3 paul p. bourbeau and brandi lynn swartz 4 geisinger...

22
First Evaluation of the WASP, a New Automated Microbiology Plating Instrument 2 Paul P. Bourbeau * and Brandi Lynn Swartz 3 Geisinger Medical Center, Danville, PA. 4 5 6 7 8 *Corresponding author. Mailing address: Division of Laboratory Medicine, Geisinger Medical 9 Center, Danville, PA 17822-0131. Phone: 570-271-7467. E-mail:[email protected] 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 ACCEPTED Copyright © 2009, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights Reserved. J. Clin. Microbiol. doi:10.1128/JCM.01963-08 JCM Accepts, published online ahead of print on 21 January 2009 on June 21, 2018 by guest http://jcm.asm.org/ Downloaded from

Upload: hatruc

Post on 12-May-2018

214 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

First Evaluation of the WASP, a New Automated Microbiology Plating Instrument 2

Paul P. Bourbeau*

and Brandi Lynn Swartz 3

Geisinger Medical Center, Danville, PA. 4

5

6

7

8

*Corresponding author. Mailing address: Division of Laboratory Medicine, Geisinger Medical 9

Center, Danville, PA 17822-0131. Phone: 570-271-7467. E-mail:[email protected] 10

11

12

13

14

15

16

17

18

19

20

21

22

23

24

ACCEPTED

Copyright © 2009, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights Reserved.J. Clin. Microbiol. doi:10.1128/JCM.01963-08 JCM Accepts, published online ahead of print on 21 January 2009

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 2: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

2

ABSTRACT 25

26

Many laboratories are experiencing growing shortages of trained microbiology technologists 27

and technicians. Consequently, there is considerable interest in new automation that could 28

potentially lessen labor demands for specimen processing. In this study, we present the first 29

published evaluation of a new microbiology instrument, the WASP (Walkaway Specimen 30

Processor) manufactured by Copan, Inc. in which we evaluated cross-contamination, accuracy of 31

plating, and quality of results. Absence of cross-contamination was demonstrated by plating a 32

total of 200 alternating inoculated and sterile specimen tubes. The ability of the WASP to 33

subculture enrichment broths was evaluated with 106 LIM broths, with results identical to 34

routine testing. Plating of urine specimens with the WASP was compared to plating with the 35

Dynacon Inoculab instrument. 300 specimens were plated in duplicate with both instruments 36

using 1-ul loops and 293 specimens were plated in duplicate on both instruments with 10-ul 37

loops. Results were compared between duplicate plating with the same instrument (replicate 38

plating) and for consensus agreement between the two instruments. Replicate plating results 39

were comparable for both instruments while the WASP had more specimens with significant 40

results than the Inoculab with the 1-ul loop only. Lastly, for plating of 113 ESwab specimens, the 41

manual method and WASP plating each yielded 90 potential pathogens. In summary, we report a 42

first evaluation of a new microbiology specimen-plating instrument, the WASP, which offers 43

opportunities to automate the plating of microbiology specimens to an extent that has not been 44

possible to date. 45

46

47

48

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 3: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

3

49

50

INTRODUCTION 51 52

Clinical microbiology laboratories have largely been bypassed in the advances in automation that 53

have benefitted other areas of the clinical laboratory in recent years. Continuously monitoring 54

blood culture systems as well as automated microbial identification and susceptibility testing 55

systems are widely utilized. However, the specific areas of specimen processing and culture 56

workup remain manual tasks with few changes in the recent past. While some larger labs utilize 57

urine-plating instrumentation, most microbiology laboratories have no automation in their 58

processing areas. In this report, we present the results of a preliminary evaluation of a new 59

microbiology plating instrument that offers the potential to automate the plating of a variety of 60

liquid-based microbiology specimens. 61

62

MATERIALS AND METHODS 63

64

Overview of WASP. The WASP (Walk Away Specimen Processor) is a new instrument 65

manufactured in Italy for Copan Diagnostics (Murrieta, CA) that is designed to plate liquid 66

specimens from a variety of different transport devices (See Figures 1 and 2). The WASP utilizes 67

two Toshiba SCARA (selective compliant assembly robot arm ) robots. The first robot moves 68

specimens and plated media and takes specimens to the decapping device while the second robot 69

does the actual inoculation and plate streaking. Barcode readers on the WASP scan the specimen 70

tube, and a printer prints specimen information and a barcode on a label that is placed on the 71

plated media. A 9-silo carousel holds 342 standard BD plates (Becton Dickinson Microbiology 72

