 Torqo Child Resistant Cap Option � 107

Page 109: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Adjusting the Down Force

Once connected, the Air Control Knob should be adjusted to

provide the required amount of down force.

Observe the down force gauge on the front of the tower

cover while adjusting the knob.

In most cases, a down force of 15 t0 20 lbs will open a child

resistant closure.

A Guide To Using Model 1502 Torqo Child Resistant Cap Option � 108

Page 110: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Down Force Mechanism Nomenclature

The names in the callouts below will be used when

describing the setup and adjustment of the Child Resistant

Down Force Mechanism.

Foot Support

Shaft Drive Shaft

Cap Chuck

Drive Shaft

Clamp Collar

Drive Shaft

Bearing

Down Force

Foot

Foot Position

Shaft Clamp

Adjusting the Down Force Foot

The Down Force Foot applies the force on the cap by

pressing down on the Drive Shaft Bearing when the Down

Force Switch is actuated.

A Guide To Using Model 1502 Torqo Child Resistant Cap Option � 109

Page 111: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

To adjust the foot put the Down Force switch in the up

position and then place the sample to be tested in the bottle

holding fixture.

Now position the Cap Chuck on the cap by lowering the

drive shaft.

Pivot the Down Force Foot over the top of the Drive Shaft

Bearing and adjust for approximately 1/16 of an inch of

clearance. (i.e. the thickness of a quarter)

Loosening the Foot Position Shaft Clamp and sliding the

Foot Support Shaft up or down and then retightening the

clamp makes this adjustment.

Operating the Child Resistant Mechanism

Once the mechanism has been adjusted, a sample is tested

by pivoting the foot into position and actuating the Down

Force Switch.

At the conclusion of the test cycle, move the switch to the up

position and pivot the foot to the rear.

Repeat this procedure for the next sample.

A Guide To Using Model 1502 Torqo Child Resistant Cap Option � 110

Page 112: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Key lock Option

What's Inside

This chapter describes the features and use of the key lock

option.

Key lock Option Description

The key lock option adds a key actuated switch on the rear

panel of drive tower.

This option was designed to prevent the inadvertent altering

of the Torqo setup.

Using the Key lock Option

The key lock switch has two positions.

In the Program position, this option is not in use and

therefore all of the torqo features can be accessed through

the control panel.

In the Run position, the user is not permitted to change the

Mode, Limits or Setup features.

If a setup key is pressed, a three-beep alarm will be sounded

to indicate that this selection cannot be performed.

A Guide To Using Model 1502 Torqo Key lock Option � 111

Page 113: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Multi-Speed Option

What's Inside

This chapter describes the use of the Multi-Speed Option.

Multi-Speed Option Description

The multi-speed option adds a selector switch on the rear

panel of drive tower.

This option was designed specifically for testing bottle caps

with tamper evident bands. The increased speed of the drive

provides for a cleaner tear of the material that connects the

band to the cap.

This option may also be used for performing more rapid

testing. However one must be careful, when using the

system in this manner, to make sure that the increased speed

is not distorting the measurements.

The Speed Selector Switch enables the user to select the

following four speeds.

Position 1: 1 RPM

Position 2: 2 RPM

Position 3: 4 RPM

Position 4: 6 RPM

A Guide To Using Model 1502 Torqo Multi-Speed Option � 112

Page 114: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

When a speed other than 1 RPM is selected, it will be used

in all the Test Modes with the exception of the Removal &

Incremental Mode.

The Removal & Incremental Mode is an industry standard

and cannot be changed.

This test is always performed at 1 RPM

When a high speed is selected, the system will drive at the

selected speed for the first 75% of a closing cycle and then

automatically revert to 1 RPM for the remaining 25%.

Using the Multi-Speed Option

To use this option the operator just has to put the Speed

Selector in the desired position and test in the normal

manner.

As mentioned previously, performing Removal tests at

speeds greater than 1 RPM may distort the data.

To insure the accuracy of the measurements, one should first

perform the tests at 1 RPM and then repeat the tests at a

greater RPM. The results should then be compared for

accuracy and repeatability.

A Guide To Using Model 1502 Torqo Multi-Speed Option � 113

Page 115: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Error Messages

What's Inside

This chapter describes the Error Messages that may be

displayed when programming and testing.

