21 ex qa

19
Physics Quality Assurance Varian 21EX Linear Accelerator (LINAC)

Upload: khlarsen

Post on 16-May-2015

3.570 views

Category:

Health & Medicine


2 download

DESCRIPTION

Quality assurance for Varian 21EX linear accelerator

TRANSCRIPT

Page 1: 21 Ex Qa

Physics Quality Assurance

Varian 21EX Linear Accelerator

(LINAC)

Page 2: 21 Ex Qa

Linear Accelerator Gantry

Electrons are Electrons are accelerated to accelerated to energies of energies of several million several million electron volts.electron volts.

Page 3: 21 Ex Qa

Linear Accelerator Block Diagram

Page 4: 21 Ex Qa

Linear Accelerator Treatment Head

The electrons strike a target to produce x-rays or The electrons strike a target to produce x-rays or they strike a scattering foil to treat directly with they strike a scattering foil to treat directly with electrons.electrons.

Page 5: 21 Ex Qa

Varian 21EX Linear Accelerator

• X-ray energies (used to treat target areas located more than a few cm below the skin surface– 6 and 10 MV energies

• Electron energies (used to treat lesions on the surface of the skin)– 6, 9, 12, 16, and 20 MeV energies

Page 6: 21 Ex Qa

l

Fields can be shaped by using custom made blocks or multileaf collimators attached to the treatment head

Page 7: 21 Ex Qa

Linear Accelerator Quality AssuranceAbsolute dose calibration is performed according to the

AAPM TG51 protocol using water phantom, ionization chamber and electrometer

Page 8: 21 Ex Qa

This shows the ionization chamber in the water tank. It is connected to the electrometer located outside the room at the control console with a long cable. The TG51 protocol

specifies the depth at which the calibration is performed.

Page 9: 21 Ex Qa

Other measurements using the water

phantom system• Tissue maximum ratios (dose dependence

on field size and depth)

• Transmission factors of beam shaping devices such as wedges, block trays and compensators

Page 10: 21 Ex Qa

In-air measurements using a build-up cap on the ion chamber

• Collimator scatter factors

• Output dependence on gantry angle

• Inverse square factor

• Collimator transmission

• Dose rate dependence

• Output linearity

• MU end effect

Page 11: 21 Ex Qa

LINAC Mechanical Tests

• Light field (used to set up patients)– Properly centered– Congruent with radiation field– Does not “walk out” when collimator is rotated– Accurate over full range of field sizes

Page 12: 21 Ex Qa

LINAC Mechanical Tests (cont)

• Crosshairs (used to set up patients)– Properly centered– Do not “walk out” when collimator is rotated

• Optical distance indicator accuracy (used to measure the distance from the target to the patient – if this changes during the treatment course, patient may have gained or lost weight and a new calculation and/or treatment plan would be done)

• Readout devices– Gantry, collimator and couch angles– Field size for both independent jaws

Page 13: 21 Ex Qa

LINAC Mechanical Tests (cont)

• Gantry axis of rotation – center of radiation field stable when gantry is rotated

• Collimator axis of rotation – center of radiation field stable when collimator is rotated

Page 14: 21 Ex Qa

LINAC Safety Tests

• Communication devices– Video– Audio

• Beam termination – Radiation off button– Emergency off buttons– Main power circuit breaker

• Time delay - when LINAC is powered up filaments must be warmed up

• Emergency couch pendant – couch can be moved out and down when all power is lost to LINAC

• Backup mechanical counter – displays dose given if power is lost during a treatment

Page 15: 21 Ex Qa

LINAC Safety Tests (cont)

• Motions controlled by hand pendant– Gantry rotation– Collimator rotation– Upper and lower jaws– Couch rotation, vertical, lateral and longitudinal movements

• Safety checks– Motion enable switch functional– Limit switches functional– Motor reversal can be initiated quickly– Timer – terminates radiation if not stopped by preset MU– Secondary dosimetry – terminates radiation if not stopped by

primary MU setting

Page 16: 21 Ex Qa

Other Tests of LINAC Components

• Multileaf collimator

• Digital portal imager

Page 17: 21 Ex Qa

This device is used each day to check the calibration, energy, symmetry and flatness of each beam energy. If any of these parameters fall outside of limits

set by the physicist, the reason must be determined and corrected before patient treatments can be performed.

Page 18: 21 Ex Qa

Lasers on the walls and ceiling are used to set up the patient on the treatment couch. This device is used to

check the accuracy of the laser positions.

Page 19: 21 Ex Qa

Physics Quality Assurance Documentation and Analysis

• Tolerance limits are set for all parameters

• If any measurement exceeds the tolerance, adjustments are made by the physicist or engineer

• All measurements are carefully documented

• Inspectors from Florida Radiation Control review the documents at least once a year