Systems, Cockeysville, MD) or 378 standard Remel plates (Remel, Lenexa, KS). Only transport 73

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 4: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

4

devices with the specimen in a liquid phase can be processed on the WASP. With the exception 74

of large urine cups (120 ml cups in our lab), all specimens are loaded on the WASP using special 75

Teflon pallets that contain holes that are sized for specific containers. For example, one pallet 76

holds 12 Vacutainer urine transport tubes, a different pallet holds 12 ESwabs, and a third type of 77

pallet holds 6 enteric transport media tubes. Up to 6 pallets can be loaded on to the instrument at 78

one time, resulting in a maximum load of 72 Vacutainers or ESwabs at one time. The WASP 79

also has a vortex and a spinner/centrifuge that can be used to prepare specimens for plating. 80

Actual plating is done by three metal loops incorporated in a triquetra (three cornered) loop 81

inoculation tool (Figure 3). The loops are available in 3 sizes: 1 ul, 10 ul, and 30 ul. The WASP 82

contains multiple sensors to verify proper operation including one sensor that is connected to a 83

camera that verifies that the loop contains specimen after it is dipped into the specimen. A touch-84

screen computer facilitates selection of inoculation protocols, media, and streaking patterns, as 85

well as other functions, and guides operation of the WASP. 86

This evaluation consisted of a preliminary evaluation of the first production version of a 87

WASP in a clinical microbiology laboratory. At the time that this evaluation was performed, the 88

WASP contained software to process only urine Vacutainer specimens, UriSwabs and ESwabs. 89

Consequently, this evaluation was limited to those specimen transport devices. 90

Cross-Contamination Studies. To determine whether cross contamination occurs when 91

streaking sequential specimens using the WASP instrument, studies were performed with both 92

Vacutainer Urine C&S Preservative Plus Plastic Tubes (B.D.) and ESwabs (Copan). A fresh 93

subculture of Escherichia coli (ATCC # 35218) was used to prepare a suspension equivalent to a 94

0.5 MacFarland standard in sterile saline. Dilutions were performed to obtain ~105 CFU/ml and 95

~106

CFU/ml organism suspensions. 96

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 5: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

5

Four ml of sterile saline was added to each of 50 Vacutainer tubes, and 4 ml of the 97

105 CFU/ml E. coli suspension was added to each of an additional 50 Vacutainer tubes. The 98

tubes were then placed into the appropriate WASP pallet by alternating a tube filled with the 99

E. coli suspension and a tube filled with sterile saline. The pallets were placed on the WASP 100

instrument, and a standard urine-streaking pattern was selected that included a Blood Agar Plate 101

(BAP) and MacConkey (MAC) Agar Plate for each of the 100 Vacutainers. All Vacutainer tubes 102

were plated first using a 1µL loop and then retested using a 10µL loop. This yielded a total of 103

200 sets of plates: 100 sets inoculated with sterile saline and 100 sets inoculated with the E. coli 104

suspension. Plates were incubated at 37°C and examined at 24 h for growth. 105

For testing of ESwabs, 0.1 ml of the106

CFU/ml E. coli suspension was added to each ESwab 106

tube (ESwabs contain 1 ml of liquid Amies transport media). The tubes were placed into the 107

appropriate WASP pallet by alternating a tube inoculated with the E. coli suspension and a tube 108

filled with only the transport media. The pallets were placed on the WASP instrument, and a 3-109

quadrant streaking pattern was selected that included a Blood Agar Plate (BAP) and MacConkey 110

