2.a.omniscan setup - calibration for sectorial scan

Upload: carlos

Post on 03-Jun-2018

261 views

Category:

Documents


8 download

TRANSCRIPT

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    1/34

    By Nick Bublitz

    Performing a Sectorial ScanSensitivity/ Velocity/Wedge Delay

    Calibration-Shear Wave PA Using

    SDHs at Known Depths 2.0 SW

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    2/34

    2

    Calibrations-Required items

    Calibration block with 2 same size reflectorsat known depths- (Navships block with

    SDHs or other calibration block with 2 same

    size reflectors at known depths)Couplant-water works best for practice-

    otherwise whatever you will test with

    Guide to keep from skewing- use a ruler, setthe block on its side, or on IHC wedges can

    screw down one sides carbide pins.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    3/34

    3

    Velocity Calibration

    The velocity calibration will fine tune the velocity to that ofthe calibration block. It will require two reflectors of thesame size at two different known depths (Depth). A radiusmay also be used in sound path overlay (Radius) .

    In many cases the default material velocity and wave type

    set by the operator in the group wizard is sufficient and thiscalibration can be skipped because wedge delay will adjustaccordingly to allow accurate measurements in the depthaxis.

    If performed the velocity calibration should be done beforethe wedge delay calibration. Performing a velocitycalibration after wedge delay will remove the wedge delaycalibration.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    4/34

    4

    Velocity Calibration

    Just like Con. UT the velocity calibration will calculate the time

    difference to two reflectors of known depth or sound path and

    calculate a material velocity. Many conventional users use an IIW

    block (radius), in PA we often use true depth rulers so a Navshipsworks well (depth)

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    5/34

    5

    Select Velocity Calibration

    Go to the Wizard-Calibration menu

    Select Ultrasound and

    Velocity

    Start the wizard

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    6/34

    6

    Select A-Scan

    Put the probe on yourblock over the first

    reflector.

    Choose an angle near the

    middle of your anglesweep using the angle

    function. The calibration

    will utilize a single beam

    as the velocity will be

    consistent across all the

    angles.

    Adjust range as needed to

    ensure you will see the 2

    reflectors you have chosen.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    7/34

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    8/34

    8

    Set Gate A on Depth 1

    Move probe back and forthover the first reflector

    building an envelope. Use

    a guide so you do not

    skew right or left. Ensure you peak the

    reflector.

    Adjust the gate to

    encompass the reflector. Get position.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    9/34

    9

    Set Gate A on Depth 2

    Repeat the peakingprocess on your

    second reflector.

    Adjust the gate to

    encompass this

    reflector.

    Get position.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    10/34

    10

    Accept or Restart

    View the velocitydetermination

    If it is sensible accept the

    calibration, if not restart.

    Depth readings may not be100% accurate after a

    velocity calibration at all

    angles due to the different

    wedge path distance ofindividual beams. Wedge

    delay calibration will

    compensate for this.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    11/34

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    12/34

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    13/34

    13

    Set Section

    Decide if you want tocalibrate the whole

    angular sweep or

    break it up into

    separate calibrations-for most setups with a

    good calibration block

    you can calibrate the

    whole sweep. Hit Next

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    14/34

    14

    Set Gate A on Depth A

    Put the probe on the block soyou can see the chosen

    depth reflector.

    Adjust the gate

    start/width/threshold toencompass the reflector.

    Move the probe forward

    and backwards to ensure

    gate is encompassing

    reflector through all angles

    and there are no other

    reflectors that will affect

    calibration.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    15/34

    15

    Calibrate and Accept

    Move the probe forwardand backward applyingeven pressure andavoiding skewing right orleft.

    Move in both directionsuntil the green line is builtacross the whole angularrange. Adjust gain asneeded.

    Hit calibrate.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    16/34

    16

    Accept or Restart

    After hitting calibrate-rerun the probe

    through the angular

    range verifying you are

    within the settolerance.

    Accept or Restart.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    17/34

    17

    Verifying Wedge Delay

    Set the DA reading fromthe Measurement-Readings menu

    Turn on the Highest %feature in the Display-

    Properties-Source menu Run the probe through the

    angular sweep whileobserving the DA reading.

    The Omniscan will trackthe highest reflector in thered gate. Return thesource to normal beforenext calibration!

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    18/34

    18

    Sensitivity Calibrations

    Sensitivity calibrations equalize the sensitivity(amplitude) to a given reflector through all the

    angles.

    This will insure no matter what angle the reflector

    is seen at the % FSH is the same for rejection or

    detection purposes as well as for the amplitude

    based color coded imaging selections. Calibration

    reflector and natural flaw response relies heavilyon angle so it is critical each beam/angle can be

    calibrated to the same sensitivity level.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    19/34

    19

    Sensitivity Calibration

    Enter Calibration menuand select Sensitivity

    Choose a reference

    screen height to equalizesensitivity to (80% FSH)and a tolerance. Thedefault tolerance of 5% willshow two linesone 2.5%

    above your set reference,one 2.5% below. This doesnot affect the quality of thecalibration it is merelyvisual reference.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    20/34

    20

    Getting Started

    Choose a SDH andcouple your probe

    above it

    Use a guide system to

    keep the probe from

    skewing side to side.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    21/34

    21

    Set Section

    Choose whether tocalibrate the whole

    angular sweep in one

    step or break it up into

    separate calibrations-if your calibration block

    has adequate spacing

    between SDHs you

    can usually do the

    whole range.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    22/34

    22

    Set Gate A on Echo A

    Set the gate to encompass theSDH reflection. Adjust the

    start/width/threshold of the gate.