Description of Error Messages

The following Error Messages are programmed in the

software to assist the user.

When an Error Message is displayed, the Metric LED will

blink rapidly. Pressing the [Clear] key will erase the

message and return the system to the Ready State.

This message will appear on the display when the user

attempts to enter more that 9 machine numbers.

This message will appear on the display when the torque

capacity of the system is exceeded during a test cycle.

A Guide To Using Model 1502 Torqo Error Messages � 114

Page 116: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

This message will appear on the display when the system

has driven 180° without finding a legitimate torque.

This message will appear on the display when the user

attempts to enter a low limit greater than the high limit for

the current machine.

This message will appear on the display when the user

attempts to enter a high limit lower than the low limit for the

current machine.

This message will appear on the display when the user

attempts to enter a high limit greater than the capacity of the

machine.

This message will appear on the display when the user

attempts to enter a number more than 1 greater than the

current count.

This message will appear on the display when the user

attempts to enter a zero as a test number.

This message will appear on the display when the user

attempts to enter a peak count greater than 100 in the

Reverse Ratchet Mode.

A Guide To Using Model 1502 Torqo Error Messages � 115

Page 117: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

This message will appear on the display when the user

attempts to enter a zero in peak count.

This message will appear when the system clock is not

performing properly.

This message will appear when the system is programmed to

close to a specific torque, which is not reached in three

revolutions.

This message will appear when the system is unable to re-

apply the cap to the Set TQ value after a Removal test.

This message will appear when the system is unable to re-

apply the cap to its original position before performing the

Incremental test.

This message will appear when the torque exceeds 30 in-lbs

at 6 RPM on a Multi Speed machine.

This message will appear when the system memory is full.

A Guide To Using Model 1502 Torqo Error Messages � 116

Page 118: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

This message will appear when the user attempts to perform

a test and the current Batch is already complete.

This message will appear when the user attempts to start a

new Batch when the current Batch is not complete.

A Guide To Using Model 1502 Torqo Error Messages � 117

Page 119: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Trouble Shooting

What's Inside

This section is provided to help the user solve some of the

minor problems that may occur with a Torqo.

Contacting Mesa Laboratories, Inc.

If you are unable to solve your problems with the system

after trying the suggestions in this manual then please call

are service department at 303-987-8000.

Before calling Mesa please obtain the serial number

from the label on the back of the control instrument and if

possible obtain the software version number by pressing

[Total] then [Display] and observing the torque readout.

If you have a printer connected to the system the software

version can be obtained by pressing [Total] then [Display] then [Print].

Trouble Shooting the System

If your Torqo does not appear to be performing properly, the

following procedures can be used to find and correct some

of the more common problems.

A Guide To Using Model 1502 Torqo Trouble Shooting � 118

Page 120: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Restoring the Factory Defaults

Restoring the factory defaults will correct errors that have

been made when programming the system.

Performing a Soft Reset

Press [Total] then [Mode]

This will return the system to the factory default state.

Re-initializing the System

The Torqo like all microprocessor-controlled systems can

experience a computer crash.

Blinking LED's

When a crash occurs, the LED's on the Control Panel will

Blink in a random pattern.

This problem is usually caused by an interruption in power

and can be avoided by using a Surge Protector.

Performing a Hard Reset

This procedure will re-initialize the microprocessor and reset

all factory settings.

The Hard Reset is performed by first turning the

instrument off and then re-powering the system with the

[Display] key depressed

A Guide To Using Model 1502 Torqo Trouble Shooting � 119

Page 121: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Erratic System Behavior

The Torqo relies on a battery backed memory chip to store

program settings, date and time settings and test results. If

your system has lost the ability to maintain its settings, the

battery in the memory chip may have degraded. This will

result in Erratic System Behavior.

The life expectancy of this device is from 5 to 10 years.

However its life is affected by the amount of time that the

system is turned off because this is when the battery is in

use.

It is not a good idea to stock these devices as backup because their life is measured from date of manufacture.

L's on the Display

When the memory chip has degraded, the Torque readout

will frequently display the letter L in one or more digit

locations.

Checking the Memory Chip

The memory chip can be checked by setting the Date and

Time and then turning the system Off for 30 minutes. If

the battery backup is working the system date and time will

be correct.