(MAC) Agar Plate for each of the 100 ESwab tubes. All ESwab tubes were processed first using 111

a 10-µL loop and then retested using a 30µL loop. This yielded a total of 200 sets of plates: 100 112

sets inoculated with sterile saline and 100 sets inoculated with the E. coli suspension. Plates were 113

incubated at 37°C and examined at 24 h for growth. 114

Enrichment broth subculture. To verify the accuracy of the WASP for subculture of 115

enrichment broths, testing was performed with LIM (Remel) broths. Routine LIM broths 116

inoculated for prenatal group B strep screens from the Geisinger Microbiology Laboratory were 117

used for this testing. Vaginal/rectals specimens are incubated in LIM broth for 18-24 h before 118

subcultured for routine testing. After the routine testing was completed, LIM broth tubes were 119

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 6: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

6

vortexed, and 1 ml was transferred to an empty, sterile ESwab tube. The ESwab tubes with LIM 120

broth were subsequently subcultured to neomycin nalidixic acid (NNA) agar plates (BD) and 121

BAP plates on the WASP using the 10-ul loop. These cultures were read and worked up at 24 h 122

and 48h, and the results were compared with the routine culture results. 123

Urine cultures. The Dynacon Inoculab (Dynacon Inc., Mississauga, ON, Canada) is used for 124

routine plating of urine cultures submitted in Vacutainer tubes to the Geisinger Microbiology 125

laboratory. Plates on the Inoculab can be inoculated with either 1-ul or 10-ul loops. For this 126

validation, routine urine specimens received in BD Vacutainers were first plated on BAP and 127

MAC agar plates using a 1-ul loop on the Inoculab for the routine microbiology lab culture, and 128

these cultures were worked up and reported using standard Geisinger Microbiology Laboratory 129

protocols. All specimens were then plated a second time with the Inoculab, and those plates were 130

labeled for study purposes. The same specimens were then subsequently plated using a 1-ul loop 131

by the WASP on BAP and MAC plates in duplicate and identified for the study. This resulted in 132

4 total platings (2 times on Inoculab and 2 times on WASP). Additional specimens were then 133

plated twice each on the Inoculab and the WASP using 10-ul loops. 134

All plates were placed in a non-CO2 incubator for a minimum of 16 hours. One set of the 135

plates from the Inoculab inoculated with a 1-ul loop was read and reported by a Geisinger 136

Microbiology Technologist and the other sets of plates were read by one of us (BLS). Standard 137

Geisinger microbiology laboratory procedures were used in the workup of all cultures. A senior 138

technologist or one of us (PPB) reviewed all discrepant results. 139

ESwabs. Routine cultures from a variety of sources submitted to the laboratory in ESwabs 140

were plated using the WASP. After routine manual plating was completed, specimens were 141

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 7: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

7

loaded onto the WASP and inoculated with a 3-quadrant streaking pattern, using BAP, MAC, 142

and CHOC agar plates. The WASP 30-ul loop was used to inoculate all plates. 143

All cultures plated by the WASP were worked up independently of the manually plated 144

cultures. Results were compared after all testing was completed on the manually plated and 145

WASP plated specimens. 146

RESULTS 147

Cross-Contamination Studies. 50 inoculated and 50 sterile Vacutainer tubes were alternately 148

loaded on the WASP. They were plated with both 1-ul and 10-ul loops for a total of 200 149

cultures. No colonies were observed from the sterile tubes, and consistent streaking patterns were 150

noted from the inoculated tubes. Similar results were observed with ESwab tubes for both the 10-151

ul and 30-ul loops with no growth from any of the sterile tubes and consistent streaking patterns 152

from the inoculated tubes. 153

Enrichment broth subculture. 106 LIM broths were subcultured with the WASP. Preliminary 154

studies (data not shown) indicated that the use of the 30-ul loop did not result in a satisfactory 155

number of isolated colonies. The use of the 10-ul loop, however, produced consistent isolated 156

colonies. Using the 10-ul loop, there was 100% concordance with the routine culture method 157

with each detecting 20 positive and 86 negative test results. 158

Urine cultures. A total of 300 specimens were processed in duplicate on the WASP and 159

