70 series - xs instruments · 2019. 2. 8. · datalogger, for pt100 rtd probes 0,01 °c from -99,99...

16
70 Series The Future Starts Now Temp70 RTD Professional Datalogger, for Pt100 RTD probes 0,01 °C from -99,99 to +99,99 °C 0,1 °C from -200,0 to +999,9 °C Thermometer Temp70 _ Ver. 3.0_ Eng 09/2020

Upload: others

Post on 18-Feb-2021

6 views

Category:

Documents


0 download

TRANSCRIPT

  • 70 Series

    The Future Starts Now

    Temp70 RTD Professional Datalogger, for Pt100 RTD probes 0,01 °C from -99,99 to +99,99 °C 0,1 °C from -200,0 to +999,9 °C

    Thermometer Temp70 _ Ver. 3.0_ Eng 09/2020

  • Content index

    1. Descprition of the Instrument ...................................................................................................... 1 1.1 Display LCD ................................................................................................................................ 1 1.2 Keypad description ..................................................................................................................... 2 1.2.1 Keypad operations ............................................................................................................... 2 1.2.2 Switching on the instrument ................................................................................................. 2 1.2.3 Switching off the instrument ................................................................................................. 2 1.3 Function of the keys ................................................................................................................... 3 2. Temperature Measurement ........................................................................................................... 4 2.1 First power-on ............................................................................................................................. 4 2.2 Measurement .............................................................................................................................. 4 2.3 Min / Max function ...................................................................................................................... 4 2.4 Hold Function.............................................................................................................................. 5 3. Store, Recall and Clear memory .................................................................................................. 6 3.1 Save the data Manually .............................................................................................................. 6 3.2 Save the data Automatically ....................................................................................................... 6 3.3 Recall of the saved data ............................................................................................................. 6 3.4 Clear Memory ............................................................................................................................. 6 4. Other Functions ............................................................................................................................. 7 4.1 Stability Criteria .......................................................................................................................... 7 4.2 Auto Power off ............................................................................................................................ 7 4.3 Offset adjustment........................................................................................................................ 8 4.4 Last calibration data ................................................................................................................... 8 5. Parameter Settings ........................................................................................................................ 9 5.1 Setup Menu ................................................................................................................................ 9 6. Software DataLink+ (for Windows 7/8/XP/10) ....................... Errore. Il segnalibro non è definito. 6.1 Functions .................................................................................................................................. 11 7. Disposal of electronic devices ................................................................................................... 13 8. Technical specifications ............................................................................................................. 14

  • Portable Thermometer

    1

    1. DESCPRITION OF THE INSTRUMENT 1.1 Display LCD

    Display

    (1) ………… Type of the probe

    (2) ………… USB icon, appears when the instrument is connected to the computer (3) ………… Data logger icon, appears during the automatic registration of values (4) ………… Date and Time (5) ………… Unit of measurement (6) (7) …….. Memory and Recall icon

    M+ — Storing values; RM — Recall values (8)…………. MIN / MAX / HOLD indicators (9) ………… Stability indicator (10) ……….. Low battery indicator, when appears replace the batteries (11) …………Measurement value

    1

    11

    3 4

    5

    7

    6

    2

    9

    10

    8

  • Portable Thermometer

    2

    1.2 Keypad description

    Keypad

    1.2.1 Keypad operations Momentary press 1,5 seconds.

    1.2.2 Switching on the instrument

    Press to power on the instrument:

    The display shows the model, the firmware version and the values of the internal settings. The instrument is now ready to measure.

    1.2.3 Switching off the instrument

    In the measurement mode, press and hold for 3 seconds to turn off the instrument.

  • Portable Thermometer

    3

    1.3 Function of the keys

    key Press Description

    Momentary press

    • With the meter turned off, press this button to turn it on. • In the measurement mode is used to turn on / off the backlight.

    Long press • During measurement, press and hold for 3 seconds to turn off the instrument.

    Momentary press

    • In the measurement mode, press to start recording the minimum and maximum values.