A new memory chip can be purchased from Mesa and it

will be supplied with detailed instructions for installing it.

A Guide To Using Model 1502 Torqo Trouble Shooting � 120

Page 122: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Torqo Drive Problems

The Torqo drive motor, drive electronics and drive belt

system are time proven, high reliability devices.

Most drive related problems can be remedied by checking

the following items.

Checking the Power

1. Verify that the power switch is on. (0=Off, 1=On)

2. Verify that the power cord is seated in the AC Input.

3. Verify that the power cord is plugged into a good outlet.

4. Verify the connection of the Drive cable at both ends.

5. If the drive still does not operate check the fuse.

Checking the Fuse

1. Turn the Power Switch off.

2. Remove the AC power cord.

3. With a small flat bladed screwdriver, pry open cover with a

twisting motion.

4. Note the direction of the arrow and then pull the white fuse

holder out of the module.

5. Check the fuse and replace if required. (MDL 1½ for

120VAC and MDL ¾. for 240VAC)

A Guide To Using Model 1502 Torqo Trouble Shooting � 121

Page 123: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Checking the Drive Belt

A broken drive belt can be detected by attempting to rotate

the drive shaft by hand.

If the Torqo is a standard Model 1502 without the Multi-

Speed option the shaft should not turn more than a few

degrees when rotated by hand.

If the Torqo has the multi-speed option, the shaft will turn

but with a great deal of resistance.

The drive belt is broken when the shaft turns freely in either of the two cases described above.

Replacing the Belt

1. Unscrew the Drive Shaft Knob.

2. Remove the Cover Screws.

3. Remove the Cover by lifting it up and over the

drive shaft.

The drive belt system in this picture is a multi-speed unit

however the procedure for removing a belt, determining the

size of a belt and installing a new belt is basically the same

in all units.

1. To remove the belt, loosen the screws on the motor

plate to take the tension off the belt and then

remove the belt.

2. To install the belt, place it around the sprockets,

slide the motor plate to tension the belt and then

tighten the motor plate screws.

A Guide To Using Model 1502 Torqo Trouble Shooting � 122

Page 124: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Determining the Belt Size

The best way to determine the correct belt size for a system

is to lay the broken belt on a flat service and count the

number of drive pins.

Then contact the factory for the correct part number.

Checking the DC Power

1. Put the AC Drive Power Switch in the Off

position.

2. Remove the 12 screws that hold the rear panel on

the tower.

3. Carefully tilt the top of rear panel out as shown in

the picture.

4. Now put the AC Drive Power Switch in the On

position.

5. Observe LED01: It should be on whenever the

tower is energized. This indicates that the +5 vdc

supply is on.

Checking for the Clock Pulses

The clock pulses come from the control instrument to the

drive tower whenever the drive motor is turned on.

With the tower energized, press [Open] and observe the

LED02.

LED02 will be fully on when the motor is not running and

will dim slightly when the drive is turned on. (It is actually

blinking quickly)

A Guide To Using Model 1502 Torqo Trouble Shooting � 123

Page 125: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

If all of the above checks do not indicate where the problem lies please contact the factory for assistance.

A Guide To Using Model 1502 Torqo Trouble Shooting � 124

Page 126: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Maintenance and Cleaning

What's Inside

This chapter describes the recommended maintenance and

cleaning procedures for a Torqo and its tooling.

General Information

The Torqo is designed to operate for many years with the

minimum amount maintenance.

However the Torqo, like any precision measuring tool,

should be periodically checked for accuracy. (See System

Calibration)

The calibration of the Torqo should be checked at least

twice a year.

A protective cover can be supplied for those applications

where washdowns are frequently performed.

General Maintenance

Keep the system clean, for this will make the use of the

Torqo more pleasurable.

Disconnect the power to the system when cleaning.

A Guide To Using Model 1502 Torqo Maintenance and Cleaning � 125

Page 127: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Clean the exterior with a mild detergent and a soft cloth.

Avoid abrasive cleaners that may scratch the lens on the

torque display and the keypad.

Cleaning the Tooling

The functionality of the bottle holding tooling and the

chucks should be periodically checked.

All bottle tooling should move smoothly and easily.