Inoculab using 1-ul loops, while 293 specimens were processed in duplicate on both instruments 160

using 10-ul loops. The same specimens were not necessarily used for the 1-ul loops and the 10-ul 161

loops so results were analyzed separately by loop size. Culture results were divided into those 162

considered likely significant and those considered likely not significant. 163

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 8: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

8

For specimens plated with 1-ul loops, single or multiple isolates with < 1000 CFU/ml and 164

cultures with 3 or more organisms each with >10,000 CFU/Ml (multiple flora) were considered 165

not significant as were isolates of Micrococcus spp. and Lactobacillus spp. in any quantity. For 166

the purposes of this study, coagulase-negative staphylococci and viridans streptococci were 167

considered likely significant if present in a concentration of ≥10,000 CFU/ml when present as a 168

single pathogen or with no more than one other organism. For tabulation purposes, if a culture 169

contained ≥10,000 CFU of a potential pathogen (e.g. E. coli) and < 10,000 CFU/ml of one or two 170

other organisms, the E. coli isolate was included with the significant isolates and the other 171

organism(s)was included with the not significant isolates. For specimens plated with 10-ul loops, 172

results obtained were interpreted in a similar manner, accounting for the 10-fold difference in the 173

dilution factor. Results obtained with the Inoculab and the WASP were evaluated in two ways. 174

First, the replicate results were compared for each instrument. For example, the result obtained 175

on specimen #1 plated with the WASP using the 1-ul loop was compared with the second plating 176

of specimen #1 on the WASP using the same size loop. Second, the recovery of significant 177

isolates was compared between the two instruments. 178

The results of the replicate results for specimens with significant isolates are summarized in 179

Table 1. Both instruments yielded similar results in replicate plating for significant isolates for 180

both sized loops. Overall, for quantitation (same CFU/ml range, e.g. both replicates >105

181

CFU/ml), there were no significant differences between the results for the two instruments. 182

For cultures with non-significant results, there was agreement for the Inoculab between paired 183

plates for the 1-ul loop for 184/223 results and 219/220 for the 10-ul loop. For the WASP, for 184

specimens with non-significant results, there was agreement for the 1-ul loop for 192/220 results 185

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 9: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

9

and 218/220 for the 10-ul loop. Most of the discrepant results involved specimens with no 186

growth on one set of plates and < 10 actual colonies on the paired culture plates. 187

Results from WASP and Inoculab for specimens with significant results not in agreement are 188

summarized in Table 2. From 74 specimens with significant results that were plated with 1-ul 189

loops, there were 11 specimens with discrepant results. For 5 of the specimens, one of the 190

Inoculab results was not in agreement with the 3 other results while for 1 specimen, one WASP 191

result was not in agreement with the 3 other results. Four significant isolates were detected only 192

by the two WASP cultures. Lastly, for one specimen, there was a difference in CFU/ml counts 193

between the two WASP cultures and the two Inoculab cultures. 194

From 65 specimens with significant results that were plated with 10-ul loops, there were 7 195

specimens with discrepant results. One each of the WASP and Inoculab specimens was not in 196

agreement with the 3 other results. For five specimens, there was a difference in CFU/ml counts 197

between the two WASP cultures and the two Inoculab cultures. 198

ESwab cultures. A total of 113 specimens that were collected in ESwabs were plated on the 199

WASP. These included 13 vaginal specimens and 100 routine cultures of mixed types (wounds, 200

drainages, fluids, nares, skin, upper respiratory and others). All plates were inoculated using a 201

30-ul loop and read at 24 and 48 h of incubation. Manual and WASP plating both yielded 90 202

potential pathogens. The yield of normal flora (skin, vaginal, and respiratory) was also the same. 203

One of the WASP cultures grew one colony of mold that was not detected on the manually 204

plated culture. 205

206

207

208

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 10: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