    Move the probe forward and

    backward to ensure the gate

    covers the reflector through all

    the angles and there is not on

    any other reflectors that may

    inhibit the calibration

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    23/34

    23

    Set Compensation Gain

    Allows you to adjustthe gain as needed toensure the reflector isseen properly

    throughout all theangles

    In most cases noadjustment or an

    adjustment of theoverall gain (if overreference level FSH%)is needed.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    24/34

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    25/34

    25

    Calibrate and Accept

    Press Calibrate Verify the calibration

    by repeating theforward and backward

    movement of the probe The green line should

    fall within the definedtolerance lines across

    all the angles.Accept or Recalibrate

    as needed.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    26/34

    26

    To Verify Sensitivity Calibration

    To verify sensitivity go toDisplay menu

    Select properties

    Change source to Highest

    Run the probe back and forthand observe FSH in the

    A% reading field as the

    data cursor tracks the

    highest amplitude signal inthe red gate. Return the

    source to normal before

    further calibrations!

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    27/34

    27

    General Tips for Calibrations-Omniscan-General

    Even pressure and couplant are key- practice

    makes perfect. Practice with water-it doesntshovel like gel couplant.

    A good calibration block helps a lot- theNavships block is a great block for PAcalibrations. Tighter spacing of reflectors mayrequire breaking up the calibrations into two ormore sections using the set section feature tokeep other reflectors from entering the gatedtracking area.

    A guide will ensure sure the same portion ofthe beam sees the reflectorthis is key

    The Omniscan must see the natural state(building of green line) between about 10%FSH and the high tolerance set.

    Instead of restarting calibrations you canreduce gain, rebuild green line and hit

    calibrate again. If the green line drops off suddenly examineyour a-scan at the point it drops, your gatethreshold and/or width may need adjustment.

    Utilize the scan view for setting gates, lookingfor interfering reflectors- it is a tendency forcon. ut operators to look at only the a-scan-remember this is only one angle in a sweep.

    max tolerance line

    10% FSH

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    28/34

    28

    General Tips for Calibrations-Omniscan settings

    Ensure the tracking function is set tonormal-Display-Properties-Source-Normal, since the wizards use a trackingsystem Highest/Thinnest/All Laws caninterfere

    Make sure you are set to run off theclock of the Omniscan, not an encoderbase-Scan-Inspection-Scan-Time-lookin right hand corner-should be inseconds.

    Make sure db reference is off- Utsettings-Advanced-db Ref-Off look in lefthand corner

    These are all set ok by default but if youcreate a setup and have issues these

    are key places to look. Sluggish-low PRF- ensure gates are all

    on screen and minimized, minimizerange etc

    All of the above usually result in nogreen line at all of a very sluggish jerkytracking response.

    No db ref ( )

    x: sec

    prf

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    29/34

    29

    General Tips for Calibrations-Omniscan

    Rounded irregular hills in green line-

    usually due to uneven pressure orskewing left/right, also excess couplantin that area- hold probe with evenpressure, use a guide and apply evencouplant between wedge and block.

    Green line not able to be set between

    reference and 10% for training- adjustgain, use comp. gain, reduce angularsweep or change wedge. Comp. Gainwill add an amount of gain per us oftravel- longer beam paths (higherangles) will thus have more gain addedto facilitate proper calibration. Lowerwedge angles will be easier bent to lowangles, the opposite for highera 45degree wedge is easier to calibrate at 30degrees, a 60 degree wedge will beeasier calibrated at 72 degrees. 55degree wedge is a good wedge for bothextremes.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    30/34

    30

    General Tips for Calibrations-Omniscan

    Sharp jagged triangular dropoutsin green line trace- usually couplantdependant- check couplantbetween wedge and block first- if itrepeats at the same angle everytime after reapplying couplant, it ismore than likely between the wedgeand probe- reapply couplant here.

    Can also be gating-look at scan toset gates over the entirety of thereflector through all the angles.

    Sharp diagonal dropout- completeloss of couplant or insufficientamplitude- view ascan at point ofdropout- reapply couplant, increasegain, lower gate threshold toencompass ascan amplitude.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    31/34

    31

    General Tips for Calibrations-Omniscan

    Data cursor tracks to other reflectornot being used for calibration-examine and reposition gate, usecalibrate by section feature, usedifferent block.

    Unable to perform wedge delay or

    improper wedge delay- if everythingelse fails, observe a single reflectorin a true depth scan- the reflectorshould not drift in depthdramatically- if it does there isprobably an error in the wedge fileparameter- Probe/Part-Wedge-check if proper wedge is selected

    and note the wedge settings-especially height- consult atechnical support specialist ordatabase of wedge parameters.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    32/34

    32

    Calibrate by Section

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    33/34

    33

    Omniscan Error Messages during Calibration-

    Sensitivity

    -At least one context has an amplitude

    peak greater than 100%. The calibrationcannot be performed (sensitivitycalibration) -at least one angle or vpa isgreater than the reference amplitudeyou set. Reduce gain until all beams fallbelow the high tolerance line but stillabove about 10% FSH.

    -at least one context has an amplitudepeak of 0%.The calibration cannot beperformed or At least one context hasno amplitude peak in gate A. Wedgedelay calibration cannot be performed..

    at least one angle or vpa is not beingpicked up due to low amplitude-increase gain, troubleshoot couplant etc

    as needed Other cause of the above two errors-not

    completing the edges (first/last angles/vpas)- green line may appear to go toedges, make sure you run the whitetrace line completely off screen andbeyond.

  • 8/12/2019 2.a.omniscan Setup - Calibration for Sectorial Scan

    34/34