Chucks should slip onto caps with the minimul amount of

force.

Cleaning the 1506A Bottle Clamp

This tool is made from corrosion resisting materials and

therefore it can be fully submerged in hot soapy water for

cleaning.

Remove the fixture by unscrewing the 2 mounting screws

and lifting straight up.

Make sure the center hole is clean before re-installing the

fixture.

Cleaning the 1614 Pin Vise

This tool can also be submerged in hot soapy water but care

should be taken to throughly dry the lead screw and then it

should be lubricated with a light oil.

A Guide To Using Model 1502 Torqo Maintenance and Cleaning � 126

Page 128: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Glossary of Terms

Bridges

The small strips of material that span the gap between the cap and a tamper evident

band.

Incremental Torque

The force required to rotate a cap to a position 6 degrees past the initial position in

the tightening direction.

Re-application Torque

The force that will be applied to a cap when re-tightening it.

Removal Torque

The force required to loosen the cap on a threaded container.

RS-232

A serial communication standard for data interchange.

A Guide To Using Model 1502 Torqo Glossary of Terms � 127

Page 129: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

Cleaning the 1614 Pin Vise 126

Clearing Memory 10, 37, 71

Communication Protocol 10, 62–63, 62–63, 62–63, 62–

63, 62–63

Index Connecting the Air Supply 107

Connecting the Drive Tower to the Readout 16

Connector Requirements 10, 62

Contacting Mesa Corporation 118

Control Panel Nomenclature 28

Creating a Calibration Verification Chart 94

D

Data Bits 63

Data Format Information 63

Deleting a Lot Number 73 1 Description of Error Messages 114

Description Of The Testing Modes 34 1506A Beverage Bottle Clamp Nomenclature 22

Determining the Belt Size 123

Down Force mechanism Nomenclature 109 A

E About the Child Resistant Option 107

AC adaptor 14, 17 Entering a Lot Number 73

Adjusting the 1506A Beverage Bottle Clamp 23 Entering Calibration Torques 100

Adjusting The 1614 Universal Vise 26 Entering High and Low Limits 75

Adjusting the Down Force 108, 109 Entering Machine Numbers 69

Adjusting the Down Force Foot 109 Entering the Calibration Mode 98

Automatically Saving Data 61 Erasing all Memory 71

Erasing all Previous User Calibrations 99 B Erasing Both the Data and the Machine Number 70

Erasing Data for a Specific Machine 71 Batch Test Mode 34–36, 34–36, 34–36, 34, 36, 46

Erasing Data from Memory in Batch Mode 48 Baud Rate 63

Erratic System Behavior 120 Blinking Metric LED 30

Error Messages 11, 56, 114 Bottle and Cap Tooling 8, 20, 32

Exiting Batch Mode 54

C G

Calibration Beam 91, 93 General Calibration Information 90

Calibration Fixture 91, 93, 98 General Maintenance 125

Calibration Frequency 90 Gold Bottle Calibration Check 11, 69, 88

Calibration Methodology 98 Gold Bottle Description 88

Calibration Verification Chart 94–95, 94–95, 94–95,

106 I Calibration Weight Chart 92

Calibration Weight Set 91 Initializing The System 18, 119

Can not Change Modes 36 Input power cable 14, 17

Changing the Date and Time 84 Inputting a Machine Number 69

Checking for the Clock Pulses 123 Inputting a Min Torque 87

Checking the DC Power 123 Installing the Calibration Beam 94

Checking the Drive Belt 122 Installing the Calibration Fixture 93

Checking the Fuse 121 Installing Tooling For Griping The Cap 21

Checking the Memory Chip 120 Installing Tooling For Holding The Bottle 20

Checking the Power 121 Interface cable 14, 16, 19

Child Resistant Cap Option 11, 107

Cleaning the 1506A Bottle Clamp 126

A Guide To Using Model 1502 Torqo Index � 128

Page 130: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

K Programming an Incremental Test 9, 41

Programming the Batch Mode 9, 45

Programming the Degrees to Drive 41–42, 41, 38–39,

41–42, 43–44, 43, 41–42, 43

Key lock Option 11, 111

Key lock Option Description 111

Keying in the Lot Number 73 Programming the Min Torque 10, 86 Keypad 8, 30, 126 Programming the Number of Samples in a Batch 45–