10

DISCUSSION 209

Many laboratories are experiencing growing shortages of trained microbiology technologists 210

and technicians. This has been exacerbated not only by growth in routine testing but also by the 211

demand for testing performed for epidemiological purposes, such as for MRSA (3). 212

Consequently, there is considerable interest in new automation that could potentially lessen labor 213

demands for specimen processing (7). 214

Current instrumentation available for microbiology processing includes streaking and plating 215

instruments. There are 3 instruments currently available that can perform plate inoculation and 216

streaking: The Dynacon Inoculab (LQ and LQH models); the bioMérieux MicroStreak 217

instrument; and the WASP. 218

Inoculab instruments are designed to plate urine specimens of one specimen type. They plate 219

specimens from one type of container that is selected at the time of instrument purchase. The 220

Dynacon model LQH that we have in our lab holds a total of 38 uninoculated Remel plates, 221

giving it a capacity of 19 specimens without reloading if two plates are used for each culture or 222

38 specimens if one plate is used. Since the LQH model has only one silo for uninoculated 223

plates, if you use more than one type of plate for a culture, the plates must be intercalated in the 224

stack. Inoculab instruments can also be used in a streak-only function 225

The MicroStreak was released in 2008 (2). The MicroStreak requires a plugged disposable 226

pipette tip for each specimen and one disposable plastic applicator for each plate. It can plate 227

either single plates or biplates, and the applicator spreads the inoculum over the entire area of a 228

standard 100 mm plate or a biplate. The smallest pipette that the MicroStreak can use is a 10-ul 229

pipette and currently, the specimen top must be removed at the time the specimen is placed on 230

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 11: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

11

the instrument. If the residual specimen is to be saved, the specimen must be recapped after plate 231

inoculation(2). 232

This study was designed to be a preliminary evaluation of a new instrument named the 233

WASP. It was restricted to two container types-Vacutainers and ESwab tubes. Software that was 234

not available when this study was performed will permit sampling of other specimen containers. 235

We compared inoculation of urine Vacutainer specimens by the WASP and the Inoculab LQH 236

instrument. We did not perform manual plating for this study as our in-house validation of the 237

Inoculab instrument in indicated that the Inoculab results were more reproducible than our 238

manual plating method. Lue has also shown that the Inoculab was more accurate than manual 239

plating in her lab using a 1-ul loop (5). Moreover, variability in manual specimen volume 240

transfer for urine specimens has been demonstrated (1).We also evaluated the accuracy of the 241

WASP to subculture Lim broths. Lastly, we compared manual plating of routine specimens 242

collected in ESwabs with plating performed by the WASP. 243

Overall, the results obtained with the Inoculab and WASP for plating of urine specimens were 244

comparable for specimens plated with 1-ul and 10-ul loops. Interestingly, there were 4 specimens 245

plated with the 1-ul loop that grew a significant pathogen at a concentration of 104-10

5 CFU/ml 246

on both WASP cultures (2 with Enterococcus sp. and one each with K. pneumoniae and B-247

hemolytic Streptococcus) while the Inoculab culture was mixed or had multiple flora. Similarly, 248

there was also one specimen plated with the 10-ul loop that grew Enterococcus sp. at a 249

concentration of 103-10

4 CFU/ml on both WASP cultures and mixed flora on the Inoculab 250

culture. 251

There are several possible explanations for the different results between the WASP and 252

Inoculab for these specimens: (1) random differences in CFUs near the breakpoints for reporting 253

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 12: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