    • While recording MIN / MAX, press to display the value of MIN / MAX.

    Long press • Hold for 3 seconds to exit the MIN / MAX recording.

    Momentary

    press

    • Used to lock / unlock the reading. • During setup and memory recall (RM), press to return to measurement mode.

    Momentary

    press

    • In the measurement mode, press to enter setup. • During setup: Press to select/confirm the program. • During the calibration, press to confirm the calibration.

    Momentary press

    • In the measurement mode, press to store the reading. • Used to scroll through and change the values of the parameters of the setup

    menu.

    Long press • Hold for 3 seconds to exit the automatic recording of values.

    Momentary

    press

    • In the measurement mode, press to recall the stored values. • Used to scroll through and change the values of the parameters of the setup

    menu.

    Momentary press

    • During the recall of memory (RM): used to scroll through the memory. • During setup: press to scroll through the main menu and sub menus. • In the sub-menu of the setup: Press to change the value of the parameter.

  • Portable Thermometer

    4

    2. TEMPERATURE MEASUREMENT

    2.1 First power-on

    At the first start you must set the following parameters: - Select the unit of the measurement °C / °F P6.1 - Set the date and time P6.4 / P6.5 - Clear the memory P6.7

    For additional settings, refer to the section Setup menu.

    2.2 Measurement

    Always connect the probe before turning on the power.

    Turn ON the meter with . After the initial self-diagnosis the instrument enters directly into the

    measurement mode. Place the probe in the measuring point and wait for the stability.

    When the value is stable and the icon of stability appears, you can take the reading.

    2.3 Min / Max function

    While measuring press to start the recording of the Min/Max values. The MIN MAX icons will stat

    flashing on the display.

    Press once to see the maximum value reached until that time, press another time to see the minimum

    value reached until that time. If you press again then the meter will again start recording the Min/Max values with the icons flashing on the display. When viewing the Min/Max value, the recording stops because the display isn’t showing the actual reading but the same value for minimum and maximum recorded.

    To exit from the recording of the Min/Max values, hold pressing for at least 3 seconds, the MIN MAX

    icons will disappear.

    If the instrument is set to the automatic shutdown then the instrument will turn off after 20 minutes. If you want to record the

    minimum and maximum values over a longer period than it’s necessary to disable the auto power off P6.8

    Warning:

  • Portable Thermometer

    5

    2.4 Hold Function

    During the measurement press key to lock the value. The display will

    show HOLD icon and the value will freeze. Press again to unlock the value. This feature can be particularly useful when after performing the measurement you must move with the instrument to write the value.

    During the recording of Min/Max values the hold function is disabled.

    Note:

  • Portable Thermometer

    6

    3. STORE, RECALL AND CLEAR MEMORY

    3.1 Save the data Manually

    When the reading is stable press the key, M+ icon and the number of the

    saved data will appear on the display. The manual save feature does not work if you set a recording interval. Set "00.00" at parameter P6.3 in the setup menu to disable automatic recording.

    3.2 Save the data Automatically

    Set the recoding interval in parameter P6.3 of setup menu, the icon will

    appear on the display. Now the meter is ready for the automatic data logging.

    Now press key (momentary press),and the meter will start recording the data

    and the icon will flash according to the interval set.

    To stop the recording press the key once, the icon will stop flashing.

    3.3 Recall of the saved data

    During measurement press to recall the last saved memory. The meter will

    show RM icon and the number of the saved data. Press and keys to

    scroll between the memory, long press these keys to speed up the scroll.

    3.4 Clear Memory

    To clear the memory go to the parameter P6.7 of the setup menu, select YES and confirm with key.

    The meter has saved 20

    data in automatic mode

    The display shows RM icon and

    the number of memory recalled

  • Portable Thermometer

    7

    4. OTHER FUNCTIONS

    4.1 Stability Criteria

    When the reading is stable the display shows . If this icon does not appear

    or blinks, do not consider the reading as the final value. Always wait for the stability of the measurement. It’s possible to select between 3 levels of the stability at parameter P7.5 of setup menu:

    Λο Low stability With small changes the icons remains fixed and indicates stable value.