47, 45–47, 45–47

L Programming to Leave the Cap Loose 38–39, 38–39,

38–39, 38 LED Indicators 28–29, 28–29, 28–29, 28–29, 28–29

Life Test Description 57 R Limits Description 74

Logging Test Results 9, 30, 58, 60 Ready State 18, 30, 36, 40, 47, 49, 73, 82, 85, 87, 95,

103, 114 Lot Number Description 72

L's on the Display 120 Re-initializing the System 119

Removal & Bridge Test Mode 35

M Removal & Incremental Test Mode 35

Removal Torque Test Mode 31, 34–35, 34–35, 34–35,

34–35 Machine Number Description 68

Manually Saving Data 60 Replacing the Belt 122

Mating Connectors 62 Required Calibration Equipment 11, 90

Metric LED 29–30, 29–30, 33, 60, 89, 114 Restoring the Factory Defaults 119

Min Torque Description 86 Retesting a Sample 9, 45–47, 45–47, 45, 47, 51, 59

Multi-Speed Option 112–13, 112–13, 112–13, 122 Retesting a Sample in Batch Mode 47, 51

Multi-Speed Option Description 112 Reverse Ratchet Test Mode 35

O S

Operating the Child Resistant Mechanism 110 Sample Printout 78–79, 78–79, 78–79

Output Port 10, 62 Selecting a Calibration to be Used 92

Selecting A Test Mode 36, 76 P Selecting Batch Mode 46

Setting the Date and Time 10, 84, 120 Parity 63

Setting the Re-application Torque 32, 38–39, 38–39 Performing a Calibration Verification 93

Speed Selector Switch 112 Performing a Hard Reset 119

Spindle Cap 20–21, 20–21, 20, 94 Performing a Life Test 9, 57

Standard Deviation Formula 83 Performing a Strip Torque Test 9, 55

Starting a New Batch Project 54 Performing A Test 8, 31

Stop Bits 63 Performing a User Calibration 11, 90, 93, 98

Strip Torque Test Description 55 Pin Out 63

System Calibration 11, 90, 125 Pin Positioning Chart 25

System Report 104–5, 104–5, 104–5, 104–5 Positioning The Chuck With The Height Adjuster 27

Powering the Drive Tower 17 T Powering the Optional Printer 19

Powering the Readout Instrument 17 Testing in Batch Mode 9, 49

Print Features 77 Torqo Drive Problems 121

Printing a List of the Machine Numbers in use 70 Torque Display 17, 29, 31, 74–75, 74–75, 74–75, 74–

75, 79, 95, 126 Printing Test Results 79

Printing the Current Machine Configuration 79 Transmitting and Saving Automatically 65

Printing the Current Machines Report 81 Transmitting Automatically in Batch Mode 67

Programming a Bridge Test 9, 43 Transmitting Data Automatically 65

Programming a Removal Test 8, 32, 38 Transmitting End of Test Results Automatically 65

Programming a Skip in Batch Mode 48 Transmitting Live Data to a PC 66, 62–63

Programming a Strip Torque Test 55 Transmitting Live Data to a PC and Saving 66

Programming an Application Torque 38 Transmitting Manually 67

A Guide To Using Model 1502 Torqo Index � 129

Page 131: A Guide To Using Model 1502 Torqo - Bottle Cap Torque Testers · 2016-04-08 · About This Manual For convenience, this manual is divided into a number of chapters. A brief description

A Guide To Using Model 1502 Torqo Index � 130

Transmitting Test Results 10, 62–63, 62–63, 62–63, 62,

63

Trouble Shooting 11, 118

Trouble Shooting the System 118

U

Using Batch Mode 49

Using Lot Numbers 10, 72

Using Machine Numbers 10, 29, 37, 46, 68

Using Specification Limits 10, 74

Using the Gold Bottle 89

Using the Key lock Option 111

Using the Multi-Speed Option 113

Using The Torqo Printer 10, 33, 77

V

View Summary Test Results 82

Viewing End of Test Results 82

Viewing Test Results 82

X

Xon / Xoff 63

Z

Zeroing the Display 31

Zeroing the Torqo in the Calibration Mode 99