12

organisms; (2) failure of a loop to pick up specimen (loss of meniscus); and (3) more 254

homogeneous specimen preparation with one instrument than the other. Differences between the 255

operation of the Inoculab and WASP operation support the latter two possibilities. The WASP 256

has a camera that detects the presence of a droplet inside the inoculating loop after the specimen 257

is sampled. If no specimen is detected, the loop returns to the tube a second time and, if 258

necessary, a third time. If the camera fails to again detect a droplet, the tube is moved to a 259

discard bin. The Inoculab has no sensor to detect the presence of specimen in the loop. A second 260

difference between the instruments is the preparation of the specimen prior to sampling. The 261

instrument manual for the Inoculab states that the specimen should be agitated prior to placing it 262

on the instrument but does not specify how that agitation should be accomplished. In contrast, 263

the WASP has a vortex that vigorously mixes the urine tube prior to sampling. The exact reasons 264

for the observed differences are clearly speculative on our part; nonetheless, the ability to verify 265

the presence of actual specimen in the inoculating loop as well as vigorous vortexing prior to 266

plating may offer more than a theoretical advantage for the WASP compared to the Inoculab 267

instrument. 268

Clearly, there is a need for studies that will assess the speed, throughput, and labor savings of 269

the WASP. At the time that this preliminary study was performed, the WASP required 27 min. to 270

plate 24 Vacutainer tubes to two plates each while the Inoculab required 24 minutes to plate 24 271

Vacutainer tubes to two plates. We anticipate collecting more data once the Sunquest interface is 272

fully validated and software upgrades have been installed. 273

There are also significant capacity differences between the Inoculab and the WASP. The 274

Dynacon model LQH, which was used in our laboratory, holds a total of 38 uninoculated Remel 275

plates compared to 378 Remel plates for the WASP. Since the Dyacon LQH model has only one 276

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 13: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

13

silo for uninoculated plates, if you use more than one type of plate for a culture, the plates must 277

be intercalated in the stack. This would not be necessary if a single plate was used. The Dynacon 278

Inoculab model LQ has 2 silos for uninoculated media that permits the use of one silo for each of 279

two types of media for each specimen. The WASP has nine silos each holding 42 Remel plates, 280

and one to nine types of media can be loaded at one time. More than one silo can be loaded with 281

the same type of media. For example, if a lab uses a BAP for most specimen types, they might 282

chose to load 3 silos with BAP and fewer silos with other types of media. 283

In our evaluation of the WASP for subculturing broths to plated media, we encountered no 284

problems with the WASP. Using a 10-ul loop, the WASP consistently produced isolated 285

colonies. For this study the Lim broths were transferred to empty ESwab tubes. We do not 286

anticipate that labs would want to do this as a standard practice. The resolution to this issue could 287

be the production of broths in smaller tubes either by Copan and or another manufacturer or 288

software changes to the WASP to accommodate taller tubes. 289

Although our evaluation of the WASP for plating of routine cultures was limited to 113 290

specimens, results obtained were comparable to the manually plated cultures. We developed a 3-291

quadrant streaking pattern for use with the 30-ul loop that was used for all of these specimens. 292

The WASP has the potential to process different size containers. The jaws on the robot have 293

three settings to accommodate 3 sizes of containers. There are three docking stations (the 294

docking station holds the bottom of the container during the uncapping process) that also 295

accommodate 3 sizes of containers. The docking station can be easily swapped out. There are 296

two decapping devices on the WASP. One is fixed and is designed for Vacutainer urine tubes. A 297

variety of other decappers can be placed in the second decapping station including an ESwab 298

decapper. For this study we used the decapper and docking station designed to hold ESwabs. 299

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 14: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

14

During this evaluation, we encountered no problems opening and closing either Vacutainers or 300

ESwab tubes. Subsequent to the performance of this study, an interface was established between 301

the WASP and the Sunquest Laboratory Information System. While now operational, it has yet 302

to be fully validated. Unlike the MicroStreak, and similar to the Inoculab, the WASP requires no 303

disposable products for plating. The option of plating urine specimens with 1-ul or 10-ul 304

provides laboratories with flexibility to address clinical needs (4, 6). 305

The only mechanical (hardware) problem that we encountered during this study was a burnt 306

out incinerator that we replaced. There were several software problems that were systematically 307

addressed by Copan as they appeared. None of those problems persisted. 308

This study was designed to be a preliminary evaluation of a new instrument, the WASP. As 309

we expand the use of the WASP to other specimen types, we are working to optimize the 310