    νορ Medium stability Intermediate stability, default choice

    Ηι High stability The icon appears only when the value is very stable and will disappear if there is even a slight change in the value.

    4.2 Auto Power off

    This meter powers off automatically after 20 minutes of inactivity. To deactivate auto power-off go to the parameter P6.8 of setup menu and select Off.

    During the automatic data storing the auto power-off function is deactivated.

    Note:

  • Portable Thermometer

    8

    4.3 Offset adjustment

    All instruments are factory pre-calibrated and remain stable over time, but in the event of an error in the reading, you can recalibrate the instrument to eliminate this error. To recalibrate the instrument at 0.0 ° C, proceed as follows:

    - Make the ice with double-distilled water. - Put the ice cubes in a 500ml beaker up to half of the volume - Add distilled water up to ¾ of the volume - Place the beaker on a magnetic stirrer with gentle stirring. - Immerse the probe in the beaker, making sure it is submerged in water and does not touch the

    wall of the beaker.

    - Cover the beaker with polystyrene to isolate the mixture of water / ice from the air as much as possible.

    - Allow to stir for about 10 minutes. - In this way you reach a temperature of 0.0 ° C

    If the meter reads 0.0 ° C (± accuracy of the measuring chain) then it means that the instrument is operating properly and no adjustment is necessary, otherwise go to parameter P7.2 of the setup menu, adjust the

    value with the arrow keys and confirm with .

    The adjustment made at 0.0°C takes effect over the entire range of measurement, for example if you correct to -1.0 ° C, then the measurement moves of -1.0°C over the entire field. The adjustment made in this way is kept in memory after turning off the meter and also when replacing the batteries. In case you change the probe you must restore the factory calibration in setup menu P7.6 and proceed, if necessary, with a new calibration.

    4.4 Last calibration data

    You can see the date and time of last calibration in the setup menu P7.3 in accordance with GLP.

  • Portable Thermometer

    9

    5. PARAMETER SETTINGS During measurement mode press to enter in the setup menu.

    Press and keys to scroll setup menu: P6.0→P7.0

    P6.0: Settings of the meter P7.0: Settings of the parameter Temperature

    Press key to enter in the selected menu and to confirm the changes.

    Press and keys to scroll the sub-menus and to change the values.

    Anytime press key to exit from setup menu and to return in the measurement mode, the changes

    made and confirmed with key will be saved.

    5.1 Setup Menu

    P6.1 Measurement unit (°C-°F)

    P6.2 Select backlight duration P6.3 Set recording interval for automatic data logging P6.4 Date setting P6.5 Time setting P6.7 Clear saved data P6.8 Activate / Deactivate auto power-off

    Sub-menu of the meter settings

    Main Menu

    P7.2 Offset adjustment P7.3 Shows date and time of last calibration P7.4 Resolution selection 0,0/0,00 P7.5 Stability criteria selection P7.6 Restore factory calibration

    Sub-menu of the parameter Temperature

  • Portable Thermometer

    10

    Parameters of the setup menu and factory default settings

    Parameter Description Display Settings Default setting

    P6.1 Measurement unit Unit °C-°F °C

    P6.2 Backlight duration bl 1-2-3-On 1 minute

    P6.3 Data recording interval - 0:00

    P6.4 Date setting / - -

    P6.5 Time setting / - -

    P6.7 Clear saved data Clr No-Yes No

    P6.8 Auto Power-off Aoff On - Off On

    P7.2 Offset adjustment ± 5 °C ----

    P7.3 Last calibration date and time

    P7.4 Resolution rES 0.01 / 0.1 0.01

    P7.5 Stability criteria SC Normal—High—Low Normal

    P7.6 Restore factory calibration Ft No-Yes No

    To power off the meter exit from SETUP!