workflow pattern for each specimen type. In our routine use of ESwabs, specimens are manually 311

plated using a disposable pipette that is manufactured to deliver approximately 28 ul/drop. Each 312

piece of media receives one drop of the inoculum, and, for specimens requiring a Gram stain, 313

one drop is placed on a slide. We envision a similar workflow with the WASP, with the 314

preparation of any necessary slides occurring prior to loading containers onto the WASP for 315

plating. It is important to understand that the only type of swab specimens that can be plated by 316

the WASP is a swab that transfers the specimen to a liquid phase. ESwab is currently the only 317

swab with that capability. ESwabs are more expensive than traditional wound swabs, and, for 318

labs that are not using ESwabs routinely, this increase in cost should be included in the 319

operational budget for the WASP. 320

We anticipate that the relative amount of specimens that a laboratory can plate on a WASP will 321

be strongly influenced by specimen volume for a particular test and the media required for that 322

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 15: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

15

particular specimen/test. For example, if a laboratory plants an average of 5 specimens each day 323

of a particular specimen/test and that test requires media not used for other tests, adding and 324

removing the required media from the silos may not be an efficient use of time. Importantly, the 325

current operation of the WASP is designed for batch testing, not random-access testing. 326

In summary, we report a first evaluation of a new microbiology specimen-processing 327

instrument, the WASP. We were satisfied by the performance of the WASP and believe that it 328

offers opportunities to automate the processing of microbiology specimens to an extent that has 329

not been possible to date. Additional studies by this and other laboratories are merited to test the 330

full potential of the instrument. 331

332

333

334

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 16: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

16

335

336

337

338

339

340

341

342

343

344

345

346

347

348

REFERENCES 349

350

1. Albers, A. C., and R. D. Fletcher. 1983. Accuracy of calibrated-loop transfer. J. Clin. 351

Microbiol. 18:40-42. 352

2. Glasson, J. H., L. H. Guthrie, D. J. Nielsen, and F. A. Bethell. 2008. Evaluation of an 353

Automated Instrument for Inoculating and Spreading Samples onto Agar Plates. J. Clin. 354

Microbiol. 46:1281-1284. 355

3. Jones, G. A., and R. B. Matthews. 2008. Abstr. 108th General Meeting American 356

Society for Microbiology., Boston, MA. 357

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 17: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

17

4. Kass, E. 1956. Asymptomatic infections of the urinary tract. Trans. Assoc. Am. 358

Physicians 69:56-63. 359

5. Lue, Y. A., C. J. Krause, and A. B. John. 2002. Abstr. 102nd General meeting of the 360

American Society for Microbiology, Salt Lake City, Utah. 361

6. Stamm, W., G. Counts, K. Running, S. Fihn, M. Turck, and K. Holmes. 1982. 362

Diagnosis of coliform infection in acutely dysuric women. N Engl J Med 307:463-468. 363

7. Tilton, R. C., and R. W. Ryan. 1978. Evaluation of an automated agar plate streaker. J. 364

Clin. Microbiol. 7:298-304. 365

366

367

368

369

Figure Legends 370 371

Fig. 1. Front view of the WASP. The dimensions of the instrument are 75 inches wide, 75 inches 372

high and 43 inches deep. 373

Fig. 2. Top view of the WASP. The two SCARA robots are referred to as Tarzan and Jane. 374

Fig. 3. Triquetra Loop. From left to right, 10-ul, 30-ul, and 1-ul loops. 375

376

377

378

379

380

381

382

383

384

385

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 18: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

19

Figure 1 407

408 409

410

411

412

413

414

415

416

417

418

419

420

421

422

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 19: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

20

Figure 2 423 424

425

426

427

428

429 430

431

432

433

434

435

436

437

438

439

440

441

442

443

444

445

446

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 20: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

21

Figure 3 447 448

449 450

451

452

453

454

455

456

457

458

459

460

461

462

463

464

465

466

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 21: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

Table 1. Summary of replicate plating results for significant isolates.

1-ul Inoculum 10-ul Inoculum Total

a Same result

b

(same CFU range)c

Total Same result

(same CFU range)

Inoculab 80 80 (75) 71 71 (70)

WASP 81 80 (79) 70 70 (70)

a Total number of significant isolates for that instrument.

b Same result indicates same organism in replicate cultures with ≥10

4 CFU/ml for 1-ul

inoculum or ≥103 CFU/ml for 10-ul inoculum.

c Same CFU range indicates results of both replicates had similar quantitation (10

3 –

104 CFU/ml, or 10

4 – 10

5 CFU/ml or > 10

5 CFU/ml)

ACCEPTED

on June 21, 2018 by guesthttp://jcm

.asm.org/

Dow

nloaded from

Page 22: 3 Paul P. Bourbeau and Brandi Lynn Swartz 4 Geisinger ...jcm.asm.org/content/early/2009/01/21/JCM.01963-08.full.pdf · 71 tube, and a printer prints specimen information and a barcode

Table 2: Results from WASP and Inoculab for specimens with significant results not in agreementa.

Inoculab WASP Loop Size Replicate 1 Replicate 2 Replicate 1 Replicate 2

Multiple Flora <10 4Mixed Flora 10

4-10

5Enterococcus

<10 4 2 other orgs.

104-10

5 Enterococcus

<104 2 other orgs

<104 Mixed Flora <10

4 Mixed Flora 10

4-10

5 Beta-strep 10

4-10

5 Beta-strep

<104 1 type <10

4 1 type 10-

410

5 Enterococcus 10

4-10

5 Enterococcus

<104 1 type No growth 10

4-10

5 Kl.

pneumoniae

<104 1 type

104-10

5 Kl

pneumoniae

<104 1 type

>105 E. coli 10

410

5 E. coli > 10

5 E. coli > 10

5 E. coli

>105 E. coli 10

410

5 E. coli > 10

5 E. coli > 10

5 E. coli

>105 Beta-strep 10

4-10

5 Beta-strep >10

5 Beta-strep > 10

5 Beta-strep

104-10

5 S. aureus 10

4-10

5 S. aureus >10

5 S. aureus 10

4-10

5 S. aureus

104-10

5 P. mirabilis 10

4-10

5 P. mirabilis >10

5 P. mirabilis >10

5 P. mirabilis

104-10

5 NLF > 10

5 NLF > 10

5 NLF > 10

5 NLF

1-ul

104-10

5 CNS

<104 1 O.T.

> 10 5CNS

<104 1O.T.

> 105 CNS

<104 1O.T.

> 105 CNS

<104 1O.T.

Multiple Flora Multiple Flora 10

3 – 10

4 CNS

<103 1 O.T.

Multiple Flora

<103 2 orgs. <10

3 2 orgs.

103 – 10

4 Enterococcus

<103 1 O.T.

103 – 10

4 Enterococcus

<103 1 O.T.

<102 Mixed Flora

103 – 10

4 CNS

<102 1 0.T.

10

3 – 10

4 CNS

<103 1 0.T.

103 – 10

4 CNS

<103 1 0.T.

103 – 10

4 E. Coli 10

3 – 10

4 E. coli > 10

4 E. coli > 10

4 E. coli

103 – 10

4 GBS 10

3 – 10

4 GBS > 10

4 GBS > 10

4 GBS

103 – 10

4 Enterococcus 10

3 – 10

4 Enterococcus > 10

4 Enterococcus > 10

4 Enterococcus

10-ul

103 – 10

4 Vir. strep.

<102 1 O.T.

103 – 10

4 Vir. strep.

<102 1 O.T.

> 104

Vir. strep.

<102 2 O.T.

> 104

Vir. strep.

<102 1 O.T.

a O.T.= other type; GBS= Group B Streptococci; CNS= coagulase negative staphylococci; NLF= unidentified non-lactose

fermenting Gram-negative bacilli.

ACCEPTED on June 21, 2018 by guest

http://jcm.asm

.org/D

ownloaded from