    Note:

  • Portable Thermometer

    11

    6. SOFTWARE DATALINK+ (FOR WINDOWS 7/8/XP/10) It is possible to connect the instrument XS Temp 70 PT100 to the PC and then use the DataLink + 1.9 software (and later versions) to perform data download, Data Logger directly on PC and exports in .xls (Excel) and .pdf. The software can be downloaded for free from the web site (pay attention to the correct installation of the drivers). • https://www.giorgiobormac.com/it/download-software_Download.htm. • Connect the USB cable inside each package to the USB port on the top panel of the instrument and the other

    end to a COM port on the computer. • Use the USB cable supplied with the instrument only. • Start the program and then switch on the instrument. • Wait for connection to be established (the connection data are shown at the bottom left of the display).

    6.1 Functions

    • Download: the data saved in the instrumental memory are downloaded to a PC and displayed in the table for processing.

    • M+: instantaneous acquisition of a value (equivalent to the manual Data Logger option). • Logger: automatic acquisition with set frequency. • Empty: emptying the data in the table. If the password is active, it will be requested. • Export to Excel / Export to PDF: export to PDF and Excel of all the data in the table, of graphs,

    calibration reports and instrumental information. • Save to file / Open from file: saving the data in the table and possibility to reload them in order to

    process them or continue recording. • Select the language: set the interface language (Eng – Ita – Deu – Esp – Fra – Cze). • Table / Graph: how to display the acquired data. The graphs are divided by parameter and can be

    printed separately.

    https://www.giorgiobormac.com/it/download-software_Download.htm

  • Portable Thermometer

    12

  • Portable Thermometer

    13

    7. DISPOSAL OF ELECTRONIC DEVICES

    The electrical and electronic equipment marked with this symbol cannot be disposed of in public landfills. According to the UE Directive 2002/96/EC, the European users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one.

    The illegal disposal of electrical and electronic equipment is punished with an administrative fine.

  • Portable Thermometer

    14

    8. TECHNICAL SPECIFICATIONS

    Model Temp 70 Professional

    For Probes Pt100

    Measuring range -200 … +999 °C

    Resolution 0,01 °C from -99,99 to +99,99 °C 0,1 °C from -200,0 to +999,9 °C

    Accuracy (meter only)

    ±0,1 °C ( -201 ... -100 °C) ±0,03 °C (-99,99 … +99,99 °C) ±0,1 °C ( +100... +999,9 °C)

    Datalogger 1000 values

    Sampling time Man/auto 1 min … 99 hours Min/Max reading Yes

    Stability indicator Yes selectable (low-med-high)

    Hold function Yes

    Working conditions Temp: -20.. +65 °C UR: 10 … 90% (no condensation)

    Protection class IP56

    CE conformity EN61326-1/A1: 1998 (EU EMC)

    Auto power-off After 20 min (deactivable)

    Display LCD backlight

    Inputs Single 3-pin round connector

    Power supply 3 x 1,5V “AA” alkaline

    Batteries life >750 hours (without backlight)

    Weight and dimensions (without probe) 86 x 196 x 33 mm / 295 g Weight and dimensions carrying case (without probe) 260 x 135 x 75 cm / 580 g

    1. Descprition of the Instrument1.1 Display LCD1.2 Keypad description1.2.1 Keypad operations1.2.2 Switching on the instrument1.2.3 Switching off the instrument

    1.3 Function of the keys

    2. Temperature Measurement2.1 First power-on2.2 Measurement2.3 Min / Max function2.4 Hold Function

    3. Store, Recall and Clear memory3.1 Save the data Manually3.2 Save the data Automatically3.3 Recall of the saved data3.4 Clear Memory

    4. Other Functions4.1 Stability Criteria4.2 Auto Power off4.3 Offset adjustment4.4 Last calibration data

    5. Parameter Settings5.1 Setup Menu

    6. Software DataLink+ (for Windows 7/8/XP/10)6.1 Functions

    7. Disposal of electronic devices8. Technical specifications